Upload
trimaptrimapcase
View
212
Download
0
Embed Size (px)
DESCRIPTION
LCMSMS
Citation preview
Results
Robust Ultra High Sensitivity Quantitative Analysis Using Low Flow LC/MS/MS and Reduced Phospholipid Matrix Effects. Myth or Reality?
David Humphries1; Roger N. Hayes1; Kevin Cook2; Mark Dreyer2; Xiang He2; Subodh Nimkar2; Patrick Bennett21MPI Research, Mattawan, MI , 2Thermo Scientific, San Jose, CA
Conclusion Micro-Flow LC provide an improvement in sensitivity which directly translated into a
gain in LLOQ peak responses of 2 to 20x in 7 out of 8 compounds tested (one outlier -Reserpine).
Micro-Flow LC did not require a change in injection volume, MS hardware changes, and more importantly there was no need to change the LC gradient when converting from the High-Flow LC method.
Ion suppression testing, while in its preliminary stages, has hinted at better signal intensity under Micro-Flow conditions while introducing plasma extracts. This can lead to reduced suppressing effects from phospholipids which will be investigated further.
Additional tests will include lower flow rates, smaller ID columns, isolating matrix components, normalizing infusion/LC flow rates and ion source comparison tests.
AcknowledgementsSpecial thanks to Laurent Rieux for his Micro-Flow expertise and allowing us to use his U3000 RSLCnano system. Also, many thanks to Jawad Pashmi for configuring the Micro-Flow system as well as his method development help. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries.
This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.
OverviewPurpose: Deliver significant sensitivity gains, reduced solvent use and less suppression for routine quantitative bioanalysis with minimal method development when compared with standard (high) flow LC methods.
Summary of Results: A simple reversed-flush trap method was developed for routine quantitation. No changes were made to the existing traditional high-flow LC gradient, eliminating the need for method development. The results were compared to traditional LC-MS/MS methods and the data supports the utility of a generic, simple to setup, robust and sensitive method using micro flow rates. For the compounds analyzed to date, the LLOQ sensitivity gain was 2-20x over traditional LC-MS/MS methods with one outlier. This is achieved in part by keeping the injection volumes consistent with the traditional LC method, e.g., if a 10L injection is utilized for a LC method, the low-flow method incorporated a 10L injection. Initial robustness testing has demonstrated excellent injection to injection reproducibility (Alprazolam-D5 CV ~7%) over 500 injections.
IntroductionThe sensitivity gains in ESI analysis using cap/micro flow LC are well documented. However, these gains are neutralized in routine quantitative bioanalysis primarily due to reduced loading capacity as well as solvent delivery pump constraints, non-reproducibility of capillary columns, lack of easy to use leak proof fittings, and the slow speed of analysis. Most importantly, the limited sample volume loaded on the column negates the sensitivity gains delivered by an ESI source operating at maximum efficiency. Here, we present data showing increased sensitivity for the quantitative bioanalytical workflow for drugs in plasma using a single valve trap LC setup with no ESI source hardware changes needed to support the high efficiency cap/micro flow that enables robust methodology using currently available instrumentation.
MethodsSample PreparationHuman plasma extracts, acetonitrile precipitated (1:3) and further diluted with water + 0.1% formic acid (1:1), were used to prepare Standard Calibrators and Quality Controls containing a mixture of eight compounds plus four internal standards. For all compounds the standard curve range was 1 pg to 10,000 pg/mL and QCs were prepared at 25 pg, 250 pg and 2500 pg/mL. In both analysis approaches, traditional phospholipid ions (m/z 496, 524, 758.6, 786.6, 806.6) were monitored to gauge potential suppression effects at low flow rates. In order to ensure broad assessment of suppression differences, the matrix samples were infused and neat study samplers were injected. Therefore, there was matrix present in the ion source continously.
LC Micro-Flow & High-FlowThe low-flow LC system was a Thermo Scientific Dionex Ultimate 3000 RSLCnano operating in Micro-Flow mode. The Micro-Flow method was run at 20 L/min for analytical separation and 80uL/min for sample loading. The analytical column used for the Micro-Flow test was a Thermo Scientific Hypersil Gold 50 x 0.5 mm, 3 m and the trap column was a Hypersil Gold 10 x 1mm, 3 m column. A single switch valve was used for this trap method. Nanoviper fittings were used to reducedead-volume.
The traditional high flow LC system wascomprised of Thermo Scientific AccelaOpen system with DLW (Dynamic Loadand Wash) coupled to an Accela1250 UHPLC pump at 500 L/min. The column used was a Hypersil Gold 100 x 2.1 mm, 1.9 m in a direct inject mode.
Mobile phases were water + 0.1% formic acid (v/v) and Acetonitrile + 0.1% formic acid (v/v). Identical gradients maintained for both assays with injection volumes of 10 L.
FIGURE 5. High-Flow representative chromatogram Standard 100 pg/mL (10 L injection, analytical flow rate of 500 L/min)Compounds are in the following order (Top to Bottom): Left Column: Oxycodone, Oxycodone-d6, Paroxetine, Paroxetine-d6, and Verapamil. Right Column: Clonazepam, Clonazepam-D6, Ketoconazole, Alprazolam, Alprazolam-d5, Reserpine, and Clopidogrel.Figure 3. Quality Control Summary Micro-Flow and Standard-Flow LC
Methods
Figure 6. Ion Suppression TestIon suppression effects were investigated under Micro-Flow and Standard-Flow LC conditions. For this experimental setup a neat Standard mixture was prepared at 100 ng/mL. Ten microliter (10L) was injected while Tee-Infusing precipitated human plasma (1:3 acetonitrile). Infusion flow rates of the extracted matrix were zero, 2 L, 5 L, and 10 L/min.
Generic ion source conditions were used for all sample collection including vaporizer temp. The micro-flow method source conditions were: spray voltage 4kV, vaporizer temp 200C, capillary temp. 325C, sheath gas 40 units, and AUX gas 5 units. The standard flow method source conditions were generic source settings of: spray voltage 4kV, vaporizer temp 475C, capillary temp. 325C, sheath gas 60, AUX gas 20, and sweep gas of 1 unit. The instrument was calibrated in positive ion mode before sample acquisition using Pierce LTQ Velos ESI Positive Ion Calibration Solution.
Data Analysis
Data was acquired using Xcalibur 2.2 with DCMSLink to control the RSLCnano front end. Data was processed using Thermo Scientifoc LCQuan 2.7 quantitation software.
FIGURE 1. Q Exactive Instrument Schematic
C-Trap
Split Lens
OrbitrapS-Lens
Capillary
HCD Collision Cell Quad Exit Lens
Quadrupole
Figure 2: LLOQ: Micro-Flow vs High-Flow
Standard calibrators and quality controls were analyzed using both micro-flow and standard flow LC methods for the same compound mixture prepared in human plasma extracts. The table below summarizes the LLOQ for each LC method using a +/- 20% accuracy cut-off, the gain achieved using micro-flow and the R^2 values.
Quality Controls were measured under Micro-Flow and Standard-Flow LC conditions. Calculated concentrations were produced using standard calibrators prepared in precipitated human plasma extracts and using LCquan 2.7 for data processing. Measured values are reported for all QCs even if they fall below the LLOQ and no QC was dropped even if accuracy was >15% from nominal. Standard accuracy was used for reporting R^2 and adopted the +/-20% cut-off for LLOQ and +/-15% cut-off for all other standards.
FIGURE 4. Micro-Flow representative chromatogram Standard 100 pg/mL (10 L injection, analytical flow rate of 40 L/min)Compounds are in the following order (Top to Bottom): Left Column: Oxycodone, Oxycodone-d6, Paroxetine, Paroxetine-d6, and Verapamil. Right Column: Clonazepam, Clonazepam-D6, Ketoconazole, Alprazolam, Alprazolam-d5, Reserpine, and Clopidogrel.
2 3 4 5Time (min)
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
RT: 2.10AA: 172900
RT: 2.09AA: 1906578
RT: 3.55AA: 1110840
RT: 3.57AA: 10716424
RT: 4.02AA: 1971460
NL: 4.91E4m/z= 316.15275-316.15591 F: FTMS + p ESI SIM msx ms [315.00-317.00, 321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 5.24E5m/z= 322.19039-322.19361 F: FTMS + p ESI SIM msx ms [315.00-317.00, 321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.57E5m/z= 330.14835-330.15165 F: FTMS + p ESI SIM msx ms [329.00-331.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.42E6m/z= 336.18597-336.18933 F: FTMS + p ESI SIM msx ms [335.00-337.00] MS ICIS CpdMix-Micro-Flow-013
NL: 2.37E5m/z= 455.28815-455.29271 F: FTMS + p ESI SIM msx ms [454.00-456.00] MS ICIS CpdMix-Micro-Flow-013
2 3 4 5 6Time (min)
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
RT: 2.95AA: 388436
RT: 2.93AA: 1252733
RT: 3.72AA: 311925
RT: 3.20AA: 703982
RT: 3.17AA: 6487955
RT: 4.77AA: 1258930
RT: 3.93AA: 748326
NL: 4.88E4m/z= 316.04677-316.04993 F: FTMS + p ESI SIM msx ms [315.00-317.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.70E5m/z= 322.06826-322.07148 F: FTMS + p ESI SIM msx ms [321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 3.75E4m/z= 531.15338-531.15870 F: FTMS + p ESI SIM msx ms [530.00-532.00] MS ICIS CpdMix-Micro-Flow-013
NL: 8.82E4m/z= 309.08860-309.09170 F: FTMS + p ESI SIM msx ms [308.00-310.00] MS ICIS CpdMix-Micro-Flow-013
NL: 8.77E5m/z= 314.11996-314.12310 F: FTMS + p ESI SIM msx ms [313.00-315.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.37E5m/z= 609.27761-609.28371 F: FTMS + p ESI SIM msx ms [608.00-610.00] MS ICIS CpdMix-Micro-Flow-013
NL: 6.61E4m/z= 322.06469-322.06791 F: FTMS + p ESI SIM msx ms [321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
1.0 1.5 2.0 2.5 3.0 3.5 4.0Time (min)
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
RT: 1.83AA: 381845
RT: 1.82AA: 2892958
RT: 3.05AA: 453247
2.79RT: 3.05AA: 2866835
RT: 3.26AA: 894462
NL: 1.86E5m/z= 316.15275-316.15591 F: FTMS + p ESI SIM msx ms [315.00-317.00, 321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.31E6m/z= 322.19039-322.19361 F: FTMS + p ESI SIM msx ms [315.00-317.00, 321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.26E5m/z= 330.14835-330.15165 F: FTMS + p ESI SIM msx ms [329.00-331.00] MS ICIS CpdMix-Micro-Flow-013
NL: 8.14E5m/z= 336.18597-336.18933 F: FTMS + p ESI SIM msx ms [335.00-337.00] MS ICIS CpdMix-Micro-Flow-013
NL: 2.73E5m/z= 455.28815-455.29271 F: FTMS + p ESI SIM msx ms [454.00-456.00] MS ICIS CpdMix-Micro-Flow-013
2.5 3.0 3.5 4.0Time (min)
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
0
50
100
RT: 3.08AA: 62697
RT: 3.06AA: 216139
RT: 3.07AA: 194289
RT: 3.20AA: 283246
RT: 3.18AA: 2487752
RT: 3.44AA: 314966
RT: 3.70AA: 491345
NL: 2.02E4m/z= 316.04677-316.04993 F: FTMS + p ESI SIM msx ms [315.00-317.00] MS ICIS CpdMix-Micro-Flow-013
NL: 7.19E4m/z= 322.06826-322.07148 F: FTMS + p ESI SIM msx ms [321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
NL: 5.60E4m/z= 531.15338-531.15870 F: FTMS + p ESI SIM msx ms [530.00-532.00] MS ICIS CpdMix-Micro-Flow-013
NL: 7.97E4m/z= 309.08860-309.09170 F: FTMS + p ESI SIM msx ms [308.00-310.00] MS ICIS CpdMix-Micro-Flow-013
NL: 6.94E5m/z= 314.11996-314.12310 F: FTMS + p ESI SIM msx ms [313.00-315.00] MS ICIS CpdMix-Micro-Flow-013
NL: 1.39E5m/z= 609.27761-609.28371 F: FTMS + p ESI SIM msx ms [608.00-610.00] MS ICIS CpdMix-Micro-Flow-013
NL: 2.04E5m/z= 322.06469-322.06791 F: FTMS + p ESI SIM msx ms [321.00-323.00] MS ICIS CpdMix-Micro-Flow-013
The LLOQ for each analyte was determined by using a +/- 20% cut-off for accuracy. In addition, out of the four replicates, >50% of the standards at the LLOQ must meet this criteria. The gain is not based on pure signal. As the summary table indicates, the improvement seen using the Micro-Flow method ranged from 2-20x. Reserpine (-2x), was the single outlier. The reason for this observation is under investigation.
The linear gradient utilized for the traditional LC Standard-Flow method was directly transferred to the Micro-Flow LC method with an additional 1-minute hold to compensate for flushing the trap column. The Standard-Flow method run time was increased by one minute to compensate for the flush step.
LC Methods (L) Micro-Flow and (R) Standard Flow
Mass Spectrometry
A Thermo Scientific Q-Exactive bench-top high resolution accurate mass Orbitrap mass spectrometer was used in full scan mode to monitor phospholipids and in targeted SIM mode for routine quantitation at resolutions of 70,000 or 140,000 (based on m/z 200) yielding spectral speeds of 3 and 1.5 Hz, respectively.
The graphs below display the peak area intensity difference for representative analyteswhile infusing precipitated human plasma extracts. Peak areas were compared under both Micro-Flow and Standard-Flow. Besides one outlier, Oxycodone, peak intensity was higher under Micro-Flow LC conditions.
1.50E+083.50E+085.50E+087.50E+089.50E+081.15E+091.35E+091.55E+091.75E+091.95E+09
No Infusion Infusion 2ul Infusion 5ul Infusion 10ul
P
a
r
o
x
e
t
i
n
e
P
e
a
k
A
r
e
a
Paroxetine Ion Suppression Test
Micro-Flow High-Flow
0.00E+00
2.00E+08
4.00E+08
6.00E+08
8.00E+08
1.00E+09
No Infusion Infusion 2ul Infusion 5ul Infusion 10ul
C
l
o
n
a
z
e
p
a
m
P
e
a
k
A
r
e
a
Clonazepam Ion Suppression Test
Micro-Flow High-Flow
0.00E+001.00E+082.00E+083.00E+084.00E+085.00E+086.00E+087.00E+088.00E+08
No Infusion Infusion 2ul Infusion 5ul Infusion 10ul
A
l
p
r
a
z
o
a
l
m
P
e
a
k
A
r
e
a
Alprazolam Ion Suppression Test
Micro-Flow High-Flow
7.50E+07
1.25E+08
1.75E+08
2.25E+08
2.75E+08
3.25E+08
3.75E+08
No Infusion Infusion 2ul Infusion 5ul Infusion 10ul
O
x
y
c
o
d
o
n
e
P
e
a
k
A
r
e
a
Oxycodone Ion Suppression Test
Micro-Flow High-Flow
Approximately 90% solvent reduction over High-Flow.
Analyte Flow RateLLOQ
(pg/mL)
Gain (Micro/
High)R^2
Alprazolam Micro 5 0.9940High 10 0.9909
Clonazepam Micro 5 0.9943High 100 0.9889
Clopidogrel Micro 5 0.9905High 50 0.9904
Ketoconazole Micro 5 0.9914High 10 0.9963
Oxycodone Micro 5 0.9907High 50 0.9908
Paroxetine Micro 5 0.9947High 10 0.9939
Reserpine Micro 10 0.9896High 5 0.9919
Verapamil Micro 5 0.9924High 10 0.9888
* all linear weighting, 1/X^2
2x
20x
10x
2x
10x
2x
-2x
2x
Component Name
LevelAvg Calc
ConcAvg % Diff % CV
Component Name
LevelAvg Calc
ConcAvg %
Diff% CV
Alprazolam QC-L 25.5 1.88 2.06 Alprazolam QC-L 24.9 -0.355 4.79QC-M 254 1.54 4.75 QC-M 272 8.65 2.55QC-H 2309 -7.65 2.26 QC-H 2005 -19.8 4.24
Clonazepam QC-L 23.8 -4.88 4.03 Clonazepam QC-L 33.8 35.2 23.4QC-M 247 -1.29 4.38 QC-M 246 -1.42 12.0QC-H 2156 -13.8 1.92 QC-H 2026 -19.0 12.0
Clopidogrel QC-L 26.3 5.13 8.45 Clopidogrel QC-L 5.43 -78.3 19.6QC-M 260 4.16 4.19 QC-M 206 -17.8 14.6QC-H 2181 -12.8 3.40 QC-H 2629 5.15 2.10
Ketoconazole QC-L 23.9 -4.43 6.18 Ketoconazole QC-L 17.7 -29.0 21.7QC-M 259 3.45 4.51 QC-M 231 -7.65 7.98QC-H 2288 -8.48 4.78 QC-H 1967 -21.3 2.21
Oxycodone QC-L 22.5 -10.0 4.05 Oxycodone QC-L 33.9 35.7 1.96QC-M 223 -10.9 3.43 QC-M 247 -1.26 1.86QC-H 2228 -10.9 0.523 QC-H 2020 -19.2 1.29
Paroxetine QC-L 22.3 -10.8 3.28 Paroxetine QC-L 28.4 13.6 4.07QC-M 252 0.929 2.23 QC-M 257 2.62 10.9QC-H 2216 -11.4 1.55 QC-H 2209 -11.6 11.9
Reserpine QC-L 24.0 -4.09 2.93 Reserpine QC-L 281 12.5 6.46QC-M 261 4.21 0.757 QC-M 2122 -15.1 2.58QC-H 2027 -19.0 0.965 QC-H 26.4 5.50 8.58
Verapamil QC-L 26.7 6.59 4.79 Verapamil QC-L 271 8.28 10.7QC-M 241 -3.60 5.60 QC-M 1990 -20.4 3.81QC-H 2055 -17.8 7.57 QC-H 27.0 8.02 11.7
Micro-Flow High-Flow
Analyte Flow BlankNo Infusion of
matrixInfusion of matrix 2ul
Infusion of matrix 5ul
Infusion of matrix 10ul
Oxycodone Micro 0 280,512,996 166,390,775 135,375,600 127,650,957High 0 330,964,625 290,799,261 259,132,534 329,191,343
Paroxetine Micro 0 1,820,691,827 1,372,468,222 1,119,092,095 984,310,228High 0 427,434,318 349,660,603 276,519,523 384,534,017
Verapamil Micro 0 947,430,503 876,700,676 870,365,398 757,313,379High 0 434,966,926 368,107,221 406,770,700 419,858,982
Clopidogrel Micro 0 655,984,091 541,468,656 515,529,087 491,253,463High 0 361,309,888 330,467,243 300,048,640 339,991,504
Clonazepam Micro 0 442,219,373 332,725,726 155,041,498 118,430,135High 0 63,854,904 56,053,157 46,440,915 57,567,712
Alprazolam Micro 0 703,177,916 568,124,244 360,452,289 287,811,884High 0 146,584,063 131,280,208 104,252,912 125,637,373
Ketoconazole Micro 0 82,806,775 88,712,167 95,468,826 90,842,707High 0 43,579,301 43,630,254 51,355,451 45,283,041