GLISLAND® TRAINING SERIES:

DESIGN CONTROL: REQUIREMENTS AND IMPLEMENTATION

Glisland, Inc.Glisland, Inc.

San Jose, California, USA

http://www.glisland.com

lgl
Glisland logo
Page 2: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Regulatory Requirements

Part I2

Regulatory Requirements

Copyright © 1999 - 2009 Glisland. All Rights Reserved. www.glisland.com

Page 3: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

The Law and Regulations

3

1. Federal Food Drug & Cosmetic (FD&C) Act

2. Code of Federal Regulations (CFR)

3. Title 21 CFR Parts 800-1299

Copyright © 1999 - 2009 Glisland. All Rights Reserved. www.glisland.com

Page 4: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

How FDA Regulates Medical Devices

1. General Controls

2. Special Controls

Premarket Approval

� Class I

� Class II

Class III

4

Three Levels of Controls Device Classification

3. Premarket Approval � Class III

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Class I: General ControlsClass II: General Controls, Special ControlsClass III: General Controls, Special Controls, PMA

Risk Based Controls

Page 5: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

FDA Quality System Regulation

5

1. 1976 Medical Device Amendments - cGMP for medical devices: 520(f) of FD&C Act (21 U.S.C. 360j(f))

2. Codified in 21 CFR Part 820, effective December 1978 (focusing on production process in 1978 GMP)

3. Added design control provision in 1990 amended by SMDA(Safe Medical Device Acts)

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

(Safe Medical Device Acts)a) Added section 803 (21 U.S.C 383)b) Revised 1978 cGMP and harmonized with EU standardc) Rename cGMP for medical device to QSRd) Design Control effective June 1, 1998

4. The new QSR applies to the entire life cycle of a device

www.glisland.com

Page 6: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

GMP & cGMP vs. QSR6

1. GMP – Good Manufacturing Practice

2. cGMP – Current Good Manufacturing Practice

3. QSR – Quality System Regulation (21 CFR 820):

“Current good manufacturing practice (CGMP) requirements

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

“Current good manufacturing practice (CGMP) requirements

are set forth in this quality system regulation. ”

“The failure to comply with any applicable provision in this

part renders a device adulterated under section 501(h) of

the act. Such a device, as well as any person responsible for

the failure to comply, is subject to regulatory action.”

www.glisland.com

Page 7: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Control in QSR

� Subpart A - General Provisions

� Subpart B - Quality System Requirements

� Subpart C - Design

Controls

� Subpart H - Acceptance Activities

� Subpart I - Nonconforming Product

� Subpart J - Corrective and Preventive Action

7

Controls

� Subpart D - Document Controls

� Subpart E - Purchasing Controls

� Subpart F - Identification and Traceability

� Subpart G - Production and Process Controls

� Subpart K - Labeling and Packaging Control

� Subpart L - Handling, Storage, Distribution, and Installation

� Subpart M - Records

� Subpart N - Servicing

� Subpart O - Statistical Techniques

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Page 8: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Development Stage Regulatory Compliance

8

Research Development Production Service Termination

Typical Product Lifecycle

GMP/QSR

Design Control

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Requirement Design Implementation TestingRelease /

Transfer

Typical Development Process

Design Control

GLP

GCP

Page 9: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

21 CFR 820 Subpart C - Design Controls

a) Required: Any class III, class II, and some class I (820.(a)(2))

b) Design and development planning

c) Design input

� Establish

� Maintain

9

Elements Action Keywords

c) Design input

d) Design output

e) Design review

f) Design verification

g) Design validation

h) Design transfer

i) Design changes

j) Design history file

Maintain

� Review

� Approve

� Document

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

D

E

M

R

A

Page 10: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

What’s Wrong with this Glass?

10

A plastic glass A closer look

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Intended use: cold beverage glass; Target market: children in the US

Page 11: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

21 CFR 829 Subpart D - Document Controls

11

Establish and maintain procedures to control all documents that are required by this part:

a) Document approval and distribution

� Review and approval prior to issuance

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Available at all locations necessary

� Remove obsolete documents promptly

b) Document changes

� Original review and approval

� Timely communication of changes

� Change records

www.glisland.com

Page 12: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Electronic Records and Electronic Signatures

12

1. 21 CFR Part 11 and Final Guidance

2. Enforced rules and requirements:a. System access controls b. Enforce Sequence Steps c. Application Access Controls

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

c. Application Access Controls d. Device checks e. Competence of People f. Accountability g. System document control h. Controls for open systems

3. e-signature requirements

www.glisland.com

Page 13: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

FDA Enforcement Actions

13

� Establishment registration and product listing

� Marketing clearance/approval

� Reports

� Inspections

� Notice of Violations (FDA-483, Warning Letters…)

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Notice of Violations (FDA-483, Warning Letters…)

� Recalls

� Civil money penalties

� Seizure

� Injunction

� Prosecution

www.glisland.com

Page 14: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Implementation

Part II14

Implementation

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Page 15: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Control Pyramid15

Policy & Process Model

Procedures

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

Procedures

Work Instructions

Records & Documentation

www.glisland.com

Page 16: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Application of Design Controls

� Waterfall Process

� Concurrent Engineering

16

What Process Model to Apply?FDA Example

� Iterative Process

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Page 17: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Deciding on Process Model

17

� Waterfall Process

� Concurrent Engineering

� Iterative Process

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

“appropriate tools and

techniques used by

competent personnel”

Page 18: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Planning

18

� Goals and objectives of the design and development

� Organizational responsibilities and interface with contractors

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

contractors

� Tasks, deliverables, resources, schedule

� Major reviews and decision points

� Policy, process, and standards

� Document and Change controls

www.glisland.com

Page 19: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Input

19

� Physical and performance requirements

� A set of principles to be followed

� Written to an engineering level of detail

� Rapid prototyping for design input development

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Rapid prototyping for design input development

� Full participation including production and service personnel, key suppliers

� Review and approval

� Change controls

www.glisland.com

Page 20: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Output

� Phase design output

� Finished design output� The device� Packaging � Labeling

� Risk management file

� Production specification

� Acceptance criteria

20

� Labeling� Device master record

� Specification� block diagrams� flow charts� software high-level code

� Procedures � Work instructions

www.glisland.comCopyright © 1999 - 2009 Glisland. All Rights Reserved.

Page 21: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Review

� Evaluate the adequacy of the design requirements

� Evaluate the capability of the design to meet these

21

How to review Cost of change

design to meet these requirements

� Identify problems

� Independent and objective reviews

� Design review meeting

� Resolution of concerns

Early stage

Middle stage

Late stage

Copyright © 1999 - 2009 Glisland. All Rights Reserved. www.glisland.com

Cost

Page 22: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Verification

22

� Establish design phase conformance: design output vs. design input

� Methods

� Tests

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Tests

� Inspections

� Analyses

� Documentation

www.glisland.com

Page 23: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Validation

23

� Validation Planning

� Validation Methods� Clinical evaluation or clinical trials

� Testing production unit in the actual or simulated use environment

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

environment

� Analysis and inspection

� Scientific literature

� Historical evidence

� Validation Review

� Documentation

www.glisland.com

Page 24: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Transfer

24

� Production Specification� assembly drawings

� component procurement specifications

� workmanship standards

manufacturing instructions,

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� manufacturing instructions,

� inspection and test specifications.

� Procedures� assessment of the completeness and adequacy of the production specifications (transfer protocols)

� Review and approval

www.glisland.com

Page 25: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design Changes

25

� Document control

� Enumeration of design documents

� Status

� Revision history

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Revision history

� Change control

� Enumeration of deficiencies and corrective actions

� Tracking of resolution

www.glisland.com

Page 26: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Design History File (DHF)

26

� Detailed design and development plan specifying design tasks and deliverables.

� Copies of approved design input documents and design output documents

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

design output documents

� Design review documentation

� Validation documentation.

� When applicable, copies of controlled design documents and change control records

www.glisland.com

Page 27: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Secret Weapons

27

Traceability

Copyright © 1999 - 2009 Glisland. All Rights Reserved. www.glisland.com

Checkpoints & Matrix Reviews

RiskManagement

Page 28: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

References28

� Fundamentals of US Regulatory Affairs, 5th Edition, RAPS, 2007

� Milestones in U.S. Food and Drug Law History,http://www.fda.gov/opacom/backgrounders/miles.html

� Significant Amendments to the FD&C Act, http://www.fda.gov/opacom/laws/

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

http://www.fda.gov/opacom/laws/

� 21 CFR Part 820, http://www.gpoaccess.gov/nara/index.html

� FDA Regulatory Procedure Manual 2007, http://www.fda.gov/ora/compliance_ref/rpm/

� FDA Guidance Documents: http://www.fda.gov/cder/guidance/index.htm

Note: the provided hyperlinks may be changed in the future by the content providers without notice.

www.glisland.com

Page 29: REQUIREMENTS AND IMPLEMENTATION - Glislandglisland.com/Pub/SampleSlides/Design_Control.pdfb) Revised 1978 cGMP and harmonized with EU standard c) Rename cGMP for medical device to

Glisland Design Control Services

29

� Process Development (including SOP & WI writing)

� Design Planning

� Project Management

� Risk Analysis

Copyright © 1999 - 2009 Glisland. All Rights Reserved.

� Risk Analysis

� Design Review

� Verification & Validation

� Compliance Audit

� Training

www.glisland.com