38
Technical Specification FLUXUS® G809**-*1 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 1 Transmitter FLUXUS G809 Measurement with transducers mounted by Variofix C Measurement with transducers mounted by PermaFiX Ultrasonic Gas Flowmeters for Permanent Installation in Hazardous Areas Especially designed for the stationary use in explosive atmosphere Features Instrument with two measuring channels for exact and reliable flow measurement under complex flow conditions Precise bi-directional and highly dynamic flow measurement with the non-intrusive clamp-on technology High precision at fast and slow flow rates, high temperature and zero point stability Transmitter housing: - Corrosionproof and suitable for offshore application - Transmitter G809**-A1 in a flameproof housing (degree of protection IP66) - Transmitter G809**-F1 in an explosionproof housing (NEMA 4X) Certifications: - G809**-A1: ATEX/IECEx - G809**-F1: FM Cl. 1, Div. 1 The transmitters can be operated by a magnet pen without opening the housing Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results User-friendly design Communication interfaces Modbus RTU and HART available Transducers available for a wide range of inner pipe diame- ters ( ) and fluid temperatures ( ) approved transducers for hazardous areas available Measurement is unaffected by gas density, viscosity, compo- sition, dust, humidity, temperature or pressure Applications Designed for industrial use in harsh environments, in gas pro- cessing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications: Measurement on natural gas pipelines and in natural gas stor- age installations Measurement of synthesized gas and injection gas Measurement for the gas supply industry 6...2100 mm -40...+200 °C ATEX/IECEx, FM Class 1 Div. 1/Div.2

Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

Technical Specification

FLUXUS® G809**-*1

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 1

Transmitter FLUXUS G809

Measurement with transducers mounted by Variofix C

Measurement with transducers mounted by PermaFiX

Ultrasonic Gas Flowmeters for Permanent Installation in Hazardous Areas

Especially designed for the stationary use in explosive atmosphere

Features

• Instrument with two measuring channels for exact and reliable flow measurement under complex flow conditions

• Precise bi-directional and highly dynamic flow measurement with the non-intrusive clamp-on technology

• High precision at fast and slow flow rates, high temperature and zero point stability

• Transmitter housing:- Corrosionproof and suitable for offshore application- Transmitter G809**-A1 in a flameproof housing (degree of

protection IP66)- Transmitter G809**-F1 in an explosionproof housing

(NEMA 4X)

• Certifications:- G809**-A1: ATEX/IECEx- G809**-F1: FM Cl. 1, Div. 1

• The transmitters can be operated by a magnet pen without opening the housing

• Automatic loading of calibration data and transducer detection for a fast and easy set-up (less than 5 min), providing precise and long-term stable results

• User-friendly design

• Communication interfaces Modbus RTU and HART available

• Transducers available for a wide range of inner pipe diame-ters ( ) and fluid temperatures ( )

• approved transducers for hazardous areas available

• Measurement is unaffected by gas density, viscosity, compo-sition, dust, humidity, temperature or pressure

Applications

Designed for industrial use in harsh environments, in gas pro-cessing and natural gas extraction, chemical industry and in the petroleum industry. Practical applications:

• Measurement on natural gas pipelines and in natural gas stor-age installations

• Measurement of synthesized gas and injection gas

• Measurement for the gas supply industry

6...2100 mm -40...+200 °C

ATEX/IECEx, FM Class 1 Div. 1/Div.2

Page 2: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

FLUXUS® G809**-*1 Technical Specification

Table of Contents

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-212

Function ........................................................................................................................................................... 3Measurement Principle ..................................................................................................................................... 3Calculation of Volumetric Flow Rate ................................................................................................................. 3Number of Sound Paths.................................................................................................................................... 4Typical Measurement Setup ............................................................................................................................. 4Standard Volumetric Flow Rate ........................................................................................................................ 5

Flow Transmitter ............................................................................................................................................. 6Technical Data .................................................................................................................................................. 6Dimensions ....................................................................................................................................................... 9Wall and 2 " Pipe Mounting Kit........................................................................................................................ 10

Transducers................................................................................................................................................... 12Transducer Selection ...................................................................................................................................... 12Transducer Order Code .................................................................................................................................. 15Technical Data ................................................................................................................................................ 16

Transducer Mounting Fixture ...................................................................................................................... 28

Coupling Materials for Transducers............................................................................................................ 30

Damping Mats (optional) .............................................................................................................................. 31

Connection Systems..................................................................................................................................... 33Transducer Cable............................................................................................................................................ 34

Junction Box (G809**-A1)............................................................................................................................. 35Technical Data ................................................................................................................................................ 35Dimensions ..................................................................................................................................................... 352 " Pipe Mounting Kit (optional)....................................................................................................................... 36Terminal Assignment ...................................................................................................................................... 36

Extension Cable (G809**-F1) ........................................................................................................................ 37Terminal Assignment for Terminal Board KFM1 ............................................................................................. 37

Page 3: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 3

Technical Specification FLUXUS® G809**-*1

Function

Measurement Principle

In order to measure the flow of a medium in a pipe, ultrasonic signals are used, employing the transit time dif-ference principle. Ultrasonic signals are emitted by a transducer installed on the pipe and received by a sec-ond transducer. These signals are emitted alternately in the flow direction and against it.

As the medium in which the signals propagate is flowing, the transit time of the ultrasonic signals in the flowdirection is shorter than against the flow direction.

The transit time difference, ∆t, is measured and allows the flowmeter to determine the average flow velocityalong the propagation path of the ultrasonic signals. A flow profile correction is then performed in order to ob-tain the area averaged flow velocity, which is proportional to the volumetric flow rate.

Two integrated microprocessors control the entire measuring process. This allows the flowmeter to removedisturbance signals, and to check each received ultrasonic wave for its validity which reduces noise.

Calculation of Volumetric Flow Rate

= kRe . A . ka . ∆t/(2 . tfl)

where

Path of the ultrasonic signal Transit time difference ∆t

- volumetric flow ratekRe - fluid mechanics calibration factorA - cross-sectional pipe areaka - acoustical calibration factor∆t - transit time differencetfl - transit time in the medium

Page 4: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Number of Sound Paths

The number of sound paths is the number of transits of the ultrasonic signal through the medium in the pipe.Depending on the number of sound paths, the following methods of installation exist:

• reflection arrangementThe number of sound paths is even. Both of the transducers are mounted on the same side of the pipe. Correct positioning of the transducers is easier.

• diagonal arrangementThe number of sound paths is odd. Both of the transducers are mounted on opposite sides of the pipe. In the case of a high signal attenuation by the medium, pipe and coatings, diagonal arrangement with 1 sound path will be used.

The preferred method of installation depends on the application. While increasing the number of sound pathsincreases the accuracy of the measurement, signal attenuation increases as well. The optimum number ofsound paths for the parameters of the application will be determined automatically by the transmitter.

As the transducers can be mounted with the transducer mounting fixture in reflection arrangement or diagonalarrangement, the number of sound paths can be adjusted optimally for the application..

Typical Measurement Setup

a - transducer distance

Reflection arrangement, number of sound paths: 2

Diagonal arrangement, number of sound paths: 3

Diagonal arrangement, number of sound paths: 1 Diagonal arrangement, number of sound paths: 1,negative transducer distance

Example of a gas flow measurement in reflection arrangement with standard volumetric flow rate output

a

a

a > 0 a < 0

transducers

transmitter

outputs

power supply

RS485(optional)

damping mat

Page 5: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 5

Technical Specification FLUXUS® G809**-*1

Standard Volumetric Flow Rate

The standard volumetric flow rate can be selected as physical quantity to be measured. It will be calculated in-ternally by:

N = . p/pN . TN/T . 1/K

where

The operational pressure p and the operational temperature T of the medium will be entered directly as fixedvalues into the transmitter.

The gas compressibility coefficient K of the gas is entered in the transmitter:

• as fixed value or

• as approximation according to e.g. AGA8 or GERG

N - standard volumetric flow rate- operating volumetric flow rate

pN - standard pressure (absolute value)p - operating pressure (absolute value)TN - standard temperature in KT - operating temperature in KK - compressibility coefficient of the gas: ratio of the compressibility factors of the gas at operating

conditions and at standard conditions Z/ZN

V· V·

Page 6: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

6 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Flow Transmitter

Technical Data

FLUXUS G809**-A1G809**-A1A

G809**-F1

design explosion proof field device, 1 or 2 measuring channelstransducers G****81, G****LI1, G***2E85, G****83, G***LI3 G**1N62

measurementmeasurement principle transit time difference correlation principleflow velocity 0.01...35 m/s, depending on pipe diameterrepeatability 0.15 % of reading ±0.01 m/smedium all acoustically conductive gases,

e.g. nitrogen, air, oxygen, hydrogen, argon, helium, ethylene, propanetemperature compensation corresponding to the recommendations in ANSI/ASME MFC-5.1-2011accuracyvolumetric flow rate ± 1...3 % of reading ±0.01 m/s depending on application

± 0.5 % of reading ±0.01 m/s with field calibrationflow transmitterpower supply 100...240 V/50...60 Hz

or20...32 V DC

power consumption < 8 Wnumber of flow measuring channels

1, optional: 2

damping 0...100 s, adjustablemeasuring cycle (1 channel) 100...1000 Hzresponse time 1 s, option: 70 mshousing material cast aluminum, special offshore coatingdegree of protection accord-ing to IEC/EN 60529

IP66

dimensions see dimensional drawingweight 6.1 kgfixation wall mounting, 2 " pipe mountingoperating temperature -30...+60 °C (< -20 °C without operation of the display)display 2 x 16 characters, dot matrix, backlightmenu language English, German, French, Dutch, Spanish

Page 7: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 7

Technical Specification FLUXUS® G809**-*1

explosion protection

ATEX/IECEx

zone 1marking G809**-A1

Ex db e IIC T6 GbEx tb IIIC T 100 °C DbTa -40...+60 °C

G809**-A1A

Ex db e ia IIC T6 GbEx tb ia IIIC T 100 °C DbTa -40...+60 °C

-

certification ATEX IBExU11ATEX1022 X -certification IECEx IECEx IBE 11.0006X -type of protection gas: electronics compartment: flameproof enclo-

sure, connection compartment: increased safetydust: protection by enclosure

-

intrinsic safety parameters G809**-A1AUm = 250 VUi = 30 V DCIi = 56 mAPi = 0.42 WCi = 3 nF

-

FM

marking - Cl. I, II, III/Div. 1/ GP. A, B, C, D, E, F, G/For Group A, conduit seal of connection compart-ment is required within 18 inches.

Cl. I, II, III/Div. 1/ GP. B, C, D, E, F, G

T4A Ta = 60 °Cmeasuring functionsphysical quantities operating volumetric flow rate, standard volumetric flow rate, mass flow rate, flow velocitytotalizer volume, masscalculation functions average, difference, sum (2 measuring channels necessary)diagnostic functions sound speed, signal amplitude, SNR, SCNR, standard deviation of amplitudes and transit timesdata loggerloggable values all physical quantities, totalized values and diagnostic valuescapacity > 100 000 measured valuescommunicationinterface - process integration (optional): Modbus RTU or HART

- diagnosis: RS2321

serial data kit (optional)software (all Windows™ versions)

- FluxData: download of measurement data, graphical presentation,conversion to other formats (e.g. for Excel™)

- FluxDiag (optional): online diagnostics and report generation

- FluxKoef: creating medium data sets

- FluxSubstanceLoader: upload of medium data setsсable RS2321

adapter RS232 - USB1

1 connection of the interface RS232 outside of explosive atmosphere (housing cover open)

FLUXUS G809**-A1G809**-A1A

G809**-F1

II2GII2D

0637

II2GII2D

0637

Page 8: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

8 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

outputsThe outputs are galvanically isolated from the transmitter.

number G809**-A1current output: 2binary output: 2 or 4

or

current output: 0 or 1binary output: 1Modbus

or

current output: 2/HARTbinary output: 2

or

frequency output: 1binary output: 1

G809**-A1Acurrent output (intrinsic safety): 1/HART

current output: 2binary output: 2 or 4

or

current output: 0 or 1binary output: 1Modbus

or

current output: 2/HARTbinary output: 2

or

frequency output: 1binary output: 1

current outputcurrent output I1, I2- range 0/4...20 mA- accuracy 0.1 % of reading ±15 μA- active output Rext < 500 Ω- passive output Uext = 4...26.4 V, depending on Rext, Rext < 1 kΩcurrent output I1 in HART mode- range 4...20 mA- passive output Uext = 7...30 V DC- active output Uint = 24 V

current output (intrinsic safety)current output I1- range 4...20 mA -- accuracy 0.04 % of reading ±3 μA -- passive output Uext = 7...30 V, depending on Rext, Rext < 1 kΩ -current output I1 in HART mode- range 4...20 mA- passive output Uext = 7...30 V DC

frequency outputrange 0...5 kHzopen collector 24 V/4 mA

optional: 30 V/100 mA or8.2 V DIN EN 60947-5-6 (NAMUR)binary output

Reed relay 48 V/100 mAopen collector 24 V/4 mA

optional: 30 V/100 mA or8.2 V DIN EN 60947-5-6 (NAMUR)

binary output as alarm output- functions limit, change of flow direction or errorbinary output as pulse output- pulse value 0.01...1000 units- pulse width 80...1000 ms

FLUXUS G809**-A1G809**-A1A

G809**-F1

Page 9: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 9

Technical Specification FLUXUS® G809**-*1

Dimensions

FLUXUS G809**-A1

FLUXUS G809**-F1

in mm

243

174

thread, cable gland:6x M20 x 1.5

136

286

material: stainless steel 304 (1.4301)

243

174

thread: 4x NPT 3/4

136

286

material: stainless steel 304 (1.4301)

thread: 2x NPT 1/2

Page 10: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

10 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Wall and 2 " Pipe Mounting Kit

Page 11: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 11

Technical Specification FLUXUS® G809**-*1

Terminal Assignment

power supply

transducers

outputs

FLUXUS G809

AC DCterminal connection terminal connectionL phase L+ +N neutral L- -PE earth PE earth

measuring channel A measuring channel Bterminal connection terminal connectionAV transducer , signal BV transducer , signalAVS transducer , internal shield BVS transducer , internal shieldARS transducer , internal shield BRS transducer , internal shieldAR transducer , signal BR transducer , signalcable gland or equipo-tential bonding termi-nal (transducers)

external shield cable gland or equipo-tential bonding termi-nal (transducers)

external shield

terminal connection1(-), 2(+) current output I1 frequency output F13(-), 4(+) current output I25(-), 6(+) binary output B1 (open collector)7(-), 8(+) binary output B2 (open collector)9(-), 10(+) binary output B3 (open collector or

Reed relay)binary output B1 (open collector)

11(-), 12(+) binary output B4 (open collector or Reed relay)

A+, B-, S RS485

L N PE(+) (-)

AR ARS AVAVS BR BRS BVBVS

equipotential bonding terminal

lower housing,front view

upper housing,back view

equipotential bonding terminal

power supply

transducers

outputs

equipotential bonding terminal (transducers)

Page 12: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

12 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Transducers

Transducer Selection

Step 1a

Select a Lamb wave transducer:

Step 1bIf the pipe wall thickness is not in the range of the Lamb wave transducers, select a shear wave transducer:

Step 2

inner pipe diameter d dependent on the flow velocity v of the medium in the pipe

The transducers are selected from the characteristics (see next page). Lamb wave transducers are selectedfrom the left column, shear wave transducers from the right column.

Lamb wave transducers: If the values d and v are not in the range, the diagonal arrangement with 1 soundpath may be used, i.e. the same characteristics can be used with doubling the inner pipe diameter. If the val-ues are still not in the range, shear waves transducers regarding the pipe wall thickness have to be selectedin step 1b.

transducer order code

GLF 14 30 35

GLG 11 23

GLH 7 15

GLK 4 9

GLM 2 5

GLP 1...3

GLQ 0.5...1

5 10 15 20 25 30 35pipe wall thickness [mm]

transducer order code

GSG 14 23

GSK 5 9

GSM 2.5 5

GSP 1.5 3

GSQ 1 1.5

5 10 15 20 25pipe wall thickness [mm]

recommended possible

Page 13: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 13

Technical Specification FLUXUS® G809**-*1

Lamb wave transducer1 shear wave transducer1

GLF

GLG GSG

GLH

GLK GSK

GLM GSM

GLP GSP

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflection arrangement with2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200

1400

Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

200

400

600

800

1000

1200Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

Page 14: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

14 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Step 3

min. medium pressure

Example

Step 4

for the characters 4...11 of the transducer order code (ambient temperature, explosion protection, connectionsystem, extension cable) see page 15

Step 5

for the technical data of the selected transducer see page 16 et seqq.

GLQ GSQ

Lamb wave transducer shear wave transducertransducer order code

medium pressure1 [bar] transducer order code

medium pressure1 [bar]metal pipe plastic pipe metal pipe plastic pipe

min. min. extended min. min. min. extended min.GLF 15 10 1GLG 15 10 1 GSG 30 20 1GLH 15 10 1 GSK 30 20 1GLK 15 (d > 120 mm)

10 (d < 120 mm)10 (d > 120 mm)3 (d < 120 mm)

1 GSM 30 20 1

GLM 10 (d > 60 mm)5 (d < 60 mm)

3 (d < 60 mm) 1 GSP 30 20 1

GLP 10 (d > 35 mm)5 (d < 35 mm)

3 (d < 35 mm) 1 GSQ 30 20 1

GLQ 10 (d > 15 mm)5 (d < 15 mm)

3 (d < 15 mm) 1

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air

d - inner pipe diameter

step1 pipe wall thickness mm 12 12 12 30

selected transducer GLG or GLH GLG or GLH GLG or GLH GS2 inner pipe diameter mm 800 600 800 300

max. flow velocity m/s 15 15 30 15selected transducer GLG GLG or GLH values not in the range

of the characteristics, but by using diagonal arrangement with 1 sound path, the inner pipe diameter in the characteristics is dou-bled:GLG

GSK

3 min. medium pressure bar 17 17 17 35selected transducer GLG GLG or GLH

influence of acoustic noise is reduced with increased transducer frequency, thus recommended:GLH

GLG GSK

Lamb wave transducer1 shear wave transducer1

1 inner pipe diameter and max. flow velocity for a typical application with natural gas, nitrogen, oxygen in reflection arrangement with2 sound paths (Lamb wave transducers)/1 sound path (shear wave transducers)

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

0 5 10 15 20 25 30 35vms

20

40

60

80

100

120

140

Dmm

Page 15: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 15

Technical Specification FLUXUS® G809**-*1

Transducer Order Code

1, 2 3 4 5, 6 7, 8 9...11 12, 13 no. of character

tra

nsdu

cer

tra

nsdu

cer

fre

quen

cy

- ambi

ent

tem

pera

ture

expl

osio

n pr

ote

ctio

n

conn

ectio

n sy

stem

- exte

nsio

n ca

ble

/ optio

n

description

GL set of ultrasonic flow transducers for gas measurement, Lamb wave

GS set of ultrasonic flow transducers for gas measurement, shear wave

F 0.15 MHz (Lamb wave only)

G 0.2 MHz

H 0.3 MHz (Lamb wave only)

K 0.5 MHz

M 1 MHz

P 2 MHz

Q 4 MHz

N normal temperature range

E extended temperature range (shear wave transducers with trans-ducer frequency M, P, Q)

A1 ATEX zone 1/IECEx zone 1

F1 FM Class I Div. 1

TS direct connection or connection via junction box

XXX 0 m: without extension cable

> 0 m: G809**-A1: with junction box, G809**-F1: with terminal board KFM1

LC long transducer cable

IP68 degree of protection IP68

OS housing with stainless steel 316

example

GL M - N A1 TS - 000 Lamb wave transducer 1 MHz, normal temperature range, ATEX zone 1/IECEx zone 1, connection system TS (direct connection)

- - /

Page 16: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

16 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Technical Data

Shear Wave Transducers (zone 1)

technical type GDG1N81 GDK1N81order code GSG-NA1TS

GSG-NA1TS/OSGSK-NA1TSGSK-NA1TS/OS

transducer frequency MHz 0.2 0.5

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70min. recommended mm 380 80max. recommended mm 810 500max. extended mm 1100 720pipe wall thicknessmin. mm 14 5max. mm - -materialhousing PEEK with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP65 IP66

transducer cabletype 1699 1699length m 5 5dimensionslength l mm 129.5 126.5width b mm 51 51height h mm 67 67.5dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +130 +130temperature compensation

x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +180 +180marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637

Page 17: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 17

Technical Specification FLUXUS® G809**-*1

technical type GDM2N81 GDP2N81 GDQ2N81order code GSM-NA1TS

GSM-NA1TS/OSGSP-NA1TSGSP-NA1TS/OS

GSQ-NA1TSGSQ-NA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PEEK with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PEEK with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP66 IP66 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +130 +130 +130temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +180 +180 +180marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 18: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

18 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Shear Wave Transducers (zone 1, IP68)

technical type GDG1LI1 GDK1LI1 GDM2LI1 GDP2LI1order code GSG-NA1TS/IP68 GSK-NA1TS/IP68 GSM-NA1TS/IP68 GSP-NA1TS/IP68transducer frequency MHz 0.2 0.5 1 2

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 250 70 30 15min. recommended mm 380 80 40 20max. recommended mm 810 500 80 40max. extended mm 1100 720 120 60pipe wall thicknessmin. mm 14 5 2.5 1.5max. mm - - - -materialhousing PEEK with stainless steel

cap 316Ti (1.4571)PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

PEEK with stainless steel cap 316Ti (1.4571)

contact surface PEEK PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP68 IP68 IP68 IP68

transducer cabletype 2550 2550 2550 2550length m 12 12 12 12dimensionslength l mm 130 130 72 72width b mm 54 54 32 32height h mm 83.5 83.5 46 46dimensional drawing

ambient temperaturemin. °C -40 -40 -40 -40max. °C +100 +100 +100 +100temperature compensation

x x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C +180 +180 +180 +180marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3 3 3

l

hb

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 19: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 19

Technical Specification FLUXUS® G809**-*1

Shear Wave Transducers (zone 1, extended temperature range)

technical type GDM2E85 GDP2E85 GDQ2E85order code GSM-EA1TS

GSM-EA1TS/OSGSP-EA1TSGSP-EA1TS/OS

GSQ-EA1TSGSQ-EA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 6min. recommended mm 40 20 10max. recommended mm 80 40 20max. extended mm 120 60 30pipe wall thicknessmin. mm 2.5 1.5 1max. mm - - -materialhousing PI with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PI with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PI PI PIdegree of protection according to IEC/EN 60529

IP66 IP66 IP56

transducer cabletype 6111 6111 6111length m 4 4 3dimensionslength l mm 64 64 40width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -30 -30 -30max. °C +200 +200 +200temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 3D1 22

gas: 2G dust: 3D1 22

gas: 2G dust: 3D1 22

explosion protection temperature (pipe surface)min. °C -45 -45 -45max. °C +225 +225 +225marking

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

Ex e q IIC T6...T2 GbEx tb IIIA TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

l

hb

hb

l

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 20: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

20 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Lamb Wave Transducers (zone 1)technical type GRF1N83 GRG1N83 GRH1N83 GRK1N83order code GLF-NA1TS

GLF-NA1TS/OSGLG-NA1TSGLG-NA1TS/OS

GLH-NA1TSGLH-NA1TS/OS

GLK-NA1TSGLK-NA1TS/OS

transducer frequency MHz 0.15 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)3 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 230 190 120 60min. recommended mm 270 220 140 80max. recommended mm 1200 900 600 300max. extended mm 2100 1600 1000 500pipe wall thicknessmin. mm 14 11 7 4max. mm 30 23 15 9max. extended mm 35 - - -materialhousing PPSU with stainless steel

cap 304 (1.4301), option OS: 316L, 316Ti (1.4404, 1.4571)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L, 316Ti (1.4404, 1.4571)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L, 316Ti (1.4404, 1.4571)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L, 316Ti (1.4404, 1.4571)

contact surface PPSU PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP66 IP66 IP66

transducer cabletype 1699 1699 1699 1699length m 5 5 5 5dimensionslength l mm 163 128.5 128.5 128.5width b mm 54 51 51 51height h mm 91.3 67.5 67.5 67.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40 -40max. °C +170 +170 +170 +170temperature compensation

x x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55 -55max. °C +140 +140 +140 +140marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 21: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 21

Technical Specification FLUXUS® G809**-*1

technical type GRM1N83 GRP1N83 GRQ1N83order code GLM-NA1TS

GLM-NA1TS/OSGLP-NA1TSGLP-NA1TS/OS

GLQ-NA1TSGLQ-NA1TS/OS

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe:3 (d < 60 mm)

metal pipe:3 (d < 35 mm)

metal pipe:3 (d < 15 mm)

min. bar metal pipe:10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

metal pipe:10 (d > 15 mm)5 (d < 15 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 90 50 22max. extended mm 150 70 35pipe wall thicknessmin. mm 2 1 0.5max. mm 5 3 1max. extended mm - - -materialhousing PPSU with stainless steel

cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

PPSU with stainless steel cap 304 (1.4301), option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP65 IP65 IP65

transducer cabletype 1699 1699 1699length m 4 4 3dimensionslength l mm 74 74 42width b mm 32 32 22height h mm 40.5 40.5 25.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +170 +170 +170temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +140 +140 +140marking

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

Ex e q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: increased safety,

powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

gas: increased safety, powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

remark on request on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 22: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

22 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Lamb Wave Transducers (zone 1, IP68)

technical type GRG1LI3 GRH1LI3 GRK1LI3order code GLG-NA1TS/IP68 GLH-NA1TS/IP68 GLK-NA1TS/IP68transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)3 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68 IP68

transducer cabletype 2550 2550 2550length m 12 12 12dimensionslength l mm 143.5 143.5 143.5width b mm 54 54 54height h mm 83.5 83.5 83.5dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +100 +100 +100temperature compensation

x x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55 -55max. °C +140 +140 +140marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

gas: powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x x

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3 3

l

hb

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637 II2GII2D

0637

Page 23: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 23

Technical Specification FLUXUS® G809**-*1

technical type GRM1LI3 GRP1LI3order code GLM-NA1TS/IP68 GLP-NA1TS/IP68transducer frequency MHz 1 2

medium pressure1

min. extended bar metal pipe:3 (d < 60 mm)

metal pipe:3 (d < 35 mm)

min. bar metal pipe:10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15min. recommended mm 40 20max. recommended mm 90 50max. extended mm 150 70pipe wall thicknessmin. mm 2 1max. mm 5 3materialhousing PPSU with stainless steel

cap 316Ti (1.4571)PPSU with stainless steel cap 316Ti (1.4571)

contact surface PPSU PPSUdegree of protection according to IEC/EN 60529

IP68 IP68

transducer cabletype 2550 2550length m 12 12dimensionslength l mm 73 73width b mm 31.6 31.6height h mm 46 46dimensional drawing

ambient temperaturemin. °C -40 -40max. °C +100 +100temperature compensation

x x

explosion protection

ATEX/IECEx

categoryzone

gas: 2G dust: 2D1 21

gas: 2G dust: 2D1 21

explosion protection temperature (pipe surface)min. °C -55 -55max. °C +140 +140marking

Ex q IIC T6...T3 GbEx tb IIIC TX Db

Ex q IIC T6...T3 GbEx tb IIIC TX Db

certification ATEX IBExU07ATEX1168 X IBExU07ATEX1168 Xcertification IECEx IECEx IBE 08.0007X IECEx IBE 08.0007Xtype of protection gas: powder filling

dust: protection by enclosure

gas: powder fillingdust: protection by enclosure

transducer mounting fixture necessary

x x

remark on request1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

3 test conditions: 3 months/2 bar (20 m)/20 °C

3 3

l

hb

l

hb

II2GII2D

0637 II2GII2D

0637

Page 24: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

24 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Shear Wave Transducers (FM Class I, Div. 1)

technical type GDG1N62 GLG1N62 GDK1N62 GLK1N62order code GSG-NF1TS

GSG-NF1TS/OSGSG-NF1TS/LCGSG-NF1TS/OS/LC

GSK-NF1TSGSK-NF1TS/OS

GSK-NF1TS/LCGSK-NF1TS/OS/LC

transducer frequency MHz 0.2 0.5

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20min. bar metal pipe: 30

plastic pipe: 1metal pipe: 30plastic pipe: 1

inner pipe diameter dmin. extended mm 250 70min. recommended mm 380 80max. recommended mm 810 500max. extended mm 1100 720pipe wall thicknessmin. mm 14 5max. mm - -materialhousing stainless steel 304 (1.4301),

option OS: 316L (1.4404)stainless steel 304 (1.4301),option OS: 316L (1.4404)

contact surface PEEK PEEKdegree of protection according to IEC/EN 60529

IP66 IP66

transducer cabletype 2549 2549 2549 2549length m 15 46 15 46dimensionslength l mm 132 132width b mm 60 60height h mm 72 72mounting length lm mm 185 185dimensional drawing

operating temperaturemin. °C -40 -40max. °C +110 +110temperature compensation

x x

explosion protection

FM

explosion protection temperaturemin. °C -40 -40max. °C +125 +125marking S/Cl. I, II, III / Div. 1 /

GP A, B, C, D, E, F, G /Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 38311 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

l

hb

lm

Page 25: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 25

Technical Specification FLUXUS® G809**-*1

Shear Wave Transducers (FM Class I, Div. 1)

technical type GDM1N62 GLM1N62 GDP1N62 GLP1N62 GDQ1N62 GLQ1N62order code GSM-NF1TS

GSM-NF1TS/OS

GSM-NF1TS/LCGSM-NF1TS/OS/LC

GSP-NF1TSGSP-NF1TS/OS

GSP-NF1TS/LCGSP-NF1TS/OS/LC

GSQ-NF1TSGSQ-NF1TS/OS

GSQ-NF1TS/LCGSQ-NF1TS/OS/LC

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe: 20 metal pipe: 20 metal pipe:3 (d < 15 mm)

min. bar metal pipe: 30plastic pipe: 1

metal pipe: 30plastic pipe: 1

metal pipe:10 (d > 15 mm)5 (d < 15 mm)plastic pipe: 1

inner pipe diameter dmin. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 80 40 22max. extended mm 120 60 35pipe wall thicknessmin. mm 2.5 1.5 0.5max. mm - - 1materialhousing stainless steel 304 (1.4301),

option OS: 316L (1.4404)stainless steel 304 (1.4301),option OS: 316L (1.4404)

stainless steel 304 (1.4301),option OS: 316L (1.4404)

contact surface PEEK PEEK PEEKdegree of protection according to IEC/EN 60529

IP66 IP66 IP66

transducer cabletype 2549 2549 2549 2549 2549 2549length m 15 46 15 46 20 46dimensionslength l mm 60 60 60width b mm 30 30 30height h mm 43 43 43mounting length lm mm 110 110 110dimensional drawing

operating temperaturemin. °C -40 -40 -40max. °C +110 +110 +110temperature compensation

x x x

explosion protection

FM

explosion protection temperaturemin. °C -40 -40 -40max. °C +125 +125 +125marking S/Cl. I, II, III / Div. 1 /

GP A, B, C, D, E, F, G /Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 3831remark on request on request

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 shear wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended/max. extended: in diagonal arrangement and for a flow velocity of 15 m/s

hb

l

lm

Page 26: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

26 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Lamb Wave Transducers (FM Class I, Div. 1)

technical type GRG1N62 GTG1N62 GRH1N62 GTH1N62 GRK1N62 GTK1N62order code GLG-NF1TS

GLG-NF1TS/OS

GLG-NF1TS/LCGLG-NF1TS/OS/LC

GLH-NF1TSGLH-NF1TS/OS

GLH-NF1TS/LCGLH-NF1TS/OS/LC

GLK-NF1TSGLK-NF1TS/OS

GLK-NF1TS/LCGLK-NF1TS/OS/LC

transducer frequency MHz 0.2 0.3 0.5

medium pressure1

min. extended bar metal pipe: 10 metal pipe: 10 metal pipe:10 (d > 120 mm)3 (d < 120 mm)

min. bar metal pipe: 15plastic pipe: 1

metal pipe: 15plastic pipe: 1

metal pipe:15 (d > 120 mm)10 (d < 120 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 190 120 60min. recommended mm 220 140 80max. recommended mm 900 600 300max. extended mm 1600 1000 500pipe wall thicknessmin. mm 11 7 4max. mm 23 15 9materialhousing stainless steel 304 (1.4301),

option OS: 316L (1.4404)stainless steel 304 (1.4301),option OS: 316L (1.4404)

stainless steel 304 (1.4301),option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP66 IP66 IP66

transducer cabletype 2549 2549 2549 2549 2549 2549length m 15 46 15 46 15 46dimensionslength l mm 132 132 132width b mm 60 60 60height h mm 72 72 72mounting length lm mm 185 185 185dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +110 +110 +110temperature compensation

x x x

explosion protection

FM

explosion protection temperaturemin. °C -40 -40 -40max. °C +125 +125 +125marking S/Cl. I, II, III / Div. 1 /

GP A, B, C, D, E, F, G /Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 38311 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

lm

Page 27: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 27

Technical Specification FLUXUS® G809**-*1

Lamb Wave Transducers (FM Class I, Div. 1)

technical type GRM1N62 GTM1N62 GRP1N62 GTP1N62 GRQ1N62 GTQ1N62order code GLM-NF1TS

GLM-NF1TS/OS

GLM-NF1TS/LCGLM-NF1TS/OS/LC

GLP-NF1TSGLP-NF1TS/OS

GLP-NF1TS/LCGLP-NF1TS/OS/LC

GLQ-NF1TSGLQ-NF1TS/OS

GLQ-NF1TS/LCGLQ-NF1TS/OS/LC

transducer frequency MHz 1 2 4

medium pressure1

min. extended bar metal pipe:3 (d < 60 mm)

metal pipe:3 (d < 35 mm)

metal pipe:3 (d < 15 mm)

min. bar metal pipe:10 (d > 60 mm)5 (d < 60 mm)plastic pipe: 1

metal pipe:10 (d > 35 mm)5 (d < 35 mm)plastic pipe: 1

metal pipe:10 (d > 15 mm)5 (d < 15 mm)plastic pipe: 1

inner pipe diameter d2

min. extended mm 30 15 7min. recommended mm 40 20 10max. recommended mm 90 50 22max. extended mm 150 70 35pipe wall thicknessmin. mm 2 1 0.5max. mm 5 3 1materialhousing stainless steel 304 (1.4301),

option OS: 316L (1.4404)stainless steel 304 (1.4301),option OS: 316L (1.4404)

stainless steel 304 (1.4301),option OS: 316L (1.4404)

contact surface PPSU PPSU PPSUdegree of protection according to IEC/EN 60529

IP66 IP66 IP66

transducer cabletype 2549 2549 2549 2549 2549 2549length m 15 46 15 46 15 46dimensionslength l mm 80 80 80width b mm 38 38 38height h mm 44 44 44mounting length lm mm 135 135 135dimensional drawing

ambient temperaturemin. °C -40 -40 -40max. °C +110 +110 +110temperature compensation

x x x

explosion protection

FM

explosion protection temperaturemin. °C -40 -40 -40max. °C +125 +125 +125marking S/Cl. I, II, III / Div. 1 /

GP A, B, C, D, E, F, G /Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 3831

S/Cl. I, II, III / Div. 1 / GP A, B, C, D, E, F, G /

Temperature Codes dwg 3831remark on request on request

1 depending on application, typical absolute value for natural gas, nitrogen, compressed air2 Lamb wave transducer:

typical values for natural gas, nitrogen, oxygen, pipe diameters for other gases on requestpipe diameter min. recommended/max. recommended: in reflection arrangement and for a flow velocity of 15 m/spipe diameter max. extended: in diagonal arrangement and for a flow velocity of 25 m/s

l

hb

lm

Page 28: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

28 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Transducer Mounting Fixture

Order Code

1, 2 3 4 5 6 7...9 10, 11 no. of character

tran

sduc

er

mou

ntin

g fix

ture

tran

sduc

er

- mea

sure

men

t ar

rang

emen

t

size

- fixat

ion

oute

r pi

pe

diam

eter

/ optio

n

description

VL Variofix L

VС Variofix C

PF PermaFiX

K transducers with transducer frequency F (VCK-*L-****/IP68), G, H, K

M transducers with transducer frequency M, P, Q

Q transducers with transducer frequency Q

D reflection arrangement or diagonal arrangement

R reflection arrangement

S small

M medium

L large

B bolts

S tension straps

W welding

N without fixation

002 10...20 mm

004 20...40 mm

T36 40...360 mm

013 10...130 mm

036 130...360 mm

092 360...920 mm

200 920...2000 mm

SK1 0.5...2.5 in

SK2 3...6 in

SK3 8...10 in

SK4 12...18 in

SK5 20...36 in

SK6 42...100 in

IP68 degree of protection IP68

OS housing with stainless steel 316

Z special design

example

VL K - D S - S 200 Variofix L and tension straps for transducers with transducer fre-quency G, H, K

PF M - D S - S 200 PermaFiX and tension straps for transducers with transducer frequency M, P, Q

- - /

Page 29: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 29

Technical Specification FLUXUS® G809**-*1

Variofix L (VLK, VLM, VLQ) material: stainless steel 304 (1.4301), 301 (1.4310), 410 (1.4006)option OS: 316 (1.4571), 316L (1.4404), 17-7PH (1.4568)

inner length:VLK: 348 mm,option IP68: 368 mmVLM: 234 mmVLQ: 176 mm

dimensions:VLK: 423 x 90 x 93 mm,option IP68: 443 x 94 x 105 mmVLM: 309 x 57 x 63 mmVLQ: 247 x 43 x 47 mm

Variofix L with bolt mounting plate (VL*-**-B) material: stainless steel 304 (1.4301), 301 (1.4310), 410 (1.4006)option OS: 316 (1.4571), 316L (1.4404), 17-7PH (1.4568)

inner length:VLM: 234 mmVLQ: 176 mm

dimensions:VLM: 309 x 57 x 63 mmVLQ: 247 x 43 x 47 mm

outer pipe diameter:max. 48 mm

Variofix C (VC) material: stainless steel 304 (1.4301), 301 (1.4310)option OS: 316 (1.4571)

inner length:VCK-*L: 500 mmVCK-*S: 350 mmVCM: 400 mmVCQ: 250 mm

dimensions:VCK-*L: 560 x 122 x 102 mm,option IP68: 560 x 126 x 120 mmVCK-*S: 410 x 122 x 102 mm,option IP68: 410 x 126 x 120 mmVCM: 460 x 96 x 80 mmVCQ: 310 x 85 x 62 mm

PermaFiX with tension straps (PF*-DS-S)

material: stainless steel 304 (1.4301), 301 (1.4310)option OS: 316 (1.4571)

inner length:PFK: 373 mmPFM: 276 mm

dimensions:PFK: 410 x 90 x 73 mmPFM: 310 x 68 x 44 mm

PermaFiX with bolts (PF*-DS-B)

Page 30: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

30 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Coupling Materials for Transducers

Technical Data

normal temperature range(4th character of transducer order code = N)

extended temperature range(4th character of transducer order code = E)

< 100 °C < 170 °C < 150 °C < 200 °C< 24 h coupling compound type N

or coupling foil type VTcoupling compound type E or coupling foil type VT

coupling compound type E or coupling foil type VT

coupling compound type E or H or coupling foil type VT

long time measurement

coupling foil type VT1 coupling foil type VT2 coupling foil type VT1 coupling foil type VT2

1 < 5 years2 < 6 months

type order code ambient temperature material remark°C

coupling compound type N

990739-1 -30...+130 mineral grease paste

coupling compound type E

990739-2 -30...+200 silicone paste

coupling compound type H

990739-3 -30...+250 fluoropolymer paste

coupling foil type VT 990739-61 -10...+200 fluoroelastomer for transducers with transducer frequency F

990739-0 for transducers with transducer frequency G, H, K

990739-6 for shear wave transducers with transducer frequency M, P

990739-14 for shear wave transducers IP68 and Lambwave transducers with transducer frequency M, P, Q

990739-5 for transducers with transducer frequency Q

Page 31: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 31

Technical Specification FLUXUS® G809**-*1

Damping Mats (optional)Damping mats will be used for the gas measurement to reduce acoustic noise influences on the measure-ment.

transducer damping matTransducer damping mats will be installed below the transducers.

pipe damping matPipe damping mats will be installed if the sound propagation is disturbed at reflection points (e.g. flange,weld). Depending on the noise, the pipe damping mats will be installed at one or both sides of the transducerdamping mat. If the local conditions are unknown, pipe damping mats should be installed.

Technical Data

type E30R4 E30R3width mm 225 50thickness mm 0.7length (per roll) m 10weight kg/m² 1.015ambient temperature °C -30...+80properties self-adhesive

outer pipe diameter ≥ 900 mm outer pipe diameter < 900 mm

transducer damping matpipe damping mat (optional)

installation length

diagonal arrangement

reflection arrangement

Page 32: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

32 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Dimensioning

transducer damping mattransducer mounting fixture

order code

type number of layers

transducer damping mat transducer damping mat+2x pipe damping mat

max. instal-lation length number of rolls1

max. instal-lation length number of rolls1

[mm] standard2 extended 2 [mm] standard extendedVarioFix LVLK GLG E30R4 3 890 4 4 1830 9 13

GSG 3 4 4 9 11GLH 2 2 3 4 7GLK 1 1 1 1 2GSK 1 1 1 2 3

VLK-**-****/IP68 GLG E30R4 3 930 5 5 1910 10 14GSG 3 5 5 10 11GLH 2 2 3 5 7GLK 1 1 1 2 2GSK 1 1 2 2 3

VLM GLM E30R3 1 660 1 1 1360 1 2GSM 1 1 1 1 2GLP 1 1 1 1 1GSP 1 1 1 1 1

VLQ GLQ E30R3 1 540 1 1 1120 1 1GSQ 1 1 1 1 1

Variofix CVCK-*L-****/IP68 GLF E30R4 3 1160 6 6 2360 13 18VCK-*LVCK-*L-****/IP68

GLG E30R4 3 1160 6 6 2360 11 15GSG 3 6 6 11 12GLH 2 3 4 5 8GLK 1 1 1 2 2GSK 1 2 2 3 3

VCK-*SVCK-*S-****/IP68

GLG E30R4 3 860 4 4 1760 7 10GSG 3 4 4 7 8GLH 2 2 3 4 5GLK 1 1 1 1 2GSK 1 1 1 2 2

VCM GLM E30R3 1 960 1 2 1960 2 3GSM 1 1 1 2 2GLP 1 1 1 1 1GSP 1 1 1 1 1

VCQ GLQ E30R3 1 660 1 1 1360 1 1GSQ 1 1 1 1 1

PermaFiXPFK GLG E30R4 3 860 4 4 1760 7 10

GSG 3 4 4 7 8GLH 2 2 3 4 5GLK 1 1 1 1 2GSK 1 1 1 2 2

PFM GLM E30R3 1 660 1 1 1360 1 2GSM 1 1 1 1 2GLP 1 1 1 1 1GSP 1 1 1 1 1GLQ 1 1 1 1 1GSQ 1 1 1 1 1

1 calculation on the base of:- max. installation length (installation of one transducer mounting fixture per transducer in reflection arrangement) and- max. recommended pipe diameter (standard) or max. extended pipe diameter (extended) (for inner pipe diameter max. recommended and max. extended see Technical Data of the Transducers from page 16)

2 calculation for the number of rolls when both transducers are mounted in one transducer mounting fixture (reflection arrangement) or in diagonal arrangement: number of rolls/2 and round up to the nearest integer

Page 33: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 33

Technical Specification FLUXUS® G809**-*1

Connection Systemsconnection system TSconnection with extension cable direct connection transducers

technical type*****8*

****LI*

*****62

transducer frequency(3d character of transducer

order code)

F, G, H, K M, P Q S

TS

x l x l x l x lcable length m 5 ≤ 300 4 ≤ 300 3 ≤ 90 2 ≤ 40

cable length (*****62) m 15 ≤ 300 15 ≤ 300 15 ≤ 90 - -cable length (option LC,

*****62)m 46 ≤ 300 46 ≤ 300 46 ≤ 90 - -

cable length (option IP68) m 12 ≤ 300 12 ≤ 300 - - - -

x - transducer cable lengthl - max. length of extension cable

tran

smitt

er

x

l

JB01

tran

smitt

er

x

tran

smitt

er

x

l

JB01

tran

smitt

er

x

extension cable

transmitter

terminal board for junction box(junction box by customer)

tran

sduc

er

tran

sduc

er

tran

smitt

er

Page 34: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

34 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

Transducer Cable

Technical Data

transducer cabletype 1699 2550 (option IP68) 6111 2549ambient temperature °C -55...+200 -40...+100 -100...+225 -100...+200properties longitudinal water

tightcable jacketmaterial PTFE PUR PFA PTFEouter diameter mm 2.9 5.2 ±0.2 2.7 5.3thickness mm 0.3 0.9 0.5 0.5colour brown grey white blackshield x x x xsheathmaterial stainless steel 304

(1.4301)- stainless steel 304

(1.4301)-

option OS: 316L (1.4404)

option OS: 316L (1.4404)

outer diameter mm 8 - 8 -

extension cabletype 2615 5245transmitter G809**-F1 G809**-A1ambient temperature °C -40...+70 -30...+70properties halogen free

fire propagation test according to IEC 60332-1

combustion test according to IEC 60754-2

halogen free

fire propagation test according to IEC 60332-1

combustion test according to IEC 60754-2

cable jacketmaterial PUR PURouter diameter mm 12 12thickness mm 2 2colour black blackshield x xsheathmaterial - steel wire braid with

copolymer sheathouter diameter mm - 15.6

Page 35: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 35

Technical Specification FLUXUS® G809**-*1

Junction Box (G809**-A1)

Technical Data

Dimensions

technical type JB01S4E3Mdimensions see dimensional drawingfixation wall mounting, optional: 2 " pipe

mountingmaterialhousing stainless steel 316L (1.4404)gasket siliconedegree of protection according to IEC/EN 60529

IP67

ambient temperaturemin. °C -40max. °C +80explosion protection

ATEX/IECEx

zone 1marking

Ex e mb IIC (T6)...T4 GbEx tb IIIC T 100 °C DbTa -40...+(70)80 °C

certification ATEX IBExU06ATEX1161certification IECEx IECEx IBE 08.0006type of protection gas:

• increased safety• decoupled network:

encapsulation dust: protection by enclosure

in mm

II2GII2D

0637

174

1192

wall mounting holder

70163.5

156

Ø 9

thread: 3x M20 x 1.5cable gland: max. 2x M20

Page 36: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

36 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

2 " Pipe Mounting Kit (optional)

Terminal Assignment

JB01

transducers

terminal strip KL1terminal connection

V transducer , signalVS transducer , internal shieldRS transducer , internal shieldR transducer , signalcable gland external shield

extension cable

terminal strip KL2terminal connection

TV signalTVS internal shieldTRS internal shieldTR signalshield terminal external shield

shield terminal

equipotential bonding terminal(at wall mounting holder)

Page 37: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 37

Technical Specification FLUXUS® G809**-*1

Extension Cable (G809**-F1)The extension cable and the transducers are connected via terminal board KFM1. The terminal board has tobe installed into a junction box (by customer) approved for hazardous areas.

Terminal Assignment for Terminal Board KFM1

transducers

terminal connectionterminal strip KL1V transducer , signalVS transducer , internal shieldVOS transducer , external shieldterminal strip KL2R transducer , signalRS transducer , internal shieldROS transducer , external shield

extension cable

terminal strip KL3terminal connection

TV signalTVS internal shieldTOS external shieldTRS internal shieldTR signal

110.9

115

73.5

73.5

Ø5.5

in mm

Page 38: Technical Specification FLUXUS® G809**-*1 · 4 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21 FLUXUS® G809**-*1 Technical Specification Number of Sound Paths The number of sound paths

38 TSFLUXUS_G809xx-x1V1-7-1EN_Leu, 2015-10-21

FLUXUS® G809**-*1 Technical Specification

FLEXIM GmbH

Wolfener Str. 36

12681 Berlin

Germany

Tel.: +49 (30) 93 66 76 60

Fax: +49 (30) 93 66 76 80

internet: www.flexim.com

e-mail: [email protected]

Subject to change without notification. Errors excepted.