82
ADMA-OPCO On-Site-Training Instrumentation & Control ADMA-OPCO ON-SITE-TRAINING PROGRAMME TECHNICAL COURSES INSTRUMENTATION AND CONTROL Temperature Measurement For Existing Instrument Technicians IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 1

Temp. Measurement

Embed Size (px)

DESCRIPTION

Temperature Measurement Course

Citation preview

Page 1: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

ADMA-OPCO ON-SITE-TRAINING PROGRAMME

TECHNICAL COURSES

INSTRUMENTATION AND CONTROL

Temperature Measurement

ForExisting Instrument Technicians

AndSenior Instrument Technicians

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 1

Page 2: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Temperature MeasurementShort course details

This course is designed for ADMA-OPCO existing instrument technicians, to provide hands on experience related to temperature measurement methods.

Objective

ContentThis course reviews types of temperature measurements used in ADMA-OPCO oil and gas facilities, Selection, Principle of operations, Maintenance and troubleshooting.

Topics

Temperature sensing elements.

Local and remote temperature control.

Pre-alarm and trip/ alarm switch-settings.

Calibration of temp transmitters using standard temp charts

Bench services of temperature instruments.

Troubleshooting and maintenance.

Other

Audience :Prerequisites :Location :Duration :Format :

Instrument technicians & senior technicians.English comprehension and communication.ADMA-OPCO VTC, DAS Island4 Days.Lecture, discussion and workshop practice.

Technical Contact IHRDC Job Advisor: Tel # 68030, Fax # 68033

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 2

Page 3: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

SAFETY REQUIREMENTS

1.GENERAL

Participant must become thoroughly familiar with the following safety requirements and first aid procedures, and must observe the safety requirements at all times. Maximum safety of personnel is of primary importance, followed closely by protection of equipment from damage. Careful observation of these safety requirements will minimize hazards or injury to personnel and equipment.

There are three types of Safety Requirements:

Warning, Cautions, and Notes, which are intended to emphasize critical information. Safety Requirements also include procedures to be observed in the event of certain operating malfunctions and important precautions to be observed when personnel are working in a special environment (such as in an explosive atmosphere) or with a special substance.

Warnings, Cautions, and Notes are listed in order of significance as follows:

WARNING

A WARNING points out a procedure, practice, condition, or precaution which, if not heeded, could result in personal injury or loss of life.

CAUTION

A CAUTION points out a precaution which, if not observed, could result in damage or destruction of equipment.

NOTE

A Note highlights information necessary to understand or follow a procedure, practice, condition, or description.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 3

Page 4: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

2. COURSE SAFETY REQUIREMENTS

Participant has to use the following safety precautions during this course:

-Coverall. -Safety helmet. -Safety shoes/boots. -Leather gloves.

COURSE CONTENTS

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 4

Page 5: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

1. Objectives P. 06

2. Course Outline P. 07-08

3. Equipment / Resources P. 09

4. Course Manual (Handout for the participant)

P. 10-54

5. Training aids P. 55

6. Lesson plans P. 56-59

7. Course Final Test P. 60

Attachments:1) Manufacturer’s instruction

Manuals for TemperatureTransmitter-Pneumatic

2) Manufacturer’s instructionManuals for TemperatureTransmitter-Electronic

3) Manufacturer’s instructionManuals for temperatureSwitch

Objectives

Upon completion of this course the participant

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 5

Page 6: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

will be able to do the following:

1. Describe the temperature scales and conversions

2. Describe the sensing elements and thermowells

3. Perform the response check of RTDs

4. Perform the response check on Thermocouples

5. Proper use of Conversion tables, T/c and RTD simulators

6. Calibration of a temperature transmitter

7. Calibration of a SMART temperature transmitter

8. Setting of a temperature switch

9. Trouble shoot the temperature measuring instruments

COURSE OUTLINE

- This course is designed for the instrument technicians of ADMA-OPCO working at offshore or onshore.

- Duration of this course is 4 working days (40 hr). The maximum number of participant in a batch shall be four

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 6

Page 7: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

- This course can be conducted in the class rooms and work shops of ADMA-OPCO vocational training center at DAS island.

- Course time plan: Every day 07.00 Hr to 17.00 Hr

- Total instruction-time 16 Hr

- Total work shop time: 20 Hr

- Total Assessment time: 2 Hr

- Total Tea/ snack/ Prayer: 3 Hr

Day –1

Time (Hr.) Activities / Topics Location4 Hr Introduction

Temp scalesMechanical sensorsTemperature switch

Class room

5 Hr Temperature transmitter calibrationTemperature switch calibration

workshop

0.5 Hr Assessment classroom

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 7

Page 8: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

DAY :02

Time Activities Location4 Hr Thermo EMF

Cold Junction CompensationThermocouples

Classroom

5 Hr Thermocouple Verification.Calibration of a transmitter

Workshop

0.5 Hr Assessment Workshop

DAY : 03

Time Activities Location4 Hr Wheatstone bridge

Resistance temperature detectorsWiring configurations

Classroom

5 Hr Resistance table verificationSmart TT calibration

Workshop

0.5 Hr Assessment Workshop

DAY –04

Time Activities Location4 Hr Burn out protection

Trouble shootingMeasuring lagsDerivative term of a controller

Classroom

5 Hr Verification of burnout protectionDerivative action verification.

Work shop

0.5 Hr Assessment Work shop

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 8

Page 9: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

EQUIPMENT & RESOURCES

The following test equipment and tools are required to conduct the short courses:

- Thermocouple type K, J, T . any one

- Resistance temperature detector Pt 100

- Filled system temperature indicating controller

- Temperature bath

- mV calibration

- Resistance decade box

- Temperature transmitter (RTD or TC input)

- Smart transmitter

- HART communication

- Reference tables (RTD, T/C)

- Digital multimeter

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 9

Page 10: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

4. Course Manual

HANDOUT FOR THE

COURSE PARTICIPANT

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 10

Page 11: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

TEMPERATURE MEASUREMENT

Temperature Scales

Heat is a from of stored energy. Temperature is the Measurement of intensity of heat. It is like measuring the pressure of a gas in cylinder, irrespective of the volume of the cylinder.

Temperature measurement is very important in oil and processing industries the components in the crude oil and gas may vary in composition due to the variations in temperature while they are treated in processing units.

The Machinery like pumps Compressors and equipment like the heating furnaces need to be monitored carefully on their heat generating parts in order to safe guard them from over heating there by preventing the damage of components and expensive break down.

Temperature is expressed in degree. There are few temperature scales commonly used in Industrial measurement the centigrade the Fahrenheit and the Kelvin are most popular scales.

The centigrade scale zero starts at the ice point of pure water and divided into 100 graduations at the temperature of boiling point of pure water each division is known as a degree centigrade.

The Fahrenheit scale zero starts below ice point. It is divided into 10 equal graduations in between pure water ice point and boiling point. The ice point is 32ºF and the boiling point is 212ºF.

The absolute or the Kelvin scale zero reference starts from a point which is theoretically derived, where all the particles in the matter stops moving and seizes to a stand still. It is 273.15 degrees below the ice point in centigrade scale. Hence, the ice point on a Kelvin scale is 273.15ºK and the boiling point is 373.15ºK.

Temperature value on a given scale can be converted to express on other Scales:

Deg C = (Deg. F-32) x 5/9

Deg F = (Deg. C x 95) + 32

Deg K = (Deg. C x 273.15)

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 11

Page 12: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Example:1 Convert 100ºC into Fahrenheit Scale?

ºF = ºC x 9/5 + 32 9

i e = (100 x ) + 32 = 212 ºF 5

Example: 2. Convert 122ºF into degrees centigrade? 5

ºC = (ºF-32) x 9

5

i e = (122-32) x = 50 ºC9

Example: 3 Convert -40ºC in degrees Fahrenheit?9

ºF = (ºC x ) + 325 9

i e = (-40 x ) + 32 = -40 5

Hence -40ºC = -40ºF !

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 12

Page 13: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Bimetallic Elements

Most substances expand when the temperature increases and contract when the temperature decreases, but different sub-stances expand and contract at different rates for a given material, the increase length per unit length per degree of temperature increase is called the coefficient of thermal expansion for that material. If two materials with different coefficients of thermal expansion are bonded together increase in temperature will cause the free end to bend toward the material with the lower coefficient of thermal expansion.

A bimetallic element can be formed in spiral or helix to increase the amount of motion available for a given temperature change.The spiral form of bimetallic element is convenient for housing in a circular flat case and is typically used is dial thermometers that measure ambient temperature. The helical form is well suited for housing in a narrow tube (stem) for insertion into a fluid directly or housing within a thermowell with a small bore.

Thermometers

Filled system thermometers have a bulb filled with an expanding substance, (usually an inter gas) and a dial, which is controlled by a bourbon tube. The bulb is connected by a capillary tube, which can be up to about 50 feet (15 meters) long, to the dial mechanism. Their accuracy is about the same as a bimetallic thermometer and they are much more expensive. Therefore, filled system thermometers are not usually unless remote installation of the gauge is desired. Figure shows filled system thermometers.

Schematic illustrating concept of bimetallic temperature elements

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 13

Page 14: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Schematic of a typical bimetallic dial thremometer

Bimetallic thermometers are available in convenient range increments for measurements between -80ºF (-50ºC) and 1000ºF (500ºC). A range should be chosen so that the normal operating temperature is near the centre and both the high and low temperatures of interest are covered. They are not very susceptible to damage from over-or under ranging. Dial calibrations are available in either Fahrenheit or Celsius or with both calibrations. An external adjustment screw is usually provided so that the thermometer can be calibrated at a single point, but there is usually no adjustment for span.

Sectional View of AnIndustrial Bimetallic Thermometer

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 14

Page 15: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 15

Page 16: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Electric Temperature Switches

An electric temperature switch is a device, which causes a contact to open or close with a change in temperature. Most switches can be used as either high temperature or low temperature sensors, depending on how they are calibrated and electrically connected.

Mechanically operated temperature switches are used more frequently in production facilities most mechanically operated temperature switches use a vapour-filled system or a liquid-filled system to operate pressure switch. Gas-filled systems generally do not develop enough power for switch use.

Filled system switches are available for both local and remote mounting. The local mounting type has the bulb rigidly attached to the switch mechanism and housing. The assembly has a threaded connection so that it can be screwed into and be supported by a thermowell. The remote mounting type has the bulb connection to the switch mechanism by a capillary tube from 6 feet (2 meters) to 25 feet (8 meters) or more long.

The local mounting type is less expensive to purchase and install, while the remote mounting type provides isolation of the switch from process vibration and more convenient access. The switch cannot be separated from the bulb in the field for either these designs.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 16

Page 17: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Schematics illustraating types of compensaating sys-Tems for filled system deuices

Thermowells

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 17

Page 18: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

A thermowell is a protective sheath, which protects a thermal sensor from the process fluid. Almost all temperature sensors in production facilities, with the exceptions stack temperature and pilot flame sensors are installed in thermowells. This relatively simple device will be discussed first because a thermowell is often apart of other types of devices.

Typical thermowells

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 18

Page 19: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Most thermowells used in production facilities are machined from a solid piece of material in a fabrication process generally referred to as drilled or bar-stock construction. They can also be made by cutting a piece of drown tubing, welding the construction end closed, and then welding on the process connection. This second type is generally known as welded, drawn, or tubing Drilled construction more expensive than welded construction, but is more reliable and durable. A very large majority of operating companies insist on drilled construction.

Schematics of threaded thermowells for screwed on connection

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 19

Page 20: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Insertion Length is the length that is exposed to the process fluid. It is the distance from the bottom of the threads, bottom of the socket to the tip of the thermowell. Insertion length is also known as the U-dimension.

Thread Allowance is the threaded length of the thermowell. This is usually 1 inch (25 mm). Thread allowance applies only to screwed thermowells.

Wrench Allowance is the length of the wrench flats above the threads. This dimension is usually 3/4 inch (18 mm). Wrench allowance applies only to screwed thermowells.

Lagging Extension (or log) is the length between the thread allowance and the wrench to allow the top of the well to be accessed when installed in an insulated pipe or vessel. The lagging extension is usually 3 inches (75mm) lagging extension is usually extension is usually only specified when insulation is present and applies only to screwed or socket-weld thermowells. The nozzle for mounting the flanged well will protrude through the insulation so that a lagging extension is not required. The lagging extension is also known as the "T" length.

Flanged and Van stone flanged thermowells extend above the face to allow for installation of the flange, but there is no special name for this dimension. This dimension is usually about 21/

4 inched (56 mm).

Element Length is the depth of the measured from the very top of the well to bottom of the bore. Usually, this dimension will be 1/4 inch (6 mm) less than the overall length of the well.

Thermocouples

A thermocouple is a junction of dissimilar metals used to measure temperature. When two different metals come in contact with each other, thermal energy is converted into electrical energy. Any two metals can be used and the amount of electrical energy created is a direct function of the absolute temperature except in circumstances. Also the amount of energy converted depends on the metals selected. Certain combinations of metals have been identified which create enough energy in a sufficiently liner manner so that they can be used to measure temperature with a high degree of accuracy.

Thermocouple, wires of the selected metals are joined together to make electrical contact.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 20

Page 21: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

The electrical limitations are that the junction, including any third metal, must be at the temperature to be measured, the wires must be insulated from each other from the junction to be receiver, and if the junction is grounded, there must be no other ground. The only physical limitation is that the wires must be able to stand the environment to which they are subjected.

Schematic depicting a copper-constantan thermocouple circuit with an external reference junctionConnection of thermocouples does present some difficulty be-cause when the selected metals are connected to any other metal, such as copper wire, another thermocouple is created and the temperature of this connection will affect the measurement as much as the temperature of the primary junction. Figure shows a thermocouple constructed from copper (Cu) wire and a copper-nickel alloy wire named constant an (c) connected to a voltmeter made of copper. The constant an wire must be connected an wire must be connected to copper somewhere in addition to the thermocouple to complete the circuit, but this will form another thermocouple. This connection is made so that the temperature can be held constant at a known temperature and is called the cold junction or reference junction (J2) A temperature that is easy to create and duplicate is that of a bath formed by pure water and the ice that water, 32ºF (0ºC) by holding the reference junction at the ice point, the temperature of the primary junction (J1) can be found by measuring the voltage it creates in reference to the voltage created by the reference junction.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 21

Page 22: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Tables of the voltage created at temperature versus the ice point are published by the Untied National Bureau of Standards (NBS) and are used worldwide. By measuring the thermocouple voltage, the temperature can be found from the table.

If the reference junction is located where the temperature is known or can be measured accurately, then the junction voltage for this temperature can be added to the measured voltage to find the temperature of the primary junction. All of the connections and the measurement are made to a thermally conductive, but electrical insulating material known as the isothermal block. This block is usually in the instrument case, but in large installations is sometimes done elsewhere by minimizing the thermocouple wire. If the temperature is computed circuit it is known as software compensation. If an electronic circuit is used to correct the reading, it is known as hardware compensation or an electronic ice point.

The thermocouple is connected tp the isothermal block by wire made from the same metals as the thermocouple, called thermocouple extension wire. A thermocouple extension wire is usually a shielded, twisted pair with the shield grounded at the instrument to minimize interference pickup. Terminal strips constructed of thermocouple material are available and should be used if intermediate connections are required. Only a few mili-volts are produced by a thermocouple, so careful attention to proper wiring and shielding is essential to good measurement.

Schematic depicting a hardware compensated thermocouple assembly

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 22

Page 23: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Three basic types of thermocouple assembly

Some thermocouple assemblies are manufactured so that the thermocouple makes electrical contact with the sheath (called ground junction) and some are manufactured where the thermocouple is electrically insulated from the sheath (called ungrounded junction) A third option is where the thermocouple extends slightly beyond the heath (called exposed junction) Exposed junction offer the fastest response, but are not used in oil and gas processing because they are subject to physical damage. They would need to be installed without a thermowell to take advantage of this faster response. Grounded junction offer faster response than ungrounded junctions because the contact area which provides the electrical connection also provides good thermal conduction. Also, grounding at thermocouple provides the most nearly symmetrical circuit, which reduced interference picked up by the wires to a minimum. Grounded thermocouples should be selected unless other components of the circuit require that the ground be at some other point, or the process fluid and piping are not at ground potential.grounding any measurement loop at more than one point will usually cause

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 23

Page 24: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

measurement errors because of potential difference in the grounding system. These errors are more acute with low voltage signals such as generated by thermocouples. These statements do not preclude grounding the extension wire shield at the receiver, which is recommended.

The most common and leat expensive thermocouple is iron versus constant an (ISA type J). Type can be used for measurement from -320°F (-195°C) to1400°F (760°C), but is normally limited to 32 to 1000°F (0-500°C). Type J is usually furnished when no specific type is specified

Chromal versus alumal thermocouples (ISA type K) offer better corrosion resistance. Type K can be used for-310°F (-190°C) to 2500°F (1370°C) but usually limited to 32 to 2000°F (0 to 1000°C) Type K does not produce as much out put as type

Copper versus constantan thermocouples (ISA type T) are usually used when temperatures below zero are to be measured. While the usable range for type T, 310 to 750 °F (- 190 to 400 °C), is the same for the lower limit and less for the higher limit than for types J and K, the recommended range is -290 to 700°F (-180 to 370°C). The materials used in type T behave more predictably at low temperatures than those used for types J and K.

Chromel versus constant an thermocouples (ISA type E) provide the largest voltage change per temperature change for standard thermocouples. An output of 40 millivolts at 1000°F can be compared to 30 mv for type J and 22 mv for type K. type E can be used for 320 to 1830°F (-195 to 1000°() and is recommended for 32 to 1600°F (0 to 870°F). Some sources extend this range downward to -100°F (-73 °C), but type T is generally considered a better choice for below freezing temperature. Type E has more tendency to changecharacteristics with time than types J, K and T.

These four types of thermocouples comprise the base metal thermocouples.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 24

Page 25: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Output versus temperature curves for the four types of base metalthermocouples. (Types J, K, T and E)

Other thermocouple types, called the noble metal types are available for measurements where the base metal types are not suitable. They are made from expensive metals such as platinum, rhodium, iridium and tungsten thus are more expensive. Also, they do not provide as much output as the base metal types.

These noble metal thermocouples are used in laboratories, for molten metals and other applications, but are rarely used in production facilities.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 25

Page 26: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

A nipple-union-nipple extension assembly for installing an RTD orthermocouple element into a thermowell

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 26

Page 27: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Resistance Temperature Detectors

The resistance of a conductor usually increase as the temperature increase. If the properties of that conductor are known, the temperature can be calculated from the measured resistance. A resistance temperature detector (RTD) is a conductor of known characteristics constructed for insertion into the medium for temperature measurement.

Any conductor can be used to construct an RTD, but a few have been identified as having more described characteristics than others. The characteristics which are desired include.

1. Stability in the temperature range to be measured. The material must not melt, corrode, embrittle or change electrical characteristics when subjected

to the environment in which it will operate.

2. Linearity. The resistance change with temperature should be as liner as possible over the range of interest to simplify readout.

3. High resistively. Less material is needed to manufacture an RTD with a specified resistance when the material has a high characteristic resistively.

4. Workability. The material must be suitable for configuring for insertion into the media.

The materials which have been identified as having acceptable characteristics are: copper, nickel, tungsten and platinum.

Copper has good linearity, workability, and is able up to 250°F (120°C), but has low resistively, thus either a long conductor or one with a very small cross-sectional area is required for a reasonable resistance. Nickel and nickel alloys have high resistively, good stability and good workability, but have poor linearity.Tungsten is brittle and difficult to work with. Platinum has been accepted asthe material which best fits all the criteria and has been generally accepted forindustrial measurement between -300 and 1200 °F (-150 and 650°C).

The effect of resistances inherent in the lead wires of the RTD circuit on the temperature measurement can be minimized by increasing the resistance of the sensor; however, the size of the sensor will also be increased. RTDS are commercially available with resistances from 50 to 1000 ohms at 32°F (0°C) and increase resistance 0.385 ohms for every °C of temperature rise.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 27

Page 28: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

This is called the European (E) standard and is in accordance with the DIN (Deutsche Institut fuer Normung) 43760 Standard.Chemically pure platinum has a rise of.392 ohms per °C for a 100 ohm RTD in accordance with the American (A) standard.The European standard is dominant,. Even in the United States the American standard is seldom used.

When the resistance of the RTD is found by measurement, the temperature canbe calculated:

°C = (Ohms reading -100)/0.385

The accuracy of this calculation is determined primarily by the accuracy of the reading. Modem instruments can measure resistance very accurately and the temperature can be determined precisely if the resistance of the connecting circuit is insignificant or is known. Unfortunately, this resistances usually not negligible or known for most practical circuits. The wire thats usually used (16 AWG stranded copper) has a resistance of approximately 4 ohms per 100 feet (305 m). If it is assumed that the RTD is connected to the instrument by a 625- foot cable as shown in Figure, the total resiatance will be 5 ohms larger than the RTD resistance, which will cause a 23.4 °F (13°C) error. Furthermore, copper wire has a temperature coefficient of about 0.0039 ohms /°C/ so the reading will vary about a degree for every 20° change in ambient temperature. These errors can be compensated for by measuring the resistance of every loop and keeping track of the ambient temperature, but fortunately there are better methods.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 28

Schematic two-wire RTD circuit

Page 29: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Schematic of four-wire RTD circuit

The most accurate method for connecting the RTD is with four wires as shown in Figure 8. A constant current source forces a known current through the RTD, which for this discussion will be assumed to be 2.6 milliamperes. The resistance of the wires conducting this current does not need to be known. By Ohm's law, the voltage across the RTD will be this current multiplied by the resistance of the RTD. This voltage is measured by a high-impedance voltmeter on the other set of wires.

The voltmeter will read 260 millivolts when the temperature of the RTD is 32°F (0°C) and its resistance is 100 ohms. For every 1.8 °F (1°C) of temperature rise, the resistance of the RTD will increase 0.385 ohms, which when multiplied by the 2.6 milliamperes flowing increases the voltage b 1 millivolt and the temperature in °C can be read by subtracting 260 from the reading. The resistance of the leads is not important, even in relation to each other as long as the current can be maintained and the resistances of the leads are small compared to the resistance of the voltmeter. Use of 4-wire RTD circuits is usually limited to laboratories and situations where very high accuracy is desired because less expansive 3-wire circuits almost always provide the needed accuracy.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 29

Page 30: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Schematic of Three-wire RTD circuit with a balanced bridge

A compromise connection method for RTDS that uses three wires and a balanced bridge circuit is shown above. For this circuit, Rl and R2 are selected to be the same resistance so that the voltage at the negative terminal of the voltmeter is one half o the supply voltage. R3 is selected to be the same resistance as the RTD at the base temperature, 100 ohms if 0°C is used as the base. For this circuit, it is important that wire A and wire B have the same resistance. The usual practice is to run the three wires as a shielded raid, thus they will all be the same length and the same resistance within manufacturing tolerance.

At the base condition, the positive terminal will also see one-half of the supply voltage and the reading will be zero. If 5.2 volts is used to power the bridge, the voltage will be 2.6 at each terminal of the voltmeter at the base temperature. When the temperature of the RTD is raised one degree C, the voltage reading will increase to one millivolt. The symmetry will be upset as the reading moves away from the base temperature and the one millivolt per degree will not continue to be exact, but various schemes of completion are available to give an acceptable reading.

The proceeding paragraphs are intended to explain the basis of two, three and four wire RTD connections. The selection of resistors and compensation schemes are left to the manufacturer of the instrument, but the facilities engineer selects which of the connection methods to use.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 30

Page 31: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

The three wire method is the proper selection for virtually all production facility applications.

Resistance temperature detectors (RTDS) are the most frequently used electronic temperature sensors for production facilities.

The industry has standardized on RTDS that are calibrated to Din standard 43760 which is also known as the European standard RTDS which meet this standard measure 100 ohms at 0°C, are made of platinum and exhibit a resistance increase of 0.385 ohms per °C temperature increase. Another standard, called the American Standard, is available but is not in wide use, even in the United States. Typical RTDS are shown in Figure.

Various methods for attaching an RTD or thermocouple sheath to athermowell fitting

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 31

Page 32: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

RTDS are usually purchased as a probe assembly consisting of the RTD sensor installed in a type 304 stainless steel sheath. The sheath is held in the thermowell by a fitting which is threaded on both ends for attachment to the thermowell and the head so that the tip of the sheath touches or is very near the end of the well. The preferred method of attachment of the sheath to the fitting is with a spring assembly which allows the fitting to be screwed into the thermowell as the spring is compressed. The spring holds the sheath firmly against the bottom of the well for good heat transfer. Another method is to sliver solder the sheath into the fitting which makes a good firm assembly, but requires a small clearance from the bottom of the well. The third popular method is with a comparison fitting so that the sheath can be pushed against the bottom of the well after the fitting is screwed into the well. The compression nut is then tightened to hold the sheath. The compression fittingallows use of a universal probe in different lengths of thermowells.

The head of the assembly is a chamber where the leads from the RTD and the leads to the receiver instrument can be terminated and connected to each other.

Temperature Transmitters

Temperature transmitters are used when it is necessary to convert the signal from a temperature sensor to one of the standard signals for transmission over a long distance or interface with other instruments. The signal is usually 4 to 20 ma. For electronic transmission and 3 to 15 psig (20 to 100 Kpa) for pneumatic if a transmitter is used. Other signals can be used if required by the receiver, but these are the most common and should be used if possible. It is also possible to bring a temperature measurement into a control room without using a transmitter A thermocouple RTD can be wired directly to an instrument in the control room and this is acceptable practice.

Temperature transmitters for new installations are predominantly electronic with 4 to 20 ma. Outputs and inputs from thermocouples or RTDS. These transmitters can be mounted in the field and on the thermowell or in the field on a support and connected to the sensor by a cable.

Temperature transmitter mounted in the field must be protected from the elements by an appropriate housing. A weatherproof (INEMA 4) housing is adequate for m most applications, even in Division 2 hazardous area because there are no arching contacts in a typical temperature transmitter. An explosion- proof (NEMA 7) housing is required for Division I area unless the installation is certified intrinsically safe. The energy level required in temperature transmitters is such that they can be used in intrinsically safe installations if isolated from the power supply and receiver by approved barriers and approved by an agency recognized in the country where installed.

Burn Out Protection

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 32

Page 33: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

As the temperature sensors are continuously subjected to process temperature, there is a likely chance of failure due to excess temperature or mechanical damage. In either event, the RTD or thermocouple circuit will be open, discontinuing the electrical path. This is called bum out.

It is a facility provided within the receiving instruments like the recorders, indicating, controllers, to respond for such loss of input signals.

Burn out Protection Up Scale.

When the input to the instrument is disconnected the instrument shows a range maximum value.

Burn out Protection Down Scale

When the input signal wiring to the instrument is disconnected. The instrumentshows scale minimum value.

The associated circuits with the above instrument like switches- alarm, shutdown logic fail safe option is selected in Bum Out Protection.

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 33

Heat / Temperature

Page 34: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 34

Temperature scales

Page 35: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 35

Page 36: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 36

Sensors

Page 37: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 37

Page 38: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 38

Thermo EMF

Thermocouples

Page 39: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Temp Renges of T. C

ThermocouplePair Type

PositiveInsulationBS/ANSI

NegativeInsulationBS/ANSI

UsefulRange

Iron-Constantan JIron (magnetic)

:yellowWhite

Constantan(dull) :Blue/

Red 0-800ºC

Copper-Constantan T

Copper (By Color):White/

Blue

Constantan(dull) :Blue/

Red

-200 to+400ºC

ChromelAlumel K

Chromel (shiny)Brown/Yellow

Alumel(Magnetic)

BlueRed

0-1100ºC

ChronelConstantan E

Chromel (shiny)Brown/Purple

Constantan(dull) Blue/

Red0-800ºC

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 39

Page 40: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 40

Page 41: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 41

Page 42: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 42

Page 43: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 43

Page 44: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 44

Page 45: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 45

Page 46: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 46

Page 47: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 47

Page 48: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 48

Page 49: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 49

Page 50: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 50

Page 51: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 51

Cables-color code

Page 52: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 52

Transfer log

Page 53: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

5. Training Aids

LIST OF TRANSPARENCIES

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 53

Derivative action

Page 54: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

T1 Objectives

T2 Heat-Energy transferT3 Temperature ScalesT4 Sensors

T5 Thermowells

T6 Thermocouples

T7 Temp Ranges of T.CT8 Promac CalibratorT9 Calibration hook up T10 T. C Verification T11 Bridge circuitT12 Resistance temperature Detector

T13 RTD wiring options

T14 RTD conversion tables

T15 Decade Resistance box

T16 RTD verification

T17 SMART transmitters

T18 Cold Junction

T19 Burn protection

T20 Cables-color code

T21 Transfer lag

T22 Derivative action

6. LESSON PLANS

Instructor 's guideLESSON I

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 54

Page 55: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Introduction -

Objectives -

Course out line -

Heat/ Temperature -

Temperature scalesConversion -

Sensors -

Thermowells -

Temperature bath -

Introduce your self and speak to theparticipants to gauge their language

Show transparency Tlread all the course objectives

Show transparency T2Read the course plan

Show transparency T3Explain about heat as an energyExplain that temperature is theintensity of heat

Show transparency T4Explain Ice point and Steam point.About deg.centigrade,about deg. fahrenheit,about deg.kelvin or Absolute scale.

Show transparency T5Explain Bimetallic sensorsFilled system bulbs

Show transparency T6Discuss thermo wellsscrewed, flanged, length etc.

Show transparency T7Explain how to use the equipment

Lesson 2

Thermo EMF - Show transparency T8Explain seebeck, Peltier andThomson effects

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 55

Page 56: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Thermocouples -

Conversion tables -

Test equipment -

Calibration -Hook-up

Thermocouple -Verification

Show transparency T9Explain different types of thermocouples,Temperature ranges

Show transparency Tl0Explain the use of tables

Show transparency Tl1explain the features of promaccalibrator

Show Transparency T12Discuss the methods to simulate themv signal

show Transparency T13Discuss the hook up

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 56

Page 57: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

LESSON 3

Wheatstonebridge -

RTD -

Wiring -

Conversiontable -

Decaderesistance box -

RTDverification -

Smarttransmitters -

Show transparency T14Explain the bridge balancing

Show transparency Tl5Function of an RTDExplain various symbols Pt 100, Cu 500

Show transparency Tl6Explain 2 wire, 3 wireand 4 wire connections

Show transparency Tl7Explain the use of tables

Show transparency Tl8Discuss the simulation methodExplain wiring

Show transparency T19Explain hook up method and verification

Show transparency T20Discuss smart communications

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 57

Page 58: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

Lesson 4

CJC -

Burn outProtection -

Cables -

Transfer lag -

Derivative action -

Show transparency T21Explain reference junctionDiscuss ambient temperature

Show transparency T22Explain reasons for bum outDiscuss fail safe circuitExplain up scale / down scale

Show transparency T23Explain extension leadsDiscuss compensating cables

- Show transparency T24Discuss the energy transfer

Show transparency T25Explain the need for a predictionExplain how the controlers work

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 58

Page 59: Temp. Measurement

ADMA-OPCO On-Site-Training Instrumentation & Control

7. Assessment

Q, 1. A temperature transmitter was calibrated for 0-100°C

(a) what will be the output when the process temperature is 348.15°K? (output range is 3 to 15 psig).

(b) what will be the local temperature gauge reading if the scale is in °F?

Q.2. A temperature transmitter uses type T thermocouple, write down the procedure to calibrate the instrument for a rang of -50 to +150°C at a room temperature of 25°C.

(a) Using thermocouple simulator.

(b) Using mV source and conversion tables.

Q.3. On a process line, RTD sensor resistance measured as 138.5 ohms. Thetransmitter gives an output signal of16 mA. The process temperature wasbrought down to 122°F by the operator.The transmitter gave an output of 8 mA. what is the calibrated range of thetransmitter?

IHRDC TM – I – 06 (Rev. 0) 26.12.99 Page 59