  • Knauer Webinar 12-01-24

    Solve your problems – KNAUER (U)HPLC Troubleshooting II

  • Knauer Webinar 12-01-24

    Solve your problems – KNAUER (U)HPLC Troubleshooting II

    Dr. Silvia Marten

    Head of Applications and Columns Dep.

    Rene Borstel

    Applications and Columns Dep.

  • Knauer Webinar 12-01-24

    Troubleshooting alocation

    System or Column ?

    mobile phase

    pump problems

    injector problems

    detector problems

    columns problems

    Problems can be everywhere in your

    HPLC system!

  • Knauer Webinar 12-01-24

    Troubleshooting categories

    major categories pressure abnormalities

    leaks

    problems with chromatogram

    other problems

    Report in your HPLC system recordbook !

    injector problems

  • Knauer Webinar 12-01-24

    Troubleshooting categories

    Problem indicators

    Retention tR

    Pressure

    Baseline

    Peak shape

    broad peaks

    split peaks

    peak tailing

    Peak area

  • Knauer Webinar 12-01-24

    broad peaks Cause change in mobile phase

    flow rate to low

    leak between column

    and detector

    extra column effects:

    - column overload

    - cell volume to large

    - tubing between column

    and detector to long

    Solutions

    check make up of mobile

    phase

    adjust flow rate;

    use higher temperatures

    check lose fittings

    - inject smaller volumes

    - dilute sample

    - use smaller flow cell

    - use 0.25 or 0.1 mm ID

    tubings

    Troubleshooting: tips and tricks

  • Knauer Webinar 12-01-24

    Troubleshooting: tips and tricks

    tailing Cause buffer concentration to

    low

    guard column

    contaminated

    column contaminated

    void at column inlet

    column temperature too

    low

    Solutions

    increase concentration

    (10 mM – 50 mM)

    replace guard column

    replace or reequilibrate

    column

    replace column

    increase temperature

  • Knauer Webinar 12-01-24

    Troubleshooting: tips and tricks

    broad peaks

    mixing

    chamber

    Avoid poorly made HPLC connections !

    leakage

  • Knauer Webinar 12-01-24

    Troubleshooting: tips and tricks

    Fittings used in HPLC

    1 Fitting only for one port !

    stainless steel

    high pressure sealing

    (re-useable)

    PEEK (black) P3860

    high pressure sealing

    frequent connection

    changing

    Dynaseal

    < 400 bar

    re-useable

    fingertight

    (U)HPLC (U)HPLC 0.5 mm ODHPLC

  • Knauer Webinar 12-01-24

    Troubleshooting; tips and tricks

    split peaks

    regular sample

    problem sample

    Cause column/precolumn

    contaminated

    sample solvent

    incompatible with mobile

    phase

    void at column inlet

    partially blocked frit

    broken column bed

    Solutions

    replace column/

    precolumn

    dilute sample in mobile

    phase

    replace or repack top of

    column

    replace frit

    replace column

  • Knauer Webinar 12-01-24

    Troubleshooting: tips and tricks

    frontingCause interference in sample

    shoulder

    column overload

    incompatible sample

    solvent

    Solutions

    Check column performance

    with standard

    increase efficiency

    try another column type

    inject smaller volumes

    inject lower concentration

    adjust sample solvent

    www. chromservice.cz

  • Knauer Webinar 12-01-24

    Minutes

    0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4

    -20

    0

    20

    40

    60

    80

    100

    120

    140

    160

    180System optimization:

    minimize tubing length

    smallest inner diameter

    low flow cell volume

    poor HPLC connections

    Minutes

    0.575 0.600 0.625 0.650 0.675 0.700 0.725 0.750 0.775 0.800 0.825 0.850 0.875 0.900 0.925

    0

    20

    40

    60

    80

    100

    120

    140

    160

    Troubleshooting: tips and tricks

    extra column dispersion

    Avoid system dead volume !

  • Knauer Webinar 12-01-24Minutes

    0,2 0,4 0,6 0,8 1,0 1,2 1,4 1,6 1,8 2,0

    mA

    U

    -50

    0

    50

    100

    150

    200

    250

    300

    350

    400

    450

    500

    550

    600

    650

    700

    750

    green: UHPLCblue: HPLC plus

    1

    2

    3

    4

    5

    Troubleshooting: tips and tricks

    Peak 5: Phenetol

    tR [min]0,982 0,979

    Area

    725858 729245

    Plates

    1567 5437

    Optimization of resolution

    Avoid system dead volume !

  • Knauer Webinar 12-01-24

    HPLC Troubleshooting Guide

    available as .pdf

    www.knauer.net

  • Knauer Webinar 12-01-24

    Troubleshooting rules

    1. Failure in system or sample?

    2. Check method parameter: are instrument settings correct?

    4. Step by step changings: modify or replace only one thing!

    3. Anything be changed in method or system/column?

    How to fix the problem?

    Instrument and method logbook can help you !

  • Knauer Webinar 12-01-24

    Troubleshooting rules

    Prevention (main rules)

    3. periodical replacement of inline filter

    1. periodical replacement of pump seals

    4. filtration of mobile phase and samples

    6. avoid pH > 8 / < 2

    Realize routine preventive maintenainance !

    5. use guard/saturator columns

    2. periodical replacement of detector lamp

    7. flush buffer from the HPLC system

  • Knauer Webinar 12-01-24

    Summary

    Troubleshooting

    1. systematic troubleshooting (step by step)

    2. report troubleshooting / maintenance

    3. avoid poor HPLC connections

    4. periodical maintenance

  • Knauer Webinar 12-01-24

    1. Operators and service Manuals for instruments

    2. experienced other people in the lab

    3. Manufacturer of instruments or columns

    4. KNAUER technical hotline: [email protected]

    5. Seminars, workshops and trainings: www.knauer.net

    6. Troubleshooting LC Systems

    by J.W. Dolan and L.R. Snyder, (Humana Press, 1989)

    7. LC-GC Magazine (http://www.chromatographyonline.com/)

    Advanstar Publishing

    8. KNAUER HPLC Troubleshooting Guide www.knauer.net

    Troubleshooting

    Additional help / References

  • Knauer Webinar 12-01-24

    Troubleshooting

    Further questions and answers via mail

    [email protected] or

    [email protected]

    Questions ?