    Customizing StandardCharacteristics of OMICRON TestUniverseSome of the OMICRON Test Universe standard characteristics can becustomized according to the users requirements. To change a characteristic,certain settings in the Windows Registry need to be modified.The following sections describe what characteristics can be changed and how.

    Section 1 What characteristics can be modified? describes the feature, thekeys and values to be set and/or created in the registry are described:

    1.1 Default CMC Device Type for Offline Mode on page........................... 4

    1.2 Ground Wire Break Warning on page.................................................. 41.3 Scanning of Parallel Ports on page ..................................................... 5

    1.4 Automatic Registration of the Kernel Mode Drivers on page............... 6

    1.5 Height of Analog Diagrams in Reports on page .................................. 6

    1.6 Software Watchdog Timeout on page ................................................. 7

    1.7 PC to Test Set Watchdog Timeout on page ........................................ 7

    1.8 Overload Disable Time on page .......................................................... 8

    1.9 Disable Hardware-Check Error Messages on page ............................ 9

    1.10 Rogowski Simulation Amplitude Dynamics on page........................ 10

    1.11 Separate Language Selection for Test Reports on page................. 11

    Section 2 How to Edit the Windows Registry provides information how these

    modifications have to be performed.2.1 Editing an entry on page.................................................................... 13

    2.2 Adding a New Entry on page............................................................. 14

  • 8/10/2019 Test Universe Registry Settings

    4/14

    OMICRON Test Universe

    4

    1 What characteristics can bemodified?

    1.1 Default CMC Device Type for Offline ModeDetermines the default device type for offline operation. When the Hardware

    Configuration dialog is opened in offline mode, Test Universe providessimulation of all device types listed below. By setting this registry entry, thedefault device type can be specified, i.e., the device type set here will beselected by default.

    Registry Settings

    1.2 Ground Wire Break WarningSpecifies whether or not the ground wire break warning is displayed.

    Note: The ground wire break warning is a crucial message! Disabling it willsuppress an important message that indicates a potential safety risk.

    Therefore we suggest to disable this warning only if the ground wirebreak warning occurs constantly due to your network topology, and allCMC users are fully aware that this is the case!

    Registry Settings

    Key name: HKEY_CURRENT_USER\Software\OMICRON\HardwareConfiguration

    Value name: DefaultDeviceTypeValue type: String (REG_SZ)

    Value data: CMC56, CMC156-EP, CMC256-3, CMC256-6, CMC151

    Key name: HKEY_CURRENT_USER\Software\OMICRON\HardwareConfiguration

    Value name: GroundWireBreakWarning

    Value type: String (REG_SZ)

    Value data: 0 or 1

    (0 = warning disabled, 1 = warning enabled. Default = 1)

    Note: When "Value name" does not exist, the warning is

    enabled.

  • 8/10/2019 Test Universe Registry Settings

    5/14

    5

    Customizing Standard Characteristics

    1.3 Scanning of Parallel PortsDetermines whether or not CM Engine scans the corresponding parallel port forconnected CMCs. The default setting is "scan port".

    Disabling a particular port from being scanned can be advantageous if ahardware device is connected to it that may conflict with the way the OMICRONkernel mode driver operates. Such a device can be a printer that, for example,every time the kernel mode driver switches, initiates an internal reset and ejects

    a sheet of paper.Registry Settings

    Key name: HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\OMICRON\ParCMC

    Value name: ScanLPTx

    (where "x" is to be replaced with the actual port number)

    Value type: DWORD (REG_DWORD)Value data: 0 or 1 (default = 1)

    0 = port is not checked for connected CMCs

    1 = the port is included when scanning for devices

    Note: When "Value name" does not exist, the port isincluded when scanning for devices.

  • 8/10/2019 Test Universe Registry Settings

    6/14

    OMICRON Test Universe

    6

    1.4 Automatic Registration of the Kernel ModeDriversDetermines whether or not CM Engine automatically tries to register the kernelmode drivers for parallel port communication on each start-up. The defaultsetting is "register kernel mode drivers".

    Note: This setting should only be used by very advanced users thatunderstand the impact of disabling kernel mode drivers.

    Registry Settings

    It may be advantageous to select not to automatically register the the kernelmode drivers if you experience a conflict with another driver. However, up todate, the only definitely known case is an incompatibility with the PPShell driverof the Windows CE development environment.

    Setting the value data to 1 prevents CM Engine from automatically registeringthe kernel mode drivers, however, results in the fact that OMICRON TestUniverse can be operated in offline mode only.

    1.5 Height of Analog Diagrams in Reports

    Specifies the height in millimeters of analog diagrams in the test reports.Registry Settings

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: RegisterAndStartKernelDrivers

    Value type: String (REG_SZ)

    Value data: 0 or 1 (default = 1)

    0 = kernel mode drivers not automatically registered1 = kernel mode drivers registration attempted automatically

    Note: when "Value name" does not exist, registration isattempted automatically.

    Key name: HKEY_CURRENT_USER\Software\OMICRON\\Options

    Value name: DiagramHeight

    Value type: String (REG_SZ)

    Value data: height in millimeters, value must be between 1 and 200

  • 8/10/2019 Test Universe Registry Settings

    7/14

    7

    Customizing Standard Characteristics

    1.6 Software Watchdog TimeoutSpecifies the time-out for the software watchdog in the device.

    Note: This setting should only be changed when advised by the OMICRONhotline.

    Registry Settings

    1.7 PC to Test Set Watchdog TimeoutSpecifies the time-out for the watchdog between the computer and theOMICRON test set (only relevant for CMC256 with NET-1 option).

    Note: This setting should only be changed when advised by the OMICRONhotline.

    Registry Settings

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: WatchdogTimeout

    Value type: String (REG_SZ)

    Value data: Watchdog timeout in milliseconds, value must be between 2000and 30000

    Note: When "Value name" does not exist, the watchdogtimeout will be set to an internal default (currentlyabout 2000ms).

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: WatchdogTimeoutAL2DevEng

    Value type: String (REG_SZ)

    Value data: Watchdog timeout in milliseconds, value must be between 1000and 30000

    Note: When "Value name" does not exist, the watchdogtimeout will be set to an internal default (currently3000ms).

  • 8/10/2019 Test Universe Registry Settings

    8/14

    OMICRON Test Universe

    8

    1.8 Overload Disable TimeSpecifies a time span in which the overload detection is disabled after a signalchange.

    Registry Settings

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: OverloadDisableTime

    Value type: String (REG_SZ)

    Value data: Overload disable time in milliseconds, value must be between15 and 25000, otherwise it is ignored.

    Note: When "Value name" does not exist, the overloaddisable time will be set to an internal default (currently50ms).

  • 8/10/2019 Test Universe Registry Settings

    9/14

    9

    Customizing Standard Characteristics

    1.9 Disable Hardware-Check Error MessagesThis setting disables one or more error messages of the hardware self-test.

    Note: Edit with extreme caution! Working with a defective test set whoseerror messages are intentionally suppressed can be very dangerous!

    Registry Settings

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: HWCheckDisableMask

    Value type: String (REG_SZ)

    Value data: The value is interpreted as 32 bit mask. Each bit in the maskdisables one error message:

    Bit Disables error message0 V1A1 V2A2 V3A3 V1B4 V2B5 V3B6 Overtemperature voltage7 Voltage overload

    8 I1A9 I2A10 I3A11 I1B12 I2B13 I3B14 Overtemperature current15 Reference current

    16 - 23 unused

    24 AD converter 25 Binary inputs26 MEA

    27 - 31 unused

  • 8/10/2019 Test Universe Registry Settings

    10/14

    OMICRON Test Universe

    10

    1.10 Rogowski Simulation Amplitude Dynamics A new setting possibility for the dynamic amplitude range using the trueRogowski coil simulation of the CMC256 test set. Two dynamic amplitude rangemodes can be selected: "normal" (default) and "high".

    "Normal" mode provides the security that the amplitudes do not exceed alimit which could result in distorted signals, even for higher frequencies, e.g.,400Hz. The amplitude, however, is limited to 5.9 times the nominal current

    (assuming that 150mV voltage is the corresponding value for the nominalcurrent).

    The "high" mode permits amplitudes of up to 37 times the nominal current (at150mV for nominal current). The price to pay for the higher dynamic range isthe danger of distorted signals if harmonics exceed certain limits (see thefollowing limits table).

    Current limits for 50Hz nominal frequency and 150mV coming from theRogowski coil at nominal current in "normal" and "high" mode:

    With this Service Release it is possible to do this setting directly in the WindowsRegistry (refer to the document "Test Universe Registry Settings.pdf" that is part

    of this release). A more user-friendly solution will be presented in the next majorTest Universe release.

    Normal Mode: High Mode:

    f max. current f max. current

    50Hz 5.9 x In 50Hz 37.0 x In

    100Hz 5.9 x In 100Hz 23.6 x In

    150Hz 5.9 x In 150Hz 15.7 x In

    200Hz 5.9 x In 200Hz 11.8 x In

    250Hz 5.9 x In 250Hz 9.4 x In

    300Hz 5.9 x In 300Hz 7.9 x In

    350Hz 5.9 x In 350Hz 6.7 x In

    400Hz 5.9 x In 400Hz 5.9 x In

  • 8/10/2019 Test Universe Registry Settings

    11/14

    11

    Customizing Standard Characteristics

    Registry Settings

    Note: When "Value name" does not exist, the amplitude dynamics will be set

    to "normal".

    1.11 Separate Language Selection for Test ReportsBy default the language for user interface and test reports is identical (set on theStart Page at S Y S T E M S E T T I N G S | L A N G U A G E ).

    This registry setting activates the possibility to select a report language differentfrom the user interface language.

    Registry Settings

    Note: This is not a 100% solution. For technical reasons, there are somestrings that will remain in the user interface language.

    Key name: HKEY_CURRENT_USER\Software\OMICRON\CM Engine

    Value name: RogowskiAmplitudeDynamics

    Value type: String (REG_SZ)

    Value data: Either "normal" or "high"

    Key name: HKEY_CURRENT_USER\Software\OMICRON\General Settings

    Value name: AllowSeparateLanguageSelection

    Value type: String (REG_SZ)

    Value data: 0 disables a separate selection

    1 enables a separate selection

  • 8/10/2019 Test Universe Registry Settings

    12/14

    OMICRON Test Universe

    12

    2 How to Edit the Windows RegistryNote: The settings in the Windows Registry are crucial for your systems

    operating stability and integrity. Proceed with caution when editing theWindows Registry. If you do not feel confident editing the Registry, askyour system administrator or a likewise qualified expert.

    To create or modify an entry, you need to start the Windows Registry Editor:

    1. Open the Windows S TA R T menu2. Select R U N

    3. Type in "regedit"

    Figure 2-1:Starting the WindowsRegistry Editor

    4. Left-click OK or hit the key. This will start the Windows RegistryEditor.

    Figure 2-2:The Windows RegistryEditor

    In the Windows Registry Editor:

    1. Left-click the + sign of HKEY_CURRENT_USER to expand the Explorer-like registry tree.

    2. Navigate to HKEY_CURRENT_USER\Software\OMICRON.

  • 8/10/2019 Test Universe Registry Settings

    13/14

    13

    Customizing Standard Characteristics

    The settings mentioned in section Customizing Standard Characteristics ofOMICRON Test Universe are located in subkeys (or subfolders, in terms of theExplorer-like tree) of the key HKEY_CURRENT_USER\Software\OMICRON.

    Clicking such a subkey will display one or more entries at the right side of theRegistry Editor window.

    Figure 2-3:The OMICRON-specificsettings in the WindowsRegistry Editor

    2.1 Editing an entryTo edit an existing entry (String or DWORD):

    1. Double-click the name of the entry (e.g. ScanLPT1).

    Depending on whether the entry is a string or a DWORD, the accordingdialog box Edit String or Edit DWORD Value opens. Here the value ofthe entry can be changed.

    Figure 2-4:The Edit dialog box

    2. Enter the value of your choice into the "Value data:" entry field, and clickO K to close the Edit String dialog box.