Leica RM2235Rotary Microtome

Instructions for useLeica RM2235V1.6 Rev A, English – 09/2011Order No. 14 0500 80101Always keep this manual with the instrument.Read carefully before working with the instrument.

Page 2: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to
Page 3: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

3Leica RM2235

The information, numerical data, notes and val-ue judgments contained in this manual repre-sent the current state of scientific knowledgeand state-of-the-art technology as we under-stand it following thorough investigation in thisfield.We are under no obligation to update thepresent manual periodically and on an ongoingbasis according to the latest technical develop-ments, nor to provide our customers with addi-tional copies, updates etc. of this manual.For erroneous statements, drawings, technicalillustrations etc. contained in this manual weexclude liability as far as permissible accordingto the national legal system applicable in eachindividual case.In particular, no liability whatsoever is acceptedfor any financial loss or consequential damagecaused by or related to compliance with state-ments or other information in this manual.Statements, drawings, illustrations and otherinformation as regards contents or technicaldetails of the present manual are not to be con-sidered as warranted characteristics of ourproducts.

These are determined only by the contractprovisions agreed between ourselves and ourcustomers.Leica reserves the right to change technicalspecifications as well as manufacturing pro-cesses without prior notice. Only in this way is itpossible continuously to improve the technolo-gy and manufacturing techniques used in ourproducts.This document is protected under copyrightlaws. Any copyrights of this document are re-tained by Leica Biosystems Nussloch GmbH.Any reproduction of text and illustrations (or ofany parts thereof) by means of print, photocopy,microfiche, web cam or other methods – includ-ing any electronic systems and media –requires express prior permission in writing byLeica Biosystems Nussloch GmbH.For the instrument serial number and year ofmanufacture, please refer to the name plate atthe back of the instrument.

© Leica Biosystems Nussloch GmbH

NOTE

Published by:Leica Biosystems Nussloch GmbHHeidelberger Str. 17 - 19D-69226 NusslochGermanyPhone: +49 (0) 6224 143-0Fax: +49 (0) 62 24 143-268Internet: http://www.leica-microsystems.com

Page 4: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

4 Instructions for use V 1.6 Rev A – 09/2011

Table of contents

1. Important notes ........................................................................................................................... 62. Safety ............................................................................................................................................ 7

2.1 Safety notes ............................................................................................................................... 72.2 Warnings ...................................................................................................................................... 72.3 Integrated safety devices ...................................................................................................... 10

3. Instrument components and specifications ........................................................................ 123.1 Overview – instrument components ................................................................................... 123.2 Instrument specifications ...................................................................................................... 133.3 Technical data ......................................................................................................................... 14

4. Startup ........................................................................................................................................ 154.1 Standard delivery ..................................................................................................................... 154.2 Installation site requirements ............................................................................................... 154.3 Unpacking and installation .................................................................................................... 164.4 Assembling the handwheel................................................................................................... 184.5 Inserting the universal cassette clamp .............................................................................. 194.6 Inserting the knife holder ....................................................................................................... 20

5. Operation.................................................................................................................................... 215.1 Operating elements and their functions ............................................................................ 215.1.1 Section thickness setting ...................................................................................................... 215.1.2 Coarse driving wheel .............................................................................................................. 215.1.4 Mechanical trimming function .............................................................................................. 225.1.3 Specimen retraction ............................................................................................................... 225.1.5 Specimen holder with precision orientation ..................................................................... 235.2 Adjusting the clearance angle ............................................................................................. 245.3 Fine adjustment of the force balance ................................................................................. 255.4 Clamping the specimen .......................................................................................................... 265.5 Clamping the knife / disposable blade ................................................................................ 265.6 Sectioning .................................................................................................................................. 275.7 Changing the specimen or interrupting sectioning ......................................................... 285.8 Finishing daily routine ............................................................................................................. 28

6. Optional accessories ............................................................................................................... 296.1 Assembly for fixture for specimen clamps ........................................................................ 296.1.1 Rigid fixture for specimen clamps ....................................................................................... 296.1.2 Directional fixture for specimen clamps ............................................................................ 296.1.3 Fine-directional fixture for specimen clamps ................................................................... 30

Page 5: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

5Leica RM2235

Table of contents

6.1.4 Quick clamping system .......................................................................................................... 316.2 Specimen clamps and holders ............................................................................................. 326.2.1 Standard specimen clamp .................................................................................................... 326.2.2 Vee insert ................................................................................................................................... 336.2.3 Foil clamp type 1 ....................................................................................................................... 346.2.4 Foil clamp type 2 ....................................................................................................................... 356.2.5 Universal cassette clamp ...................................................................................................... 36

Universal cassette clamp, ice-cooled ................................................................................ 376.2.6 Super mega-cassette clamp ................................................................................................ 376.2.7 Holder for round specimens ................................................................................................. 396.3 Knife holder base and knife holder ..................................................................................... 406.3.1 Knife holder base, without lateral movement feature ..................................................... 406.3.2 Knife holder E/E-TC .................................................................................................................. 416.3.3 Knife holder N/NZ..................................................................................................................... 446.4 Section waste tray ................................................................................................................... 466.5 Backlighting .............................................................................................................................. 466.6 Tray ............................................................................................................................................. 476.7 Freezer pack ............................................................................................................................. 476.8 Universal microscope carrier ............................................................................................... 486.9 Magnifying lens ........................................................................................................................ 506.10 Additional accessories .......................................................................................................... 516.11 Ordering information ............................................................................................................... 68

7. Customized solutions ............................................................................................................... 707.1 Possible faults ........................................................................................................................... 707.2 Instrument malfunctions ........................................................................................................ 71

8. Cleaning and maintenance .................................................................................................... 728.1 Cleaning the instrument ......................................................................................................... 728.2 Maintenance instructions ..................................................................................................... 748.3 Lubricating the instrument .................................................................................................... 75

9. Warranty and service .............................................................................................................. 76

Page 6: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

6 Instructions for use V 1.6 Rev A – 09/2011

Notes, i.e. important user informationappear in a gray box and are marked

by an information symbol .

Dangers, warnings and cautions ap-pear in a gray box and are marked bya warning triangle .

Symbols in the text and their meanings

1. Important notes

(5)

Intended use of instrument

The Leica RM2235 is a manually operated ro-tary microtome for creating thin sections ofspecimens of varying hardness for use in rou-tine and research laboratories in the fields of bi-ology, medicine and industry.It is designed for sectioning soft paraffin-em-bedded specimens as well as harder speci-mens, as long as they are suitable for beingcut manually.

Any other use of the instrument is consideredimproper!

Instrument type

All information provided in this manual appliesonly to the instrument type indicated on thecover page.An identification label with the serial number isfastened to the left side of the instrument (thisfigure is only symbolic).

Fig. 1

Qualification of personnel

• The Leica RM2235 may be operated bytrained laboratory personnel only.

• All laboratory personnel designated to oper-ate the Leica instrument must read these

Numbers in parentheses refer to itemnumbers in illustrations.

REF

SN

Manufacturer

This product fulfills the requirementsof the Council's Directive 98/79/ECconcerning in vitro diagnostics (IVD)medical devices.

In vitro diagnostics (IVD) medical de-vice

Observe the Instructions for Use

Order No.

Serial number

instructions for use carefully and must befamiliar with all technical features of theinstrument before attempting to operate it.

Page 7: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

7Leica RM2235

Be sure to comply with the safety instructions and warnings provided in this chapter.Be sure to read these instructions, even if you are already familiar with the operation and useof other Leica products.

2. Safety

2.1 Safety notes

This instructions for use contains importantinstructions and information regarding the op-erational safety and maintenance of the instru-ment.The instructions for use is an important part ofthe product, which must be read carefully priorto startup and use and must always be keptnear the instrument.

If additional requirements on accidentprevention and environmental protec-tion exist in the country of operation,this instructions for use must be sup-plemented by appropriate instructionsto ensure compliance with suchrequirements.

This instrument is built and inspected accordingto the Safety requirements for laboratory instru-ments.For current information about applicable stan-dards, please refer to the CE declaration of con-formity on our Internet site:

www.leica-microsystems.com

In order to maintain this condition and ensuresafe operation, the operator must observe allthe instructions and warnings contained in thisinstructions for use.

The protective devices on both instrument and accessories may neither be removed nor modi-fied. Only service personnel qualified by Leica may repair the instrument and access the in-strument’s internal components.

2.2 Warnings

The safety devices installed in this instrument by the manufacturer only constitute the basisfor accident prevention. Primarily responsible for accident-free operation is above all the own-er of the instrument and, in addition, the designated personnel who operates, services orcleans the instrument.To ensure trouble-free operation of the instrument, make sure to comply with the followinginstructions and warnings.

Page 8: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

8 Instructions for use V 1.6 Rev A – 09/2011

Warnings – Working at the instrument

• Take care when handling microtome knives and disposable blades. The cutting edge isextremely sharp and can cause severe injury!

• Always remove the knife / blade before detaching the knife holder from the instrument. Alwaysput the knives back into the knife case when not in use!

• Never place a knife anywhere with the cutting edge facing upwards and never try to catch afalling knife!

• Always clamp the specimen block BEFORE clamping the knife.

Warnings – Safety instructions / warning labels attached to the instrument

• Safety instruction labels on the instrument marked with a warning triangle indicate that thecorrect operating instructions (as described in this manual) must be followed when operatingor replacing the instrument component bearing the label.

Failure to adhere to these instructions may result in an accident, personal injury, damage tothe instrument or accessory equipment.

Warnings – Transport and installation

• Once removed from the crate, the instrument may only be transported in an upright position.

• Do not transport the instrument by holding it by the handwheel grips, coarse driving wheel orthe knob for setting the section thickness.

• The protective devices on both instrument and accessories must neither be removed nor mod-ified.

2. Safety

Page 9: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

9Leica RM2235

Warnings – Working at the instrument

• Prior to manipulating the knife and specimen, or changing the specimen or knife, and duringbreaks, always lock the handwheel and cover the cutting edge with the knife guard!

• ALWAYS turn the handwheel clockwise; otherwise, the brake will not work properly.

• Always wear protective glasses when sectioning brittle specimens. Specimens may splinter!

• Specimen blocks must not be oriented during the retraction phase! If a block is oriented duringretraction, the block will advance by the retraction value PLUS the selected section thicknessbefore the next section. This may cause damage to both specimen and knife!

• Prior to sectioning, check that the specimen is securely clamped in the specimen clamp –failure to observe this poses the risk of damaging the specimen.

Warnings – Cleaning and maintenance

• Always lock the handwheel before cleaning!

• Do not use any solvents containing acetone or xylene for cleaning!

• Ensure that no liquids enter the interior of the instrument when cleaning!

• When using detergents please comply with the safety precautions of the manufacturer.

2. Safety

Page 10: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

10 Instructions for use V 1.6 Rev A – 09/2011

• To lock the handwheel, press the lever (5)outwards and continue to turn the hand-wheel slowly until it locks exactly in the12 o'clock position.

When using both brake systems at thesame time, always move the lever (3)to position first.Otherwise, it may be not be possibleto release the lever (5).

2. Safety

2.3 Integrated safety devicesLocking the handwheelThere are two ways of locking the handwheel(12):Using the lever (3) on the right side of the micro-tome base plate, the handwheel can be brakedin almost any position.• To brake, rotate the lever in a counterclock-

wise direction to position .

Caution!The braking lever (3) must be exactlyin position , so that the handwheelbrake is applied correctly.If the lever is moved beyond this point,it is possible that the handwheel is nolonger braked.

• To unlock the handwheel brake turn thelever (3) back to its original position.Position .

Fig. 2

3

5

12

Page 11: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

11Leica RM2235

2. Safety

Knife guard on the knife holder

Each knife holder is equipped with a tightlymounted knife guard (8, 9). This makes it possi-ble to cover completely the cutting edge inevery knife or blade position.

Knife holder N/NZ

The knife guard (8) of the knife holder N/NZ canbe easily positioned via the two handles (7)(Fig. 3).To cover the knife edge, push both cover stripsof the knife guard to the center.

Knife holder E/E-TC

The knife guard on knife holder E/E-TC consistsof a red foldaway handle (9). To cover thecutting edge, fold the knife guard handle (9)upwards as illustrated in Fig. 4.

Fig. 4

Fig. 3

7

78

9 1011

The clamping levers on the knife hold-er E are not interchangeable.The two clamping levers (10, 11) mustremain in the position shown at alltimes, as otherwise isolated malfunc-tions of the knife holder can occur.Clamping lever for the blade (10) atthe right, clamping lever for the later-al displacement (11) at the left.

9

Knife holder E-TC

Knife holder E

Knife holder N

Page 12: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

12 Instructions for use V 1.6 Rev A – 09/2011

3.1 Overview – instrument components

3. Instrument components and specifications

Adjusting knobfor setting the

sectionthickness

Window fordisplaying the

sectionthickness

Tray

Removable sectionwaste tray

Smooth-turning

handwheel

Lever foractivating the

handwheel brake

Handwheellocking

mechanism

Lever foractivating the

mechanicaltrimming function

Directionalfixture for specimen

clamps

Coarse drivingwheel

Fig. 5

Knife holder E

Knife holder base,without lateralmovement feature

Clamping lever forlateral movementfunction of knifeholder

Universalcassette clamp

Clamping leverof the knife

holder base

Page 13: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

13Leica RM2235

3. Instrument components and specifications

3.2 Instrument specifications

Basic instrument with mechanical trim function, lateral coarse driveThe Leica RM2235 rotary microtome is equipped with a low-mainte-nance, slack-free micrometer drive, with vertical and horizontal speci-men feed realized via low-maintenance cross roller bearings.The instrument is equipped with two independent handwheel lockingsystems for even greater operating safety.Leica's patented, user-adjustable force balance system compensatescentrifugal forces arising while cutting via a pretensioned spring forextremely light handwheel action.Benefit: a heavy counterweight in the handwheel is no longer needed. Thespring tension is individually adjustable, corresponding to the weight ofthe respective attached specimen clamp or specimen.The patented specimen retraction system can be switched on and off bythe user. The instrument thus offers all of the advantages of specimenretraction while supporting work in "rocking mode", i.e. without a fullhandwheel rotation.We recommend disabling specimen retraction when working in "rockingmode".The coarse drive wheel is ergonomically positioned.(For more information on the rotating direction, see chapter 5.1.2)

Page 14: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

14 Instructions for use V 1.6 Rev A – 09/2011

3.3 Technical data

GeneralApprovals: The instrument-specific marks are located on the

rear panel of the instrument next to the name plate.Operating temperature range: +10 °C to +35 °CTemperature range during storage: +5 °C to +55 °CRelative humidity: max. 80 % non-condensingHumidity during storage: < 80 %Section thickness range: 1.0 - 60.0 µm

Section thickness settings: from 1.0 - 10.0 µm in 1.0 µm incrementsfrom 10.0 - 20.0 µm in 2.0 µm incrementsfrom 20.0 - 60.0 µm in 5.0 µm increments

Specimen feed: approx. 24 mm, ± 2 mmVertical stroke: 70 mm Max. sectioning range without retraction: 69 mm (without specimen orientation at 1 µm) Max. sectioning range with retraction: 62 mmThe specimen retraction can be turnedoff manually: approx. 40 µm

Dimensions and weightWidth (including handwheel): 413 mmDepth (including waste tray): 618 mmHeight (total): 305 mm (with tray on the hood)Working height (knife blade): 168 mm (measured from the table)

Weight (without accessories): approx. 37 kg

Optional equipment and optional accessoriesObject orientation (option)horizontal: 8°vertical: 8 °Rotating range: ± 90°

Trimming increments: 10 µm, 50 µm

Repositioning of knife holder basenorth-south: ± 25 mm

3. Instrument components and specifications

Page 15: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

15Leica RM2235

4. Startup

4.1 Standard deliveryThe Leica RM2235 standard delivery includes:

1 Leica RM2235 basic instrument1 handwheel, complete .............................................................. 14 0500 381811 section waste tray .................................................................... 14 0502 379311 tool set – consisting of: ............................................................ 14 0500 38600

1 Allen key with handle, size 5 ............................................. 14 0194 047601 Allen key with handle, size 4 ............................................. 14 0194 047821 Allen key, size 3 .................................................................... 14 0222 041381 screwdriver 3x50, 186 long ............................................... 14 0170 115681 bottle (50 ml) of oil for drives, type 405 ........................... 14 0336 060861 brush w/magnet .................................................................. 14 0183 40426

1 dust protective cover .............................................................. 14 0212 303501 operating manual D/E + language CD.................................. 14 0500 800011 CE Declaration of Conformity (engl.) ..................................... .. 14 0500 80011

4.2 Installation site requirements• Stable, vibration-free laboratory bench with

horizontal and even stage plate; practicallyvibration-free floor.

• No other instruments nearby which mightcause vibrations.

• Room temperature permanently between+10 °C and +35 °C.

• Obstruction-free access to the handwheel.

Never operate the instrument in roomswith explosion hazard.

The accessories ordered are includedin a separate box.Carefully check the delivery againstthe packing list and the delivery note.Should there be any discrepancy,please contact the Leica selling unithandling your order.

Page 16: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

16 Instructions for use V 1.6 Rev A – 09/2011

4.3 Unpacking and installation

When the instrument is delivered, check the tilt indicators on thepackaging.If the arrowhead is blue, the shipment was transported laying flat,was tilted at too great an angle or fell over during transport.Note this on the shipping documents and check the shipment forpossible damage.

Fig. 6

Fig. 7

11111

22222

33333

44444

44444

44444

44444

55555

• Loosen and unscrew the six upperscrews (2).

• Remove the cover (1).• Take the accessory carton (optional

accessories) (3) and the cartons (4)from the standard scope of delivery.

The transport crate and includ-ed retaining elements shouldbe kept in case a return ship-ment is necessary later.

4. Startup

Page 17: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

17Leica RM2235

6666655555

4.3 Unpacking and installation (cont.)

Fig. 8

77777

88888

9

• Take out the fixing module (5). To do so, holdit by the top edge of the module and in therecessed grip (6) and pull it out by pullingupwards.

• Lift the instrument* (7) by holding it by thebase plate and under the instrument on theback and lift it out of the formed cushion(8).

Never hold the instrument fortransport by the handwheel or therotary knob for section thicknessadjustment.

Observe the correct resting angleto the table to avoid pinchingyour fingers.

• Place the instrument on a stable laboratorytable.The two sliding elements (9) located on therear of the base plate make it easier tomove the instrument on the table.

• To move the instrument, hold it by the frontof the base plate, lift it up gently and slideit on its slides.

4. Startup

• (*= the instrument illustration is providedas an example only)

Page 18: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

18 Instructions for use V 1.6 Rev A – 09/2011

4. Startup

The feather key (4) is looselyplaced in the handwheel shaft (1)and fixed in place with a tie-rapduring transport.

• Remove the cable tie (3).Caution!Make sure not to lose thefeather key!

• Place the handwheel (2) on thehandwheel shaft (1) as shown.

• Tighten the screw (2a) locatedin the center hole of the hand-wheel with an Allen key size 4(5).

• Remove the cover foil from theself-adhesive cover disk andfix the cover disk on the hand-wheel.

4.4 Assembling the handwheel

Fig. 9

The handwheel has to be assembled before attemptingto use the instrument.The necessary parts and tools can be found in the tool-kit supplied in the delivery.

1

3

2

2a

5

4

Page 19: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

19Leica RM2235

4. Startup

There are two versions of the specimen holder,one with and one without specimen orientation,which are interchangeable.The object orientation allows for simple positioncorrection of the specimen surface when thespecimen is clamped into place.

You can use the quick clamping system (64) tohold all available accessory specimen clamps(for more information, see Chapter 6 "Optionalaccessories").

To do so, proceed as follows:

• Move the object head (60) to the upper endposition by turning the handwheel (12) andengage the handwheel lock.

• To release the clamping system, turn thescrew (61) of the quick clamping system (64)counterclockwise using an Allen key size 4(71).

• Push the guide (63) of the universal cassetteclamp (62) from the left into the quick clamp-ing system (64) as far as it will go.

• To clamp the cassette clamp turn the screw(61) clockwise as far as it will go.

Fig. 10

4.5 Inserting the universal cassette clamp

71

63

62

60

12

61

64

60

Since all stage clamps available as accessories are equipped with the same kind of guide onthe back, they are inserted in the same way described here using the example of the cassetteclamp.

Page 20: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

20 Instructions for use V 1.6 Rev A – 09/2011

4. Startup

Setting up the knife holder base

• Release the clamping lever (50) by rotating itcounterclockwise.

• Insert the knife holder base (51) with thenotch (52) on the bottom into the T-piece (55)of the microtome base plate (53).

• To secure the knife holder base, turn theclamping lever (50) clockwise.

The knife holder base (51) can be moved backand forth on the microtome base plate. Thisallows bringing the knife holder to optimal sec-tioning position in relation to the specimen.There is a scale (54) on the right side of the mi-crotome base plate. This enables faster andbetter positioning of the knife holder at thespecimen if various combinations of standardspecimens and specimen holders are used. Therear edge of the knife holder base (51) functionsas the scale reference.

Fig. 11

55

50

51

52

53

4.6 Inserting the knife holder

Inserting the knife holder• Loosen the screw (58) using an Allen key

size 4 (71) until the knife holder (57) can bemoved.

• Place the knife holder (57) with the under-side groove onto the T-piece (56) of the knifeholder base (51).

• To clamp, retighten the screw (58).

Fig. 12

56

57

58

Enlarged detail:Scale for betterrepositioning of theknife holder forvarying specimenheights. 54

71

Page 21: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

21Leica RM2235

5. Operation

5.1 Operating elements and their functions

Fig. 13

The section thickness is set by turning the ad-justing knob (33) at the front of the microtomeon the right.The adjusting knob has a notch for each valuethat can be set.Setting range: 1 - 60 µmfrom 1 - 10 µm in 1 µm incrementsfrom10 - 20 µm in 2 µm incrementsfrom20 - 60 µm in 5 µm increments.

The section thickness set in each case isdisplayed in the window (34).The selected section thickness (on the scale)must agree with the red pointer (38)

5.1.1 Section thickness setting

33

38

34

Fig. 14

5.1.2 Coarse driving wheel

The instrument can be ordered with clock-wise or counterclockwise rotation. The givendirection of rotation means "forwards" and re-lates to the feed movement of the specimentowards the knife.The coarse motion serves for a fast horizontalforwards movement of the object - towards theknife - and backwards - away from the knife.When reaching the rear/front end positions, thecoarse driving wheel can only be turned withdifficulty. In the front end position, no more feedmotion takes place.

The coarse driving wheel also turnsduring sectioning. Therefore it mustnot be blocked while the handwheelis being turned during sectioning;otherwise, no feed motion can takeplace and thus also no sectioning.

Page 22: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

22 Instructions for use V 1.6 Rev A – 09/2011

5. Operation

5.1.3 Specimen retraction

The specimen retraction serves for protectingthe knife and the specimen. When the retrac-tion is switched on, the object is drawn back200 µm into the upper end position after thesectioning stroke during the return movement.Before the feed motion of the new sectionthickness, the feed motion for the retractionvalue takes place.The specimen retraction can be switched offmanually at the back of the instrument (Fig. 15),if required, using the slotted-screwdriver sup-plied with the delivery.Before switching the specimen retraction Onand Off, run the object head to the upper endposition by turning the handwheel.

5.1.4 Mechanical trimming function

The RM2235 is fitted with a mechanical trimming function. Thetrimming lever has 3 notching positions (0, 10 µm, 30 µm).The points (36) mark the two trimming stages: = 10 µm = 30 µm

• For activating the trimming function, press the lever downwardsinto one of the two notching positions and keep depressed. Aftereach rotation of the handwheel, a feed motion of 10 µm or 30 µmtakes place.

• After letting go of the lever, it automatically springs back to itsoriginal position (zero position). The trimming function is therebydeactivated.

The section thickness that has been set is not added to theselected trimming value.If the section thickness that has been set is greater than theselected trimming value, the section thickness is fed.

Magnet forholding thewrenchsize 4

For turning the retractionOff, turn the slot into thehorizontal position (off).

For turning the retractionOn, turn the slot into thevertical position (on).

Fig. 15

Fig. 16

36

Page 23: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

23Leica RM2235

5. Operation

In the quick clamping device of the specimen holder fixture with precision orientation, allobject clamps available as optional accessories can be used (implemented).

5.1.5 Specimen holder with precision orientation

Orienting the specimen

Specimen blocks must not be orient-ed during the retraction phase!If a block is oriented during retrac-tion, the block will advance by the re-traction value PLUS the selected sec-tion thickness before the next section.This may cause damage to both speci-men and knife!

• Raise the object head to the upper end posi-tion and activate the handwheel lock.

• To release the clamp, turn the eccentriclever (29) forwards.

• Turn setscrew (30) to orient the specimen innorth-south direction. Turn setscrew (31) toorient the specimen in east-west direction.Each complete turn of the screw inclines thespecimen by 2°. A total of 4 complete turns =8° are possible in every direction. The accu-racy is approximately ± 0.5°.For ease of estimation, there is a whitemarking on the handle and a click stopthat is noticeable during turning.

• To lock the current orientation, turn theeccentric lever (29) backwards.

Fig. 17

When the large standard specimenclamp (50 x 55 mm) is used, the speci-men orientation of 8° in north-southdirection is no longer possible.The usable angle is only about 4° inthis case.

Display of the zero positionFor better display of the zero position, the ori-entation has two red indicators (32).When both indicators are visible and bothsetscrews are in zero position at the sametime (click stop, white marking on " "), thespecimen is in zero position.

The object orientation allows for simple positioncorrection of the specimen surface when thespecimen is clamped into place.32

30 32

31

29

Page 24: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

24 Instructions for use V 1.6 Rev A – 09/2011

5. Operation

• The index marks (0°, 5° and 10°) for adjust-ment of the clearance angle (59.1) are locat-ed on the right side of the knife holder (57).

• There is also an index mark (59.2) on theright side of the knife holder base (51)which serves as a reference point whenadjusting the clearance angle.

• Loosen the screw (58) using an Allen keysize 4 (71) until the knife holder (57) can bemoved.

• Move the knife holder until the index mark ofthe desired clearance angle coincides withthe reference line on the knife holder base.Example:Enlarged detail showing a clearance anglesetting of 5°.

The recommended clearance anglesetting for knife holder E is approx. 5°.

• Hold down the knife holder in this positionand retighten the screw (58) for clamping.

5.2 Adjusting the clearance angle

Fig. 18

51

58

57

59.1

59.2

Enlarged detail:Index marks foradjusting theclearance angle.

71

Page 25: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

25Leica RM2235

Fig. 19

Fig. 20

33

34

Important!Never turn thescrew morethan 1/2 turn ata time.

5.3 Fine adjustment of the force balance

If another accessory of a different weight ismounted on the object head (33), you mustcheck whether it is necessary to readjust theforce balance.Checking the correct setting:• Attach the new accessory and clamp the

specimen.• Set the object head to half the height of the

vertical travel range by turning the hand-wheel (Fig. 19).

If the object head remains in this exact position,the setting is correct.If the object head moves, i.e. it is raised or low-ered, fine adjustment is necessary.

Failure to adjust the force balancemay result in injury while working.

The force balance is adjusted using the screw(34), which can be accessed by removing thesection waste tray on the bottom of the baseplate of the microtome. Use the Allenkey provided, size 5 (with handle!) for theadjustment.

• If the object head moves downwards, turn thescrew approx. 1/2 turn clockwise.

• If the object head moves upwards, turn thescrew (34) approx. 1/2 turn counterclock-wise.

• Continue this procedure until the objecthead no longer moves once released.

5. Operation

Page 26: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

26 Instructions for use V 1.6 Rev A – 09/2011

5. Operation

5.4 Clamping the specimen

Always clamp the specimen block BEFORE clamping the knife.Lock the handwheel and cover the knife edge with the knife guardprior to any manipulation of knife or specimen, prior to changingthe specimen block and during all work breaks!

• Rotate the handwheel until the specimen clamp is in the uppermost po-sition.

• Lock the handwheel (allow lever (5) Fig. 2 to notch) and activate thebrake.

• Insert a specimen block into the specimen clamp.

A detailed description for inserting the specimen into variousspecimen clamps and specimen holders is provided in Chapter 6"Optional accessories".

5.5 Clamping the knife / disposable blade

Be very careful when handling microtome knives or blades. Thecutting edge is extremely sharp and can cause severe injury!

• Carefully insert knife or disposable blade into the knife holder andclamp.

• Make sure that the blade is clamped parallel to the upper edge of thepressure plate.(For more information see Chapter 6.3.2, Fig. 36)

A detailed description for inserting the blade or the knife intothe individual knife holders is provided in Chapter 6, "Optionalaccessories".

Page 27: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

27Leica RM2235

5.6 Sectioning

Always turn the handwheel evenly in clockwise direction;otherwise, the brake will not work properly. The rotation speedof the handwheel must be adapted to suit the hardness of the spec-imen.For harder specimens, use a slower speed.

Cutting into the specimen (trimming)

• Run the specimen to the rear end position by turning the coarse drivingwheel.

• Push the knife holder on the knife-holder base almost until it is justbefore the object.

• Orientate the position of the specimen surface (only in the case ofspecimen holders that can be orientated).

• Release the handwheel lock, or handwheel brake, respectively.• Using the trimming lever select the required trimming stage.• Begin the cutting process by turning the handwheel.• Stop the cutting process when the required specimen level has been

reached.• Let go of the trimming lever.

Remove the sections• Set the required section thickness, or check the value setting on the

display, respectively.Always use a different area of the cutting edge for trimming and sectioning.• To do so, laterally displace the blade or knife in the knife holder.

When using the knife holder E with lateral movement, it is sufficient tomove the knife holder sideways.

• For sectioning, turn the handwheel evenly in a clockwise direction.

Take care not to block the coarse driving wheel when turning thehandwheel!Otherwise there will be no feed motion of the section thicknessand thus no sectioning will occur.

• Pick up the sections and mount them on microscope slides.

5. Operation

Page 28: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

28 Instructions for use V 1.6 Rev A – 09/2011

5.7 Changing the specimen or interrupting sectioning

Lock the handwheel and cover the knife edge with the knife guardprior to any manipulation of knife or object head, as well as priorto changing the specimen block and during all work breaks!

• Raise the specimen to the upper end position and activate themechanical handwheel lock.

• Cover the sectioning edge with the knife guard.

• Remove the specimen from the specimen clamp and mount a newsample to continue.

• Run the object clamps with the coarse driving wheel back far enoughuntil the new specimen can start being cut.

5.8 Finishing daily routine

• Raise the specimen to the upper end position by turning the handwheeland engage the handwheel lock.

Always remove the knife / blade before detaching the knife holderfrom the instrument.Always put the knives back into the knife case when not in use!Never place a knife anywhere with the cutting edge facing up-wards and never try to catch a falling knife!

• Remove the blade from the knife holder and insert it in the receptacleat the bottom of the dispenser, or remove the knife from the knife hold-er and put it back in the knife case.

• Remove the specimen from the specimen clamp.• Push all section debris into the section waste tray and empty the tray.• Clean the instrument (see Chapter 8.1).

5. Operation

Page 29: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

29Leica RM2235

6.1 Assembly for fixture for specimen clamps

Depending upon the purchase order, the basic instrument is delivered with the directional orrigid fixture for specimen clamps which must be assembled first. All specimen clamps avail-able as accessories can be used in both fixtures for specimen clamps.Before assembling the fixture for specimen clamps, activate the mechanical handwheel lock!

6.1.1 Rigid fixture for specimen clamps

• Screw the rigid fixture for specimen clamps(4) onto the object head (3):Remove the screw (1), place the fixture forspecimen clamps (4) onto the object head (3)from the front and tighten the screws (2) withan Allen key size 3.Next, insert the screw (1) from the side andbriefly tighten it with an Allen key size 4.

Remove the rubber ring only afterattaching the object head!

Fig. 211

2

4

5

3

6.1.2 Directional fixture for specimenclamps

• Loosen the eccentric bolt (6) by turning itcounterclockwise.

• Completely unscrew the thrust piece (5) with aflat-tip screwdriver and pull it out with spring(5a) and pin (5b).

• Completely unscrew the setscrews (3) and (4).• Attach the directional fixture for specimen

clamps as shown.• Insert the screws (7+8) in the bore (2 screws

(8) are accessible through the bore (9a)) andevenly screw them in using a size 3 Allen key.

• Insert the spring (5a) and pin (5b) with the flatterside into the thrust piece (5). Completely screwin the thrust piece with a flat-tip screwdriver.

• Completely screw in the setscrews (3+4).

Fig. 22

1

2

35b5a5 6

4

7 + 8

7 + 8 9b 9b

9a 9a

• Finally, place the dovetail holder (2) and fastenby screwing in the 4 screws (1) using an Allenkey size 3.

6. Optional accessories

Page 30: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

30 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

6.1.3 Fine-directional fixture for specimenclamps

• Before the fine-directional fixture for speci-men clamps can be mounted, loosen4 screws (10) (Allen key size 3) and carefullyremove the fixture for specimen clamps fromthe baseplate (9).

Fig. 23

10

11

• Using the 4 supplied screws (11) and theAllen key size 3, fasten the baseplate to thespecimen head (12).

11

• Now, screw the fine-directional fixture forspecimen clamps with the 4 screws (10) andthe Allen key size 3 onto the object head.

If the fine-directional fixture for spec-imen clamps is not used, retain thebaseplate and 4 screws (11) togetherwith the fine-directional fixture forspecimen clamps!

9

12

Page 31: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

31Leica RM2235

6.1.4 Quick clamping system

It is used as specimen holder for use with thefine-directional fixture for specimen clampswith zero point indicators or the directionalfixture for specimen clamps.• Screw the 4 screws (13) into bore A with an

Allen key size 2.5 and tighten them.

Fig. 22

13

A

A

6. Optional accessories

Page 32: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

32 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

6.2.1 Standard specimen clamp

The standard specimen clamps are available intwo sizes: 40 x 40 mm and 50 x 55 mm.They are designed for direct clamping of rect-angular blocks. In addition, they accommodatethe foil clamps.

• Turn the knurled screw (66) counterclock-wise to move the movable jaw (68) down-ward.

• Mount the sample (67) as required.• Turn the knurled screw (66) clockwise to

move the movable jaw upward against thefixed jaw to securely clamp the sample.

6.2 Specimen clamps and holders

All specimen clamps available as accessories can be integrated into either the directional ornon-directional specimen holder fixture.

Fig. 25

66

6867

When clamping cassettes, makesure that they are not clamped tootightly, since the cassette bodiescan bend and result in sections thatare too thick or thin, or the entirespecimen may fall out and becomedamaged.

Page 33: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

33Leica RM2235

6.2.2 Vee insert

Fig. 26

70

66

69

70.1 68

The vee insert (70) is mounted in the hole pro-vided in the lower movable jaw of the stan-dard specimen clamp.This makes it possible to clamp round speci-mens in the standard specimen clamp.• Turn the knurled screw (66) counterclock-

wise to move the movable jaw (68) down-ward.

• Insert the pin (70.1) of the vee insert (70) inthe hole (69) of the lower jaw (68).

• Mount the sample as required.• Turn the knurled screw (66) clockwise to

move the movable jaw with the vee insertupward against the fixed jaw to securelyclamp the sample.

6. Optional accessories

Page 34: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

34 Instructions for use V 1.6 Rev A – 09/2011

Clamping of foil pieces• Move the movable jaw (74) to the right as re-

quired by turning the set screw with an Allenkey size 4 (71).

• Place the foil (72) between the movable jaw(74) and the fixed jaw (73).

• To clamp the foil, screw the movable jaw(74) against the fixed jaw (73) by using theAllen key.

• Insert the foil clamp (75) in the standardspecimen clamp as shown.

• Turn the knurled screw (66) clockwise toclamp the foil clamp in the standard speci-men clamp.

6.2.3 Foil clamp type 1The foil clamp type 1 is appropriate both forclamping very small foil pieces and flat, angularsamples. It is mounted in the standard speci-men clamp.

Clamping of flat, angular samplesTo clamp angular samples, replace the long setscrew (76) with the short set screw (77) provid-ed with the foil clamp.• Unscrew the long set screw (76) to the left

with an Allen key size 4 (71).• Screw the short set screw (77) in the hole.• Place the sample (67) between the movable

jaw (74) and the fixed jaw (73).• To clamp the sample, screw the movable

jaw (74) by turning the set screw (77) againstthe fixed jaw (73).

• Insert the foil clamp in the standard speci-men clamp as shown.

• Turn the knurled screw (66) clockwise toclamp the foil clamp in the standard speci-men clamp.

Fig. 27

Fig. 28

66

74

72

75

71

73

76

77

67

7473

6. Optional accessories

Page 35: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

35Leica RM2235

• To open the jaws (78) and (79) lightly loosenthe 3 screws (81) with an Allen key size 4(71).

• Insert the foil ribbon (72) from behind toposition it between the movable jaw (79) andthe fixed jaw (78).

• To clamp the foil, first tighten the screw inthe middle and then the other two screws(81) with the Allen key (71).

• Place the foil clamp (82) in the standardspecimen clamp so that the beveled surface(80) on the back of the foil clamp points tothe right or left.

• Turn the knurled screw (66) clockwise tosecurely clamp the foil clamp in the standardspecimen clamp.

6.2.4 Foil clamp type 2

The foil clamp type 2 is appropriate for large foil ribbons.It is mounted in the standard specimen clamp.

Fig. 29

71

72 7978

66

81

80

82

6. Optional accessories

Page 36: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

36 Instructions for use V 1.6 Rev A – 09/2011

Prior to sectioning, laboratory per-sonnel MUST check that the cas-sette is seated securely in the uni-versal cassette clamp.

• Push the lever (60) forwards.• Mount the cassette (65) horizontally or verti-

cally as required.• To clamp the cassette, release the lever (60).

6.2.5 Universal cassette clamp

Fig. 30

60

65

6. Optional accessories

Leica / Surgipath cassettes with minimum dimensions of 39.8 x 28 mm and maximum dimen-sions of 40.9 x 28.8 mm may be clamped in the universal cassette clamp (UCC) horizontally aswell as vertically.

When using other – particularly thin-walled – cassettes, the cassette may become de-formed or other problems may arise with the clamping system. If the user tries clamping thecassette and realizes that it is not securely clamped into place, an alternative tensioningclamp must be used.

When using cassettes whose lid is molded on, make sure that the broken edge left by re-moving the lid does not prevent the specimen from being securely clamped – if necessary,the specimen must be clamped horizontally.

Before clamping the cassette into the universal cassette clamp, remove excess wax on theoutside of the cassette to ensure that the cassette clamps in securely.

Wax deposits on the outside of the cassette can make the universal cassette clamp dirty.The dirt prevents the cassette from clamping in securely and can lead to sections being toothick or thin, chattering within the section and, in the worst-case scenario, damage to thespecimen. Prior to sectioning, the user has to verify that the specimen is clamped securelyand, if necessary, remove wax deposits from the universal cassette clamp according to thespecifications in Chapter 8.1 "Cleaning and maintenance – Universal cassette clamp".

Page 37: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

37Leica RM2235

Assembly of the super mega-cassette clampThe super mega-cassette clamp should preferably be used with therigid fixture for specimen clamps.To do so, proceed as follows:

Remove the rubber ring (5) only after having attached the cassetteclamp on the object head!

6.2.6 Super mega-cassette clamp

Fig. 321

2

34

5

6. Optional accessories

Use the ice-cooled UCC with the non-directional specimen holder fixtureonly!

• Ice cubes can be made using the includedParaflex mold (a).

• Quick clamping system (b), see page 31.

Universal cassette clamp, ice-cooled

Fig. 31

a

b

Page 38: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

38 Instructions for use V 1.6 Rev A – 09/2011

• Screw the rigid fixture for specimen clamps (4) onto the objecthead (3):Remove the screw (1), place the fixture for specimen clamps (4) ontothe object head (3) from the front and tighten the screws (2) with anAllen key size 3.Next, insert the screw (1) from the side and briefly tighten it with anAllen key size 4.

• Insert the super mega-cassette clamp from the side on the left into thedovetail guide of the rigid fixture for specimen clamps and tightenscrew (1).

If the directional fixture for specimen clamps is used with the rigidknife holder base, the orientation must be in position "0" and thecover for backlighting illumination must be detached. (Danger ofcollision if not observed!)

NEVER use the super mega-cassette clamp with backlightingillumination!

6. Optional accessories

Page 39: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

39Leica RM2235

• To mount the required insert (89.1-3) turn theclamping ring (90) counterclockwise andremove it.

• Place the required insert into the tensionring (90) and screw the tension ring onto thethread (91) by turning it clockwise.

• Mount the sample (67) and fix by turning theclamping ring (90) clockwise.

• To orient the inserted sample, insert the pin(92) into the bore (93) and rotate it counter-clockwise to release the clamp. You cannow rotate the specimen so that the sideyou want faces upwards.

• To lock it in the position you have chosen,tighten the pin (92) by turning it clockwise.

6.2.7 Holder for round specimens

The holder for round specimens is designed to accommodate cylindrical samples.Inserts for specimens of 6, 15 and 25 mm diameter are available.

Fig. 33

89.3

67

89.1

93

90

89.2

91

6. Optional accessories

92

Page 40: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

40 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

6.3 Knife holder base and knife holder

Repositioning the knife holder baseThe one-piece knife holder base (rigid) (51) canbe moved forwards and backwards on themicrotome base plate.This vertical displacement allows bringing theknife holder into the optimal cutting position inrelation to the specimen.

• To release, rotate the clamping lever (50) onthe right side of the microtome base platecounterclockwise.

• Reposition the knife holder together with theknife holder base forward or backward asappropriate.

• Secure the clamping mechanism by rotatingthe lever (50) clockwise.

The plastic handles of all clamping levers on theinstrument and knife holders can be turned tothe position that is most convenient for eachuser.Pull the grip (94) out of the lever, hold it in thisposition, and rotate it to the desired position. Itwill then lock automatically when released.

6.3.1 Knife holder base, without lateralmovement feature

Fig. 34

Fig. 35

51

50

94

Page 41: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

41Leica RM2235

6. Optional accessories

Prior to inserting the blade, both knifeholder and knife holder base musthave been installed on the instrument!

Inserting the blades, knife holder E and E-TC

• Fold knife guard (9) downward.

• To insert the blade, flip the right clampinglever (10) forward and down.

• Carefully insert the blade from the side.Make sure that the blade is clamped parallelto the upper edge of the pressure plate.

• To clamp the blade, rotate clamping lever(10) back upwards.

6.3.2 Knife holder E/E-TC

The knife holder E is designed for conventional disposable blades from all current manufac-turers. It is available in two models: one for low profile, blades and one for high profile,blades. The knife holder has a lateral movement, so that the entire width of the blade canbe used.

Fig. 37

9

1110

Fig. 36

Knife holder E

The knife holder E-TC is designed forthe Leica TC-65 tungsten carbideblades.

The clamping levers on the knife hold-er are not interchangeable. The twoclamping levers (10, 11) must remainin the position shown at all times, asotherwise isolated malfunctions ofthe knife holder can occur.Clamping lever for the blade (10) atthe right, clamping lever for the later-al displacement (11) at the left.

Knife holder E-TC

9 10

Page 42: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

42 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

Lateral displacement (only for knife holder E)The lateral movement feature of the knife hold-er base enables the use of the entire length ofthe blade, eliminating the need for readjust-ing the knife holder. The knife holder E con-sists of a segment arch A (with lever (11))and the clamp mount B (with lever (10) andejector (13)).The extreme left and right positions as well asthe middle position are each marked with aclick stop.• To release the clamp, rotate the lever (11)

on the left side of the knife holder for-wards.

• Move the knife holder sideways.• To clamp, rotate the lever (11) back.

Fig. 38

11

Note on the knife holder E:

The knife holder E is an important precision component, the quality andprecise adjustment of which have a lasting effect on the entire function ofthe microtome. In case of malfunctions or damages to the clamp mount, itis always required to replace the clamp mount, including the pertinentclamping lever.Leica Biosystems offers special prices for new clamp mounts in case ofdamages to the clamp mount after the warranty has expired.In this way, perfect function of the device can be ensured over the courseof many years.

Knife holder E

A B

10

To ensure a proper sectioning result, clampmount B must be securely clamped onto seg-ment arch A.The clamping is carried out using an eccentricat the lever (11). The clamping force is adjustedwith the setscrew (12) on the underside of thesegment arch. The setting of the clamping iscarried out so that the clamping lever can berotated to the stop with constantly increasingresistance.

Setting of the clamping system of the clamp mount on the segment arch

12

Adjust the clamping with a 2.5 mm Allen key atthe setscrew (12) so that the lever initially"slips" when the lever is activated. Continueturning the setscrew (12) a little bit (approx. 1/4turn to the left or right), then check that the le-ver no longer "bars", but also does not jam"heavily".(The language CD features a short video film forthis purpose.)

13

Page 43: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

43Leica RM2235

6. Optional accessories

Knife holder E with water bath for narrow-band and broad-band blades

Fig. 39

Knife holder E with water bath is available forboth low profile and high profile blades.The knife guard on knife holder E consists ofa red foldaway handle (9). To cover the cut-ting edge, fold the knife guard handle (9) up-wards as illustrated in Figure.

911 10

The clamping levers on the knife holder are not interchangeable. The two clamping levers(10 and 11) must remain in the position shown at all times, as otherwise isolated malfunctionsof the knife holder can occur.Clamping lever for the blade (10) at the right, clamping lever for the lateral displacement (11) atthe left.

UseFloating thin paraffin sections (for example, for subsequent immunostain-ing procedures) on the surface of the water. Flat paraffin sections can beremoved from the surface of the water using glass slides.

The vessel is filled with water up to the blade.After trimming, remove the section waste fromthe tray and create the sections to be prepared.The sections floating on the surface of thewater can be removed using the glass slide.

Fig. 40

Page 44: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

44 Instructions for use V 1.6 Rev A – 09/2011

6.3.3 Knife holder N/NZ

The knife holders N and NZ are appropriate for standard steel and tungsten carbide knives,profile c and d, up to 16 cm long. The integrated height adjustment feature allows you to alsouse knives that have been resharpened numerous times.

Fig. 41

8

46

47

4948

49

Enlarged detail:Knife inserted andheight-adjusted

39

6. Optional accessories

Mounting the knife support bar• Push knife guard (8) to the center.• Set the knife support bar (46) onto the height adjustment screws (not vis-

ible) in the position shown. The flat ends of the height adjustment screwsmust be located in the slots at each end of the knife support bar.

Prior to inserting the knife, both knife holder and knife holder basemust have been installed on the instrument!

Knife holder NFor holding convention-al knives up to 16 cm inlength.

Knife holder NZFor holding conventional and hardmetal knives up to 16 cm in length.Knife pressure plate (56) for extremestability and full utilization of theknife blade.

56

Page 45: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

45Leica RM2235

6. Optional accessories

Inserting the knife

• Rotate the knurled nuts (48) on the right and left of the knife holderforward in opposite directions, lowering the knife support bar to thelowest possible position, thus ensuring that the knife edge will not bedamaged when inserting the knife.

• Unscrew the clamping screws (49) as far out as possible (rotate coun-terclockwise).

• Hold the knife (47) at the base back and carefully insert it in theholder from the side as shown with the cutting edge facing upward.

Knife height adjustment

When adjusting the clearance angle, the knife edge should be positionedas exactly as possible in the actual center of rotation of the knife holder.The lay-on edge (39) of the rear clamping chucks serves as a referenceposition for correct knife height adjustment. The knife edge should be par-allel with the locating edges.• Rotate the knurled nuts (48) uniformly and backwards until the knife

blade is parallel to the lay-on edge (39) (see detailed illustration) of therear clamping chucks.

• To clamp the knife (47) evenly screw the two knife clamping screws(49) inward (rotate clockwise).

Lateral repositioning of the knife

• Push knife guard (8) to the center.• Loosen the clamping screws (49) by turning them counterclockwise.• Push the knife (47) to the left or right as required.• To clamp the knife (47), always tighten the clamping screw (49) first

which is located on the side to which the knife has been repositionedby turning it clockwise.

Page 46: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

46 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

• Push the section waste tray (18) from thefront to the microtome base plate (53) until itis held in place by the two magnets (39) (onthe front of the microtome base plate).To remove the section waste tray, lift itslightly and pull it off towards the back.

Fig. 42

6.4 Section waste tray

53

18

39

The backlighting is inserted into thefront of the one-piece knife holderbase.

• Remove the two screws (1) using a slottedscrewdriver and then remove the cover plate (2).

• Insert the backlighting (3) in the recess atthe rear of the knife holder base.

• Insert the plug (4) for the backlighting intothe socket (5) of the microtome and connectthe plug of the power adapter to an AC pow-er socket.

The backlighting illuminates once the micro-tome is turned on with the power switch.

NEVER use the backlighting illumina-tion with the super mega-cassetteclamp!

Fig. 43

2

1

3

4

1

5

6.5 Backlighting

Page 47: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

47Leica RM2235

6. Optional accessories

6.7 Freezer pack

Fig. 45

98

99

98

The freezer pack consists of the freezer plate(98) and the insulation jacket (99). It is for cool-ing (or keeping cold) paraffin specimens. Up to35 standard cassettes can be cooled at thesame time. The specimens require approx.20 min. for being cooled from room temperature(approx. 20 °C) to "sectioning temperature".The cooling performance depends on the ambi-ent temperature and the volume of the speci-mens being cooled.The insulation jacket can be placed on the hoodof the microtome in the same manner as thetray, or it can be placed next to the instrument.

When using the freezer pack on theinstrument the tray (Fig. 44) must beremoved.

• Remove the freezer plate from the insulationjacket and place it in a deep freezer, ideallyovernight (but for at least six hours) atapprox. -23 °C.

• Reassemble the freezer plate and insulationjacket and place them on the bench or theinstrument.

Fig. 44

6.6 TrayThe tray is mounted on the hood of the micro-tome so that the small feet on the underside fitinto the cutouts on the hood.It is for storage of the utensils used duringsectioning as well as the sectioned specimens.

Page 48: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

48 Instructions for use V 1.6 Rev A – 09/2011

bc

23

e

d

4

5

6

78

6.8 Universal microscope carrier

6. Optional accessories

1 - Baseplate with drilling (a)2 - Vertical column with size 8 cap screw (b) and

lock washer (c)3 - Horizontal arm with cross-member (d) and

support ring (e)4 - Support plate, large (for RM2235, RM2245 and

RM2255)5 - Support plate, small (for RM2265)6 - Allen key size 37 - 4 countersunk screws to install support plate8 - Allen key size 8

Unpack all accessories in the packageand check for completeness.

Assembling the universal microscopecarrier

1. Attach baseplate• Select the large (5) or small (4) support plate,

depending on the microtome to be used.• Attach the support plate to the baseplate with

the included 4 countersunk screws (7) usingthe Allen key size 3 (6).

1

a

Fig. 46

Fig. 48

Fig. 47

Page 49: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

49Leica RM2235

2. Attach vertical column• Insert the cap screw (b) into the hole of the

baseplate from below. Place the lock washer(c) on the cap screw from above.

• Thread the silver vertical column (2) onto thebaseplate from above and tighten with theAllen key size 8.

The lock washer must be positionedbetween the baseplate and verticalcolumn to prevent unintentional turn-ing of the column.

A

Extreme danger of tipping! After install-ing the vertical column, immediatelyplace the microtome onto the baseplateso that the front feet of the microtome arepositioned in the shallow recesses (A).

3. Attach horizontal arm• Slide the support ring (e) onto the vertical

column and position it so that the lock nut (f)is facing to the back. Tighten the lock nut.

• Slide the cross-member (d) onto the column.Ensure that the lock screw (g) is facing to theright of the baseplate. The horizontal arm mustbe centered over the microscope.

• Slide the horizontal arm (3), flat side facing thelock screw (g), into the cross-member (d) andtighten.

For more information on connectingand using the microscope, magnifyinglens or cold-light source, please seethe appropriate manual.

6. Optional accessories

Fig. 51

Fig. 50

Fig. 49

bc

2

e

d2g

f

h

Page 50: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

50 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

6.9 Magnifying lens

Fig. 52

Fig. 54

Fig. 53

• Open the screw (3) on the horizontal arm ofthe microscope carrier in a counterclock-wise direction.

• Insert the silver connection piece (1) as faras it will go. Tighten the screw (3).

• The adapter (2) allows a fiber-optic lightguide to be installed.

• Adjust the position of the magnifying lens tothe specimen being processed.The magnifying lens can be swiveledcompletely to the side if necessary.

1

3

The magnifying lens provides a 2xmagnification and can be used withall Leica 2200 series rotary micro-tomes.

2

4

• Use the provided protective cap (4) to coverthe magnifying lens.

Always protect the magnifier glass fromexposure to direct sunlight! Exposure tosunlight may cause a burning glasseffect.Danger of fire!

• The instrument illustrations in Fig. 53 andFig. 54 are an example only.

Page 51: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

51Leica RM2235

Knife holder base, without lateral movementfeatureSilver for knife holders N, NZ, E, and E-TC

Order No. ........................................... 14 0502 37962

Knife holder N,silver, for holding conventionalknives up to 16 cm in length,height adjustment of the knife blade,separate clearance angle adjustment,moveable knife guard.

Order No. ........................................... 14 0502 37993

6. Optional Accessories

Knife holder NZ, silverfor holding conventional and carbide metalknives up to 16 cm in length, knife pressureplate for extreme stability and full utilization ofthe knife blade, height adjustment of the knifeblade, separate clearance angle adjustment,moveable knife guard.

Order No. ........................................... 14 0502 37994

Fig. 55

Fig. 57

Fig. 56

6.10 Additional accessories

Page 52: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

52 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

Knife holder E,for low-profile disposable blades,(80 x 8 x 0.25 mm), silver, for RM22xxQuick clamp system withability to shift the blade laterally.3 click-stop positions make it easier to utilizethe entire blade.Color-coded, foldaway knife guard.With blade ejector

Order No. .......................................... 14 0502 40508

For high-profile disposable blades,(80 x 14 x 0.25 mm), silver, for RM22xx

Order No. .......................................... 14 0502 40509

Knife holder E,for low-profile disposable blades,(80 x 8 x 0.25 mm), silver, for RM22xxwith water bath,Quick clamp system withability to shift the blade laterally.3 click-stop positions make it easier to utilizethe entire blade.Color-coded, foldaway knife guard.With blade ejector

Order No. .......................................... 14 0502 38961

For high-profile disposable blades,(80 x 14 x 0.25 mm), silver, for RM22xxwith water bath,

Order No. .......................................... 14 0502 39114

Fig. 59

Fig. 58

Page 53: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

53Leica RM2235

Knife holder E-TC,for carbide metal disposable blades TC-65, sil-verQuick clamp system,rustproof clamping plate made of stainlesssteel, rear pressure plate made of carbide metal

Order No. ........................................... 14 0502 37997

Fig. 60

Ralph insertfor using Ralph knivesin knife holder GD -with clamping frame for 25-mm Ralph knifeOrder No. ........................................... 14 0502 33795

with clamping frame for 38-mm Ralph knifeOrder No. ........................................... 14 0502 33796

Trimming insert,for knife holder GD,for clamping the universal flat specimen holder

Order No. ........................................... 14 0502 33794

Fig. 62

Fig. 61

6. Optional Accessories

Page 54: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

54 Instructions for use V 1.6 Rev A – 09/2011

Fig. 63

Disposable blades – low-profile (819)80 x 8 x 0.25 mm01 package of 50 pcs. ...................... 14 0358 3892510 packages of 50 pcs. .................... 14 0358 38382

Disposable blades – high-profile (818)80 x 14 x 0.317 mm01 package of 50 pcs. ...................... 14 0358 3892610 packages of 50 pcs. .................... 14 0358 38383

Disposable blades Leica TC-65Leica TC-65 microtome, disposable blade sys-tem for sectioning hard specimen materials.The Leica TC-65 carbide metal disposableblades were specially developed for the re-quirements in labs where hard, blunt materialsare routinely sectioned. The one-of-a-kind fine-grain carbide metal guarantees sections to ap-prox. 1 µm. The blades are fully recyclable.Length: 65 mmThickness: 1 mmHeight: 11 mm

1 pack of 5 pcs.Order No. ........................................... 14 0216 26379

Fig. 64

Fig. 65

6. Optional accessories

Page 55: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

55Leica RM2235

Knife 16 cm – profile c – steelKnife, 16 cm long, profile cNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 07100

Knife 12 cm – profile c – steelKnife, 12 cm long, profile cNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 07092

Profile

Knife 16 cm, profile d – steelKnife, 16 cm long, profile dNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 07132

Knife 12 cm, profile d – steelKnife, 12 cm long, profile dNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 07130

Knife, 16 cm, profile d, tungsten carbideKnife, 16 cm long, tungsten carbide, profile dNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 04813

Knife 16 cm, profile c, tungsten carbideKnife, 16 cm, tungsten carbide, profile cNote: Knife case 14 0213 11140 includedOrder No. ........................................... 14 0216 04206

Serial numbercarbide metalknife

Fig. 68

Fig. 66

Fig. 67

6. Optional Accessories

Page 56: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

56 Instructions for use V 1.6 Rev A – 09/2011

Knife caseVariable knife case (plastic),for 1 or 2 knives: 10 – 16 cm long(Carbide metal or SM2500 knife: only for1 knife!)

Order No. ........................................... 14 0213 11140

Universal cassette clamp, ice-cooledwith adapter and one-hand clamp,also includes:quick clamping system for specimen holder, silver

For use with standard cassettes with dimen-sions that are 39.8 x 28 mm at minimum and40.9 x 28.8 mm at maximum.

Note:Use the ice-cooled universal cassette clampwith the non-directional specimen holder fixture14 0502 38160 only!Recommendation:Ice cubes can be made using the includedParaflex mold 14 0339 10963.

Order No. ......................................... 14 0213 11140

Universal cassette clampwith adapterRM2125 RTS, silver

For use with standard cassettes with dimen-sions that are 39.8 x 28 mm at minimum and40.9 x 28.8 mm at maximum.

Order No. ......................................... 14 0502 37999

Fig. 71

Fig. 70

Fig. 69

6. Optional accessories

Page 57: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

57Leica RM2235

Standard specimen clamp50 x 55 mm, with adapter, silver

Order No. ........................................... 14 0502 38005

Specimen holder fixture, directionalsilverFor holding the standard clamp,universal cassette clamp,round specimen holder and EM specimenholder,including installation hardware

Note: Quick clamping system14 0502 37718 has to be ordered separately

Order No. ........................................... 14 0502 38949

Standard specimen clamp40 x 40 mm, with adapter, silver

Order No. ........................................... 14 0502 37998

Fig. 74

Fig. 73

Fig. 72

6. Optional Accessories

Page 58: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

58 Instructions for use V 1.6 Rev A – 09/2011

Specimen holder fixture, non-directionalsilver, for holding the standard clamp, universalcassette clamp, round specimen holder and EMspecimen holder, including installation hard-ware

Order No. ........................................... 14 0502 38160

Quick clamping system,for specimen holder for use with the fine-direc-tional fixture for specimen clamps with zeropoint indicators 14 0502 37717 or the directionalfixture for specimen clamps 14 0502 38949

Order No. ........................................... 14 0502 37718

Specimen holder fixture, fine directional, silver,with 2 zero point indicators, XY orientation 8°per direction, click stops every 2°

Note: Quick clamping system 14 0502 37718 hasto be ordered separately.

Order No. ........................................... 14 0502 37717

Fig. 77

Fig. 76

Fig. 75

6. Optional accessories

Page 59: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

59Leica RM2235

Round specimen holderwith adapter, without clamping rings, silver

Order No. ........................................... 14 0502 38001

Round specimen holderwith adapter, with 3 clamping rings, silver

Order No. ........................................... 14 0502 38002

Vee insert for standard specimen clamp,silver

Order No. ........................................... 14 0502 38000

Fig. 79

Fig. 80

Fig. 78

6. Optional Accessories

Page 60: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

60 Instructions for use V 1.6 Rev A – 09/2011

Foil clamp Type Ifor standard specimen clamp, blackMaximum specimen size: 25 x 13 mm

Order No. ........................................... 14 0402 09307

Clamping ringfor round specimen holder, 6 mm diameter

Order No. ........................................... 14 0356 08322

Clamping ringfor round specimen holder, 15 mm diameter

Order No. ........................................... 14 0356 09200

Clamping ring,for round specimen holder, 25 mm diameter

Order No. ........................................... 14 0356 08320

Fig. 84

Fig. 83

Fig. 82

Fig. 81

6. Optional accessories

Page 61: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

61Leica RM2235

Foil clamp Type IIfor standard specimen clamp, black

Order No. ........................................... 14 0402 26922

Special wrench,for EM specimen holder

Order No. ........................................... 14 0356 10869

EM specimen holder fixture,for EM specimen holder, blackFor samples with a 10mm diameter

Order No. ........................................... 14 0502 29968

Fig. 86

Fig. 85

Fig. 87

6. Optional Accessories

Page 62: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

62 Instructions for use V 1.6 Rev A – 09/2011

EM universal specimen holder,For samples with a 8.5mm diameter

Order No. ........................................... 14 0356 10868

EM flat specimen holder,Opening width up to 4.5 mm

Order No. ........................................... 14 0355 10405

Super Mega cassette clamp,with adapter, silver

Note:Use only together with non-orientable specimenholder fixture 14 0502 38160, which has to be or-dered separately.Backlighting 14 0502 38719 cannot be used inconnection with a Super Mega cassette clamp.We recommend use with Surgipath SuperMega cassettes and covers, white (VSP59060B-BX, VSP 59060-CS) and Super metal em-bedding molds (VSP58166)(L x W x H) 75 x 52 x 35 mmOrder No. ........................................... 14 0502 38967

Fig. 90

Fig. 89

Fig. 88

6. Optional accessories

Page 63: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

63Leica RM2235

Leica RM CoolClamp,Electrically cooled universal cassette clamp withadapter for the Leica RM2200 series rotary micro-tomes. For use with standard cassettes withdimensions that are 39.8 x 28 mm at minimumand 40.9 x 28.8 mm at maximum.Uniform microtome sections through electriccooling of the universal cassette clamp using theLeica RM CoolClamp. Energy-savingcooling through the heat dissipation system.The patented force balance system of the LeicaRM2200 series rotary microtomes enables thespecimen clamps to be changed quickly and reli-ably without modifying the handwheel. Antistaticmaterial makes cleaning easier. Can be retrofit-ted for all rotary microtomes from the LeicaRM2200 series.

Please note:The RM CoolClamp cannot be used in con-nection with a microscope or magnifier.Technical data:Precooling time until work begins: 30 minutesTemperature: 20 K below ambient temperatureAmbient temperature range: +10 °C to +35 °CRelative humidity: max. 80%, non-condensingWeight: approx. 650 gDimensions (D x W x H): 80 x 114 x 204 mmPower connection: 100-240 V / 50-60 HzCertificates: CE, c_CSA_US

Standard delivery:RM CoolClampPower supply with cable and 3 adapters (UK, US,EU)5 cable clampsInstructions for Use and DVD

Order No. ........................................... 14 0502 46573

Fig. 91

6. Optional Accessories

Page 64: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

64 Instructions for use V 1.6 Rev A – 09/2011

Magnifier,for assembly on the microscope carrier14 0502 40580, 2x magnificationNote:Magnifier includes an adapter for holding theoptional fiber-optic light guide14 0502 30028

Order No. ........................................... 14 0502 42790

Universal microscope carrier, universalassembly

Order No. ........................................... 14 0502 40580

Light guide,gooseneck

Order No. ........................................... 14 0386 31352

Fig. 94

Fig. 92

Fig. 93

6. Optional accessories

Page 65: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

65Leica RM2235

Leica CLS 100, cold light source,with power adapter – USA 100-120 V / 50-60 HzOrder No. ........................................... 14 0502 30214

with power adapter – Europe andthe United Kingdom 230-240 V / 50-60 HzOrder No. ........................................... 14 0502 30215

with power adapter – Australia 240V / 50-60 HzOrder No. ........................................... 14 0502 30216

Backlighting,Note:For RM2235 and RM2245 only together with ex-ternal power supply 14 0500 31244.Backlighting cannot be used in connection witha Super Mega cassette clamp14 0502 38967.

Order No. ........................................... 14 0502 38719

Fig. 97

External power supply unit,for use with backlighting 14 0502 38719, for rota-ry microtome seriesLeica RM2235 and RM2245

The following adapters are included:UK, Europe, USA/Japan, Australia

Order No. ........................................... 14 0386 31352

Fig. 96

Fig. 95

6. Optional Accessories

Page 66: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

66 Instructions for use V 1.6 Rev A – 09/2011

Tray,for rotary microtome series Leica RM2200

Order No. ........................................... 14 0502 37932

Handrest,for the section waste tray

Order No. ........................................... 14 0502 38770

Fig. 100

Cooling plate,with one freezer pack

Order No. ........................................... 14 0386 38325

Fig. 99

Fig. 98

6. Optional accessories

Page 67: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

67Leica RM2235

Section waste tray,large, included in the standard scope of deliveryof RM2235, RM2245, and RM2255.Order No. ........................................... 14 0502 37931

small, included in the standard scope of deliv-ery of RM 2265

Order No. ........................................... 14 0503 39060

Brush, 'Leica'with magnet for blade removal tool for knifeholder E.

Order No. ........................................... 14 0183 40426

Fig. 102

Fig. 101

6. Optional Accessories

Page 68: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

68 Instructions for use V 1.6 Rev A – 09/2011

6. Optional accessories

6.11 Ordering information

Knife holder-base, non-orientable, silver ............................................................................ 14 0502 37962Knife holder N RM22xx, silver .................................................................................................. 14 0502 37993Knife holder NZ RM22xx, silver ............................................................................................... 14 0502 37994Knife holder E narrow band RM22xx, silver, with ejector ................................................. 14 0502 40508Knife holder E broad band RM22xx, silver, with ejector ................................................... 14 0502 40509Knife holder E narrow band Run-off tray RM22 .................................................................. 14 0502 38961Knife holder E broad band Run-off tray RM22 ..................................................................... 14 0502 39114Knife holder E-TC RM22xx, silver ............................................................................................ 14 0502 37997Ralph insert, for 25 mm Ralph knife ........................................................................................ 14 0502 33795Ralph insert, for 38 mm Ralph knife ........................................................................................ 14 0502 33796Trimming insert ........................................................................................................................... 14 0502 33794Disposable blades – low-profile, 1 pack of 50 (80 x 8 x 0.25 mm) .................................... 14 0358 38925Disposable blades – low-profile, 10 packs of 50 ................................................................ 14 0358 38382Disposable blades – high-profile, 1 pack of 50 (80 x 14 x 0.317 mm) .............................. 14 0358 38926Disposable blades – high-profile, 10 packs of 50 ............................................................... 14 0358 38383Disposable blades Leica TC - 65 ............................................................................................ 14 0216 26379Knife 16 cm – profile c – steel .................................................................................................. 14 0216 07100Knife 12 cm – profile c – steel .................................................................................................. 14 0216 07092Knife 16 cm – profile d – steel .................................................................................................. 14 0216 07132Knife 12 cm – profile d – steel .................................................................................................. 14 0216 07130Knife 16 cm – profile d – carbide metal ................................................................................. 14 0216 04813Knife 16 cm – profile c – tungsten carbide .......................................................................... 14 0216 04206Universal cassette clamp, with adapter RM2125 (silver) .................................................. 14 0502 37999Universal cassette clamp – ice-cooled................................................................................ 14 0502 37793Standard specimen clamp with adapter 50x60, silver ...................................................... 14 0502 38005Standard specimen clamp with adapter 40x40, silver ...................................................... 14 0502 37998Fixture for specimen clamps, directional, silver ................................................................. 14 0502 38949Fixture for specimen clamps, non-orientable, silver ......................................................... 14 0502 38160Fixture for specimen clamps, directional, with 2 zero point indicators, silver ............ 14 0502 37717Quick clamping system assembly .......................................................................................... 14 0502 37718Round specimen holder, silver ............................................................................................... 14 0502 38001Round specimen holder with adapter and 3 clamping rings, silver .............................. 14 0502 38002V-insert for standard specimen clamp, silver ..................................................................... 14 0502 38000Clamping ring for round specimen holder, D = 6 mm ......................................................... 14 0356 08322Clamping ring for round specimen holder, D = 15 mm ....................................................... 14 0356 09200Clamping ring for round specimen holder, D = 25 mm ....................................................... 14 0356 08320

Page 69: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

69Leica RM2235

Foil clamp – type 1, black ............................................................................................................ 14 0402 09307Foil clamp – type 2, black ............................................................................................................ 14 0402 26922Special wrench for EM specimen holder ................................................................................. 14 0356 10869EM specimen holder fixture ........................................................................................................ 14 0502 29968EM universal specimen holder ................................................................................................... 14 0356 10868EM flat specimen holder ............................................................................................................. 14 0355 10405Megaclamp assembly RM22xx, silver ....................................................................................... 14 0502 38967CoolClamp ...................................................................................................................................... 14 0502 46573Magnifying lens assembly .......................................................................................................... 14 0502 42790Universal microscope carrier ..................................................................................................... 14 0502 40580Fiber optic light guides................................................................................................................. 14 0502 30028Light guide, gooseneck ................................................................................................................ 14 0386 31352Cold light sources

Leica CLS 100X, 100-120V/50-60Hz ....................................................................................... 14 0502 30214Leica CLS 100X, 230V/50-60Hz .............................................................................................. 14 0502 30215Leica CLS 100X, 240V/50-60Hz .............................................................................................. 14 0502 30216

Backlighting assembly ................................................................................................................. 14 0502 38719External power supply assembly ............................................................................................... 14 0500 31244Tray ................................................................................................................................................. 14 0502 37932Handrest for waste tray ............................................................................................................... 14 0502 38770Cooling device assembly ............................................................................................................. 14 0502 38325Section waste tray, large ............................................................................................................ 14 0502 37931Section waste tray, small ............................................................................................................ 14 0503 39060‘Leica’ brush w/magnet ............................................................................................................... 14 0183 40426Demo Box RM22xx series ........................................................................................................... 14 0502 38930

6. Optional accessories

Page 70: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

70 Instructions for use V 1.6 Rev A – 09/2011

In the following table there is a list of the most common problems which can arise whileworking with the instrument, along with possible causes and troubleshooting procedures.

7.1 Possible faults1. Thick/thin sections

The sections alternate betweenbeing thick and thin, or there ischattering in the sections, or thespecimen is torn out of the embed-ding. In extreme cases, there areno sections whatsoever.

• Reclamp the blade.

• Check that the cassette is securelyclamped in the universal cassetteclamp.

• If the universal cassette clamp issoiled by wax, clean the universalcassette clamp (refer to Chapter 8.1– "Cleaning and maintenance –Universal cassette clamp").

• When using cassettes whose lid ismolded on, check that the brokenedge allows the cassette to beclamped securely; if necessary,remove burrs or clamp the cassetteinto the universal cassette clamphorizontally instead of vertically.

• If the cassette dimensions arewithin the specified tolerances andthe cassette still cannot be securelyclamped into place, it may be thatthe universal cassette clamp isconfigured incorrectly or is defec-tive. In this case, have TechnicalService inspect and reconfigure theuniversal cassette clamp.

• When using cassettes, particularlythin-walled cassettes, from a com-pany other than Leica or Surgipath,the cassette may become deformedor other clamping problems mayarise. If while trying to clamp thecassette you realize that it is notsecurely clamped into place, analternative tensioning clamp mustbe used.

7. Customized solutions

Problem Corrective actionPossible cause

• The blade, knife holder or orienta-tion is not clamped properly.

• The specimen is not clampedproperly.

Page 71: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

71Leica RM2235

• Reclamp the blade.

• Laterally displace the knife holderor insert a new blade.

• Insert a new pressure plate or usea new knife holder.

• Readjust the pressure plate.

• Methodically experiment withlarger clearance angle settingsuntil you have found the optimumangle.

• The blade is not clamped properly.

• The blade is dull.

• The pressure plate is damaged orincorrectly adjusted.

• Clearance angle of the knife/bladetoo small.

• The blade is dull.

• The specimen is too warm.

• The sectioning speed is too fast.

• There is an accumulation of paraf-fin on the rear pressure plate of theknife holder.

7.2 Instrument malfunctions1. There is no further feed motion and

thus no sectioning.

3. "Stripes" in sectionsFor knife holder E

2. Section compressionThe sections are very compressed,show folds or are squeezedtogether.

• Adjust the sectioning speed and/orsection thickness when trimming.Select a smaller section thickness,rotate the handwheel more slowly.

• Run the specimen back by turningthe coarse driving wheel.

2. High blade consumption

• Remove paraffin from this area on aregular basis.

• Use another area of the blade or anew blade.

• Cool the specimen before section-ing.

• Reduce the sectioning speed.

• The front end position has beenreached.

• Too great of a sectioning force wasapplied.

4. Noises during sectioningThe knife "sings" when sectioninghard specimens. The sections showscratches or chatter marks.

• The sectioning speed is too fast.

• The clearance angle is too wide.

• Insufficient clamping of object and/or knife holder.

• Turn the handwheel at a slowerspeed.

• Methodically reduce the clearanceangle setting until you have foundthe optimum angle.

• Check all screw and clamp connec-tions on the object holder systemand the knife holder. If necessary,tighten the levers and screws.

7. Customized solutions

Problem Corrective actionPossible cause

Page 72: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

72 Instructions for use V 1.6 Rev A – 09/2011

8.1 Cleaning the instrument

Always remove the knife / blade before detaching the knife holder from the instrument.Always put the knives back into the knife case when not in use!Never place a knife anywhere with the cutting edge facing upwards and never try to catch afalling knife!When using cleaners, comply with the safety instructions of the manufacturer and the labor-safety regulations of your laboratory!When cleaning the outer surfaces, do not use xylene, scouring powders or solvents containingacetone or xylene. Xylene or acetone will damage the finished surfaces!Ensure that liquids do not enter the interior of the instrument during cleaning!

Before each cleaning carry out the following preparatory steps:

• Raise the specimen clamp to the upper end position and activate thehandwheel lock.

• Switch the unit off and unplug it.• Remove the blade from the knife holder and insert it in the receptacle

at the bottom of the dispenser, or remove the knife from the knife hold-er and put it back in the knife case.

• Remove knife holder base and knife holder for cleaning.• Remove the specimen from the specimen clamp.• Remove section waste with a dry brush.• Remove specimen clamp and clean separately.

Instrument and outside surfaces

If necessary, the varnished outside surfaces of the control panels can becleaned with a mild commercial household cleaner or soap water and thenbe wiped with a moist cloth.To remove paraffin residue, xylene substitutes, paraffin oil, or paraffin remov-ers such as "Para Gard" (Polysciences) can be used.The instrument must be completely dry before it can be used again.

Cleaning the knife

Always wipe the knife from the back of the knife to the cutting edge. NEVER wipe in the oppo-site direction – risk of injury!

Clean using an alcohol-based solution or acetone.

8. Cleaning and maintenance

Page 73: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

73Leica RM2235

Take the knife holder apart for cleaning. To do so,proceed as follows:• Fold knife guard (9) downward.• Rotate the clamping lever (11) of the lateral

displacement forwards and pull it out side-ways.

• Push the knife holder base plate (86) with thepressure plate (83) until it can be removedfrom the segment arch (87).

• Rotate the blade clamping lever (10) down-ward and pull it out sideways.

• Remove pressure plate (83).• Clean all parts of the knife holder.

For cleaning and removal of paraffin, donot use xylene or cleaning fluids con-taining alcohol (e.g. glass cleaner).

• Lay the removed parts on an absorbent clothinto the drying chamber (up to a max. of65 °C) and let the paraffin contamination runoff.

There is a burn hazard when removingthe parts from the drying chamber(65 °C). Wearing safety gloves is rec-ommended!

• Dry knife holder and reassemble.• After cleaning the moving parts, apply a

thin coat of drive part oil to them (see alsoChapter 8.3).

• When installing the pressure plate (83) makesure that the upper edge is parallel and levelwith the rear edge of the knife holder baseplate (86).

8. Cleaning and maintenance

Knife holder E

Fig. 103

10

11

9

Fig. 104

86

83

83

86

87

11

10

If several knife holders are cleaned atthe same time, the parts must NOT bemixed up! Failure to adhere to this mayresult in sectioning problems!

Page 74: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

74 Instructions for use V 1.6 Rev A – 09/2011

Universal cassette clamp• Detach cassette clamp (13) for a thorough cleaning, removing all

paraffin residues.• For cleaning, do not use xylene. Use xylene substitutes or paraffin

removers such as "Para Gard."• The cassette clamp (13) can also be placed in an oven heated to a

maximum of 65 °C, until the liquid wax escapes.• Remove paraffin residues with a dry cloth.• After such a cleaning procedure in an oven, always be sure to lu-

bricate the spiral springs (91) of the tension lever (60) (refer also toChapter 8.3) and move it back and forth several times.

8. Cleaning and maintenance

Fig. 105

60

13

8.2 Maintenance instructions

Only authorized and qualified service personnel may access the internal components of the in-strument for service and repair!

The instrument is basically maintenance-free.To ensure trouble-free operation of the instrument over a long period of time, the following is rec-ommended by Leica:

3. Have the instrument inspected on a regularbasis by a qualified service technician au-thorized by Leica.The intervals depend on how heavily the in-strument is used.

1. Thoroughly clean the instrument on a dailybasis.

2. Enter into a service contract at the latest atthe end of the warranty period. For more in-formation, please contact your local Leicatechnical service center.

We recommend the following inspection intervals depending on the general workload of the instru-ment as defined in the two categories listed in the table:

Category I

> 8.000 sections per day

> 5 hours per day

Predominantly high sectioning speed

Working with soft and hardspecimens

Every 12 months

Number of sections per day:

Workload (hours per day):

Sectioning speed:

Specimen material:

Maintenance:

Category II

< 8.000 sections per day

< 5 hours per day

Slow to medium sectioning speed

Predominantly soft specimenmaterial

Every 24 months

91

Page 75: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

75Leica RM2235

8. Cleaning and maintenance

97

881110

Knife holder E:• Clamping lever (11) for the lateral displace-

ment.• T-piece (88) and guide (97) of the knife hold-

er for the lateral displacement.• Clamping lever (10) for the blade.

Fig. 106

Knife holders N and NZ:• Sliding surfaces of the finger guard (8).• Knurled nuts (73) for measuring height ad-

justment.

95

96

55

8.3 Lubricating the instrument

Once a month, lubricate the following partswith the included drive parts oil No. 405 (1 – 2drops are enough).

Instrument and specimen holder:• The clamping key (95) of the quick clamping

system.• The T-piece (55) on the microtome base

plate.• Guide rails (96) for the knife holder base on

the microtome base plate.• The T-piece on the knife holder base (56).

56

73

8

Knife holder N

Knife holder NZ

Page 76: RM2235 1v6 RevA en - Histopathology Service · Knife guard on the knife holder Each knife holder is equipped with a tightly mounted knife guard (8, 9). This makes it possi-ble to

76 Instructions for use V 1.6 Rev A – 09/2011

9. Warranty and service

Warranty

Leica Biosystems Nussloch GmbH guarantees that the contractual prod-uct delivered has been subjected to a comprehensive quality control pro-cedure based on the Leica in-house testing standards, and that the prod-uct is faultless and complies with all technical specifications and/oragreed characteristics warranted.The scope of the warranty is based on the content of the concludedagreement. The warranty terms of your Leica sales organization or the or-ganization from which you have purchased the contractual product shallapply exclusively.

Service information

If you are in need of technical customer service or spare parts, pleasecontact your Leica representative or the Leica dealer where you pur-chased the instrument.Please provide the following information:

• Model name and serial number of the instrument.• Location of the instrument and name of the person to contact.• Reason for the service call.• Delivery date.

Decommissioning and disposal

The instrument or parts of the instrument must be disposed of accordingto existing local applicable regulations.