Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries, Marcus L. Scott, and Shilesh Jani J Bone Joint Surg Am Volume 83(2 suppl 2):S116-122 November 1, 2001 ©2001 by The Journal of Bone and Joint Surgery, Inc.

Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Embed Size (px)

Citation preview

Page 1: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared

with Cross-Linked Polyethylene

by Michael D. Ries, Marcus L. Scott, and Shilesh Jani

J Bone Joint Surg AmVolume 83(2 suppl 2):S116-122

November 1, 2001

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 2: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Display of particle analysis.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 3: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Gravimetric wear testing showed a significant improvement in wear as the level of radiation increased.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 4: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particle-generation rates did not correlate with gravimetric wear rates.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 5: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particles of nonirradiated, ethylene-oxide-sterilized ultra-high molecular weight polyethylene (10,000).

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 6: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particles of inert-gas gamma-irradiated ultra-high molecular weight polyethylene (10,000).

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 7: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particles of highly cross-linked ultra-high molecular weight polyethylene subjected to 5 Mrad of gamma irradiation (10,000).

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 8: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particles of highly cross-linked ultra-high molecular weight polyethylene subjected to 10 Mrad of gamma irradiation (10,000).

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 9: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particle size histogram.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 10: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particle surface area versus size histogram.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 11: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Particle volume versus size histogram.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.

Page 12: Relationship Between Gravimetric Wear and Particle Generation in Hip Simulators: Conventional Compared with Cross-Linked Polyethylene by Michael D. Ries,

Total wear volume per Mcycle of particles generated from each type of liner.

Michael D. Ries et al. J Bone Joint Surg Am 2001;83:S116-122

©2001 by The Journal of Bone and Joint Surgery, Inc.