2
Characterized By Ingenuity | 1 Replacing the Desiccant Filter 3 Using the flathead screwdriver pry open the bottom grey latch holding the filter. Using needle nose pliers, A. press up on the quick release and B. pull down on the filter assembly. 6 Press the key to power on the instrument and perform Desiccant Filter Replacement Verification. Replace with the new desiccant filter assembly and press up into the quick disconnect and leave an 1/8”gap for the brass fitting to be easliy remove on the next replacement. Snap the grey latch closed. Press the key to power off the instrument. When the desiccant beads are dark in color, the desiccant filter needs replacing. 1 4 5 2

Replacing the Desiccant Filter

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Replacing the Desiccant Filter

Characterized By Ingenuity | 1

Replacing the Desiccant Filter

3

Using the flathead screwdriver pry open the bottom grey latch holding the filter.

Using needle nose pliers,

 A. press up on the quick release and

 B. pull down on the filter assembly.

6

Press the key to power on the instrument and

perform Desiccant Filter Replacement

Verification.

Replace with the new desiccant filter assembly and press up

into the quick disconnect and leave an 1/8”gap for the brass fitting to be easliy remove on

the next replacement.

Snap the grey latch closed.

Press the key to power off the instrument.

When the desiccant beads are dark in color, the desiccant filter needs replacing.

1 4

52

Page 2: Replacing the Desiccant Filter

Desiccant Filter Replacement VerificationAfter replacing the desiccant filter, perform this procedure to ensure there are no air leaks.

2

3

4

5

6

7

8

Plug in the 1/4” hose to the air input fitting from a clean dry air source.

Press the key to power on the instrument.

From the PODS+ Homescreen, press the menu button next to .

press the menu button next to to display the manual entry screen.

press on the keypad to pressurize the system. Listen for leaks around the desiccant filter assembly.

Press on the keypad again to depressurize.

If there is a leak re-seat the desiccan filter assembly.

If there are no leaks the system is ready to process samples.

1

1

© 2017 Beckman Coulter, Inc. All rights reserved. Beckman Coulter, the stylized logo, and the Beckman Coulter product and service marks mentioned herein are trademarks or registered trademarks of Beckman Coulter, Inc. in the United States and other countries.

For Beckman Coulter’s worldwide office locations and phone numbers, please visit “Contact Us” at beckman.com

PART-3213FLY11.17