SIMPLELOGGER®
RMSVOLTAGEMODULEL205L230L260
E N G L I S H User Manual
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.
We guarantee that at the time of shipping your instrument has met its published specifications.
An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.
The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer. For recalibration, please use our calibration services. Refer to our repair and calibration section at www.aemc.com.
Serial #: ________________________________
Catalog #: 2116.05 / 2113.93 / 2113.94
Model #: L205 / L230 / L260
Please fill in the appropriate date as indicated:
Date Received: _________________________________
Date Calibration Due: _______________________
Chauvin Arnoux®, Inc.d.b.a AEMC® Instrumentswww.aemc.com
Table of Contents
1. INTRODUCTION ............................................................................... 21.1 InternationalElectricalSymbols................................................41.2 DefinitionofMeasurementCategories.....................................51.3 ReceivingYourShipment..........................................................51.4 OrderingInformation.................................................................5
1.4.1AccessoriesandReplacementParts...............................5
2. PRODUCT FEATURES ...................................................................... 42.1 IndicatorsandButtons..............................................................42.2 InputsandOutputs...................................................................52.3 Mounting...................................................................................5
3. SPECIFICATIONS............................................................................. 63.1 ElectricalSpecifications............................................................63.2 MechanicalSpecifications........................................................73.3 EnvironmentalSpecifications....................................................73.4 SafetySpecifications................................................................7
4. OPERATION .................................................................................... 84.1 SoftwareInstallation.................................................................84.2 RecordingData.........................................................................94.3 UsingtheSoftware.................................................................10
5. MAINTENANCE ............................................................................. 115.1 BatteryInstallation..................................................................115.2 Cleaning..................................................................................11
APPENDIX A ....................................................................................... 12Importing.TXTFilesintoaSpreadsheet.........................................12
OpeningaSimpleLogger.TXTfileinExcel...........................12FormattingtheDateandTime................................................13
RepairandCalibration...........................................................................15TechnicalandSalesAssistance............................................................15LimitedWarranty...................................................................................16WarrantyRepairs...................................................................................16
� RMS Voltage Simple Logger® L205/L230/L260
CHAPTER 1
INTRODUCTION
Warning Thesesafetywarningsareprovidedtoensurethesafetyofper-sonnelandproperoperationoftheinstrument.• Readtheinstructionmanualcompletelyandfollowallsafety
informationbeforeoperatingthisinstrument.• Usecautiononanycircuit:Potentiallyhighvoltagesandcur-
rentsmaybepresentandmayposeashockhazard.• Readthespecificationssectionpriortousingthedatalogger.
Neverexceedthemaximumvoltageratingsgiven.• Safetyistheresponsibilityoftheoperator.• Formaintenance,useonlyoriginalreplacementparts.• NEVERopenthebackoftheinstrumentwhileconnectedto
anycircuitorinput.• ALWAYSconnecttheleadstotheloggerbeforeinsertingthe
leadstothetestvoltage• ALWAYS inspect the instrument and leads prior to use.
Replaceanydefectivepartsimmediately.• NEVERusetheSimpleLogger®ModelsL205,L230,L260on
electrical conductors ratedabove600V inovervoltagecat-egoryIII(CATIII).
1.1 International Electrical Symbols
This symbol signifies that the instrument is protected by double or reinforced insulation.This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol preceding instructions indicates that if the instructions are not followed, bodily injury, installation/sample and product damage may result.Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.
In conformity with WEEE 2002/96/EC
RMS Voltage Simple Logger® L205/L230/L260 5
1.2 Definition of Measurement CategoriesCat. IV: For measurements performed at the primary electrical supply
(
6 RMS Voltage Simple Logger® L205/L230/L260
CHAPTER 2
PRODUCT FEATURES
Models L205, L230 and L260:
2.1 Indicators and ButtonsTheSimpleLogger®hasonebuttonandoneindicator.Botharelocatedonthefrontpanel.ThePRESSbuttonisusedtostartandstoprecordingsandtoturntheloggeronandoff.
TheRedLEDindicatesthestatusofthelogger:
• Single Blink: STAND-BYmode• Double Blink: RECORDmode• Continuously On: OVERLOADcondition• No Blinks: OFFmode
1
3
2
(1) Start/Stop Button(2) Input Safety Plugs
(3) Red LED Indicator(4) RS-232 Interface
PRESS
4
LOGGER OPERATION:
1. Press to Start Logging(Logging = Two Blinks)
2. Press to Stop Logging(Stand-By = One Blink)
3. Press & Hold 5 sec. to Shut Off(Off = No Blinks) Warning: All Data will be Deleted!
! Continuous LED indicates OUT OF RANGE
SIMPLE LOGGER®RMS VOLTAGE MODULE L260
RS
-232IN
PU
T(6
00 V
rms m
ax)
RMS Voltage Simple Logger® L205/L230/L260 �
2.2 Inputs and OutputsTheleftsideoftheloggerincorporates4mmsafetybananajackinputcon-nectorscompatiblewiththecurrentprobesforwhichyourSimpleLogger®wasdesigned.Therightsidehasafemale9-pin“D”shellserialconnectorusedfordatatransmissionfromtheloggertoyourcomputer.
2.3 MountingYourSimpleLogger®isequippedwithclearanceholesinthebaseplatetabsformounting.Forlesspermanentmounting,theVelcro®pads(sup-plied loose)canbeattachedto the loggerandthesurfacetowhichtheloggerwillbemounted.
� RMS Voltage Simple Logger® L205/L230/L260
CHAPTER 3
SPECIFICATIONS
3.1 Electrical SpecificationsNumber of Channels: 1
Measurement Range:L205:0-25Vrms(strayvoltage)L230:0-300VrmsL260:0-600Vrms
Input: L205:0-25VrmsL230:0-300VrmsL260:0-600Vrms
Input Connection:RecessedSafetyBananaJacks
Input Impedance: L205:1MΩL230andL260:2MΩ
Resolution:8Bit
L205Scale Range Maximum Input Resolution
100% 25V 0.1V50% 12.5V 0.05V25% 6.25V 0.025V
12.5% 3.125V 0.0125V
L230Scale Range Maximum Input Resolution
100% 300V 2V50% 250V 1V25% 125V .5V
12.5% 62.5V .25V
L260Scale Range Maximum Input Resolution
100% 600V 4V50% 300V 2V25% 250V 1V
12.5% 125V 0.5V
Reference condition: 23°C ± 3K, 20 to 70% RH, Frequency 50/60Hz, No AC external mag-netic field, DC magnetic field ≤ 40A/m, battery voltage 9V ± 10%.Accuracy: 1% ± Resolution
RMS Voltage Simple Logger® L205/L230/L260 �
Sample Rate: 4096/hrmax;decreasesby50%eachtimememoryisfull
Data Storage: 8192readings
Data Storage Technique: TXR™TimeExtensionRecording™
Power:9VAlkalineNEDA1604,6LF22,6LR61
Battery Life Recording: Upto1yearofrecording@77°F(25°C)
Output: RS-232viaDB9connector(1200Baud)
3.2 Mechanical SpecificationsSize: 2-7/8x2-5/16x1-5/8"(73x59x41mm)
Weight (with battery):5oz(140g)
Mounting: BaseplatemountingholesorVelcro®pads
Case Material:PolystyreneULV0
3.3 Environmental SpecificationsOperating Temperature: -4to158°F(-20to70°C)
Storage Temperature:-4to176°F(-20to80°C)
Relative Humidity: 5to95%non-condensing
3.4 Safety Specifications
Working Voltage: EN61010600VCatIII
*All specifications are subject to change without notice
10 RMS Voltage Simple Logger® L205/L230/L260
CHAPTER 4
OPERATION
4.1 Software Installation
Minimum Computer Requirements• Windows®7, 8, 8.1, 10• Processor-486orhigher• 8MBofRAM• 8MBofharddiskspaceforapplication,400Kforeachstoredfile• One9-pinserialport;oneparallelportforprintersupport• CD-ROMdrive
1. InserttheSimpleLogger®CDintoyourCD-ROMdrive.If auto-run isenabled, theSetupprogramwill start automatically. Ifauto-runisnotenabled,selectRunfromtheStartmenuandtypeinD:\SETUP (ifyourCD-ROMdrive isdriveD. If this isnot thecase,substitutetheappropriatedriveletter).
2. TheSet-up windowwillappear.
Figure 1
RMS Voltage Simple Logger® L205/L230/L260 11
Thereareaseveraloptions tochoosefrom.Someoptions(*) requireaninternetconnection.
• Simple Logger, Version 6.xx-InstallstheSimpleLogger®soft-waretothecomputer.
• *Acrobat Reader-LinkstotheAdobe®websitetodownloadthemostrecentversionofAdobe®AcrobatReader.Acrobat Reader isrequired for viewing PDF documents supplied on the CD-ROM.
• *Check for Available Software Updates - Opens the AEMCSoftwareupdatewebsite,whereupdatedsoftwareversionsareavailablefordownloading,ifnecessary.
• View User Guide and Manuals-OpensWindows®Explorerforviewingofdocumentationfiles.
3. Toinstallthesoftware,selectSimple Logger Software SetupinthetopsectionoftheSet-upwindow,thenselectSimple Logger,Version 6.xxintheOptionssection.
4. ClicktheInstallbuttonandfollowtheon-screenpromptstoinstallthesoftware.
4.2 Recording Data• Connecttheleadstotheloggerandtheotherendoftheleadstothe
conductortobemeasured.
Overload Warning: If the LED is continuously lit, disconnect your logger immediately
• PressthePRESSbuttononthetopoftheloggertobegintherecordingsession.TheLEDindicatorwilldouble-blinktoindicatethattherecord-ingsessionhasstarted.
• Whentherecordingsessionhasbeencompleted,pressthePRESSbuttontoendtherecording.TheLEDindicatorwillsingle-blinktoindi-catetherecordingsessionhasendedandtheloggerisinStand-by.
• Disconnecttheleadsfromtheconductorandconnecttheloggertothecomputerfordatadownloading.SeetheUserGuideontheCD-ROMfordownloadinginstructions.
12 RMS Voltage Simple Logger® L205/L230/L260
4.3 Using the SoftwareLaunchthesoftwareandconnecttheRS-232cablefromyourcomputertothelogger.
NOTE: A language will need to be selected the first time the program is launched.
Select Port from themenu bar and select theCom port (COM 1, 2 3or4)youwillbeusing(seeyourcomputermanual).Once thesoftwareautomatically detects the baud rate, the logger will communicate withthecomputer. (IDnumberof the loggerandnumberofpoints recordeddisplayed).Select Download to display the graph (download takes about 90 sec-onds).
RMS Voltage Simple Logger® L205/L230/L260 13
CHAPTER 5
MAINTENANCE
5.1 Battery InstallationUndernormalconditions,thebatterywill lastuptoayearofcontinuousrecordingunlesstheloggerisrestartedveryfrequently.IntheOFFmode,theloggerputsalmostnoloadonthebattery.UsetheOFFmodewhentheloggerisnotinuse.Replacethebatteryonceayearinnormaluse.Iftheloggerwillbeusedattemperaturesbelow32°F(0°C)orisfrequentlyturnedonandoff,replacethebatteryeverysixtoninemonths.
1. Makesureyourloggeristurnedoff(noblinkinglight)andallinputsaredisconnected.
2. Turntheloggerupsidedown.RemovethefourPhillipsheadscrewsfromthebaseplate,thentakeoffthebaseplate.
3. Locatethetwo-wire(red/black)batteryconnectorandattachthe9Vbatterytoit.Makesurethatyouobservepolaritybyliningupthebatterypoststotheproperterminalsontheconnector.
4. Oncetheconnectorispluggedontothebattery,insertthebatteryintotheholdingcliponthecircuitboard.
5. Iftheunitisnotinrecordmodeafterinstallingthenewbattery,dis-connectitandpressthebuttontwicethenreinstallthebattery.
6. ReattachthebaseplateusingthefourscrewsremovedinStep2.
Yourloggerisnowrecording(LEDblinking).PressthePRESSbuttonforfivesecondstostoptheinstrument.
NOTE: For long-term storage, remove the battery to prevent discharge effects.
5.2 CleaningThe body of the logger should be cleanedwith a clothmoistenedwithsoapywater.Rinsewithaclothmoistenedwithcleanwater.Donotusesolvent.
1� RMS Voltage Simple Logger® L205/L230/L260
APPENDIX A
Importing .TXT Files into a Spreadsheet
Opening a Simple Logger .TXT file in Excel
ThefollowingexampleusedwithExcelVer.7.0orhigher.
1. After opening theExcel program, select “File” from the mainmenuandthenselect“Open”.
2. In the dialog box that appears, browse and open the folderwhere your logger .TXT files are stored.Thiswill be locatedinC:\Program Files\Simple Logger 6.xx ifyouacceptedthedefaultchoiceofferedbytheloggerinstallationprogram.
3. Next,changethefiletypeto“Text Files”inthefieldlabeledFilesofType.Allthe.TXTfilesintheloggerdirectoryshouldnowbevisible.
4. Double-clickonthedesiredfiletoopentheTextImportWizard.
5. Reviewtheselectionsinthefirstwizardscreenandmakesurethatthefollowingchoicesareselected:
OriginalDataType: DelimitedStartImportatRow: 1FileOrigin: Windows(ANSI)
6. Clickthe“NEXT”buttonatthebottomoftheWizarddialogbox.Thesecondwizardscreenwillappear.
7. Clickon“Comma” intheDelimitersbox.Acheckmarkshouldappear.
8. Clickthe“NEXT” buttonatthebottomoftheWizarddialogbox.Thethirdwizardscreenwillappear.
9. Aviewoftheactualdatatobeimportedshouldappearinthelowersectionofthewindow.Column1shouldbehighlighted.IntheColumnDataFormatwindow,select“Date”.
10. Next,clickon“Finish”tocompletetheprocessandimportthedata.
11. Thedatawillnowappearinyourspreadsheetintwocolumns(AandB)andwilllooksimilartothatshowninFigureA-1.
RMS Voltage Simple Logger® L205/L230/L260 15
A B
8 Arms 35401.49 3.535401.49 535401.49 935401.49 13.535401.49 1735401.49 2035401.49 23.535401.49 27.535401.49 3135401.49 34.535401.49 38
Figure A-1. Sample Data Imported into Excel.
Formatting the Date and Time
Column‘A’containsadecimalnumberthatrepresentsbothdateandtime.Excelcanconvertthisnumberdirectlyasfollows:
1. Clickoncolumn‘B’atthetopofthecolumntoselectthedata,thenclickon“Insert”fromthemainmenuandselect“Columns”fromthedrop-downmenu.
2. Next,clickoncolumn‘A’atthetopofthecolumntoselectthedata,thenclickon“Edit”fromthemainmenuandselect“Copy”tocopytheentirecolumn.
3. Clickoncell1ofcolumn‘B’andthenclickon“Edit” andselect“Paste”toinsertaduplicateofcolumn‘A’intocolumn‘B’.Thisisnecessaryifyouwanttoshowthedateandtimeintwoseparatecolumns.
4. Next,clickonthetopofcolumn‘A’,thenclickon“Format”andselect“Cells”fromthedrop-downmenu.
16 RMS Voltage Simple Logger® L205/L230/L260
5. Inthedialogboxthatopens,selectthe“Date”optionfromthecategorylistontheleft.Selectthedateformatyoudesireandclickon“OK”toformatthecolumn.
6. Clickonthetopofcolumn‘B’,thenclickon“Format”andselect“Cells” fromthedrop-downmenu.
7. Inthedialogboxthatopens,selectthe“Time”optionfromthecategorylistontheleft.Selectthetimeformatyoudesireandclickon“OK”toformatthecolumn.
FigureA-2showsatypicalspreadsheetwithdate,timeandvaluedisplayed.Itmaybenecessarytochangethecolumnwidthtoseeallthedata.
A B C
12/02/04 11:45AM 1712/02/04 11:45AM 2012/02/04 11:45AM 23.512/02/04 11:45AM 27.512/02/04 11:45AM 3112/02/04 11:45AM 34.512/02/04 11:45AM 3812/02/04 11:45AM 41.512/02/04 11:45AM 45.512/02/04 11:46AM 4912/02/04 11:46AM 52
Figure A-2. Shows Date, Time and Value
RMS Voltage Simple Logger® L205/L230/L260 1�
Repair and Calibration
Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthatitbescheduledbacktoourfactoryServiceCenteratone-yearintervalsforrecalibration,orasrequiredbyotherstandardsorinternalprocedures.
For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknowifyouwantastandardcalibration,oracalibrationtraceabletoN.I.S.T.(Includescalibrationcertificateplusrecordedcalibrationdata).
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments15FaradayDriveDover,NH03820USAPhone:(800)945-2362(Ext.360)
(603) 749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]
(Orcontactyourauthorizeddistributor)Costsforrepair,standardcalibration,andcalibrationtraceabletoN.I.S.T.areavailable.NOTE: You must obtain a CSA# before returning any instrument.
Technical and Sales Assistance
Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:
ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments200FoxboroughBoulevardFoxborough,MA02035USAPhone:(800)343-1391
(508) 698-2115Fax: (508)698-2118E-mail: [email protected]
NOTE: Do not ship Instruments to our Foxborough, MA address.
1� RMS Voltage Simple Logger® L205/L230/L260
Limited Warranty
The Simple Logger® Model L205/L230/L260 is warranted to the ownerfor a period of two years from the date of original purchase againstdefects in manufacture. This limited warranty is given by AEMC®Instruments, not by the distributor from whom it was purchased. Thiswarrantyisvoidiftheunithasbeentamperedwith,abusedorifthedefectisrelatedtoservicenotperformedbyAEMC®Instruments.
For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.
What AEMC® Instruments will do:If a malfunction occurs within thewarranty period, you may return the in-strument to us for repair, provided we have your warranty registrationinformation on file or a proof of purchase. AEMC® Instruments will, at itsoption,repairorreplacethefaultymaterial.
REGISTER ONLINE AT:www.aemc.com
Warranty Repairs
What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:
Ship To: ChauvinArnoux®,Inc.d.b.a.AEMC®Instruments15FaradayDrive•Dover,NH03820USAPhone:(800)945-2362(Ext.360)
(603) 749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]
Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyourreturnedmaterial.
NOTE: You must obtain a CSA# before returning any instrument.
08/18
99-MAN 100210 v16
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346
www.aemc.com
INTRODUCTIONPRODUCT FEATURES2.1Indicators and Buttons2.2Inputs and Outputs2.3Mounting
SPECIFICATIONS3.1Electrical Specifications3.2Mechanical Specifications3.3Environmental Specifications3.4Safety Specifications
OPERATION4.1Software Installation4.2Recording Data4.3Using the Software
MAINTENANCE5.1Battery Installation5.2Cleaning
APPENDIX A: Importing .TXT Files into a SpreadsheetRepair and CalibrationTechnical and Sales AssistanceLimited WarrantyWarranty Repairs