58
GRAPHICAL ANALYSIS OF SYSTOLIC PRESSURE VARIATIONS AND RELATED NONINVASIVE INDICATORS OF BLOOD VOLUME STATUS. by Richard Stewart Shelton A thesis submitted to the faculty of the University of Utah in partial fulfillment of the requirements for the degree of Masters of Science in Bioengineering Department of Bioengineering University of Utah July 2002

Richard Shelton - School of Medicine - U of U School of

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Richard Shelton - School of Medicine - U of U School of

GRAPHICAL ANALYSIS OF SYSTOLIC PRESSURE VARIATIONS AND

RELATED NONINVASIVE INDICATORS OF BLOOD VOLUME STATUS.

by

Richard Stewart Shelton

A thesis submitted to the faculty of the University of Utah in partial fulfillment of the

requirements for the degree of

Masters of Science in Bioengineering

Department of Bioengineering

University of Utah

July 2002

Page 2: Richard Shelton - School of Medicine - U of U School of

Copyright © Richard Stewart Shelton 2002

All Rights Reserved

Page 3: Richard Shelton - School of Medicine - U of U School of

SUPERVISORY COMMITTEE APPROVAL FORM

Page 4: Richard Shelton - School of Medicine - U of U School of

FINAL READING APPROVAL FORM

Page 5: Richard Shelton - School of Medicine - U of U School of

ABSTRACT

Systolic Pressure Variation or SPV is a useful indicator of blood volume status.

SPV is characterized as an initial increase in systolic pressure followed by a subsequent

decrease in systolic pressure due to positive pressure ventilation. With blood loss, the

systolic pressure is further decreased and the overall variation in systolic pressure

increases.

SPV is similar to the variation of other indicators of blood volume status in that it

corresponds to a simultaneous variation in stroke volume. The cause of these variations

was studied through graphical analysis using different models of the circulation. These

models showed that positive intrathoracic pressure causes an initial increase in stroke

volume followed by a decrease in stroke volume, which causes the increase and decrease

in systolic pressure observed in SPV. However, these models were not able to predict a

greater decrease in stroke volume after blood loss. Therefore, a further decrease in the

systolic pressure was not predicted.

Nevertheless, some more recent studies of the compensation mechanisms of the

body affecting these models suggest that a further decrease in systolic pressure with

blood loss is possible. Currently, these compensation mechanisms are not well enough

understood to completely explain the further decrease in systolic pressure in SPV after

blood loss. Therefore, an animal study is proposed to better understand the compensation

mechanisms that cause the further decrease in systolic pressure with blood loss.

Page 6: Richard Shelton - School of Medicine - U of U School of

v

TABLE OF CONTENTS ABSTRACT………………………………………………………………………….….iv ACKNOWLEDGEMENTS……………………………………………………………..vi Chapter

I. SYSTOLIC PRESSURE VARIATIONS………………………………..1 II. PHOTO-PLETHYSMOGRAPHIC PULSE WAVEFORM VARIATIONS

……………………………………………………………………..…….7 III. CARDIOGENIC OSCILLATION VARIATION ……………………..10 IV. THE THEORY OF SPV ..………………………………………………15 V. ∆UP …………………………..………………………………………...17 VI. ∆DOWN……..………………………………………………………….21 Guyton’s Graphical Model …………………………………………….21 SPV is Caused by Positive Pressure Breaths .………………………….28 VII. SPV INCREASES WITH BLOOD LOSS ……………………………..30

The Effect of Blood Loss and Compensation Mechanisms on the Function Curves…………………………………………………………………..31

Will ∆down Increase with Blood Loss According to Guyton’s Model? ………………………………………………………………………….32 What Went Wrong?…………………………………………………….34

VIII. ADDITIONAL STUDIES OF PHOTO-PLETHYSMOGRAPHIC PULSE

WAVEFORM VARIATIONS……………………. …………………...38 IX. FURTHER WORK …………………………………………………….45 X. CONCLUSION………………………………………………..………..47

REFERENCES………………………………………………………………….………49

Page 7: Richard Shelton - School of Medicine - U of U School of

vi

ACKOWLEDGEMENTS

Page 8: Richard Shelton - School of Medicine - U of U School of

1

CHAPTER 1

SYSTOLIC PRESSURE VARIATIONS

What Is It?

Systolic Pressure Variation (SPV) is the variation in systolic blood pressure

during one positive pressure breath initiated by ventilation [1-13].

Figure 1. Systolic Pressure Variations (SPV). Perel A, Pizov R, Cotev S: Systolic blood

pressure variation is a sensitive indicator of hypovolemia in ventilated dogs subjected to

graded hemorrhage. Anesthesiology. 67: 499, 1987.

SPV is measured by first establishing a baseline systolic pressure during apnea

[1]. Then, after initiation of a positive pressure breath, the increase and decrease in

systolic pressure from this baseline is measured as ∆up and ∆down respectively [1].

Page 9: Richard Shelton - School of Medicine - U of U School of

2

Figure 2. ∆up and ∆down. Pizov R, Ya’ari Y, Perel A: Systolic pressure variation is

greater during hemorrhage than during sodium nitroprusside-induced hypotension in

ventilated dogs. Anesth Analg. 67:171, 1988.

In general, at the initiation of a positive pressure breath, the systolic pressure will

initially increase and ∆up is measured [1]. Then, within the same positive pressure

breath, the transient increase in systolic blood pressure is followed by a decrease in

systolic blood pressure below the baseline systolic blood pressure and ∆down is

measured [1].

Figure 3. Isolated breath with ∆up and ∆down. Perel A, Pizov R, Cotev S: Systolic

blood pressure variation is a sensitive indicator of hypovolemia in ventilated dogs

subjected to graded hemorrhage. Anesthesiology. 67: 499, 1987.

Page 10: Richard Shelton - School of Medicine - U of U School of

3

Measurement of SPV has been successfully used as a minimally invasive

indicator of blood volume status for over a decade [1-9]. Most of the initial successful

studies where SPV was used as an indicator of blood volume status were performed on

dogs [1-3]. In these studies done by Perel and associates, SPV was measured at normal

blood volume, and after blood loss of 5, 10, 20, and 30% of their estimated blood

volume. Then, SPV was measured again after retransfusion. As blood volume was

decreased, SPV and especially the ∆down portion of SPV, increased [1].

Figure 4. SPV at normal and decreased blood volume. Rooke GA, Schwid HA, Shapira

Y: The effect of hemorrhage and intravascular volume replacement on systolic pressure

variation in humans during mechanical and spontaneous ventilation. Anesth Analg. 80:

926, 1995.

In these first studies, increase in SPV was found to be a more accurate indicator

than other commonly used indicators of blood volume status [1-2]. Below are the results

of a study comparing SPV and ∆down with other indicators of blood volume status.

%SPV, shown below in the results, is the measured SPV divided by the systolic pressure

at apnea.

Page 11: Richard Shelton - School of Medicine - U of U School of

4

Figure 5. SPV Results. Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a

sensitive indicator of hypovolemia in ventilated dogs subjected to graded hemorrhage.

Anesthesiology. 67: 501, 1987.

Similar studies have confirmed the usefulness of SPV in monitoring blood

volume status in varied circumstances [4-13]. Some studies have even demonstrated the

usefulness of automating the measurement of SPV where the systolic pressure at the end

of a positive pressure breath was approximated as the baseline blood pressure at apnea

[14].

Page 12: Richard Shelton - School of Medicine - U of U School of

5

Figure 6. Schwid study comparison of end-expiration pressure to apnea pressure.

Schwid HA, Rooke GA: Systolic blood pressure at end-expiration measured by the

automated systolic pressure variation monitor is equivalent to systolic blood pressure

during apnea. Journal of Clinical Monitoring and Computing. 16: 118, 2000.

As a partial explanation of the origin of these variations in systolic pressure, Perel

and associates performed a study in which they concluded that variations in systolic

pressure were closely related to variations in stroke volume [15]. Furthermore, these

decreases in systolic pressure in ∆down were accompanied by simultaneous decreases in

stroke volume [15]. These results imply that the variations in systolic pressure are a

direct result of the variations in stroke volume beat to beat. Below are the results

showing the correlation between the variations in systolic pressure and stroke volume

measured through velocity time index (VTI) detected by a doppler probe.

Page 13: Richard Shelton - School of Medicine - U of U School of

6

Figure 7. Correlation of SPV and Stoke Volume Variation or % change in VTI.

Beaussier M, Coriat P, Perel A, Lebret F, Kalfon P, Chemla D, Lienhart A, Viars P:

Determinants of systolic pressure variation in patients ventilated after vascular surgery.

Journal of Cardiothoracic and vascular anesthesia. 9(5):550, 1995.

Page 14: Richard Shelton - School of Medicine - U of U School of

7

CHAPTER 2

PHOTO-PLETHYSMOGRAPHIC PULSE WAVEFORM VARIATIONS

These variations in stroke volume can also be detected through monitoring

variations in other indicators of hemodynamics. One of these is photo-plethysmographic

pulse waveform variations (PWV) or the variations of the waveforms of pulse oximetry.

The first study to show the usefulness of PWV in monitoring blood volume was

performed by Partridge and associates [16]. They showed that an increase in PWV was a

sensitive indicator of blood loss or hypovolemia [16]. An example of the change in PWV

during blood loss and fluid resuscitation is shown below. After blood loss, PWV

increased; and after fluid resuscitation, the PWV decreased to near its original variation.

Figure 8. Examples of PWV in Partridge study. Partridge BL: Use of pulse oximetry as

a noninvasive indicator of intravascular volume status. J Clin Monit. 3: 265, 1987.

In addition, Perel and associates also found PWV to be effective as an indicator of

blood volume status [17]. PWV was measured by Perel and associates in a similar way

Page 15: Richard Shelton - School of Medicine - U of U School of

8

to SPV [17]. The photo-plethysmographic pulse waveform was first recorded during

apnea [17]. From this waveform the signal strength of the peak plateau was recorded as a

baseline [17]. Next, at the initiation of a positive pressure breath, the signal strength from

the maximal peak to the apneic plateau was measured as the ∆up [17]. Finally, the signal

strength from the minimal peak to the apneic plateau was measured as the ∆down [17].

An example of PWV compared to SPV during normal blood volume and during

hypovolaemia is shown below.

Figure 9. Examples of PWV in Shamir study. Shamir M, Eidelman LA, Floman Y,

Kaplan L, Pizov R: Pulse oximetry plethysmographic waveform during changes in blood

volume. British Journal of Anaesthesia. 82(2): 180, 1999.

The results of their studies matched closely with the results of SPV studies. They

show that PWV is an effective monitor of blood volume status.

However, PWV has several advantages over SPV. Because pulse oximetry is less

invasive and more commonly monitored than an arterial line, PWV is more convenient

and more readily available than SPV [16]. Also, PWV introduces less injury and risk to

Page 16: Richard Shelton - School of Medicine - U of U School of

9

the patients. For these reasons, we saw PWV as a practical method of automatically and

continuously monitoring blood volume status during surgery.

Page 17: Richard Shelton - School of Medicine - U of U School of

10

CHAPTER 3

CARDIOGENIC OSCILLATION VARIATION

Another method of noninvasively monitoring variations in stroke volume is to

monitor the flow signal waveform of the ventilator providing positive pressure ventilation

to the animal. Often in these waveforms, oscillations can be found that correspond to the

heart rate of the animal.

-60-50-40-30-20-10

010203040

0 2 4 6 8 10Time (sec)

Flo

w (

L/m

in)

Figure 10. Cardiogenic Oscillations.

These cardiogenic oscillations are the result of the heart pumping and creating a

negative pressure sufficient to suck air into the lungs during a positive pressure breath.

The magnitude of these oscillations was found to correlate well to stroke volume in our

lab by a fellow student, Kai Breuger. An example of the correlation between these

cardiogenic oscillations and stroke volume are shown below.

Page 18: Richard Shelton - School of Medicine - U of U School of

11

0.0

10.0

20.0

30.0

40.0

50.0

60.0

70.0

80.0

90.0

1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39 41 43 45

indexstroke volume (ml of blood)

0.00

2.00

4.00

6.00

8.00

10.00

12.00

14.00

air volume (ml of air)

Thermodilution

Cardiogenic Oscillations

y = 0 . 1 1 4 8 x + 0 . 6 7 5 9R 2 = 0 . 5 4 0 1

0 . 0 0

2 . 0 0

4 . 0 0

6 . 0 0

8 . 0 0

1 0 . 0 0

1 2 . 0 0

1 4 . 0 0

0 . 0 1 0 . 0 2 0 . 0 3 0 . 0 4 0 . 0 5 0 . 0 6 0 . 0 7 0 . 0 8 0 . 0 9 0 . 0

s t r o k e v o l u m e ( m l o f b l o o d )

air

volu

me

(ml o

f ai

r)

Figure 11. Correlation between cardiogenic oscillations and stroke volume.

Furthermore, in one dog, the variations in the heights of these cardiogenic

oscillations or cardiogenic oscillation variations (COV) were found to increase after

blood loss.

Page 19: Richard Shelton - School of Medicine - U of U School of

12

8480 8500 8520 8540 8560 8580 8600 8620

24

26

28

30

32

34

36

flow

d28jun99flow16005750 5800 5850 5900

24

26

28

30

32

34

36

Before Blood Loss After Blood Loss

Figure 12. COV before and after blood loss.

In fact, COV increased until approximately 50% of the estimated blood volume

were shed. The values of COV with the approximate blood loss are shown below.

Estimated % Blood Loss

0% 19% 27% 75%

Variation Mean StDev Mean StDev Mean StDev Mean StDev

in Height 0.7 2.0 1.5 1.5 3.1 0.81 2.4 0.78

Table 1. The height of the COV as blood is removed.

COV

01234

0% 20% 40% 60% 80%

% Estimated Blood Loss

Hei

gh

t o

f C

OS

(L

/min

)

Figure 13. The height of the COV as blood is removed.

Monitoring COV was attempted in many other dogs and pigs and even on the

flow signal waveforms of several humans but often times the cardiogenic oscillations in

Page 20: Richard Shelton - School of Medicine - U of U School of

13

these waveforms were not identifiable. An example is shown below of a pig flow signal

waveform compared to a dog flow signal waveform.

3300 3400 3500 3600 3700 3800-50

-40

-30

-20

-10

0

10

20

30

40

flow

p06oct99flow1200

2250 2300 2350 2400 2450 2500 2550 2600 2650-50

-40

-30

-20

-10

0

10

20

30

d28jun99flow1600

flow

Pig Flow Signal Dog Flow Signal

Figure 14. Pig and Dog flow signal.

As you can see, the cardiogenic oscillations are not very prominent. One possible

reason that the cardiogenic oscillations are not prominent is that pigs and humans have a

different ratio of lung compliance to chest compliance than dogs. In fact, Perel and

associates note that dogs have a greater chest compliance relative to their lung

compliance than humans [1]. Because chest compliance in dogs is greater relative to

lung compliance than in pigs and humans, the lungs are more easily influenced by the

negative pressures created by the pumping of the heart in dogs. Thus, the effects of the

pumping of the heart in pigs and humans do not cause a noticeable oscillation on the flow

signal.

One other problem with monitoring the cardiogenic oscillations of the pigs and

humans is that the ventilation scheme was different than the dog studies. First of all,

constant pressure ventilation was used instead of constant flow ventilation. Constant

flow ventilation is needed to provide a stable baseline from which to measure cardiogenic

Page 21: Richard Shelton - School of Medicine - U of U School of

14

oscillations. However, constant flow is becoming a less popular method of ventilation

for various reasons [18]. The second problem with the ventilation scheme was that the

tidal volumes were too large and introduced pressures in the lungs that overwhelmed the

cardiogenic oscillations.

So, in summary, although monitoring COV is completely noninvasive,

continuously available in surgery, and a promising method of monitoring blood volume,

the cardiogenic oscillations were often not identifiable or discernable. Perhaps, if a

ventilation scheme were used that employed smaller, more frequent tidal volumes with

constant flow, cardiogenic oscillations would be more often identifiable and useful for

monitoring blood volume status.

Page 22: Richard Shelton - School of Medicine - U of U School of

15

CHAPTER 4

THE THEORY OF SPV

What causes SPV and the variation of other indicators and why do they increase

with blood loss? Perel gives one explanation [5]. His explanation is based on the

influence of the positive pressure breaths on the preload and afterload of the right and left

ventricle respectively. First, the increased intrathoracic pressure of the positive pressure

breath hinders the blood from entering the right ventricle and decreases the preload of the

right ventricle [5]. Secondly, after the initiation of the positive pressure breath, the

resistance of the pulmonary vasculature increases and thereby increases the afterload of

the right ventricle [1,5]. Eventually, both of these effects serve to decrease the stroke

volume of the right ventricle after one or two beats of the heart.

Another effect of the positive pressure breath is that the positive pressure

squeezes the blood out of the lungs and into the left ventricle [1,5]. The extra blood in

the left ventricle increases the preload of the left ventricle and the left ventricle stroke

volume initially increases. In addition, the systolic wall stress increases and the aortic

impedance decreases [1,5]. These effects serve to decrease left ventricular afterload and

also provide for the initial increase in stroke volume. In Summary, these effects cause an

initial increase in left ventricular stroke volume followed by a decrease in stroke volume

after one or two beats.

Perel continues to explain that the increase in SPV after blood loss is caused by a

further dependence of stroke volume on preload after blood loss [1,5]. Therefore, when

Page 23: Richard Shelton - School of Medicine - U of U School of

16

preload is decreased by the positive pressure breaths, the decrease in stroke volume is

greater after blood loss. Thus, ∆down and SPV as a whole are increased.

These theories may all be true. However, they have never been tested by

scientific principles. First of all, an explanation based only on afterload and preload

depends on a number of questionable assumptions like constant twitch tension and

contractility for different preloads. They also ignore the role of the body’s vasculature in

providing venous return and preload to the right ventricle. In reality, the cardiac output

of the heart and the venous return of the vasculature are mutually dependent on each

other. Therefore, an adequate explanation of SPV should describe the effects of positive

pressure breaths on both the heart function and the vascular function.

In the following chapters, a graphical analysis of the occurrence of SPV and its

relationship to blood volume is attempted.

Page 24: Richard Shelton - School of Medicine - U of U School of

17

CHAPTER 5

∆UP

First, an explanation of ∆up is here supplied. ∆Up in particular represents a

transient increase in systolic pressure. Perel proved in his study of stroke volume

variation that the variation of systolic pressure in SPV is closely related to the variation

of stroke volume [15]. Thus, this transient increase in systolic pressure of ∆up followed

by a decrease in systolic pressure is indicative of a temporary augmentation of stroke

volume caused by a positive pressure breath. This augmentation is followed by a

decrease in stroke volume and systolic pressure that is representative of the steady state

response of the circulation to a positive pressure breath. The transient response of the

increased stroke volume can be explained by relating positive pressure breathing to

another study.

This study done by Bromberger-Barnea explains the transient response of the

circulation due to changes in intrathoracic pressure [19]. Bromberger-Barnea showed

that according to graphical analysis, a decrease in intrathoracic pressure causes a decrease

in stroke volume [19]. Bromberger-Barnea’s graphical analysis is shown below. The

first graph is a representation of a normal beat of the heart and the other three graphs are

representations of a beat of the heart under different negative intrathoracic pressures.

Page 25: Richard Shelton - School of Medicine - U of U School of

18

Figure 15. Bromberger-Barnea’s graphical analysis of decreased intrathoracic

pressure. Bromberger-Barnea B: mechanical effects of inspiration on heart functions: a

review. Federation Proc. 40: 2173, 1981.

Bromberger-Barnea then confirmed her analysis through experiments that

decreased the intrathoracic pressure and thus decreased the stroke volume [19]. The

results one of those experiments is shown below. As can be seen, the QAO or cardiac

output decreases with decreased intrathoracic pressure. This effect is not shown to be

transient here because the QPA or the cardiac output of the right heart is held constant.

Page 26: Richard Shelton - School of Medicine - U of U School of

19

Figure 16. Experiment confirming graphical analysis of decreased intrathoracic

pressure. Bromberger-Barnea B: mechanical effects of inspiration on heart functions: a

review. Federation Proc. 40: 2176, 1981.

Therefore, using a similar graphical analysis of increased intrathoracic pressure

shows that stroke volume should transiently increase with positive pressure breaths. A

graphical analysis of an increase in intrathoracic pressure of approximately 9-mmHg is

shown below.

Page 27: Richard Shelton - School of Medicine - U of U School of

20

0 20 40 60 80 100 1200

20

40

60

80

100

120

140

160

PL

V-P

PL

(m

mH

g)

LV VOL. (ml)

AA1BB1

C

C1

D

D1

Increase in SV = 7%

PLAM = 7.2mm Hg

Increased Intrathoracic Pressure

Figure 17. Graphical analysis of increased intrathoracic pressure.

This graph shows an increase of stoke volume of approximately 7% with an

increase in intrathoracic pressure of approximately 9-mmHg. This increase is similar to

the increase in stoke volume and systolic pressure of approximately 3-7% in SPV [1-

3,17]. Therefore, we can see that the ∆up of SPV or the transient increase in stroke

volume is caused by the increase in intrathoracic pressure of positive pressure breathing.

Page 28: Richard Shelton - School of Medicine - U of U School of

21

CHAPTER 6

∆DOWN

In order to understand the ∆down portion of SPV, the steady state response of the

circulation must be explored. This response is observed after the transient augmentation

of stroke volume. Guyton and associates describe a model of the circulation that

provides a graphical analysis of the steady state response of the circulation to

physiological changes [20-26].

Guyton’s Graphical Model

Guyton’s representation of the circulation is based on two curves, the Cardiac

Function Curve and the Vascular Function Curve, both shown below [20-22].

Figure 18. Cardiac and Vascular Function Curves. Guyton AC: Graphical analysis of

cardiac output regulation. In: Circulatory Physiology: Cardiac Output and Its

Regulation. Saunders: Philadelphia, Pennsylvania, 238, 1973.

Page 29: Richard Shelton - School of Medicine - U of U School of

22

These curves represent the function of the heart and the vasculature of the body

that Guyton and associates established through varied experiments [20,22]. The

experiments involved isolating the heart from the vasculature and testing their cardiac

output, venous return, and respective atrial pressures independent of each other [20]. An

example of their experiments is shown below.

Figure 19. Guyton’s Experiments. Guyton AC: Effect of right atrial pressure on venous

return – the normal venous return curve. In: Circulatory Physiology: Cardiac Output

and Its Regulation. Saunders: Philadelphia, Pennsylvania, 189, 1973.

The Cardiac Function Curve represents the ability of the heart to deliver a specific

cardiac output to the body at a particular atrial pressure [21].

Page 30: Richard Shelton - School of Medicine - U of U School of

23

Figure 20. Cardiac Function Curve. Modified from Guyton AC: Graphical analysis of

cardiac output regulation. In: Circulatory Physiology: Cardiac Output and Its

Regulation. Saunders: Philadelphia, Pennsylvania, 238, 1973.

The Vascular Function Curve represents the ability of the vasculature to deliver a

specific venous return to the heart at a particular atrial pressure [22].

Page 31: Richard Shelton - School of Medicine - U of U School of

24

Figure 21. Vascular Function Curve. Modified from Guyton AC: Graphical analysis of

cardiac output regulation. In: Circulatory Physiology: Cardiac Output and Its

Regulation. Saunders: Philadelphia, Pennsylvania, 238, 1973.

Just as the venous return and cardiac output must equal each other in steady state,

the steady state response of the circulation to the body’s physiological state is determined

by the intersection of these two curves [25]. At the intersection of these two curves, we

can see the atrial pressure and the cardiac output or the venous return [25]. An example

is shown below.

Page 32: Richard Shelton - School of Medicine - U of U School of

25

Figure 22. Intersection of the Cardiac and Vascular Function Curves. Guyton AC:

Graphical analysis of cardiac output regulation. In: Circulatory Physiology: Cardiac

Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 238, 1973.

These curves change in predictable ways in response to physiological changes

[21-26] Physiological effects that were determined through experimentation by Guyton

and associates include sympathetic stimulation, cardiac hypertrophy, decreased load on

the heart, myocardial infarction, parasympathetic stimulation, decreased vascular

resistance, changes in vascular capacitance, blood loss, transfusion, positive pressure

breathing, negative pressure breathing, opening of the chest, muscular compression of the

vascular system, and vasomotor tone [21-26]. In response to higher sympathetic

stimulation, the Cardiac Function Curve is augmented due to the increased contractility

of the heart [21]. Shown below is the augmentation of the Cardiac Function Curve by

sympathetic stimulation.

Page 33: Richard Shelton - School of Medicine - U of U School of

26

Figure 23. Effects of sympathetic stimulation on the Cardiac Function Curve. Guyton

AC: Patterns of cardiac output curves. In: Circulatory Physiology: Cardiac Output and

Its Regulation. Saunders: Philadelphia, Pennsylvania, 163, 1973.

The effect of an instantaneous change in blood volume is to shift the Vascular

Function Curve right or left and increase or decrease the slope. The Vascular Function

Curve shifts to the right and increases slope for transfusion and shifts to the left and

decreases slope for blood loss [23,26]. The reason that the curve shifts is that mean

circulatory pressure changes with blood volume changes [23]. Guyton tells us that the

slope changes because with greater volume the vasculature distends and the resistance to

flow decreases [23,26]. Below is a graph showing the effects of instantaneous changes in

blood volume.

Page 34: Richard Shelton - School of Medicine - U of U School of

27

Figure 24. Effects of instantaneous blood volume changes on the Vascular Function

Curve. Guyton AC: Effect of blood volume changes and orthostatic factors on cardiac

output. In: Circulatory Physiology: Cardiac Output and Its Regulation. Saunders:

Philadelphia, Pennsylvania, 357, 1973.

Another physiological change that we will discuss is the effects of positive

pressure breathing [21,27]. Positive pressure causes mostly a shift in the Cardiac

Function Curve to the right and a slight decrease in slope of the Venous Return Curve

due to a slight increase in resistance in the vasculature [21,27]. The results of the

experiments done with positive pressure are shown below.

Page 35: Richard Shelton - School of Medicine - U of U School of

28

Figure 25. Effects of positive pressure on the Cardiac and Vascular Function Curve.

Fermoso JD, Richardson TQ, Guyton AC: Mechanism of decrease in cardiac output

caused by opening of the chest. Am J Physiol. 207(5): 1115, 1964.

SPV is Caused by Positive Pressure Breaths

The positive pressure used in the experiments above was equivalent to

approximately 4-mmHg [27]. According to Gilbert and Glantz, the contact pressure

associated with positive pressure ventilation with a typical tidal volume of 15-20 ml/kg is

less than 2-mmHg [28]. Thus, the effects of positive pressure in positive pressure

ventilation are less than half of those shown on the graph above. So, using extrapolation

of halfway in between the two curves, the following curves were acquired.

Page 36: Richard Shelton - School of Medicine - U of U School of

29

Figure 26. Approximate effects of positive pressure of positive pressure ventilation on

function curves.

There is an approximate decrease of 160 cc/min in cardiac output or stroke

volume from point A to point C. This is a decrease of about 9.4%. This would

correspond to an approximate 9.4% decrease in systolic pressure in SPV. The actual

decrease in systolic pressure of SPV at normal blood volume is approximately 5-12%

depending on the tidal volume used, which matches fairly closely [1-3].

Page 37: Richard Shelton - School of Medicine - U of U School of

30

CHAPTER 7

SPV INCREASES WITH BLOOD LOSS

Guyton and associates also conducted experiments with changes in blood volume

[26]. The effects of transfusion on the function curves are shown below.

Figure 27. Effects of transfusion on function curves. Guyton AC: Effect of blood volume

changes and orthostatic factors on cardiac output. In: Circulatory Physiology: Cardiac

Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 363, 1973.

The first effect is that the Vascular Function Curve shifts to the right and

increases slope because of the increased blood volume, shown at point B [26]. Then, the

compensatory effects of the body to the increased blood volume cause further changes to

the curves [26]. One of these changes is a reflex relaxation of the vasculature and the

heart [26]. This relaxation shifts the Vascular Function Curve to the left and decreases

in slope while the Cardiac Function Curve shifts down and to the right, shown at point C

Page 38: Richard Shelton - School of Medicine - U of U School of

31

[26]. After ten minutes, the curves further change because of stress relaxation, shown at

point D [26]. In response to stress relaxation, the Vascular Function Curve shifts further

to the left and further decreases in slope [26]. The Cardiac Function Curve recovers a bit

and shifts back up and to the left [26].

The Effect of Blood Loss and Compensation Mechanisms on the Function Curves

The effects of blood loss on the function curves are nearly opposite that of

transfusion. First, the Vascular Function Curve shifts to the left and decreases in slope

due to the decrease in blood volume [23,26]. This is due to the reduction in mean

circulatory pressure and the increase in vascular resistance [23-24,26]. The

compensation mechanisms of the body for blood loss include sympathetic stimulation,

nervous reflexes, and slower compensations of stress relaxation recovery, and

readjustment of blood volume [26]. The effect of the sympathetic stimulation is to

increase the contractility of the heart and augment the Cardiac Function Curve. The

nervous reflex of the vasculature is to constrict. This causes an increase in mean

circulatory pressure and a shift to the right of the Vascular Function Curve. The stress

relaxation recovery has the effect of the vasculature recovering its elasticity, mostly in the

veins, and more tightly clamping down on the reduced blood volume of the body [26].

This also has the effect of returning the Vascular Function Curve more towards normal

[26]. The readjustment of blood volume involves the movement of fluid from the tissues

to the plasma because of the greater osmalarity of the plasma after blood loss [26,29-31].

Guyton also explains that this effect serves to return the Vascular Function Curve

towards normal [26]. The overall change in the function curves after 10% blood loss was

Page 39: Richard Shelton - School of Medicine - U of U School of

32

determined from graphical analysis of the curves provided by Guyton and associates.

This overall change is shown at point D in the figure below.

Figure 28. Graphical analysis of the effects of 10% blood loss on the function curves.

Will ∆down Increase with Blood Loss According to Guyton’s Model?

In order for ∆down to increase with blood loss, the change in stroke volume due

to positive pressure after blood loss must be greater than the change in stroke volume due

to positive pressure at normal blood volume. Because the positive pressure shifts the

Cardiac Function Curve the same amount at both blood volume states, a greater change

would require an increase in the slopes of the functional curves after blood loss.

However, in the model shown above, there is an overall decrease in slope between the

two curves, even though the Cardiac Function Curve does increase in slope slightly.

Furthermore, to compare ∆down after 10% blood loss to normal blood volume, a

Page 40: Richard Shelton - School of Medicine - U of U School of

33

graphical analysis of the effects of positive pressure ventilation on the 10% blood loss

was done similar to the one for normal blood volume. Below is an approximation of the

effects of positive pressure on the function curves after 10% blood loss. The point D1

represents the cardiac output with positive intrathoracic pressure.

Figure 29. Graphical analysis of the effects of positive pressure on the functional curves

after 10% blood loss.

The decrease in stroke volume in this case is 140 cc/min. This decrease of 140

cc/min is actually less than the decrease in stroke volume of 160 cc/min at normal blood

volume. These decreases in stroke volume are approximately proportional to ∆down for

each case respectively. Therefore, after 10% blood loss, ∆down would have actually

decreased by approximately 10%. This is exactly opposite the result we expected.

Below is shown a comparison of the decrease in stroke volume for normal blood volume

and the decrease in stroke volume after 10 % blood loss. As can be seen through careful

Page 41: Richard Shelton - School of Medicine - U of U School of

34

measuring, the decrease in stroke volume from A to A1 is greater than the decrease from

D to D1.

Figure 30. Comparison of the effects of positive pressure ventilation on the normal

function curves and the function curves after 10% blood loss.

What Went Wrong?

Guyton admits that our understanding of the compensation mechanisms to blood

loss is only partially understood [26]. In later publications, Holcroft and Jacobsohn

provide a greater understanding of the compensation mechanisms for blood loss [30-31].

They both discuss how compensation mechanisms of sympathetic stimulation,

vasoconstriction, stress relaxation, and readjustment of blood volume all serve to

augment cardiac output and venous return for acute blood loss [30-31]. However, they

note that these mechanisms fail to augment cardiac output after severe blood loss [30-31].

In addition, they both mention that after some time the movement of the fluid from the

Page 42: Richard Shelton - School of Medicine - U of U School of

35

tissues and interstitial spaces to the plasma serves to dilute the blood and make the blood

less viscous [30-31]. Holcroft states that this reduces the afterload on the heart and

further augments the Cardiac Function Curve [30]. Jacobsohn goes even further to say

that this dilution of the blood serves to reduce the resistance of the blood to flow and

increases the slope of the Vascular Function Curve [31]. If the Vascular Function Curve

did actually increase in slope for acute blood loss, then the slopes for both functional

curves would increase. If both function curves increase in slope, then positive pressure

ventilation would cause a greater decrease in stroke volume after blood loss than at

normal blood volume. This would cause ∆down to increase with blood loss as was

observed by Perel and many others.

Unfortunately, no data exists which show the Vascular Function Curve with an

increase in slope after blood loss. But, there are studies that suggest that the vascular

resistance decreases and therefore the slope should increase [29-31]. Jacobsohn provides

a graph with his publication that shows the effects of blood loss on the function curves,

which is shown below.

Page 43: Richard Shelton - School of Medicine - U of U School of

36

Figure 31. Functional curves provided by Jacobsohn. Jacobsohn E, Chorn R, O’Conner

M: The role of the vasculature in the regulating venous return and cardiac output:

historical and graphical approach. Can J Anaesth. 22(8): 859, 1997.

However, he does not state the amount of blood loss that he is assuming and does

not infer that the body has had enough time for the slow effect of blood readjustment to

take place. Nevertheless, if we use his graph and apply the same graphical analysis for

positive pressure ventilation that we performed previously on Guyton’s graphs, we find

that stroke volume or cardiac output is indeed decreased more after blood loss. This

means that ∆down increases with blood loss as was observed by Perel and others.

Page 44: Richard Shelton - School of Medicine - U of U School of

37

Figure 32. Graphical analysis of the effects of positive pressure on the normal function

curves and the function curves after blood loss.

As can be seen through careful measuring, delta2 is greater than delta1. The

stroke volume actually decreased approximately 45% more after blood loss than at

normal blood volume. This greater decrease in stroke volume accounts for the greater

∆down and greater SPV after blood loss than at normal blood volume.

Page 45: Richard Shelton - School of Medicine - U of U School of

38

CHAPTER 8

ADDITIONAL STUDIES OF PHOTO-PLETHYSMOGRAPHIC PULSE WAVEFORM

VARIATIONS

In our lab we performed our own experiments which attempted to quantify the

effects of blood loss on photo-plethysmographic pulse waveform variations. Our study

attempted to establish a continuous monitor of blood volume status by only monitoring

the total PWV and not establishing a baseline plateau of peak plethysmographic pulse

signal at apnea. The protocol of our study was simply to continuously record the best

available PWV on a tongue, foot or ear of the animal. Then, the animal was bled of 75%

of their estimated blood volume.

Results

The results of our study turned out to be somewhat different than what was

expected. We expected that the PWV would increase with blood loss at least until about

30% of the estimated blood volume was shed. However, in a study of four animals, the

PWV held constant or increased for a time then dramatically decreased. An example of

this dramatic decrease in PWV is shown below.

Page 46: Richard Shelton - School of Medicine - U of U School of

39

Before Blood Loss After Blood Loss

Figure 33. Example of dramatic decrease in PWV after blood loss.

This dramatic decrease happened at various estimated blood losses, sometimes

before even 10% blood loss. The magnitude of PWV and %PWV at various estimated

blood losses are shown below for all four animals.

POWV 28-Jun 16-Sep 23-Sep 6-Oct Normal 5416.667 1111 1461.333 108

10% 3969.667 1071 1461.333 139.5333 25% 2727.5 838.3333 1132.667 144.0667 50% 2336.667 621 452.3333 94.53333 75% 726.6667 550.5 409.5 41.01333

Table 2. The magnitude of PWV at various blood volumes.

Page 47: Richard Shelton - School of Medicine - U of U School of

40

PWV 28-Jun

01000

20003000

40005000

6000

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

Mag

nit

ud

e o

f P

WV

PWV 16-Sep

0200400600800

10001200

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

Mag

nit

ud

e o

f P

WV

PWV 23-Sep

0

200

400

600

800

1000

1200

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

Mag

nit

ud

e o

f P

WV

PWV 6-Oct

0

50

100

150

200

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

Mag

nit

ud

e o

f P

WV

Figure 34. The magnitude of PWV at various blood volumes.

%POWV 28-Jun 16-Sep 23-Sep 6-Oct Normal 0.023807 0.02221 0.044942 0.003553

10% 0.017056 0.020549 0.045624 0.00443 25% 0.011883 0.015753 0.036431 0.004352 50% 0.009376 0.010462 0.013493 0.002615 75% 0.00651 0.00778 0.013829 0.001079

Table 3. The %PWV at various blood volumes.

Page 48: Richard Shelton - School of Medicine - U of U School of

41

%PWV 28-Jun

00.0050.01

0.0150.02

0.025

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

%P

WV

%PWV 16-Sep

0

0.0050.01

0.0150.02

0.025

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

%P

WV

%PWV 23-Sep

0

0.010.020.030.04

0.05

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)

%P

WV

%PWV 6-Oct

0

0.0010.002

0.0030.004

0.005

0 20 40 60 80

Estimated Blood Loss (% Blood Volume)%

PW

V

Figure 35. The %PWV at various blood volumes.

In the first animal, the PWV decreased immediately. In the second animal, the

PWV held constant until 10% blood loss. In the third animal, the %PWV increased until

10% blood loss. And finally, in the last animal, PWV increased until 25% blood loss.

Discussion

As discussed above, Holcroft and Jacobsohn state that some major compensation

mechanisms are only effective in augmenting the Cardiac Function Curve and the

Vascular Function Curve for acute and moderate blood loss [30-31]. In a more severe

state of hypovolemic shock, these mechanisms are no longer effective and the slopes of

the function curves will decrease [30-31]. This will cause PWV to decrease also.

However, we did not expect these effects to happen so soon after initiating bleeding as

happened in our study.

Page 49: Richard Shelton - School of Medicine - U of U School of

42

Holcroft and Jacobsohn both mention that these compensation mechanisms also

have a slow response time [30-31]. In some animals of our study, these compensation

mechanisms were not allowed to take effect because we bled the animals too quickly.

The second animal in particular was bled 75% of the estimated blood volume in only 5

minutes. The third and fourth animals were bled a little slower at 8 and 9 minutes,

respectively. In contrast, Perel waited 15 minutes after each blood loss before measuring

SPV in his studies [1]. This explains why in the second animal PWV never increases and

dramatically decreases after only 25% blood loss.

Another of the biggest problems that we had in our study was that we could not

get a consistent signal strength on the photo-plethysmographic signal. This caused

problems in that the magnitude of PWV varied with the signal strength, which would

change when bumping, moving, or otherwise changing the position of the animal. In

addition, another compensation mechanism of the body not mentioned above is that

during blood loss, we can expect the body to react to hypovolemic shock by shutting

down blood perfusion to the periphery or our site of monitoring PWV. This would cause

the signal strength of the photo-plethysmographic signal to decrease and thus decrease

the measured PWV. This undoubtedly caused the PWV to decrease sooner than

expected. Furthermore, the signal strength on a fifth animal was so weak that the pulse

could never be determined through the signal and the results were not used in this study.

Interestingly, the catheter of the first animal was changed shortly after the initiation of

bleeding. This dramatically changed the position of the animal and caused the signal

strength to greatly decrease. This may have been responsible for the immediate decrease

in PWV in the first animal.

Page 50: Richard Shelton - School of Medicine - U of U School of

43

In summary, the PWV or %PWV increased in two animals until 25% or 10%

blood loss respectively. In the other two animals, the PWV decreased almost

immediately due to either bleeding the animal too fast or decreasing the signal strength

by dramatically changing the position of the animal. In the animal that PWV did

increase, the increase in PWV at 10% blood loss was approximately 28.8%.

In addition, a graphical analysis was done for the effects of a positive pressure

breath after a 75% blood loss without the compensation effects present after acute blood

loss. The graph below shows how the function curves change after 75% blood loss

without the effects of compensation mechanisms.

Figure 36. The change in the function curves after 75% blood loss without compensation

mechanisms.

Page 51: Richard Shelton - School of Medicine - U of U School of

44

The graphical analysis of the effects of positive pressure ventilation on the

function curves of 75% blood loss without compensation mechanisms is also shown

below.

Figure 37. Graphical analysis of the effects of positive pressure on the function curves

after 75% blood loss but with no compensation mechanisms.

The point B1 represents the decrease in stroke volume with positive pressure.

This analysis showed that in comparison to normal blood volume, the stroke volume

decreased approximately 62% less after 75% blood loss. Thus, PWV should also

decrease by approximately 62%. This is similar to the average 67% decrease in PWV

after 75% blood loss found in our studies.

In the end, our lab abandoned the pursuit of a continuous monitor of blood

volume through PWV because of the inconsistencies in the signal strength of photo-

plethysmographic signal.

Page 52: Richard Shelton - School of Medicine - U of U School of

45

CHAPTER 9

FURTHER WORK

In order to better understand SPV and other similar noninvasive indicators of

monitoring blood volume, we must better understand the compensation mechanisms that

effect these indicators. To better understand these compensation mechanisms, we must

understand how they affect the function curves and the response times of the curves to

these compensation mechanisms.

An effective way to obtain this understanding of these mechanisms would be to

perform a new set of experiments on animal models. First of all, in an animal model, the

Cardiac Function Curve and the Vascular Function Curve would be established using the

experimental methods used by Guyton and associates [20-22]. Also, the SPV and PWV

would be recorded using Perel’s method [1,17]. Then, we would bleed the animal of a

10% of their blood volume and allow time for the compensation mechanisms to take

effect. Next, the SPV and PWV would be measured using Perel’s methods to confirm an

increase in SPV and PWV. Finally, the Cardiac Function Curve and the Vascular

Function Curve would be determined again through Guyton’s methods to determine the

change in the curves that caused SPV and PWV to increase. This process could be

repeated again for 25% blood loss and greater blood loss to establish when these

mechanisms were no longer effective and no longer caused SPV and PWV to increase.

In addition, the function curves could be determined at different times after each blood

loss to determine how long before these mechanisms were effective in increasing SPV

and PWV.

Page 53: Richard Shelton - School of Medicine - U of U School of

46

In this way, our understanding of blood loss, compensation mechanisms, and

indicators of blood volume status would be greatly enhanced. This understanding would

help us better diagnose blood loss noninvasively and help us to better treat it effectively.

Page 54: Richard Shelton - School of Medicine - U of U School of

47

CHAPTER 10

CONCLUSION

Positive Pressure Ventilation causes a transient increase in systolic pressure

termed ∆up and a subsequent decrease in systolic pressure termed ∆down. The increase

in systolic pressure can be explained by a temporary augmentation of stroke volume.

This augmentation is caused by the effects of increased intrathoracic pressure better

understood through a graphical analysis similar to those done by Bromberger-Barnea on

decreased intrathoracic pressure. The decrease in systolic pressure can be explained by

graphical analysis using the model of the circulation established by Guyton and

associates. This model shows that the steady state response of the circulation to positive

intrathoracic pressure is a decrease in stroke volume and thus a similar decrease in

systolic pressure.

However, the current models of circulation do not adequately explain the greater

decrease in stroke volume and the increase in SPV with blood loss. Guyton’s model

actually predicts a decrease in SPV with blood loss. Later revisions of this model imply

that an increase in SPV with blood loss is possible through compensation mechanisms

that increase the slope of both the Cardiac Function Curve and the Vascular Function

Curve. However, the increase in these slopes have never been quantified, therefore a

concrete conclusion as to the cause of the increase in SPV is not currently available.

In addition, a similar noninvasive monitor of blood volume, PWV, was found to

often decrease with blood loss. This increase was caused by not allowing compensation

Page 55: Richard Shelton - School of Medicine - U of U School of

48

mechanisms to have an effect on PWV or through a decrease in signal strength due to

either moving the animal or the shut down of peripheral blood perfusion.

These compensation mechanisms could be greater understood through further

studies into the function curves after varying degrees of blood loss. These studies could

assist in more effectively diagnosing and treating blood loss.

Page 56: Richard Shelton - School of Medicine - U of U School of

49

REFERENCES

1. Perel A, Pizov R, Cotev S: Systolic blood pressure variation is a sensitive indicator of hypovolemia in ventilated

dogs subjected to graded hemorrhage. Anesthesiology. 67: 498-502, 1987.

2. Pizov R, Ya’ari Y, Perel A: Systolic pressure variation is greater during hemorrhage than during sodium

nitroprusside-induced hypotension in ventilated dogs. Anesth Analg. 67: 170-174, 1988.

3. Szold A, Pizov R, Segal E, Perel A: The effect of tidal volume and intravascular volume state on systolic pressure

variation in ventilated dogs. Intensive Care Medicine. 15: 368-371, 1988.

4. Pizov R, Ya’ari Y, Perel A: The arterial pressure waveform during acute ventricular failure and synchronized

external chest compression. Anesth Analg. 68: 150-156, 1989.

5. Perel A: Cardiovascular assessment by pressure waveform analysis. ASA Refresher Course Handbook. 264: 1-7,

1991.

6. Marik PE: The systolic blood pressure variation as an indicator of pulmonary capillary wedge pressure in

ventilated patients. 21: 405-408, 1993.

7. Rooke GA, Schwid HA, Shapira Y: The effect of hemorrhage and intravascular volume replacement on systolic

pressure variation in humans during mechanical and spontaneous ventilation. Anesth Analg. 80: 925-932, 1995.

8. Preisman S, DeSigni E, Vered Z, Perel A: Left ventricular preload and function during graded haemorrhage and

retransfusion in pigs: analysis of arterial pressure waveform and correlation with echocardiography. British

Journal of Anaesthesia. 88(5): 716-718, 2002.

9. Pizov R, Cohen M, Weiss Y, Segal E, Cotev S, Perel A: Positive end-expiratory pressure-induced hemodynamic

changes are reflected in the arterial pressure waveform. Crit Care Med. 24(8): 1381-1387, 1996.

10. Pinsky MR: The hemodynamic consequences of mechanical ventilation: an evolving story. Intensive Care Med.

23: 493-503, 1997.

11. Vedrinne JM, Duperret S, Decaillot F, Gratadour P, Motin J: Haemodynamic changes induced by two I:E ratios: a

transoesophageal echocardiographic study. Can J Anaesth. 44(4): 354-359, 1997.

12. Perel A: Assessing fluid responsiveness by the systolic pressure variation in mechanically ventilated patients.

Anesthesiology. 89: 1309-1310, 1998.

13. Dalibon N, Schlumberger S, Saada M, Fischler M, Riou B: Haemodynamic assessment of hypovolaemia under

general anaesthesia in pigs submitted to graded haemorrhage and retransfusion. British Journal of Anaesthesia.

82(1): 97-103, 1999.

Page 57: Richard Shelton - School of Medicine - U of U School of

50

14. Schwid HA, Rooke GA: Systolic blood pressure at end-expiration measured by the automated systolic pressure

variation monitor is equivalent to systolic blood pressure during apnea. Journal of Clinical Monitoring and

Computing. 16: 115-120, 2000.

15. Beaussier M, Coriat P, Perel A, Lebret F, Kalfon P, Chemla D, Lienhart A, Viars P: Determinants of systolic

pressure variation in patients ventilated after vascular surgery. Journal of Cardiothoracic and vascular anesthesia.

9(5):547-551, 1995.

16. Partridge BL: Use of pulse oximetry as a noninvasive indicator of intravascular volume status. J Clin Monit. 3:

263-268, 1987.

17. Shamir M, Eidelman LA, Floman Y, Kaplan L, Pizov R: Pulse oximetry plethysmographic waveform during

changes in blood volume. British Journal of Anaesthesia. 82(2): 178-181, 1999.

18. Natalini G, Facchetti P, Dicembrini MA, Lanza G, Rosana A, Baruardini A: Pressure controlled versus volume

controlled ventilation with laryngeal mask airway. J Clin Aesth. 12(6):436-9, 2001.

19. Bromberger-Barnea B: mechanical effects of inspiration on heart functions: a review. Federation Proc. 40: 2172-

2177, 1981.

20. Guyton AC: The pumping ability of the heart as expressed by cardiac function curves. In: Circulatory Physiology:

Cardiac Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 147-157, 1973.

21. Guyton AC: Patterns of cardiac output curves. In: Circulatory Physiology: Cardiac Output and Its Regulation.

Saunders: Philadelphia, Pennsylvania, 158-170, 1973.

22. Guyton AC: Effect of right atrial pressure on venous return – the normal venous return curve. In: Circulatory

Physiology: Cardiac Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 188-190, 1973.

23. Guyton AC: Mean circulatory pressure, mean systemic pressure, and mean pulmonary pressure and their effect on

venous return. In: Circulatory Physiology: Cardiac Output and Its Regulation. Saunders: Philadelphia,

Pennsylvania, 205-221, 1973.

24. Guyton AC: Effect of peripheral resistance and capacitance on venous return. In: Circulatory Physiology: Cardiac

Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 222-233, 1973.

25. Guyton AC: Graphical analysis of cardiac output regulation. In: Circulatory Physiology: Cardiac Output and Its

Regulation. Saunders: Philadelphia, Pennsylvania, 237-249, 1973.

26. Guyton AC: Effect of blood volume changes and orthostatic factors on cardiac output. In: Circulatory Physiology:

Cardiac Output and Its Regulation. Saunders: Philadelphia, Pennsylvania, 353-364, 1973.

27. Fermoso JD, Richardson TQ, Guyton AC: Mechanism of decrease in cardiac output caused by opening of the

chest. Am J Physiol. 207(5): 1112-1116, 1964.

Page 58: Richard Shelton - School of Medicine - U of U School of

51

28. Gilbert JC, Glantz SA: Determinants of left ventricular filling and of the diastolic pressure-volume relation.

Circulation Research. 64(5): 827-852, 1989.

29. Rothe CF, Drees JA: Vascular capacitance and fluid shifts in dogs during prolonged hemorrhagic hypotension.

Circulation Research. 38(5): 347-356, 1976.

30. Holcroft JW: Impairment of Venous Return in Hemorrhagic Shock. Surg Clin of North Amer. 62(1): 17-29,

1982.

31. Jacobsohn E, Chorn R, O’Conner M: The role of the vasculature in the regulating venous return and cardiac

output: historical and graphical approach. Can J Anaesth. 22(8): 849-867, 1997.