Real World Experience: Developing Dose and Protocol Monitoring from Scratch Ingrid Reiser, PhD DABR Department of Radiology The University of Chicago

Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

RealWorldExperience:DevelopingDoseandProtocolMonitoring

fromScratch

IngridReiser,PhDDABRDepartmentofRadiologyTheUniversityofChicago

Page 2: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Outline

• CTprotocolmonitoring– Let’sbuildaprotocolbook!

• CTdosemonitoring

Page 3: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Fromscratch:statusasof~2015

• Fleetof8-9PhilipsCTscanners,mixofinpatient,outpatient,dedicatedpediatric• Radimetrics installed,about60%ofCTstudiesrecorded

• Events:– Majorsoftwareupgradeonsomescanners(PhilipsiPatient)– 3scannersreplaced,whichaddedGE– Openingofremotefacility

• Personneltransitions(CTmanager)• NodedicatedCTphysicist

Page 4: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

CTProtocolMonitoringandOptimization

• Joint Commission Standard PC.01.03.01: The [critical access] hospital plans the patient’s care. A 25. The [critical access] hospital establishes or adopts diagnostic computed tomography (CT) imaging protocols based on current standards of practice, which address key criteria including clinical indication, contrast administration, age (to indicate whether the patient is pediatric or an adult), patient size and body habitus, and the expected radiation dose index range.A 26. Diagnostic computed tomography (CT) imaging protocols

are reviewed and kept current with input from an interpreting radiologist, medical physicist, and lead imaging technologist to make certain that they adhere to current standards of practice and account for changes in CT imaging equipment. These reviews are conducted at time frames identified by the [critical access] hospital.

Page 5: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

JCrequirements

• CTQualityPolicywasestablished– CTQualityCoreTeamcreated,taskedwithprotocolanddosereview

Page 6: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:
Page 7: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theprotocols

“Routine critical and systematic review of protocols are necessary not only to ensure that appropriate radiation dose levels are being used but also to verify that other aspects of are

acceptable and consistent with best practice standards.”

Protocolreview:Whatare ourprotocols?

Whataretheparameters?

J Am Coll Radiol 2014;11:267-270.

Page 8: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theprotocols

“Routine critical and systematic review of protocols are necessary not only to ensure that appropriate radiation dose levels are being used but also to verify that other aspects of are

acceptable and consistent with best practice standards.”

– Typically>200imagingprotocolsperscanner– NeedtheactualCTprotocols

Page 9: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:
Page 10: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theprotocols

“Routine critical and systematic review of protocols are necessary not only to ensure that appropriate radiation dose levels are being used but also to verify that other aspects of are

acceptable and consistent with best practice standards.”

– Typically>200protocolsperscanner– NeedtheactualCTprotocols– Notallvendors/modelsproduceaCOMPLETEhumanreadablelistingofprotocolparameters

Page 11: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Oursolution:gettingscannerprotocols

• Haveserviceengineerretrieveprotocolbackupfiles(Philips)• PhilipsiPatient softwareproducesatext-file(htmlformat)• GERevolutionCTseriesproducecsvfiles– incomplete

• noserieslabels• RevolutionCT:Nopitchfactor,protocolnumberingomitted

Page 12: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Usingtheprotocols(i.e.,reviewlogistics)

• Textprotocolformattingmaynotbeamenableforreview

• Radimetrics hasaprotocoleditor– Mayormaynotimportyourprotocolparameters

• Oursolution:– Parseprotocolsusingascriptinglanguage(python,matlab,etc)– Generateexcelmasterprotocolforeachscanner

Page 13: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Acquisitionparameters

Page 14: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Reconstructionparameters

Page 15: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Multiplevendors:Consistencyofprotocols

• Perfectparametermatchnotalwayspossible– GE:3.75mmsliceinterval– Philips:4mmsliceinterval

Page 16: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Currentbestpracticestandards?Acceptable

doselevels?

Radiologistsinindividualsections:

Physicist: Technologist/Radiologist:

protocolsparameters

Scaninstructions Consolidation,

Standardization

Notificationvalues Review:

appropriatedoseandimagequality

Approvalofprotocols

Page 17: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:
Page 18: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theteam

“Protocol review and improvement efforts should be undertaken by a core team consisting of at least one radiologist, one physicist, and one technologist.”

• Radiologistcanonlyoverseeprotocolsrelevanttotheirsection– GenerallycannotapproveCTprotocolsforadifferentsection(i.e.,thoracicradiologistcannotapproveneuroprotocols)

J Am Coll Radiol 2014;11:267-270.

Page 19: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theteam

“Protocol review and improvement efforts should be undertaken by a core team consisting of at least one radiologist, one physicist, and one technologist.”

• CTclinicalmanager(s)overseesCT– IsultimatelyresponsibleforthequalityofCTscansperformedbyhis/hertechnologists

– Has/shouldhaveaninterestinwelldocumented,standardizedCTimagingprotocols

– Supervisesandcommunicateswithalltechnologists• Hasfirepower

Page 20: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Theteam

“Protocol review and improvement efforts should be undertaken by a core team consisting of at least one radiologist, one physicist, and one technologist.”

• CTclinicalmanager(s)overseesCT– Cangetaccesstoscanners(masteroftheschedule)– Knowstheprotocols– ControlshowCTscannersaresetup(i.e.,passwordprotectionofprotocoleditor)

..DOpasswordprotectyourCTprotocols- allchangesaremadeinthebestintent..

Page 21: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Must-haves:

• Supportfromtheadministration:– Dedicated,protectedtechnologists’time($$)

• CTmanagersupportcritical– Appointsonededicatedtechnologist(“technicalcoordinator”)– Scheduling:protectedtimeOFFTHEFLOORtohelpwithprotocols

• Difficultiftherearestaffingshortages

– ControlsCTschedule(blockscannertimeforprotocolupdates,notificationsettings)

• Leadtechnologistmaynotbeideal,astheyalreadyhavemanyotherresponsibilities

Page 22: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:
Page 23: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Aprotocolbook

• Protocolbook:– Scaninstructions:Technologist/radiologist– Scanparameters:Physicist

• Protocolbookhelpsestablishstandard– Technologists’willchangescanparametersontheflybecauseA radiologistrequestedscanA tobedoneinacertainway..

Page 24: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:
Page 25: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Ourapproachtobuildingaprotocolbook

• Goal:Buildaprotocolbookthatisflexible,editable,allowingforautomatedupdatingofscanparameters

• Requirements:– Easyexchangewithtechnologists/clinicians(MSoffice)– Automatedscanparameterupdate

• Todisseminate:– UtilizeMSsharepoint

• Use.aspx extensiontodisplayhtml

Page 26: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Softwaretool

Data entry:clinicalinstructions

Data entry:scanparameters

Data entry:Scanparameterupdate

Clarityoflayout

Access/Publishing

Versioncontrol

MSWord Good Manualorlinkedfields

Manual Userdefined Uploadindividually,openinWord

NeedstobemaintainedbyuserLinked

MSExcel Canbecumbersome(celllayout)

Manualorlinkedcells

Manual Userdefined Uploadindividually,openinExcel

NeedstobemaintainedbyuserLinked

HTML Needspecialeditor,ordirecthtml

Via scripting Via scripting, Userdefined Via scripting,displaysaswebsite(sharepoint)

Needstobemaintainedbyuser

Radimetricsprotocoleditor

Good Automated(depends onCTmanfr)/Manual

Manual Notuseradjustable,poorlayout

Exporttopdfanduploadindividually

Tracksprotocolversion

Page 27: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

CTProtocolMonitoringandOptimization

• Joint Commission Standard PC.01.03.01: The [critical access] hospital plans the patient’s care.

A 26. Diagnostic computed tomography (CT) imaging protocols are reviewed and kept current with input from an interpreting radiologist, medical physicist, and lead imaging technologist to make certain that they adhere to current standards of practice and account for changes in CT imaging equipment. These reviews are conducted at time frames identified by the [critical access] hospital.

A 25. The [critical access] hospital establishes or adopts diagnostic computed tomography (CT) imaging protocols based on current standards of practice, which address key criteria including clinical indication, contrast administration, age (to indicate whether the patient is pediatric or an adult), patient size and body habitus, and the expected radiation dose index range.

Page 28: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

CTdosemonitoring

• Order?Protocol?StudyDescription?Examcard?

• Terminologyisnotdefined.Becarefulwhencommunicating.

Page 29: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Understandyourworkflow

Referringphysicianordersstudy

CTCHESTW

Radiologyresidentreviewsorderandclinicalhistory,selectsscanprotocol

• Standardcontrast-enhancedchest• Interstitiallungdisease• …

Technologistchoosessize-specificprotocol,ormayadjusttechniqueforpatientsize

• StandardChest• Interstitiallungdisease• …

instructureddosereport“order”,”studydescription”

”protocol”,“examcard”retrievedfromimages

visibleonERMonly“protocolling”

->requiresimagestobesenttomonitoringsoftware

Page 30: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Dosereporting

• Bystudydescription/order– Imagingprotocolnotknown– Notasusefulforscanparameteroptimization

• Byscanprotocol– Sameimagingprotocolmaybeusedfordifferentorders

• DIR:byorder

Page 31: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

CTdosemonitoring

Wherehaveallthestudiesgone..

Page 32: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

UnderstandyourIT:Whatgetssendwhere?

CTexamiscompletedonscanner

Technologistsendsstudy

DirectlytoPACS

“Smartrouter”

KnowstosendCTstudytoradimetrics

CTexamiscompletedonscanner

Autosend toradimetrics

Initialsetup:

Newsetup:

Page 33: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

ITneedstobepartoftheequation

CTexamiscompletedonscanner

Autosend toradimetrics

Newsetup:

Drawback:Thislinkcanbreakeasily

Page 34: Real World Experience: Developing Dose and Protocol ...amos3.aapm.org/abstracts/pdf/127-35538-418554... · CT Protocol Monitoring and Optimization • Joint Commission Standard PC.01.03.01:

Finalwords

• Establishingprotocol/dosemonitoringisnottrivial• Administrationsupport(dedicatedtechnologists’timecosts$$)• Knowyoursystems,yourworkflow

• BuildaDEDICATEDteam– Needtobeonthesamepagewithtechnologists/radiologists– GetsomeonefromITwhoCANHELP

• Knowyourmedicalphysicscolleagues,reachouttosomeonewithsimilarenvironment

Acknowledgements:UChicago CTmanagerandtechnologists,ITgroup,MedicalPhysicsGroup,NickBevins,PhD