Room for flexibility
The EasyDiagnost Eleva is a R/F system designed to perform a wide range of
applications from gastro-intestinal, various iodine and vascular procedures to
standard radiography. By combining the EasyDiagnost Eleva with a choice of a
vertical stand and/or a digital detector in the table you can create your own
2-in-1 room solution. Set up a radiography and fluoroscopy room that perfectly
fits your needs – from a single detector version to a high-end full digital room
solution with all advantages of filmless and even cassetteless workflow.
Philips EasyDiagnost Eleva with digital Radiography/Fluoroscopy room solutions to combine both applications in one room
�
Content
04 Main components and options
06 EasyDiagnost Eleva
09 DoseWise
10 X-ray generation
12 Image detection
13 Image display
14 Eleva workspot
16 Extended Digital Imaging
20 Over-table technique
22 Digital vertical stand VS
24 Vertical stand VS advanced
26 Ceiling suspension CS
27 Height adjustable trolley TA-M
29 Image plate reader PCR Eleva S Plus
30 DRF room solution – Configuration proposals
32 Room planning
35 General
�
Main components and options*
EasyDiagnost Eleva ✓
Spot-film device (Digital Camera) ✓
Under-table X-ray tube assembly ✓
Tube lift ❍
Image detection, incl. image intensifier and TV camera ✓
Image display, incl. Eleva Examination Control ✓
LCD monitor in examination room ✓
Monitor trolley for LCD monitor ✓
Ceiling suspension for one or two LCD monitors ❍
Eleva workspot ✓
Extended Digital Imaging, incl. image processing and storage with ViewForum ✓
Angiography (with Bolus Chase) ❍
Footswitch for fluoroscopy and exposure ✓
Accessories ✓/❍
DoseWise ✓/❍
High voltage generator Velara RF ✓
DICOM Worklist Management ❍
DICOM MPPS ❍
DICOM image export ✓
DICOM print ✓
DICOM Media on CD-R ❍
mShield ✓
Bucky tray ❍
EasyLat ❍
Digital in-table detector ❍
� ✓ = standard ❍ = optional
Main components and options
* Please contact your local sales specialist for additional system extensions and accessories
�
Main components and options
Ceiling suspension with X-ray tube assembly with control grip and collimator (CS2/4)
❍
Digital vertical stand VSMotorized tilting – 20º to +90ºDigital flat detectorRemovable gridAutomatic exposure control (AEC)� AEC measuring fieldsTrackingAutomatic image stitching incl. Patient supportServo supported CS movements (with automatic image stitching)
❍
❍
✓
✓
✓
✓
❍
❍
❍
Vertical stand VS advancedMotorized tilting – 20º to +90ºCassette trayRemovable gridAutomatic exposure control (AEC)� AEC measuring fields
Tracking (detector in vertical stand with tilt)
❍
❍
✓
✓
✓
✓
❍
Height adjustable trolley TA-M ❍
UNIQUE image processing ✓
Reject analysis ❍
PCR Eleva S Plus ❍
Integration in existing PCR Eleva network ❍
Philips CAD Chest solution xLNA ❍
✓ = standard❍ = optional
�
The EasyDiagnost Eleva stand
consists of:
• Table
• Cassette tray, digital detector unit or
EasyLat (option)
• Spot-film device (Digital Camera)
• Compressor
• Tube assembly
• Tube lift (option)
Table
DimensionsTable heightTable top
8�0 mm2,000 mm x �90 mm
(�2.7")(78.7" x 27.2")
✓
✓
Table tiltTilt angle
Speed
90° / –20°90° / –�0°90° / –��°90° / –8�°1°/s to �°/s variable
✓
❍
❍
❍
✓
Table top
MovementsLateralLongitudinal
–100 mm to +90 mm±8�0 mm
(–�.9" to +�.�")(±��")
✓
✓
Movement speedLateralLongitudinalwith bolus chase acquisition
�2 mm/s�0 mm/s0–1�0 mm/s
(1.7"/s)(2.�"/s)(0–�.9"/s)
✓
✓
✓
X-ray attenuation aquivalent0.7 mm Al typical(at 100 kV, 2.7 mm Al HVL)
✓
Maximum patient weightfor all movementsstatic, in horizontal or verticaltable position (incl. footrest)
180 kg2�0 kg
(�00 lbs)(��0 lbs)
✓
✓✓ = standard❍ = optional
EasyDiagnost Eleva
7
All dimensions in mm (feet/inches)
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")750
(29.5")340
(13.4") 520(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")
2115
(6'
11.
3")
2950
(9'
8.1
")
1558 (5' 1.3")
1775
(5'
9.9
")
1500 (4' 11.1")1075 (3' 6.3")
554
(21.
8")
2399
for
23 c
m I.
I. (7
' 10.
5" fo
r 9"
I.I.)
2390
for
38 c
m I.
I.
(7' 1
0.1"
for
15"
I.I.)
2950
(9'
8.1
")
8
Spot-film device (Digital Camera)
MovementsLateralParkingLongitudinal
220 mm�20 mm7�0 mm
(8.7")(20.�")(29.�")
✓
✓
✓
Gridwithout tube lift option
with tube lift option
�0 l/cmr = 10:1, fo = 800 mm�0 l/cmr = 10:1, fo = 900 mm
✓
✓
Grid parking Automatic movement to parking position ✓
Compressor
Compression cone Integrated in spot-film device ✓
Compression movement �0� mm (12") ✓
Compression stop 10 steps, manually selectable ✓
Compression pressure servo supported ✓
Tube lift (option)
Speed �0 mm/s (2"/s) ✓
Focus – Table top distance �10 mm to ��0 mm (20" to 2�.�") ✓
✓ = standard❍ = optional
EasyDiagnost Eleva (contd.)
750 (29.5")
340 (13.4") 520
(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
All dimensions in mm (feet/inches)
750 (29.5")
340 (13.4") 520
(20.5")
1340 (4' 4.8")
620 (24.4")
film
38 cm I.I. (15")
975 (38.4") 1
2000 (6' 6.7")
835 (32.9")
835 (32.9")
690
(27.
2")
100
(3.9
")85
(3
.3")
554
(21.
8")
833
(32.
8")
max
. 210
4 (m
ax. 6
' 10.
8")
626
(24.
6")
155
(6.1
")49
0(1
9.3"
)
512
190
750
(29.
5")
140
1075 (3' 6.3") 885 mm (34.8") – with EasyLat
108.5 108.5 (4.3")
1000 (3' 3.4")
1
9
DoseWise
✓ = standard❍ = optional
Less radiation time
FluoroGrab ✓
Dynamic FluoroGrab (filming) ✓
Preset collimation ✓
Collimation on LIH ✓
Automatic parameter setting ✓
High resolution LCD monitor (flicker-free) ✓
More awareness
Dose calculation ✓
Dose measurement ❍
Dose display ✓
Dose reporting ✓
X-ray disable function ✓
X-ray shieldingfixed and flexible lead curtains againstscattered radiation
✓
Smart beam
Pulsed Fluoroscopy (PF), In-pulse controlled ❍
Grid Controlled Fluoroscopy (GCF), In-pulse controlled
❍
Intelligent Exposure (IQX), In-pulse controlled ✓
SpectraBeam RF ✓
Adaptive Measuring Field ✓
Tube lift ❍
Iris shutter and two pairs of parallel shutters ✓
Parkable Grid (in spot-film device) ✓
✓ = standard❍ = optional
10
X-ray generation
Generator Velara RF
Power pack �0 kW�� kW80 kW
✓
❍
❍
Tube support SRO 2��0SRM 22�0 GSSRO ��100 (for CS)
✓
❍
❍
Exposure techniques • Manual: kV-mAs or kV-mA-s• Automatic Exposure Control (AEC), based on
falling load and fixed current techniques• Intelligent Exposure (IQX), In-Pulse controlled• Automatic kV reduction techniques
✓
✓
✓
❍
Fluoroscopy techniques • Automatic kV/mA control during fluorscopy• Pulsed Fluoroscopy (PF), In-pulse controlled• Grid Controlled Fluoroscopy (GCF),
In-pulse controlled
✓
❍
❍
Tube voltage exposureTube voltage fluoroscopy
�0 to 1�0 kV�0 to 110 kV
Tube current exposureTube current fluoroscopy
1 to 1100 mA0.2 to � mA
depending on power pack
mAs product 0.� to �00 mAs0.� to 8�0 mAs
with AECwithout AEC
Exposure times 1 ms to � s with AEC
Access times (tube dependent)Standby to fluoroscopyFluoroscopy to radiographyRadiography to fluoroscopy
< 0.� s< 0.8 s< 0.8 s
Dose managementDose calculationDose measurement
Entrance dose, entrance dose rate, dose-area productDose-area product
✓
❍
Safety Tube overload protectionAutomatic mains voltage compensation
✓
✓
✓ = standard❍ = optional
Integrated generator
control
11
Tube SRO 2550 SRM 2250 GS SRO 33100 (for CS)
Focal spot 0.� / 1.0 0.� / 1.0 0.� / 1.0
Anode diameter 90 mm 100 mm 90 mm
Anode angle 1�º 1�º 1�º
Anode heat storage capacity �00 kHU (220 kJ) �80 kHU (280 kJ) �80 kHU (280 kJ)
Maximum voltage 1�0 kV 12� kV 1�0 kV
Max. continuous heat dissipation 2�0 W �00 W 2�0 W
Max. heat content of assembly 1700 kHU (12�0 kJ) 20�0 kHU (1�10 kJ) 1700 kHU (12�0 kJ)
Grid switch No Yes No
Collimator for Under-table technique Over-table technique
Collimator controlBuilt-in µ-controller and motor controllersPower-up self testAuto calibration at power-upError logging, reporting
✓ ✓
SpectraBeam RFProgrammable filter pre-setsMax. number of pre-filters
✓
�✓
�
Rectangular collimationMotor drivenTwo pairs of parallel shuttersMax. opening �1 cm (20")Positioning accuracy 0.�% of SID
✓ ✓
Circular collimationOctagonal ShapeMotor drivenPositioning accuracy 0.�% of SID
✓ –
Collimation on LIHFor regular shuttersOn last fluoroscopy imageOn last exposure image
✓
✓
✓
–––
✓ = standard❍ = optional
12
Image Intensifier (I.I.)
Mode2�0 mm, 170 mm, 1�0 mm �10 mm, 2�0 mm, 200 mm, 170 mm�80 mm, �10 mm, 2�0 mm, 170 mm
(9", 7", �.�") or(12", 10", 8", 7") or(1�", 12", 10", 7")
Input window Titanium
Detective Quantum Efficiency ��%
Modulation Transfer Factor ��% at 1 lp/mm in �8 cm (1�'') mode
Central spatial resolution �80 mm (1�'') mode:170 mm (�.7'') mode:
� lp/mm (1�.7 lp/ '')� lp/mm (2�.� lp/ '')
Conversion factor in �8 cm (1�'') mode,typical
�00 cd/m2/(mR/s); ��,8 cd/m2/(µGy/s)
Structure noise Negligible
Contrast ratio170 mm (�.7'') (10% disc): �80 mm (1�'') (10% disc):
��:122:1
Image detection
Camera
Type CCD
Matrix 102� x 102� and �12 x �12
Dynamic range Between �� and 70 dB
Digital output 12 bit
Horizontal / Vertical scan reverse
Yes
Dose control, automatic In-pulse controlled
Automatic Gain Control In-frame gain control
Measurement fields 2 fields configurable
Application range (dose) � to 10 000 nGy/image
1�
Image display
Monitor
Type LCD (Liquid Crystal Display) ✓
Monochrome display ��0 mm (18 ) ✓
Progressive display (flicker free) ✓
Native format 1280 x 102� SXGA ✓
Nominal brightness �00 Cd/m2 ✓
Monitor carriers
Trolley for one LCD monitorfor two LCD monitors
✓
❍
Ceiling suspension for one LCD monitorfor two LCD monitors
❍
❍
✓ = standard❍ = optional
1�
Eleva workspot
✓ = standard❍ = optional
Eleva workspot
Hard disk • 1�0 GB total• 108 GB for the image data (up to approximately �,700 images)
RAM storage capacity 2 GB
Interfaces • Ethernet 10/100/1000 Base-T Gigabit• DICOM interface• Detector interface• Memory stick support for QC
CD writer �8x CD/DVD reader, CD writer
Monitor 19''-LCD color touchscreen monitor EEC-A ✓with hardware frame for ergonomic workflow
19''-LCD color touchscreen monitor EEC-E ❍(space saving, recommended just for configurations without second plane)
Keyboard with mouse and function buttons
For entering administrative patient data and for operating the screen menus
Barcode reader Error free input of patient data ❍and patient selection
Generator control Integrated ✓EPX-programmed ✓
Image data
Data volume Up to 18 MB/image
Matrix depth 1� bit/pixel
1�
468 (18.4")
510.
75 (
20.1
")34
2 (1
3.5"
)
456.75 (18.0")
All dimensions in mm (feet/inches)
293 (11.5")
61 (2.4")
386
(15.
2")
207 (8.2")
23 (2.4")
All dimensions in mm (feet/inches)
429 (16.9")35
9 (1
4.1"
)
EEC-A EEC-E
1�
Extended Digital Imaging
Fluoroscopy
Digital high resolution continuous fluoroscopy ✓
Grid Controlled Fluoroscopy: 0,�-�0 pulses/sec (�0/�0Hz) ❍
Pulsed fluoroscopy, 0,�-� pulses/sec (�0/�0Hz) ❍
FluoroGrab ✓
Dynamic FluoroGrab (up to �0 images/sec) ✓
Fluoro images freeze (last image hold) ✓
Trace subtract fluoroscopy ❍
Dynamic noise reduction ✓
Acquisition
First time right IQ with harmonization and auto windowing ✓
Matrix size 102�2 and �122 ✓
Acquisition speed: up to 8 images/sec (102�2 and �122) ✓
Acquisition speed: up to �0 images/sec (�122) ❍
Adjustable acquisition speed (during run) ✓
Live subtraction ❍
Standard acquisition RAM capacity (0.� GB) ✓
Max. acquisition RAM capacity (1 GB) ❍
External video display ✓
Bolus chase (non subtracted) ❍
✓ = standard❍ = optional
17
Image processing
Harmonization ✓
Processing depth (bit) 1�
Video invert ✓
Zoom variable up to x �.7 ✓
Magnifying glass and panning ✓
Manual adjustable contrast/brightness/edge enhancement ✓
Electronic shutters (automatic & manual) ✓
Flip/rotate ✓
Annotation 'free text and advanced' ✓
Measurements: length, angle and area ✓
Vascular image processing
Subtract (single mask) ❍
Run subtract (multiple mask) ❍
Measurements extended: area, histogram, vessel analysis (automatic) ❍
Re-masking ❍
Pixel shift ❍
Split screen pixel shift ❍
Auto warp (automatic pixel shift) ❍
Land marking/road mapping: acquisition ❍
View trace (max. opacification) ❍
CO2 imaging (min. opacification) ❍
✓ = standard❍ = optional
18
Viewing
View switch time between patients stored in RAM: 0.1 sec ✓
View switch time between runs stored in RAM (within one patient): 0.1 sec ✓
Dynamic view & run cycle: up to �0 fps ✓
Overview (user selectable number of images) ✓
Pictorial index ✓
Parallel working ❍
Multimodality viewing (CT/MR/…) in control- and exam room ❍
Direct Mouse Manipulation (DMM) ✓
Default display protocols ✓
User log-on ✓
Image storage
Hard disk ✓
DVD/CD-recordable (background storage and retrieval) ❍
Flagging for selective storing, printing and export of images ✓
(Automatic) export of image to network servers (e. g. PACS) ❍
Image reconstruction
ColonMap (complete abdominal overview) ❍
Bolus Chase Reconstruction (complete overview of peripheral vasculature) ❍
✓ = standard❍ = optional
Extended Digital Imaging (contd.)
19
Printing
DICOM Print SCU ✓
Paper printing ✓
Manual compose print (user dependent layout) ✓
Protocol print (user selects layout from a list) ✓
Auto protocol print ✓
Print protocol editor ❍
Automatic annotation of hardcopies ✓
Networking
DICOM store SCP ✓
DICOM store SCU/SCP ❍
DICOM storage commit SCU ❍
DICOM query/retrieve SCU/ SCP ❍
DICOM worklist management SCU ❍
DICOM MPPS SCU ❍
DICOM presentation state ✓
DICOM grey scale display standard ✓
Multitasking (background communication) ✓
Serviceability❍
Remote service support ✓
Digital image quality measurement tool ✓
✓ = standard❍ = optional
20
Over-table technique
EasyLat
Type Fold-out cassette holder forhigh quality lateral exposures
Amplimat measuring chamber ✓
Cassette-size sensing ✓
Grid 70 l/cm, r = 1�:1,f0 = 110 cm, stationary
✓
Cassette formatsfold-in
fold-out
1� cm x 18 cm to �� cm x �� cm(�" x 7" to 1�" x 17")�� cm x �� cm to �� cm x �� cm(1�" x 1�" to 1�" x 17")
✓ = standard❍ = optional
All dimensions in mm (feet/inches)
900 (35.4")
min
. cei
ling
heig
ht 2
720
(8' 1
1.1"
) *
min
. 860
(33
.9")
mov
emen
t ra
nge
0...1
480
(4' 1
0.3"
)
mov
emen
t ra
nge
0...1
495
(4' 1
0.9"
) (w
ithou
t ß-
stop
)m
in. 8
45 (
33.3
")
(with
out
ß-st
op)
max
. 234
0 (7
' 8.1
")38
0(1
5")
1200 (3' 11.2")
510
(20.
1")
1500 (4' 11.1")2653 (8' 8.5")
3000 (9' 10.1")
2000 (6' 6.7")1322 (4' 4.1") 1678 (5' 6.1")
1875 (6' 1.8")1090 (3' 6.9")
588(23.1")
375(14.8")
72(2.8")
Bucky unit in table
Amplimat measuring chamber ✓
Automatic cassette size sensing ✓
Removable grid Choice of � exchangable
oscillating grid types
Cassette formats 1� cm x 18 cm to �� cm x �� cm(�'' x 7'' to 1�'' x 17'')
✓ = standard❍ = optional
Digital detector unit
Amplimat measuring chamber ✓
Preset collimation ✓
Exam-type dependent collimation presets ✓
Type Flat detector made of amorphous silicon
Detector size �� cm x �� cm (17" x 17")
Image matrix size �,000 pixels x �,000 pixels
Pixel size 1�� µm
Image resolution up to �.� Lp/mm
Dimension �9� mm x �7� mm x 18� mm
(2�.�" x 22.�" x 7.2")
Removable grid Choice of � exchangable oscillating
grid types
✓ = standard❍ = optional
21
Digital vertical stand VS
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.8" to �’11")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip (option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
22
Detector unit
Type Flat detector made of amorphous silicon
Detector size �� cm x �� cm (17" x 17")
Image matrix size �,000 pixels x �,000 pixels
Pixel size 1�� µm
Image resolution up to �.� Lp/mm
Dimension �9� mm x �7� mm x 18� mm (2�.�" x 22.�" x 7.2")
Angle of tilt (option)
– 20º to +90º, motorized tilting
Automatic exposure control (AEC)
� AEC measuring fields
Weight approx. 2�.� kg (�8.� lbs)
Removable grid Choice of � exchangable
oscillating grid types
Grid storage For up to 2 grids within the detector unit
Digital vertical stand VS
with/without tilt
2�
The digital vertical stand VS
consists of:
• Column
• Detector unit
• Grid
• 2 user interfaces (left & right)
• 2 patient grips (left & right)
• Vertical stand display
All dimensions in mm (feet/inches)
64 (2.5") 107 (4.2")
575
(22.
6")
406 (16")710 (28")
825
(32.
5")
1173
(3'
10.
2")
1819
(5'
11.
6")
596 (23.5") 227 (8.9")
160(6.3")
1990
(6'
6.3
")20
80 (
6' 9
.9")
≤150
0 (4
' 11.
1")
≥300
(11.
8")
≤210
5 (6
' 10.
9")
543 (21.4")
dete
ctor
pla
ne
Buck
y un
it fr
ont
side
596
(23.
5")
50 (2.0") 54 (2.1")
902
(35.
5")
1286 (4' 2.6")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
596
(23.
5")
300(11.8")
902
(35.
5")
with spacer (option)
spacer
64 (2.5")107 (4.2")
1990
(6'
6.3
")20
80 (
6' 9
.9")
986 (38.9")
596
(23.
5")
120
(4.7
")
902
(35.
5")
712 (28.1")
20°
≥300
≤150
0 (4
' 11.
1")
≤210
5 (6
' 10.
9")
≥549
(21.
6")
dete
ctor
pla
ne
Buck
y un
itfr
ont
side
54 (2.1")
with tilting (option)
with tilting (option)
2�
Vertical stand VS advanced
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip(option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
Vertical stand VS advanced
with/without tilt
Detector unit (cassette tray)
Dimensions (w x h) �9� mm x �7� mm x 18� mm (2�.�" x 22.�" x 7.2")
Angle of tilt (option) – 20º to +90º, motorized tilting
Automatic exposure control (AEC)
� AEC measuring fields
Operating 2 control units (left & right)
Removable grid Choice of � exchangable oscillating grid types
Grid storage For up to 2 grids within the detector unit
Cassette formats 18 cm x 2� cm to �� cm x �� cm(8" x 10" to 1�" x 17")
2�
Column
Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit
Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")
Installation Floor and wall attachment or floor only (option)
Grips
Two patient grips On left and right of the detector unit
Patient stretch grip(option)
• Arranged on left or right of the detector unit• Rotatable
Brakes
Brakes Electromagnetic, self-locking
All dimensions in mm (feet/inches)
The vertical stand VS advanced
consists of:
• Column
• Detector unit (cassette tray)
• Grid
• 2 user interfaces (left & right)
• 2 patient grips (left & right)
64 (2.5")
107 (4.2")
139 (5.5")
50 (2.0")
Ref. cross point
543 (21.4")
fron
t si
de b
ucky
uni
t
cass
ette
film
pla
ne
902
(35.
6")
120
(4.7
")19
90 (
6' 6
.3")
2080
(9'
2.2
")
max
. 150
0 (4
' 11.
1")
max
. 210
5 (6
' 10.
9")
min
. 300
(11
.9")
1) space for cassette and grid exchange
139 (5.5") 54 (2.1")
spacerRef. cross point
1286 (4' 2.6")
300 (11.9")
902
(35.
5")
fron
t si
de b
ucky
uni
t
cass
ette
film
pla
ne
120
(4.7
")
1173
(3'
10.
2")
1513
(4'
11.
6")
535
(21.
1")
575
(22.
7")
710 (28.0")550 (21.7")
406 (16.0")
227 (8.9")
160 (6.3")
596 (23.5")
min. 1000 (3' 3.4")1
902
(35.
5")
cass
ette
film
pla
ne
fron
t si
de b
ucky
uni
t
120
(4.7
")
Ref. cross point
986 (38.9")
54 (2.1")139 (5.5")
64 (2.5")
107 (4.2")
708 (27.9")
2080
(6'
9.9
")
1990
(6'
6.3
")
max
. 150
0 (4
' 11.
1")
max
. 210
5 (6
' 10.
9")
min
. 300
(11
.8")
min
. 550
(21
.7")
20°
2�
The ceiling suspension CS2 / CS4
for over-table technique consists of:
• Column
• Rail system
• X-ray tube assembly with collimator
and control grip
• Optional: Motor for automatic image
stitching (see room layout on page ��)
Ceiling suspension CS2 / CS4
Column �-part Al telescopic column;adaptable to individual room heights
Weight-balanced vertical movement 1,�80 mm (�8.�") ✓
Format-sensing to control the automatic collimator
✓
Tube support SRO 2��0SRO ��100
✓
❍
Longitudinal travelwith extension rails
�,��0 mm�,0�0 mm
(1�9")(2��")
✓
❍
Transverse movementCS2CS�
1,�7� mm�,19� mm
(�0")(118")
✓
✓
Longitudinal/transverse centering up to 7 lock-in positions can be set
Servo supported longitudinal and tube movement (with automatic image stitching)
❍
Automatic tube alignment for tracking of vertical stand
❍
Tube assembly rotationaround horizontal axisaround vertical axis
±12�º+1��º / –180º
✓
✓
Laser slit light ✓
Ceiling suspension CS
✓ = standard❍ = optional
All dimensions
in mm (feet/inches)
brake rail
max
. cei
ling
heig
ht 3
110
(10'
2.4
") * min
. 125
0 (4
' 1.2
")m
in. 1
235
(4' 0
.6")
(w
ithou
t ß-
stop
)
mov
emen
t ra
nge
0...1
480
(4' 1
0.3"
)m
ovem
ent
rang
e 0.
..149
5 (4
' 10.
9")
(with
out
ß-st
op)
max
. 273
0 (8
' 11.
5")
380
(15"
)
* for max. maneuverability
27
Height adjustable trolley TA-M
Trolley
Height of table adjustment �00 mm to 870 mm (2�.�" to ��.�")
Dimensions with accessories, max. 2,270 mm x 900 mm (7' �.�" x ��.�")
Total weight 1�0 kg (28� lbs)
Brakes At each side
Bariatric table top
Type Floating table top of sandwich design with carbon fiber overlay
Dimension table top (l x w) 2,200 mm x �70 mm (7' 2.�" x 2�.�")
X-ray transparent area 1,7�0 mm x �70 mm (�' 8.1" x 2�.�")
Thickness of table top �� mm (2.1")
Range of horizontal movementstransverselongitudinal
±100 mm (�.�")floating by fixing the vertical axes of the wheels
Attenuation equivalent ≤1.2 mm Al equivalent at 100 kV
Patient load 22� kg (�9� lbs)
28
All dimensions in mm (feet/inches)
Height adjustable trolley TA-M (contd.)
2270 (7' 5.4")
670
(26.
4")
100
(3.9
")10
0(3
.9")
600
(23.
6")
870
(34.
3")
890
(35"
)
54 (
2.1"
)
Image plate reader PCR Eleva S Plus
Image plate reader PCR Eleva S Plus
Type The image plate reader PCR Eleva S Plus consists of• Cassette feeding unit for the cassette• Reader unit
Plate througput Up to 97 plates/h (18 cm x 2� cm) (high-speed mode allowing up to 9� plates/h for �� cm x �� cm)
Cassette release �7 s to �� s, depending on cassette size and type
Cassette feeding unit • Removal of exposed image plate from the cassette• Reading of the image plate (cassette) barcode• Transport of the exposed image plate to and from the
reader unit• Loading the cassette with the erased plate• Release of newly loaded cassette
Reader unitScanning resolution
Read functionErase function
100 µm or 200 µm(200 µm through either software re-sizing or high-speed mode)
12 bit analog/digital conversion of informationExposure-dependent erasure of residual information on the image plate using white light
All dimensions in mm (inches)
29
590 (23.3")
810
(31.
9")
43 (
1.7"
)
380 (15.0")
DRF room solution – Configuration proposals
Configuration 1
EasyDiagnost Eleva with Bucky tray or EasyLat• Ceiling suspended tube carrier (CS2/�)• Digital vertical stand VS with/without tilt
Configuration 2
EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)
Configuration 3
EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)• Digital vertical stand VS with/without tilt
�0
Optional available*:
• CR-integration
• Height adjustable trolley TA-M
• Vertical stand VS advanced
with/without tilt
• Tracking
• Automatic image stitching
* Please contact your local sales specialist for additional configurations and accessories
�1
�2
Room planning
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control advanced (EEC A)
� Keyboard
� Mouse
� Generator cabinet (Velara)
7 EasyDiagnost Eleva main cabinet
8 DI cabinet rear cover
9 Generator cabinet rear cover
10 Main cabinet rear cover
11 Indication box (IBO)
12 Acquisition work spot
Key (contd.)
1� Infra red receiver
1� ViewForum computer
1� Mouse
1� Keyboard
17 Viewing monitor (18") (maybe combined with moitor for shared mode)
18 Footswitch
19 Monitor carriage
20 Tilting axis
21 Center of table top
22 Speed control
2� Handswitch colon map
2� Exposure handswitch
Key (contd.)
2� Area for service purposes
2� Space required for installing the X-ray module
27 Space required for inserting the table top (alternatively: 2.0� m at the head end in 0° position)
28 If necessary, order an insulating kit for ceiling rails locally
29 For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
Key (contd.)�0 Min. space �–7 mm between
2 Cabinet Rear Cover
�1 EasyDiagnost Eleva
Layout proposal for EasyDiagnost Eleva with under-table technique (single plane)
All dimensions in mm (feet/inches)
Remote serviceEthernet or Telefon connection (Ethernet preferred)
RIS/PACSEthernetconnection
29
1 7 6
30 30
8 910
19
20
18
13
21
11
28 25
31
26
27
cubicle
lavatory
24
23
22
151654
17
1214
3
2
5500 (18' 0.5")
824
(32.
6")
2080
(6'
9.9
")
5200
(17
' 0.7
")
3100 (10' 2.0")2400 (7' 10.5")
1050 (3' 5.3")
1000
(3'
3.4
")
600
(23.
6") 12
00 (
3' 1
1.2"
)
1150 (3' 9.3")
91 (3.6")280 (11.0")
240
(9.5
")
��
Room planning
Layout proposal for EasyDiagnost Eleva with over-table technique (second plane)
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control advanced (EEC A)
� Keyboard
� Mouse
� Ceiling rails ceiling suspension
7 Ceiling rails MCS
8 Generator cabinet (Velara)
9 EasyDiagnost Eleva main cabinet
10 DI cabinet rear cover
11 Monitor ceiling suspension
Key (contd.)
12 Generator cabinet rear cover
1� Main cabinet rear cover
1� Acquisition work spot
1� Indication box (IBO)
1� BuckyDiagnost VS advanced tilt
17 Infra red receiver
18 Examination monitor (18")
19 Reference monitor (18")
20 ViewForum computer
21 Mouse
22 Keyboard
Key (contd.)
2� Viewing monitor (18") (maybe combined with monitor for shared mode)
2� Speed control
2� Handswitch colom map
2� Exposure handswitch
27 Footswitch
28 Ref. cross point
29 If necessary, order an insulating kit for ceiling rails locally
�0 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision
Key (contd.)
�1 SID max. 1800 mm with table in upright position
�2 For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
�� Min. space �–7 mm between 2 Cabinet Rear Cover
�� Tube in position “free exposure”
�� EasyDiagnost Eleva
�� Depend on room size
All dimensions in mm (feet/inches)
Remote serviceEthernet or Telefon connection (Ethernet preferred)
RIS/PACSEthernetconnection
cubicle
lavatory
26
25
24
212254
23
1420
3
2
15
33
1
13
12
32
33
8
9
10
17
19
18
117
29
34
27
30
1628
31
35
6
36
5800
(19
' 0.3
")
1515
(4'
11.
6")
1500
(4'
11.
1")
≥ 6665 (21' 10.4")
≥ 3525 (11' 6.8") 3140 (10' 3.6")
835 (32.9")986 (38.8") 835 (32.9")
2200 (7' 2.6") ext. rails, max. 2700 (8' 10.3")
1000 (3' 3.4") 1000 (3' 3.4")
SID max. 1800 (5' 10.9")
80 (3.2")
4300 (14' 1.3")
479
(18.
9")
��
Key
1 DI cabinet
2 Image Receiver Module
� Eleva examination control advanced (EEC A)
� Keyboard
� Mouse
� Ceiling rails ceiling suspension
7 Ceiling rails MCS
8 CS
9 Generator cabinet (Velara)
10 EasyDiagnost Eleva main cabinet
11 DI cabinet rear cover
Room planning
Layout proposal for DRF room solution with digital vertical stand (minimum room layout without applicational restrictions)
Key (contd.)
12 Monitor ceiling suspension
1� Generator cabinet rear cover
1� Main cabinet rear cover
1� ViewForum computer
1� Indication box (IBO)
17 Acquisition work spot
18 BuckyDiagnost VS digital with motorized tilting
19 Infra red receiver
20 Examination monitor (18")
21 Reference monitor (18")
22 Mouse
2� Keyboard
Key (contd.)2� Viewing monitor (18")
(maybe combined with monitor for shared mode)
2� Footswitch
2� Speed control
27 Handswitch colon map
28 Exposure handswitch
29 If necessary, order an insulating kit for ceiling rails locally
�0 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision
�1 External rails (optional)
Key (contd.)
�2 Scopomat in work position
�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)
�� Min. space �–7 mm between 2 Cabinet Rear Cover
�� EasyDiagnost Eleva
�� Only with option stitching
All dimensions in mm (feet/inches)
Remote serviceEthernet or Telefon connection (Ethernet preferred)
RIS/PACSEthernetconnection
33
1 10 9
34 34
11 1314
28
27
26
222354
24
1715
3
2
29
16
19
21
20
7
31
32
1825
8
30
6
cubicle
lavatory
12
35
≥ 6100 (20' 0.2")
5800
(19
' 0.3
")
≥ 3430 (11' 3.0") 2670 (8' 9.1")
SID 1800 (5' 10.9")SID min. 1000 (3' 3.4")
835 (32.9")1000 (3' 3.4")1000 (3' 3.4")835 (32.9")
986 (38.8")
1580
(5'
2.2
")
1500
(4'
11.
1")
1075
(3'
6.3
")
1150
(3'
9.3
")
3636
��
General
Power supply Verticalstand VS
Eleva workspot EasyDiagnost Eleva Generator 50 kW
Mains voltage 2�0 V ±10% 11� V to 2�0 V +10%; -1�%
2�0 V ±10% �00 V ±10%; � phase
Mains frequency �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz
Current input max. 1.7 A max. 2.� A max. 7.� A max. 1�� A
Fuse �A 10 A 10 A � x �� A
Heat dissipation (average)
100 W 1�0 W �00 W - max. 1�00 W �00 W
Ambient conditions
Working temperature +18˚C to +�0˚C
Storage temperature –10˚C to +��˚C
Rel. humidity (in operation) 20% to 7�% (without condensation)
Rel. humidity (storage) 10% to 9�%
Air pressure 700 hPa to 10�0 hPA
Conformity
The EasyDiagnost Eleva system and the related room solutions comply with the provisions of the Medical Device Direktive 9�/�2 EEC (9�) and with IEC, FDA and CSA standards.
Philips Healthcare is part of
Royal Philips Electronics
How to reach us
www.philips.com/healthcare
fax: +�1 �0 27 �� 887
Asia
+8�2 2821 �888
Europe, Middle East, Africa
+�9 70�1 ��� 22��
Latin America
+�� 11 212� 07��
North America
+ 1 �2� �87 7000
800 28� ��8� (toll free, US only)
Philips Healthcare
Global Information Center
P.O. Box 128�
��02 BG Eindhoven
The Netherlands
© 2009 Koninklijke Philips Electronics N.V.All rights are reserved.