hyperion colorimeter specification

specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

hyperion colorimeter

specification

Page 2: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

2

Contents 1 Hyperion .............................................................................................. 3

2 Highlights ............................................................................................. 3

3 General specifications ......................................................................... 4

4 Typical spectral sensitivity ................................................................... 4

5 10mm measurement specifications .................................................... 5

6 5mm improved measurement specifications ....................................... 6

7 10mm improved measurement specifications ..................................... 7

8 10mm WA measurement specifications .............................................. 8

9 20mm measurement specifications .................................................... 9

10 20mm improved measurement specifications ............................... 10

11 10mm measurement data ............................................................. 11

12 10mm improved measurement data ............................................. 13

13 20mm measurement data ............................................................. 15

14 20mm improved & 10mm WA measurement data ........................ 17

15 Xtreme low luminance performance data ..................................... 19

15.1 Hyperion 10mm ......................................................................... 19

15.2 Hyperion 10mm improved ......................................................... 19

16 Hyperion 10mm dimensions .......................................................... 20

17 Hyperion 10mm & 5mm improved dimensions ............................. 21

18 Hyperion 10mm WA dimensions ................................................... 22

19 Hyperion 20mm dimensions .......................................................... 23

20 Hyperion 20mm improved dimensions .......................................... 24

21 Hyperion M8 fibre dimensions ....................................................... 25

Page 3: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

3

1 Hyperion The Hyperion colorimeter offers a unique combination of high speed and

accurate colour measurement capabilities packed in a robust package. In

comparison to its predecessors, the Hyperion has a significant improvement

on filter characteristics and an incredible speed upgrade. It is actually four

times faster than previous models making accurate colour measurements

possible at 0.3cd/m2 in only 50ms.

The Hyperion colorimeter is available with a 10mm spot size. A fiber version

is also available with several optics, custom optics can be applied on

request. Added to the filter characteristics the high sensitivity, ultra-low

noise electronics and a huge dynamic range make it the ideal device for

display measurements even at low levels.

2 Highlights Highly accurate colour measurement according to human eye

(CIE1931)

Fast colour measurement even at low luminance level

Flicker luminance (Y) function: 2000 samples/second

Auto-range function

Powerful MCU enables internal JEITA flicker calculation

Mechanical shutter

USBMTC standard compliant

Windows, Linux and MAC OSX compatible

Directly supported in Labview, LabWindows, Visual Studio via

VISA library

Page 4: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

4

3 General specifications Interfaces

USB 2.0 USBMTC compliant, SCPI command set, high speed device

RS 232 For PC and embedded purposes, using same command set as USB

Trigger in & out 5V compliant

Power ratings

Min voltage

Typical voltage

Max voltage

Max current

USB power 4.75V 5.00V 5.25V 300mA

9V power 8.75V 9.00V 9.25V 300mA

Measurement system

Photo detector 3 silicon photo diode using XYZ interference filter

Spectral response Approximates CIE 1931 colour matching functions

Measurement parameters

XYZ, Yxy, Yuv, correlated colour temperature (CCT), dominant wavelength DWL, Flicker, Response time

Size (HxWxD) 53x70x147 mm (without lens system)

Weight 670 gram

Mounting 12 M3 threat holes spread over four sides

4 Typical spectral sensitivity

Page 5: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

5

5 10mm measurement specifications

Measurement system

Optical system Acceptance angle 5º (+/- 2.6)

Measurement spot size 14 mm at 50 mm distance 19 mm at 100 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.005 cd/m² - 20,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.5% at 0.1 cd/m² ¹ Y +/- 0.2% at 1 cd/m² ¹ Y +/- 0.15% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.001 for Y at 0.1 cd/m² ¹ x,y +/- 0.0002 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 6: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

6

6 5mm improved measurement specifications

Measurement system

Optical system Acceptance angle 5º (+/- 2.5)

Measurement spot size 9.6 mm at 50 mm distance 14 mm at 100mm distance 6.8 mm at 20 mm distance 5.2 mm at 1 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.012 cd/m² - 48,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.3% at 0.4 cd/m² ¹ Y +/- 0.15% at 4 cd/m² ¹ Y +/- 0.1% at 20 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.00075 for Y at 0.4 cd/m² ¹ x,y +/- 0.0003 for Y at 4 cd/m² ¹ x,y +/- 0.0003 for Y at 20 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.4 cd/m² ¹ 10-20 samples / s for Y at 4 cd/m² ¹ 40 samples / s for Y at 20 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 15 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 15 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 7: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

7

7 10mm improved measurement specifications

Measurement system

Optical system Acceptance angle 5º (+/- 2.5)

Measurement spot size 14.4 mm at 50 mm distance 18.7 mm at 100 mm distance 10.2 mm at 1 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.003 cd/m² - 12,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.3% at 0.1 cd/m² ¹ Y +/- 0.15% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.00075 for Y at 0.1 cd/m² ¹ x,y +/- 0.0003 for Y at 1 cd/m² ¹ x,y +/- 0.0003 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 8: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

8

8 10mm WA measurement specifications

Measurement system

Optical system Acceptance angle 10º (+/- 5)

Measurement spot size 17.4 mm at 50 mm distance 26.1 mm at 100 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.0004 cd/m² - 4,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.0005 for Y at 0.1 cd/m² ¹ x,y +/- 0.00025 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 4 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 4 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 9: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

9

9 20mm measurement specifications

Measurement system

Optical system Acceptance angle 3.5º (+/- 1.75)

Measurement spot size 25 mm at 50 mm distance 28 mm at 100 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.0005 cd/m² - 5,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.0005 for Y at 0.1 cd/m² ¹ x,y +/- 0.00025 for Y at 1 cd/m² ¹ x,y +/- 0.0002 for Y at 5 cd/m² ¹ x,y +/- 0.0002 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 5 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 5 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 10: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

10

10 20mm improved measurement specifications

Measurement system

Optical system Acceptance angle 5º (+/- 2.5)

Measurement spot size 22.1 mm at 50 mm distance 26.7 mm at 100 mm distance

Flicker measurement speed (sample mode)

Luminance 2000 samples / second, XYZ 2000 samples / second Correct detected frequency of at least 1 Hz

Colour measurement speed Colour 22 ms or higher, depending on luminance level (including communication) 150 cd/m² with DC level light at 16 ms. PWM requires longer integration (multiple frames)

Sample mode signal frequency response

Parameter F3dba

Gain 1 DC – 500 Hz

Gain 2 DC – 500 Hz

Gain 3 DC – 500 Hz

Colorimeter specification

Parameter Range Accuracy Repeatability

Resolution 16 bit for X, Y and Z >80 dB without averaging

Luminance (Y) 0.0004 cd/m² - 4,000 cd/m² integration time between 0.5ms – 1s

+/-2% of measured value, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

Y +/- 0.1% at 0.1 cd/m² ¹ Y +/- 0.1% at 1 cd/m² ¹ Y +/- 0.1% at 5 cd/m² ¹ Y +/- 0.1% at 150 cd/m² ¹

Chromaticity (x,y) +/- 0.001 after calibration, measured at white image of LCD display, Luminance of app. 150 cd/m², x=0.300 y= 0.325

x,y +/- 0.0004 for Y at 0.1 cd/m² ¹ x,y +/- 0.0002 for Y at 1 cd/m² ¹ x,y +/- 0.00016 for Y at 5 cd/m² ¹ x,y +/- 0.00016 for Y at 150 cd/m² ¹

Measurement speed 4-10 samples / s for Y at 0.1 cd/m² ¹ 10-20 samples / s for Y at 1 cd/m² ¹ 40 samples / s for Y at 5 cd/m² ¹ 40 samples / s for Y at 150 cd/m² ¹

Flicker (contrast method) 4 cd/m² or higher +/- 0.3% flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2%

Flicker (JEITA method) 4 cd/m² or higher +/- 0.3dB flicker frequency 30Hz AC/DC 10% sine wave. Sine wave at 10 cd/m²

+/- 0.2dB

Operating temperature 10-35°C ²

Shutter lifetime >1000000

Shutter speed 250-300 ms depending on temperature and lifetime

1 All measurements are performed 20 times on a LED display with sufficient signal noise ratio, value is based on 2 sigma. Sample speed depends on the measured sample.

If the sample uses PWM, it will take longer. Use of lower rated values is strongly recommended to ensure repeatability.

2 Operating temperature reaches from 0 to 40 degrees. Dark level compensation is optimized for operating in temperatures between 10 and 35 degrees.

Other temperature ranges can be calibrated using the mechanical shutter if necessary.

Page 11: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

11

11 10mm measurement data

Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²) to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

0.005 20000 0.005 20000 0.005 20000

Accuracy (for white)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

Repeatability (2 sigma)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-4% +/-1% +/-1.2% +/-0.6% +/-0.3% +/-0.3% +/-0.2% +/-0.3%

+/-0.0005 +/-0.001 +/-0.0036 +/-0.006 +/-0.015 +/-0.045 +/-0.10 +/-0.30

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-3% +/-0.6% +/-0.5% +/-0.5% +/-0.3% +/-0.3% +/-0.12% +/-0.08%

+/-0.0003 +/-0.0006 +/-0.0015 +/-0.005 +/-0.015 +/-0.045 +/-0.06 +/-0.08

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 20000

+/-1% +/-0.5% +/-0.2% +/-0.2% +/-0.15% +/-0.13% +/-0.07% +/-0.07%

+/-0.0001 +/-0.0005 +/-0.0006 +/-0.002 +/-0.0075 +/-0.0195 +/-0.035 +/-0.07

Page 12: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

12

Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level

0.1 20000 0.1 20000 0.1 20000

Accuracy (for white)

0.1 0.3 1 5 15 50 100

- - - - - - 20000

+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 20000

+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 20000

+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

Repeatability (2 sigma)

0.1 0.3 1 5 15 50 100

- - - - - - 20000

0.006 0.0026 0.00056 0.00066 0.00035 0.00017 0.00014

0.1 0.3 1 5 15 50 100

- - - - - - 20000

0.0037 0.0014 0.00061 0.00067 0.00028 0.00012 0.00012

0.1 0.3 1 5 15 50 100

- - - - - - 20000

0.001 0.00022 0.00033 0.00066 0.00033 0.00022 0.0001

Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec

xyLv 0.1 0.3 1 5 15 50 100

- - - - - - 20000

8 8 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 20000

3.7 3.7 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 20000

0.6 – 0.9 0.6 – 0.9 10 40 40 40 40

Page 13: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

13

12 10mm improved measurement data

Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

0.003 12000 0.003 12000 0.003 12000

Accuracy (for white)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

Repeatability (2 sigma)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-3% +/-0.9% +/-0.9% +/-0.6% +/-0.3% +/-0.3% +/-0.2% +/-0.3%

+/-0.0003 +/-0.0009 +/-0.0027 +/-0.006 +/-0.015 +/-0.045 +/-0.10 +/-0.30

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-2% +/-0.5% +/-0.5% +/-0.4% +/-0.3% +/-0.3% +/-0.12% +/-0.08%

+/-0.0002 +/-0.0005 +/-0.0015 +/-0.004 +/-0.015 +/-0.045 +/-0.06 +/-0.08

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 12000

+/-0.6% +/-0.4% +/-0.2% +/-0.15% +/-0.13% +/-0.11% +/-0.06% +/-0.06%

+/-0.0001 +/-0.0004 +/-0.0006 +/-0.0015 +/-0.0065 +/-0.062 +/-0.030 +/-0.06

Page 14: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

14

Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level

0.003 12000 0.003 12000 0.003 12000

Accuracy (for white)

0.1 0.3 1 5 15 50 100

- - - - - - 12000

+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 12000

+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 12000

+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

Repeatability (2 sigma)

0.1 0.3 1 5 15 50 100

- - - - - - 12000

0.004 0.0020 0.00041 0.00056 0.00032 0.00017 0.00014

0.1 0.3 1 5 15 50 100

- - - - - - 12000

0.0025 0.0009 0.00043 0.00040 0.00028 0.00012 0.00012

0.1 0.3 1 5 15 50 100

- - - - - - 12000

0.00075 0.00022 0.00033 0.00066 0.00033 0.00022 0.0001

Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec

xyLv 0.1 0.3 1 5 15 50 100

- - - - - - 12000

8 8 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 12000

3.7 3.7 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 12000

1.25 1.5 15 40 40 40 40

Page 15: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

15

13 20mm measurement data Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²) to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

0.0005 5000 0.0005 5000 0.0005 5000

Accuracy (for white)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

Repeatability (2 sigma)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-2% +/-0.45% +/-0.45% +/-0.3% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-1% +/-0.2% +/-0.2% +/-0.15% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-0.3% +/-0.2% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

Page 16: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

16

Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level

0.0005 5000 0.0005 5000 0.0005 5000

Accuracy (for white)

0.1 0.3 1 5 15 50 100

- - - - - - 5000

+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 20000

+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 5000

+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

Repeatability (2 sigma)

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.001 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec

xyLv 0.1 0.3 1 5 15 50 100

- - - - - - 5000

8 8 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 5000

3.7 3.7 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 5000

3.7 10 40 40 40 40 40

Page 17: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

17

14 20mm improved & 10mm WA measurement

data

Luminance Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²) to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²) from (cd/m²)

to (cd/m²)

level (%)

level (cd/m²)

0.0004 4000 0.0004 4000 0.0004 4000

Accuracy (for white)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-5% +/-5% +/-4% +/-3% +/-3% +/-2% +/-2% +/-2%

- - - - - - - -

Repeatability (2 sigma)

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-1.5% +/-0.45% +/-0.45% +/-0.3% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-0.6% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

0.01 0.1 0.3 1 5 15 50 100

- - - - - - - 5000

+/-0.2% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1% +/-0.1%

Page 18: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

18

Chromaticity Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

Measurement range from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level from (cd/m²)

to (cd/m²)

level

0.0004 4000 0.0004 4000 0.0004 4000

Accuracy (for white)

0.1 0.3 1 5 15 50 100

- - - - - - 5000

+/-0.005 +/-0.0025 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 20000

+/-0.005 +/-0.0025 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

0.1 0.3 1 5 15 50 100

- - - - - - 5000

+/-0.003 +/-0.002 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015 +/-0.0015

Repeatability (2 sigma)

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.001 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

0.1 0.3 1 5 15 50 100

- - - - - - 5000

0.0003 0.0002 0.0002 0.0002 0.0002 0.0002 0.0001

Speed Setting 1: max integration time 100ms Setting 3: max integration time 240ms Setting 2: max integration time 1000ms

from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec from (cd/m²)

to (cd/m²)

times/sec

xyLv 0.1 0.3 1 5 15 50 100

- - - - - - 5000

8 8 10 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 5000

3.7 10 40 40 40 40 40

0.1 0.3 1 5 15 50 100

- - - - - - 5000

3.7 10 40 40 40 40 40

Page 19: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

19

15 Xtreme low luminance performance data In below sub-chapters the repeatability data for extreme low luminance

can be found. This data is non-binding, so no claims can be made based

on this data, this data is purely for indication. When measuring assumed is

that dark is kept under control

15.1 Hyperion 10mm

15.2 Hyperion 10mm improved

Chromaticity Setting 3: max integration time 1000ms

Measurement range at (cd/m²) level

Accuracy (for white)

0.005 0.0015

- -

+/-0.02 +/-0.07

Repeatability (2 sigma)

0.005 0.0015

- -

0.012 0.050

Luminance Setting 3: max integration time 1000ms

Measurement level at (cd/m²) level (%)

level (cd/m²)

Accuracy (for white)

0.005 0.0015

- -

+/-6% +/-10%

- -

Repeatability (2 sigma)

0.005 0.0015

- -

+/-2% +/-5%

- -

Luminance Setting 3: max integration time 1000ms

Measurement level at (cd/m²) level (%)

level (cd/m²)

Accuracy (for white)

0.005 0.0015

- -

+/-8% +/-15%

- -

Repeatability (2 sigma)

0.005 0.0015

- -

+/-4% +/-11%

- -

Chromaticity Setting 3: max integration time 1000ms

Measurement range at (cd/m²) level

Accuracy (for white)

0.005 0.0015

- -

+/-0.01 +/-0.07

Repeatability (2 sigma)

0.005 0.0015

- -

0.005 0.015

Page 20: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

20

16 Hyperion 10mm dimensions

Page 21: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

21

17 Hyperion 10mm & 5mm improved dimensions

Page 22: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

22

18 Hyperion 10mm WA dimensions

Page 23: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

23

19 Hyperion 20mm dimensions

Page 24: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

24

20 Hyperion 20mm improved dimensions

Page 25: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

25

21 Hyperion M8 fibre dimensions

Page 26: specification · Gain 1 DC – 500 Hz Gain 2 DC – 500 Hz Gain 3 DC – 500 Hz Colorimeter specification Parameter Range Accuracy Repeatability Resolution 16 bit for X, Y and Z >80

26

Admesy B.V. Sleestraat 3 6014 CA Ittervoort The Netherlands T +31 (0)475 600 232 F +31 (0)475 600 316 www.admesy.com [email protected]

The material in this document is subject to change. No rights can be derived from the content of this document. All rights reserved. No part of this document may be reproduced, stored in a database or retrieval system, or published in any form or way, electronically, mechanically, by print, photo print, microfilm or any other means without prior written permission from the publisher.

Version 1.2.1 02/2019