Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Synchronous Multi-Directional Motion Encoding
in Magnetic Resonance Elastography
BY
DAVID A. BURNS
B.S., University of Illinois, Urbana-Champaign, 2010
THESIS
Submitted as partial fulfillment of the requirements
for the degree of Master of Science of Mechanical Engineering
in the Graduate College of the
University of Illinois at Chicago, 2014
Chicago, Illinois
Defense Committee:
Dieter Klatt, Chair and Advisor
Thomas J. Royston
Michael J. Scott
ii
ACKNOWLEDGMENTS
I would like to thank the members of my thesis defense committee; Dr. Dieter Klatt, Dr. Thomas
Royston, and Dr. Michael J. Scott; for their support and assistance.
Additionally, I am sincerely thankful to the following individuals for their tireless support during the
course of my research: Steve Kearney, Spencer Brinker, Altaf Khan, Dr. Weiguo Li, Dr. Andrew Larson,
Dr. Daniel Procissi, Sol Misener, Mark Brown, Andrew Gordon, and Dr. Kaya Yasar.
Much of the background explanation in this work follows the direction of Dr. Dieter Klatt’s “MR
Elastography” and “Advances in MR Elastography” courses, held at the University of Illinois at Chicago
(Fall 2013 – Spring 2014). In completing this thesis, I am greatly indebted to his intensive and succinct
coverage of, and patient guidance through, an enormous body of knowledge.
DAB
iii
TABLE OF CONTENTS
CHAPTER PAGE
I. INTRODUCTION……………………………………………………………………... 1
II. CONCEPTUAL FRAMEWORK……………………………………………………… 3
A. Magnetic Resonance Imaging Background……………………………………….. 3
1. Nuclear Magnetic Resonance…………………………………………….. 3
2. Magnetic Field Gradients………………………………………………… 5
3. Signal Echoes…………………………………………………………….. 6
4. Pulse Sequence Diagrams………………………………………………… 8
B. Mechanical Wave Theory Background…………………………………………… 13
1. Equation of Motion of a Continuum…………………………………….... 13
2. Mechanical Stiffness Retrieval…………………………………………… 16
C. Magnetic Resonance Elastography Background………………………………….. 18
1. Phase Accumulation……………………………………………………… 19
2. Magnetic Resonance Elastography Pulse Sequence……………………… 27
3. Data Sampling……………………………………………………………. 30
III. MOTIVATION………………………………………………………………………... 32
A. Three-dimensional Encoding……………………………………………………… 32
B. Selective spectral Displacement Projection - Magnetic Resonance Elastography... 32
C. SampLe Interval Modulation – Magnetic Resonance Elastography……………… 38
IV. METHODS…………………………………………………………………………….. 44
A. Spin Echo Programming…………………………………………………………... 44
B. Magnetic Resonance Elastography Programming……………………………….... 46
C. Selective spectral Displacement Projection Programming………………………... 53
D. SampLe Interval Modulation Programming………………………………………. 54
E. Frequency Correction……………………………………………………………... 55
V. EXPERIMENTAL SETUP……………………………………………………………. 59
A. Physical Vibration Source………………………………………………………… 59
B. Experimental Gel Sample…………………………………………………………. 60
C. Sequence Parameters……………………………………………………………… 62
VI. RESULTS……………………………………………………………………………… 65
iv
TABLE OF CONTENTS (continued)
CHAPTER PAGE
VII. DISCUSSION…………………………………………………………………………. 70
A. Limitations………………………………………………………………………… 70
B. Conclusion………………………………………………………………………… 71
CITED LITERATURE………………………………………………………………… 72
VITA…………………………………………………………………………………... 74
v
LIST OF TABLES
TABLE PAGE
I. TIME-DOMAIN FUNCTIONS AND THEIR FREQUENCY-DOMAIN
COUNTERPARTS, OBTAINED VIA THE FOURIER TRANSFORM………… 15
II. SELECTIVE SPECTRAL DISPLACEMENT PROJECTIN – MAGNETIC
RESONANCE ELASTOGRAPHY PARAMETERS, INCLUDING MOTION-
ENCODING GRADIENT FREQUENCY AND NUMBER OF CYCLES FOR
EACH PROJECTION……………………………………………………………... 35
III. SELECTIVE SPECTRAL DISPLACEMENT PROJECTIN – MAGNETIC
RESONANCE ELASTOGRAPHY PARAMETERS COMPATIBLE WITH
TRAPEZOIDAL MOTION-ENCODING GRADIENTS, INCLUDING
FREQUENCY AND NUMBER OF CYCLES FOR EACH PROJECTION……... 37
IV. MOTION-ENCODING GRADIENT START TIMES FOR SAMPLE
INTERVAL MODULATION – MAGNETIC RESONANCE ELASTOGRAPY
WITH VIBRATION FREQUENCY AND NUMBER OF OFFSETS,
WITHOUT AND WITH CONSIDERATION OF THE SYMMETRY AND
PERIODICITY OF MOTION-ENCODING GRADIENTS………........................ 42
V. MAGNETIC RESONANCE IMAGING PARAMETERS USED FOR ALL
MAGNETIC RESONANCE ELASTOGRAPHY SCANS……………………….. 62
VI. MAGNETIC RESONANCE ELASTOGRAPHY PARAMETERS FOR ONE-
DIMENSIONAL SCANS IN THE SLICE, READ, AND PHASE
DIRECTIONS, AND FOR THE SAMPLE INTERVAL MODULATION
SCAN……………………………………………………………………………… 63
VII. MAGNETIC RESONANCE ELASTOGRAPHY PARAMETERS FOR ONE-
DIMENSIONAL SCANS IN THE SLICE, READ, AND PHASE
DIRECTIONS, AND FOR THE SELECTIVE SPECTRAL
DISPLACEMENT PROJECTION SCAN………………………………………... 64
vi
LIST OF FIGURES
FIGURE PAGE
1. The response of magnetic spin vectors with no external field, with an external magnetic
field applied, during excitation by an RF pulse, and emitting a Free Induction Decay
signal during relaxation …………………………………………………………………… 5
2. The state of a macroscopic magnetization vector after excitation by a 90-degree RF
pulse, after magnetization vectors have dephased due to local field inhomogeneities at
one-half echo time, after application of a 180-degree refocusing RF pulse, and after
magnetization vectors have rephrased to produce a signal echo at echo time……………... 7
3. The Nuclear Magnetic Resonance signal intensity decay during Free Induction Decay,
during a Spin Echo, and during a Gradient Echo………………………………………….. 8
4. The slice, read, and phase directions as they relate to a specimen in a Magnetic
Resonance Imaging scanner, to the acquired k-space data, and to an image produced
through the inverse two-dimensional Fourier Transform of such k-space data …………… 9
5. A Gradient Echo Magnetic Resonance Imaging pulse sequence with echo time defined…. 11
6. A Spin Echo Magnetic Resonance Imaging pulse sequence with both half-echo times
defined……………………………………………………………………………………… 13
7. A zeroth-moment-nulled function K(t) with two periods of oscillation and amplitude
of A. ……………………………………………………………………………………….. 20
8. The encoding efficiency of sinusoidal motion-encoding gradients for vibration
frequencies surrounding the motion-encoding gradient frequency. ……………………… 23
9. A realistic trapezoidal motion-encoding gradient with finite ramp up, ramp down, and
flat top times. ……………………………………………………………………………… 24
10. A Gradient Echo Magnetic Resonance Elastography pulse sequence diagram with echo
time defined. ………………………………………………………………………………. 27
11. A Spin Echo Magnetic Resonance Elastography pulse sequence diagram with both
half-echo times and the motion-encoding gradient gap defined. ………………………….. 29
12. Motion-encoding gradients applied at different starting phases of the physical vibration
to produce different snapshots of the propagating wave. …………………………………. 31
vii
LIST OF FIGURES (continued)
FIGURE PAGE
13. Pulse sequence diagram for a Selective spectral Displacement Projection – Magnetic
Resonance Elastography sequence with motion-encoding gradient frequencies of
500, 1000, and 1500 Hertz and cycle numbers of 1, 2, and 3 cycles, respectively. ………. 35
14. Pulse sequence diagram for a Selective spectral Displacement Projection – Magnetic
Resonance Elastography sequence with motion-encoding gradient frequencies of
500, 1000, and 1500 Hertz and cycle numbers of 3, 6, and 9 cycles, respectively. ………. 36
15. Pulse sequence diagram for a SampLe Interval Modulation – Magnetic Resonance
Elastography sequence with all motion-encoding gradient offsets superimposed. ……….. 41
16. Each of eight motion-encoding gradient start times in a SampLe Interval Modulation -
Magnetic Resonance Elastography sequence depicted in real time and in discretized
time. ……………………………………………………………………………………….. 43
17. One sGRAD_PULSE_SIN structure with Amplitude, Ramp Up Time, Ramp Down
Time, and Duration defined. ……………………………………………………………… 47
18. One period of a sinusoidal motion-encoding gradient composed of two
sGRAD_PULSE_SIN structures with opposite amplitudes. ……………………………… 48
19. One sGRAD_PULSE structure with Amplitude, Ramp Up Time, Ramp Down Time,
and Duration defined. ……………………………………………………………………… 49
20. One period of a trapezoidal motion-encoding gradient composed of two
sGRAD_PULSE structures with opposite amplitudes. ……………………………………. 50
21. Quarter-period motion-encoding gradient pulses defined for the leading end of a
sinusoidal motion-encoding gradient, the trailing end of a sinusoidal motion-encoding
gradient, and either end of a trapezoidal gradient. ………………………………………… 51
22. One period of a flow-compensated motion-encoding gradient composed of two
quarter-pulses and one half-pulse for sinusoidal and trapezoidal motion-encoding
gradients. …………………………………………………………………………………... 51
23. Applied frequency versus input frequency for motion-encoding gradients when using
fSDSRoundUpGRT() and fSDSRoundDownGRT() rounding functions. ………………… 57
viii
LIST OF FIGURES (continued)
FIGURE PAGE
24. Applied frequency versus input frequency for motion-encoding gradients when
alternating use of both fSDSRoundUpGRT() and fSDSRoundDownGRT() rounding
functions. …………………………………………………………………………………... 57
25. Maximum erroneous frequency trend lines for motion-encoding gradients when using
fSDSRoundUpGRT(), fSDSRoundDownGRT(), and alternating both rounding
functions……………………………………………………………………………………. 58
26. Experimental setup to produce physical vibration in test sample. ………………………… 59
27. Schematic for experimental inhomogeneous gel sample. …………………………………. 60
28. Orientation of vibration actuation for experimental gel sample as it relates to slice
direction in scanner. ………………………………………………………………………. 61
29. Acquired displacement images for SampLe Interval Modulation – Magnetic Resonance
Elastography sequence in slice, read, and phase directions, and for one-dimensional
Magnetic Resonance Elastography sequences in slice, read, and phase directions. ………. 66
30. Calculated complex shear modulus images using curl-operator inversion method for
SampLe Interval Modulation – Magnetic Resonance Elastography and one-dimensional
Magnetic Resonance Elastography. ……………………………………………………….. 67
31. Acquired displacement images for Selective spectral Displacement Projection –
Magnetic Resonance Elastography sequence in slice, read, and phase directions,
and for one-dimensional Magnetic Resonance Elastography sequences in slice,
read, and phase directions. ………………………………………………………………... 68
ix
LIST OF ABBREVIATIONS
1D One-dimensional
3D Three-dimensional
ADC Analog-to-Digital Converter
.CPP File extension for programs designed using the C++ language
FID Free Induction Decay
FOV Field of View
GE Gradient Echo
IDEA Integrated Development Environment for Applications (Siemens Software)
MRE Magnetic Resonance Elastography
MRI Magnetic Resonance Imaging
NMR Nuclear Magnetic Resonance
RF Radio Frequency
RTE Real-Time Event
RTEB Real-Time Event Block
SDP Spectral selective Displacement Projection
SE Spin Echo
SLIM SampLe Interval Modulation
TE Echo Time
TE/2 One-half Echo Time
TTL Transistor-transistor logic
TR Repetition Time
x
SUMMARY
Magnetic Resonance Elastography is a non-invasive method of acquiring biological tissue
mechanical stiffness data. Such data is useful in the detection of a number of diseases and medical
conditions. By applying a physical vibration to a tissue sample, encoding this displacement into the phase
of the nuclear magnetic resonance signal, and applying relevant mechanical wave theory, a map of the
complex shear modulus of a tissue can be produced.
Conventionally, three-dimensional displacement data requires three subsequent Magnetic
Resonance Elastography scans. The novel techniques of Selective spectral Displacement Projection and
SampLe Interval Modulation, recently developed at the University of Illinois at Chicago, encode three-
dimensional displacement data in a single scan, thereby decreasing the total required time by a factor of
three. Such an improvement is relevant to a clinical application, where the delay between subsequent
scans can include alterations to the physiological state and bulk motion of the experimental target,
potentially resulting in inaccurate three-dimensional data. Further, the rapid data acquisition may improve
clinical acceptance of MRE.
In this study, Selective spectral Displacement Projection and SampLe Interval Modulation pulse
sequences were designed using the previously-untested system of a Bruker 7T Clinscan MRI Scanner
operating with Siemens IDEA VB15 software. The novel sequences were tested with scans of an
inhomogeneous EcoFlex gel sample. The acquired displacement images and calculated shear modulus
elastograms of the novel sequences were found similar to those of related conventional Magnetic
Resonance Elastography scans, verifying the effectiveness of the novel sequences. Additionally, adding to
the existing implementations of the novel sequences, this study serves as further evidence that these
sequences can be implemented successfully independent of the modality.
1
I. INTRODUCTION
One of the most prevalent diagnostic techniques available to doctors is simple manual palpation.
By touching and deforming a tissue, the mechanical properties of the tissue are being tested. A tissue’s
stiffness can, in many cases, be related to its health. Several illnesses have been shown to correlate to a
change in an affected tissue’s mechanical stiffness. One well-defined case of this effect is found in human
diffuse liver disease. As the disease progresses, increasing liver fibrosis has been shown to cause an
increase in the liver’s mechanical stiffness, boosting both shear and Young’s moduli [1] [2].
Traditionally, a tissue’s stiffness is tested through tactile methods: the physical touching and
manipulation of a tissue by a doctor. This method has several drawbacks. Palpation cannot access regions
deep in the body, making this technique unusable for many organs. Furthermore, the method is
subjective; its accuracy is highly dependent on the experience and judgment of the doctor, and offers no
quantitative results.
Elastography is the mapping of the mechanical properties of a material. Magnetic Resonance
Elastography (MRE) is the acquisition of this mechanical data through the non-invasive technique of
Magnetic Resonance Imaging (MRI). Through the application of both forced vibrations and carefully
selected oscillating magnetic gradients to a tissue during an MRI scan, physical shear waves propagating
through the tissue can be detected and measured. These waves can be analyzed through relevant
mechanical models to calculate mechanical stiffness values for the tissue. In short, MRE allows for the
quantitative measurement of the mechanical properties of a material, giving doctors another indicator for
a patient’s health.
MRE requires three distinct steps. The first is the introduction of a physical vibration to a target
material. The second is the measurement of this vibration as it propagates through the material using
MRI. Finally, the spatio-temporal characteristics of this vibration are used to calculate the mechanical
properties of the material.
2
When a physical vibration is applied to the target material, the protons embedded in the material’s
atoms are displaced through harmonic oscillation. MRE uses an MRI machine as a tool to visualize this
displacement. The movement must be encoded into the Nuclear Magnetic Resonance (NMR) signal of the
protons.
Accurate mechanical stiffness analysis requires three-dimensional displacement data, but
conventional MRE scans acquire one direction of displacement data at a time. To acquire a fully three-
dimensional data set using conventional MRE, three subsequent scans must be performed, which is a
lengthy process and risks inaccuracy due to potential changes in the physiological state and bulk motion
of the target between scans.
Two novel techniques for acquiring three directions of displacement data simultaneously are
implemented in this study: Selective spectral Displacement Projection (SDP) – MRE, introduced by
Yasar et al. [3], and SampLe Interval Modulation (SLIM) – MRE, introduced by Klatt et al [4]. These
novel techniques have previously been introduced and tested using Agilent MRI software on an animal
scanner [3] [4], and Siemens IDEA VB17 on a 3T human scanner [5], but the modality-independence of
the novel MRE techniques has not yet been verified for all systems.
The purpose of this study is to apply the novel three-dimensional MRE techniques of SDP-MRE
and SLIM-MRE to a previously-untested Siemens IDEA VB15 pulse sequencing system on a 7T Bruker
Clinscan MRI scanner. By comparing the displacement images and calculated mechanical stiffness maps
generated by the novel techniques to those of conventional MRE sequences, the successful
implementation of the novel three-dimensional MRE sequences as time-saving innovations is verified.
Additionally, success in applying the novel techniques to another MRI pulse sequencing software further
confirms their modality-independence.
3
II. CONCEPTUAL FRAMEWORK
A. Magnetic Resonance Imaging Background
1. Nuclear Magnetic Resonance
NMR is a process dependent on the naturally-occurring spin of protons. Because protons
are charged particles, their spin produces an associated magnetic moment. The collection of protons
spinning in a tissue produces a measureable macroscopic magnetization vector. The dynamics of this
macroscopic magnetization vector are described by the Bloch Equations, summarized in equations 1.1,
1.2, and 1.3:
(1.1)
(1.2)
(1.3)
where M(t) is the macroscopic magnetization vector with x-, y-, and z-direction
components of Mx(t), My(t), and Mz(t), respectively, B(t) is the magnetic field experienced by the
magnetization vector, γ is the Gyromagnetic Ratio (a property unique to the nucleus), T2 is the transversal
(x- and y-direction) relaxation time, T1 is the longitudinal (z-direction) relaxation time, and M0 is the
macroscopic magnetization at thermal equilibrium [6].
One important result of the Bloch equation is that when a magnetic field is applied to a
tissue, the material’s naturally misaligned magnetization vectors tend to align along and undergo
precession about the direction of the field. In the presence of a magnetic field vector B, nuclear spins
tend to undergo precession about the B axis with the “Larmor Frequency”, given by equation 1.4:
(1.4)
4
where ω is the Larmor Frequency [6].
An NMR signal is generated by first exciting the magnetization vectors with a radio-
frequency (RF) pulse. This pulse, when produced with a frequency equal to the Larmor Frequency and in
a direction perpendicular to the applied magnetic field vector, causes the target magnetization vectors to
misalign with the applied magnetic field vector (or “tilt”) by an angle dependent on the duration and
amplitude of the RF pulse, as given in equation 2:
(2)
where α is the flip angle of the magnetization vector, B1 is the strength of the magnetic
field produced by the RF pulse, and τRF is the duration of the RF pulse [6].
After the application of the RF pulse, the magnetization vectors will “relax” by returning
to their positions before the RF pulse, that is, aligned with the applied magnetic field B0. This relaxation
produces an RF signal, known as the Free Induction Decay (FID), which can be detected by MRI
equipment. Each magnetization vector produces a signal with a magnitude and phase.
Figure 1 displays the macroscopic magnetization vector of a material as it is placed in an
applied magnetic field, excited with an RF pulse, and allowed to relax to its original state, producing a
FID signal.
5
Figure 1. The response of magnetic spin vectors with no external field (a), with an external
magnetic field applied (b), during excitation by an RF pulse (c), and emitting an FID signal during
relaxation (d).
2. Magnetic Field Gradients
MRI records the phase and magnitude of NMR signals emitted by precessing
magnetization vectors after being excited by an RF pulse to produce a spatially-resolved image of the
target material. Equation 1 shows that the precession frequency of a nuclear spin is directly proportional
to the strength of the applied magnetic field. By changing the magnitude of the magnetic field vector, the
frequency of nuclear spin precession can be controlled, and thus the phase of the nuclear spin vector can
be controlled. This process is performed in conventional MRI using magnetic field gradients. When
applied within the permanent magnetic field vector B0, a magnetic field gradient vector K alters the total
magnetic field strength. Magnetic field gradients produce a magnetic field with a spatially-dependent
magnitude, as described by equation 3.1:
(3.1)
where B(r) is the spatially-dependent total magnetic field vector, B0 is the permanent
magnetic field vector, K is the magnetic field gradient vector, and r is the spatial position vector.
6
The Larmor Frequency of a nuclear spin can be controlled through the strength of the
magnetic field gradient and the position of the nuclear spin. Equations 1.4 and 3.1 can be combined to
form equation 3.2, which gives the Larmor Frequency ω of nuclear spins within a permanent magnetic
field B0 and magnetic field gradient K.
(3.2)
3. Signal Echoes
The FID signal produced by a material as its nuclear spins relax after RF excitation is
very brief, often on the order of microseconds. This is due to the existence of magnetic field
inhomogeneities, which tend to cause nuclear spins in close proximity to undergo precession with slightly
different frequencies, resulting in a larger distribution of nuclear spin vectors and, thus, a smaller
macroscopic magnetization vector. To remove this inhomogeneous dephasing, it is necessary to produce
an “echo” of the signal, occurring later than the original FID signal and with lower amplitude.
One way to produce an NMR signal echo is to apply a second RF pulse after the initial
excitation RF pulse. This second pulse, known as the “refocusing pulse”, is conventionally calibrated via
equation 2 to produce a flip angle of 180° (π radians). The refocusing pulse has the effect of inverting the
magnetization vectors. This inversion causes a phenomenon known as the “spin echo” of an FID, as
summarized in Figure 2.
7
Figure 2. The state of a macroscopic magnetization vector after excitation by a 90-degree RF pulse
(a), after magnetization vectors have dephased due to local field inhomogeneities at time TE/2 (b),
after application of a 180-degree refocusing RF pulse (c), and after magnetization vectors have
rephrased to produce a signal echo at time TE (d).
After an excitation pulse, the affected magnetization vectors are rotated away from the
permanent magnetic field vector, where they proceed to begin relaxation while undergoing precession
about the permanent field vector. The decay of the subsequent NMR signal is caused by two effects: static
effects, dephasing caused by magnetic field inhomogeneities, and spin-spin interactions, the relaxation
effects of interactions between magnetic spins. The former causes some magnetization vectors to go out
of phase with each other, the effect of which is reversed by inverting the magnetization vectors with a
180° refocusing RF pulse. This allows magnetization vectors dephased by magnetic field inhomogeneities
to rephase, producing a spin echo signal at echo time (TE). The relaxation effects due to spin-spin
interactions, however, cannot be reversed in this way.
Another means of producing an NMR signal echo is through the use of magnetic
gradients rather than a refocusing RF pulse. Using the “gradient echo” technique involves the application
of a dephasing magnetic gradient immediately following the excitation RF pulse, followed by a
rephrasing magnetic gradient to generate a signal echo. After the dephasing gradient selectively changes
the frequencies of precession for the nuclear spin vectors, forcing the decay of the NMR signal, the
8
rephrasing gradient reverses this effect by giving opposite precession frequencies to the nuclear spin
vectors, resulting in the rephrasing of the magnetization vectors. The time between excitation pulse and
generated echo is known as the echo time, TE. The gradient echo process produces an NMR signal echo
faster than the spin echo process, but does not remove the dephasing effect due to magnetic field
inhomogeneities.
Figure 3 compares the NMR signal characteristics of FID, spin echo, and gradient echo
processes. The signal decay due to both static field inhomogeneities and spin-spin interactions is known
as T2* decay, whereas the signal decay due to static field inhomogeneities alone is known as T2 decay.
The forced decay used to produce a gradient echo is known as TG decay.
Figure 3. The NMR signal intensity decay during Free Induction Decay (a), during a Spin Echo (b),
and during a Gradient Echo (c).
4. Pulse Sequence Diagrams
The acquisition of an image using MRI is possible through the coordinated operation of
magnetic gradients and RF coils. The precise timing of these elements is controlled by a piece of software
called a pulse sequence. A pulse sequence, when input the specific parameters of a desired scan, outputs
9
directions to the scanner of when, how long, and with what strength to apply magnetic gradient and RF
pulses. Magnetic gradient pulses can be applied in three mutually orthogonal directions, known as “slice”,
“read”, and “phase”. Acquired images are always displayed in the read-phase plane, with subsequent
images aligned along the slice direction. By convention, the spatial directions corresponding to slice,
phase, and read are z, x, and y, respectively. Figure 4 depicts the directions as they are conventionally
defined in a scanner and image. By default, the slice direction is parallel to the permanent magnetic field
vector, but this is not required.
MRI scanners acquire data in complex-valued “k-space”, which is the two-dimensional
Fourier Transform of the image being measured. The two-dimensional Fourier Transform is given in
equation 4.1. An example of k-space data, which is in the frequency domain, is given in Figure 4b. To
convert raw k-space data into an image, the inverse two-dimensional Fourier Transform is used, as given
by equation 4.2. An example of the result is the image given in Figure 4c.
Figure 4. The slice, read, and phase directions as they relate to a specimen in an MRI scanner (a), to the acquired k-space data (b), and to an image produced through the inverse two-dimensional Fourier Transform of such k-space data (c).
10
(4.1)
(4.2)
MT* is the spatially-dependent transverse magnetization vector, S is the k-space-
dependent acquired signal, and kx and ky are the spatial frequencies in x and y directions, respectively.
Figure 5 is a “pulse sequence diagram”, a means of displaying the relative timing of the
magnetic field gradients, RF elements, and associated equipment signals for a specific pulse sequence.
The horizontal axis is time, whereas each of the rows represents a different element-controlling signal.
The pulse sequence diagram depicts a single echo, which corresponds to a single pixel-wide “line” of
image data. This sequence must be repeated for each line required to produce an image of the desired
resolution. The time required for all MRI components during the acquisition of a single line of k-space
data is called Repetition Time (TR).
Figure 5 depicts a Gradient Echo (GE) pulse sequence. The top row describes the RF
signal. In a GE sequence, one RF pulse per echo is required. This excitation pulse usually produces a flip
angle of 90° or less via equation 2.
The second, third, and fourth rows depict the magnetic field gradients for the slice, phase,
and read directions, respectively. Three slice direction pulses per echo are required. The first pulse, the
“slice select” gradient, is positive in polarity and coincides with the excitation RF pulse. The purpose of
this pulse is to spatially select the slice being excited with the RF pulse via equation 3.2. The second slice
direction gradient, the “slice rephase” gradient, is negative in polarity and occurs immediately after the
slice select gradient. This second gradient has the purpose of re-aligning the phase of the precession of the
nuclear spins within the slice. The third pulse, the “spoiler” gradient, occurs after the gradient echo. This
gradient destroys any residual phase alignment in the slice direction. The purpose of the spoiler is to
rapidly dephase the nuclear spins to prepare for another repetition of the pulse sequence
11
The third row depicts the phase direction magnetic field gradient. Two phase direction
pulses per echo are required. The first pulse, occurring directly after the slice select gradient, is the “phase
encode” gradient. This pulse encodes the phase-direction spatial location of each nuclear spin into its
precession phase. By convention, phase-direction encoding usually corresponds to the x-direction location
of the affected nuclear spins. After a scan has been completed, the phase of each NMR signal can be used
to assign its phase-direction spatial location. For each echo, a unique amplitude of phase encoding
gradient is used. Each line of k-space data corresponds to a unique precession phase, indicating a unique
phase-direction spatial location. In the pulse sequence diagram, this is represented by the phase encode
gradient appearing with multiple amplitudes visible. The second phase direction pulse is the “phase
rewind” gradient. This gradient is identical to the phase encode gradient, except that it has an opposite
amplitude. The phase rewind gradient eliminates the phase encoding of the nuclear spins in preparation
for the next phase encoding gradient.
Figure 5. A Gradient Echo MRI pulse sequence with Echo Time TE defined.
12
The fourth row depicts the read direction magnetic field gradient. Two read direction
pulses per echo are required. The first pulse, coinciding with the phase encode gradient, is the “read
dephase” gradient. This pulse rapidly removes the alignments of the nuclear spins’ precession phases in
the read direction, which conventionally corresponds to the y-direction location of the affected nuclear
spins. Following the read dephase pulse is a gradient pulse of longer duration and smaller amplitude. This
pulse, the “readout” gradient, has three purposes. First, it rephrases the frequency-direction phases of the
nuclear spins’ precession, producing the gradient echo signal halfway through its duration from which the
NMR data can be extracted. In addition, the readout gradient encodes the read-direction spatial location of
each nuclear spin in its precession frequency via equation 3.2. After a scan has been completed, the
frequency of each NMR signal can be used to assign its read-direction spatial location. Finally, after the
gradient echo has occurred, the second half of the readout gradient removes any residual phase encoding,
in preparation for the next echo.
The fifth row depicts the Analog-to-Digital Converter (ADC) signal. Only one ADC
pulse is required per echo: to turn on the ADC when acquiring the NMR signal via the RF receiver.
The sixth row depicts the NMR signal produced by the target material. Immediately
following the excitation pulse, the material produces a signal, which decreases in amplitude
exponentially. The gradient echo is visible after the application of the readout gradient. The peak signal
occurs halfway through the readout gradient at Echo Time TE.
Figure 6 is the pulse sequence diagram for a Spin Echo (SE) sequence. The main
difference between a SE sequence and a GE sequence is the 180° refocusing RF pulse required to
generate a spin echo. This pulse is visible on the first row of Figure 6. The refocusing pulse occurs at a
time halfway between the excitation RF pulse and the signal echo, called TE/2. Due to the inversion of
magnetization vectors caused by the refocusing pulse, the polarities of both the phase rewind and readout
gradients are reversed from their values in the GE sequence. Additionally, a slice-direction gradient
13
associated with the refocusing RF pulse is required. This pulse, called the “slice refocus” gradient, is
visible on the second row of Figure 6. Its amplitude is set so that the 180° RF pulse applies only to the
desired slice via equation 3.2.
B. Mechanical Wave Theory Background
1. Equation of Motion of a Continuum
The stiffness of a material is directly related to the wave properties of a mechanical wave
propagating through it. The deflection caused by a mechanical wave is related to the material’s resistance
Figure 6. A Spin Echo MRI pulse sequence with both half-echo times defined.
14
to shear and longitudinal deformation. When the material is considered to be linearly elastic and isotropic,
this relationship is summarized by the Navier equation, as derived by Chandrasekharaiah et al. [7] in
equation 5.1:
(5.1)
where ρ is the density of the material, u is the three-dimensional deflection vector of the
wave as it propagates through the material, is the divergence operator, Δ is the Laplace operator,
defined by equation 5.2, and λ and μ are the first and second Lamé Parameters, respectively. Each of the
variables in equation 5.1 are spatially-dependent.
(5.2)
The two Lamé Parameters are used to quantify the mechanical stiffness of a material: the
second Lamé Parameter, μ, is also referred to as the shear modulus of the material; the first Lamé
Parameter, λ, is a factor with magnitude correlated to a material’s resistance to compression (longitudinal)
waves. The Lamé Parameters and Young’s Modulus E are related by equation 6.
(6)
Biological tissues are generally regarded as viscoelastic materials. Therefore, it makes
sense to convert equation 5.1 to the frequency domain to account for the frequency-dependent
characteristics of the Lamé Parameters. The Fourier Transform, given in equation 7, is used to convert a
temporal function to the frequency domain [8].
(7)
F(ω) is the frequency-domain function corresponding to time-domain function f(t) and i
is the imaginary unit. The Fourier transform allows frequency-domain counterparts to be found for most
15
time-domain functions. Table I gives a number of frequency-domain counterparts to useful time-domain
functions, computed using equation 7.
Table I
Time-Domain functions and their Frequency-Domain counterparts, obtained via the Fourier
Transform of equation 7
Time-Domain function Frequency-Domain function
Applying the Fourier Transform pairs in table I allows the conversion of equation 5.1 to
the frequency domain. This yields equation 8:
(8)
where ω is the angular frequency of the mechanical vibration, GL and GS are the complex
frequency-dependent moduli of the first and second Lamé Parameters, respectively, and U is the
frequency-dependent deflection vector. The real parts of the complex moduli are related to a material’s
tendency to store energy, and the imaginary parts are related to its damping characteristics. Each of the
variables in equation 8 are spatially-dependent.
16
2. Mechanical Stiffness Retrieval
The retrieval of mechanical properties from equation 8 can be performed with or without
consideration of the first Lamé Parameter complex modulus GL. Sinkus et al. demonstrated a technique
involving the vector curl operator [9]. The curl operator can be used to define vector Q:
(9)
which simplifies equation 8 to equation 10, effectively removing the longitudinal wave
contribution to the Navier equation.
(10)
Equation 10 can be used to solve for the complex shear modulus GS using the method of
Least-Squares solution to the overdetermined problem, resulting in equation 11:
(11)
where GS and Q are both functions of space and angular frequency ω, and superscript T
and -1 denote the transpose and inverse of a vector, respectively.
Equation 11 provides a method for obtaining the complex shear modulus GS from
displacement data, but if compression waves are not ignored, the solution can follow the method
presented by Oliphant et al. [10], as follows. Rearranging equation 8 yields equation 12:
17
or,
where:
(12)
where U1, U2, and U3 are the first, second, and third orthogonal components of deflection
vector U, respectively, and x1, x2, and x3 are the first, second, and third orthogonal spatial directions,
respectively. Equation 12 is solved using the method of Least-Squares solution to the overdetermined
problem to yield equation 13.
(13)
Equation 13 can be used to solve for both the complex shear modulus GS and the complex
first Lamé Parameter modulus GL.
If only two-dimensional data is available, the further assumption must be made that the
material is incompressible and that deflection occurs in a single direction. This simplification leads to the
reduced form of equation 8 given in equation 14.
(14)
18
Equation 14 decouples the spatial directions and the complex shear modulus can be
solved for each component of the deflection vector. This is possible through the scalar Helmholz
Inversion shown in equation 15:
(15)
where i = 1, 2, or 3.
Equations 11, 13, and 15 represent three different methods of obtaining complex shear
modulus values from physical deflection vectors, all being conventional techniques used in today’s MRE
research.
C. Magnetic Resonance Elastography Background
MRE combines the technique of MRI with mechanical wave theory to evaluate the stiffness properties of
a given tissue. To generate an elastogram, the following steps are taken:
1. A mechanical vibration is applied to a tissue
2. The displacement caused by the mechanical wave is encoded into the phase of the NMR signal during an
MRI scan
3. The resulting displacement map is used to determine mechanical stiffness properties of the tissue
Mechanical vibration application and the calculation of mechanical stiffness properties are
generally performed in MRE with standardized techniques. This study deals primarily with the particulars
of encoding motion into the phase of an NMR signal, and the bulk of the work herein is towards this end.
19
1. Phase Accumulation
Equation 1.4 indicates that the precession frequency of magnetic spins depends on the
magnetic field strength. When magnetic field gradients are applied for a finite time, the phase shift in the
NMR signal is given by equation 16:
(16)
where φ is the phase shift of the NMR signal, K(t) is the time-dependent magnetic field
gradient vector, u(t) is the deflection vector of the nuclear spins, and τK is the duration of application of
the magnetic field gradient [11]. Equation 16 demonstrates that the physical movement of a nuclear spin
causes a phase shift in the NMR signal. In MRE, the magnetic fields are varied in a controlled manner to
ensure that oscillating nuclear spins experience a position-dependent Larmor Frequency.
For the case of an oscillating magnetic field gradient vector, the duration of application
τK can be rewritten in the form of equation 17.
(17)
Here, τ is the period of oscillation and n is the number of cycles of oscillation applied.
To convert a standard MRI sequence to an MRE sequence, oscillating Motion-Encoding
Gradients (MEGs) must be added. These gradients have the effect of encoding physical motion into the
phase of the magnetization vectors. To avoid encoding constant displacement into the NMR phase, an
MEG must obey the “0th-moment-nulled” condition: it must be an oscillating function which, over any
integer number of periods, has equal positive and negative areas. This condition is summarized by
equation 18:
(18)
20
where τ is the period of the gradient’s polarity-switching oscillation and N is any integer.
Such a gradient function is shown in Figure 7 with N = 2 and an amplitude of A, and with a single period
of oscillation described by equation (19).
for
(19)
Figure 7. A zeroth-moment-nulled function K(t) with two periods of oscillation and amplitude of A.
The process described by equation 16 allows intra-voxel coherent motion to be encoded
into the phase of the NMR signal. Conventionally, two types of MEGs are used: sinusoidal and
trapezoidal pulses.
Sinusoidal gradient pulses provide a magnetic field gradient which is spatially constant
and shaped like a sinusoidal function in time, as described by equation 20:
21
(20)
where the amplitude, frequency, and initial phase (with respect to an arbitrary but fixed
time point) of the gradient waveform are represented by K0, ωK, and , respectively. Considering the
physical motion of nuclear spins as harmonic oscillation yields equation 21:
(21)
where the amplitude, frequency, and initial phase of the physical vibration are
represented by Yn, ωn, and , respectively.
The phase accumulated by a nuclear spin undergoing harmonic motion in a magnetic
field gradient is given by equation 16. This equation is adapted to account for an arbitrary start time of the
magnetic field gradient K, yielding equation 22:
(22)
where s is the start time of the magnetic field gradient. To obey the 0th-moment-nulled
condition of equation 18, the duration of magnetic field gradient application τK is taken to be an integer
multiple of the gradient’s period of oscillation, as in equation 23:
(23)
where q is an integer number of cycles of the magnetic field gradient with angular
frequency .
When the magnetic field gradient takes the form of a sinusoidal function as in equation
20 with a phase of zero at its onset, which is provided by the condition: , then equation 22 can
be solved with the method used by Muthupillai et al. [12] for the accumulated phase given by equation
24:
22
where
and
for
(24)
Equation 24 indicates that the accumulated phase in the NMR signal is proportional to the
number of applied cycles of the magnetic field gradient q, the amplitude of the magnetic field gradient K0
and the physical vibration amplitude Yn. Additionally, the sinusoidal function in equation 24 involving
MEG start time s gives the accumulated phase a spatial dependence of equal frequency to that of the
physical vibration. The spatially-dependent phase accumulation is therefore a “snapshot” (a single
moment in time) of the physical wave with a certain phase offset as it propagates through the
target material.
In equation 24, the factor ξ is called the “encoding efficiency”, as it represents the ratio of
phase accumulation amplitude to physical vibration amplitude. As equation 24 indicates, the encoding
efficiency exists for two cases: matching or mismatched MEG and physical vibration frequencies. For
matching MEG and vibration frequencies, the encoding efficiency is linearly proportional to the strength
and number of cycles of the MEG, and inversely proportional to the vibration frequency. For mismatched
MEG and vibration frequencies, the encoding efficiency has a sinusoidal dependence on the product of
the number of MEG cycles, pi, and the ratio of vibration to MEG frequencies. This results in an encoding
efficiency function as it appears in Figure 8.
Figure 8 demonstrates that, when sinusoidal MEGs are used, no vibration is encoded for
frequencies that obey equation 25, the “filter condition” [13]:
23
when
(except: ) (25)
where N is any integer.
Figure 8. The encoding efficiency of sinusoidal motion-encoding gradients for vibration frequencies
surrounding the motion-encoding gradient frequency.
Trapezoidal gradient pulses provide a magnetic field gradient shaped like a trapezoidal
function, oscillating regularly between positive and negative amplitude. Figure 9 is an example of such a
function.
24
Figure 9. A realistic trapezoidal motion-encoding gradient with finite ramp up, ramp down, and
flat top times.
A practical trapezoidal MEG must have finite rise and fall times, but for the evaluation of
a trapezoidal encoding efficiency, the rise times are considered instantaneous. With this simplification,
the trapezoidal MEG takes the form of equation 26.
(26)
where τK is the MEG period of oscillation.
To obtain the trapezoidal encoding efficiency, equation 22 can be solved for the
accumulated phase when the magnetic field gradient takes the form of equation 26. When the magnetic
25
field gradient obeys the 0th-moment-nulled condition as in 23, the accumulated phase takes the form of
equation 27:
where
and
for
(27)
Equation 27 gives the encoding efficiency for trapezoidal MEGs for two cases: matched
and mismatched MEG and vibration frequencies. For matched frequencies, the trapezoidal encoding
efficiency is identical to the sinusoidal encoding efficiency besides a factor of
. This is the primary
benefit to using a trapezoidal MEG instead of a sinusoidal MEG. For mismatched MEG and vibration
frequencies, however, the trapezoidal encoding efficiency function has sinusoidal and trapezoidal factors.
The trapezoidal encoding efficiency can be solved for several cases using L’Hôpital’s Rule [14], shown in
equation 28:
(28)
where f(x) and g(x) represent functions of independent variable x, and f’(x) and g’(x)
represent the first derivatives with respect to x of f(x) and g(x), respectively. Using L’Hôpital’s Rule, the
trapezoidal encoding efficiency of equation 27 can be solved for several cases, as shown in equations 29,
30, and 31:
when
even integer (29)
when
odd integer (30)
26
when
any integer (31)
The trapezoidal encoding efficiency is similar to the sinusoidal encoding efficiency for
mismatched MEG and vibration frequencies, besides one major difference. Equation 30 demonstrates the
trapezoidal encoding efficiency is non-zero for cases where the ratio of vibration to MEG frequency is an
odd integer. Unlike sinusoidal MEGs, trapezoidal MEGs encode vibration frequencies of higher odd-
integer harmonics.
Additionally, MEGs can be flow-compensated. Flow-compensated MEGs are designed to
resist encoding constant-velocity motion. This is useful for situations where encoding of vibration is
desired, but constant-velocity flow (such as the flow of blood through veins) should be suppressed.
An MEG is flow-compensating if it is 1st-moment-nulled. As with the 0
th-moment-nulling
condition of equation 18, a function K(t) is said to be 1st-moment-nulled if it obeys equation 32:
(32)
where τ is the period of the gradient’s polarity-switching oscillation and N is any integer.
A flow-compensated MEG must be 0th- (equation 18) and 1
st-moment-nulled (equation
32). One function that fills both conditions is a cosine-shaped function, like the one described by equation
33:
(33)
where K0 and ωK are the amplitude and frequency of the function, respectively.
Effectively, a cosine-shaped MEG is flow-compensated, thereby suppressing constant-velocity flow
during motion encoding.
27
2. Magnetic Resonance Elastography Pulse Sequence
To construct an MRE pulse sequences, MEGs are added to an MRI pulse sequence. To
encode motion, the MEGs must be active after the excitation RF pulse and before the ADC readout
signal. Additionally, for SE sequences, the MEGs cannot be active during the refocusing RF pulse, as this
would result in an incorrect location of the refocusing event.
Figure 10 is a Gradient Echo Magnetic Resonance Elastography (GE MRE) pulse
sequence diagram for motion encoding in the slice direction, using three cycles of a sinusoidal MEG.
Figure 10. A Gradient Echo Magnetic Resonance Elastography pulse sequence diagram with Echo
Time TE defined.
28
The one-dimensional motion encoding is performed by the sinusoidal MEG visible on the
second row of Figure 10, corresponding to the slice direction magnetic field gradient. The effect of this
gradient is to encode the slice direction motion (out-of-plane with respect to the magnitude image
obtained by the sequence) into the phase of the generated NMR signal. Because only a single direction of
motion is encoded, this type of sequence is further referred to as One-Dimensional Magnetic Resonance
Elastography (1D MRE). The physical vibration of the sample is visible in the last raw of Figure 10.
A similar 1D MRE pulse sequence can be made for either of the other two directions by
applying MEG along a different directional magnetic field gradient. In this manner, a phase-direction or
read-direction 1D MRE sequence is attainable. Each 1D MRE sequence encodes a single direction of
deflection into the phase of the NMR signal.
Figure 11 is a Spin Echo MRE (SE MRE) pulse sequence diagram for a motion encoding
gradient in the slice direction, using four cycles of a sinusoidal MEG.
As in the GE MRE pulse sequence diagram of Figure 10, the SE MRE pulse sequence
diagram in Figure 11 includes a sinusoidal MEG on the second row, indicating one-dimensional motion
encoding in the slice direction. Unlike the GE MRE sequence, however, SE MRE requires a 180°
refocusing RF pulse to generate the spin echo. The MEG cannot be active during this RF pulse, as it
would cause the refocusing event to occur in an incorrect location. As a result, the MEG is split so that
even portions occur before and after the refocusing RF pulse. For clarity, the portions of the MEG before
and after the refocusing RF pulse will be referred to as the “first run MEG” and “second run MEG”,
respectively.
29
The splitting of the MEG is done to minimize TE. Because the refocusing pulse must
occur at exactly TE/2, it is most efficient to apply exactly half of the MEG cycles as the first run MEG,
and the other half of the cycles as the second run MEG. Additionally, care must be taken to ensure that
the MEG provides continuous motion encoding before and after the refocusing pulse. Because the
refocusing RF pulse inverts the magnetization vectors, the second run MEG must also be inverted to
continue encoding vibration into the phase of the signal. This is visible in the MEG of Figure 11. The
second run MEG is applied with a reversed polarity to that of the first run MEG: in the case of a
sinusoidal function, this is equivalent to a phase offset of π radians. Without this reversal of MEG
polarity, the second run MEG would remove the accumulated phase of the first run MEG. Additionally,
care must be taken to ensure that the second run MEG is initiated at a point where the physical vibration
Figure 11. A Spin Echo Magnetic Resonance Elastography pulse sequence diagram with both half-echo times and the motion-encoding gradient gap defined.
30
is at the same phase as it was when the first run MEG was stopped. This is ensured by requiring the time
span between the end of the first run MEG and beginning of the second run MEG, referred to as the
“MEG Gap”, is an integer multiple of periods of the physical vibration. This condition is expressed in
Figure 11.
3. Data Sampling
The Fourier Transform, shown in equation 7 in its continuous form, is also useful in the
analysis of discrete functions. The discrete Fourier Transform is used to decompose a function into its
spectral components. Among other uses, this technique is used in conventional MRE to remove any
constant-displacement signal from a deflection vector, which would otherwise cause error in the
calculation of stiffness properties.
The discrete Fourier transform is useful for converting a signal in the time domain to one
in the frequency domain. To acquire a time-depended displacement vector using MRE, several scans are
performed, each with the MEG applied at a different phases of the physical vibration. Conventionally, this
is done by shifting the starting time of the MEG relative to the physical vibration, though in theory, it
would also be possible to vary the phase of the MEG at its onset for a fixed onset time to achieve the
same effect. This will result in multiple displacement vector images, each representing a unique
“snapshot” of the wave as it prorogates the material, as demonstrated in Figure 12. Together, these
snapshots represent a time-dependent displacement field.
Conventionally, a time-dependent displacement vector is given full temporal resolution
by acquiring N snapshots each with an MEG starting time shifted by Δt from the previous MEG starting
time, that add up to a complete period of the physical vibration. This allows the discrete Fourier transform
[15] to provide spectral data at each of the discrete frequencies, multiples of Δf, given in equation 34.
for
(34)
31
The discrete Fourier transform, when applied to a time-dependent displacement vector
u(t), yields a frequency-dependent displacement vector U(f). The discrete frequency corresponding to n =
0 in equation 34 is the constant-displacement signal, which does not contribute towards the calculation of
stiffness of a material. By discarding the value of U(f) at this frequency, the non-oscillating signal is
effectively removed from the displacement vector.
Figure 12. Motion-encoding gradients applied at different starting phases of the physical vibration
to produce different snapshots of the propagating wave.
In conventional 1D MRE, the discrete frequency corresponding to n = 1 in equation 34,
the base sampling frequency, is the only spectral data point of U(f) required to calculate the stiffness of
the material.
32
III. MOTIVATION
A. Three-Dimensional Encoding
Realistically, waves must propagate through any medium in three dimensions. When motion is
encoded in only a single direction, two-thirds of the wave’s behavior is lost. To encode three spatial
directions using 1D MRE, three individual scans must be completed consecutively, with the MEG applied
in one of the three directions during each scan. This has two major drawbacks. First, the time required to
acquire 3D data this way is three times longer than a single 1D MRE scan. Additionally, there is danger
of the target changing between single-directional scans, especially when the target is a biological system.
This would result in inaccuracy between results of different directions for the same target.
Two schemes for three-dimensional MRE (3D MRE) will be implemented in this study. The first,
developed by Yasar et al., is Spectral selective Displacement Projection – MRE (SDP-MRE) [3]. This
modality exploits the filter condition of equation 24 to simultaneously encode three independent
directions of displacement values using a multifrequency vibration. The second 3D MRE technique,
developed by Klatt et al., is SampLe Interval Modulation – MRE (SLIM-MRE) [4]. This modality utilizes
a monofrequency vibration and simultaneously encodes three independent directions of displacement
values by using different sample intervals, resulting in each direction encoding as a different apparent
frequency. In both SDP-MRE and SLIM-MRE, the three individual components of the displacement
vector can be obtained through decomposition of the generated NMR phase signal using a discrete
Fourier transform. Both techniques have not previously been applied to the Siemens IDEA VB15 system
used for this study.
B. Selective spectral Displacement Projection – Magnetic Resonance Elastography
SDP-MRE encodes three directions of displacement simultaneously by exploiting the filter
condition of equation 25.
33
The phase accumulation of an oscillating magnetization vector within a magnetic field gradient,
given by equation 22, can be adapted for the case where three orthogonal MEGs are applied
simultaneously [3]. This yields equation 35:
(35)
where s is the (simultaneous) start time of the three MEGs K1, K2, and K3, T is the duration of the
MEGs, and , , and represent the NMR signal phase accumulated by the MEGs corresponding to j
= 1, 2, and 3, respectively. It is further assumed that the physical vibration u(t) takes the form of the
superposition of three sinusoidal waveforms, each taking the form of equation 21, and each of the
orthogonal MEG components take the form of a sinusoidal waveform [3]. The accumulated phase of
equation 35 is therefore solved to yield equation 36:
with
(36)
where physical vibration amplitude , encoding efficiency , and phase offset are
specified for each physical vibration frequency component q and encoding direction j. In equation 36, fq
represents the frequency of the physical vibration’s qth component, and
, τj, and Nj are the amplitude,
period of oscillation, and integer number of cycles of the MEG in the jth direction, respectively.
34
As with equation 25, the filter condition can be acquired for equation 36. For the case of
mismatched MEG and vibration frequencies, no phase is accumulated for vibration frequencies
corresponding to an encoding efficiency of zero, which yields the filter condition of equation 37:
when
(37)
for all integers n besides n = Nj. In equation 37, T represents the total MEG duration [13]. The
reciprocal of T is the “base frequency” of the SDP-MRE experiment. Equation 37 enforces the additional
constraint of SDP-MRE that each of the three MEGs must be applied with a period of oscillation and
number of cycles that have a product equal to T. For each MEG frequency, equation 37 demonstrates that
all vibration frequencies that are integer multiples of the base frequency are not encoded, besides the case
where the MEG and vibration frequencies match. Through careful selection of the MEG frequencies and
number of cycles for each of the three MEG projections, it is possible for each direction to encode a
single frequency and filter all others.
Figure 13 is an SDP-MRE pulse sequence diagram based on a GE sequence. In this case, the
MEG frequencies and number of cycles are summarized in Table II. With these parameters, the base
frequency is 500 Hz. Equation 34 is used to show that, for the sequence parameters in Table II, 8
snapshots are required for the discrete Fourier transform to generate enough resolution for all three
discrete frequencies: 500, 1000, and 1500 Hz. The discrete Fourier transform of the acquired time-
dependent displacement vector provides direction-specific displacement vectors at each discrete
frequency.
35
Table II
SDP-MRE Parameters of Figure 13, including MEG frequency and number of cycles for each
projection
MEG Projection MEG Frequency MEG Number of Cycles
Slice 500 Hz 1
Read 1000 Hz 2
Phase 1500 Hz 3
Figure 13. Pulse sequence diagram for an SDP-MRE GE sequence with MEG frequencies of 500, 1000, and 1500 Hz and cycle numbers of 1, 2, and 3 cycles, respectively.
36
While the selection of SDP-MRE parameters, rather than being arbitrary, must follow the
constraints set by the filter condition, it is possible to boost the encoding efficiency of the MEGs once
values of MEG frequencies and number of cycles have been chosen. In the matching MEG and vibration
frequency case of equation 36, it is clear that encoding efficiency increases with MEG number of cycles.
By multiplying the number of cycles of each MEG component by the same integer, the encoding
efficiency can be increased without altering the ratios necessary to fulfill the filter condition. Such an
increase is made on the pulse sequence of Figure 13 by multiplying the number of MEGs by 3, as shown
in Figure 14.
Figure 14. Pulse sequence diagram for an SDP-MRE sequence with MEG frequencies of 500, 1000,
and 1500 Hz and cycle numbers of 3, 6, and 9 cycles, respectively.
37
The use of trapezoidal MEGs is possible with SDP-MRE, but one additional constraint must be
considered. Equation 30 demonstrates that, unlike sinusoidal MEGs, trapezoidal MEGs encode vibration
at frequencies that are odd integer multiples of the MEG frequency. With this consideration, trapezoidal
MEGs in an SDP-MRE sequence must be carefully chosen so that each of the three MEG frequencies
encodes only one of the three physical vibration frequencies. The example in Table II would not be
compatible for SDP-MRE with trapezoidal MEGs, as the 500 Hz MEG would encode both 500 Hz and
1500 Hz physical vibration frequencies. An example of trapezoidal-MEG-compatible SDP-MRE
parameters is given in Table III, which has a base frequency of 250 Hz.
Table III:
SDP-MRE Parameters compatible with trapezoidal MEGs, including MEG frequency and number
of cycles for each projection
MEG Projection MEG Frequency MEG Number of Cycles
Slice 750 Hz 3
Read 1000 Hz 4
Phase 1250 Hz 5
38
C. SampLe Interval Modulation – Magnetic Resonance Elastography
SLIM-MRE encodes three directions of displacement simultaneously through the use of different
sample intervals for each MEG. This causes each orthogonal displacement projection to encode with a
different “apparent frequency”, which can be retrieved through the use of the discrete Fourier transform.
In an MRE experiment, the sample interval is the time span by which the MEG’s start time is
shifted between successive snapshot acquisitions. The total number of snapshots, also called the number
of “offsets”, denoted by N, and the sample interval, denoted by Δt, dictate the discrete frequencies Δf that
are given values in the discrete Fourier transform of the displacement vector according to equation 34
(repeated below).
for
(34) [repeated]
In conventional 1D MRE, Δt is defined so that the MEG is shifted over one complete period of
the physical vibration with frequency f, as in equation 38.
(38)
When combined with equation 34, this constraint guarantees that the physical vibration frequency
will correspond to the first discrete frequency of the discrete Fourier transform of the displacement vector,
that is, with n = 1. In conventional 1D MRE, all other discrete frequencies are discarded.
The accumulated phase generated by an MEG, given by equation 16, can be extended to include
three different start times for the three directional MEGs, given in equation 39:
(39)
39
where subscripts j = 1, 2, and 3 correspond to the slice, read, and phase directions, respectively,
and Kj and uj stand for direction-specific MEG and displacement projection, respectively [4]. Assuming
sinusoidal-shaped MEGs and displacement functions as governed by equations 20 and 21, respectively,
equation 39 can be solved [4]. Adding the additional constraint that each directional MEG starts with zero
phase simplifies the solution to equation 40:
where
(40)
where ξj, Yj, sj, , Kj0, and qj are the direction-specific encoding efficiency, physical vibration
amplitude, phase offset, MEG amplitude, and number of MEG cycles, respectively. By using different
sampling intervals, Δtj, for each direction, it is possible to encode different directions with different
apparent frequencies. This yields the direction-specific sampling intervals of equation 41:
(41)
and the direction-specific MEG start times of equation 42:
for (42)
where N is the number of offsets, f is the physical vibration frequency, and subscript j = 1, 2, and
3 corresponds to the slice, read, and phase directions, respectively. Equation 42 demonstrates that the
MEGs are shifted over one, two, and three times the vibration frequency for slice, read, and phase
direction encoding, respectively [4]. Equation 42 is used to evaluate the discrete form of equation 39,
which yields equation 43.
40
where
(43)
Equation 43 demonstrates that each offset acquisition produces an NMR signal phase with
contributions from all three directions, each with a different apparent frequency. As with 1D MRE, the
NMR signal phase vector can be divided by the encoding efficiency of equation 43 to yield the
displacement vector. The discrete Fourier transform decomposes the displacement vector into discrete
frequencies via equation 34. Each directional displacement projection, having been acquired with a
unique apparent frequency, is stored independently in discrete frequencies corresponding to n = 1, 2, and
3 for slice-, read-, and phase-direction displacement, respectively. It is clear from equation 34 that the
number of offsets N must be at least 8 for three direction-dependent discrete frequencies to appear in the
discrete Fourier transform of the displacement vector.
Figure 15 depicts a pulse sequence diagram for a SLIM-MRE experiment, with MEGs from all of
the offsets superimposed into one diagram.
To minimize the increase in TE required by the increased sample intervals of SLIM-MRE, the
periodicity of harmonic waveforms is taken into account [4]. The MEG start times can be condensed
through the use of the modulo function, as in equation 44.
for (44)
Furthermore, the symmetry of harmonic waveforms is taken into account by recognizing that, for
N offsets divided evenly over period of oscillation 1/f, equation 45 must hold.
41
Figure 15. Pulse sequence diagram for a SLIM-MRE sequence with all MEG offsets superimposed.
(45)
Therefore, any MEG start time above the period of oscillation 1/f can be reduced by reversing the
polarity of the MEG amplitude and increasing the phase by π.
Table IV lists MEG start times for a SLIM-MRE sequence, both without (a) and with (b)
consideration of periodicity/symmetry, for vibration frequency f. MEGs with reversed amplitude
polarities are marked with (-).
It is apparent from Table IV that the consideration of symmetry and periodicity, shown in the
rows marked with (b), requires a much smaller increase in TE than otherwise. The maximum start time is
for number of offsets N and vibration frequency f, whereas without symmetry and periodicity
consideration this value is, at highest,
.
42
Table IV
MEG Start Times for SLIM-MRE with vibration frequency f and number of offsets N, without (a)
and with (b) consideration of the symmetry and periodicity of MEGs.
Offset n Slice MEG
start time (a)
Read MEG
start time (a)
Phase MEG
start time (a)
Slice MEG
start time (b)
Slice MEG
start time (b)
Slice MEG
start time (b)
0 0 0 0 0 0 0
1
2
3
4
5
6
7
Figure 16, devised by Klatt et. al [4], gives a visual representation of the apparent frequency
modulation of SLIM-MRE. Figure 16a represents the timing of the eight offsets for each MEG in real
time domain t. Figure 16b represents the timing of the eight offsets for each MEG in discretized time
domain t’, due to each offset sampling simultaneously for all three MEGs, as defined by equation 43.
43
Figure 16. Each of eight MEG start times in a SLIM-MRE sequence depicted in real time (a) and
discretized time (b).
44
IV. METHODS
A. Spin Echo Programming
The MRE pulse sequences are programmed using Siemens Integrated Development Environment
for Applications (IDEA) VB15 software for use on a Bruker 7T Clinscan MRI Scanner. MiniFLASH
pulse sequences, distributed with IDEA VB15 software, are used as the base sequence to be upgraded to
MRE sequences [16].
IDEA software is designed to simplify pulse sequence programming and testing. Pulse sequences
in IDEA are in the form of one or more C++-language files. IDEA-specific syntax calls for each pulse
sequence to be a distinct .CPP file containing five functions: fSEQInit, fSEQPrep, fSEQCheck, fSEQRun,
and fSEQRunKernel. Each of these functions performs a specific task in calculating and generating the
various outputs required to produce a pulse sequence on an MRI scanner: fSEQInit declares the limits on
user-defined parameters, fSEQPrep calculates the required timing for the pulse sequence by defining the
shape and duration of each of the RF / gradient pulses, fSEQCheck guarantees that the sequence does not
violate safety protocols, fSEQRun defines the loops over which the pulse sequence will be repeated, and
fSEQRunKernel (which is called by fSEQRun) sends the output signals to the MRI scanner that produce
RF and magnetic gradient pulses.
MiniFLASH is a very basic GE sequence. For each line of data acquired, MiniFLASH generates
a pulse sequence as seen in the pulse sequence diagram of Figure 5. To upgrade this sequence to a SE
sequence, thereby converting it to “MiniSE”, the following steps must be taken:
1. Add a 180° Refocusing RF pulse centered about time TE/2.
2. Add a Slice Refocus gradient pulse coinciding with the Refocusing RF pulse
3. Reverse the polarity of the Phase Rewind and Readout gradient pulses
45
4. Ensure that the readout gradient is centered about time TE, a time occurring exactly TE/2 after the center
of the RF pulse.
The addition of RF/Gradient pulses to a sequence using IDEA is performed in three steps. First,
the pulse is declared as an IDEA-specific C++ structure, either an sRF_PULSE or sGRAD_PULSE, for
RF or magnetic gradient pulse, respectively. This declaration must be made “globally”, that is, outside of
the four required functions in the .CPP file. Each structure defining a pulse is called a Real-Time Event
(RTE). Next, the RTE is prepared in the fSEQPrep function through IDEA-specific method calls that
define the pulse’s shape, duration, and amplitude. These parameters can be calculated or obtained through
user input in fSEQInit. Additionally, fSEQPrep organizes the timing of all RTEs. This guarantees that no
two pulses in the same direction are applied at the same time, and that the assigned values of TE and TR
are sufficient for a consistent pulse sequence. Finally, the RTE is applied in fSEQRunKernel within a
method call known as a Real-Time Event Block (RTEB). The RTEB is a list of all of the RTEs to be
applied during the pulse sequence along with their corresponding start times and directions of application.
As fSEQRunKernel is inside a loop within fSEQRun, the new pulse can be applied as many times as
required for the entire scan.
The above procedure can be used to include a refocusing RF pulse and a slice refocus gradient in
the MiniFLASH pulse sequence. The reversal of the phase rewind and readout gradient pulses’
amplitudes is performed during these pulses’ definition in fSEQPrep.
To guarantee that the timing conditions are fulfilled, two calculations must be made. First, the
minimum time required for all pulses between the midpoint of the excitation RF pulse and the midpoint of
the refocusing RF pulse is determined. Next, the minimum time required for all pulses between the
midpoint of the refocusing RF pulse and the midpoint of the ADC event is determined. The greater of
these two times is taken as the minimum required TE/2 for the pulse sequence, and is multiplied by two to
determine the minimum required TE. Finally, the excitation RF pulse, refocusing RF pulse, and ADC
46
event are given start times that guarantee their midpoints hit t = 0, TE/2, and TE, respectively. Thus, the
timing of the sequence is aligned.
These steps create a sequence, MiniSE, which generates a pulse sequence for each line of data
required identical to that of Figure 6.
B. Magnetic Resonance Elastography Programming
To upgrade MiniFLASH or MiniSE to an MRE sequence, the following steps must be taken:
1. Guarantee there is enough time between non-MEG gradients to apply MEGs without any gradient
overlap.
a. For GE sequences, MEGs are applied between the RF excitation pulse and the Readout gradient pulse.
b. For SE sequences, MEGs are applied in two locations: between the RF excitation pulse and the RF
refocusing pulse, and also between the RF refocusing pulse and the Readout gradient pulse.
2. Add oscillating MEGs to pulse sequence
3. For SE sequences, ensure that the MEG Gap is an integer multiple of half the vibration period.
4. Add an external trigger event to initiate the physical vibration
An MRE sequence requires several new user-controlled parameters. For each MEG, the user
specifies the frequency, amplitude, and number of cycles to be applied, as well as whether the gradient
pulses are sinusoidal or trapezoidal in shape, and flow-compensated or not flow-compensated.
Furthermore, the number of offsets acquired to resolve a time-dependent displacement function, as well as
the frequency corresponding to the time period over which the offsets are shifted, are given as user-
defined parameters.
User-controlled parameters can be added to a sequence by defining a global variable for each
parameter and assigning a value to it within the fSEQInit function. User control of these variables is made
convenient with the Siemens-supplied Parameter Map function, designed by Maxim Zaitsev [16]. Within
47
fSEQInit, user-controlled values are added to the new variables, and within fSEQPrep, these values are
used to prepare the various RF and gradient pulses, or rejected if found invalid for the sequence.
MEGs are declared as gradient pulses with the method used previously. For sinusoidal MEGs,
there is a specialized IDEA-specific C++ structure, sGRAD_PULSE_SIN, which is shaped like a positive
half-period of a sinusoidal function with an extended “flat top” time, as shown in Figure 17. Public
methods allow the ramp up time, duration, ramp down time, and amplitude to be defined, with each
parameter as described in Figure 17.
Figure 17. One sGRAD_PULSE_SIN structure with Amplitude, Ramp Up Time, Ramp Down
Time, and Duration defined.
48
A single period of a sinusoidal MEG is composed of two sGRAD_PULSE_SIN structures. By
assigning the ramp up time, duration, and ramp down time all equal to a quarter of the intended MEG
period, a sinusoidal half-pulse is generated. When one half-pulse is placed immediately before a second
half-pulse with opposite amplitude, one complete sinusoidal MEG period is generated, as shown in Figure
18.
Figure 18. One period of a sinusoidal MEG composed of two sGRAD_PULSE_SIN structures with
opposite amplitudes.
Similarly, a trapezoidal MEG can be constructed using the sGRAD_PULSE structure, which is
shaped like the function in Figure 19.
49
Figure 19. One sGRAD_PULSE structure with Amplitude, Ramp Up Time, Ramp Down Time, and
Duration defined.
To construct a trapezoidal MEG, two sGRAD_PULSE structures are used. Unlike sinusoidal
pulses, the ramp up and ramp down times of a trapezoidal pulse should be as close to zero as possible, so
that the NMR signal phase encoded by the MEG approximates to equation 27. The minimum ramp up and
ramp down times are defined by the system. Therefore, each trapezoidal half-pulse is assigned with ramp
up and ramp down times equal to the system’s minimum, and duration equal to the difference between
half of the intended MEG period and the assigned ramp down time. When one half-pulse is placed
immediately before a second half-pulse with opposite amplitude, one complete trapezoidal MEG period is
generated, as shown in Figure 20.
50
Figure 20. One period of a trapezoidal MEG composed of two sGRAD_PULSE structures with
opposite amplitudes.
To construct flow-compensated MEGs, quarter-pulses must be combined with the existing half-
pulses. A sinusoidal quarter-pulse can be generated by constructing an sGRAD_PULSE_SIN structure
with one ramp time equal to the minimum rise time of the system, and the other ramp time equal to the
difference between a quarter of the intended MEG period and the minimum system ramp time. A
trapezoidal quarter-pulse can be generated by constructing an sGRAD_PULSE structure with ramp up
and ramp down times equal to the minimum system ramp time and the duration equal to the difference
between a quarter of the intended MEG period and the assigned ramp down time. Both sinusoidal and
trapezoidal quarter-pulses are shown in Figure 21.
To construct a full period of a flow-compensated MEG, two quarter-pulses are applied on either
side of a half-pulse of opposite amplitude, as in Figure 22.
51
Figure 21. Quarter-period MEG pulses defined for the leading end of a sinusoidal MEG (a), the
trailing end of a sinusoidal MEG (b), and either end of a trapezoidal MEG (c).
Figure 22. One period of a flow-compensated MEG composed of two quarter-pulses and one half-
pulse for sinusoidal (a) and trapezoidal (b) MEGs.
52
To ensure that a sequence gives proper time for MEGs to be applied, a simple calculation is
made. The total time required by the MEGs for each acquisition line is the product of the MEG period and
number of MEG cycles, which are both user-defined parameters. This total is added to the calculation of
the minimum required TE. In the case of a SE sequence, half of the total MEG time is added to both the
calculation of the TE/2 before the refocusing pulse and the TE/2 after the refocusing pulse.
MEGs differ from most other gradient pulses in that MEGs are applied an arbitrary number of
times, namely, the number of cycles specified by the user. To accomplish this, a loop is created within the
fSEQRunKernel function, repeating the application of both halves of an MEG as many times as required.
For each instance of the loop, the MEG half-pulses are given a greater start time, so that no one period of
the MEG overlaps another.
For SE MRE sequences, special care must be taken to ensure that the time between the end of the
first run MEG and the start of the second run MEG, called the MEG Gap, must be an integer number of
half-periods of the vibration frequency. Within fSEQPrep, the minimum possible MEG Gap is computed
for the given value of TE as the time between the end of the first run MEG and the end of the slice
refocus gradient. Next, the smallest multiple of the vibration period that is greater than the minimum
possible MEG Gap is determined. This value is stored as the actual MEG Gap, and the start time of the
second run MEG is set as the sum of the end of the first run MEG and the actual MEG Gap.
To initiate the physical vibration, an external signal must be sent from the scanner to the
equipment producing the vibration. This is programmed into an MRE pulse sequence using an RTE called
sSYNC_EXTTRIGGER. This structure, when applied by a pulse sequence, produces a signal not to a
magnetic gradient or RF coil, but instead to an output port located on the scanner. This signal can be
directed to any external equipment.
Applying an sSYNC_EXTTRIGGER structure is achieved similarly to all other RTEs: the
structure must be defined globally, prepared within fSEQPrep, and applied within fSEQRunKernel. When
53
preparing an sSYNC_EXTTRIGGER structure, two parameters must be defined: the external port to
which the signal should be sent, and the duration of the RTE. The standard values for these parameters are
0 (the first output port) and 10 (the minimum acceptable duration in microseconds), respectively. Using
these parameters, every time this RTE is applied within fSEQRunKernel, a 10-microsecond transistor-
transistor logic (TTL) signal is produced on the first output port.
The above methods convert MiniFLASH and MiniSE to GE 1D MRE and SE 1D MRE
sequences, respectively.
C. Selective spectral Displacement Projection Programming
To upgrade 1D MRE sequences to SDP-MRE, the following steps must be taken:
1. Three simultaneous MEGs must be defined and applied, each with independent frequencies, number of
cycles, and amplitudes
2. For a given set of MEG frequencies, optimal number of cycles, number of offsets, and base frequency
must be calculated.
Defining and applying three simultaneous MEGs is achieved identically to the 1D case, except
performed in triplicate. Each parameter must be individually defined for slice-, read-, and phase-direction
MEGs.
To achieve proper SDP multi-directional motion encoding, each MEG’s combination of
frequency and number of cycles must selectively filter all other MEG frequencies by the process of
equation 37. To determine optimal SDP parameters, an additional calculation must be performed within
fSEQPrep. First, the greatest common divisor of the three user-defined MEG frequencies is calculated.
This is achieved by looping over all integers between zero and the smallest given MEG frequency,
inclusive, and saving the largest of these integers that divides evenly into all three given MEG frequencies
without remainder. This value is the base frequency of the SDP sequence. The minimum number of
54
cycles for each MEG is calculated by dividing each MEG frequency by the base frequency. If this
algorithm reports that the minimum number of cycles for any MEG is greater than 50, which is much too
large for any practical MRE sequence, the function exits and reports that the chosen MEG frequencies are
not SDP-compatible. Finally, the user can select an integer “MEG cycle factor”, which is multiplied by
the number of minimum cycles for each MEG, to boost MEG cycles without breaking the SDP
conditions.
The above methods convert MiniFLASH and MiniSE to SDP-MRE GE and SDP-MRE SE
sequences, respectively.
D. SampLe Interval Modulation Programming
To upgrade MiniFLASH and MiniSE to SLIM-MRE sequences, the following steps must be taken:
1. Three simultaneous MEGs must be defined and applied, all with identical frequencies
2. Each MEG is applied with a different sampling interval, resulting in different start times, as displayed in
Table IV
3. To account for the periodicity and symmetry of harmonic MEGs, the polarity of the MEGs is switched for
some offsets, as displayed in Table IV
Defining and applying three simultaneous MEGs is achieved identically to the 1D case, except
performed in triplicate. Each parameter must be individually defined for slice-, read-, and phase-direction
MEGs, except frequency, which for SLIM-MRE is identical in all directions.
To coordinate the starting times and polarities for each MEG during each offset, six arrays with
eight members each are defined globally. Next, in fSEQPrep, the arrays are filled with values. Three of
the arrays contain starting times for each of the three directional MEGs for each of eight offsets. The
other three arrays contain values of 0 or 1, corresponding to positive- or negative-polarity MEGs for each
of eight offsets. The values stored in these six arrays correspond to the starting time and polarity
55
assignments found in Table IVb. Finally, within fSEQRunKernel, the active offset being applied
corresponds to one of the eight members in each of the six arrays, which define the starting time and
polarity for each of the three directional MEGs.
E. Frequency Correction
One condition of IDEA pulse sequence programming is the system’s temporal resolution of 10
microseconds. All magnetic gradient pulses must have ramp up, duration, and ramp down times that are
integer multiples of 10 microseconds. This generates error when arbitrary MEG frequencies are used.
When the user of an MRE sequence inputs an MEG frequency, the quarter-frequency in
microseconds is calculated in fSEQPrep. Because this quarter-frequency time is used for the ramp up and
ramp down times of the MEG, it must be rounded to the nearest integer multiple of 10 microseconds. This
is performed using either the fSDSRoundUpGRT or fSDRoundDownGRT, which round the function up
or down to the nearest integer multiple of 10, respectively. These rounded values are used to prepare the
MEG pulses. If the quarter-frequency of the MEG is not already an integer multiple of 10 microseconds,
this rounding induces error in the actual applied frequency of the MEG. This error is considerable for
large values of MEG frequency. Since each quarter-period of the MEG must be rounded to the nearest 10-
microsecond multiple, each period has a maximum error of 40 microseconds. Thus, the maximum percent
error when fSDSRoundUpGRT is used follows equation 46:
(46)
and the maximum percent error when fSDSRoundDownGRT is used follows equation 47:
(47)
where MPE is the maximum percent error generated when rounding quarter-period values for
frequency f in Hertz.
56
To minimize this error, fSDSRoundUpGRT and fSDSRoundDownGRT are both used during
each MEG preparation. By preparing each MEG with half its quarter-pulse durations rounded using
fSDSRoundUpGRT and fSDSRoundDownGRT, the maximum percent error is decreased by nearly a
factor of 2, following equation 48.
(48)
Figures 23, 24, and 25 display the effect of the rounding error and its partial correction. Figure 23
displays the actual MEG frequencies generated for intended frequencies between 1 and 5000 Hz when
fSDSRoundUpGRT or fSDSRoundDownGRT are used alone. Figure 24 displays the actual MEG
frequencies generated for intended frequencies between 1 and 5000 Hz when partial correction is
performed by alternating fSDSRoundUpGRT and fSDSRoundDownGRT. Figure 25 displays the
bounding lines of erroneous frequencies for the cases described in figures 23 and 24, demonstrating that
the partial correction of Figure 24 limits the erroneous frequencies to a smaller bandwidth about the
intended frequency.
57
Figure 23. Applied frequency versus input frequency for MEGs when using fSDSRoundUpGRT()
and fSDSRoundDownGRT() rounding functions.
Figure 24. Applied frequency versus input frequency for MEGs when alternating use of both
fSDSRoundUpGRT() and fSDSRoundDownGRT() rounding functions.
58
Figure 25. Maximum erroneous frequency trend lines for MEGs when using fSDSRoundUpGRT(),
fSDSRoundDownGRT(), and alternating both rounding functions.
59
V. EXPERIMENTAL SETUP
A. Physical Vibration Source
To produce the vibration necessary for an MRE experiment, the experimental setup of Figure 26 is
used.
Figure 26. Experimental setup to produce physical vibration in test sample.
Four pieces of equipment are illustrated in Figure 26:
1. Agilent 33250A 80MHz Function / Arbitrary Waveform Generator
2. Elenco XP-581A DC Variable Voltage Supply
3. Yamaha P3500S Power Amplifier
4. PI Ceramic P-887.91 Stack Multilayer Piezoelectric Actuator
An external trigger event in an MRE pulse sequence generates a TTL signal that is fed into the
function generator. This signal triggers the function generator to output a voltage waveform matching the
60
shape of the desired vibration, which is fed into the power amplifier. The vibration waveform is amplified
to 40 Volts peak-to-peak. This amplified signal is given a DC offset of 20 Volts by the variable voltage
supply, guaranteeing that the voltage signal never drops below zero Volts. Finally, the amplified and
offset vibration waveform is fed into the piezoelectric actuator, which vibrates with frequency equal to the
input voltage waveform. The actuator vibrates the sample with a desired frequency, allowing an MRE
experiment to take place.
B. Experimental Gel Sample
To test the MRE sequences, an inhomogeneous gel sample is produced. An inhomogeneous
design is chosen to guarantee three-dimensional wave characteristics. For the purpose of validating the
efficacy of the 3D MRE sequences, the specific stiffness of the inner and outer gels are not relevant; it is
more important that the gels are of different stiffness, thereby producing 3D wave patterns caused by
refraction and reflection. The cylindrical gel sample followed the general format of Figure 27.
Figure 27. Schematic for experimental inhomogeneous gel sample.
61
The container is a hollow Garolite cylinder of length 50 mm, inner diameter 32 mm and wall
thickness of approximately 1 mm. One end of the cylinder is closed, and a threaded hole is bored into the
closed end to attach to the actuator. The actuation of the sample occurs mainly parallel to the slice
direction, as displayed in Figure 28.
Figure 28. Orientation of vibration actuation for experimental gel sample as it relates to slice
direction in scanner.
The gel sample is composed of two masses of Smooth-On EcoFlex 00-10 Silicone rubber with
slightly different densities. The inner gel sample, of irregular shape and approximate diameter of 25 mm,
is first produced using EcoFlex 00-10 with standard density of 1040 kg/m3 [17]. After curing for 24
hours, this inner mass is placed in the cylinder container and surrounded with EcoFlex of lower density,
produced by mixing silicone oil with the EcoFlex 00-10 reagents at a volume ratio of 1:9 before curing.
After curing for 24 hours, the sample is ready for testing.
62
C. Sequence Parameters
The MRE sequences were tested using a 7T Bruker ClinScan MRI Scanner operated using
Siemens IDEA VB15 pulse sequencing software. To obtain the maximum amount of signal, SE
sequences were used. Sinusoidal, non-flow-compensating MEGs were used for all results presented in
this work, although other MEG shapes were implemented as part of this thesis as well. The general MRI
parameters used are collected in Table V. The MRE-specific parameters used for the SLIM-MRE scan
and associated 1D MRE scans are collected in Table VI, and those of the SDP-MRE scan and associated
1D MRE scans are collected in Table VII.
Table V
MRI Parameters used for all MRE scans
Parameter Value
TR 1000 ms
Field of View (FOV) 32 mm (read) x 32 mm (phase) x 10 mm (slice)
Image Resolution 64 pixels (read) x 64 pixels (phase)
Number of Slices 20
Slice Thickness 0.5 mm
Flip Angle 90°
63
Table VI
MRE Parameters for 1D MRE scans in the Slice (a), Read (b), and Phase (c) directions, and for the
SLIM-MRE scan (d)
1D MRE –
Slice (a)
1D MRE – Read
(b)
1D MRE – Phase
(c)
SLIM-MRE
(d)
Slice MEG Frequency 1000 Hz N/A N/A 1000 Hz
Slice MEG Number of
Cycles
3 N/A N/A 3
Slice MEG Amplitude 250 mT/m N/A N/A 250 mT/m
Read MEG Frequency N/A 1000 Hz N/A 1000 Hz
Read MEG Number of
Cycles
N/A 3 N/A 3
Read MEG Amplitude N/A 250 mT/m N/A 250 mT/m
Phase MEG Frequency N/A N/A 1000 Hz 1000 Hz
Phase MEG Number of
Cycles
N/A N/A 3 3
Phase MEG Amplitude N/A N/A 250 mT/m 250 mT/m
Number of Offsets 8 8 8 8
TE 12.99 ms 12.99 ms 12.99 ms 11.96 ms
Vibration Frequency 1000 Hz 1000 Hz 1000 Hz 1000 Hz
64
Table VII
MRE Parameters for 1D MRE scans in the Slice (a), Read (b), and Phase (c) directions, and for the
SDP-MRE scan (d)
1D MRE –
Slice (a)
1D MRE – Read
(b)
1D MRE – Phase
(c)
SDP-MRE
(d)
Slice MEG Frequency 1000 Hz N/A N/A 1000 Hz
Slice MEG Number of
Cycles
3 N/A N/A 3
Slice MEG Amplitude 250 mT/m N/A N/A 250 mT/m
Read MEG Frequency N/A 2000 Hz N/A 2000 Hz
Read MEG Number of
Cycles
N/A 6 N/A 6
Read MEG Amplitude N/A 250 mT/m N/A 250 mT/m
Phase MEG Frequency N/A N/A 3000 Hz 3000 Hz
Phase MEG Number of
Cycles
N/A N/A 9 9
Phase MEG Amplitude N/A N/A 250 mT/m 250 mT/m
Number of Offsets 8 8 8 8
TE 12.99 ms 12.99 ms 12.99 ms 13.52 ms
Vibration Frequency 1000 Hz 2000 Hz 3000 Hz Superposition
of 1000, 2000,
and 3000 Hz
65
VI. RESULTS
Figure 29 compares the temporally-resolved displacement images acquired from one SLIM-MRE
experiment (Figures 29a, 29b, and 29c) to those of three related 1D MRE experiments (Figures 29d, 29e,
and 29f). Twenty slices total were acquired, and one example slice (slice number 7), is displayed in
Figure 29. The real parts of the complex displacement images are shown. Of note, all wave images were
noise-filtered using a 4-pixel Butterworth lowpass filter of order 2 [8], and therefore may still contain the
contribution due to compression waves.
Figure 29 demonstrates that the SLIM-MRE sequence encodes similar displacement images to
that of 1D MRE sequences. Because the three SLIM-MRE displacement projections were acquired in
one-third the time of using 1D MRE, SLIM-MRE represents a significant acceleration in acquiring 3D
MRE data. The additional noise visible in the 1D MRE data is a product of the increased TE of the 1D
MRE sequences, which was not optimized for TE minimization.
66
Figure 29. Acquired displacement images for SLIM-MRE sequence in slice (a), read (b), and phase
(c) directions, and for 1D MRE sequences in slice (d), read (e), and phase (f) directions.
Figure 30 compares the calculated stiffness factor images obtained from SLIM-MRE (Figure 30a)
and 1D MRE (Figure 30b) sequences. It was found that the inversion technique not using the curl
operator, shown in equation 13, was unsuitable for processing of the acquired images. The compression
waves propagating through the sample cause substantial error in the calculation of shear modulus values
using this inversion technique. However, the technique that involves use of the curl operator, shown in
equation 11, removes the effects of compression waves in the displacement images, resulting in clearer
67
shear modulus images. This inversion technique is used to generate the images in Figure 30. Of the
twenty acquired slices, an example slice (number 7) was chosen for display in Figure 30. The real parts of
the complex modulus images are shown.
Figure 30. Calculated complex shear modulus images using curl-operator inversion method for
SLIM-MRE (a) and 1D MRE (b).
Figure 30 demonstrates that the elastograms generated by the SLIM-MRE sequence are
comparable to those of 1D MRE sequences. The slight difference in stiffness between the outer and inner
gel masses is visible. While the characteristics of the elastograms are similar, differences in magnitude are
most likely due to extra noise in the 1D MRE sequences, compounded by the multiple spatial derivatives
necessary during inversion.
68
Figure 31 compares the temporally-resolved displacement images acquired from one SDP-MRE
experiment (Figures 31a, 31b and 31c) to those of three related 1D MRE experiments (Figures 31d, 31e,
and 31f). Twenty slices total were acquired, and one example slice (number 7) is displayed in Figure 31.
The real parts of the complex displacement images are shown. Of note, a Butterworth bandpass filter
(with low and high filter limits of 4 and 32 pixels, respectively) was applied to all wave images for
filtering noise and the contribution due to compression waves.
Figure 31. Acquired displacement images for SDP-MRE sequence in slice (a), read (b), and phase
(c) directions, and for 1D MRE sequences in slice (d), read (e), and phase (f) directions.
69
Figure 31 demonstrates that the SDP-MRE sequence encodes similar displacement images to that
of 1D MRE sequences. Because the three SDP-MRE displacement projections were acquired in the time
it took to acquire one 1D MRE displacement projection, SDP-MRE represents a significant improvement
in acquiring 3D MRE data. One notable shortcoming, however, of the SDP-MRE displacement data is the
decreased intensity of displacement encoding as compared to that of 1D MRE sequences. This is due to
the fact that in the SDP-sequence, by simultaneously exciting three frequencies of vibration
simultaneously, a decreased amplitude of vibration for each single frequency is experienced. In effect, the
multifrequency vibration, applied with identical amplitude to the monofrequency vibration of 1D MRE,
exhibits one-third the amplitude for each frequency component than that of the monofrequency vibration.
The decreased amplitude results in less motion encoding, visible in the lower intensity of the SDP-MRE
displacement images as compared to the 1D MRE displacement images in Figure 31.
The usual 3D inversion techniques of equations 11 and 13 are unusable with our SDP-MRE data
set because they require the 3D displacement for each frequency. Still, the wave patterns depicted in
Figure 31 indicate a successful implementation of SDP-MRE.
70
VII. DISCUSSION
A. Limitations
The SLIM-MRE displacement images of Figure 29 resemble the associated 1D MRE images, but
several improvements could be made. The noise apparent in the 1D MRE images could be mitigated
through optimization of the 1D MRE sequence. Instead of shifting the MEG relative to the other gradient
and RF events in the 1D MRE sequence, all other events could be shifted relative the external trigger
event. This would allow for minimization of TE, resulting in much less noise. Furthermore, it was
observed for the inhomogeneous gel sample that vibration in the read and phase directions had amplitudes
several orders of magnitude smaller than that of vibration in the slice direction. To normalize the
vibration amplitudes, actuation can be performed in a non-axial direction. Such actuation would result in
increased vibration amplitude in non-slice directions, which would increase the amount of NMR phase
signal encoded in these directions.
The SDP-MRE displacement images of Figure 31 resemble the associated 1D MRE images, but
several improvements could be made. As with the 1D MRE images of Figure 29, the noise level apparent
in both the SDP-MRE and 1D MRE images of Figure 31 could be decreased by optimizing the sequences
for TE minimization. In addition, the SDP-MRE images exhibit decreased intensity as compared to the
1D MRE images. This is due to the decreased vibration amplitude inherent in multifrequency vibrations:
each frequency component ends up with a smaller amplitude than that of a monofrequency vibration. A
solution to this would be an increase in multifrequency vibration amplitude: either by increasing the
voltage sent to the piezoelectric actuator, or adding more actuators in series with the first to boost
displacement.
71
B. Conclusion
SLIM-MRE and SDP-MRE generate displacement images comparable to those of standard 1D
MRE sequences, and do so in one third of the time. Therefore, SLIM-MRE and SDP-MRE represent
significant time-saving 3D MRE techniques, and may improve the acceptance of MRE in clinical settings.
This work demonstrates the viability of SLIM-MRE and SDP-MRE implementation on a
Siemens/Bruker MRI system, adding to the existing deployment of these sequences demonstrated by
Klatt et al. [4] [5] and Yasar et al. [3]. This work serves as further verification that the new 3D motion-
encoding techniques of SLIM-MRE and SDP-MRE are modality-independent, equally viable on any MRI
operating system or scanner.
72
CITED LITERATURE
1. Mazza, E., et al.: The mechanical response of human liver and its relation to histology: An in vivo
study. Med. Image Analysis, 11: 663-672, 2007.
2. Yeh, W., et al.: Elastic modulus measurements of human liver and correlation with pathology.
Ultrasound in Med. & Biol., 28: 467-474, 2002.
3. Yasar, K., et al.: Selective spectral displacement projection for multifrequency MRE. Phys. Med.
Biol., 58: 5771-5781, 2013.
4. Klatt, D., et al.: Sample interval modulation for the simultaneous acquisition of displacement
vector data in magnetic resonance elastography: theory and application. Phys. Med. Biol., 58:
8663-8675, 2013.
5. Klatt, D., et al.: Simultaneous, multidirectional acquisition of displacement fields in magnetic
resonance Elastography of the in vivo human brain. Journal of Magnetic Resonance Imaging, In
Press.
6. Liang, Z., and Lauterbur, P.: Principles of Magnetic Resonance Imaging: A Signal Processing
Perspective. New York, Wiley-IEEE Press, 2000.
7. Chandrasekharaiah, D., and Debnath, L.: Continuum Mechanics. New York, Academic Press,
1994.
8. Buttkus, B.: Spectral Analysis and Filter Theory in Applied Geophysics. New York, Springer,
2000.
9. Sinkus, R., et al.: MR Elastography of Breast Lesions: Understanding the Solid/Liquid Duality
Can Improve the Specificity of Contrast-Enhanced MR Mammography. Magnetic Resonance in
Medicine, 58: 1135-1144, 2007.
10. Oliphant, T., et al.: Complex-Valued Stiffness Reconstruction for Magnetic Resonance
Elastography by Algebraic Inversion of the Differential Equation. Magnetic Resonance in
Medicine, 45: 299-310, 2001.
11. Hahn, E.: Detection of sea-water motion by nuclear precession. J. of Geophys. Res., 65: 776-777,
1960.
12. Muthupillai, R., et al.: Magnetic Resonance Elastography by Direct Visualization of Propagating
Acoustic Strain Waves. Science, 269: 1854-1857, 1995.
73
CITED LITERATURE (continued)
13. Sack, I., et al.: Magnetic resonance Elastography and diffusion-weighted imaging of the sol/gel
phase transition in agarose. J. of Magn. Reson., 166: 252-261, 2004.
14. Taylor, A.: L’Hospital’s Rule. The American Mathematical Monthly, 59: 20-24, 1952.
15. Oppenheim, A., and Schafer, R.: Digital Signal Processing. Upper Saddle River, Prentice Hall,
1975.
16. IDEA: Integrated Development Environment for Applications. Siemens AG, Healthcare,
Erlangen, Germany.
17. Smooth-On Inc.: EcoFlex Series Super-Soft, Addition Cure Silicone Rubbers, 2011. Available:
<http://www.smooth-on.com/tb/files/ECOFLEX_SERIES_TB.pdf> Accessed October 01, 2014.
74
VITA
NAME:
EDUCATION:
RESEARCH EXPERIENCE:
HONORS:
CERTIFICATION:
David Arthur Burns
B.S., Nuclear, Plasma, and Radiological Engineering, University
of Illinois, Urbana-Champaign, Illinois, 2010
M.S., Mechanical Engineering, University of Illinois at Chicago,
Chicago, Illinois, 2014
Department of Bioengineering, University of Illinois at Chicago,
Chicago, Illinois: Motion-Sensitive MRI Laboratory, 2014
Department of Nuclear, Plasma, and Radiological Engineering,
University of Illinois, Urbana-Champaign, Illinois: Center for
Plasma-Material Interactions, 2007-2010.
Roy A. Axford Undergraduate Scholarship, University of Illinois,
Urbana-Champaign, Illinois, 2009
George H. Miley LENR Undergraduate Scholarship, University
of Illinois, Urbana-Champaign, Illinois, 2009-2010
Chancellor’s Scholar, University of Illinois, Urbana-Champaign,
Illinois, 2006-2010
James Scholar, University of Illinois, Urbana-Champaign,
Illinois, 2006-2010
Dean’s List, University of Illinois, Urbana-Champaign, Illinois,
2006-2010
Siemens IDEA Pulse Sequence Programming Training, Cary,
North Carolina, 2014
U.S. Peace Corps Pre-Service Training, Loitokitok, Kenya, 2010