High Temperature Mechanical Testing Committee
NPL Thermal Measurement Awareness Network
UKAS
Temperature Measurement in High Temperature Mechanical Testing
Wednesday 15th March 2006
Programme
09:00 Registration and coffee 09:55 Welcome - Peter Quested, NPL 10:00 The requirements of temperature measurement:
Fatigue, tensile, thermomechanical, limitations of temperature measurement, uncertainties of measurement Phil G Jones, ALSTOM Power Paul R McCarthy, Paul McCarthy Consulting
10.30 Meeting the requirements of temperature measurement – The role of UKAS Phil Endersby, UKAS
10:45 Thermocouples: How they work : Best practice : Are present uncertainties acceptable ?David Head, NPL
11:00 Practical considerations using thermocouples for measuring high temperatures during long duration mechanical testing Malcolm Loveday / Peter Quested, NPL
11:35 Discussion 12:00 Lunch 13:00 Radiation thermometry
Helen McEvoy, NPL 13:30 Recent developments in industrial non-contact thermometry at NPL, including
Laser Absorption Radiation Thermometry (LART) Andrew Levick, NPL
13:50 Dynamic temperature measurement and use of ribbon thermocouples Peter Haehner, DG-JRC
14:30 Comparison: Thermocouples and optical pyrometers Matt Brooks, NPL
14:50 Concluding discussion
15:10 Laboratory Tour
Temperature Measurement in High Temperature Mechanical Testing
Wednesday 15th March 2006 Abstracts
The requirements of temperature measurement Phil G Jones - ALSTOM Power
Paul R McCarthy - Paul McCarthy Consulting The need for accurate, reliable temperature measurement is key to the successful testing of materials at elevated temperatures. Currently, the bulk of temperature measurement is performed using thermocouples complemented, in limited cases, by pyrometry and PRT’s (Platinum Resistance Thermometers). This paper reviews the physical and operational characteristics of the various sensor types for isothermal and dynamic applications. It examines a range of inherent properties and considers the interactions between measurement sensors and the associated elements of the measurement system / electronics. The requirements of a number of testing procedures / standards are reviewed, producing a set of “ground rules” within which the temperature measurement sensors need to operate. The ability of the various sensors to meet these requirements is explored and some conclusions drawn as to suitability for use in isothermal or dynamic applications. The paper concludes with an assessment of the application of uncertainties of measurement of sensors and measurement systems to the permitted temperature ranges within national / international standards, technical consensus documents and codes of practice for elevated temperature testing.
Meeting the requirements of temperature measurement – the role of UKAS
Phil Endersby - Assessment Manager, UKAS UKAS assesses the competence of all aspects of high temperature mechanical testing, including appropriate calibration and use of temperature measuring equipment. This paper addresses some of the factors which need to be considered when specifying requirements for internal or external calibration of temperature measuring equipment. This is illustrated for the calibration of a base metal thermocouple. The schedule of accreditation of a specialist temperature calibration laboratory is used to illustrate how the characteristics of PRTs, noble metal thermocouples and base metal thermocouples affect the uncertainty of measurement that can be achieved in practice. Consideration is then given to the process of UKAS’ assessment of internal calibration of temperature measuring systems, to confirm that they are fit for purpose.
1
Thermocouples David Head - NPL
An introduction into how thermocouples work and the actual signal source will be given. Some example uncertainty budgets will be reviewed. There will be a discussion on how a lab comparison calibrates thermocouples. The presentation will conclude with some thoughts on how to set up "industrial" measurements to avoid some obvious pitfalls.
Practical considerations using Thermocouples for measuring high temperatures during long duration mechanical testing
Malcolm Loveday, Bryan Roebuck, Peter Quested - NPL The measurement of temperature during long duration mechanical tests is a key parameter and can contribute significant uncertainties in the measurements with consequences designing with the data, The thermocouple is the most frequently used temperature sensor because it is relatively cheap and small and if a few simple precautions are taken it is capable of achieving the required uncertainties of measurement. A list of factors, which need to be considered. For the design of the assembly:
• Wire types (use of type N and noble metal thermocouples). • Wire diameter (thin wires are more prone to surface reactions but thicker
wires act as heat shunts). • Sheath materials must be stable. (danger in reducing atmospheres of
reducing some oxide ceramics). Or operating conditions:
• Temperature range, which effects choice of wire materials. • Time. The longer the duration of the test, the more likely to degrade. • Environmental gas. (In oxidising environments Rh can migrate in noble
metal thermocouples and under reducing conditions need stable sheath materials).
• Pressure levels. (special interface design. Pressure has little effect on output of thermocouples).
• Thermal cycling • Ionising radiation (transmutations of material change the thermoelectric
voltage characteristic). • Strong electromagnetic alternating fields. (induction heating). • Strong inhomogeneous magnetic fields along thermoelectric wires can
influence thermoelectric voltages. Despite its simplicity the thermocouple is a precision instrument which needs to be handled with care. It has been recognised for many years that the characteristics of the wire can be changed by subjecting it to mechanical work and this is the reason that the wires are annealed at high temperatures. Therefore care needs to be taken minimising the bending of the wires. Exit wires from the furnace with the associated high temperature gradients means that any damage in this area has a disproportionate effect on the output voltages. Good practice is not only to calibrate the thermocouple but also examine it for inhomogeneities by for example making sure that in the calibration furnace the depth of immersion of the
2
thermocouples is identical to the testing set up or passing a temperature spike down the assembly maintaining both junctions at 0 °C monitoring any changes in the voltages.
Radiation thermometry Helen McEvoy - NPL
Radiation thermometry, or ‘pyrometry’, is the technique whereby the temperature of a surface is determined by measuring the amount of thermal energy or radiance emitted by that surface. Measurements are made using a detector-based instrument, situated remotely from the surface. The technique is non-contact, meaning that it can be used to measure temperatures in situations where a contact probe would be unsuitable or impractical; for example, if the target is very hot, moving, in an hostile environment or where contamination of a product needs to be avoided. Also, thermal imaging cameras allow many points over a large area to be measured simultaneously. However, care is required when using the method and there are issues that need to be taken into consideration to prevent large errors in the temperature reading; for example, there must be a clear line-of-sight between the target and the thermometer and the emissivity of the surface needs to be known or understood. The talk will give the principles behind radiation thermometry, describe the construction of a typical radiation thermometer and discuss some of the calibration issues, applications and advantages of using the technique. It will also cover the disadvantages and areas of potential error, and give examples of how these errors could be minimised.
Recent developments in industrial non-contact thermometry at NPL including Laser Absorption Radiation Thermometry (LART)
Andrew Levick - NPL Temperature is the most frequently measured parameter in process engineering, and affects the quality of the product and energy consumption of the process. Radiation pyrometry is a well-established non-contact thermometry technique used in a wide range of industrial processes. A radiation pyrometer collects and detects thermal radiation emitted from the target over a narrow band of wavelengths, and infers the temperature via the Planck relationship. However, the radiation pyrometry is prone to systematic errors if the emissivity of the target is uncertain or there is reflected background radiation present. This talk will describe recent developments in non-contact thermometry that are potentially immune to emissivity and background radiation, focusing on two methods:
The Laser Absorption Radiation Thermometry (LART) method is based on photothermal radiometry, and involves the detection of modulated thermal radiances from the target irradiated by modulated high power laser beams. The Radiation thermometry in High Ambient Temperatures (RTHAT) method involves separate direct measurements of the background and target radiances. A least squares analysis is utilised to obtain the surface emissivity
3
and temperature. The technique relies on the extra information provided by fluctuating components in the background radiation that can be correlated with the reflected radiation from the target itself.
We will discuss the pros and cons of each method and possible industrial applications.
Dynamic Temperature Measurement and Use of Ribbon Thermocouples
Peter Haehner - DG-JRC
Comparison: Thermocouples and optical pyrometers Matt Brooks - NPL
Accurate temperature measurements via pyrometry require the characterisation of emissivity as a function of temperature for a given detection wavelength. Surface emissivity for a given material can be affected by dynamic surface conditions such as roughness, oxide layer and contamination. Such variations can be the cause of significant error in the temperature measured by a pyrometer. A small scale mechanical test system, the ETMT, has been used at NPL to investigate the effect of the time dependent oxide scaling process on the temperature measured by a two colour pyrometer when controlled by thermocouple. The time dependent variation of indicated pyrometer temperature was found in some instances be as large as 100°C when compared to the temperature of the controlling thermocouple.
4
March 2006March 2006
The Practicalities of Temperature Measurement
Mr Philip JonesMr Paul McCarthy
Date of last change Reference/Name of Presentation/SN 2
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Throughout the materials testing world, the need for accurate, reliable temperature measurement is a key component to successful testing of materials at elevated temperatures.
TensileCreepRuptureStress relaxationLow & high cycle fatigue
High Temperature Mechanical TestingHigh Temperature Mechanical Testing
Date of last change Reference/Name of Presentation/SN 3
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
RobustAccurate and easy to useFast responseCheapHeating equipment compatibilityAge insensitive
The Ideal Temperature Sensor
Date of last change Reference/Name of Presentation/SN 4
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Thermocouples–Base–Noble
Pyrometers
Platinum Resistance Thermometers (PRT’s)
Other Novel Methods
The Traditional OptionsThe Traditional Options
Date of last change Reference/Name of Presentation/SN 5
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
GoodSpecial needs
GoodOKHeating compatibility
HighHighHighCheapPrice
OKGoodGoodGoodSpeed of response
GoodOK to Good
GoodOKAccuracy
PoorSpecial needs
GoodGoodHandling & Installation
PRTPyroNobleBase
Comparative PerformanceComparative Performance
Date of last change Reference/Name of Presentation/SN 6
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
T/C’s – well developed, extensive experience
Pyrometers – unusual, unless surface contact / contamination issues
PRT’s – rare, due to mounting issues. Excellent for long term, very stable sensor (v. low drift)
Static ApplicationsStatic Applications
Date of last change Reference/Name of Presentation/SN 7
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
T/C’s – attachment issues, also thermal mass and r.f. field considerationsPyrometers – non-contacting , able to access areas difficult for t/c’s, emissivity changes can be problematicPRT’s – thermal mass and contact issues
All are critically dependent upon the associated electronics
Dynamic ApplicationsDynamic Applications
Date of last change Reference/Name of Presentation/SN 8
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
T/C’s– bead size & wire path– strip t/c’s– r.f. interference
Pyrometers– Changes in surface emissivity ( pre-oxidise to
minimise impact)– Electronics - immunity to r.f. interference
RF Heating ConsiderationsRF Heating Considerations
Date of last change Reference/Name of Presentation/SN 9
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
The Critical Issue :- sampling rate vs. accuracy / discrimination
– Greater accuracy requires longer integration times
– For high rate testing this controls temperature uncertainty
• e.g. a test at 50ºC/s, reading at 8 integration's per sec, gives 6.25ºC “steps”
• More integration's possible, but lower discrimination.
Dynamic Testing Dynamic Testing -- ElectronicsElectronics
Date of last change Reference/Name of Presentation/SN 10
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
But we usually test to recognised standard
What do the testing standards call for?
– Test temperature accuracy
– Measuring system capability
– Calibration Frequency
– Test temperature stability
Complying with StandardsComplying with Standards
Date of last change Reference/Name of Presentation/SN 11
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Document number Subject
EN10002-5 : 1992Tensile testing of metallic materials - Part 5: Method of test at elevated temperatures.
ASTM E 21 – 03aStandard test methods for Elevated Temperature Tension Tests of Metallic Materials
BS EN 10291 : 2000 Metallic materials – Uniaxial creep testing in tension - Methods of test
ASTM E 139 - 00є1Standard test methods for Conducting Creep, Creep-Rupture, and Stress- Rupture Tests of Metallic Materials
BS EN 10319-1 : 2003Metallic materials – Tensile stress relaxation testing - Part 1: Procedure for testing machines
ASTM E 328 - 02 Standard Test Methods for Stress Relaxation for Materials and StructuresECCC Recommendations - Volume 3 Part 1
Data acceptability criteria and data generation - Generic recommendations for creep, creep rupture, stress rupture and stress relaxation data.
ASTM E 606 – 92 (Re-approved 2004) Standard Practice for Strain-Controlled Fatigue TestingBS 3518 part 1 1993 Fatigue testing - Guide to general principles
BS 7270 : 1990 Constant amplitude strain controlled fatigue testing - Method for
Review of Testing StandardsReview of Testing Standards
Date of last change Reference/Name of Presentation/SN 12
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test temperature accuracy and deviations.
Thermal monitoring sensors at the specimen surface.
– Temperature sensor attachment good practice.
Measuring equipment accuracy, resolution and stability.
Calibration of the complete monitoring system.
Taking account of uncertainties within the whole system.
Items of Key ImportanceItems of Key Importance
Date of last change Reference/Name of Presentation/SN 13
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test type Procedure Limit type < 600 °C
Max specimen gradient <
600°C 601 – 800
°C
Max specimen
gradient 601 – 800 °C
801 – 1000 °C
Max specimen
gradient 801 – 1000 °C
1001 – 1100 °C
1101 – 1200 °C
1201 – 1350 °C
Max specimen gradient >
1000 °C
ET Tensile EN10002:5 Deviations +/- 3Not
specified +/- 4 Not specified +/- 5Not
specifiedNot
specified
Creep EN 10291 Deviations +/- 3 3 +/- 4 4 +/- 5 5By
agreement
Creep / RuptureASTM E139 Deviations +/- 2
Not specified +/- 2 Not specified +/- 2
Not specified +/- 3 +/- 3 +/- 3
Not specified
Stress Relax EN 10319 Deviations +/- 3 2 +/- 4 3 +/- 5 3By
agreement
Creep & Rupture ECCCMax total deviation +/- 3 +/- 3 +/- 4 +/- 4 +/- 5 +/- 5 +/- 6 +/- 7 +/- 8
Not specified
Uniaxial Stress Relax ECCC
Max total deviation +/- 3 +/- 3 +/- 4 +/- 4 +/- 5 +/- 5
Not specified
Strain-Controlled Fatigue
ASTM E606 – 92 Reapp 2004 Deviations
+/- 2 or reported if greater
+/- 2 or 1% if greater
+/- 2 or reported if greater
+/- 2 or 1% if greater
+/- 2 or reported if greater
+/- 2 or 1% if greater
Not specified
Fatigue testingBS 3518 pt1 Deviations +/- 2 2 +/- 2 2 +/- 2 2 +/- 2 +/- 2 +/- 2 2
+/- 2 or reported if greater
By agreement
By agreement
By agreement
Not specified
This table indicates a number of common accuracy and deviation levels
Test temperature accuracy and deviations
Date of last change Reference/Name of Presentation/SN 14
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test type Procedure Limit type < 600 °C
Max specimen gradient <
600°C 601 – 800
°C
Max specimen
gradient 601 – 800 °C
801 – 1000 °C
Max specimen
gradient 801 – 1000 °C
1001 – 1100 °C
1101 – 1200 °C
1201 – 1350 °C
Max specimen gradient >
1000 °C
ET Tensile EN10002:5 Deviations +/- 3Not
specified +/- 4 Not specified +/- 5Not
specifiedNot
specified
Creep EN 10291 Deviations +/- 3 3 +/- 4 4 +/- 5 5By
agreement
Creep / RuptureASTM E139 Deviations +/- 2
Not specified +/- 2 Not specified +/- 2
Not specified +/- 3 +/- 3 +/- 3
Not specified
Stress Relax EN 10319 Deviations +/- 3 2 +/- 4 3 +/- 5 3By
agreement
Creep & Rupture ECCCMax total deviation +/- 3 +/- 3 +/- 4 +/- 4 +/- 5 +/- 5 +/- 6 +/- 7 +/- 8
Not specified
Uniaxial Stress Relax ECCC
Max total deviation +/- 3 +/- 3 +/- 4 +/- 4 +/- 5 +/- 5
Not specified
Strain-Controlled Fatigue
ASTM E606 – 92 Reapp 2004 Deviations
+/- 2 or reported if greater
+/- 2 or 1% if greater
+/- 2 or reported if greater
+/- 2 or 1% if greater
+/- 2 or reported if greater
+/- 2 or 1% if greater
Not specified
Fatigue testingBS 3518 pt1 Deviations +/- 2 2 +/- 2 2 +/- 2 2 +/- 2 +/- 2 +/- 2 2
+/- 2 or reported if greater
By agreement
By agreement
By agreement
Not specified
Conclusion - for both accuracy and gradient requirements, there are typically two levels specified, the most common (3,4,5°C) or a tighter level.
Test temperature accuracy and deviations
Date of last change Reference/Name of Presentation/SN 15
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Testing laboratory applied control levels, distilled from the various standards
– Temperature deviation limit <600°C: +/- 3°C– Gradient along specimen at <600°C: +/- 2°C
– Temperature deviation limit 601-800°C: +/- 4°C– Gradient along specimen at 601-800°C: +/- 2°C
– Temperature deviation limit 801-1200°C: +/- 5°C– Gradient along specimen at 801-1200°C: +/- 3°C
Test temperature accuracy and deviations
Date of last change Reference/Name of Presentation/SN 16
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test temperature accuracy and deviations.
Thermal monitoring sensors at the specimen surface.
– Temperature sensor attachment good practice.
Measuring equipment accuracy, resolution and stability.
Calibration of the complete monitoring system.
Taking account of uncertainties within the whole system.
Thermal monitoring sensors at the specimen surface.Thermal monitoring sensors at the specimen surface.
Date of last change Reference/Name of Presentation/SN 17
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test type ProcedurePosition for GL < 50mm
Position for GL > 50mm Attachment to specimen
Shielding of thermocouple
T/C wire hot zone insulation
Elevated Temperature Tensile EN10002:5
2 off - 1 near each end.
NOTE: 3 off – arranged at identical intervals along the parallel length.
Measured at the surface of the parallel length. NOTE: T/C junctions should make thermal contact with the surface.
NOTE: suitably screened from direct radiation Not specified
Creep ASTM E21 – 03a2 off - 1 near each end.
3 off - 2 near each end and 1 near the centre. Intimate contact Shielded from radiation Ceramic insulators
Creep EN 102912 off - each end of the parallel length.
3 off - each end of the parallel length third in the middle region.
T/C junctions should make thermal contact with the surface.
screened from direct radiation
Thermally shielded and electrically insulated
Creep ASTM E139 - 002 off - 1 near each end.
3 off - 2 near each end and 1 near the centre. Intimate contact Shielded from radiation Ceramic insulators
Stress Relax EN 103192 off at each end of the parallel length.
3 off - each end of the parallel length third in the middle region.
T/C junctions should make thermal contact with the surface.
screened from direct radiation
Thermally shielded and electrically insulated
Stress Relax ASTM E 328 - 022 off - 1 near each end.
3 off - 2 near each end and 1 near the centre. Intimate contact Shielded from radiation Ceramic insulators
Creep / Rupture ECCC Not specified Not specified Not specified Not specified Not specifiedStress Relax ECCC Not specified Not specified Not specified Not specified Not specifiedStrain-Controlled Fatigue
ASTM E606 – 92 Reapp 2004 Not specified Not specified Not specified Not specified Not specified
Fatigue BS 3518 pt1 Not specified Not specified Not specified Not specified Not specified
Conclusion - with reference to gradient, large samples are penalised.Size matters to the customer.How do you achieve intimate contact? Large samples are fine, small are difficult!
Thermal monitoring sensors at the specimen surface
Date of last change Reference/Name of Presentation/SN 18
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
– Specimen of < 50mm gauge length– 2 sensors positioned near either end
of G.L.– Specimen of > 50mm gauge length– 3 sensors positioned one near either
end of G.L. plus one at near centre.
– Ensure good thermal contact of sensors with specimen.
– Shield sensors from direct radiant heat.
Thermal monitoring sensors at the specimen surface
Date of last change Reference/Name of Presentation/SN 19
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test temperature accuracy and deviations.
Thermal monitoring sensors at the specimen surface.
– Temperature sensor attachment good practice.
Measuring equipment accuracy, resolution and stability.
Calibration of the complete monitoring system.
Taking account of uncertainties within the whole system.
Measuring equipment accuracy, resolution and stability.Measuring equipment accuracy, resolution and stability.
Date of last change Reference/Name of Presentation/SN 20
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test type Procedure Resolution minimum (sensitivity) °CAccuracy
(Precision) °CSurrounding air temperature
max variation °C
ET Tensile EN10002:5 1 +/- 2 Part 1:
CreepASTM E21 –03a
Sufficiently sensitive and reliable to ensure that the temperature of the
specimen is within the limits specified. 6
Creep EN 10291 0.5 +/- 1 +/- 3
CreepASTM E139 -00
Sufficiently sensitive and reliable to ensure that the temperature of the
specimen is within the limits specified. +/- 3
Stress Relax EN 10319 0.5 +/- 1 +/- 3
Creep / Rupture ECCC Not specified Not specified +/- 3
Interrupted Creep / Rupture ECCC Not specified Not specified +/- 2
Uniaxial Stress Relax ECCC Not specified Not specified +/- 3
Strain-Controlled Fatigue
ASTM E606 –92 (2004) Not specified Not specified Not specified
Fatigue BS 3519 pt1 Not specified Not specified Not specified
Measuring equipment accuracy, resolution and stability
Date of last change Reference/Name of Presentation/SN 21
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Is test temperature deviation, independent from measuring system accuracy?Examples:
– Temperature deviation limit: +/- 3°C - (BSEN10291)
– Measuring equipment accuracy (precision): +/- 2°C - (BSEN10291)
– Temperature deviation limit: +/- 2°C - (ASTM E139)
– Measuring equipment accuracy (precision): +/- 3°C - (ASTM E139)
Measuring equipment accuracy, resolution and stability
Date of last change Reference/Name of Presentation/SN 22
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test temperature accuracy and deviations.
Thermal monitoring sensors at the specimen surface.
– Temperature sensor attachment good practice.
Measuring equipment accuracy, resolution and stability.
Calibration of the complete monitoring system.
Systematic error correction.
Taking account of uncertainties within the whole system.
Calibration of the complete monitoring system.Calibration of the complete monitoring system.
Date of last change Reference/Name of Presentation/SN 23
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Document number
Sensor calibration Frequency Tracability system errors
System Cal Frquency
EN10002-5 : 1992 12 monthstraceable to the international unit (SI)
verification of temperature- measuring system 12 months
ASTM E 21 – 03a
As frequently as is necessary to comply with test method
against secondary standard / E 220
verified against secondary standard Periodically
BS EN 10291 : 2000
12 months or before and after test
traceable to the international unit (SI)
Measurement system (cable, the connection, cold junction, indicator or recorder) 12 months
ASTM E 139 - 00є1 Each lot run Method E 220Calibration of temperature measuring equipment / system 3 Months
BS EN 10319-1 : 2003
12 months or before and after test
traceable to the international unit (SI)
Measurement system (cable, the connection, cold junction, indicator 12 months
ASTM E 328 - 02 Each lot run Method E 220Calibration of temperature measuring equipment / system 3 Months
ECCC Volume 3Temperature dependent National standard
Corrected, where necessary, for all systematic errors not specified
ASTM E 606 – 92 (2004)
no greater than 6 months
National Institue of Standards and Technology
If automated system is used computer output calibration is necessary
no greater than 6 months
BS 3518 pt1 not specified not specified not specified not specified
Conclusion – Maximum system calibration frequency 12 months.Sensor calibration frequency material and temperature / time dependant.
Calibration of the complete monitoring systemCalibration of the complete monitoring system
Date of last change Reference/Name of Presentation/SN 24
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Test temperature accuracy and deviations.
Thermal monitoring sensors at the specimen surface.
– Temperature sensor attachment good practice.
Measuring equipment accuracy, resolution and stability.
Calibration of the complete monitoring system.
Systematic error correction.
Taking account of uncertainties within the whole system.
Taking account of uncertainties within the whole system.Taking account of uncertainties within the whole system.
Date of last change Reference/Name of Presentation/SN 25
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
All areas within the temperature measuring system that may contribute to errors of the actual temperature, should be taken in to account.
– Um Max error in the measurement sensor (e.g. Thermocouple)– Ucs Max error for coupling of the thermocouple to the specimen.– Ums measurement system (including “cold junction" or Reference temp)
– Uu along specimen uniformity (Gradient)
– Ud Sensor drift with time– Uid change in depth of immersion from calibration– Uet environmental temperature influences on measuring system– T Test temperature
Taking account of uncertainties within the whole system
Date of last change Reference/Name of Presentation/SN 26
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Formula for uncertainty of temperature measurement is as follows:-U = √[Um(T)2 + Ucs(T)2 + Ums(T)2 + Uu(T)2 + Ud(T)2 + Uid(T)2 + Uet(T)2] / √ 3
– Um - from thermocouple calibration– Ucs - will be ‘zero’ for intimate contact– Ums - from system calibration check– Uu - from T/C gradient measurements– Ud - drift published information– Uid - this can be ignored if the calibration immersion is similar– Uet - calculated from voltage readings at 2 ambient temps– T - defined Test temperature
Taking account of uncertainties within the whole system
Date of last change Reference/Name of Presentation/SN 27
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
For further details please refer to the following documents:-
– UKAS M3003
– Uncert Codes
– CEN Workshop agreement
Taking account of uncertainties within the whole system
Date of last change Reference/Name of Presentation/SN 28
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Practical ramifications an example– Test temp 550°C– Standard allows +-3°C– T/C cal lab uncertainty = +-1.5°C– Cal of measuring system = +-0.5°C– This leaves +- 1.4°C for gradient, furnace control band
and drift– Is this sufficient?– Can the cal lab uncertainties be reduced?
Taking account of uncertainties within the whole system
Date of last change Reference/Name of Presentation/SN 29
The Practicalities Of Temperature MeasurementHTMTC & TMAN, NPL, March 2006
Thank you
Questions or comments?
The EndThe End
www.alstom.com
HTMTC 15 March 2006 1
Meeting the requirements of temperature measurement
The role of UKAS
Phil EndersbyUKAS
HTMTC 15 March 2006 2
Overview
• Illustration of the effects of thermocouple characteristics
• Specifying calibration requirements
• UKAS assessment of internal temperature calibration
HTMTC 15 March 2006 3
Factors limiting uncertainty in thermocouple calibration
• Thermocouple homogeneity
• Thermocouple output sensitivity
• Thermocouple ageing
• Furnace uniformity
• Traceability
HTMTC 15 March 2006 4
Device Range BMCPlatinum resistance thermometers
0 °C0.01 °C-196 °C-90 °C to 0 °C-0 °C to 250 °C250 °C to 300 °C300 °C to 420 °C420 °C to 660 °C
0.005 °C0.003 °C0.007 °C0.007 °C0.010 °C0.015 °C0.03 °C0.05 °C
Thermocouples -base metal
-196 °C-90 °C to 0 °C0 °C to 40 °C40 °C to 80 °C80 °C to 350 °C350 °C to 420 °C420 °C to 660 °C660 °C to 1100 °C1100 °C to 1300 °C
0.2 °C0.15 °C0.10 °C0.15 °C0.2 °C0.3 °C0.4 °C0.9 °C2.3 °C
Thermocouples -noble metal
0 °C to 280 °C280 °C to 660 °C660 °C to 1100 °C1100 °C to 1300 °C
0.5 °C0.45 °C0.9 °C2.3 °C
Accredited BMC for
specialist temperature calibration laboratory
HTMTC 15 March 2006 5
Factors limiting uncertainty Areas of most significance in italics
• Thermocouple homogeneitybase metal thermocouples
• Thermocouple output sensitivitynoble metal thermocouples, lower temperatures
• Thermocouple ageingbase metal thermocouples, after high temperature
• Furnace uniformityhigher temperatures
• Traceabilityabove 1100°C
HTMTC 15 March 2006 6
Considerations for optimum calibration
• The effects of immersion depth• Annealing• Ageing• Re-calibration• Repeat points• Homogeneity• Batch characterisation
HTMTC 15 March 2006 7
Thermocouple immersionRe-calibration of base metal orcheck prior to annealing for noble metal…
…unlikely to be of any benefit
Immersion during use
Immersion during calibration
…may sometimes be useful
Immersion during use
Immersion during calibration
HTMTC 15 March 2006 8
Repeat pointscan be useful but…
Immersion up to 250°C
Immersion above 250°C
…a repeat point as 250°C for this calibration
Information about repeatability/homogeneity
X Information about ageing/annealing
HTMTC 15 March 2006 9
Ageing, homogeneity and batch characterisation
• Base metal thermocouples can be aged at an appropriate temperature prior to calibration, to give greater stability
• Homogeneity effects can be estimated by changing immersion during calibration
• Batches of suitably annealed/aged base metal thermocouples can be characterised using a representative sample of the batch
HTMTC 15 March 2006 10
UKAS assessment of internal calibration
• Fit for purpose• TPS52 and AC Supp
Internal calibration must be assessed to confirm that it is competent and fit for purpose
• UKAS technical assessorsUKAS will generally use out own specialist staff for assessment of internal calibration in test labs, because of ‘flexibility’ required
Thermocouples : Effects of Long Term Drift, Atmosphere & Strain
Peter Quested, Bryan Roebuck Matt Brooks & Malcolm Loveday
Materials Centre, Division of Engineering & Process Control
NPL
March 2006
Creep Testing
Temperature accuracy is important in test reproducibility.
Source: M.S Loveday and R Morrell ‘Standardisation of Mechanical Testing and Quality Control’ in “Mechanical Testing of Engineering Ceramics at High Temperatures” edited by B. F. Dyson, R.D. Lohr & R. Morrell, Pub:Elsevier Applied Science, London, 1989.
Thomas Johann Seebeck1770 - 1831
Undated engraving,
Deutsches Museum, Munich
T. J. Seebeck’s discovery 1821The emf* is proportional to the temperature difference
http://chem.ch.huji.ac.il/~eugeniik/history/seebeck.html* emf – electromotive force ( voltage)
V » T1 – T2
V
T1 T2
Note: He even found that electrical currents flowed if one portion of a wire was hammered or twisted while the other portion of the same wire was not reshaped, ie influence of Cold Working
J. C. A. Peltier’s discovery 1834Passing a current through the couple generates
a different temperature at the junctions & depends on the polarity of the applied voltage
( Peltier cooling )
I
+ -
Jean Charles Peltier ( French ) 1785 - 1845
T1 T2
William Thomson (Lord Kelvin)British
Sir William Thompson1824 - 1907
( later Lord Kelvin )
Discovered in 1854 that the emf* is generated in Temperature Gradients
rather than at the junctions
* emf – electromotive force ( voltage)
Operating conditions.
• Temperature range, effects choice of wire materials.• Time. The longer the duration of the test, the more likely
thermocouples are to degrade.• Environmental gas. (In oxidising environments Rh can migrate in
noble metal thermocouples and under reducing conditions need stable sheath materials.)
• Pressure. (special interface design. Pressure has little effect on output of thermocouples.)
• Thermal cycling (degrade the thermocouples)• Ionising radiation (transmutations of material change the
thermoelectric voltage characteristic.)• Strong electromagnetic alternating fields. (induction heating)• Strong inhomogeneous magnetic fields along thermoelectric wires
can influence thermoelectric voltages.• Mechanical working This can effect the thermoelectric output.
Design of assembly
• Wire types (use of type N and noble metal thermocouples.)
• Wire diameter (thin wires are more prone to surface reactions but thicker wires can act as heat shunts.)
• Sheath materials must be stable. (danger in reducing atmospheres of reducing some oxide ceramics)
• Avoid mechanical work in thermocouples.
Choice of thermocouples.
There are five base metal and three precious metal thermocouplesThe base metal types are T,J,E,K and N. Types T, J,and E either have restricted temperature ranges or are unstable
from 300°C.
Type + leg - leg Range (°C) long term use
Comments
K Ni-Cr Ni-Al 0 to 1100 High sensitivity; cheap; easy to use but unstable.
N Ni-Cr-Si Ni-Si-Mg 0 to 1100 More expensive and more stable than Type K.
Long term performance of types K and N Thermocouples.
Type N thermocouples have better long term performance than Type K.N A Burley and T P Jones “Practical Performance of Nicrosil-Nisil Thermocouples” Inst.Phys.Conf, Ser, 26,1975, 172-180.
Precious metal thermocouples.
Type +leg -leg Approximate long term °C*
Approximate short term °C
Comments
R Pt-13Rh Pt 0 to 1600 -50 to 1700 Stable in air or oxidising conditions. Care in vacuum especially stability of ceramic sheaths. High cost. Relatively low sensitivity. Considered to be very stable. Easily contaminated
S Pt-10Rh Pt 0 to1550 -50 to 1750 See type R
B Pt-30Rh Pt-6Rh 100 to1600 100 to 1820 Similar to R and S but not as popular. Used in glass industry. Requires no cold junction.
* Also depends on the thickness of the wire.
Selected Ranges for Best Measurement Capability at NPL for Various Types of Thermometer.
Measured Quantity Range Best MeasurementCapability Expressed as an Expanded Uncertainty (k=2)
Resistance thermometers, fixedpoint calibrations
419 °C to 660 °C660 °C to 961 °C
±0.001 °C to ±0.003 °C±0.003 °C to ±0.005 °C
Resistance thermometers,calibration by comparison
200 °C to 550 °C ±0.02 °C
Liquid-in-glass thermometers 0 °C to 50 °C450 °C to 550 °C
±0.003 °C±0.5 °C
Thermocouples, noble metal types 0 °C to 1100 °C1100 °C to 1600 °C
±0.3 °C±1.0°C to ±1.5°C
Thermocouples, base-metal types 200 °C to 550 °C550 °C to 1100 °C1100 °C to 1200 °C
±0.5 °C±1.0 °C±2.0 °C
Infrared Thermometers 500 °C to 1000 °C1000 °C to 3000 °C
±0.4 °C0.05% of Celsius temperature
Source: Schedule of Accreditation for National Physical Laboratory issued by UKAS, October 2004.
• Drift is a change in thermocouple output while it is within the test environment which in turn remains constant,– Chemical contamination; compositional change; straining of
both limbs.• Depth of Immersion Error is a difference in output of the
thermocouple relative to a known fixed standard when the thermocouple is placed in a different from that of the test furnace.
Calibrate working thermocouples against a Standard T/C in a furnace with similar
depth of immersion to that used on a testing machine
H Robson (1977) ‘Temperature measuring techniques in a large creep laboratory’ J Phys E : Sci Instruments 10 384 - 389
Thermocouples
Temperature versus
position
8.02 mV
Voltageversus
position
glass tube
The reproducibility of the platinum thermocouple at the freezing points of gold, silver and antimony.M de Selincourt. Proc Phys Soc 51 145- 158 ( 1939)
Pure Platinum Wire( data from Selincourt, 1939)
0
2
4
6
8
10
12
0 1 2 3 4 5Number of Strainings
Tem
pera
ture
D
iffer
ence
°C
New Wire
Old wire annealed 1600°CNew Wire
New Wire
Platinum Alloy Wire( Selincourt, 1939 )
-3
-2
-1
0
1
2
0 1 2 3 4 5
Number of Strainings
Tem
pera
ture
D
iffer
ence
°C
New wire
Old wire annealed at1600°COld wire annealed at1600°CNew wire
Old wire annealed at1600°C
8 mm
Influence of straining (1)
[ Early NPL work in 1930’s ]
Annealing Time & Temperature
Strained Alloy Wire: Influence of annealing time ( Selincourt 1939)
0
1
2
0 60 120
Annealing Time, minutes
Tem
pera
ture
Di
ffere
nce
°C
Annealed at 800°CAnnealed at 1000°CAnnealed at 1400°C
Re-plotted by Loveday assuming 10 µV / °CThe reproducibility of the platinum thermocouple at the freezing points of gold, silver and antimony.
M de Selincourt. Proc Phys Soc 51 145- 158 ( 1939)
[ Early NPL work in 1930’s ]
Temperature : 100°C 500°C 1000°CThermocouple Type µV / °C µV / °C µV / °C
Iron- Constantan( Fe-Cu/Ni ) Type J
55 56 -
Chromel- Alumel ( Ni/Cr – Ni/Al) Type K
40 40 40
Noble(10%Rh/Pt – Pt) Type S
8 10 12
Noble(13% Rh/Pt – Pt) Type R
7 11 13
Examples of approximate Seebeck coefficients for some thermocouples
Influence of straining (2)
Thermocouple Error due to Strain (Fenton 1969)
-15
-10
-5
0
5
10
15
0 50 100 150
Strain %
Ap
rox
ima
te
Te
mp
era
ture
Err
or
°C Pt 600°CPt 800°CPt 1000°CPt 1200°CPt-13%Rh 600°CPt-13%Rh 800°CPt-13%Rh 1000°CPt-13%Rh 1200°C
A.W.Fenton ( 1969) ‘ Errors in thermoelectric thermometers’.
Proc. Inst. Engrs. 116 (7) 1277 – 1285.(Data re-plotted by Loveday 2006)
In-situ calibration of thermocouples using a phase transformation
Ti - phase change at 885°C ( α → β )
Eutectoid steel – phase change on heating at 745°C- 750°C Pearlitic ( bcc Fe-C / Fe3C) → austenitic ( fcc Fe-C )
Phase change on cooling ~ 655°C
ETMT Measurements
600 650 700 750 800 850 900600
650
700
750
800
850
900
950
1000
1050
1100
1150
1200R
esis
tivity
n
Wm
Temperature oC
e1 heat e1 cool a1 heat a1 cool
Pt - undeformedPt/Rh - undeformed
Heat to 850°C at 2°C / sCool to RT at 2°C / s
In-situ calibration of thermocouples using a phase transformationRepeat measurements on different thermocouples
( Roebuck et al 2006, NPL )
ETMT Measurements : Strained Thermocouples ( Roebuck et al 2006)
0 200 400 600 800 1000
200
400
600
800
1000
1200
Res
istiv
ity
nΩ
m
Temperature oC
c1 heat c1 cool
Pt - flattenedPt/Rh - undeformed
0 200 400 600 800 1000
200
400
600
800
1000
1200
Res
istiv
ity
nΩ
m
Temperature oC
d2 heat d2 cool
Pt - undeformedPt/Rh - flattened
p
0 200 400 600 800 1000
200
400
600
800
1000
1200
Res
istiv
ity
nΩ
m
Temperature oC
a1 heat a1 cool
Pt - undeformedPt/Rh - undeformed
U – Un-flattened F – Flattened
100 µm diameter 30 µm x 250 µm
Testpiece 1 x 2 x 40 mm
750°C
722°C
702°C
In-situ calibration of thermocouples
using a phase transformation
(~ 60% cold work)
Thermocouples: Influence of straining Measurement on ETMT ( Roebuck et al, NPL March 2006)
Table 2 - Phase transformation temperatures
Pt/Pt-13%Rh U/U*
Pt/Pt-13%Rh U/F*
Pt/Pt-13%Rh F/U*
heating (1) 750 oC 702 oC 722 oC cooling (1) 655 oC 623 oC 623 oC heating(2) 750 oC 702 oC cooling (2) 655 oC 611 oC
U – Un-flattened F – Flattened100 µm diameter 30 µm x 250 µm
Eutectoid Steel . In-situ calibration of thermocouples using a phase transformation
Thermocouples: Influence of straining Measurement on ETMT ( Roebuck et al, NPL March 2006)
In-situ calibration of thermocouples
using a phase transformation
500 550 600 650 700 750 800 850 900600
650
700
750
800
850
900
950
1000
1050
1100
1150
1200
Res
istiv
ity
nΩ
m
Temperature oC
f1 heat and cool d2 heat d2 cool
Pt - undeformedPt/Rh - flattened
Good repeatability inCold worked thermocouple
Inhomogeneity
Inhomogenity of SeebeckCoefficient
Rig with traversing furnace todetermine variation of
response along the length of the wire.
Zvizdic,D et al. “Estimation of Uncertainties in Comparison Calibrationof Thermocouples” TMCSI 7, 2003, 529.
Stability of Thermocouples.Pt-13%RhPt in oxidising atmosphere.
In oxidising conditions at high temperatures above about 1050°C calibration changes which occur in platinum metal thermocouples are associated with selective volatilisation of the metal oxides. Also Rhodium oxide can exist as crystalline form from prolonged exposure at 400-600°C
Darling, A S and Selman, G L “Some Aspects of Environment on thePerformance of Noble Metal Thermocouples” TMCSI,4,Part 3, 1633-1644.
Recommendations by Quinn on construction of thermocouples.
• Use high quality alumina.• Anneal wire for 1 hour at 1250°C in air.• Alumina tube heated at 1200°C in air.• Alumina tube should be sufficiently long
so no gaps between successive sections. • Avoid contamination with bare hands.
Heywood, A.E. “Thermocouples and resistance Thermometers”Proc. Comf. On Temperature Measurement and Control,The Chameleon Press Ltd; 22-29 (1979)
100 days at 1,400°C
Stability of Thermocouples.Pt-13%RhPt in reducing atmosphere.
• Under inert or reducing conditions severe reactions can occur between the thermocouple limbs and refractory insulants.
• Showed that alumina; zirconia and thoria can dissociate to give the respective metal which reacts with the platinum to form a low melting point phase.
• Magnesia is very resistant to these effects.Darling, A S and Selman, G L “Some Aspects of Environment on the Performance of Noble Metal Thermocouples”
TMCSI,4,Part 3, 1633-1644
• Impurities in alumina such as Fe and Si (often as silica) can alloy with Pt and effect output.
• Demonstrated higher drifts (about a factor 2) with thinner wires 0.33 mm versus 0.5 mm.
Reported in Glawe, G L and Szaniszlo *Long Term Drift of some noble and Refractory’Metal Thermocouples at 1600 K in Air; Argon and Vacuum” TMCSI,4,Part 3, 1645-1662.
Thermocouples: Long term drift
‘Calibration Drift of PR13 Thermocouples During Long-Time Creep Testing’
S Yokoi, H Iton, T Murata, M Egashira & H Miyazaki ( 1980)
Trans ISIJ 20 (6) 243( also reproduced in Desvaux, 1982)
Note: at 600°C more drift in multi-specimen machines –
due to more handling ( cold-work) ?
~ 5°C –10°C drift may beencountered after ~ 5 years
( Diffusion of Rh across the junction ? )
Thermocouples: Long term drift
~ 5°C –10°C drift may be
encountered after ~ 5 years
( Diffusion of Rh
across the junction ? )
Compensate during test ?
Data from ERA Creep Laboratory,M.P.E. Desvaux ‘The practical realisation of
temperature measurement standardsin high temperature mechanical testing’
Chapt. 7 in Loveday, Day & Dyson, 1982.Based on H Robson (1977) ‘Temperature measuring techniques
in a large creep laboratory’ J Phys E : Sci Instruments 10 384 - 389
Concluding RemarksRemarks
• Avoid introducing cold work into thermocouples. • Compensate for drift in very long term tests ? • Calibrate with similar depth of immersion, or check for
inhomogeneities in the wire. • Consider using in-situ calibration in the testing machine
using a material with well defined phase change. • Remember the Seebeck emf is generated in the
temperature gradient.
R: Thompson
( Lord Kelvin )L: Seebeck
Useful References
M.S.Loveday, M.F.Day & B.F.DysonPublished : HMSO 1982
ISBN 0 11 480049 9
Useful References
• VDI/VDE-RICHLINIEN VDI/VDE 3511 on Temperature Measurement in Industry.
Part 1 Principles and Special Methods of Temperature Measurement
Part 2 Contact thermometersPart 3 Measuring systems and measuring quantity
processing for electrical contact thermometersPart 4 Radiation thermometersPart 4.1 Specification for radiation thermometersPart 4.2 Maintenance of the specification for radiation thermometersPart 4.3 Radiation thermometry. Standard test methods
for radiation thermometers with one wavelength rangePart 4.4 Radiation thermometry Calibration of Radiation
thermometers Part 5 Installation of thermometers.
Useful References
www.evitherm.organd Temperature>Measurement methods>Thermocouples.
http://en.wikipedia.org/wiki/Peltier-Seebeck_effectQuinn, T J “Temperature” Academic Press London. 1990, ISBN 0-12-569681-7 Chapter 6 Thermocouples.Burley,N; Powell, R; Burns,G and Scroger,M “The NICROSIL versus NISIL Thermocouple: Properties and Thermoelectric Reference Data” NBS Monograph 161, 1978.BS1041:Temperature Measurement Part 4 Robin E. Bentley “Thermocouples in Temperature Measurement” CSIRO (Australia) ISBN 0-9750744-4-XJ V Nicholas and D R White “Traceable Temperature: An Introduction to Temperature Measurement and Calibration”, 2001L Michalski, K Eckersdorf and J McGhee, “Temperature Measurement,” 2002
“Temperature Measurement and Calibration course”, National Physical Laboratory
Thank you for listening Thank you for listening
•• Any Questions ? Any Questions ?
HT Double SidedHT Double SidedAveraging TensileAveraging TensileExtensometerExtensometer
( INSTRON)( INSTRON)
Additional Thermocouples
There are also W; W3 and W5 known as “Tungsten-Rhenium.” Poor performance in oxidising atmosphere.
A number of others without designation types:Ir40Re; Ir Very high temperatures 2100°C in air.
Pt:Pd Under investigation for 660-1500°C. Capable of high accuracy.
Mo:Nb Used in nuclear industry because of resistance to nuclear bombardment. Not used in oxidising atmosphere.
See www.evitherm.org and Temperature>Measurement methods>Thermocouples.
Most high temperature mechanical testing uses the noble metals because of their relative stability in oxidising atmospheres; readily available and guaranteed quality of initial homogeneity.
Thermocouples: How they work
• No voltage generated at the junction• Voltage generated along the entire length of both wires by
temperature gradients– Uniform temperature: electrons have no preferred direction of
motion– Temperature gradient: electrons diffuse in a preferred
direction (either along or against the gradient)
• Any changes in structure or composition of the wires where there is a temperature gradient effects the output.
High Temperature Mechanical Testing meeting - Radiation Thermometry
Helen McEvoy
15 March 2006
Restricted Commercial
The talk will cover:
• What is radiation thermometry?
• Applications and advantages
• Calibration issues
• Disadvantages
• Practical considerations
Restricted Commercial
Concept
• All surfaces radiate thermal energy
• Radiation thermometry (radiation pyrometry)
– measures the thermal energy
– converts that measurement into a temperature
• For objects at ordinary temperatures most of the thermal radiation is at infrared wavelengths
Restricted Commercial
Advantages of radiation thermometry
• Non contact, non-intrusive– can measure remote objects, moving objects
• Large temperature range– can measure very hot objects
• Good sensitivity, Lλ∝ exp(-c2/λT)
• Can measure large areas and images
• It is fast
• Fundamental laws can be applied
Restricted Commercial
Examples:Fibre optic extrusion
Restricted Commercial
Steel processing
Restricted Commercial
Plant maintenance
Restricted Commercial
Maintenance – thermal image of an electrical connector
Restricted Commercial
Turbine blade monitoring
Restricted Commercial
Sea surface temperatures
Restricted Commercial
What is a radiation thermometer?
It has:
• an optical system (lenses, mirrors) to collect the radiation
• a detector to measure it
• often a filter to select the wavelength
• the means to relate the detector signal to the temperature
Restricted Commercial
An NPL radiation thermometer
Restricted Commercial
Fundamental principles
Planck's law for spectral radiance distribution
Or Wien's approximation to it
]1 - T)/c(exp[c = L 1-2
5-λ λλ
πελ 1
c1 and c2 are constants; ε is the emissivity; specify λ, measure Lλ, determine T
T)/c(c = L 25-
λ λλπελ −exp1
Restricted Commercial
Plot of the Planck function
Restricted Commercial
Calibration
• To relate radiation thermometer output to temperature requires well-characterised sources of radiation
– (usually) practical blackbodiesradiance closely similar to ideal Planckian radiator
• The radiation field inside a closed, isothermal cavity depends only on the temperature and is independent of the material properties of the cavity
Restricted Commercial
Cavity (blackbody) radiators
• An aperture (hole) is used to observe the radiation in the cavity
• By careful design (e.g. small aperture compared to length) the disturbance this aperture creates can be minimised
• Cavity sources are therefore ideal standard sources of radiation, governed by Planck’s law, and with an emissivity very close to 1
Restricted Commercial
Practical blackbody radiators
• Need:
– blackbody cavity of a good design (large compared to size of aperture, good wall emissivity)
– means of keeping cavity isothermal (stirred liquid bath, heat-pipe)
– means of measuring the temperature (e.g. PRT placed close to, and in good thermal contact with, back radiating surface or calibrate in terms of radiance temperature)
Restricted Commercial
The NPL blackbody radiation sources
Blackbody type Temperature range (°C)
Ammonia heatpipe -40 to +50
Stirred liquid 0 to 80
Water heatpipe 50 to 250
Caesium heatpipe 300 to 600
Sodium heatpipe 500 to 1000
High temperature 1000 to 3000
Restricted Commercial
The NPL blackbody sources (50 °C to 600 °C)
41 mm
Single zone furnace
680 mm
SPRT
Compacted ceramic fibre insulation (caesium heat pipe only)
heat pipe
Compacted ceramic fibre insulation. (caesium heat pipe only) Temperature profiling
tube ( 2 off)
Restricted Commercial
High temperature blackbody source
Restricted Commercial
The NPL fixed-point blackbody cavity
Restricted Commercial
Au point melting and freezing plateaux
5.88E-11
5.89E-11
5.90E-11
5.91E-11
5.92E-11
5.93E-11
5.94E-11
0:00 0:30 1:00 1:30 2:00 2:30time hrs:min
Iph / A
Restricted Commercial
Calibration uncertainties
For a primary standard instrument (includes characterisation, stability, calibration):
< 0.1 °C at 1000 °C increasing to ~1.5 °C at 3000 °C
For a customer instrument (best measurement capability; depends on performance of instrument):
0.2 °C (–40 to 200 °C); 0.3 °C (200 to 500 °C);0.4 °C (500 to 1000 °C);0.05 % of Celsius temperature (1000 °C to 3000 °C)
Restricted Commercial
Disadvantages of radiation thermometry
• A clear line-of-sight is needed
• Radiation may be:– absorbed in the path– scattered into or out of the line of sight– reflected from the object surface or from surrounding
objects
• Radiation emitted depends on the properties and condition of the material’s surface – its ‘emissivity’
• For good results the system must be well understood
Restricted Commercial
• Select a thermometer operating at a suitable wavelength for your application (consider required temperature range, effect of atmospheric absorption, are you viewing through a window?)
• Select a thermometer with a suitable target size for your application
• Be aware of the problems with the emissivity of your surface (more later)
• Consider whether purging of the line-of-sight might be required
Selecting a radiation thermometer
Restricted Commercial
1) Select a thermometer that operates at a suitable wavelength for your application
2) Select a thermometer with a suitable target size for your application
Selecting a radiation thermometer
Restricted Commercial
Plot of the Planck function
Restricted Commercial
Detector response
Restricted Commercial
Atmospheric transmission
Restricted Commercial
1) Select a thermometer operating at a suitable wavelength for your application (consider required temperature range, effect of atmospheric absorption, are you viewing through a window?)
2) Select a thermometer with a suitable target size for your application
Selecting a radiation thermometer
Restricted Commercial
The size-of-source effect (SSE)
• Caused by scattering and imperfections in the thermometer optics
• Instrument sees radiation from outside its nominal target size (field-of-view)
• Result - as the size of the target increases its apparent temperature (generally) also increases
• Amount depends on quality of optics and how manufacturer defines target size
• Effect can be large for poor quality instruments
Restricted Commercial
Examples of SSE measurements
SSE comparison
40.00
50.00
60.00
70.00
80.00
90.00
100.00
0 2 4 6 8 10 12 14 16 18 20 22 24 26Aperture size (mm)
SSE
(%)
NPL primary pyrometercommercial instrument
Restricted Commercial
Example of some SSE results (source at ~800 °C)
Aperture diameter (mm)
Temperature reading (°C)
4 618.0
25 786.530 792.040 796.6
6 751.89 763.810 765.112 768.2
15 772.820 780.0
50 798.1
Restricted Commercial
So, in summary:
• For high temperature applications use a short wavelength, narrow band filter; for low temperature applications, use a long wavelength, broad band filter
• Avoid wavelengths that correspond to atmospheric absorption bands
• Use as short a wavelength as possible (uncertainties proportional to λ)
• Make sure that the stated target size of the instrument is small enough for your application
Restricted Commercial
Real surfaces and emissivity
• Radiation emitted from a surface depends on the ‘emissivity’ ελ where:
• Blackbody radiators are perfect emitters of thermal radiation and also nearly perfect absorbers of it – their emissivity is 1
)(blackbodyL(surface)/L = λλλε
Restricted Commercial
• Highly polished surfaces are nearly perfect reflectors and radiate very little energy – their emissivity is 0
• For most surfaces 0 < ε < 1
• Emissivity is a material property which depends on:
temperature, wavelength, direction of view and surface conditions
Real surfaces and emissivity
Restricted Commercial
Explaining emissivity
Restricted Commercial
The variation of emissivity with temperature
Restricted Commercial
The variation of emissivity with wavelength
Restricted Commercial
Effect of emissivity
• A radiation thermometer can only respond to radiation it receives
• It will therefore give different answers when viewing surfaces with different emissivities
• For real surfaces the Planck law must be modified:
]1 - T)/c([c= L 1-2
5- λλπ
ελλ exp1
For accurate temperatures the emissivity must be known or measured
Restricted Commercial
Coping with Emissivity
• Estimate and compensate
• Enhance emissivity to ~1
• Paint surface with known emissivity substance
• Ratio two spectral radiances
• Measure emissivity by a secondary technique
• Live with it: if reproducibility is the main concern
Restricted Commercial
Ratio or two colour thermometers
• Principle - take the ratio of the thermometer signals at two wavelengths
• Emissivity term cancels out, can calculate T
BUT:
• The emissivity at the two wavelengths must be the same, or the ratio must be known and remain constant. This assumption is usually not valid.
• Some sensitivity is lost
Restricted Commercial
The variation of emissivity with wavelength
Restricted Commercial
Ratio or two colour thermometers
Generally, uncertainties are difficult to evaluate due to the emissivity uncertainty and such instruments should only be used with caution
However,
•allows some compensation for absorption in the line-of-sight
•Useful if source size is smaller than the field-of-view
Restricted Commercial
Summary
• Introduced fundamental principles
• Described applications and advantages
• Described fundamental disadvantages– and how these might be overcome
• Outlined practical radiation thermometry
Restricted Commercial
Any questions?
Recent developments in applied non-contact thermometry at NPL
A Levick, M Broussely and G Edwards
Thermal MetrologyNational Physical Laboratory
United Kingdom
Non-contact thermometryTechniques
• Photothermal radiometry for temperature and thermal properties measurement including Laser Absorption Radiation Thermometry (LART)
• Radiation thermometry employing background correction algorithms to compensate for reflected radiation, so called “RTHAT”
• Microwave radiometry for non-invasive internal temperature profiling of human tissue or foods
• Multi-wavelength pyrometry – does it work?
Laser Absorption Radiation Thermometry (LART)
Photothermal Radiometry for Temperature Measurement
LART is independent of:
•Emissivity•Reflected radiation•Gaseous absorptions•Laser intensity profile
At least in principle
Photothermal Radiometry
Wavelength λ1
Diode laser
Wavelength λ2
InGaAsphotodiode
Temperature
Hot Sample
Laser
Detector
2
1S λωλ
1
2S λωλ
Temperature
( )( )
2
1
1
2
1
2
,,LART
S L TC
S L T
λωλλ
ωλ
λλ
′= ⋅
′
2
1
1
2
SS
λωλλ
ωλLARTC ( )
( )1
2
,,
L TL T
λλ
′′
2
0 2 1
1 1626
1
CTe λ λλ
λ
⎛ ⎞−⎜ ⎟
⎝ ⎠= ⋅
Temperature
( )( )
2
1
1
2
1
2
,,LART
S L TC
S L T
λωλλ
ωλ
λλ
′= ⋅
′LARTC
T
LART1
The modulated thermal radiance at λ1 due to photothermal heating induced by a laser at λ2 is: -
1
2
11 1
( , ). . .L TS C TT
λω λ
λε ∂= ∆
∂
Measure two modulated thermal radiances at λ1 and λ2 due to lasers at λ2 and λ1, and take the ratio:-
1
2
2
1
1 2
2 1
'( , ) ( ). .'( , ) ( )LART
S L T PCS L T P
λω λλ
ω λ
λ λλ λ
= No emissivity terms!LEFT1
2 2 2 2. ( ). ( )T C Pα λ λ∆ =where
Kirchhoff’s law ( ):-
1
2
11 2 1 1 2 2 2
( , ). . ( ). ( ). . ( )L TS C C PT
λω λ
λε λ ε λ λ∂=
∂
ε α=
LART2
Measure two modulated thermal radiances at λ3 due to lasers at λ2 and λ1
No emissivity terms! – but still sensitive to reflected background radiation
Also, measure two dc thermal radiances at λ2 and λ1
3 2
1
3 1
2
2 1
1 2
. ( , ) ( ). .. ( , ) ( )LART
S S L T PCS S L T P
λ λω λλ λ
ω λ
λ λλ λ
=
Take the quantity:
LART Instrument Overview
LART Capabilities
•Measurement of temperature and thermal properties
•Single instrument
•Transportable
•Non-invasive
•Flexible
•Independent of emissivity, reflected radiation, absorption
Photograph of Instrument
LART Instrument Specifications
• Temperature range: 500°C and above
• Measurement time: 1 second
• Operates using thermal radiation laser wavelengths of 840 nm, 970 nm,
1320 nm, 1550 nm and 1970 nm
• Laser powers: 1 Watt
• Target materials include opaque solids and liquids; some semi-transparent materials and moving targets
• Target distance: 200mm – 2000mm
• Transportable in an a small car
Temperature – Results
-20
-15
-10
-5
0
5
10
15
20
0 5 10 15
InconelPlatinum
Dev
iatio
n fro
m T
herm
ocou
ple
Rea
ding
(Cel
sius
)
Point
847 oC 940 oC 1033 oCInconel vsplatinum
• RF Levitated metal drops
• Laser-heated aerodynamic levitated particles
• Silicon wafers in a Chemical Vapour Deposition (CVD) apparatus
LART thermometry -Fieldwork
Having demonstrated the LART technique in the laboratory, the next stage is to investigate “real” field applications. So far investigated:-
How to we verify the temperature readings?There are no other suitable thermometry methods
for cross-comparison
LART fieldwork - Electromagnetically Levitated drop
•Metals – liquid or solid
•RF heats and levitates
•Contactless/containerless
•Inert or reducing purge gas
•Measure various properties
•Radiation thermometry•Emissivity problem
•Work by Brooks et al, NPL UK and Egry et al at DLR Germany
LART Fieldwork - RF Levitated drop
LART – levitated iron drop Results
Levitated drop – Solid Temperature gradients
Drop
LART
Two colourpyrometer
TemperatureD
ista
nce
Solid
Levitated drop – Liquid No temperature gradients
Drop
LART
Two colourpyrometer
TemperatureD
ista
nce
Liquid
Convection
LART – Levitated iron drop Results
Temperature profile on the drop surface
Solid state (1260 C)
Liquid state (1538 C)
LART – Considerations
Weakness of photothermal signals is an issue for field work – Ideally need higher power lasers
Difficult finding alternative thermometry methods for cross-comparison
Temperature range limited by laser wavelengths available
Expensive due to high cost of high power lasers
Eye hazards associated with high power lasers
Conclusions
Technique Immune to problems associated with radiation thermometry
In-situ multi-property measurements
Instrument – Use of fibre optics
NowLART proven in the laboratory
LART demonstrated on some field applications
Promising results for multi-properties
Next… Single fibre-optic transportable instrument
Further field LART measurements
Tests on well-characterised reference targets
Commercialisation – cheap high power lasers?
Radiation Thermometry
Radiation Thermometry in HighAmbient Temperatures (RTHAT)
RTHAT is independent of:
•Emissivity•Reflected radiation
At least in principle
RTHAT apparatus
M
SPyrometer 1
Pyrometer 2
Background at TB
Target surface at TS
fibre-coupled pyrometer 1
fibre-coupled pyrometer 2
Incandescent lamp
concave mirror
Furnace & targetCorrelation: S1=a•S2+Starget
Ripple technique - background radiation subtraction from measured signal
0
0.5
1
1.5
2
2.5
3
0 20 40 60 80 100 120Time
Rad
iatio
n / a
rb.
Background (S2) Sample signalSample+reflection (S1) Recovered
Pyrometer 1
Pyrometer 2
a=0.6094 (0.6)Ssample=0.9797 R2=0.9865
Hot target with reflected radiation
Hot target with no reflected radiation
Experimental procedure:
Apply least
squares fit
Surface radiance (em itted + re flected)
0.00
0.02
0.04
0.06
0.08
0.10
0.12
0 0.05 0.1 0.15 0.2
t / s
Si/ V
Background Radiance
0.0
0.5
1.0
1.5
2.0
2.5
0 0.05 0.1 0.15 0.2
t / s
Mi /
V
Correlation between background radiance and surface radiance.
0.00
0.02
0.04
0.06
0.08
0.10
0 0.5 1 1.5 2 2.5
Mi / V
Si /
V
. ( )i iS L T r Mε= + ⋅
Results for temporal RTHAT
1r ε= −
Therefore get T and ε
Results for temporal RTHAT
Least squares background correction method• Inconel target
• Diffuse reflector• High emissivity• ∆T < 1%
Platinum target• Specular reflector• Low emissivity• ∆T <10%• Better for high T
0
10
20
30
40
50
60
70
80
90
100
-20 -10 0 10 20 30 40 50
The ta / °
T(FI
T)-T
(DC
) / K T=789K
T=882K
T=976K
T=1103K
-3
-2
-1
0
1
2
3
4
5
6
7
-20 -10 0 10 20 30 40 50
The ta / °
T(FI
T)-T
(DC
) / K T=789K
T=882K
T=976K
T=1103K
Radiation Thermometry
Multi-wavelength Pyrometry
Claimed to be independent of emissivity!
Or is it?
Multi-wavelength Pyrometry
Measure thermal radiation signals at m different wavelengths:
15
2
1( , )exp( ) 1
cL T cT
λλ
λ
=−
nε
where( , )n n nS L Tε λ=
= emissivity
We have m equations and m+1 unknownsTherefore introduce an “emissivity relationship”, e.g.
11
1
mi
n i ni
cε λ−
−
=
= ∑ We now have m equations and m unknowns
However……
Multi-wavelength PyrometryHowever…..
• The “emissivity relationship” is not very valid and may lead to large errors in the estimated temperature
• The set of equations is often “ill-conditioned”. Thus, small errors in the measured signals lead to large errors in measured temperature value.
• Increasing the number of wavelengths does not reduce the errors
2
1 1 ln nT T cλ
λ ε= − extrapolating to λ=0
Multiwavelength PyrometryTungsten target – errors due to emissivity relationship
Two wavelengths:Delta T = 260 CLambda1 = 1.000e-006Lambda2 = 2.000e-006
Three wavelengths:Delta T = 89.23 CLambda1 = 1.000e-006Lambda2 = 1.500e-006Lambda3 = 2.000e-006
Five wavelengths:Delta T = -123.67Lambda1 = 1.000e-006Lambda2 = 1.270e-006Lambda3 = 1.500e-006Lambda4 = 1.800e-006Lambda5 = 2.000e-006 Contour plot of versus T and c1 – note the valley!2χ
Conclusions
• LART is independent of emissivity, reflected radiation and gaseous absorptions
• RTHAT is independent of reflected radiation and possibly emissivity
• Multi-wavelength pyrometry can measure tagets that underfill the field of view and possibly immume to gaseous absorptions. It is NOT immune to emissivity
However all the techniques in their current form only work with a limited number of applications
Thermal Measurement Awareness Network, NPL, 15 March 2006
Issues of Dynamic Temperature Measurement and Control
in Thermo-Mechanical Fatigue Testing
Peter HähnerEuropean Commission, Joint Research Centre, Institute for Energy
NL-1755 ZG Petten, The Netherlands
Klaus Rau, Tilmann BeckUniversität Karlsruhe, Institut für Werkstoffkunde I,
D-76128 Karlsruhe, Germany
Thermal Measurement Awareness Network, NPL, 15 March 2006
Thermo-Mechanical Fatigue (TMF):
Design & residual life analysis of safety critical components
exposed simultaneously to thermal & mechanical loads
(gas turbines, aero engines, automotive & process industries):
Internal thermal constraints of representative volume element
simulated by uniform temperature and mechanical strain fields
cyclically imposed on a specimen with uniform gauge section.
Thermal Measurement Awareness Network, NPL, 15 March 2006
temperature: T = T0 + δT F(ωt) , mech. strain: εm = εm,0 + δε F(ωt − ϕ)
phase angle ϕ , F(x) = F(x + 2π) , ∀ x
Thermal Measurement Awareness Network, NPL, 15 March 2006
0 50 100 150 200-0.8
-0.7
-0.6
-0.5
-0.4
-0.3
-0.2
-0.1
0.0
Mec
hani
cal S
trai
n [%
]
Time [sec]
350
450
550
650
750
850
950
Tem
pera
ture
[o C
]
0 50 100 150 2000.0
0.2
0.4
0.6
0.8
1.0
1.2
Mec
hani
cal S
trai
n [%
]
Time [sec]
350
450
550
650
750
850
950
Tem
pera
ture
[o C
]
Triangular TWF waveforms (without dwell):
in phase ϕ = 0, Rε = 0
out-of-phase ϕ = 180o, Rε = − ∞
mech. strain ratio:
Rε = εm,min/εm,max
Thermal Measurement Awareness Network, NPL, 15 March 2006
-800
-600
-400
-200
0
200
400
600
800
-0,4 -0,3 -0,2 -0,1 0 0,1 0,2 0,3 0,4
Mechanical strain %
Stre
ss M
Pa
OP cycleIP cycle
400°C
900°C
650°C
650°C
400°C
900°C
Mechanical Response ?
Hysteresis loops:
Number of TMF Cycles to Failure ?
Out-of-phase
In-phase
εm = εel + εin
Thermal Measurement Awareness Network, NPL, 15 March 2006
• Load train alignment• Precision extensometry
LCF Thermalfatigue
Load-contr.TMF
Strain-contr.TMF
Dynamic temperature measurement & control
Therm. – mech. load phasing
• Pre-cycling, start-up procedures• Thermal strain compensation
εtot = εm + εth
Thermal Measurement Awareness Network, NPL, 15 March 2006
Outline
Introduction
TMF-Standard Project
Dynamic T Measurement & Control• heating method• specimen geometry• type of thermocouple (TC), methods of TC attachment• pyrometry Recommendations and Concluding Remarks
Thermal Measurement Awareness Network, NPL, 15 March 2006
TMF-Standard GRD2-2000-30014Thermo-mechanical fatigue – the route to standardisation(funded within the GROWTH Programme of the Europ. Comm.)
Objectives: to establish a TMF testing platform in Europeto issue a validated code-of-practiceto disseminate and exploit the results
Starting date: 1. October 2001Duration: 48 monthsBudget: ~ 1.800.000 Euro (~ 1.080.000 EC contribution)Consortium: 10 principal contractors
9 assistant contractors, 1 external participantNo. of tests: ~ 400 incl. 120 validation tests
Thermal Measurement Awareness Network, NPL, 15 March 2006
Three distinct phases of the TMF-Standard project:
1. Preparatory phase:• Inventory of TMF testing procedures in Europe (WP1)• Material procurement, characterization & distribution (WP2)
2. R&D phase:• Pre-normative testing (WP3, Task 1 – 5)• Validation testing (WP4)• Statistical sensitivity analysis (WP5)
3. Exploitation and dissemination phase:• Drafting of Code-of-Practice (WP6)• Dissemination of CoP, workshop (WP7)
Thermal Measurement Awareness Network, NPL, 15 March 2006
WP 3: Pre-normative research on TMF procedures & tolerances
Task 1: Pre-cycling, start, interrupt, restart procedures
Task 2: Dynamic T measurement and control
Task 3: Thermal strain comp., deviations from nominal T
Task 4: T gradient effects in 3 different sample geometries(solid cylindrical, hollow cyl., solid rectangular)
Task 5: Deviations from nominal phase angle
Reference testing as preparation for WP4
Thermal Measurement Awareness Network, NPL, 15 March 2006
Challenges in temperature measurement and control for TMF:
Temperature rates (heating & cooling) up to 50°C/s to be controlled with sufficient accuracy (± 5°C or 1% of ∆T)
-> specimen design, heating system, fast temperature control system
T measurement must not affect crack initiation and TMF life
-> Avoid any microstructural damage, e.g. by thermocouple attachment
Heating method (induction, radiation, ...) must not influence temperature readout in the absence of thermal equilibrium
-> no cold spot at thermocouple, nor direct over-heating of thermocouple
Long-term stability of dynamic temperature fields must be ensured
-> no drift of T profile
Repeatable & reproducible results with desired accuracy of ± 5°C
Thermal Measurement Awareness Network, NPL, 15 March 2006
Issues affecting dynamic T measurement and control:
• heating method
• specimen geometry
• type of thermocouple (TC)
• method of TC attachment
All these issues are interrelated and cannot be optimized independently !
Thermal Measurement Awareness Network, NPL, 15 March 2006
Heating methods:
Resistance furnace: too slow
Direct ohmic heating: practicable for miniaturized, e.g. thin-walled tubular, specimens only
radiation (bulb) furnace: appropriate, if
- excessive radial temperature gradients are avoided
- issue of reflectivity differences: specimen – TC is addressed
Induction heating: most commonly used in TMF
- skin effect to be kept sufficiently low
- cold spot to be avoided by proper TC type and attachment method
Thermal Measurement Awareness Network, NPL, 15 March 2006
Radiation heating:• surface heating: use of thin-walled tubular specimens• ensure good thermal contact of TC
(tends to remain colder due to its higher reflectivity)
Set-up of Centre des Materiaux / Ecole des Mines de Paris:
Thermal Measurement Awareness Network, NPL, 15 March 2006
Induction heating:
• heating of subsurface layer: Skin effect d = (4πσµ0ν)−1/2
• ensure very good thermal contact of TC (not heated directly)
TC spot-welded outside GL Ribbon TC wrapped around centre
Water coolingfor T profile stability
Thermal Measurement Awareness Network, NPL, 15 March 2006
Measurement by spot welded thermocouples in the gauge length:+ Good thermal contact+ Direct measurement in the gauge length− Damage (crack initiation) at specimen surface possible? Formation of „cold spots“ due to thermal conduction in
thermocouple wires, in particular, for Pt-based TCs:
? Which configuration of welding gives best accuracy:separately welded TC wires!
κ heat conductivityρ densityc specific heat
TCTCTCTCcycle rc
tTT
21 ρκ
κπδ
&≈
Thermal Measurement Awareness Network, NPL, 15 March 2006
Measurement by ribbon thermocouples in the gauge length:
+ Direct measurement in the gauge length
+ No significant damage in the gauge length
+ No formation of „cold spots“ by thermal conduction
? Thermal contact to specimen surface may be insufficient and prone
to degradation by oxidation, surface rippling, micro-cracking…T measurement point
Cylindrical specimen Flat specimen
Thermal Measurement Awareness Network, NPL, 15 March 2006
Measurement by spot welded thermocouples outside the gauge length:+ Good thermal contact+ No damage of test piece within the gauge section− Additional temperature calibration necessary
? Formation of „cold spots“ due to thermal conduction in thermocouple wires, in particular, for Pt-based TCs
? Configuration of welding: wires to be welded separately !
Thermal Measurement Awareness Network, NPL, 15 March 2006
Non-Contacting Measurement by pyrometry:
+ Direct measurement in the gauge length
+ Non-contacting, hence no damage in the gauge length
+ No formation of „cold spots“ by thermal conduction
? Temperature drift due to oxidation (One Colour Pyrometry)
? Interference effects due to oxide scale growth (Two Colour Pyrometry)
Thermal Measurement Awareness Network, NPL, 15 March 2006
Overview of tests performed at IWK
TC-types: R(Pt-PtRh), S(Pt-PtRh), K(Ni-CrNi), N(NiCrSi-NiSi)
TC-designs: - (Coaxial (type R, d = 1 mm))
- Ribbon type (type S, as developed by MTU)
- Spot welded (all types with d = 0.25 mm, type K additionally with d = 0.5mm)
Spot welded TCs with different contacting lengths to the specimen surfaceby wrapping the TC wires circumferentially around the gauge length
(Coaxial TCs placed in holes with different depths)
Thermocouple position inside and outside the gauge length
Pyrometry: One-Colour (IWK) and Two-Colour (NPL)
Thermal Measurement Awareness Network, NPL, 15 March 2006
Test Setup & Calibration at IWK
Optimization of induction coil
and air-cooling
- dT/dt up to 50°C/s
- Axial and circumferential
gradients below +/− 5°C
Tmax = 850°C, Tmin = 400°C
Inconel 617
One type R-TC every 5 mm
T1
T5T4
T3T2
X: Thermocouple locations
X X
X
Thermal Measurement Awareness Network, NPL, 15 March 2006
Ribbon and Spot-Welded Thermocouples
lw<
1mm
∅
d
Spot-WeldedRibbon
Thermal Measurement Awareness Network, NPL, 15 March 2006
0 10 20 30 40 50 60 70 80 90 100200
300
400
500
600
700
800
900dT/dt = 10°C/s
Spotweld, 90°-wrapping Ribbon-Type Spotweld, non-wrapping
T [°
C]
t [s]30 35 40 45 50 55 60
750
800
850
900 Spotweld, 90°-wrapping Ribbon-Type Spotweld, non-wrapping
dT/dt = 10°C/s
T [°
C]
t [s]
-> Using spot welded TCs the development of „cold spots“ at the welding point is avoided by wrapping the TC wires along at least 90° of the circumference of the specimen.
Ribbon and Spot-Welded Thermocouples(Pt based)
lw
∅
d
Thermal Measurement Awareness Network, NPL, 15 March 2006
Comparison SW vs ribbon TC (Pt based)dT/dt [K/s] SW (non-wrapping) SW (90° wrapping)
Tmax - 14 K (1.6% of Tmax) - 3 K (0.4% of Tmax) 2
Tmin - 9 K (2.2% of Τmin) - 3 K (0.8% of Tmin)
Tmax - 21 K (2.5% of Tmax) -3 K (0.4% of Tmax) 5
Tmin - 7 K (1.8% of Tmin) -2 K (0.5% of Tmin)
Tmax - 24 K (2.8% of Tmax) - 1 K (0.1% of Tmax) 10
Tmin - 8 K (2.0% of Tmin) - 1 K (0.3% of Tmin)
Tmax - 46 K (6.4% of Tmax) - 6 K (0.7% of Tmax) 50
Tmin - 11 K (2.8% of Tmin) - 1 K (0.3% of Tmin)
-> Spot welded TCs must have a sufficiently high contacting length to the specimen to give accurate results
-> Using ribbon TCs, sufficient thermal contact to the test piece must be ensured in presence of oxidation and/or surface rippling by plastic deformation
-> For Ni-based TCs the cold-spot effect tends to be less pronounced
Reference:Ribbon TC
Thermal Measurement Awareness Network, NPL, 15 March 2006
GROWTH Project n° GRD2-2000-30014
TMF-Standard
T-control using thermocouples outside the gauge length
? Possible accuracy? Achievable heating and cooling rates? Accuracy of thermocouple location and specimen position in test rig? Best practice procedure
Thermal Measurement Awareness Network, NPL, 15 March 2006
GROWTH Project n° GRD2-2000-30014
TMF-Standard
0 50 100 150 200 250 300 350 400200
300
400
500
600
700
800
900
ribbon type TC in the gage section controll: spot welded TC on the
specimen shoulder
T [
°C]
t [s]
dT/dt = 5°C/s desired T-path in the gauge section
0 20 40 60 80 100 120 140 160 180 200200
300
400
500
600
700
800
900dT/dt = 10°C/s
controll: spot welded TC on the specimen shoulder
ribbon type TC in the gage section
T [
°C]
t [s]
desired T-path in the gauge section
T-control using thermocouples outside the gauge length (specimen shoulder)
-> Using appropriate methods for the optimisation of the set point signal, temperature control using a TC located at the specimen shoulder is possible for dT/dt up to 5°C/s (or even 10°C/s for control TC sufficiently close to GL)
-> The accuracies of the TC position within the specimen and of the specimen position with respect to the induction coil have to be respected within +/- 0.5 mm
-> Long term stability of T field has to be assured by appropriate cooling
Thermal Measurement Awareness Network, NPL, 15 March 2006
Temperature Measurement by Pyrometry
0 10 20 30 40 50 60 70 80 90 100200
300
400
500
600
700
800
900dt/dT = 10 K/s
IN 617,preoxidised for 2h @ 850°C
increasing N = 1 ... 50
T [°
C]
t [s]
One-Colour (IWK)
0 5000 10000 15000 20000500
600
700
800
900
1000
Tem
pera
ture
oC
Time s
Thermocouple Pyrometer
IN718
Two-Colour (NPL)
-> One colour pyrometry: stable emissivity of surface must be ensured
-> Two-colour pyrometry: oscillations due to interferences must be avoided
These requirements will exclude the pyrometric T control in most cases !
Thermal Measurement Awareness Network, NPL, 15 March 2006
Further Recommendations• To reduce the effects of cold spots on the specimen, the diameter of the spot-
welded and ribbon TC wires should not exceed 0.5mm.• K (Ni-CrNi) and N (NiCr-NiAl) type TCs are recommended for Tmax up to
850°C because of their higher thermo-voltage and lower thermal diffusivity.Above 900°C type R (Pt-PtRh) or S (Pt-PtRh) TCs are recommended.
• Thermocouple wires should be spot welded separately with parallel non contacting wires.
• Observe heater output and machine stroke to make sure that there is no long term drift of the T profile.
• More comprehensive set of recommendations can be found in the TMF-Standard Code-of-Practice to be published in Int. J. Fatigue,or upon request to [email protected]
Thermal Measurement Awareness Network, NPL, 15 March 2006
Final Remark
At present there is no generally accepted, optimized method of dynamic
temperature measurement & control:
- awareness of critical issues is most important;
- use of complementary means is necessary to ensure accuracy of method;
However, what does „accuracy“ mean in the absence of appropriate methods of
Dynamic Temperature Calibration
? ? ?