Upload
others
View
8
Download
0
Embed Size (px)
Citation preview
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
J UOEH21 (2):107-118 {1999) 107
(Origina])
Subjectivethe Isehial-Evaluations
and Objective Measurements of
Ramal Containment Prosthesis
Kenii HAcHIsuKAL, Yuichi UMEzuL, Hajime OGATAi,
Saburo OHMINE', Koichi SHiNKoDA2 and Hideaki ARIzoNosiDepartment
ofRehabigitation Medicine, School ofMedicine,2Rehahilitation
Unit, University HbspitaZ, U}ziversity of Occupational and
Japan. Ykehatanishi-ku, Kitakyushu 807-8555, Jcrpan'IArizono's
Smplies. }khatahigas'hi-ku, Kitakyu,sht{ 805-O061, Japan
Environmentai Health,
Abstract: We examined 12 transfemoral amputees, 6 using the IRC socket and 6 the QL socket, to confirm whether the ischial-ramal containment (IRC) socket is truly su-
perior to the quadrilateral (QL) socket. In subjective evaluation, the IRC group was significantly better in the totai score and in items of comfort, that is, to sit on a
chair and lumbar lordosis at heel off (Mann-Whitney test, P <O.05) , better but not
significant in the items of comfbrtable to wear, comfortable to go up and down
stairs, and truncal sway during stance phase. By computed tomography, the fe- mur of the IRC group was kept in a position significantly more medial than that of
the QL group (Mann-Whitney test, P<O.05), but no signjfjcant difference in
gluteal medial muscle atrophy ratios between the two groups was found (Mann- Whitney test, P >O.05). By X-ray, the stump of the IRC group was maintained
significantly more adducted during one foot standing on the prosthesis (Mann- Whitney test, P<O.05), but the lateral fbrce ratio during mid-stance of the IRC
group was smaller, but not significantly, than that of the QL group. Physiological
cost index (PCI) , an indirect simple method fOr evaluating oxygen consumption of
gait, had no significant difference between the two groups (Mann-Whitney test, P
>O.05), and a multiple regression analysis revealed that the stump length ratio
and lateral fbrce ratio during mid-stance were significant explanatory variables for
predicting PCI (adjustedR square : O.87, F-value : 11.85, P <O.05) , The results
of this study have revealed that the advantage of the IRC socket is a tender feeling
of the stump, but that the rnetabolic efficiency is not superior to the QL socket at
the most comfortable speed.
KL7 worcZs:ischial-ramal containment socket, transfemoral amputation, prosthesis, evaluation.
<Received 22 January 1999, accepted 15 April 1999 )
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
leg K HAcHisuKA et al
Introduction
The quadrilateral (QL) socket design provides excellent support and stability for trans-
femoraL amputees during walking, and has replaced the p]ug fit socket [1,2]. The QLsocket has a contour with narrow anterior-posterior and wide medial-lateral dimensions, and
an ischial seat is attached to the top of the posterior wal] to bear the arnputee's body weight
(Fig. 1 ). Although the QL prosthesis is standard all over the world, we have occasionally
seen a transfemoral amputee using the QL socket who has his or her stump slightly abducted
[3] , complains of a feeling of being pushed up by the edge of the posterior wal1 at hip ex-
tension [4], and experiences pain on the ischial tuberosity, especially, at the beginning ot:
gait training. In recent years, new socket designs fora transfemoral amputee have been in-
troduced : a narrow medial-lateral socket known as Normal Shape-Noirnai Alignment [5] ,Contoured Adducted TTochanteric-Controlled Alignment Method (CAT-CAM) [6], and
Narrow Medial-Lateral socket (Narrow M-L) [7] . These new socket designs are generally
called the ischial-ramal containment socket (IRC) [8-10], because they have the common
feature that the' socket contains the ischial tuberosity and ramus inside the socket. The es-
sential structure of the RC socket is l. ) narrow medial-lateral and wide anterier-posterior di-
mensions, 2 ) no definite ischial seat at the top of the posterior wall, and containment of the
ischial tuberosity and ramus inside the socket.
Fig. 1 . Top and antcrior vicws ofthe quadrilateral socket for a left transfemoral amputee,
A : anterior, L : lateral, M : medial, * : ischial seat.
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
IRC Prosthcsis for Transfcmoral Amputation i09
The initjal idea of the IRC socket was derived frem Long's line [3 , 5] , Long noticed
that a transfemoral amputee wearing the QL socket had his or her stump abducted during
walking, and advocatedanew alignment of the socket so that the tip of the femur on the am-
putated side was set en a line perpendicular from the hip joint to the ground to maintain the
stump in a normal position. Sabolich developed this concept and adopted a bony lock to
prevent the stump from abducting by adding pressures to the lower femur and pelvis toward
the inside and to the ischial tuberosity and ramus toward the outside [6]. A transfemora]
amputee wearing a CAT-CAM prosthesis reported increased comfort, ease of donning and
doffing, and increased gait efficiency [11]. Flandry described the diminished compensa-
tory lateral trunk lean during walking, increased customary gait velocities, and reduction in
the quantity ofconsumed oxygen per meter by up to 50 % [12]. Gailey indjcated a signifi-
cantly lower energy expenditure during walking with the CAT-CAM socket at a faster pacethan that with the QL socket [13]. In our previous survey of the IRC prosthesis, amputees
wearing the IRC socket noticed that there was a more comfortable feelillg at the ischial tuber-
osity during walking after they had their socket converted from the QL socket to the IRC
socket, and that lateral trunk lean during the stance phase was reduced [14], However,
some amputees wearing the IRC socket still showed lateral trunk lean, and could not actually
realize any reduction of oxygen consumption during walking, fbr example, less fatigue from
walkjng on the prosthesis. This study, therefbre, was perTformed to confirm whether the IRC
socket is better than the QL socket in subjective assessment and objective measurement, and
whether metabolic efficiency of the IRC socket evaluated with the physiological cost index
(PCI) [15] is really superior to the QL socket
Materials and method
Twelve transfemoral amputees ( 6 for the IRC group and 6 fbr the QL group) were se-
lected as 36 subjects according to the inclusion and exclusion criteria from amputees who had
been rehabilitated in the UniveTsity Hospital of Occupational and Environmental Health and
aff11iated hospitals. The inclusion criteria for selecting the IRC group were as follows : 1 )unilateral transfemoral amputees, 2 ) being prescribed andf'or checked at their brace clinics
by one of the authors, 3 ) being able to walk usefully with the IRC prosthesis, 4 ) having
once used the QL socket, and S ) agreeing to join this clinical study. The exclusion criteria
were as fbllows: 1) amputees wearing aplug-fit type socket, 2) amputees suffering from
is¢ hemic heart diseases, diabetes mellitus, renal insufficiency, and or degenerative joint dis-
ease, and 3 ) taking any medicine affecting cardio-respiratory functions, that is, 5-blocker,and Ca-antagonist, For selecting the QL group,
`CIRC"
in the third item of the inclusion cri-
teria was replaced by "QL"
and the fourth item was omitted.
Befbre evaluating the IRC and QL sockets, all the prostheses were carefuIly checked by
the authors, and were remade or modified to obtain optimal fit, if necessary. Six transfemo-
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
110 K TIAcHtsuKA et at
ral amputees ( 5 males and 1 female) were using the IRC socket which was made by an ex-
perienced certified clinical prosthetist according to the manual [7] . The IRC sockets were
the Scandinavian flexible type with a Surlyn plastic inner socket and acrylic resin hard frame
(Fig. 2 ) , afld were attached to the endoskeletal prostheses. Their parts were a safety knee
for 5 prostheses and a manual ]ock kncc fbr onc, and a sing]c axis foot for 6. Six trans-
femoral amputees ( 6 males) were using the QL sockets which were also made by expcricnced
cenified clinical prosthetists, Five QL sockets were attached to the endoskeletal and one to
the shell prostheses, and their parts were a safety knee for 5 prostheses and a manual lock
knee for one, anda single axis foot fbr3 and an energy storing foot fbr3. Stump length ra-
tios were obtained by dividing the length from the ischial tuberosity to the tip ol' the stump by
the Iength from the ischial tuberosity to the lateral femoral condyle on the non-amputated side.
Subje¢ tive evaluation of the sockets consisted of 6 items of self-assessment and 3 items
of specialist-assessment, which wcrc based on the results of our previous study [14],A selflassessment questionnaire, including
"comfortable
to wear," "easc
to swing the prosthe-sis,""comfortablc to go up and down stairs,""comfortable to sit on a chair,']"appearance of
the socket," and "donning
and doffing thc sockct," was handed to the subjects at the brace
clinic, and these items were rated as 4 for good, 3 fbr slightly good, 2 fbr slightly poor, and 1
for poor by the subjects. Specialist-assessment items, "hip
abduction during swing phase,"C`truncal
sway during stancc phasc," and "lumbar
lordosis at heel off," were rated as 4 for no
abduction (sway or lordosis),3 fbr slight, 2 for somewhat, and 1 fbr severe by three of the
Fig.2. Top and antcrior
amputce,
A : an terior, M :
view of the ischial-rainal containment socket for a right Lransfemorul
medial, 'i:
: indentation of the posLcrior svul] bearing ischial tuberosity.
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
IRC Prosthesis for Transi'emeral Amputation 111
authors (KH, SO, and HA) , while the subjects walked on a walkway in the training room two
or three times. VVhen there was a difference in judgment among the evaluators, they ac-
cepted a majority decision.
Cemputed tomograms of every 2 cm of transverse s]ice between the hip afld upper
thighs were taken to reconfirm whether the socket fitted well and to examine the femur posi-tion which was one of the features of the IRC socket. During a scan, the amputees wearing
their prosthesis lay on their back with their feet pressing against a footrest to maintain total
contact between the stump and socket. Femur posjtion at the anterior-posterior direction
was determined by the ratio of distance between the anterior edge of the thigh and femur di-
vided by the anterior-posterior dimension of the thigh 4 cm below the great trochanter andmultiplied by 1OO. The femur position at the medial-lateral direction was defined as a ratio
of distance between the medial edge of the thigh and femur divided by the medial-lateral di-
mension of the thigh 4 cm below the great trochanter and multiplied by leO.
To ascertain atrophy of the gluteal medial muscle, we measured the areas of the bilateral
gluteal medial muscles at the slice containing the inferior anterior iliac spine by an image ana-
lyzer (Avionics, Japan) . Gluteal medial muscle atrophy ratio was defined as the area on the
amputated side divided by that on the non-amputated side.
Anterior-posterior X-ray photographs of the pelvis and femurs weTe taken while the sub-
.iects were standing on one foot to examine the hip adduction angle simulating mid-stance of
gait. The angle between the longitudinal axis of the femur and the perpendicular line to a
line connecting the upper borders of the acetabulum of the bilateral hip joint was rneasured.
Residual femur abduction during standing was defined as the adduction angle of the non-
amputated hip joint during standing on a non-arnputated foot subtracted from that of the arn-
putated hip during standing on the prosthesis.
To measure lateral stability of the socket, the amputees were asked to walk on a walk-
way containing a force plate (Kistler, Switzerland) at their most comfortable speed until five
appropriate ground-reaction fbrce patterns were obtained. The signals from the force plate
were transmitted to an A/D converter and personal computer to calculate the forces of the
vertical, right-left and front-back directions. Data were normalized by the individual's body
weight and tirne of the stance phase and were then averaged for each amputee [16]. Lat-
eral force ratio was defined as the lateral fOrce during mid-stance on the amputated side di-
vided by that on the non-amputated,
To examine the metabolic efficiency of gait with the presthesis, the physiological cost
index (PCI) was obtained [15] : [heart rate at the endofa three minute gait (beats/min) -
heart rate at rest (beats/min)] / velocity (m/min). Heart rates at rest and at the end ofa
three minute walk were monitored by a telemetric electrocardiograph (NEC-Sanei, Japan) ,
axid velocity was obtained by the distance of a three minute gait at the subject's most comfort-
able speed.
Data were presented as mean ± standard deviation, and statistical analyses were proc-
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
112 K HAcmsuKA et aJ
essed with SPSS 8.0 J for Windows. A t -test was apptied to compare the differences in age,
duration after amputation, and stump length ratio between the IRC and QL groups, and Mann-
Whitney test was used to compare the differences in the subjective evaluation and objective
measurement between the two groups. A multiple regression analysis was perfbrmed to de-
termine what factors were related to PCI. Its predicting variable was PCI, and explanatory
variables were stump length ratio, lateral force ratio during mid-stance, socket type, femur
position at the medial-lateral direction, residual femur abduction during standing, gluteal me-
dial muscle atrophy ratio, and age. As the socket type was c'ategorical, the data were con-
verted to 1 fbr the IRC socket and 2 fbr the QL socket befbre analysis, Differences with P-
value of less than O.05 were regarded as significant.
Results
1 . Subjects'characteristics
There were no significant differences in age and stump length ratio between the IRC and
QL groups (Table 1 I t-test,P >O.05) . No difference in causes ofamputation was fbund :
trauma fbr 5 subjects and atherosclerosis obliterans for1 subject in each group. Although
duration after amputation of the IRC group was longer than that of the QL group <t -test, P<O.05) ,
subjects of the two groups were eligible for this comparative study,
2 . Subjective evaluation of the IRC and QL sockets
Selfassessments revealed that the IRC socket was significantly more comfortable to sit
on a chair but had a significantly poorer appearance than the QL socket (Table 2 ; Mann-
Whitney test, P<O.05). The IRC socket was also comfbrtable to wear and to go up and
Table L Subjects'characteristics
RC greup
Numbcr dn]afc:fomaJe)Age (years)Causesofarnputatien
Trauma
ASO
Duration after amputation (years) t11Stump length ratio
6 (6:O)46.2 ± 13.2
5
1
7.7 ± 5.5 *
O.70 ± O.14
QL group
6 (5:1)38,8 ± 14.2
5
1
3.9 ± 5.6
O.67 ± O.11
Age, duTation, and ratio are presented as mean ± standard deviation, 1 ) : ratio of the length from the ischia] tu-
berosity to the tip of the stump divided by the length fron] the ischial tuberosity to the lateral femoral condyle on
the non-arnputated side, IRC group : amputees wearing the ischial-ramal containment socket, QI. group : ampu-
tees wearing the quadrilateral socket. " : t-test ; P < O.OS, the IRC group vs. the QL greup.
ASO : atherosclerosis obliterans.
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
IRC Prosthesis for Transfenioral Amputation 113
Table2. Sub;]¢ ctiye evaluation of the IRC and QL sockets
IRC group QL group
Self-assessment
Comfortable to wear
Ease to swing the prosthesis
Comtbrtable to go up and down staiTs
Comfortable to sit on chair
Appearance of thc socket
Donning and doffing the socket
Specialist-assessrnent
Hip abduction during swing phase
Truncal sway during stance phase Lumbar lordsis at heel off
3,8 ± O,44.o ± e.o
4.0 ± O.O4.0 ± O.O *
3.0 ± O.6 *
4,O ± O.O
3.8 ± O,4
3.7 ± O.5
4.e ± o.o *
3,3 ± O,53,7 ± O,8
3.2 ± 12
3.2 ± O,4
4.0 ± O.O
4.0 ± O.O
3.7 ± O,5
3,2 ± O,4
3.2 ± 0,4
Total score 34.2 ± 12 * 31.3 ± 2.5
Data are presented as mean ± standard deviatjon.
Mann-Whitney test, the IRC group vs, the QL group ;*:p<e.os.
down stairs, but not significantly.
Specialist-assessments indicated the IRC socket induced less tmncal sway durifig stance
phase and less lumbar ]ordosis at heel off than the QL socket (Mann-Whitney test, P <O.05) .
3 . 0bjective measurement of amputees wearing the IRC and QL sockets
By computed tomography, femurs in the IRC and QL sockets were kept in almost the
same position at the anterior-posterior direction (Table 3 ; Mann-Whitney test, P >O.05) , but
the QL socket sifted the femur significantly more laterally at the medial-lateral direction
(Mann-wnitney test, P<O.05), As the femur position in the IRC socket and no space be-
tween the stump and socket were confirmed by computed tomography, the IRC sockets were
regarded as fitting well,
There was no significant difference in gluteal medial muscle atrophy between the groups
(Mann-Whitney test, P >O.e5) ,
During one foot standing on the prosthesis, subjects wearing the IRC socket had their re-
sidual femur significantly less abducted than those wearing the QL socket (Mann-Whitneytest, P <O.05) ,
and when walking on the force plate, subjects wearing the IRC socket had a
tendency of being less pushed toward the amputated sjde during mid-stance. However, no
significant difference in PCI was fbund between the IRC and QL groups (Mann-Whitney test,
P >O.05) ,
4 . Factors affecting PCI
Multiple regression analysis disclosed that stump length ratio and lateral fOrce ratio dur-
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
114 K HACHTSUKA et at
ing mid-stance were significant explanatory variables fbr PCI (adjusted R square : O.87,
F-value:11,85,P <O.05), that is, the longer the stump length and the less the lateral fbrce
toward the amputated side during mid-stance, the less oxygen consumption of gait (Table 4 ) .Socket type was not an important factor at the most comfortable speed.
Table 3. Measurements of amputees wearing the IRC and QI・ sockets
IRC group (2L group
Femurposition in thethigh (%) 1) Anterior-postcriordircction
Mcdial-lateral direction 2)
3)Gluteal mcdial musc]e atrophy ratio
Residual t'emurabduction during standing Cdegree)4)Lateral force ratio during mid-stance 5)
Physiological eost index 6}
42.9 ±- 3.4
57.4 ± 4,5 *
O.75 ± O.10
3.0 ± 2.5 *
128 ± O.11
O,48 ± O.23
46,7 ± 2.2
67.4 ± O,9O.74 ± O,08
11.0 ± 6.0
1.44 ± O.16
055 ± O.18
1) : Distance betwcen the anterior edge of the thigh and femur divided by the anterior-posterior dimension of the
thigh 4 cm below the great trochanter on thc amputated side and multiplied by 100, 2) : distance betwccn the me-
dial edge of the thigh and femur divided by the medial-lateral dimension of the high 4 cm below the great tro-
chanter on the amputated side and multiplied by 1OO, 3) : ratio of the gluteal medial muscle area on the amputated
side at thc slicc containing the inferior anterior iliac spine to that on the non-amputated side, 4) : adduction angle
of the non-amputatcd hip joint during standing subtractcd from that of the amputated hip joint, 5) : ratio of the
lateral force on the force plate during mid-stance on the amuputatcd side to that on thc non-amputated side, 6) :
(heart rate during walking-heart rate at rest)/walking speed (m,lmin). Mann-Whitncy test, the IRC group vs.
theQL group;* : P<O.05.
rl"able 4. Factors aiTecting the physiological eost index
Explanatory variables B SEB Beta P-value
Stump length ratio
Lateral force ratio during mid-stance
Socket type *
Femur position in the inedial-lateraldircction
Hip abduction during standing
Gluteal medial muscle atrophy ratio
Age (year)
-O.953
O.788
-O,328
1.721
O,O16
-O.268
<-0.UOI
O.275O.262O,148O.835O.O08O,327O,O02-O.575
O.621-O,869
o,s2e
O,500-O.120-O.079
O.026O.040O.091O.108O.104O.459O.580
Multiple regression analysis for predieting PCIlmultipie R : O.977, R square : O,954, adjusted R square : O.873 ;
F-value:11,849, and P-value:O,Ol5, ':socket
types were converted into 1 fOr the IRC socket and 2 for the
QL socket,
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
IRC Prosthesis for Transfcmoral Amputation 115
Discussion
Subjects in this study were carefu11y selected without prejudice according to the inclu-
sion and exclusion criteria, and no physical or mental disturbances except transfemoral ampu-
tation affected the subjective eyaluation and objective measurement of the sockets. All the
IRC and QL prostheses fitted very well, because some sockets were remade or modified be-
fbre the beginning of this study. The evaluation and measurement used in this study are es-
tablished methods or based on preceding studies.
However, there are still two issues : small number of subjects and comparison of self-
assessments. With regard to the number of subjects, we are unlikely to add new amputees to
the subjects of this study within a few years, because recent patients who are Qbliged to un-
dergo a transfemoral amputation are old, have one or two complications hindering their walk-
ing, and do not fu1fi11 our inclusion criteria. However, although the sample size is small, the
results obtained from this study properly explain features of these two types of sockets, and
because of this, the limited number of the subjects may be permitted. In the self-assessment,
the IRC group had already used the QL socket before a trial of the IRC socket, and were able
to compare the features of the two sockets, whereas the QL group only had experience using
the QL socket. As judgement criteria of the IRC and QL groups may not be identical in
comparison with seliassessments, it may not be rational to examine the difference in self-
assessments between the two groups. HQwever, this bias does not cause a false positive er-
ror and allows us to compare self-assessments between the groups. The reasons are as fol-
lows : the QL group possibly had a tendency toward rating self-assessment items higher and
the IRC group lower, because the amputee generally prefers a familiar prosthesis, even if it
does not fit well, and the QL group was accustomed to the QL socket;this tendency may re-
duce differences between the two groups.
The IRC socket is comfortable to wear, to go up and down stairs, and to sit on a chair,
and decreases lumbar lordosis at heel off, These results may derive from features of the
socket design (Fig. 2):wide anterior-posterior dimension and no typical ischial seat [4].The wide anterior-posterior dimension of the IRC socket reduces pressure against the Scarpa' s
triangle and maintains a sufficient space for quadriceps, hamstrings, and gluteal muscles.
An indentation of the posterior wall instead of the ischial seat gently supports the ischial tu-
berosity. This is one of the most important advantages of the IRC socket.
While lying on one's back and during one foot standing on the prosthesis, the stump is
maintained significantly more rnedially in the IRC socket, but while actually walking, the ]at-
eral fOrce toward the amputated side during mid-stance is not significantly less in the IRC
group. This means that the IRC socket design is usefuI for preventing the stump from ab-
ducting during the stance phase but is unable to prevent abduction completely. The stump
abduction may be affected by an amputee's habitual walking pattem, gluteal medial muscle
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
116 K HAcmsuKA et at
atrophy, and method of transfemoral amputation. If an amputee has already acquired trunk
lateral lean, which is one of the abnormal gait patterns frequently observed in transfemoral
amputees, the IRC socket can not improve the gait pattern. All the subjects in this study had
already had gluteal medial muscle atrophy by computed tomography, and a transfemoral am-
putee with gluteal medial muscle atrophy may not gain complete lateral stability during the
stance phase, even if the IRC socket is applied, As the adductor masnus muscle has a major
mechanical advantage in holding the thigh in its normal anatomical position, loss of function
of the adductor magnus muscle after a transfemoral amputation leads to abduction of the re-
sidual femur [17]. A preserved muscie at the time of transfemoral amputation, which
maintains the function of the ,residual femur adduction, prevents excessive abduction of the
stump, and the IRC socket in addition to the preserved muscle will brjng sufficient lateral sta-
bility during the stance phase.
PCI, an indirect simple method to evaJuate the approximate oxygen consumption of gait,
showed no difTerence between the IRC and QL groups. Although PCI may not be sensitive
enough to detect a subtle difference of oxygen consumption, we think that there is no great
difference in the metabolic efficiency of gait at the most comfortable speed, because truncal
sway during the stance phase and lateral force ratio during mid-stance are not significantly
improved by the IRC socket. By multiple regression analysis, it was seen that the most im-
portant factor re]ated to PCI is stump length ratio and lateral force ratio during mid-stance,
and the socket type is not so impertant as these factors. On the other hand, improved meta-
bolic efficiency of gait by the IRC prosthesis has been reported [12 , 13]. If the QL pros-thesis is remade or modified befOre entry of measurements and its optimal fit is confirmed
clinicatiy and by computed tomography, this process will possibly diminish any difference in
metabolic efiiciency between the IRC and QL groups. However, a significant difference be-
tween the IRC and QL sockets may be obta{ned if oxygen consumption of gait with the IRC
socket is measured at a faster speed.
Some subj'ects complained of the poor appearance of the IRC socket. The standard
IRC socket has a lateral wall high above the great trochanter, which puts pressure on the pel-vis. A bulge is noticed at this portion, especially when a transfemoral amputee puts onjeans.
As the ponion is not rnandatory to keep the stump in a normal position from our clinical ex-
perience, we propose that the height of the lateral wall to be within 5 cm above the great tro-
chanter, and that its hard frame at the great trochanter be trimmed away,
The results of this study reveal that the advantages of the IRC socket are a comfbrtable
feeling, that is, comfortable to wear, to go up and down stairs, and to sit on a chair, and a low
pressure against the ischia] tuberosity, but that metabolic eflliciency of the IRC socket evalu-
ated with the PCI at the most comfortable speed is not superior to the QL socket. These re-
sults adequately explain clinical impressions of the IRC prosthesis. In a future study, the
IRC prosthesis including an Intelligent Knee Joint, which enables a transfemoral amputee to
walk at various speeds with ease, will clarify the metabolic efficiency of gait.
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEHAssociation ofHealth Sciences
IRC Prosthesis for Transl'emoral Amputation 117
Acknowledgements
The authors thank T.Shukuwa, CPO for participation in the clinical study, and
O. Yoshimura, Dr. I. Shigenaga and H. Ohkawa, RPT for cooperation in data collection.Dr,
References
1 . MooncyV&Quigley MJ (1981) : Prostheticmanagement, In : Atlas ofLimbProstheties (Ameri- canAcademyofOrthopedicSurgcons,ed). CVMosby,St. Louis pp384-401
2 . Leonard JA Jr & MeierRH III (1990) : Prosthetics. In : Rehabilitation medicine : principles and
practice (DeLisaJA,ed). Lippincott,Philadelphia pp330-34S
3 . LongIA(1975) : Allowingnormaladductionoffemurinabove-kneeamputations. OrthotProsthet
29: 53-54
4 . LehneisHR (1985) : Beyondthequadiilateral. C]inProsthetOrthot 9: 6-8
5. LonglA (1985): Normal Shape-Normal Alignment (NSNA) above-kneeprosthesis. ClinPros-
thetOrthot 9: 9-14
6 . SabolichJ (1985) : Contouredadductedtrochanteric-controlledalignmentmethod (CAT-CAM) :
introductionandbasicprinciples. ClinProsthetOrthot 9: 15-26
7 . Prosthetic Consultants Incorporated of Akron (1986) : Manual for use of the Shamp brim fbr the
Narrow M-L above-knee prosthetic socket. The Ohio Willowwood Co., Ohio
8 ・ Schuch CM (1988) : Report flrom international workshop on above-knee fitting and aligninent tech-
niques. ClinProsthetOrthot 12: 81-98
9 . Schuch CM (1988) : Modem above-knee fitting practice (a report on the ISPO workshop on above-
kneefittingandalignmcnttechniquesMayl5r19,1987). ProsthetOnhotInt 12: 77-'90
10. Pritham CH (1990) : Biomechanies and shape of the above-knee socket considered in light of the is-
chialcontainmentconcept, ProsthetOrthotlnt 14: 9-21
11 ・ Mitchell CA &Verslu{sTL (1990) : Management of an abovc-knee amputee with complex medical
problemsusingtheCAT-CAMprosthesis. PhysTher 70: 389-393
12. Fiandry F, Beskin J, Chambers RB et al (1989) : The effect of the CAT-CAM abovc-knee prosthe-
sisonfunctionalrehabilitation. ClinOrthop 239: 249-262
l3・ Gailey RS, Lawrence D, Burditt C, Spyropoulos P, Newell C & Nash MS (1993) : The CAT-CAM
socket and quadrilateral socket : a comparison of energy cost during ambulation. Prosthet Orthot '{nt
17: 95-100
14. Hachisuka K (199e) : Clinical indication of the ischial-ramal containment socket for the above-kncc
arnputation. JpnjProsthetOrthot 6: 303"308 (inJapanese)15 . MacGregor J (1979) : The objective measurement of physical performance with long terrn arnbula-
tory physiological surveillance equipment (LAPSE) . in : Proceedings of3 rd Internationa] Sympo-
sium on Ambulatory Monitoring (Stoot FD, Raftery EB & Goulding L, ed), Academic Press,
The UOEH Association of Health Sciences
NII-Electronic Library Service
The UOEH Assooiation of Health Soienoes
118 KHA 〔:HISUKA et ai
London pp 29L39
16 .Ehara Y (1989): Analysis of biomcchanical factors. In : Clinical a皿alysis o 「gait(Tsuchiya K ,
ed ), Ishiyaku−Shuppan, Tokyo pp 61 − 94 (jn Japanese)
17.Gottschalk FA & Stills M (1994): The biemechanics of transfemoral amputation . Prosthet Orthot
lnt l8 : 12 − 17
坐骨収納型大腿義足の 主観的お よび客観的評価
蜂須賀研二1,梅津 祐
一
,緒方 甫L
, 大峯
有薗 秀昭3
⊥
産業医科大学医学部 リハ ビ リテーシ ョ ン 医学教室
2産業医科大学病 院 リハ ビ リテ
ーシ ョ ン 部
3有薗製作所
三 郎2
, 新小 田幸一
要 旨 : 坐 骨収納型 (IRC )大腿義足 が 四 辺形 (QL)大腿義足 よ りも優 れ て い る か 否 か を明
らか に す る 目的で ,IRC義足使用者 6 名, QL 義足使用者 6 名 を対象に して,義足 の
主観的評価,CT を用 い た大腿骨の 位置, X 線写 貞に よ る大腿骨内転角 , 立脚 中期 の
床反力側方成分 , physiological cest index (PCI )を 用 い た エ ネル ギー
効率 の 測定を行
っ た.H之c 義足 は 有意に主観的装着感 に優 れ,背臥位 と 立位 で は 断端 は内転位 に 保た
れ て い た (Mann −Whitney tcst, P 〈 0.1)5),実際 の 歩行 で は,立脚 1「1期の 側方成分は減
少す る傾向 に あ っ たが有意差はな く,また,PCI は両群問で有意差 は なか っ た (Mann −
Whinlcy tcst, P > 0.05).重 回帰分析 に よ れ.ば
, PCI を予測 す る 宥意 な 因了
.は ソ ケ ッ ト
の タ イ プ で は な く,断端長 と立 脚中期成分 の 2 因子 で あ っ た.従 っ て ,IRC 義足 は装
着感 に優 れ,断端 を内転位 に保 つ が,最適 歩行速度 の もとで はエ ネル ギ ー効率が優
れ て い る わけで は なか っ た.
JUOEH (産業医大誌)21 (2);107− 118 (1999)
N 工工一Eleotronio Library