69
Stereotactic Breast Biopsy Design and Testing Robert J. Pizzutiello, Jr., F.A.C.M.P. Upstate Medical Physics, Inc. 716-924-0350

Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Stereotactic Breast BiopsyDesign and Testing

Robert J. Pizzutiello, Jr., F.A.C.M.P.Upstate Medical Physics, Inc.

716-924-0350

Page 2: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Overview

uWhat is a SBB?u Purpose of ACR-SBBAPu Requirements for accreditationu Technologist’s QC Testsu Medical Physicists QC Tests

Page 3: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Detection vs. Diagnosisu Detection

BSEPhysician Physical ExamMammographyUltrasound

u Tissue DiagnosisBiopsyCytologyHistology

333

Page 4: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

SBB- Imaging andDose Considerations

u Localization (not detection) ofabnormalities

u Limited numbers of “normals”u Complex backgroundu Microcalcifications, massesu Limited FOV neededu Radiation risk to selected “at risk”

population

Page 5: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

SBB X-ray Systemsu Dedicated prone units: similar to

Diagnostic Mammography Unitsu Smaller FOV required, 16o target

angleu Light Field may be replaced by two

illuminating lampsu SID may be longeru AEC may or may not be available

Page 6: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Types of SBB Equipment

uDedicated prone tablesuAdd-on stereo unitsuScreen-film imaginguDigital Image receptors

Page 7: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

The CCD ImageReceptor

The CCD ImageReceptor

u Charge-Coupled Deviceu An integrated circuit (chip)

silicon wafer

u About the size of a postage stampu Converts light into electronic

image

detectors

amplifiers

Page 8: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

LGE NL 11-22-96 #4

CCD ImageReceptors

CCD ImageReceptorsu5cm x 5cm FOV CCD, typical

uLoRad DSM (below) 5 cm x 5 cmuGE Senovision (right) 8 cm x 8 cm

Page 9: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Display Gradient

1 2 5 10 20 50 1000

0.05

0.1

0.15

0.2

0.25

0.3

Relative Exposure

Gra

dien

t

Characteristic Curve

1 2 5 10 20 50 1000

1

2

3

4

Relative Exposure

Opt

ical

Den

sity

Higher Film Contrast means less tolerance for exposurevariation

Exposure Latitude

Page 10: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Conventional x-rayexposure creates an aerialimage

Readout CCD to computer

Minify light image to CCD size

Intensifying screen convertslatent x-ray image to visiblelight image

Display, manipulate, archivedigital image

Page 11: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Side View

Focal Spot X-rayTube

Small areaCollimator

CompressedBreast

Phosphor

Mirror

CCD

Front View

DMAA/D

Optical coupling/mirror systemLight reflection from phosphor

Lens

CCD

Page 12: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Side View

Focal Spot X-rayTube

Small areaCollimator

CompressedBreast

PhosphorFiber Optic

CCD

Front View

DMAA/D

2:1 fiberoptic taper demagnificationLight transmission through phosphor

Page 13: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Digital Image Quality

u Contrastu Bluru Noiseu Artifactsu Dose

Page 14: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)
Page 15: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)
Page 16: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Factors Affecting BreastDoseu kVp, mAs

u exposure time (film only)u breast thicknessu breast composition (dense or fatty)umultiple exposures

u digital image processing does NOT affectdose

u optical density of film (if hardcopy is used)does NOT affect dose

Page 17: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

To Minimize Breast Doseu Develop and maintain a good technique chartu Obtain manufacturer’s suggested techniquesu Evaluate image quality at different mAs values

(Technologist and Medical Physicist)u Moderately higher mAs will reduce image

noise, but increase doseu Insufficient mAs will produce a noisy (grainy)

image, but can be made to appear “wellexposed” with window/level control

u Excessive mAs images may also appear “OK”with window/level adjustment

u Minimize retakes

Page 18: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Principle ofStereotacticLocalization

u2-D localization -planar view

u3-D localization -triangulation neededfor depth coordinate

Page 19: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Our first exposure to StereoLocalization Techniques

Page 20: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

How many pins are left?

Page 21: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

The origin of the 15o stereo shift?

Page 22: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

+ 15° - 15°

+ 15° View - 15° View

+ 15° - 15°

+ 15° View - 15° View

Before Lesion After Lesion

Page 23: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Methods of ImagingGuided Breast Biopsy

u Ultrasound guided, hand-held needleu Stereotactically guided core biopsy

Not visible on ultrasoundLocalize with millimeter precision

Page 24: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Core Biopsy

Page 25: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)
Page 26: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

ACR-SBBAPHistory

u Committee convened Fall, 1995Develop professional standardsDevelop SBBAP materials for facilities

u Pilot program 1st quarter, 1996u Announced at ACR Breast Cancer

Meeting (April, 1996)u Reviewers trained

Page 27: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

ACR-SBBAPu Modeled after ACR-MAPu 1996 vs. 1987u Personnel qualificationsu Equipment performanceu QCu Procedure verification

(through clinical image evaluation)u Image quality (phantom images)u Dose

Page 28: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Personnel QualificationsMedical Physicist

u Board Certification or alternate requirementsu 15 hours CE in Mammo Physics every 3 yearsu < 6/1/97

3 hands-on SBB MP Surveys, or1 hands on SBB MP Survey under guidance ofQualified SBB MP

u > 6/1/971 hands-on SBB MP Survey under guidance

u At least 1 SBB MP Survey per yearu 3 hrs CE in SBB Physics every 3 years

Page 29: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Physician Qualificationsvs.

Practice Model

Where a radiologist or surgeon (or otherphysician) are practicing independently, the

expertise in the diagnosis and management ofbreast disease of an individual physician mayprovide the patient with an equivalent benefit.

In a collaborative practice, the patient derives thebenefit of consultation and collaboration from the

radiologist and surgeon (or other physician)working together.

Collaborative Independent

Page 30: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Physician Credentialsu Mammogramsu SBBu Training, Experienceu Category I SBB coursesu QAu Radiation Physics Trainingu Supervision of RT and MPu Post biopsy recommendationsu Lesion identification at time of biopsy

Page 31: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Approximate StatusSeptember, 1998

u 377 facilities appliedu 278 facilities accreditedu 382 units applied - activeu 105 units applied - pendingu 182 units accredited on initial attempt

Page 32: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

147 units with deficiencies oninitial evaluation

Clinical onlyClinical in combinationPhantom onlyPhantom in combinationDose

62 (42%)34 (23%)31 (21%) 4 ( 3%)16 (11%)

96 units passed after re-application

Page 33: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

The latest word...u No longer accepting optical disk or

diskette. Hard copy images only.u FDA will implement regulations

mandating accreditation of facilitiesif they do not comply voluntarily

u Check TLD technique (10% failurerate for dose)

u QC Manual available early 1998

Page 34: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Darkroom CleanlinessProcessor Quality ControlScreen CleanlinessViewboxes & Viewing Conditions

Phantom ImagesVisual ChecklistRepeat AnalysisAnalysis of Fixer RetentionDarkroom FogScreen-Film ContactCompression

DailyDaily

WeeklyWeeklyWeeklyMonthly

QuarterlyQuarterly

Semi-annuallySemi-annuallySemi-annually

D (SF only)D (SF only)W (SF only)W (SF only)

WeeklyMonthly

QuarterlyQ-(SF only)S - (SF only)S - (SF only)

Semi-annually

Mammo SBB

QC Tests Common to Mammographyand SBB Minimum Testing Frequencies

Page 35: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Zero Alignment Test(only on some units)

Localization Accuracy Test (in Air)

Phantom Image Quality TestHardcopy Output Quality(if hard copy is produced from digital data)

Visual Equipment CheckRepeat AnalysisCompression Force Test

Before each patient

Daily

WeeklyMonthly

Monthly

Semi-annuallySemi-annually

QC Tests Unique to SBBMinimum Testing Frequencies

Page 36: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Zero Alignment Test(only on some units)

Localization Accuracy Test (in Air)

Phantom Image Quality TestHardcopy Output Quality(if hard copy is produced from digital data)

Visual Equipment CheckRepeat AnalysisCompression Force Test

An overview of the QC Tests Unique to SBB(Radiologic Technologist)

RT

Page 37: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Zero Alignment Test

u Perform before each patientu Verify that zero coordinate is

accurateu Assures that stereotactic unit is not

improperly installed

B

RT

Page 38: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Localization Accuracyu Closed loop system testu Position needle to a known

coordinateu Digitize position of needle tipu Targeting software calculates

position of needle tipu Coordinates should be identicalu ± 1.0 mm sphere

D

RT

Page 39: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Phantom ImageQuality Evaluation

FiberDiameters

(mm)1.561.120.890.750.540.40

SpeckDiameters

(mm)0.540.400.320.240.16

Mass Diameters (mm) 2.00, 1.00, 0.75, 0.50, 0.25

D-102

Page 40: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Phantom ImageQuality Evaluation

NuclearAssociates Digital

Mini Phantom

MammographyAccreditation

Phantom

W

RT

Page 41: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Fibers

Specks

Masses

1.561.120.80.750.540.540.40.320.240.16

21

0.750.50.25

xx

0.930.740.540.54

x0.320.240.2x1

0.750.50.25

ACRAccreditation NA Digital

Page 42: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Minimum PassingPhantom Image Scores

FibersSpecksMasses

ACR-MAP

Screen/film

4.03.03.0

AccreditationPhantom

Digital

5.04.03.5

Mini-Phantom

Digital

3.03.02.5

Page 43: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Phantom Imaging: a common avoidablefailureu NAD Digital Mini Phantom

1st image (image quality)2nd image (TLD)

u Mammo Accreditation Phantom4 images for image quality5th image for TLD

u OK to window/level digital imagesu Use grid (or not) per clinical technique

Page 44: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)
Page 45: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

HardcopyOutput Qualityu Laser or multiformat camerau Evaluate SMPTE Test Pattern, if

availableu Record window width, levelu Produce hardcopyu Measure OD at 4 consistent

locationsu Record and monitor for consistency

M

RT

Page 46: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Visual Checklist

u Use ACR checklist or equivalentu Lights, switches, motion, accessoriesu Customize for your

machine/roomu Documentation

(date, initials)

M

RT

Page 47: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Repeat Analysisu Count repeated and rejected film by

category and tabulateu Include a log of images repeated, but

not discardedu Overall repeat rate should be < 5%u Document analysis and corrective

action - even if your repeat rate islow

Q

RT

Page 48: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Compression Forceu Bathroom scale or

compression gaugeu Measure maximum

compression in manualand power modes

u The scale should read25-40 pounds inautomatic mode

u Documentation

S

RT

Page 49: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Additional Technologist’sQC Tests (Screen-Film

only)TEST

Darkroom Cleanlinessprocessor QC

Screen CleanlinessViewboxes & Viewing

ConditionsFixer Retention Analysis

Screen-Film ContactDarkroom Fog

FREQUENCYDailyDaily

Weekly

WeeklyQuarterly

Semi-AnnuallySemi-Annually

Page 50: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

SBB Annual Medical Physics Surveyu SBB Unit Assembly Evaluationu Collimation Assessmentu Focal Spot Performance and System Limiting

Resolutionu kVp Accuracy and Reproducibilityu Beam Quality Assessment (HVL)u Automatic Exposure Control System

Performanceu Uniformity of Screen Speed or Digital Fieldu Breast ESE, AGD, AEC Reproducibilityu Image Quality Evaluation (phantom)u Artifact Evaluationu Localization AccuracyMP

Page 51: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Assembly Evaluationu Free-standing unit is mechanically stable

u All moving parts move smoothly, withoutobstructions to motion

u All locks and detents work properly

u Image receptor holder is free from vibrations

u Image receptor is held securely by assembly inany orientation

MP

Page 52: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Assembly Evaluationu Image receptor slides smoothly into holder

assembly

u Compressed breast thickness scale is accurateto ± 0.5 cm, reproducible to ± 2 mm

u Patient or operator is not exposed to sharp orrough edges or other hazards

u Operator technique charts are posted

u Operator protected by adequate radiationshielding

MP

Page 53: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

CollimationuDoes the x-ray beam exceed the image

receptor?

Note: X-rays beyond the digital imagereceptor will not be seen on the monitor

uDoes the biopsy window align with the imagefield of view?

MP

Page 54: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Performance - SystemLimitingResolutionu Line Pair Test Patternu Use film to test x-ray

machineu Use CRT image to test

“system”u Technique, clinical kVpu Scoring the image

Lines distinct over 1/2 length

Page 55: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

kVp Accuracy -Reproducibility

u Verify that actual kVp’s are the same as the

indicated kVp’s

u Range of clinical kVp values

u Accuracy within 5%

u Reproducible CV < 0.02

MP

Page 56: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Beam Quality (HVL)u Thickness of aluminum

to reduce radiationexposure by one-half

u Affects contrast anddose

u Used in dosecalculation

u minimum = kVp/100

MP

Page 57: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

AEC System Performanceu AEC available on some digital SBB unitsu Performance Capability

Record signal level asfunction of thicknessand technique

u Monitor exposure timeu Performance Capability

(4,6,8 cm)u Provide suggested

technique chart

MP

Page 58: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Varying thicknesses ofbreast equivalent material

Page 59: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Uniformity of ScreenSpeed or Digital Field

u Image a uniform phantomu Screen Film systems

Each cassette produces the sameoptical density under the sameconditions

u Digital SystemsDigital detector produces uniform signalvalues across the field of view

MP

Page 60: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)
Page 61: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Phantom Image Qualityu Same procedure as

for technologistsu Medical Physicist

reviews scoringprocedure andchecks forconsistency

u Uses techniquefactors for dosedeterminationMP

Page 62: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Breast Entrance Exposure,AGD

u Data per technique chartu Measure ESEu HVL determines DgNu AGD = ESE * DgNu AGD < 300 mradu Dose and Optical Density

MP

Page 63: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Artifact EvaluationUnwanted irregularity not caused

by structures of interestCauses (Digital)

Digital Image ReceptorCauses (Screen-Film)

Lint, dust, static, filter,compression device, grid,BSD, screen, film,cassette, phantom

Common CausesUnwanted objects in x-ray beamMP

Page 64: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Source of Artifacts (Film)

Direction of Film Feed

Page 65: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Source of ArtifactsChange of orientation

Processor related

Either parallel orperpendicular todirection of travel

Plus or minus density

Rollers

Guide shoes

Replenishment stream

No change of orientation

X-ray tube

Filter

Compression device

BSD

Grid (stop motion)

Cassette

Page 66: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Targeting Accuracyu Performed annually by technologist

under supervision of medical physicistu Position gel-type phantomu Image, target and sampleu Result: was the lesion collected?

MP

Page 67: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

QC Program Review

u Review procedures(ACR SBB-QC Manual)

u Review documentationu Answer questionsu Written recommendations

MP

For all Technologist QC Tests

Page 68: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Role of the Surgeon inQuality Control

u Understand the importance of QC in SBBu Assures that personnel remain qualifiedu Support QC activities

Allow enough time for QCProvide for QC trainingPeriodically check that QC is done as required

u Confer with medical physicist annuallyu Assure that follow-up is done if the QC

program indicates corrective action is requiredu Accreditation

Page 69: Stereotactic Breast Biopsy Design and Testing · Localization Accuracy Test (in Air) Phantom Image Quality Test Hardcopy Output Quality (if hard copy is produced from digital data)

Summary

uWhat is a SBB?u Purpose of ACR-SBBAPu Requirements for accreditationu Technologist’s QC Testsu Medical Physicists QC Tests