SERVO-i WITH HELIOX OPTION
GROWING WITH THE CUSTOMER’S NEEDS
SERVO-i System Version 4.0
Heliox option
Low MV alarm audio off option
Possibility to switch off back-up ventilation
Other usability improvements
HELIOX – GAS PROPERTIES
What is Heliox?
Heliox – HeO2 – is a mixture of helium (He) and oxygen (O2) (e.g. 80/20; 70/30).
Helium is an inert, low density gas (~1/7 the density of air and oxygen).
Lower density leads to a more laminar flow (less turbulence) and thus reduces resistance in the airways.
Laminar gas flow (A) is more efficient than turbulent flow (B).
HELIOX – GAS PROPERTIES
HELIOX – EFFECTS ON VENTILATION
Effects of Heliox therapy according to the literature
Established positive effects on ventilation
Reduces work of breathing
Increases CO2 diffusion
Potential positive effects on ventilation
Reduces the peak pressure needed to maintain flow
Increases CO2 elimination and improves gas exchange
Increases expiratory flow
Reduces respiratory rate and auto PEEP
Increases drug deposition in distal parts of the airways.
Upper airway obstruction
Lower airway obstruction
HELIOX – USES AND APPLICATIONS
For which patients is Heliox most commonly used?
HELIOX – POTENTIAL BENEFITS
Avoiding endotracheal intubation
Protecting the lungs
Improving patient outcome
Enhancing patient comfort
Potential benefits of Heliox therapy according to the literature
Available as an option for the SERVO-i
Field upgrade
Factory installed
The option consists of:
Heliox SW
Heliox adapter kit
SERVO-i WITH HELIOX OPTION
General
Can be used with all pediatric and adult patients, as well as neonates
Compatible with all invasive/non-invasive ventilation modes
Supporting mobility from emergency room to MR environment
Connection of 80/20; 79/21; 78/22 mixtures possible
Adjustable O2 concentration (21 to 100%).
SERVO-i WITH HELIOX OPTION
Functionality
Unique solution for automatic gas identification
Automatic compensation of flow delivery
Automatic compensation of CO2 and volume monitoring
SERVO-i WITH HELIOX OPTION
Key benefits of using SERVO-i for Heliox therapy
Quick and easy start of therapy
Quick and easy switch from air to Heliox (and vice versa) without going to standby
User friendly interface
Cost efficient Heliox therapy thanks to the SERVO-i’s low gas consumption.
Key benefits of using SERVO-i for Heliox therapy
SERVO-i WITH HELIOX OPTION
SERVO-i WITH HELIOX OPTION
Four different adapters are available:
DISS (both air (A) and Heliox (B))
NIST/NIST
AGA/NIST
French/NIST
Heliox adapters
SERVO-i WITH HELIOX OPTION
Color marked for Heliox
Brown/green for Heliox DISS
Brown/white for Heliox NIST
Attached to ventilator with safety wire
Attach to air module when using Heliox
Heliox adapters
SERVO-i WITH HELIOX OPTION
Connecting Heliox to the ventilator – Heliox adapters NIST / AGA / French
Remove adapter from holder.
Connect adapter to Heliox high pressure hose.
Connect adapter to air module inlet.
SERVO-i WITH HELIOX OPTION
Preparations before using Heliox with SERVO-i
SERVO-i WITH HELIOX OPTION
Automatic gas identification
Performed by the ultrasonic expiratory flow sensors
Available during Pre Use Check and during ventilation
Active as long as the set O2 concentration is 75% or below
Time for identification is determined by tidal/minute volume and O2 concentration used
SERVO-i WITH HELIOX OPTION
Pre Use Check
SERVO-i WITH HELIOX OPTION
Automatic identification when changing from air to Heliox during ventilation
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION
Gas type change – manually initiated compensation for Heliox gas
SERVO-i WITH HELIOX OPTION
Non-invasive application
SERVO-i WITH HELIOX OPTION
Infant application
Part Description Part NumberHeliox software, SERVO-i 66 75 580Heliox adapter kit, new SERVO-iDISSNIST/NISTAGA/NISTFrench/NIST
66 75 58566 75 58266 75 59066 75 587
Heliox adapter kit, onsite upgradeDISSNIST/NISTAGA/NISTFrench/ NIST
66 77 55266 77 55566 55 56066 77 557
SERVO-i up to serial number 40.000 also requires:PC board 1784 rev 16 or laterPC board 1785 rev 3 or later
64 67 98464 68 016
High pressure Heliox hose DISS, 5 m 66 75 592High pressure Heliox hose NIST, 5 m 66 75 607
Order information
SERVO-i WITH HELIOX OPTION
The Heliox option is not compatible with the O2 Sensor and Y Sensor measuring.
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
Low Minute Volume alarm audio off
SERVO-i 4.0 – OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Possibility to disable back-up ventilation
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Showing previous Pre Use Check/Suction Support FiO2 retention
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Maintenance notifications/Changed calculation of time for next PM
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Previous mode information/Initiating a Pre Use Check
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
New alarm symbol/Clearer difference between medium and high priority alarms
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
New patient category symbols
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Time scale on the waveforms
SERVO-i/s 4.0 – OTHER IMPROVEMENTS/CHANGES
Art. no. 6679957 rev. 01