 INFUSION PUMP TE-172 INSPECTION RECORD N o Check item Result  Y es No 1 Is ther e an y damage, etc in the mai n uni t or pole c lamp 2 Can it be operated with ba tter y po wer 3 Is norma l self -check i s pr ovided when th e po wer swit ch is O ! If t he " C power cable i s connected, does the #"$$% &' indicator go O ( ) as t he * "C+ C i ndicat or l it when "C power was turn ed on . / )ith th e door open, d id the 0I &% nger sect ion move momentarily 4 If t he d oor is closed and then opened5 Is the tube clamp closed 6 oes the $ "& $+ $O+ I7%C% button functi on normally. 8 Is the t ube rmly clamped 19 oes the air- in- line sensor work normally 11 ) as th e Occ lusio n ala rm a ctiv ated in th e foll owing rang e . :pstream ; - 39 < - 69 ka = - 953 < - 956 kgf+cm 2  > ownstream ; 39 < 89 ka = 953 < 958 kgf+cm 2 > etting on ump; ?low rate ; 129 m7+h elivery 7imit ; 199 m7 istance from pump to pressure gauge ; o (9 cm for :pstream Occlusion ensor o 199 cm for ownstream Occlusion ensor  @ alue; :pstream; 5555555555555555555555555 ownstream; 555555555555555555555555 12 ) as th e * ?7O) &" $% accur acy withi n the foll owing range . 8-11 m7 etting on ump ; ?low &ate ; 399 m7+h elivery 7imit ; ( m7 $est $ime ; 1 minutes  @ alue; 555555555555555555555555 &emarks; A5555555555555555555555555555555555555555555555555555555555555555555555555555555555555AAAAAAAAA55 AAAAAAAAA AAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAAA

TE172 NEO

Embed Size (px)

DESCRIPTION

doc

Citation preview

TE-331/332 INSPECTION RECORD

INFUSION PUMP TE-172

INSPECTION RECORD

NoCheck itemResult

YesNo

1Is there any damage, etc in the main unit or pole clamp

2Can it be operated with battery power

3Is normal self-check is provided when the power switch is ON

4If the AC power cable is connected, does the BATTERY indicator go ON

5Was the [AC/DC] indicator lit when AC power was turned on ?

6With the door open, did the MIDPRESS finger section move momentarily

7If the door is closed and then opened. Is the tube clamp closed

8Does the START/STOP/SILENCE button function normally?

9Is the tube firmly clamped

10Does the air-in-line sensor work normally

11Was the Occlusion alarm activated in the following range ?

Upstream : - 30 ~ - 80 kPa ( - 0.3 ~ - 0.8 kgf/cm2 )

Downstream : 30 ~ 90 kPa ( 0.3 ~ 0.9 kgf/cm2 )

Setting on Pump:

Flow rate : 120 mL/h

Delivery Limit : 100 mL

Distance from pump to pressure gauge :

50 cm for Upstream Occlusion Sensor

100 cm for Downstream Occlusion Sensor

Value:

Upstream:.........................Downstream:........................

12Was the [ FLOW RATE ] accuracy within the following range ?

9-11 mL

Setting on Pump :

Flow Rate : 300 mL/h

Delivery Limit : 5 mL

Test Time : 1 minutes

Value:

........................

Remarks:

.......................................................................................

..cription:

or work normallyork normallyified rangeV Set )

sedeven r

ringe is moved to the end of its travel1111111111111111111

Serial Number:Date:Checked by:Signature: