TE-331/332 INSPECTION RECORD
INFUSION PUMP TE-172
INSPECTION RECORD
NoCheck itemResult
YesNo
1Is there any damage, etc in the main unit or pole clamp
2Can it be operated with battery power
3Is normal self-check is provided when the power switch is ON
4If the AC power cable is connected, does the BATTERY indicator go ON
5Was the [AC/DC] indicator lit when AC power was turned on ?
6With the door open, did the MIDPRESS finger section move momentarily
7If the door is closed and then opened. Is the tube clamp closed
8Does the START/STOP/SILENCE button function normally?
9Is the tube firmly clamped
10Does the air-in-line sensor work normally
11Was the Occlusion alarm activated in the following range ?
Upstream : - 30 ~ - 80 kPa ( - 0.3 ~ - 0.8 kgf/cm2 )
Downstream : 30 ~ 90 kPa ( 0.3 ~ 0.9 kgf/cm2 )
Setting on Pump:
Flow rate : 120 mL/h
Delivery Limit : 100 mL
Distance from pump to pressure gauge :
50 cm for Upstream Occlusion Sensor
100 cm for Downstream Occlusion Sensor
Value:
Upstream:.........................Downstream:........................
12Was the [ FLOW RATE ] accuracy within the following range ?
9-11 mL
Setting on Pump :
Flow Rate : 300 mL/h
Delivery Limit : 5 mL
Test Time : 1 minutes
Value:
........................
Remarks:
.......................................................................................
..cription:
or work normallyork normallyified rangeV Set )
sedeven r
ringe is moved to the end of its travel1111111111111111111
Serial Number:Date:Checked by:Signature: