Kambič d.o.o. Metliška cesta 16, SI-8333 Semič, Slovenia - EU,
T: +386 (0)7 35 65 220, info©kambic.com, [email protected] www.kambic.com, www.kambicmetrology.com
• Maintaining superior temperature stability
• Data loggers and sensors calibration
• World class metrology performance • Accelerated ageing
• Stress tests
TEMPERATURE CHAMBER
(- 90 °C …+ 90 °C)
TK-64 CKULT
Kambič d.o.o. Metliška cesta 16, SI-8333 Semič, Slovenia - EU,
T: +386 (0)7 35 65 220, info©kambic.com, [email protected] www.kambic.com, www.kambicmetrology.com
Device description:
All exterior in galvanized and powder coated
steel RAL-7035
Extensive heat insulation shield on all chamber sides and doors
Parallel sealed with soft profile silicone seal to ensure perfect
tightness
Fan for internal air circulation with speed
control
Heavy duty closing
mechanism
Access port with both end plugs
Ø50 as standard, Ø40 or Ø90 as
accessory
Fully stainless steel chamber
2 level over temp. protection
1. 5°C over set point
heating disabled
2. 10°C above max .temp
mechanical shut off
Full 4 side maintenance access to ref.
system
Heavy duty swivel wheels with brake for
simple manipulation
Height adjustable stainless steel wire shelves
(1 wire shelve included, extra as accessory, also
perforated shelves as accessory)
Touch ontroller with user friendly interface, advance settings options, RS-232, (USB communication port
as accessory)
Kambič d.o.o. Metliška cesta 16, SI-8333 Semič, Slovenia - EU,
T: +386 (0)7 35 65 220, info©kambic.com, [email protected] www.kambic.com, www.kambicmetrology.com
Control panel with touch display:
Controller functions:
Menu, Exit, Clear, Enter
buttons
Time and date
Parameters display
Numerical keyboard
Increase value
Parameter selection (forward/
backward)
Decrease value
Display power switch
Start temperature
control
Stop temperature
control
LCD capacitive
touch display
System process flow
diagram
Kambič d.o.o. Metliška cesta 16, SI-8333 Semič, Slovenia - EU,
T: +386 (0)7 35 65 220, info©kambic.com, [email protected] www.kambic.com, www.kambicmetrology.com
Technical data:
*All performance in controlled environment (Tambient = 22 °C ± 3 °C)! *Accessories might affect performance!
TK-64 CKULT
External dimensions (WxHxD) in mm
850 x 1720 x 1100
Internal dimensions (WxHxD) in mm
450 x 400 x 400
Volume (L) ~ 72 Temperature range (°C) - 90 …+ 90 Heat up rate (°C/min) EN 60068-3-5
3
Cool down rate (°C/min) EN 60068-3-5
1
Temperature display resolution (°C) 0.1 Temperature set resolution (°C) 0.1
Temperature stability (°C)
± 0,08 °C @ - 90 °C ± 0,08 °C @ - 40 °C ± 0,05 °C @ 0 °C ± 0,05 °C @ + 90 °C
Temperature uniformity (°C)
± 0,8 °C @ - 90 °C ± 0,8 °C @ - 40 °C ± 0,6 °C @ 0 °C ± 0,8°C @ + 90 °C
Temperature control PID Power supply 3x400V 50/60Hz Wattage (W) 4800
Interface RS 232
USB (accessory) Shelve 1 (max 6) Shelve capacity (kg) 20 Max capacity (kg) 60
Access port (mm) Ø 40 standard,
(Ø 50, Ø 90 as accessory) Noise (dBA) @ 1 m distance 67 Weight (kg) ̴ 280
Kambič d.o.o. Metliška cesta 16, SI-8333 Semič, Slovenia - EU,
T: +386 (0)7 35 65 220, info©kambic.com, [email protected] www.kambic.com, www.kambicmetrology.com
Ordering information and accessories:
Description Part no.
Temperature chamber TK-64 CKULT 2728 Shelve wire TK-64 CKULT 2998 Shelve perforated TK-64 CKULT 2999 Access Port w. Plug Ø 40 mm right 2629 Access Port w. Plug Ø 50 mm right 608 Access Port w. Plug Ø 90 mm right 1731 USB interface 1466 TK-Tool (Software Monitoring & History) 1663 Password protection 1718 Evaluation report 9 points, 3 temperatures - performed by Kambic 1719 Evaluation report 9 points, 3 temperatures - Accredited 1852