REF AP-07014 Refill Kit–Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS WARNINGS The Pump should be implanted and refilled only by qualified medical personnel, knowledgeable in the surgical use and servicing of implantable devices and catheters, and trained specifically to implant or refill the Pump. Use of the Pump by personnel not properly trained in its implantation and/or servicing may lead to serious consequences involving either under or over-delivery of drug to the patient. In the event of an over-delivery of drug refer to the approved drug labeling for appropriate action. Utilization of the Pump requires the proper handling (filling, storage and dispensing) of a significant volume/dosage of drug. This amount of drug may be extremely harmful to the patient if delivered suddenly or inappropriately. Bolus access and Pump refill procedures must be performed using the correct access needle. Never attempt to refill the Pump using a Special Bolus Needle. This use will result in giving a bolus injection to the patient and may cause a fatal drug overdose. The Model 3000 Series Constant Flow Implantable Pumps contain a safety valve in-line with the bolus pathway to insure that a bolus procedure can only be performed when using a properly positioned Special Bolus Needle (REF AP-04013 or AP-04032). 1. Attach the Needle and Stopcock to the tubing set. Stopcock and clamp are in the OPEN position. Next attach the 50 mL cali- brated syringe barrel to the Stopcock. Tighten all connections. 2. Re-palpate the Pump site and locate the raised septum. Insert the non-coring needle PERPENDICULAR to the Pump septum. 3. Advance needle until it is in contact with the needlestop. 4. Allow the Pump reservoir to empty. Keep downward pressure on the needle throughout the procedure. If no fluid returns to the syringe, refer to Troubleshooting Guide for assistance. NOTE: After allowing the Pump to "empty" into the syringe barrel, approxi- mately 1.5 mL of fluid from the previous refill will remain in the Pump. 5. CLOSE STOPCOCK. Note return volume (mL) and record on refill data sticker provided. Discard syringe barrel and stopcock leaving the needle and refill set in place. 6. Expel air from the 10 mL syringe of saline. Attach the syringe to the proximal end of the refill set and confirm that the needle is still in contact with the needle stop. Keep downward pressure on the needle and inject 5 mL of saline into the Pump. 7. Release pressure on the plunger and allow the 5 mL injected to return to the syringe. This procedure reconfirms correct positioning of the needle. If no fluid returns, refer to Troubleshooting Guide for assistance. Disconnect the syringe from the refill set. 8. Expel air from refill syringe. Attach syringe to proximal end of refill set. Confirm that the needle is still in contact with the needle stop. 9. Keep downward pressure on the needle and begin to inject refill solution into the Pump. Release pressure on plunger at 5 mL increments and allow 1 mL of solution to return to syringe. This will verify that the needle is in the correct position and the Pump reservoir is being filled. Continue to inject and check needle placement until the syringe is emptied. PRECAUTION: If no fluid returns to the syringe upon release of the plunger, DO NOT CONTINUE TO INJECT REFILL SOLUTION UNTIL YOU HAVE VERIFIED THE NEEDLE PLACEMENT PER THE PROCEDURE IN THE TROUBLESHOOTING GUIDE. 10. After injecting the entire contents of the refill syringe, close the clamp on the tubing set and pull needle out of Pump septum. Remove drape and iodine. Apply adhesive bandage to access site. Troubleshooting Guide: If no fluid returns during Pump emptying, or during the Refill Injection Procedure, proceed as follows; A. Confirm needle position and that an 22 gauge, non-coring needle is being used. Check that the needle is PERPENDICULAR to the Pump septum and is fully depressed and in contact with the needlestop. If no volume returns proceed with steps B, C, D. B. Attach a 10 mL syringe containing normal saline solution to the refill tubing set. C. Keep downward pressure on the needle and inject 5 mL of saline into the Pump. Release pressure on the syringe plunger and allow the fluid to return to the syringe. If fluid returns then the needle is in the proper position and indicates that the Pump was empty at the start of the refill procedure. Replace the 10 mL syringe with a syringe containing the Pump refill solution and continue with the refill procedure (step 8 to 10). D. If there is still no fluid return after injection of the 5 mL of saline, remove needle and refill tubing set. Flush the refill set to confirm that the system is patent. Re-insert the needle perpendicular to the Pump septum until it is in contact with the needlestop. Repeat step C. E. Contact Codman & Shurtleff, Inc. for 24-Hour Technical Assistance at 508 660-1122 or 800 660-2660. DDS-06-001 KIT CONTENTS [1] 50 mL Calibrated Syringe Barrel [1] Tubing Set [1] Stopcock [2] Huber Point Needles; 22 Gauge x 1.5" (3.8 cm) [3] POVIDONE IODINE swabsticks [3] Alcohol Swabsticks [6] Gauze Pads [1] Fenestrated Drape REVIEW ALL INSTRUCTIONS BEFORE PROCEEDING: PRECAUTION 1. A non-coring needle is provided in this kit. It is critical to the safe functioning of the Pump that no other needle be used to penetrate the septum. PRECAUTION 2. Do not aspirate fluid from the Pump. Aspiration may cause blood to be drawn into the catheter and result in occlusion. PRECAUTION 3. It is important to precisely follow the Pump refill instructions as detailed in this pamphlet in order to successfully complete the Pump refill procedure. If the needle is not properly positioned, and verified as detailed in the Pump Refill Procedures, there is a possibility that a drug extrava- sation will occur. Use the refill data sticker provided in this kit to maintain accurate Pump refill records. Place patient in a supine position. Expose Pump pocket site. Palpate the Pump site and locate the raised septum. Fill a syringe with appropriate refill solution (16 mL for the Model 3000-16, 30 mL for the Model 3000 and 50mL for the Model 3000-50) and a 10 mL syringe with saline solution. Using sterile technique, expose kit contents. Don sterile gloves. Use three iodine swabsticks to prep the Pump site. Allow prepped area to dry. Place fenestrated drape over Pump site. KIT CONTENTS [1] Data Sticker [2] Adhesive Bandages [1] Refill Worksheet [1] INSTRUCTIONS FOR USE [2] INSTRUCTION POSTERS ITEMS NOT PROVIDED IN KIT: Sterile Gloves and Syringes Required for Procedure. For 24-Hour Technica Assistance or to arrange an in-service Training Program, contact: Codman & Shurtleff, Inc. 508 660-1122 or 800 660-2660. STOPCOCK OPEN LID INCORRECT NEEDLE POSITION INCORRECT NEEDLE POSITION REFILL PATH BOLUS PATH BOLUS SAFETY VALVE NEEDLE STOP CORRECT NEEDLE POSITION RETURN VOLUME STOPCOCK OPEN LID REFILL DATA Patient Name _______________________________________________________________________ Date Time Volume mL Volume No. of Flow Rate Refill substance, concentration and total volume in mL. Returned Infused Days From mL/Day From Pump Last Refill 1. Add 1.2 mL to the return volume in syringe barrel to account for the fluid in the tubing set. 2. Subtract return volume from volume at last refill typically 30 mL. 3. Flow rate per day. Divide volume infused by number of days from last refill. TLB1234X STOPCOCK CLOSED Codman SALINE SOLUTION 5 mL SALINE SOLUTION 5 mL REFILL SOLUTION 1 mL REFILL SOLUTION 1 mL 5 mL 5 mL 5 mL Refill Kit Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS ARROW Refill Kit Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS ARROW Refill Kit Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS Refill Kit Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS REFILL DATA Patient Name Date Time Volume Volume no of Flow Rate Refill substance, concentration and total volume in mL. Returned Infused Days From mL / Day From Pump Last Refill 1. Subtract return volume from volume at last refill (16 ,L Model 3000-16,, 30 mL Model 3000 2. (30 mL), 50 mL Model 3000-50. Flow Rate /Day. Divide volume infused by number of days. Codman Codman Codman CE 0 1 2 3 Refill Volume: Model 3000-16.... 16 mL Model 3000........ 30 mL Model 3000-50.... 50 mL

REF AP-07014 Refill Kit–Model 3000 Series step C. E. Contact Codman & Shurtleff, Inc. for 24-Hour Technical Assistance at 508 660-1122 or 800 660-2660. DDS-06-001 KIT CONTENTS [1]

  • Upload
    lehanh

  • View
    217

  • Download
    2

Embed Size (px)

Citation preview

REF AP-07014

Refill Kit–Model 3000 SeriesCONSTANT FLOW IMPLANTABLE PUMPS

WARNINGSThe Pump should be implanted and refilled only by qualified medical personnel, knowledgeable in the surgical use and servicing of implantable devicesand catheters, and trained specifically to implant or refill the Pump. Use of the Pump by personnel not properly trained in its implantation and/or servicingmay lead to serious consequences involving either under or over-delivery of drug to the patient. In the event of an over-delivery of drug refer to theapproved drug labeling for appropriate action.Utilization of the Pump requires the proper handling (filling, storage and dispensing) of a significant volume/dosage of drug. This amount of drug may beextremely harmful to the patient if delivered suddenly or inappropriately.Bolus access and Pump refill procedures must be performed using the correct access needle. Never attempt to refill the Pump using a Special Bolus Needle.This use will result in giving a bolus injection to the patient and may cause a fatal drug overdose.The Model 3000 Series Constant Flow Implantable Pumps contain a safety valve in-line with the bolus pathway to insure that a bolus procedure can onlybe performed when using a properly positioned Special Bolus Needle (REF AP-04013 or AP-04032).

1. Attach the Needle and Stopcock to the tubing set. Stopcockand clamp are in the OPEN position. Next attach the 50 mL cali-brated syringe barrel to the Stopcock. Tighten all connections.

2. Re-palpate the Pump site and locate the raised septum. Insert thenon-coring needle PERPENDICULAR to the Pump septum.

3. Advance needle until it is in contact with the needlestop.

4. Allow the Pump reservoir to empty. Keep downward pressure on the needle throughout the procedure. If no fluid returns to the syringe,refer to Troubleshooting Guide for assistance.NOTE: After allowing the Pump to "empty" into the syringe barrel, approxi-mately 1.5 mL of fluid from the previous refill will remain in the Pump.

5. CLOSE STOPCOCK. Note return volume (mL) and record on refilldata sticker provided. Discard syringe barrel and stopcock leaving the needle and refill set in place.

6. Expel air from the 10 mL syringe of saline. Attach the syringe to theproximal end of the refill set and confirm that the needle is still in contactwith the needle stop. Keep downward pressure on the needle and inject5 mL of saline into the Pump.

7. Release pressure on the plunger and allow the 5 mL injected to returnto the syringe. This procedure reconfirms correct positioning of the needle.If no fluid returns, refer to Troubleshooting Guide for assistance. Disconnectthe syringe from the refill set.

8. Expel air from refill syringe. Attach syringe to proximal end of refillset. Confirm that the needle is still in contact with the needle stop.

9. Keep downward pressure on the needle and begin to inject refill solutioninto the Pump. Release pressure on plunger at 5 mL increments andallow 1 mL of solution to return to syringe. This will verify that the needleis in the correct position and the Pump reservoir is being filled. Continueto inject and check needle placement until the syringe is emptied.

PRECAUTION: If no fluid returns to the syringe upon release of the plunger, DO NOT CONTINUE TO INJECT REFILL SOLUTION UNTIL YOU HAVE VERIFIED THE NEEDLE PLACEMENT PER THE PROCEDUREIN THE TROUBLESHOOTING GUIDE.

10. After injecting the entire contents of the refill syringe, close the clamp on the tubing set and pull needle out of Pump septum. Remove drape and iodine. Apply adhesive bandage to access site.

Troubleshooting Guide: If no fluid returns during Pump emptying, or during the Refill Injection Procedure, proceed as follows;

A. Confirm needle position and that an 22 gauge, non-coring needle isbeing used. Check that the needle is PERPENDICULAR to the Pumpseptum and is fully depressed and in contact with the needlestop.If no volume returns proceed with steps B, C, D.

B. Attach a 10 mL syringe containing normal saline solution to the refilltubing set.

C. Keep downward pressure on the needle and inject 5 mL of salineinto the Pump. Release pressure on the syringe plunger and allow thefluid to return to the syringe. If fluid returns then the needle is in theproper position and indicates that the Pump was empty at the start ofthe refill procedure. Replace the 10 mL syringe with a syringe containingthe Pump refill solution and continue with the refill procedure(step 8 to 10).

D. If there is still no fluid return after injection of the 5 mL of saline,remove needle and refill tubing set. Flush the refill set to confirm thatthe system is patent. Re-insert the needle perpendicular to the Pumpseptum until it is in contact with the needlestop. Repeat step C.

E. Contact Codman & Shurtleff, Inc. for 24-Hour Technical Assistanceat 508 660-1122 or 800 660-2660.

DDS-06-001

KIT CONTENTS[1] 50 mL Calibrated Syringe Barrel[1] Tubing Set[1] Stopcock[2] Huber Point Needles;

22 Gauge x 1.5" (3.8 cm)[3] POVIDONE IODINE swabsticks[3] Alcohol Swabsticks[6] Gauze Pads[1] Fenestrated Drape

REVIEW ALL INSTRUCTIONS BEFORE PROCEEDING:PRECAUTION 1. A non-coring needle is provided in this kit.It is critical to the safe functioning of the Pump that no otherneedle be used to penetrate the septum.PRECAUTION 2. Do not aspirate fluid from the Pump.Aspiration may cause blood to be drawn into the catheterand result in occlusion.

PRECAUTION 3. It is important to precisely follow the Pumprefill instructions as detailed in this pamphlet in order tosuccessfully complete the Pump refill procedure. If the needleis not properly positioned, and verified as detailed in the PumpRefill Procedures, there is a possibility that a drug extrava-sation will occur.

Use the refill data sticker provided in this kit to maintain accurate Pump refill records.Place patient in a supine position. Expose Pump pocket site. Palpate the Pump site and locate the raised septum. Fill a syringe with appropriate refill solution (16 mL forthe Model 3000-16, 30 mL for the Model 3000 and 50mL for the Model 3000-50) and a 10 mL syringe with saline solution. Using sterile technique, expose kit contents.Don sterile gloves. Use three iodine swabsticks to prep the Pump site. Allow prepped area to dry. Place fenestrated drape over Pump site.

KIT CONTENTS[1] Data Sticker[2] Adhesive Bandages[1] Refill Worksheet[1] INSTRUCTIONS FOR USE[2] INSTRUCTION POSTERS

ITEMS NOT PROVIDED IN KIT:Sterile Gloves and SyringesRequired for Procedure.

For 24-Hour Technica Assistance or to arrange an in-service TrainingProgram, contact: Codman & Shurtleff, Inc.

508 660-1122 or 800 660-2660.

STOPCOCKOPEN

LID

INCORRECT NEEDLE POSITION INCORRECT NEEDLE POSITION

REFILL PATH

BOLUS PATH

BOLUS SAFETY VALVE

NEEDLE STOP

CORRECT NEEDLE POSITION

RETURN VOLUME

STOPCOCK OPEN

LID

REFILL DATA Patient Name _______________________________________________________________________

Date Time Volume mL Volume No. of Flow Rate Refill substance, concentration and total volume in mL.

Returned Infused Days From mL/Day

From Pump

Last Refill

1. Add 1.2 mL to the return volume in syringe barrel to account for the fluid in the tubing set.

2. Subtract return volume from volume at last refill typically 30 mL.

3. Flow rate per day. Divide volume infused by number of days from last refill.

TLB12

34X

STOPCOCKCLOSED

Codman

SALINE SOLUTION

5 mL

SALINE SOLUTION

5 mL

REFILL SOLUTION

1 mL

REFILL SOLUTION

1 mL

5 mL5 mL 5 mL

Refill KitModel 3000 SeriesCONSTANT FLOW IMPLANTABLE PUMPS

ARROWRefill KitModel 3000 Series

CONSTANT FLOW IMPLANTABLE PUMPS

ARROWRefill KitModel 3000 Series

CONSTANT FLOW IMPLANTABLE PUMPS

Refill KitModel 3000 SeriesCONSTANT FLOW IMPLANTABLE PUMPS

REFILL DATA Patient Name

Date Time Volume Volume no of Flow Rate Refill substance, concentration and total volume in mL. Returned Infused Days From mL / DayFrom Pump Last Refill

1. Subtract return volume from volume at last refill (16 ,L Model 3000-16,, 30 mL Model 3000 2. (30 mL), 50 mL Model 3000-50. Flow Rate /Day. Divide volume infused by number of days.

Codman

Codman

Codman

CE 0 1 2 3

Refill Volume: Model 3000-16.... 16 mLModel 3000........ 30 mLModel 3000-50.... 50 mL

Refill Kit - Model 3000 Series CONSTANT FLOW IMPLANTABLE PUMPS

Instructions for Use

Caution: U.S. Federal Law limits this device to sale by or on order of a physician. Contents of unopened, undamaged package are sterile. Disposable. Refer to package insert for current warnings, precautions, and instructions for use.

© 2002 Codman & Shurtleff, Inc. All rights reserved.

Implantable Drug Delivery Systems

GERMANYEthicon GmbHGeschäftsbereich CodmanOststrasse 1D-22844 NorderstedtGermany+49 (0) 40 52207 623

FRANCECodman7 allée Irène Joliet CurieBP 25669801 SAINT PRIEST CedexFrance+33 (0) 4 72 79 29 34

SPAINJohnson & Johnson ProductosProfesionales S.A.Paseo de las Doce Estrellas, No5 y 7CAMPO DE LAS NACIONES – 28042MADRIDSpain+34 91 7 22 83 00

ITALYEthicon SPAVia Del Mare, 5600040 Pratica Di Mare RMItaly+39-06-911.94.500 or 911.94.313

Codman & Shurtleff, Inc.24 Hour Technical Assistance:508 660-1122800 660-2660

Implantable Drug Delivery Systems

Codman & Shurtleff, Inc.24 Hour Technical Assistance 508 660-1122 or 800 660-2660