Requirements of the ESS BLM by Lali Tchelidze CERN, Geneva, 18 Oct. 2011

Requirements of the ESS BLM

Embed Size (px)

DESCRIPTION

Requirements of the ESS BLM. by Lali Tchelidze CERN, Geneva, 18 Oct. 2011. The ESS. ESS: Ion source Linac Target Neutron instruments 5 MW average power 2.5 GeV protons 2.86 ms pulse 50 mA pulse current 14 Hz repetition rate - similar to SNS, but higher power and long pulse. - PowerPoint PPT Presentation

Citation preview

Page 1: Requirements of the ESS BLM

Requirements of the ESS BLM

by Lali Tchelidze

CERN, Geneva, 18 Oct. 2011

Page 2: Requirements of the ESS BLM

The ESS

ESS

Max-IV

ESS:

Ion sourceLinacTargetNeutron instruments

5 MW average power2.5 GeV protons2.86 ms pulse50 mA pulse current14 Hz repetition rate

- similar to SNS, but higher power and long pulse.

Page 3: Requirements of the ESS BLM

The ESS

Page 4: Requirements of the ESS BLM

BLM requirements

• BLM should produce a beam abort signal in less than 2 μs.– Faster than SNS detector is needed (?);– We’ve redesigned the current SNS ionization chamber (shorter

gap in between the electrodes) to achieve faster ion collection time;

– Prototype quoting/ordering under process;– Will test and compare the SNS, LHC and the prototype

ionization chamber charge collection times as soon as we get all three;

Risetime of Original FNAL and New Prototype BLM

0

0.2

0.4

0.6

0.8

1

1.2

0 500 1000 1500 2000

Time [microseconds]

Cu

rre

nt

[No

rma

lize

d]

FNAL Neg Bias

New Prototype

3 kV bias, electrons to center electrode, Argon gas

Page 5: Requirements of the ESS BLM

Dynamic range and Sensitivity

• DR – large enough to detect slow signals, yet not saturate at high losses.

– 1 W/m (at 1 foot) -> 100 mR/h beam-off dose rate -> 100 R/h beam-on dose rates!

– 2 mR/pulse -> 0.69 R/sec during 2.86 msec pulse.

• @ SNS they never use more than 1000.

• Sensitivity - 70 nA/R/s (or even less) is probably enough.

Page 6: Requirements of the ESS BLM

Cryogenic requirements

• A hybrid design of a cryostat is foreseen for ESS.• an operating loss detector might be needed at 70 K / 2 K!

• Physical size of the detector should not be very small – to cover a “large” part of the loss area.