25
Temperature Measurements According to the International Temperature Scale of 1990 and Associated Uncertainties Steffen Rudtsch 230 th PTB Seminar Conductivity and Salinity PB 7.42

Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

  • Upload
    lehanh

  • View
    220

  • Download
    5

Embed Size (px)

Citation preview

Page 1: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Temperature Measurements According to the International

Temperature Scale of 1990 and Associated Uncertainties

Steffen Rudtsch

230th PTB Seminar Conductivity and Salinity

P B

7.42

Page 2: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Content

Steffen Rudtsch Physikalisch-Technische Bundesanstalt

Thermodynamic temperatures and temperature scales

Definition of the unit kelvin

ITS-90 and fixed points

Resistance thermometry

Error sources and uncertainty budget

Page 3: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Definition of a Temperature ScalePhysikalisch-Technische

Bundesanstalt

W. Thomson (later Lord Kelvin, 1848) Absolute zero and a second fixed point Efficiency of reversible Carnot-cycle T1=T2(1-η)

Problem: Temperature is an intensive thermodynamic property (value does not depend on the amount of the substance).

How can we define a universal temperature scalewhose properties does not depend on specific substances or thermometers?

Page 4: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Measurement of Thermodynamic Temperatures Physikalisch-Technische

Bundesanstalt

Gas thermometer: equation of state of an ideal gas

tpwtpwtpw

bzw.VpVp

TTTkNVp == k = 1,38065·10-23 J/K

(Boltzmann constant)

Dielectric constant gas thermometer (DCGT)

VN

00 αεε += α0- atomic polarisibility

V=const., p→0 N: Number of atoms

0

0tpw

tpwtpw εεεε

−−

⋅=pp

TT

0

0

αεε −= Tkp

Measurement of the change in capacitance of a gas-filled capacitor after evacuation

Page 5: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Thermodynamic TemperaturesPhysikalisch-Technische

Bundesanstalt

Basis: Fundamental Physical Laws

Gas Properties:- Equation of state of an ideal gas- Dielectric constant- Speed of sound- Spectral-line Doppler broadening

Electron Gas:- Thermal noise

Photons:- Total radiation of a blackbody- Spectral emission of a blackbody

Page 6: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

International Temperature Scale of 1990 Physikalisch-Technische

Bundesanstalt

Definition of the unit kelvin

Fixed points of the ITS-90

Interpolation instruments

Interpolation equations

Page 7: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Definition of the Kelvin and Triple Point of Water Physikalisch-Technische

Bundesanstalt

Ttpw = 273,16 K (Definition)ptpw = 611,66 Pa

The kelvin, unit of thermodynamic temperature, is the fraction 1/273.16 of the thermodynamic temperature of the triple point of water.

13. CGPM (1967): Metrologia, 4 (1968), 147

Page 8: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Fixed Points of the ITS-90

←←←←←←←←←←←←←←←

T90

∆ Tth.dyn /mK ∆ TITS-90 /mK Cu (fp) 1084,62 °C 60 15 Au (fp) 1064,62 °C 50 10 Ag (fp) 961,78 °C 40 3 Al (fp) 660,323 °C 25 2 Zn (fp) 419,527 °C 13 1 Sn (fp) 231,928 °C 5 1 In (fp) 156,5985 °C 3 1 Ga (mp) 29,7646 °C 1 0,25 H2O (tp) 273,16 K 0 0,1 Hg (tp) 234,3156 K 1,5 0,25 Ar (tp) 83,8058 K 1,5 0,3 O2 (tp) 54,3584 K 1 0,4 Ne (tp) 24,5561 K 0,5 0,4 e-H2 (vp) ≈ 20,3 K 0,5 0,4 e-H2 (vp) ≈ 17,0 K 0,5 0,4 e-H2 (tp) 13,8033 K 0,5 0,4 4He (vp) 4,2221 K 0,3 0,1

Page 9: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Fixed-Point MeasurementPhysikalisch-Technische

Bundesanstalt

Page 10: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

ITS-90: Interpolation Instruments and EquationsPhysikalisch-Technische

Bundesanstalt

0,65 K to 5,0 K:Vapour pressure 3He und 4He

( )( )[ ]

( )[ ]( )[ ]( ) 1exp

1XexpX 1

902

1902

90

90

−−= −

TcTc

TLTL

λλ

λ

λabove 961,78 °C:Radiation Thermometer

X: Ag, Au, Cu; c2=0,014388 m·K

( ) ∑=

+=

9

1

90090r 481

754,15-K/i

i

iTCCTW13,8033 K bis 961,78 °C:

Standard Platinum Resistance Thermometer

( ) VNTBcpbpaT

/1 90x

2

90 +++=3,0 K to 24,5561 K:

Gasthermometer (He)

i

ii C

BpAAT ∑=

+=

9

1090

-Pa/)ln(K/

Page 11: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Standard Platinum Resistance ThermometerPhysikalisch-Technische

Bundesanstalt

Platinum wire of highest purity Induction free winding, Free of mechanical stressSpecific gas filling in order to control oxidation Special heat treatment....

Page 12: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

IPRTSPRT

Page 13: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Reference Function for Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

( ) ∑=

+=

9

1

90090r 481

754,15-K/i

i

iTCCTW

Reference Function0 °C to 961.78 °C:

( ) i

i

ri

TWDDT ∑=

−++=

9

1

90090 64,1

64,215,273K/

Callendar-van Dusen: ( ) [ ]32 )100(1)C0( ttCBtAtRtR −+++⋅°=

ITS-90 (SPRT calibration)

Determination of resistance ratio W is required.- extrapolated to I = 0 mA

( )0SPRT

SPRT

(TPW)PointFixed

RR

RRW ==

Page 14: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Calibration of Platinum Resistance ThermometersPhysikalisch-Technische

Bundesanstalt

Purity of the Platinum: W(29,7646 °C) ≥ 1,1807 W(-38,8344 °C) ≤ 0,844235

Calibration Certificate for SPRTs:

Deviation Function (deviation from reference function)

[ ]∑=

++−+−=−5

1

2290r90 )ln()1()1()()(

i

ii WcWbWaTWTW

Page 15: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

SPRT-Calibration: Uncertainty BudgetPhysikalisch-Technische

Bundesanstalt

0,920,710,660,450,450,13Standard Uncertainty0,300,300,150,150,200,05Type A0,870,650,640,430,400,12Type B (combined)0,200,200,060,060,060,03Fixed Point Value0,100,100,100,100,200,01Immersion0,200,200,200,200,150,05Self-heating0,280,200,150,110,090,08TPW0,250,200,160,120,110,02Bridge0,010,010,010,010,010,01Standard Resistor0,300,300,120,080,100,01Gas Pressure0,080,020,020,020,020,01Hydrost. Pressure0,650,400,540,310,250,06Impurities

u(TAg) mK

u(TAl) mK

u(TZn) mK

u(TSn) mK

u(TIn) mK

u(TGa) mK

Quantity

PTB-Budget

0,54

0,66

Page 16: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance Measurement Physikalisch-Technische

Bundesanstalt

Resistance Thermometertypically between 25 Ω and 10 kΩ

1. Digital Multimeter (8 ½ digits) U(T) ≈5 mK Problems: - Stability reference resistor

- Self-heating- Offset thermovoltages

2. Potentiometric MethodU(T) < 1 mKHigh-precision standard resistor RS including its temperature controldifferent currents are used

17

rel

S a102d

d −−⋅<

tR

RS

RT

I ≈ 1 mA

Page 17: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Resistance MeasurementPhysikalisch-Technische

Bundesanstalt

AC-BridgesInductive voltage divider (ratio transformer)typically 25 Hz or 30 Hz Manufacturer: Automatic Systems Laboratories (ASL)

“DC”-BridgesCurrent comparator (current reversal necessary)Manufacturer: Measurement International, Guildline

8

S

T 102 −⋅<

RRU U(T) < 5 µKHighest level with

commercial equipment

Page 18: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Measurement Errors and Corrections Physikalisch-Technische

Bundesanstalt

1. Self Heating of the Resistance Thermometer

Page 19: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Typical Measurement Errors Physikalisch-Technische

Bundesanstalt

1. Stability of the Thermometer- resistance might be changed by mechanical shock or oxidation (SPRT, IPRT)- Drift even after careful heat treatment (Thermistors)

6. Hysteresisup to ∆T = 500 mK at IPRTs (not at SPRTs)- depend on thermal “history” of the IPRT- caused by thermal expansion of the ceramic substrate / powder

Page 20: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Thermal Measurement ErrorsPhysikalisch-Technische

Bundesanstalt

4. Change of the temperature of the investigated object- heat conduction- heat storage/capacity

5. Heat dissipation and thermal couplingi.e. thermometer and object have different temperatures

6. Dynamic behavior and thermal lag

7. Radiation errors

A contact thermometer measures only its own temperature.

Page 21: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Heat Dissipation and Thermal Coupling Physikalisch-Technische

Bundesanstalt

TFurnace

TEnvironment

Page 22: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Dynamic Behavior and Thermal Lag Physikalisch-Technische

Bundesanstalt

T1

T2

The dynamic behavior depend on the construction ofthe thermometer and the thermal coupling between object and thermometer.

Page 23: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Example of an Uncertainty Budget Physikalisch-Technische

Bundesanstalt

Calibration of a resistance thermometer in a liquid bath(more details in DKD-R 5-1)

tx = tN + δtKal + δtDrift + cRδRR + δtBr + δtWaN + δtWAP + δtEWN + δtHom + δtStab

tx: ITS-90 temperature of the resistance thermometer to be calibratedtN: Mean of the temperature of the SPRT (reference)δtKal: Correction based on the SPRT calibrationδtDrift: Correction based on the drift of the SPRTcRδRR: Correction based on the calibration of the standard resistorcR: Sensitivity of the bridgeδtBr: Correction based on the bridge calibrationδtWaN: Correction based on heat conduction of the SPRTδtEWN: Correction based on self-heating of the SPRTδtWAP: Correction based on heat conduction of the resistance thermometerδtHom: Correction based on the in-homogeneity of the bathδtStab: Correction based on the stability of the bath

Page 24: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Example of an Uncertainty Budget Physikalisch-Technische

Bundesanstalt

10 mK180,234 °CTemperaturetx

3.5 mK1Rectang.3.5 mK0 KStability bathδtStab

4.6 mK1Rectang.4.6 mK0 KHomogeneity bathδtHom

1.2 mK1Rectang.1.2 mK0 KHeat cond. SPRTδtWaN

1.5 mK1Normal1.5 mK0 KBridgeδtBr

0.7 mK10 K/ΩNormal0.07 mΩ0 ΩStandard ResistorδRR

3.5 mK1Rectang.3.5 mK0 KDrift SPRTRefδtDrift

7.5 mK1Normal7.5 mK0 KCalibration SPRTδtKal

1.2 mK1Normal1.2 mK180.234 °CTemperature SPRTtN

Uncertaintycontribution

Sensitivitycoefficient

Probabilitydistribution

Standarduncertainty

EstimateQuantity

Calibration of a resistance thermometer in a liquid bath(more details in DKD-R 5-1)

Page 25: Temperature Measurements According to the International ... · PDF fileTemperature Measurements According to the International ... Measurement of the change in capacitance of a gas-filled

Summary Physikalisch-Technische

Bundesanstalt

Definition of the unit kelvin

International Temperature Scale of 1990

Sensors and instruments

Sources of uncertainty

Uncertainty budget