SimMan 3G Software Installation Instructor PC and Patient Monitor PC: First upgrade the Instructor PC, then repeat the procedure for the Patient Monitor PC. 1 - Power on computer and ensure no SimMan 3G applications are running. 2. Un-install any previous versions of SimMan 3G software: Use the Windows Control Panel or select: <Start> <All Programs> <SimMan 3G> <Uninstall SimMan 3G> 3. Insert the software DVD and allow auto-run. If auto-run is disabled on your computer, please manually execute the <autorun.exe> file on the DVD. 4. In the installation menu – click the <Install SimMan 3G Software complete> button once. This will start copying files to your computer. Note that the copying process may take several minutes. Please wait. Once the copying process is finished, please follow instructions in the installation wizard: Click on <Next> three times (3X). Then click on <Install> Manikin Update: CAUTION! Please note that you must not power off the manikin during the following process! Never update the simulator over WLAN (wireless): 1 - Make sure you have installed the software on the Instructor PC as described above. 2 - Turn off WLAN (wireless) using the slider switch on the side of the Instructor PC tablet. Set switch to 0. 3 - Power on the manikin. Wait until manikin is fully powered up and the top orange light turns green. Do not start the Instructor Application or Patient Monitor Application. 4 - Connect a network (ethernet) cable from the LAN port on manikin to the LAN port on the back of the Instructor PC. Allow 60sec for a connection to be established. 5 - Make sure the manikin is ON and breathing. Then execute < Start><All Programs><SimMan 3G><Manikin Update>. 6 - Identify and click on the manikin you wish to update from the list that appears and click on <OK> 7 - The Manikin Update program will start searching the software versions currently installed. This will take about 2 minutes. 8 - All programs and software versions are listed. The update status for each program is displayed under <Update>. 9 - Click OK to begin updating the software. Do not disconnect the manikin, power supply or network cable. This may cause serious damage to the manikin. Please wait while the manikin is prepared for update Note: manikin might stop and start breathing intermittently. SimMan 3G DFU Update - Installing SimMan 3G software

SimMan 3G - Microsoftlaerdalcdn.blob.core.windows.net/downloads/f221/SimMan3G_Upgrad… · SimMan 3G Software Installation Instructor PC and Patient Monitor PC: First upgrade the

Embed Size (px)

Citation preview

Page 1: SimMan 3G - Microsoftlaerdalcdn.blob.core.windows.net/downloads/f221/SimMan3G_Upgrad… · SimMan 3G Software Installation Instructor PC and Patient Monitor PC: First upgrade the

SimMan 3G Software Installation

Instructor PC and Patient Monitor PC:First upgrade the Instructor PC, then repeat the procedure for the Patient Monitor PC.

1 - Power on computer and ensure no SimMan 3G applications are running.

2. Un-install any previous versions of SimMan 3G software: Use the Windows Control Panel or select:

<Start>

<All Programs>

<SimMan 3G> <Uninstall SimMan 3G>

3. Insert the software DVD and allow auto-run. If auto-run is disabled on your computer, please manually execute the <autorun.exe> file on the DVD.

4. In the installation menu – click the <Install SimMan 3G Software complete> button once. This will start copying files to your computer.

Note that the copying process may take several minutes. Please wait.

Once the copying process is finished, please follow instructions in the installation wizard:

Click on <Next> three times (3X). Then click on <Install>

Manikin Update:CAUTION!

Please note that you must not power off the manikin during the following process! Never update the simulator over WLAN (wireless):

1 - Make sure you have installed the software on the Instructor PC as described above.

2 - Turn off WLAN (wireless) using the slider switch on the side of the Instructor PC tablet. Set switch to 0.

3 - Power on the manikin. Wait until manikin is fully powered up and the top orange light turns green. Do not start the Instructor Application or Patient Monitor Application.

4 - Connect a network (ethernet) cable from the LAN port on manikin to the LAN port on the back of the Instructor PC. Allow 60sec for a connection to be established.

5 - Make sure the manikin is ON and breathing. Then execute < Start><All Programs><SimMan 3G><Manikin Update>.

6 - Identify and click on the manikin you wish to update from the list that appears and click on <OK>

7 - The Manikin Update program will start searching the software versions currently installed. This will take about 2 minutes.

8 - All programs and software versions are listed. The update status for each program is displayed under <Update>.

9 - Click OK to begin updating the software. Do not disconnect the manikin, power supply or network cable. This may cause serious damage to the manikin. Please wait while the manikin is prepared for update

Note: manikin might stop and start breathing intermittently.

SimMan 3G DFU Update - Installing SimMan 3G software

Page 2: SimMan 3G - Microsoftlaerdalcdn.blob.core.windows.net/downloads/f221/SimMan3G_Upgrad… · SimMan 3G Software Installation Instructor PC and Patient Monitor PC: First upgrade the

10 - Complete the update wizard.

Screens similar to the one shown above will display as the SimMan 3G modules update.

This phase may take several minutes

Note that some components might fail updating on first attempt. Please ignore this and allow the update process to continue.

The SimMan 3G Update Manager - Report will show all updates and show which ones were performed succesfully or failed. Click <OK>

If any updates have failed, run Manikin Update again. Should the same module fail repeatedly, check to see whether all internal wiring is connected and functioning properly.

If problem persists, please contact your local technical service centre.

11- Wait for manikin to resume spontaneous breathing.

Troubleshooting

1 - Should the following message appear on the screen during the update, click <Yes> to continue installing updates.

2 - In the SimMan 3G Manager - Report: If any updates have failed, run Manikin Update again. Should the same module fail repeatedly, check to see whether all internal wiring is connected and functioning properly.

If problem persists, please contact your local technical service centre.

Advanced Video Server (AVS):If you use an AVS with your SimMan 3G system, please make sure that it uses the latest SW available.

Internal no: 20-00610

18.11.2009