885
Spectra Optia ® APHERESIS SYSTEM SERVICE MANUAL

Spectra Optia Service Manual - English, EN...Spectra Optia Apheresis System Service Manual v Fan 3 failed. ..... 4-180 Fan 3 Fluid balance change caused fluid balance to exceed limits

  • Upload
    others

  • View
    31

  • Download
    0

Embed Size (px)

Citation preview

  • Spectra Optia®APHERESIS SYSTEM

    SERVICE MANUAL

  • Spectra Optia®Apheresis SystemService Manual

    Part No. 777401-0422011-01

  • ©2011CaridianBCT, Inc.

    CaridianBCT, Inc.10811 W. Collins AvenueLakewood, Colorado 80215USAUSA Phone: +1.877.339.4228Phone: +1.303.231.4357USA Fax: +1.866.715.6768Fax: +1.303.542.5215

    CaridianBCT Europe N.V.Ikaroslaan 411930 ZaventemBelgiumPhone: +32.2.715.05.90Fax: +32.2.721.07.70

    www.caridianbct.com

  • Contents

    Preface ................................................................................................................. 15

    1: Operational DescriptionOperational Description ..................................................................................................................................... 1-2

    System Components ........................................................................................................................................... 1-3

    Modes of Operation and Procedure States .......................................................................................................... 1-5

    User Interface ..................................................................................................................................................... 1-7

    Exchange Disposable Tubing Set ...................................................................................................................... 1-11

    Collection Disposable Tubing Set .................................................................................................................... 1-15

    Therapeutic Plasma Exchange .......................................................................................................................... 1-19

    Mononuclear Cell Collection ........................................................................................................................... 1-22

    Red Blood Cell Exchange Procedure Types ...................................................................................................... 1-24

    RBCX Functional Description ......................................................................................................................... 1-26

    2: System DescriptionSystem Description ............................................................................................................................................. 2-2

    E-Box and Computer Systems ............................................................................................................................ 2-3

    Control CCA ............................................................................................................................................. 2-5

    Control Computer CCA ............................................................................................................................ 2-7

    Control Ethernet CCA .............................................................................................................................. 2-8

    Other Control Stack CCAs ........................................................................................................................ 2-9

    Safety CCA .............................................................................................................................................. 2-11

    Safety Computer CCA ............................................................................................................................. 2-12

    Other Safety Stack CCA .......................................................................................................................... 2-14

    AIM System CCAs .................................................................................................................................. 2-14

    Motor Driver CCA .................................................................................................................................. 2-18

    Display CCA ........................................................................................................................................... 2-26

    Top Cap and Motherboard CCAs ........................................................................................................... 2-28

    Spectra Optia Apheresis System Service Manual i

  • Power System .......................................................................................................................................... 2-29

    Cooling Fans ........................................................................................................................................... 2-32

    Display System ................................................................................................................................................. 2-33

    Display Assembly ..................................................................................................................................... 2-35

    Pump System ................................................................................................................................................... 2-38

    Pump Assembly ....................................................................................................................................... 2-41

    Valve System .................................................................................................................................................... 2-43

    Valve Assembly ........................................................................................................................................ 2-45

    Sensor System ................................................................................................................................................... 2-48

    Pressure Sensors ....................................................................................................................................... 2-48

    Reservoir Level Sensors ............................................................................................................................ 2-52

    Reservoir Management ............................................................................................................................ 2-55

    Fluid Sensors ........................................................................................................................................... 2-56

    Return Line Air Detector ......................................................................................................................... 2-58

    Set ID Detector ....................................................................................................................................... 2-60

    RBC Detector .......................................................................................................................................... 2-63

    Fluid Leak Detector ................................................................................................................................. 2-66

    Linear Actuator System .................................................................................................................................... 2-68

    AIM System ..................................................................................................................................................... 2-71

    Camera System ........................................................................................................................................ 2-71

    Lighting System ....................................................................................................................................... 2-73

    Centrifuge System ............................................................................................................................................ 2-81

    Centrifuge Drive Assembly ...................................................................................................................... 2-85

    Filler Assembly ......................................................................................................................................... 2-87

    Centrifuge Motor Controller ................................................................................................................... 2-89

    Mechanical System ........................................................................................................................................... 2-92

    Dual IV Pole ............................................................................................................................................ 2-92

    Wheel and Brake System ......................................................................................................................... 2-95

    3: Software DescriptionSoftware Description .......................................................................................................................................... 3-2

    Base Layer Software States .................................................................................................................................. 3-4

    Power Control State ................................................................................................................................... 3-4

    Startup State .............................................................................................................................................. 3-7

    T1 Pump Power Test State ...................................................................................................................... 3-12

    T1 Device Test State ............................................................................................................................... 3-14

    T1 APC Test State ................................................................................................................................... 3-16

    T1 Centrifuge Power Test State ............................................................................................................... 3-18

    Load Disposable State .............................................................................................................................. 3-21

    RBC Detector Calibration State .............................................................................................................. 3-26

    ii Spectra Optia Apheresis System Service Manual

  • Disconnect Test State .............................................................................................................................. 3-28

    Unload Disposable State .......................................................................................................................... 3-30

    Set Layer Software States .................................................................................................................................. 3-32

    Collection States ...................................................................................................................................... 3-32

    Exchange States ....................................................................................................................................... 3-54

    Protocol Layer Software States .......................................................................................................................... 3-77

    MNC Run State ...................................................................................................................................... 3-77

    RBCX Run State ..................................................................................................................................... 3-82

    TPE Run State ......................................................................................................................................... 3-88

    4: TroubleshootingNon-Alarm Troubleshooting .............................................................................................................................. 4-2

    Touchscreen Troubleshooting ................................................................................................................... 4-2

    General AIM Troubleshooting ................................................................................................................... 4-3

    Troubleshooting Failures During Start Up ................................................................................................ 4-5

    Datalog Files ...................................................................................................................................................... 4-7

    Alarm Troubleshooting ...................................................................................................................................... 4-8

    Spectra Optia Alarms ....................................................................................................................................... 4-11

    AC fluid detector detected fluid too soon. ................................................................................................ 4-12

    AC fluid detector did not detect fluid. ..................................................................................................... 4-14

    AC infusion data was invalid. ................................................................................................................... 4-16

    AC infusion data was invalid. ................................................................................................................... 4-17

    AC occlusion test failed. ........................................................................................................................... 4-18

    AC pump did not stop when commanded. .............................................................................................. 4-20

    AC pump malfunctioned. ........................................................................................................................ 4-22

    AC pump software error occurred. ........................................................................................................... 4-24

    AC pump speed was too high. ................................................................................................................. 4-25

    AC pump turned faster than inlet pump. ................................................................................................. 4-26

    AC volume limit was exceeded. ................................................................................................................ 4-28

    AC was not detected. ............................................................................................................................... 4-29

    AIM subsystem failed. .............................................................................................................................. 4-31

    AIM system's interface measurement data was not updated. .................................................................... 4-33

    AIM system could not establish target interface. ...................................................................................... 4-35

    AIM system detected RBC interface near top of channel. ........................................................................ 4-37

    AIM system did not complete calibration test. ......................................................................................... 4-40

    AIM system did not complete channel imaging test. ................................................................................ 4-42

    AIM system did not complete illumination test. ...................................................................................... 4-44

    AIM system did not complete interface measurement. ............................................................................. 4-46

    AIM system did not detect interface. ....................................................................................................... 4-48

    AIM system did not respond to system command. ................................................................................... 4-51

    AIM system failed calibration test. ........................................................................................................... 4-53

    Spectra Optia Apheresis System Service Manual iii

  • AIM system failed channel imaging test. .................................................................................................. 4-55

    AIM system failed illumination test. ........................................................................................................ 4-58

    AIM system failed to measure interface. ................................................................................................... 4-61

    Air may not have been removed from return line. .................................................................................... 4-64

    Air removal is complete. ........................................................................................................................... 4-66

    Air was detected in return line. ................................................................................................................ 4-67

    Analog sensor data was not available. ....................................................................................................... 4-69

    Bolus volume caused fluid balance to exceed limits. ................................................................................. 4-70

    Calculation of pump rpm was not consistent. .......................................................................................... 4-72

    Canceling custom prime recommendation. .............................................................................................. 4-74

    Cassette mechanism that raises and lowers cassette malfunctioned. .......................................................... 4-75

    Cassette sensor malfunctioned. ................................................................................................................ 4-77

    Cassette software error occurred. .............................................................................................................. 4-79

    Cassette tray was not correctly positioned. ............................................................................................... 4-80

    Cassette was not detected in all positions. ................................................................................................ 4-82

    Cassette was not lowered. ......................................................................................................................... 4-84

    Centrifuge control malfunctioned. ........................................................................................................... 4-85

    Centrifuge did not stop during alarm. ...................................................................................................... 4-87

    Centrifuge did not stop when commanded. ............................................................................................. 4-88

    Centrifuge door lock malfunctioned. ....................................................................................................... 4-90

    Centrifuge door lock power circuit failed. ................................................................................................ 4-91

    Centrifuge door lock sensor malfunctioned. ............................................................................................. 4-92

    Centrifuge door lock software error occurred. .......................................................................................... 4-94

    Centrifuge door lock test malfunctioned. ................................................................................................. 4-95

    Centrifuge door position sensor malfunctioned. ....................................................................................... 4-97

    Centrifuge door was not closed. ............................................................................................................... 4-99

    Centrifuge door was opened or unlocked when centrifuge was spinning. ............................................... 4-100

    Centrifuge exceeded maximum allowed speed. ....................................................................................... 4-102

    Centrifuge motion detected was improper. ............................................................................................ 4-104

    Centrifuge power could not be disabled. ................................................................................................ 4-106

    Centrifuge power could not be enabled. ................................................................................................. 4-107

    Centrifuge power did not turn off. ......................................................................................................... 4-108

    Centrifuge pressure exceeded limit. ........................................................................................................ 4-110

    Centrifuge pressure sensor failed calibration test. ................................................................................... 4-113

    Centrifuge pressure sensor failed post-load test. ..................................................................................... 4-115

    Centrifuge pressure sensor failed zero differential test. ............................................................................ 4-117

    Centrifuge pressure sensor failed zero limit test. ..................................................................................... 4-119

    Centrifuge pressure sensor failed zero range test. .................................................................................... 4-121

    Centrifuge pressure sensor malfunctioned. ............................................................................................. 4-123

    Centrifuge software error occurred. ........................................................................................................ 4-125

    iv Spectra Optia Apheresis System Service Manual

  • Centrifuge speed sensors malfunctioned. ................................................................................................ 4-126

    Centrifuge speed was not sufficient. ....................................................................................................... 4-128

    Centrifuge speed was too high while procedure was paused. .................................................................. 4-130

    Centrifuge was commanded to spin during alarm. ................................................................................. 4-132

    Centrifuge was spinning in reverse. ........................................................................................................ 4-133

    Chamber took too long to fill. ............................................................................................................... 4-135

    Change replacement fluid to RBC for exchange. .................................................................................... 4-137

    Change to patient data caused fluid balance to exceed limits. ................................................................. 4-138

    Check connection from return line to blood warmer tubing. ................................................................. 4-141

    Clamp inlet and return lines. ................................................................................................................. 4-142

    Clamp inlet line. .................................................................................................................................... 4-143

    Clamp return line. ................................................................................................................................. 4-144

    Clamp return line. ................................................................................................................................. 4-145

    Clear saline from drip chamber. ............................................................................................................. 4-146

    Close inlet saline line. Clamp return line. .............................................................................................. 4-147

    Close lines to prepare for tubing set test. ................................................................................................ 4-148

    Close return saline line. .......................................................................................................................... 4-149

    Close return saline line. .......................................................................................................................... 4-150

    Close saline lines and clamp extra remove line. ...................................................................................... 4-151

    Collection bag reached capacity. ............................................................................................................ 4-152

    Collect recirculation flow path was not primed. ..................................................................................... 4-153

    Confirming selection to begin rinseback. ............................................................................................... 4-155

    Confirming selection to disable AIM system. ......................................................................................... 4-156

    Confirming selection to disable RBC detector. ...................................................................................... 4-157

    Confirming selection to disconnect patient and raise cassette. ................................................................ 4-158

    Confirming selection to disconnect patient and raise cassette. ................................................................ 4-159

    Confirm return line did not contain air. ................................................................................................ 4-160

    Confirm saline lines were closed. ........................................................................................................... 4-161

    Control flow was not consistent. ............................................................................................................ 4-162

    Could not confirm saline lines were closed. ............................................................................................ 4-163

    Current run targets could not be attained with run value entered. ......................................................... 4-164

    Current run targets could not be attained with value entered. ................................................................ 4-165

    Current run targets could not be attained with value entered. System paused. ....................................... 4-166

    Custom prime volume was not sufficient. .............................................................................................. 4-167

    Door-closed sensor was not consistent. .................................................................................................. 4-168

    Door lock sensor was not consistent. ...................................................................................................... 4-170

    Fan 1 failed. ........................................................................................................................................... 4-172

    Fan 1 malfunctioned. ............................................................................................................................. 4-174

    Fan 2 failed. ........................................................................................................................................... 4-176

    Fan 2 malfunctioned. ............................................................................................................................. 4-178

    Spectra Optia Apheresis System Service Manual v

  • Fan 3 failed. ........................................................................................................................................... 4-180

    Fan 3 malfunctioned. ............................................................................................................................. 4-182

    Fluid balance change caused fluid balance to exceed limits. .................................................................... 4-184

    Fluid balance no longer exceeded limits. ................................................................................................ 4-186

    Fluid leak detector failed self-test. .......................................................................................................... 4-188

    Hall sensor and encoder in AC pump malfunctioned. ............................................................................ 4-189

    Hall sensor and encoder in inlet pump malfunctioned. .......................................................................... 4-191

    Hall sensor and encoder in plasma pump malfunctioned. ...................................................................... 4-193

    Hall sensor and encoder in replace pump malfunctioned. ...................................................................... 4-195

    Hall sensor and encoder in return pump malfunctioned. ....................................................................... 4-197

    High-level reservoir sensor detected excess fluid. .................................................................................... 4-199

    High-level reservoir sensor detected fluid too soon. ................................................................................ 4-202

    High-level reservoir sensor did not detect fluid. ..................................................................................... 4-204

    High-level reservoir sensor failed fluid check. ......................................................................................... 4-206

    High-level reservoir sensor failed timing test. ......................................................................................... 4-208

    High-level reservoir sensor malfunctioned. ............................................................................................. 4-210

    Hypervolemia condition was possible. ................................................................................................... 4-212

    Hypervolemia limit was exceeded. ......................................................................................................... 4-213

    Hypovolemia condition was possible. .................................................................................................... 4-214

    Hypovolemia limit was exceeded. .......................................................................................................... 4-216

    Identification of tubing set did not match selected procedure. ............................................................... 4-217

    Improper state transition occurred. ........................................................................................................ 4-219

    Inlet:AC ratio data was invalid. .............................................................................................................. 4-220

    Inlet:AC ratio data was invalid. .............................................................................................................. 4-221

    Inlet:AC ratio was above maximum. ...................................................................................................... 4-222

    Inlet:AC ratio was below minimum. ...................................................................................................... 4-223

    Inlet line prime failed. ............................................................................................................................ 4-224

    Inlet line was not clamped. .................................................................................................................... 4-225

    Inlet pressure sensor failed calibration test. ............................................................................................. 4-226

    Inlet pressure sensor failed post-load test. ............................................................................................... 4-228

    Inlet pressure sensor failed zero differential test. ..................................................................................... 4-230

    Inlet pressure sensor failed zero limit test. .............................................................................................. 4-232

    Inlet pressure sensor failed zero range test. ............................................................................................. 4-234

    Inlet pressure sensor malfunctioned. ...................................................................................................... 4-236

    Inlet pressure was too low. ..................................................................................................................... 4-238

    Inlet pump did not stop when commanded. .......................................................................................... 4-240

    Inlet pump malfunctioned. .................................................................................................................... 4-242

    Inlet pump software error occurred. ....................................................................................................... 4-244

    Inlet saline line was obstructed. .............................................................................................................. 4-245

    Inlet saline line was obstructed during rinseback. ................................................................................... 4-247

    vi Spectra Optia Apheresis System Service Manual

  • Interface still took too long to establish. ................................................................................................. 4-248

    Interface took too long to establish. ....................................................................................................... 4-251

    Internal timing was not consistent. ........................................................................................................ 4-254

    Leak was detected in centrifuge. ............................................................................................................. 4-256

    LED sensor malfunctioned. ................................................................................................................... 4-259

    Low-level reservoir sensor continued to detect excess fluid. .................................................................... 4-260

    Low-level reservoir sensor detected air. ................................................................................................... 4-263

    Low-level reservoir sensor detected excess fluid. ..................................................................................... 4-265

    Low-level reservoir sensor detected fluid too soon. ................................................................................. 4-268

    Low-level reservoir sensor did not detect fluid. ....................................................................................... 4-270

    Low-level reservoir sensor failed timing test. .......................................................................................... 4-272

    Low-level reservoir sensor malfunctioned. .............................................................................................. 4-274

    Low-level reservoir sensor malfunctioned. .............................................................................................. 4-276

    Low-level reservoir sensor malfunctioned. .............................................................................................. 4-278

    Low-level reservoir sensor persisted in detecting excess fluid. ................................................................. 4-279

    Minus twelve volt power supply was out of range. ................................................................................. 4-282

    NVRAM safety test failed. ..................................................................................................................... 4-283

    NVRAM test failed. ............................................................................................................................... 4-284

    Open and close centrifuge door. ............................................................................................................ 4-285

    Open inlet saline line. ............................................................................................................................ 4-286

    Open return saline line. ......................................................................................................................... 4-287

    Open saline lines. ................................................................................................................................... 4-288

    Patient's fluid balance may be 10% higher than reported. ...................................................................... 4-289

    Patient's fluid balance may be 10% lower than reported. ....................................................................... 4-291

    Patient's fluid balance may be 15% higher than reported. ...................................................................... 4-293

    Patient's fluid balance may be 15% lower than reported. ....................................................................... 4-295

    Patient's fluid balance may be 5% higher than reported. ........................................................................ 4-297

    Patient's fluid balance may be 5% lower than reported. ......................................................................... 4-299

    Patient fluid balance may be lower than reported. .................................................................................. 4-301

    Patient height was invalid. ..................................................................................................................... 4-303

    Patient height was invalid. ..................................................................................................................... 4-304

    Patient hematocrit data was invalid. ....................................................................................................... 4-305

    Patient hematocrit data was invalid. ....................................................................................................... 4-306

    Patient sex was invalid. .......................................................................................................................... 4-307

    Patient sex was invalid. .......................................................................................................................... 4-308

    Patient TBV was invalid. ....................................................................................................................... 4-309

    Patient TBV was invalid. ....................................................................................................................... 4-310

    Patient weight was invalid. ..................................................................................................................... 4-311

    Patient weight was invalid. ..................................................................................................................... 4-312

    Pause button was touched. ..................................................................................................................... 4-313

    Spectra Optia Apheresis System Service Manual vii

  • Plasma bag reached capacity. .................................................................................................................. 4-314

    Plasma pump did not stop when commanded. ....................................................................................... 4-315

    Plasma pump malfunctioned. ................................................................................................................ 4-317

    Plasma pump software error occurred. ................................................................................................... 4-319

    Plasma recirculation flow path was not primed. ..................................................................................... 4-320

    Plasma remove flow path was not primed. ............................................................................................. 4-322

    Plasma valve malfunctioned. .................................................................................................................. 4-324

    Plasma valve sensor malfunctioned. ........................................................................................................ 4-325

    Plasma valve software error occurred. ..................................................................................................... 4-326

    Plasma valve was not correctly positioned. ............................................................................................. 4-327

    Predicted fluid balance exceeded limits. ................................................................................................. 4-328

    Predicted procedure time exceeded 8 hours. ........................................................................................... 4-330

    Prediction software malfunctioned. ........................................................................................................ 4-331

    Pressure in collect line was too high. ...................................................................................................... 4-333

    Pressure sensor 4 failed calibration test. .................................................................................................. 4-335

    Pressure sensor 4 failed post-load test. .................................................................................................... 4-337

    Pressure sensor 4 failed zero differential test. .......................................................................................... 4-339

    Pressure sensor 4 failed zero limit test. ................................................................................................... 4-341

    Pressure sensor 4 failed zero range test. .................................................................................................. 4-343

    Prime saline has been diverted. Close return saline line. ......................................................................... 4-345

    Prime saline was not diverted. ................................................................................................................ 4-346

    Procedure configuration data was corrupt. ............................................................................................. 4-347

    Procedure data error detected. ................................................................................................................ 4-348

    Procedure data error detected. ................................................................................................................ 4-349

    Procedure time exceeded 8 hours. .......................................................................................................... 4-350

    Pump 1 failed rotation test. .................................................................................................................... 4-351

    Pump 2 failed rotation test. .................................................................................................................... 4-353

    Pump 3 failed rotation test. .................................................................................................................... 4-355

    Pump 4 failed rotation test. .................................................................................................................... 4-357

    Pump 5 failed rotation test. .................................................................................................................... 4-359

    Pump did not function correctly. ........................................................................................................... 4-361

    Pump motion detected was improper. ................................................................................................... 4-363

    Pump power could not be disabled. ....................................................................................................... 4-364

    Pump power could not be enabled. ........................................................................................................ 4-365

    Pumps and centrifuge did not stop during alarm. .................................................................................. 4-366

    Pumps continued to turn during alarm. ................................................................................................. 4-367

    Pumps did not stop during alarm. .......................................................................................................... 4-368

    Pumps have been paused for 10 minutes. ............................................................................................... 4-369

    Pumps have been paused for 3 minutes. ................................................................................................. 4-370

    Pumps were commanded to turn during alarm. ..................................................................................... 4-371

    viii Spectra Optia Apheresis System Service Manual

  • Pumps were commanded to turn instead of stop. ................................................................................... 4-372

    Pumps were paused for 3 minutes. ......................................................................................................... 4-373

    Pumps were turned during alarm. .......................................................................................................... 4-375

    RBC detector calibration error occurred. ............................................................................................... 4-376

    RBC detector did not respond. .............................................................................................................. 4-378

    RBC detector did not transmit data. ...................................................................................................... 4-380

    RBC detector software error occurred. ................................................................................................... 4-382

    RBC recirculation flow path was not primed. ........................................................................................ 4-383

    RBC remove flow path was not primed. ................................................................................................ 4-385

    RBC valve malfunctioned. ..................................................................................................................... 4-387

    RBC valve malfunctioned. ..................................................................................................................... 4-388

    RBC valve sensor malfunctioned. ........................................................................................................... 4-389

    RBC valve software error occurred. ........................................................................................................ 4-390

    RBC valve was not correctly positioned. ................................................................................................ 4-391

    Read-type communication error from RBC detector occurred. .............................................................. 4-392

    Recommending custom prime. .............................................................................................................. 4-394

    Red blood cells were detected in collect line too soon. ........................................................................... 4-395

    Red blood cells were detected in plasma line from centrifuge. ................................................................ 4-398

    Remove bag reached capacity. ................................................................................................................ 4-401

    Remove valve malfunctioned. ................................................................................................................ 4-402

    Remove valve sensor malfunctioned. ...................................................................................................... 4-403

    Remove valve software error occurred. ................................................................................................... 4-404

    Remove valve was not correctly positioned. ............................................................................................ 4-405

    Replacement fluid detector detected fluid too soon. ............................................................................... 4-406

    Replacement fluid was not detected. ...................................................................................................... 4-408

    Replace pump did not stop when commanded. ..................................................................................... 4-410

    Replace pump malfunctioned. ............................................................................................................... 4-412

    Replace pump software error occurred. .................................................................................................. 4-414

    Reservoir did not contain enough fluid. ................................................................................................. 4-415

    Reservoir sensors malfunctioned. ........................................................................................................... 4-416

    Reservoir was full during air removal. .................................................................................................... 4-417

    Return line air detector detected fluid too soon. ..................................................................................... 4-419

    Return line air detector failed fluid check. .............................................................................................. 4-421

    Return line prime failed. ........................................................................................................................ 4-423

    Return line was not clamped during air removal. ................................................................................... 4-425

    Return line was not primed. ................................................................................................................... 4-427

    Return pressure sensor failed calibration test. ......................................................................................... 4-429

    Return pressure sensor failed post-load test. ........................................................................................... 4-431

    Return pressure sensor failed zero differential test. ................................................................................. 4-433

    Return pressure sensor failed zero limit test. ........................................................................................... 4-435

    Spectra Optia Apheresis System Service Manual ix

  • Return pressure sensor failed zero range test. .......................................................................................... 4-437

    Return pressure was too high. ................................................................................................................ 4-439

    Return pressure was too high when system was diverting saline. ............................................................. 4-441

    Return pump did not stop. .................................................................................................................... 4-442

    Return pump did not stop when commanded. ....................................................................................... 4-443

    Return pump malfunctioned. ................................................................................................................ 4-445

    Return pump malfunctioned. ................................................................................................................ 4-447

    Return pump software error occurred. ................................................................................................... 4-449

    Return pump speed was too high. .......................................................................................................... 4-450

    Return pump turned too fast. ................................................................................................................ 4-452

    Return saline line failed integrity test. .................................................................................................... 4-454

    Return saline line was not closed during air removal. ............................................................................. 4-456

    Return saline line was obstructed. .......................................................................................................... 4-458

    Return saline line was obstructed during air removal. ............................................................................ 4-460

    Run target will be attained before next collection phase. ........................................................................ 4-462

    Safety system could not confirm position of collect valve. ...................................................................... 4-463

    Safety system could not confirm valve position. ..................................................................................... 4-464

    Safety system could not read backup configuration file. ......................................................................... 4-466

    Safety system could not read default configuration file. .......................................................................... 4-467

    Safety system could not read primary configuration file. ........................................................................ 4-468

    Safety system detected air in return line. ................................................................................................ 4-469

    Safety system detected air in return line. ................................................................................................ 4-470

    Safety system detected excess AC infusion. ............................................................................................. 4-471

    Safety system detected internal timing inconsistency. ............................................................................. 4-473

    Safety system detected software communication error. ........................................................................... 4-475

    Safety system detected that patient hematocrit was lower than minimum. ............................................. 4-477

    Safety system detected too much fluid in collection bag. ........................................................................ 4-478

    Safety system detected too much fluid in plasma bag. ............................................................................ 4-479

    Safety system detected too much fluid in remove bag. ............................................................................ 4-480

    Safety system detected valve open longer than expected. ........................................................................ 4-481

    Safety system did not respond to power request. .................................................................................... 4-483

    Safety system did not respond to request to turn off power. ................................................................... 4-484

    Safety system did not respond to request to turn on power. ................................................................... 4-485

    Safety system failed power test. .............................................................................................................. 4-486

    Safety system failed start process. ........................................................................................................... 4-488

    Safety system lost internal communication capability. ............................................................................ 4-490

    Safety system stopped air removal. ......................................................................................................... 4-492

    Saline lines were not closed. ................................................................................................................... 4-493

    Saline line was obstructed. ..................................................................................................................... 4-494

    Sixty-four volt power supply was out of range. ....................................................................................... 4-496

    x Spectra Optia Apheresis System Service Manual

  • Software consistency error occurred. ...................................................................................................... 4-497

    Software control data was not updated. .................................................................................................. 4-498

    Software CRC failed. ............................................................................................................................. 4-500

    Software memory for AIM system was not sufficient. ............................................................................. 4-501

    Software memory for control system was not sufficient. ......................................................................... 4-502

    Software memory for safety system was not sufficient. ........................................................................... 4-503

    Software module was not consistent. ...................................................................................................... 4-504

    Stop button was touched. ...................................................................................................................... 4-505

    System cannot verify that patient was disconnected. .............................................................................. 4-506

    System could not load correct software. ................................................................................................. 4-507

    System could not read backup configuration file. ................................................................................... 4-508

    System could not read default configuration file. ................................................................................... 4-509

    System could not read primary configuration file. .................................................................................. 4-510

    System did not correctly monitor AC infusion. ...................................................................................... 4-511

    System malfunctioned when it raised or lowered cassette. ...................................................................... 4-513

    System paused: Centrifuge door was not closed. .................................................................................... 4-514

    System was reset. .................................................................................................................................... 4-516

    System was reset. .................................................................................................................................... 4-518

    System was reset. .................................................................................................................................... 4-520

    TBV restricted volume available to confirm volume in reservoir. ........................................................... 4-521

    Test of negative inlet pressure failed. ...................................................................................................... 4-522

    Test of negative return pressure failed. ................................................................................................... 4-524

    Timing discrepancy in low-level reservoir sensor occurred. .................................................................... 4-526

    Too much AC was detected during prime. ............................................................................................. 4-528

    Tubing set did not match selected procedure. ........................................................................................ 4-529

    Tubing set failed dwell test for negative inlet pressure. ........................................................................... 4-531

    Tubing set failed dwell test for negative return pressure. ........................................................................ 4-533

    Tubing set failed dwell test for positive inlet pressure. ............................................................................ 4-535

    Tubing set failed dwell test for positive return pressure. ......................................................................... 4-537

    Tubing set failed test for negative inlet pressure. .................................................................................... 4-539

    Tubing set failed test for positive inlet pressure. ..................................................................................... 4-541

    Tubing set failed test for positive inlet pressure. ..................................................................................... 4-543

    Tubing set failed test for positive return pressure. .................................................................................. 4-545

    Tubing set failed test of negative return pressure. ................................................................................... 4-547

    Tubing set identification error occurred. ................................................................................................ 4-549

    Tubing set may have contained fluid. .................................................................................................... 4-552

    Twelve volt power supply was out of range. ........................................................................................... 4-554

    Twenty-four volt power supply was out of range. ................................................................................... 4-555

    Unclamp return line. ............................................................................................................................. 4-556

    Unclamp return line. ............................................................................................................................. 4-557

    Spectra Optia Apheresis System Service Manual xi

  • Valve position error occurred. ................................................................................................................ 4-558

    Vent bag contained fluid during prime. ................................................................................................. 4-560

    Vent bag contained fluid during run. ..................................................................................................... 4-562

    Verification test that patient was disconnected from inlet line failed. ..................................................... 4-564

    Verification test that patient was disconnected from return line failed. ................................................... 4-565

    Voltage of fluid leak detector was above normal range. .......................................................................... 4-566

    Voltage of fluid leak detector was below normal range. .......................................................................... 4-567

    Volume entered for custom prime was too low. ..................................................................................... 4-569

    Volume in collection bag may be lower than reported. ........................................................................... 4-570

    Write-type communication error to data logging subsystem occurred. ................................................... 4-572

    Write-type communication error to RBC detector occurred. ................................................................. 4-573

    5: Maintenance and CalibrationMaintenance and Calibration ............................................................................................................................. 5-2

    Removing the Side Panels .......................................................................................................................... 5-2

    Raising the Pump Panel ............................................................................................................................. 5-5

    Removing the Linear Actuator ................................................................................................................... 5-8

    Calibrating the Centrifuge Motor Controller ........................................................................................... 5-10

    Making a Test Cassette ............................................................................................................................ 5-12

    Positioning the Linear Actuator Sensors ................................................................................................... 5-14

    Printer and FTP Connectivity .......................................................................................................................... 5-17

    Connecting a Printer ............................................................................................................................... 5-18

    Connecting a Computer .......................................................................................................................... 5-20

    6: Spectra Optia Spare PartsLiterature Group ................................................................................................................................................ 6-2

    Documentation Group ....................................................................................................................................... 6-2

    Display Head Group .......................................................................................................................................... 6-3

    Pump Panel Group ............................................................................................................................................. 6-4

    Centrifuge Basin Group ..................................................................................................................................... 6-6

    E-box Group ...................................................................................................................................................... 6-7

    Seal Safe Group .................................................................................................................................................. 6-9

    Exterior Parts Group ........................................................................................................................................ 6-10

    Tools Group ..................................................................................................................................................... 6-11

    Other Parts Group ........................................................................................................................................... 6-11

    Software Group ................................................................................................................................................ 6-12

    Software Multipack Group ............................................................................................................................... 6-14

    xii Spectra Optia Apheresis System Service Manual

  • 7: SpecificationsSpecifications ...................................................................................................................................................... 7-2

    Glossary

    Index

    Spectra Optia Apheresis System Service Manual xiii

  • xiv Spectra Optia Apheresis System Service Manual

  • PrefaceThe Spectra Optia Therapeutic Apheresis System Service Manual provides the information neededto service and troubleshoot the system.

    Spectra Optia Apheresis System Service Manual 15

  • Who Should Read This ManualThis manual is intended for CaridianBCT service technicians and employees, trained and qualifiedcustomer technical staff, and CaridianBCT service partners.

    How to Use This ManualThis manual is divided into sections that can be read and used separately.

    Title Description

    OperationalDescription

    Describes the operating principles of the machine. Includes information about systemcomponents, operator controls, disposables, and procedures.

    System Description Describes the functional and electronic principles of the machine

    Software Description Describes the software that the machine uses to operate.

    Troubleshooting Describes non-alarm and alarm troubleshooting. Provides alarm text and alarminformation.

    Maintenance andCalibration

    Describes maintenance and service procedures for the machine.

    Spare Parts Describes the spare parts available for the machine.

    Preface

    16 Spectra Optia Apheresis System Service Manual

  • 1Operational Description

    Spectra Optia Apheresis System Service Manual 1-1

  • Operational DescriptionThe Spectra Optia Apheresis System is a transportable, automatic blood component separator thatuses centrifugation and optical detection (automated interface management or AIM) to performexchange and collection procedures.

    The apheresis system operator uses the Spectra Optia system’s touch screen to communicate with thesystem. The operator follows the instructions on the screens to enter patient and procedure data,load and prime the disposable tubing set, and perform and troubleshoot procedures. During theprocedure, the system displays detailed procedural information, enabling the operator to adjustvalues to achieve a specific outcome or troubleshoot a condition. After the procedure, the systemprovides a summary of procedure data for the patient record. The operator may copy the data fromthe screen, or print the data if the system is connected to a printer or a computer.

    Operational Description

    1-2 Spectra Optia Apheresis System Service Manual

  • System ComponentsThe Spectra Optia system components include the video display, front panel, centrifuge chamber,and electronics compartment.

    Figure 1-1: Spectra Optia system, front and back views

    Table 1-1: Spectra Optia System Components

    Component Function

    1 Monitor with touch screen Allows you to communicate with the system.

    2 Dual IV pole Contains hooks for hanging bags and containers. Adjusts up anddown for transport.

    3 Front panel Holds the pumps, valves, sensors and detectors.

    4 Seal Safe system Seals the lines of the tubing set.

    5 Cassette tray Holds the tubing set cassette in place.

    System Components

    Spectra Optia Apheresis System Service Manual 1-3

  • Table 1-1: Spectra Optia System Components (continued)

    Component Function

    6 Power switch Allows you to turn power to the system on and off.

    7 View port Allows you to look through the centrifuge door to see the interfacein the channel.

    8 Centrifuge door Allows access to the centrifuge chamber.

    9 Wheel pedal Allows you to adjust the direction of the wheels for moving thesystem, or lock the wheels in place so that the system does not roll.

    10 Wheels Used to transport the system.

    11 Handle wheels (2) Facilitates transport of the system in a horizontal position.

    12 System serial number Unique number that identifies the system.

    13 IV pole release button Allows you to lower the IV pole by pressing the button.

    14 Power cord Connects the system to a power source.

    15 Power cord holder Secures the power cord during transport.

    16 Ethernet connection Allows you to connect the system to a printer or a computer to printa procedure report.

    17 Blood warmer bracket (notshown)

    Holds the blood warmer, if used.

    Operational Description

    1-4 Spectra Optia Apheresis System Service Manual

  • Modes of Operation and Procedure StatesThe Spectra Optia system has three modes of operation: Automatic mode, Semi-Automatic andService mode. In Automatic mode, the AIM system controls the interface. While in Semi-Automaticmode, AIM system control of the interface is disabled. There are nine procedure states that performmany individual steps to complete a procedure.

    Automatic ModeDuring Automatic mode, the system uses configured parameters and entered data to set and controlthe pump flow rates, centrifuge speed, packing factor, run targets, and establish the interface. Foroptimal separation in the channel, the target packing factor is 20. The system then uses feedback itreceives from the AIM system to control the interface, so it can collect a specific blood componentand return the rest to the patient.

    Semi-Automatic ModeDuring Semi-Automatic mode, the operator monitors the interface and adjusts it, as needed. Thesystem displays the interface on the screen, but does not control it. When certain alarm conditionsoccur, you may be required to disable the AIM system before you continue the procedure in Semi-Automatic mode. Under those conditions, the interface position does not appear on the screen andyou must use the viewport to see the interface.

    Service ModeService mode allows technicians to connect to the Spectra Optia system using service software fortroubleshooting and maintenance.

    Procedure StatesThe procedure states listed in the following table describe system activity from the perspective of theoperator. For specific information on the software states, see the Software Description section of thismanual.

    Table 1-2: Procedure States

    State Description

    Load Operator selects the procedure and loads theappropriate tubing set.

    Tubing set test System verifies that the correct tubing set is loadedproperly, and confirms the function of the pumps,valves, and sensors.

    Prime System primes the tubing set to remove air from theset, and performs volumetric tests.

    Modes of Operation and Procedure States

    Spectra Optia Apheresis System Service Manual 1-5

  • Table 1-2: Procedure States (continued)

    State Description

    Run Patient is connected, and the system performsapheresis, according to the selected procedure.

    Rinseback System returns the blood remaining in the tubing set tothe patient after the run is complete to reduce thepatient’s RBC loss. Operator may choose to perform orskip rinseback.

    Disconnect test System verifies that the lines to the patient are clampedbefore it raises the cassette.

    Unload System raises the cassette and unloads the pumps.

    Procedure summary System displays the final procedure values.

    Next procedure System prepares for the next procedure.

    Operational Description

    1-6 Spectra Optia Apheresis System Service Manual

  • User InterfaceThe Spectra Optia system’s user interface allows you to perform, adjust, and monitor procedures.

    Navigating the Screen

    Figure 1-2: Example screen

    Table 1-3: User Interface elements

    Section Description

    1 Menu bar Contains menu buttons that you touch to show or hide menu tabs.

    2 Menu tabs Appear according to the menu button that you touch. Tabs provideaccess to screens that allow you to view information or modify theprocedure.

    User Interface

    Spectra Optia Apheresis System Service Manual 1-7

  • Table 1-3: User Interface elements (continued)

    Section Description

    3 Message bar Displays information about what is currently occurring, andoccasionally displays instructions.

    4 Activity area Displays tasks to complete, buttons that allow you to

    • Navigate to another screen

    • Change procedure conditions

    • Change values or information about the procedure.

    5 Action bar Contains buttons that you touch to perform the action stated on thebutton, or to navigate to a different screen.

    Screen ColorsThe Spectra Optia system uses a color scheme to indicate the different states and conditions of theprocedure, allowing you to quickly determine the progress of the procedure.

    Table 1-4: Corresponding color schemes for operating states and conditions

    State or Condition Color

    Load, prepare procedure, tubing set test, prime (patientnot connected)

    Run, rinseback, disconnect test (patient connected)

    Unload, procedure summary, next procedure (patientdisconnected)

    Caution

    Screen ButtonsTable 1-5: Active and Inactive Buttons

    Active button Active buttons are colored andappear raised. Touch an activebutton to enter, change, or savedata, or to proceed to the nextactivity or action.

    Inactive button Inactive buttons are gray. Theyrepresent an action or informationthat is not available.

    Operational Description

    1-8 Spectra Optia Apheresis System Service Manual

  • The buttons and icons described in the following table appear on the monitor or the screens, andeither allow you to operate and adjust the Spectra Optia system during an apheresis procedure, orcommunicate important information.

    Table 1-6: Button and Icons

    Graphic Description Location Function

    Go back button Bottom right side of thescreen

    Returns you to theprevious screen. Availableonly when it is safe toreturn to the previousscreen. You must confirmany data you changedbefore touching the goback button or the datawill not be saved.

    Mute button Bottom left side of thescreen

    Allows you to mute thesound during an alarmcondition. The sound isrestored if another alarmoccurs.

    Pause button Bottom right corner of themonitor

    Pauses the pumps.

    Stop button Bottom left corner of themonitor

    Stops the centrifuge andthe pumps.

    Active alarm button Bottom left side of thescreen Indicates an activealarm.

    Takes you to the activealarm screen, or to a listof active alarms if morethan one has occurred.

    Clock button Run and the exchangestatus screens

    Allows you to view clocktime or elapsed time forthe procedure.

    Up/down arrow button Main run screen Allows you to adjust theposition of the interfaceduring Semi- Automaticmode.

    User Interface

    Spectra Optia Apheresis System Service Manual 1-9

  • Table 1-6: Button and Icons (continued)

    Graphic Description Location Function

    Left/right arrow button Strobe screen Adjusts the timing of thestrobe. Allow you to viewdifferent areas of thechannel.

    Up/down scroll button Active alarm, alarmhistory, and reportscreens

    Allows you to viewadditional information onthe screen.

    Left/right scroll button Alarm action screens Allows you to scrollthrough the differentalarm action screens, soyou do not have to returnto the main alarm screento troubleshoot an alarm.

    Check mark icon Buttons on the prepareprocedure screens

    Indicates completion of arequired task, or confirmsa selection.

    Do not connect patienticon

    Bottom right side of thescreens

    Safety feature thatreminds you that thepatient should not beconnected during thecurrent operating state.

    Operator override icon Total Blood Volumebutton on the data entryscreen

    Indicates that the operatorentered the patient’s totalblood volume whennecessary, or instead ofusing the volume that thesystem calculated.

    Semi-Automatic modeicon

    Bottom right side of therun screens

    Indicates that the systemis operating in Semi-Automatic mode.

    Operational Description

    1-10 Spectra Optia Apheresis System Service Manual

  • Exchange Disposable Tubing SetThe Exchange disposable tubing set is used for therapeutic exchange procedures.

    Table 1-7: Exchange Set Catalog and Set ID Numbers

    Catalog Number Set ID

    10200 • 0020 (4 L remove bag)

    • 0120 (6 L remove bag)

    10220 0122

    Note: The 10200 Exchange set can only be used with TPE procedures. The 10220 Exchange set can be usedwith both TPE and RBCX procedures.

    Figure 1-3: The Exchange set (10200)

    Exchange Disposable Tubing Set

    Spectra Optia Apheresis System Service Manual 1-11

  • Figure 1-4: The Exchange set (10220)

    Table 1-8: Exchange Set Components

    Component Function

    1 Inlet line Carries anticoagulated blood from the patient to the channel. Theinlet luer (male) connects to the patient inlet site. The inlet lineclamp (red) is used to clamp the inlet line.

    2 Inlet line manifold Consists of an injection site and connections for the inlet line,anticoagulant line, and inlet saline line.

    3 Inlet saline clamp (red) Used to close the inlet saline line or control the flow of saline to thepatient.

    4 Flow restrictor Allows AC to be pumped through the tubing but prevents the freeflow of AC.

    5 Return line Carries blood components from the reservoir to the patient. Theluer (male) connects to the patient return access.

    Operational Description

    1-12 Spectra Optia Apheresis System Service Manual

  • Table 1-8: Exchange Set Components (continued)

    Component Function

    6 Return line manifold Consists of an injection site and connections for the return line andreturn saline line.

    7 Return saline clamp (blue) Used to close the inlet saline line or control the flow of saline to thepatient.

    8 Anticoagulant (AC) line (orange) Carries AC to the inlet manifold. The spike (orange) is used to spikethe AC container.

    9 Saline line (green) Line that carries saline to the inlet manifold and return manifold fordelivery to the patient. The spike (green) is used to spike the salinecontainer.

    10 Inlet line trap Prevents large, solid cellular matter from entering the channel.

    11 Inlet pressure sensor Diaphragm that attaches to the inlet pressure sensor on the frontpanel to monitor the pressure to the patient.

    12 Centrifuge pressure sensor Diaphragm that attaches to the centrifuge pressure sensor on thefront panel to monitor centrifuge pressure.

    13 Reservoir Contains the fluid that is returned to the patient. Includes a 200-micron filter.

    14 Return pressure sensor Diaphragm that attaches to the return pressure sensor on the frontpanel to monitor return pressure to the patient.

    15 Centrifuge loop Four-lumen tubing that carries fluids in and out of the channel. Thesleeves reinforce the tubing at flex points. The collars fix the twoends of the loop to the centrifuge. The bearings are the contactpoints between the centrifuge arm and the loop.

    16 Channel Used during centrifugation to separate cellular components fromplasma.

    17 Connector Holds the lines where the separated components exit the channel.

    18 Vent bag Holds air displaced from the system.

    19 Remove bag Holds removed blood components.

    Exchange Disposable Tubing Set

    Spectra Optia Apheresis System Service Manual 1-13

  • Table 1-8: Exchange Set Components (continued)

    Component Function

    20 Remove line Carries the removed blood components to the remove bag.

    21 Extra remove line Provides a connection (female luer) for an extra remove bag. Theclamp (yellow) is used to clamp the extra remove line.

    22 Replace line Carries the replacement fluid from the containers to the reservoir.The spikes (white) are used to spike the replacement fluidcontainer. The clamps (white) are used to clamp the replacementline. The luer (male/female) allows connection of a filter or bloodwarmer tubing.

    Operational Description

    1-14 Spectra Optia Apheresis System Service Manual

  • Collection Disposable Tubing Set

    Figure 1-5: The Collection set

    Catalog number: 10110. Set ID number: 0011.

    Table 1-9: Collection Set Components

    Component Function

    1 Inlet line Inlet line: Used to carry anticoagulated blood from the patient to thechannel. Also includes:

    • Luer connector (male): Connects the inlet line to the needle line.

    • Inlet line clamp (red): Used to clamp the inlet line.

    • Clamp (white): Used to clamp the needle line.

    2 Diversion bag Used to capture a skin plug after performing a peripheralvenipuncture, or to collect a blood sample. Also includes:

    • Clamp (yellow): Used to clamp the line to the diversion bag.

    Collection Disposable Tubing Set

    Spectra Optia Apheresis System Service Manual 1-15

  • Table 1-9: Collection Set Components (continued)

    Component Function

    3 Inlet line manifold Provides a connection point for the inlet line, the AC line, and theinlet saline line. Includes an injection site.

    4 Inlet saline line clamp (red) Used to close the inlet saline line or to control the flow of saline tothe patient.

    5 AC check valve Allows the system to pump AC through the tubing, but prevents thefree flow of AC.

    6 Return line Used to carry fluid from the reservoir to the patient. Also includes:

    • Luer connector (male): Connects the return line to the patientreturn access.

    • Return line clamp (blue): Used to clamp the return line.

    7 Return line manifold Provides a connection point for the return line and the return salineline. Includes an injection site.

    8 Return saline line clamp (blue) Used to close the return saline line or to control the flow of saline tothe patient.

    9 Anticoagulant (AC) line (orange) Used to carry AC from the AC container to the inlet line manifold.Also includes:

    • Spike (orange): Used to spike the AC container.

    • Sterile barrier filter (0.2 microns): Prevents bacteria fromentering the system to maintain a functionally closed system.

    10 Saline line (green) Used to carry saline from the saline container to the inlet linemanifold and the return line manifold. Also includes:

    • Spike (green): Used to spike the saline container.

    • Sterile barrier filter (0.2 microns): Prevents bacteria fromentering the system to maintain a functionally closed system.

    11 Inlet line trap Prevents larger solid cellular matter from entering the channel.

    12 Inlet pressure sensordiaphragm

    Attaches to the inlet pressure sensor on the front panel to monitorpressure in the inlet line.

    13 Centrifuge pressure sensordiaphragm

    Attaches to the centrifuge pressure sensor on the front panel tomonitor pressure in the