Revisimo
Remote Vital Sign Monitor
About us
Dr. Thomas Morris
Advisor
CoryZywno
EEData Transmission
MatthewGalloway
EEADC & Temperature Sensor
MatthewGallups
EEAlarm & ADC
AaronCheeseman
EEMonitor &PCB design
PaulJang
CPEData Transmission& Website
Overview
• Introduction• Problem/Solution• Constraints
– Technical– Practical
• Technical Approach• Progress• Timeline• Questions
Introduction
• What is Revisimo?– Digital wireless Thermometer
Current Problem
• Hospital equipment– Expensive– Hard to use
• Market (oral, tympanic, scanning)– Inaccurate & No Alarm– Difficulty of Continuously Monitoring
Solution
• Bandage– Wireless– Temperature Sensing Device
• Base station– Alarm– Temperature set by the user
System Overview
Constraints
• Revisimo must meet a set of predetermined technical and practical constraints ensuring that the device will be competitive in the home-healthcare market.
• Technical Constraints outline Revisimo’s functionality.
• Practical Constraints define usability.
Technical ConstraintsBattery Life Revisimo must have a minimum battery life of seven
days. Wireless The bandage and monitor must communicate with
each other within a 100-foot radius (indoors). Alarm System The alarm must be triggered if a user defined
temperature is reached. Also, the alarm must be audible from at least 30 feet away.
Temperature Precision
Revisimo must calculate the core body temperature within 0.5 °F. Revisimo must read temperatures within the range of 65 °F to 110 °F.
Visual Distance The display must present data to the user at a distance of up to 3 feet.
Manufacturability
• The bandage must fit on the temple and the monitor must be handheld.
Political
• Revisimo must abide by Federal Communications Commission regulations and allocations of the radio spectrum (9kHz to 3THz).
Wireless Protocols in Unlicensed Bands
Name Frequency Band
Data Rate Nominal Range Application
ZigBee 900 MHz; 2.4 GHz
250 kbit/s 10m - 100 m WSNs
Bluetooth 2.4 GHz 1 Mbit/s 10 m PANs
Wi-Fi 2.4 GHz 54 Mbit/s 100 m Wireless Internet
Ultra Wideband
3.1 – 10.6 GHz 110 Mbit/s 10 m Multimedia
Microprocessors
Name Size Integrated Radio
ADC Comm. Interfaces
Resources Available
CC2430 7 mm x 7 mm
2.4 GHz DSSS
12 bits
UART, SPI
Prototyping Boards/Stack
PIC24 6 mm x 6 mm
None 10 bits
UART, I2C, SPI
Code examples for PIC architecture
Temperature Sensing DevicesName # of wires
or pinsCost Sensitivity Linear
Thermistor 2 Low-Mid Good No
Resistance Temperature Detector (RTD)
2-4 Low-High Poor Yes
Thermocouple 2 Low- High Poor No
Digital IC’s Multiple Low-Mid Good Yes
Fourier’s Law of Heat Conduction
• Bandage reduces ambient, evaporative, and environmental heat loss
• TsTbTshhTc
skin
bandage
Progress: Radio Testing
Progress: ADC Noise0x1327: 0
0x1328: 3
0x1329: 29
0x132A: 2656
0x132B: 2656
0x132C: 7773
0x132D: 11340
0x132E: 8045
0x132F: 2576
0x1330: 320
0x1331: 9
0x1332: 0
Total: 33232
• Reduces accuracy of measurements• Measure a thermistor multiple times
TimelineJanuary February March April
Research
Radio
Hardware
Assembly
Testing
References[1] http://www.longbeachrealestatehome.com/m/blogs/lbreh/bandaid.jpg, Accessed:
February 18, 2009[2] Information Technology Strategic Plan,
http://www.fcc.gov/omd/strategicplan/InfoTechnologyPlansandReports/itsp-07-23-02.pdf, Accessed: February 18, 2009
[3] http://www.ussensor.com/pgthermistors.html, Accessed: February 23, 2009[4] http://www.omega.com/toc_asp/subsectionsc.asp?book=Temperature&subsection=C
11&flag=1, Accessed: February 23, 2009[5] http://www.omega.com/guides/thermocouples/Unsheathed.html, Accessed: February
23, 2009[6] http://media.digikey.com/photos/Texas%20Instr%20Photos/296-SOT-23-6.jpg,
Accessed: February 23, 2009[7] I. G. Finvers, J. W. Haslett, and G. Jullien, “Wireless Temporal Artery Bandage
Thermometer.” Biomedical Circuits and Systems Conference, 2006. BioCAS 2006. IEEE, pp.166-169, Nov. 29 2006-Dec. 1 2006. [Online] Available http://ieeexplore.ieee.org/stamp/stamp.jsp?arnumber=4600334&isnumber=4600290
Questions?