If the bulb does not stay depressed, a
to the air vent tubing as indicated (if required by a clinician).
ttach LP-suction
Expanded bulb
Attach low pressure (LP) suction
Depress the bulb. If the bulb stays depressed it
applies suction to assist fluid / air to drain.
suction
Depress the bulb. If it re-expands there is an air leak.
If it re-expands, syringe 30ml clean water into air-vent tubing.
Look for bubbling or water level movement.
Bubbling or water level movement = air leak
Air leak?
AIR VENT TUBING
Suction tubing
water
Good news
Potential reasons: - Pneumothorax
- Fistula (air leak)
- Tube connections not secure
= air leaks into the system
Reason:
(suction / negative
pressure is achieved
provided there is no
blockage in tubing)
Depressed bulb
Monitoring
X Bad news
Depress the bulbDepressed bulb
Npr
egative
re
ssu
e
1000
ml
900
100
200
300
400
500
600
700
800
150
250
350
450
550
650
750
850
950
ml
6080
40
20
100
30
10 5
MAX
50
20
H O2 level
bubbling
7654321
water levelmovement
SINAP
Iche
st d
rain
MO
DE
L N
O: X
L 1
00
0 (
00
1)
(pa
ten
t p
en
din
g)
AIR LEAK SIZE
INDICATOR
ega
ve
rr
Ntip
ess
ue
1000
ml
900
100
200
300
400
500
600
700
800
150
250
350
450
550
650
750
850
950
ml
6080
40
20
100
30
10 5
MAX
50
20
H O2 level
bubbling
7654321
water levelmovement
SINAP
Iche
st d
rain
MO
DE
L N
O: X
L 1
00
0 (
00
1)
(pa
ten
t p
en
din
g)
AIR LEAK SIZE
INDICATOR
7654321
Small
Large
AIR LEAK SIZEINDICATOR
Look for bubbles
H O level2
bubbling
water levelmovement
Do not use if unit package is opened
or damaged.
Single UseAvoid Extreme Temperatures
o50 C
o0 C
SINAPI biomedical Arc / Infruitec North CampusLelie Road, Stellenbosch, 7599, South AfricaTel: +27 21 887 5260 Fax: +27 21 887 3059email: [email protected]: Suite 214, Pbag X5061, Stellenbosch 7599
STERILE EO LATEX
Latex Free
INSTRUCTIONS FOR USE - NURSING CARESINAPI CHEST DRAIN
CAUTION: Federal law restricts this device to sale by or on the order of a physician. Doc. Rev 01
If re-used the sterilityof the device is
compromised, which willgreatly increase infection risk
Replace deviceafter 7 days
For hemothorax: Look for blood visible in the drainage tubing.
For pneumothorax: Depress the bulb - the bulb must initially re-
expand. If it does not re-expand it confirms an absence of air at the catheter tip,
thus the catheter is not placed in the pleural space (or pneumothorax is resolved).
How to confirm chest catheter position
Hang device lower than the chest. Routinely drain
fluids via the bottom outlet tap before it overflows. If
necessary, you may lie device flat on the bed, as long as the
liquid in the reservoir remains less than the maximum volume
level. Hang next to the bed when unattended.
Gravity drainageMobileShorten the chest catheter to
prevent loops in tubing. For
pneumothorax, remove the
drainage bag.
A guide when to remove the chest catheter
Evaluate the above clinically and / or with chest X-ray
Depress the bulb until the bulb stays depressed. If the bulb re-expands, pinch
the tubing above the valve and then depress the bulb. Repeat the process until the
bulb stays depressed (up to 30x). For BP fistula (air-leak), attach external suction.
Increase the drainage rate (suction)
Depressed bulb
X
depress bulb
pinchtubing
You have to confirm two things:
1. Drainage volume must be low.
2. Depress the bulb: the bulb must stay depressed.
if bulb re-expands it confirms air in the pleural space / if bulb
stays depressed it confirms the opposite.
Negative
pre
ssure
1000
ml
900
100
200
300
400
500
600
700
800
150
250
350
450
550
650
750
850
950
ml
0680
40
20
010
30
10 5
MAX
50
20
HO2 level
bubbling
7654321
water levelmovement
SINAP
Iche
st d
rain
MO
DE
L N
O: X
L 1
00
0 (
00
1)
(pa
ten
t p
en
din
g)
0
120
AIR LEAK SIZE
INDICATOR
Pneumothorax
asNe
gti v
e pr
esu
re
1000
ml
900
100
200
300
400
500
600
700
800
150
250
350
450
550
650
750
850
950
ml
6080
40
20
100
30
10 5
MAX
50
20
HO2 level
bubbling
7654321
movementwater level
SINAP
Iche
st d
rain
MO
DEL
NO
: XL
1000
(001
)
(pate
nt pe
nding
)
0
120
SIZEAIR LEAK
INDICATOR
Hemothorax
Ngat
i e p
ress
ure
ev
1000
ml
900
100
200
300
400
500
600
700
800
150
250
350
450
550
650
750
850
950
ml
6080
40
20
100
30
10 5
MAX
50
20
HO2 level
bubbling
7654321
movementwater level
SINAP
Iche
st d
rain
MO
DE
L N
O: X
L 10
00 (0
01)
(pate
nt pe
nding
)
0
120
SIZEAIR LEAK
INDICATOR
Replace device after 7 days!!
Expanded bulb
Confirms air atcatheter tip
Depressed bulb
SINAPI CHEST DRAININSTRUCTIONS FOR USE - INSERTION & REMOVAL
Confirms NO air atcatheter tip
NOTE: Device volumes are approximate.