RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Embed Size (px)

Citation preview

Page 1: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

RADIATION PROTECTION INDIAGNOSTIC AND

INTERVENTIONAL RADIOLOGY

L 6: X Ray production

IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

Page 2: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 2

Basic elements of the X Ray source assembly

• Generator : power circuit supplying the required potential to the X Ray tube

• X Ray tube and collimator: device producing the X Ray beam

Page 3: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 3

X Ray tubes

Page 4: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 4

X Ray tube components

• Cathode: heated filament which is the source of the electron beam directed towards the anode• tungsten filament

• Anode (stationary or rotating): impacted by electrons, emits X Rays

• Metal tube housing surrounding glass (or metal) X Ray tube (electrons are traveling in vacuum)

• Shielding material (protection against scattered radiation)

Page 5: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 5

X Ray tube components

1: long tungsten filament2 : short tungsten filament3 : real size cathode

1: mark of focal spot

housing cathode

Page 6: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 6

Cathode structure (I)

• Cathode includes filament(s) and associated circuitry

• tungsten material : preferred because of its high melting point (3370°C)

• slow filament evaporation

• no arcing

• minimum deposit of W on glass envelope

• To reduce evaporation the emission temperature of the cathode is reached just before the exposure

• in stand-by, temperature is kept at ± 1500°C so that 2700°C emission temperature can be reached within a second

Page 7: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 7

• Modern tubes have two filaments • a long one : higher current/lower resolution

• a short one : lower current/higher resolution

• Coulomb interaction makes the electron beam divergent on the travel to the anode• lack of electrons producing X Rays

• larger area of target used

• focal spot increased lower image resolution

Focalisation of electrons is crucial !

Cathode structure (I)

Page 8: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production

X Ray tube characteristics

• Anode mechanical constraints• Material : tungsten, rhenium, molybdenum, graphite• Focal spot : surface of anode impacted by electrons• Anode angle• Disk and annular track diameter (rotation frequency

from 3,000 to 10,000 revolutions/minute)• Thickness mass and material (volume) heat

capacity

• Anode thermal constraints• Instantaneous power load (heat unit)• Heat loading time curve• Cooling time curve

Page 9: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 9

Anode angle (I)

• The Line-Focus principle• Anode target plate has a shape that is more

rectangular or ellipsoidal than circular • the shape depends on :

• filament size and shape

• focusing cup’s and potential

• distance between cathode and anode

• Image resolution requires a small focal spot

• Heat dissipation requires a large spot

• This conflict is solved by slanting the target face

Page 10: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 10

Anode characteristic

1 : anode track2 : anode track

Page 11: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Induction motor• Works on the principle similar to the transformer.

• Electromagnetic induction.

• Current flowing in the stator develops a magnetic field.

• Stator windings are sequentially energized so that the induced magnetic field rotates on the axis of the stator.

• This causes the rotor to rotate.

Page 12: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Line focus principle

• The area of the x-ray tube anode from which the x-ray photons are emitted.

• This is called the actual focal spot

Page 13: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 13

THE SMALLER THE ANGLETHE SMALLER THE ANGLETHE BETTER THE RESOLUTIONTHE BETTER THE RESOLUTION

Anode angle (II)

Angle

Incident electron beam width

Apparent focal spot size

Actual focal spot size

Film

Angle

Incident electron beam width

Increased apparent

focal spot size

Actual focal spot size

Film

‘ 

Page 14: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Line focus principle

• Was incorporated into x-ray tube targets to allow a large area for heating while maintaining a small focal spot.

• The effective focal spot is the area projected onto the patient and film.

Page 15: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Line focus principle

• Focal spot sizes always make reference to the effective focal spot.

• The lower the target angle, the smaller the effective focal spot size.

Page 16: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Line focus principle

• The advantage of the line-focus principle is that it provides the detail of a small focal spot while allowing a large amount of heat dissipation.

Page 17: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Line focus principle

• The unfortunate bi-product of the line-focus principle is the “anode heel effect”

Page 18: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Anode heel effect

• Construction phenomenon that causes the x-ray photons exiting the tube on the cathode side to have a greater energy value than those exiting the tube on the anode side.

Page 19: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Anode heel effect• More energy absorption

occurs at the anode heel resulting in less energy value from the incident photons at the anode heel.

• This is used to advantage when imaging anatomical parts that are unequal in thickness and densities throughout their respective lengths.

Page 20: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Using the anode heel effect

• The following anatomical parts may be imaged using the anode heel effect:• Thoracic vertebrae• Humerus • Femur • Tibia & fibula • Forearm

Page 21: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 21

Anode heel effect (I)

• Anode angle (from 7° to 20°) induces a variation of the X Ray output in the plane comprising the anode-cathode axis

• Absorption by anode of X photons with low emission angle

• The magnitude of influence of the heel effect on the image depends on factors such as :

• anode angle• size of film• focus to film distance

• Anode aging increases heel effect

Page 22: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 22

• The heel effect is not always a negative factor

• It can be used to compensate for different attenuation through parts of the body

• For example:• thoracic spine (thicker part of the patient

towards the cathode side of the tube) • mammography

Anode heel effect (II)

Page 23: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 23

Focal spot size and imaging geometry

• Focal spot finite size image unsharpened

• Improving sharpness small focal spot size

• For mammography focal spot size 0.4 mm nominal

• Small focal spot size reduced tube output (longer

exposure time)

• Large focal spot allows high output (shorter exposure time)

• Balance depends on organ movement (fast moving organs

may require larger focus)

Page 24: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 24

X-ray generator (I)

It supplies the X-ray tube with : Current to heat the cathode filament

Potential to accelerate electrons

Automatic control of exposure (power application time)

Energy supply 1000 X-ray beam energy (of which 99.9% is

dissipated as thermal energy)

Page 25: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 25

• Generator characteristics have a strong influence on the contrast and sharpness of the radiographic image

• The motion unsharpness can be greatly reduced by a generator allowing an exposure time as short as achievable

• Since the dose at the image plane can be expressed as:

D = k0 . Un . I . T• U: peak voltage (kV)• I: mean current (mA)• T: exposure time (ms)• n: ranging from about 1.5 to 3

X-ray generator (II)

Page 26: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 26

Tube potential wave form (I)

• Conventional generators• single 1-pulse (dental and some mobile systems)

• single 2-pulse (double rectification)

• three 6-pulse

• three 12-pulse

• Constant potential generators (CP)

• HF generators (use of DC choppers to convert 50Hz mains into voltages with frequencies in the kHz range) “Inverter technology”

Page 27: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 27

100%

13%

4%

Line voltage

Single phase single pulse

Single phase 2-pulse

Three phase 6-pulse

Three phase 12-pulse

0.02 s

0.01 s

kV ripple (%)

Tube potential wave form (II)

Page 28: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 28

The choice of the number of pulses (I)

• Single pulse : low power (<2 kW)

• 2-pulse : low and medium power

• 6-pulse : uses 3-phase mains, medium and high power (manual or automatic compensation for voltage drop)

• 12-pulse : uses two shifted 3-phase system, high power up to 150 kW

Page 29: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 29

• CP : eliminates any changes of voltage or tube current • high voltage regulators can control the voltage

AND switch on and off the exposure• voltage can be switched on at any moment

(temporal resolution)• kV ripple <2% thus providing low patient exposure

• HF : combines the advantages of constant potential and conventional generator• reproducibility and consistency of tube voltage • high frame rate possible

The choice of the number of pulses (II)

Page 30: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 30

Automatic exposure control

• Optimal choice of technical parameters in order to avoid repeated exposures (kV, mA)

• Radiation detector behind (or in front of) the film cassette (with due correction)

• Exposure is terminated when the required dose has been integrated

• Compensation for kVp at a given thickness

• Compensation for thickness at a given kVp

Page 31: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 31

Automatic exposure control

X Ray tube

Collimator

Beam

Soft tissue

BoneAir

Patient

Table

Grid

Cassette

AEC detectors

Page 32: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 32

X-ray equipment operation mode and application (II)

Radiography and Tomography• Single and 3 generators (inverter technology)

• output : 30 kW at 0.3 focus spot size

• output : 50 - 70 kW at 1.0 focus spot size

• selection of kV and mAs , AEC

Radiography and Fluoroscopy• Under couch equipment, three generator (inverter

technology) - continuous output of 300 - 500 W• output : 50 kW at 1.0 focus size for spot film

• output : 30 kW at 0.6 for fluoroscopy (high resolution)

• priority given to contrast

• automatic settings of kV

Page 33: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

6: X Ray production 33

X-Ray equipment operation mode and application (III)

• Radiography and Fluoroscopy• Over couch equipment, three phase generator (inverter

technology) - continuous output of at least 500 W

• output : 40 kW @ 0.6 focus size for spot film

• output : 70 kW @ 1.0 for fluoroscopy (high resolution)

• priority given to contrast

• automatic settings of kV

• Cardiac angiography • Three phase generator - continuous output 1kW

• output : 30 kW @ 0.4 focus size

• output : 80 kW @ 0.8 focus size

• frame rate : up to 120 fr/s

Page 34: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Protective housing

Page 35: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Protective housing• X-ray tube is always mounted inside a lead-

lined protective housing that is designed to:• Prevent excessive radiation exposure.

• Prevent electric shock to the patient and operator (technologist).

Page 36: RADIATION PROTECTION IN DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY L 6: X Ray production IAEA Training Material on Radiation Protection in Diagnostic and

Protective housing• Incorporates specially designed high-voltage

receptacles.• Provides mechanical support for the x-ray tube and

protects it from damage.• Some tube housings contain oil in which the tube

is bathed.• Some tube housings contain a cooling fan to air-

cool the tube.• When properly designed, they reduce the level of

leakage radiation to less than 100 mR/hr at 1 meter when operated at maximum conditions.