3
The Analysis of Smokestack Samples for Dioxins by Gas Chromatography/Mass Spectrometry and Gas Chromatography/Tandem Mass Spectrometry Colin Moore, John Moncur, David Jones and Barry Wright* Kratos Analytical, Barton Dock Road, Wrmston, Manchester, UK A CONCEPT 1 SQ hybrid mass spectrometer has been used to show the potential of high resolution mass spectrometry and hybrid tandem mass spectrometry (MSIMS) for dioxin analysis. The usual method chosen for gas chromatographyhnass spectrometry (GUMS) determination of polychlorinated dibenzodioxins (PCDDs) and polychlorinated dibenzofurans (PCDFs) is selected-ion monitoring (SIM) at high resolution, the high resolutionimparting a degree of specificity to the analysis. An alternative technique is to use low resolution (1000) gas chromatography/tandemmass spectrometry on a hybrid mass spectrometer, where the specificity of MSlMS is utilized. Data are presented illustrating the potential of both techniques from a single instrument, the CONCEPT 1 SQ. The determination of chlorinated dibenzo-p-dioxins (PCDDs) and dibenzofurans (PCDFs) in complex matrices often requires extensive sample clean-up prior to instrumental analysis. Samples may contain chemical interferences at concentrations several orders of magni- tude greater than those of the PCDD and PCDF, placing heavy demands on the instrumentation. Difficulties in the analysis of environmental samples make selectivity and sensitivity important consider- ations for the method of choice. Capillary column high resolution gas chromatography/high resolution mass spectrometry has been the ‘standard method’ to date for the analysis of samples such as sediments, effluents and biological tissues. Tandem mass spectrometry (MS/ MS) has been able to meet the requirements of selecti- vity but has not yet reached the ultimate detection levels of selected-ion monitoring (SIM). However, our MSlMS data do show some very useful features when it comes to the analysis of real samples. EXPERIMENTAL Instrumentation Figure 1 shows a diagram of the Kratos CONCEPT 1 SQ mass spectrometer. It is of EBQQ geometry, that is a double-focusing, forward geometry sector instrument (MSI), followed by a quadrupole collision cell and a second quadrupole for fragment mass analysis (MS2). A resolution of 10 000 (10% valley definition) is needed for dioxin analysis using SIM. Figure 2 shows that the product of resolution and sensitivity vs resolution remains relatively constant below 10 000 resolution. This indicates that the transmission of MS1 at 10000 resolution is about 10% of that at 1000 resolution. To circumvent the problem of sensitivity loss, encountered by monitoring at substantially reduced accelerating voltage, the ions to be monitored are split into groups of similar mass (tetra-, penta-, hexa-, hepta- and octa- chloro) and a different magnetic field strength is used for each group. By this means, mass discrimination Author to whom correspondence should be addressed Magnet Rf only Quadrupoles Analyser Quadrupoles Final Detector Figurel. Schematic diagram of a CONCEPT 1 SQ hybrid GClMSlMS instrument. amongst the groups is minimized. Figure 3 shows the analysis of equal concentrations of penta-, hexa-, hepta- and octa-chloro compounds by this method. MS/MS For low resolution MS/MS to attain equal sensitivity to high resolution SIM, no less than 10% of the precursor ions originating in the source must reach the final collector (MS2) as fragment ions. Such sensitivities are not generally achieved in MS/MS, however, because of further losses in transmission through MS2 and because of ion losses in the collision process itself. 0951-4198/90/0418-0420 $05 .OO 418 RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 4, NO. 10, 1990 @ John Wiley & Sons Limited, 1990

The analysis of smokestack samples for dioxins by gas chromatography/mass spectrometry and gas chromatography/tandem mass spectrometry

Embed Size (px)

Citation preview

Page 1: The analysis of smokestack samples for dioxins by gas chromatography/mass spectrometry and gas chromatography/tandem mass spectrometry

The Analysis of Smokestack Samples for Dioxins by Gas Chromatography /Mass Spectrometry and Gas Chromatography /Tandem Mass Spectrometry Colin Moore, John Moncur, David Jones and Barry Wright* Kratos Analytical, Barton Dock Road, Wrmston, Manchester, UK

A CONCEPT 1 SQ hybrid mass spectrometer has been used to show the potential of high resolution mass spectrometry and hybrid tandem mass spectrometry (MSIMS) for dioxin analysis. The usual method chosen for gas chromatographyhnass spectrometry (GUMS) determination of polychlorinated dibenzodioxins (PCDDs) and polychlorinated dibenzofurans (PCDFs) is selected-ion monitoring (SIM) at high resolution, the high resolution imparting a degree of specificity to the analysis. An alternative technique is to use low resolution (1000) gas chromatography/tandem mass spectrometry on a hybrid mass spectrometer, where the specificity of MSlMS is utilized. Data are presented illustrating the potential of both techniques from a single instrument, the CONCEPT 1 SQ.

The determination of chlorinated dibenzo-p-dioxins (PCDDs) and dibenzofurans (PCDFs) in complex matrices often requires extensive sample clean-up prior to instrumental analysis. Samples may contain chemical interferences at concentrations several orders of magni- tude greater than those of the PCDD and PCDF, placing heavy demands on the instrumentation.

Difficulties in the analysis of environmental samples make selectivity and sensitivity important consider- ations for the method of choice. Capillary column high resolution gas chromatography/high resolution mass spectrometry has been the ‘standard method’ to date for the analysis of samples such as sediments, effluents and biological tissues. Tandem mass spectrometry (MS/ MS) has been able to meet the requirements of selecti- vity but has not yet reached the ultimate detection levels of selected-ion monitoring (SIM). However, our MSlMS data do show some very useful features when it comes to the analysis of real samples.

EXPERIMENTAL Instrumentation Figure 1 shows a diagram of the Kratos CONCEPT 1 SQ mass spectrometer. It is of EBQQ geometry, that is a double-focusing, forward geometry sector instrument (MSI), followed by a quadrupole collision cell and a second quadrupole for fragment mass analysis (MS2). A resolution of 10 000 (10% valley definition) is needed for dioxin analysis using SIM. Figure 2 shows that the product of resolution and sensitivity vs resolution remains relatively constant below 10 000 resolution. This indicates that the transmission of MS1 at 10000 resolution is about 10% of that at 1000 resolution. To circumvent the problem of sensitivity loss, encountered by monitoring at substantially reduced accelerating voltage, the ions to be monitored are split into groups of similar mass (tetra-, penta-, hexa-, hepta- and octa- chloro) and a different magnetic field strength is used for each group. By this means, mass discrimination

Author to whom correspondence should be addressed

Magnet

Rf only Quadrupoles

Analyser Quadrupoles

‘ Final Detector

Figurel. Schematic diagram of a CONCEPT 1 SQ hybrid GClMSlMS instrument.

amongst the groups is minimized. Figure 3 shows the analysis of equal concentrations of penta-, hexa-, hepta- and octa-chloro compounds by this method.

MS/MS For low resolution MS/MS to attain equal sensitivity to high resolution SIM, no less than 10% of the precursor ions originating in the source must reach the final collector (MS2) as fragment ions. Such sensitivities are not generally achieved in MS/MS, however, because of further losses in transmission through MS2 and because of ion losses in the collision process itself.

0951-4198/90/0418-0420 $05 .OO

418 RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 4, NO. 10, 1990 @ John Wiley & Sons Limited, 1990

Page 2: The analysis of smokestack samples for dioxins by gas chromatography/mass spectrometry and gas chromatography/tandem mass spectrometry

GUMS AND GClMSlMS ANALYSIS OF SMOKESTACK SAMPLES FOR DIOXINS

11. -I I I I U I U -

Figure 2. Curve of the product of resolution and sensitivity vs resolu- tion for a CONCEPT.

When these loss factors are taken into account, the detected ion currents at MS2 are typically between 1% and 0.5% of the source ions, leading to detection limits between 10 and 20 times poorer than the high resolu- tion SIM mode of operation. Figures 4 and 5 show these limits at 2 ions s-l for high resolution SIM and 4 ions s-l for MS/MS.

Despite this reduced sensitivity, MS/MS may, how- ever, provide superior results where heavily-polluted samples are to be analysed. In particular, PCBs (especially the hexa- and hepta-chlorinated congeners) are known to interfere with the measurement of TCDDs and its ion fragments formed in the ion source. Additional interferences are often experienced from methyl and ethyl tetrachlorinated dibenzofurans, even when using SIM at 10000 resolution. MUMS consti- tutes the method of choice in these cases. In a case where the analyst has no knowledge of possible inter- fering peaks, it may be impossible to decide which approach is more appropriate so that both methods may be necessary.

60

40

80

20

10

8c.n 400 Ed0 R.T. l O d 8 14:44

I000 2233

L I 1200

2930

Figure 3. Analysis of penta-, hexa-, hepta- and octa-chloro PCDDs by group selection in SIM.

Figure 4. Analysis of 10 fg of TCDD by HRSIM (10 000 RP).

Figure 5. Analysis of 350 fg of TCDF by hybrid GCIMSIMS using COCl loss.

RESULTS In the analysis of smokestack samples undertaken during the course of this work, the MS/MS conditions used are shown in Table 1. Several standard solutions and blanks were run during the course of this investi- gation. The data obtained showed good linearity of response over the concentration range 500 pg to 350 fg and suggested an ultimate detection limit of 200 fg with 2 : 1 signal-to-noise ratio. All unknown samples were prepared in iso-octane, and 1 yL amounts of these solutions were injected into the GC.

For heavily-contaminated samples such as the smokestack sample of Fig. 6 , MS/MS shows clear, strong ions from components simultaneously eluting from the GC column. Many isomers are present, as is typical of real samples. The series shows all 5 groups

Table 1. MSlMS conditions MS1 resolution 1000 MS2 resolution unit mass Collision gas argon Pressure set for 50% attenuation Collision energy 160 eV

Number of ions monitored native or I3C spike

Precursor ion rate

5 groups of 8 ions all COCl losses from

8 ions s.-'

@ John Wiley & Sons Limited, 1990 RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 4. NO. 10, 1990 419

Page 3: The analysis of smokestack samples for dioxins by gas chromatography/mass spectrometry and gas chromatography/tandem mass spectrometry

GUMS AND GC/MS/MS ANALYSIS OF SMOKESTACK SAMPLES FOR DIOXINS

Figure 6. Hybrid GClMSIMS analysis of the heavily-contaminated smokestack sample for PCDDs and PCDFs.

TCDF 304 * 24 1 II

Figure 7. GCIMSIMS trace for TCDD and TCDF components in the same analysis shown in Fig. 6.

from tetra- to octa-chloro. Figure 7 shows the isomers of TCDD and TCDF included in Fig. 6. 8 ions s-' were monitored, 4 from the 13C spike parent ions and 4 from the native parent ions; only the 4 native ions are shown.

Figure 8 shows an MS/MS ion chromatogram of a composite smokestack sample, this time showing a predominance of tlie hepta-chloro isomers. Figure 9 shows an MS/MS mass chromatogram of a 1000 m3 air sample with TCDD and TCDF isomers at.pg levels. Subsequent isomer quantitation of these samples has been performed using our standard software package, with both internal or external calibration. Results of this fall outside the scope of this paper and are to be reported elsewhere.

420 RAPID COMMUNICATIONS IN MASS SPECTROMETRY, VOL. 4, NO. 10, 1990

Figure 8. GC/MS/MS trace for the analysis of a composite smoke- stack sample.

320 -) 257

Y 322 -t 259

I* _h

v .

n .

9.

* - Ah-. ~

Sn rn m lY !u IL &* C l L1 h) 16 911 ?a w 'IP s 1 $5 11 l i I; IS n n * a 11 i l :: i , i? i l i l 2 II i? t i '3 i? il 'lo at1 > , I

Figure 9. GCIMSIMS trace for the analysis of a 1000 m3 air sample.

CONCLUSION Overall sensitivity in PCDD and PCDF analysis is primarily a function of the selectivity of the analytical process. The wide variety of potentially interfering matrices encountered in everyday samples precludes the use of any one analytical method. This study illus- trates an alternative approach, providing reasonable sensitivity in situations where traditional SIM is inade- quate.

Received 30 July 1990; accepted 10 August 1990.

0 John Wiley & Sons Limited, 1990