WSC highWSC highWSC high---level workshop onlevel workshop onlevel workshop onInternational Standards for Medical International Standards for Medical International Standards for Medical
TechnologiesTechnologiesTechnologies
Standards andStandards andStandards andNew TechnologyNew TechnologyNew Technology
or or or Choosing what Choosing what Choosing what you want to be you want to be you want to be
when you when you when you grow up.grow up.grow up.
Charles SidebottomFebruary 2004
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
• When should standards lead technology?– Case study – Cardiac pacemakers and mobile
telephones– Case Study – Cardiac valve prostheses
• What is necessary for success?
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Medtronic's first office building and manufacturing facility in Minneapolis, Minnesota
Medtronic is the world leader in medical technology providing lifelong solutions for people with chronic disease.
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Guiding principles• Write a standard only when there is
a clear patient safety need validated by risk assessment.
• Focus on requirements that address basic safety and essential performance.
• Ensure that the stakeholders are involved from the beginning.
Case Studies:• Cardiac Pacemakers and Mobile Telephones• Cardiac valve prostheses
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Case Study: Cardiac Pacemakersand Mobile Telephones
Typical ECG input electrodes(on opposite internal sides)
- Y
Typical DUT outputmonitoring electrodes(on opposite internal sides)
+ X
- X
+ Y
X,Y = 0,0
DUT and leadimmersed in saline bath
DUTsupport grid
AntennaAntennasupport grid
nonconducting string(20-pound test monofilimentfishing line)
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Input Output
Forwardpower
Reflectedpower
RFsignalgenerator
Powermeter A
Antennacable
Dualdirectionalcoupler
Powermeter B
ECG signalsource
Monitoringequipment
Dipoleantenna
Cutoutarea
+x
-x
-y +y
Case Study: Cardiac Pacemakersand Mobile Telephones
Torso simulator showing the position of the pulse generator and a dipole antenna simulating the mobile telephone
The test system uses a dipole antenna and worst-case modulation at mobile phone operating frequencies to test for interactions.
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Case Study: Cardiac Pacemakersand Mobile Telephones
Dipole antenna tuned to the frequency of the particular test.
Computer controlled robot arm that precisely positions and orients the dipole antenna relative to the pulse generator
Case Studies:• Cardiac Pacemakers and Mobile Telephones• Cardiac valve prostheses
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Case Study: Cardiac valve prostheses
Mechanical (rigid) valve prosthesis
Tissue (flexible) valve prosthesis
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
Stented valve prosthesis
Case Study: Cardiac valve prostheses
Unstented valve prosthesis
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
• ISO 5840– First edition published in 1984– Revised three times (1989, 1996 and 2004)– The first three editions were written using a
“traditional” requirements-based model– In recent years, this model has proved unsatisfactory
because of the speed of technological innovation
Case Study: Cardiac valve prostheses
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
• ISO 5840 (4th edition)– Risk-based model– Includes best practice methods for verification testing– Provides guidance on a risk classification system
Case Study: Cardiac valve prostheses
Standards and New TechnologyStandards and New TechnologyStandards and New Technology
• When should standardsattempt to lead technology?– When there is a clear patient safety need that can be
validated
• What is necessary for success?– A focus on requirements that address basic safety
and essential performance– Ensuring that all the key stakeholders are involved
from the beginning