SVAC Instrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

  • View
    216

  • Download
    0

Embed Size (px)

Citation preview

Page 1: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 1

GEM Discarded EventsGEM Discarded Events

Anders W. Borgland

ScienceVerification,Analysis andCalibrations

Page 2: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 2

OutlineOutline

• What are the GEM discarded events• Number of discarded events• Time distribution of the discarded events• STR 33 results• STR 35 anticipation• Summary

Page 3: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 3

Why You Should Be Looking At The Data!Why You Should Be Looking At The Data!

• This all started with Mutsumi coming into my office with a distribution that looked strange.

• A lot of people have been involved in what followed:

– Warren, Jim, Mike, Eric, Eric, Eric .....• Here I will try to summarize what we know about this effect.

SVAC ntuple variables!

Run 135004445B28 tower

Page 4: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 4

GEM Discarded EventsGEM Discarded Events

• GEM discarded events:

– A counter that keeps track of how many times the trigger window turned, but the LAT was busy.

– Called 'GemDiscarded' in the SVAC ntuple.• We may have discarded events under normal data taking:

– Two physics events coming close to each other, and the LAT is busy reading out the first one

– Periodic trigger fires while we read out an event• But the rate should be very low!

GEM Discarded counter for first 10 events of run 135004445

Value of GemDiscarded

We have

We have 20 discarded events for the first 10 events that were read out!There is also a pattern here .....

Page 5: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 5

Plots: Part IPlots: Part I

• I will plot how much the GEM discarded counter increases from one read out event to the next read out event:

– Delta GemDiscarded• Will show you plots for:

– Different number of towers in the grid:

– 1, 2, 4, 6 and 8 towers

– Different run types:

– B2 and 1-1: Flight configuration

– B10: CAL 4-range readout

– B13: CAL 4-range readout and no zero-suppression• Deadtime is different for the different run types (see Warren's talk

from August 5):

– B2 : 529 ticks

– B10: 1309 ticks

– B13: 13129 ticks• Remember:

– It's only a discarded event if it occurs inside the dead time!

Page 6: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 6

One Tower Data: End2End 1-1 One Tower Data: End2End 1-1

Delta GemDiscarded peaks at 0.0.5% of the read out events have discarded events

Page 7: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 7

Two Tower Data: End2End B2 Two Tower Data: End2End B2

Delta GemDiscarded still peaks at 0.1% of the read out events have discarded events.

Page 8: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 8

Two Tower Data: B13Two Tower Data: B13

More discarded events here than for B2, but remember the deadtime is larger.

Page 9: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 9

Four Tower Data: B2Four Tower Data: B2

We nearly always have one discarded event per read out event!

Page 10: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 10

Four Tower Data: B10Four Tower Data: B10

Delta GemDiscarded peaks at 0!

Page 11: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 11

Six Tower Data: B2 And B10Six Tower Data: B2 And B10

B2: Nearly always 2 discarded events per read out event! And we always have discarded events!

B10: Nearly no discarded events at all!

Page 12: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 12

Eight Tower Data: B2 And B10Eight Tower Data: B2 And B10

B2

B10We now have discarded events all the time for both run types!

Page 13: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 13

Time Structure Of The Discarded EventsTime Structure Of The Discarded Events

• Will plot:

– GemDeltaEventTime - GemDeltaWindowOpenTime• This gives us the time between the previous read out event and the

LAST discarded event:

– See cartoon on the next page.

– NB! As the number of discarded events increases this time quantity will increase since we always look at the time of the last discarded event.

– This quantity will be 0 if there are no discarded events• We don't know anything about the other discarded events except how

many there are (from GemDiscarded).• This time quantity is only useful for events where GemDeltaEventTime

is not saturated (3.3 ms).

Page 14: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 14

Time Structure Of The Discarded EventsTime Structure Of The Discarded Events

Previously read out event

GEM Delta Event Time

GEM Delta Window Open Time between current event and the last retriggered event

Trigger window width (12 ticks)

N retriggered events:Only info available is GEM discarded counter

Currentread out event:Time of event is time of window closing

Page 15: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 15

One Tower Data: 1-1One Tower Data: 1-1

Remember: Not many discarded events for 1 tower data, but there is a peak around 140 ticks .....

529: Deadtime for 1-1 runs

Page 16: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 16

Two Tower Data: B2Two Tower Data: B2

Definitively a peak around 140 ticks.

Page 17: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 17

Four Tower Data: B2 And B10Four Tower Data: B2 And B10

B2

B10

New peak around 350 ticks.Peak at 140 is still there.

Page 18: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 18

Six Tower Data: B2 Six Tower Data: B2

B2

Remember: Now we always have discarded events.

Page 19: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 19

Eight Tower Data: B2 And B10Eight Tower Data: B2 And B10

Two peaks: 350 and 365 ticks

B2

B10

Another new peak ....

Page 20: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 20

STR 33STR 33

• We don't know what causes these discarded events.• Suspect the CAL:

– “Known” that when you read out the CAL you 'retrigger'.• Special Test Request: STR 33

– Take muon data in a B2 configuration

– Turn off CAL LO

– Turn off CAL HI

– Turn of both CAL LO and CAL HI• 'Turn off':

– We don't allow it to open the trigger window

Page 21: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 21

STR 33: Baseline B2 RunSTR 33: Baseline B2 Run

B2

B2

STR Baseline B2 run

Old 8 tower B2 run

For the same configuration,the number of discarded events changed!

Page 22: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 22

STR 33: Baseline Run B2STR 33: Baseline Run B2

Time distribution did not change!

Page 23: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 23

STR 33: Turn Off CAL LOSTR 33: Turn Off CAL LO

A small decrease in the number of discarded events

Page 24: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 24

STR 33: Turn Off CAL LO - TimeSTR 33: Turn Off CAL LO - Time

We see only the '350' peak.

Page 25: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 25

STR 33: Turn Off CAL HISTR 33: Turn Off CAL HI

A lot less discarded events!

Page 26: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 26

STR 33: Turn Off CAL HI - TimeSTR 33: Turn Off CAL HI - Time

We see our old friend at 140 again .....(consequence of having less discarded events). And we still have the 365 peak.

Page 27: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 27

STR 33: Turn Off CAL LO And CAL HISTR 33: Turn Off CAL LO And CAL HI

Very few discarded events left!

Page 28: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 28

SummarySummary

• We now always have discarded events when we take data.• The are coming from the CAL triggers:

– Mainly CAL HI• The discarded events occur at very specific times.• The effect seems to be worse with increasing number of towers in the

grid:

– Current hypothesis is that some CAL modules are more sensitive than others and we happened put one of them in when we went to 4, 6 and 8 towers.

• STR 35:

– We will redo STR 33 but with only one tower powered up at a time.

– Will tell us if some modules are more sensitive than others

– Hope to do it next week.

Page 29: SVACInstrument Analysis Meeting, September 23, 2005 Anders W. Borgland 1 GEM Discarded Events Anders W. Borgland Science Verification, Analysis and Calibrations

SVAC Instrument Analysis Meeting, September 23, 2005

Anders W. Borgland 29

Summary cont'Summary cont'

• Source of the problem:

– Seem to come from the range selector in the Mux (see Fig. 1 in LAT-SS-01972).

– According to the experts these events should never leak outside the dead time and become real retriggers.

• It looks like we have to live with this effect.• Makes some GEM variables useless:

– Gem Discarded

– Gem Delta Window Open Time• And we discovered this because somebody plotted an SVAC ntuple

variable and thought it looked strange ......

Looking at the Data is doubleplusgood!