QUICK-START GUIDE AND SAFETY INSTRUCTIONS

Transient Immunity Generators: compact NX NSG 3000A And Accessories: coupling NX CDN 30x0A DCD CCI CNI 508N2 HSC 4-8 SPN 508N2

Page 2: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 2 / 50

AMETEK CTS GmbH Sternenhofstrasse 15 4153 Reinach BL1 Switzerland Phone: +41 61 204 41 11 Fax: +41 61 204 41 00 URL: http://www.ametek-cts.com Copyright © 2021 AMETEK CTS GmbH All right reserved.

Specifications subject to change

Page 3: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 3 / 50

Contents

1. Safety................................................................................................................................ 6 1.1. Safety Aspects .................................................................................................................................... 6 1.2. Safety and warning label on the device .............................................................................................. 6 1.3. Responsibility of the operator ............................................................................................................. 7 1.4. General hazard ................................................................................................................................... 7 1.5. Qualification of personnel ................................................................................................................... 8 1.6. Warranty Terms .................................................................................................................................. 8 1.7. Prohibition of unauthorized conversions and modifications ............................................................... 8 1.8. Specific accessories required for safety reason ................................................................................. 8 1.9. Procedure in case of hazard ............................................................................................................... 8 1.10. Temperatures at the surface .............................................................................................................. 8 1.11. Lifting and carrying instructions .......................................................................................................... 8

2. Before Activating the Equipment ................................................................................... 9 2.1. Damage due to Shipment ................................................................................................................... 9 2.2. Power requirements ........................................................................................................................... 9 2.3. 3-phase EUT power lines ................................................................................................................. 11 2.4. Fuse for the EUT power supply ........................................................................................................ 11

3. Safety functions ............................................................................................................ 12 3.1. Safety circuit ..................................................................................................................................... 12 3.1.1. Safety circuit for compact NX and NSG 3000A series ..................................................................... 12 3.1.2. Safety circuit on Rack Systems ........................................................................................................ 12 3.2. Warning lamp.................................................................................................................................... 13 3.2.1. Warning lamp for compact NX and NSG 3000A series ................................................................... 13 3.3. Earthing of devices ........................................................................................................................... 14 3.4. FI Fault current protection ................................................................................................................ 14

4. Testing and Precautions ............................................................................................... 15 4.1. Precautions to be taken .................................................................................................................... 15 4.2. Failures and Heavy Strain ................................................................................................................ 15 4.3. Coupling Networks ........................................................................................................................... 15 4.4. Interference to the environment........................................................................................................ 16

5. Testing............................................................................................................................ 17 5.1. Safety Precautions ........................................................................................................................... 17 5.2. Danger from EUT ............................................................................................................................. 17

6. Quick start guide compact NX - NSG 3000A ............................................................... 18 6.1. Scope of delivery .............................................................................................................................. 18 6.1.1. Accessories ...................................................................................................................................... 18 6.1.2. Options ............................................................................................................................................. 19 6.2. Installation put in service .................................................................................................................. 20 6.3. Safety instructions for installation and initial installation .................................................................. 20 6.3.1. Qualifications of the staff .................................................................................................................. 20 6.3.2. Installation ......................................................................................................................................... 20 6.4. Installation of the compact NX / NSG 3000A .................................................................................. 21 6.4.1. Installation site .................................................................................................................................. 21 6.4.2. Voltage free switching ...................................................................................................................... 21 6.4.3. Grounding and power connection..................................................................................................... 21 6.4.4. Mains Switch and fuse ...................................................................................................................... 22 6.4.5. Connecting the generator to the ground reference .......................................................................... 22 6.5. Fuses for EUT with smaller nominal currents ................................................................................... 23 6.6. EUT power ON/OFF with supply Line to Line (400V) ...................................................................... 23 6.7. Front view (model-specific distinctions) ............................................................................................ 24 6.8. Rear view (model-specific distinctions) ............................................................................................ 25 6.9. Phase indicator (compact NX-models only) ..................................................................................... 25 6.10. Power ON the compact NX - NSG 3000A generator ....................................................................... 25

7. How to use compact NX - NSG 3000A generator ........................................................ 27 7.1. 1. General safety instructions ........................................................................................................... 27 7.2. 2. Front connections ......................................................................................................................... 27 7.3. 3. Rear connections .......................................................................................................................... 27 7.4. 4. Control elements .......................................................................................................................... 28 7.5. 5. Home screen ................................................................................................................................ 28

Page 4: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 4 / 50

7.6. 6. Quick start..................................................................................................................................... 28 7.7. 7. Pulse settings ............................................................................................................................... 29 7.8. Info: Pulse settings ........................................................................................................................... 29 7.9. Info: Pulse settings ........................................................................................................................... 29 7.10. 8. Starting a test ............................................................................................................................... 30 7.11. 9. The test run................................................................................................................................... 30 7.12. Info: Button Status indicators ............................................................................................................ 30

8. Quick start guide coupling NX - CDN 30x3A ............................................................... 31 8.1. Grounding and power connection..................................................................................................... 31 8.2. Mains Switch and fuse ...................................................................................................................... 32 8.3. Connecting coupling NX with compact NX generator ...................................................................... 32 8.4. Connecting CDN 30x3A-C32 with NSG 30x0A generator ............................................................... 33 8.5. Front view (model-specific distinctions) ............................................................................................ 35 8.6. Rear view (model-specific distinctions) ............................................................................................ 35 8.7. Operation with external CDN or generator internal CDN ................................................................. 36

9. Quick start guide DCD sr- and st-series ...................................................................... 37 9.1. Test Setup for DCD application ........................................................................................................ 37 9.2. AE Protection for auxiliary equipment .............................................................................................. 38 9.3. Device functions and operating ........................................................................................................ 38 9.4. Coupling network DCD sr ................................................................................................................. 38 9.5. Coupling Line to Line ........................................................................................................................ 39 9.5.1. Coupling with 40 Ω resistor and 0.5 µF capacitor ............................................................................. 39 9.5.2. Coupling with 40 Ohm and Gas Discharge Tube (GDT) .................................................................. 39 9.6. Coupling to GND .............................................................................................................................. 40 9.6.1. Test setup Coupling to PE ................................................................................................................ 40 9.7. Coupling network DCD st, 4 lines ..................................................................................................... 40 9.8. Coupling network DCD st, 8 lines ..................................................................................................... 41 9.9. Test setup for 10/700 µs impulses ................................................................................................... 41

10. Quick start guide capacitive coupling clamp CCI ...................................................... 43

11. Quick start guide HSC 4-8, CNI 508N2, SPN 508N1 .................................................... 44 11.1. HSC 4-8 Coupling / decoupling network .......................................................................................... 44 11.2. CNI 508 N2 Coupling / decoupling network ..................................................................................... 44 11.3. SPN 508 N1 Surge Protection Network ........................................................................................... 44 11.4. Front and rear view HSC 4-8 ............................................................................................................ 45 11.5. Front and rear view CNI 508N2 ........................................................................................................ 45 11.6. Front and rear view SPN 508N1....................................................................................................... 45 11.7. General ............................................................................................................................................. 46 11.8. Test methods for Surge and Ring Wave to high speed datalines .................................................... 46 11.9. Test setup ......................................................................................................................................... 47 11.10. Surge test to shielded datalines ....................................................................................................... 47 11.11. Surge test to datalines ...................................................................................................................... 49 11.12. Surge test to data-lines with SPN 508N1 protection ........................................................................ 49 11.13. Burst test to data-lines ...................................................................................................................... 49

12. Maintenance, Adjustments, Replacement of Parts..................................................... 50 12.1. Disposal of devices ........................................................................................................................... 50 12.2. Calibration and Verification ............................................................................................................... 50

Page 5: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 5 / 50

Page 6: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 6 / 50

1. Safety

1.1. Safety Aspects Observe all precautions to assure your personal safety. The generators comply with Installation Category II (excess voltage section).

Read the user / operation manual carefully. Pay special attention to safety and operation details!

1.2. Safety and warning label on the device

Please take note of the following explanations of the symbols used in order to achieve the optimum benefit from this manual and to ensure safety during operation of the equipment.

This symbol warns of a potential risk of shock hazard. The symbol on an instrument shows that that it can source 1000 volt or more, including the combined effect of normal and common mode voltages. Use standard safety precautions to avoid personal contact with these voltages.

This symbol indicates where a caution is required. Refer to the operating instructions located in the manual in order to protect against personal injury or damage the equipment.

GROUND Indicates protective Ground Terminal

CAUTION The CAUTION symbol indicates a potential hazard. It calls attention to a procedure, practice or condition which, if not followed, could possibly cause damage to equipment. Such damage may invalidate the warranty. If a CAUTION is indicated, do not proceed until its conditions are fully understood and met.

WARNING

The WARNING symbol indicates a potential hazard. It calls attention to a procedure, practice or condition which, if not followed, could possibly cause bodily injured or death. If a WARNING is indicated, do not proceed until its conditions are fully understood and met.

Power Supply The equipment is intended to operate with a power supply not to exceed 250 volts between phase and neutral or between phase and ground. A proper ground connection through the ground connector of the power cord is essential for safe operation.

Grounding the Generators The generators are grounded through the power cord. To avoid electric shock, plug the power cord into a properly installed receptacle which was tested by a qualified electrician. Have the test performed before connecting equipment.

Without the protective ground connection, all parts of the generators are potential electric shock hazards. This may include components which appear to be insulated. The equipment MUST NOT BE USED if this protection is altered.

Use the Proper Power Cord Use only power cords and connector specified for your product. Use only power cords in good condition.

Use Proper Fuses To avoid fire hazard, use only fuses as specified in the parts listing for your product - matching type, voltage and current rating.

Do Not Remove Covers or Panels To avoid personal injury, do not operate the generators without panels and covers.

Do Not Operate in an Explosive Environment

Electric Overload Never apply power to a connector which is not specified for that particular voltage/current.

Page 7: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 7 / 50

CAUTION

Before using this equipment, read the user manual and the

separate delivered safety manual carefully

1.3. Responsibility of the operator

These operating instructions form an essential part of the equipment and must be available to the operator at all times. The user must obey all safety instructions and warnings.

CAUTION

The purpose of this instrument is the generation of defined interferences signals for EMI immunity testing. Depending on the arrangement of the test rig, the configuration, the cabling and the properties of the EUT itself, a significant amount of electromagnetic radiation may result that could also affect other equipment and systems.

The equipment is designed to operate in industrial environment. For operating in other or sensitive environment, such as light industry, airport area…, the user may use a shielded room for operate.

The user himself or herself is ultimately responsible for the correct and controlled operation of the rig. In case of doubt, the tests should be carried out in a Faraday cage.

1.4. General hazard

Before applying power to the system, verify that your product is configured properly for your particular application.

WARNING

The generators and their accessories operate at high voltages.

Hazardous voltages may be present when covers are removed. Qualified personnel must use extreme caution when servicing this equipment.

Circuit boards, test points, and output voltages also may be floating above (below) chassis ground. The design of external insulation must be such that it exceeds the maximum impulse voltages of the generator (5.5 kV or 7.0 kV).

Only qualified personnel who deal with attendant hazards in impulse generators, are allowed to perform installation and servicing.

Ensure that the AC power line ground is connected properly to the Power Rack input connector or chassis. Similarly, other power ground lines including those to application and maintenance equipment must be grounded properly for both personnel and equipment safety.

Always ensure that facility AC input power is de-energized prior to connecting or disconnecting any cable.

The user must ensure that the output power lines are labeled properly as to the safety hazards and that any inadvertent contact with hazardous voltages is eliminated. Guard against risks of electrical shock during open cover checks by not touching any portion of the electrical circuits. Even when power is off, capacitors may retain an electrical charge. Use safety glasses during open cover checks to avoid personal injury by any sudden component failure. Neither AMETEK CTS GmbH, nor any of the subsidiary sales organizations can accept any responsibility for personnel, material or inconsequential injury, loss or damage that results from improper use of the equipment and accessories.

Page 8: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 8 / 50

CAUTION

Personnel fitted with a heart pacemaker must neither operate the instrumentnor approach the test setup while a test is being executed.

Only approved accessories, connectors, adapters, etc. are to be used to ensure safe operation.

1.5. Qualification of personnel

The compact NX5 / NX7 must be operated only by authorized and trained specialists.

1.6. Warranty Terms

AMETEK CTS provides this written warranty covering the product stated above, and if the buyer discovers and notifies AMETEK CTS in writing of any defect in material or workmanship within the applicable warranty period stated above, then AMETEK CTS may, at its option: repair or replace the product; or issue a credit note for the defective product; or provide the buyer with replacement parts for the product.

The buyer will, at its expense, return the defective product or parts thereof to AMETEK CTS in accordance with the return procedure specified below. AMETEK CTS will, at its expense, deliver the repaired or replaced product or parts to the buyer. Any warranty of AMETEK CTS will not apply if the buyer is in default under the purchase order agreement or where the product or any part thereof:

• is damaged by misuse, accident, negligence or failure to maintain the same as specified or required by AMETEK CTS;

• is damaged by modifications, alterations or attachments thereto which are not authorized by AMETEK CTS; • is installed or operated contrary to the instructions of AMETEK CTS; • is opened, modified or disassembled in any way without AMETEK CTS’s consent; or • is used in combination with items, articles or materials not authorized by AMETEK CTS.

The Buyer may not assert any claim that the products are not in conformity with any warranty until the buyer has made all payments to AMETEK CTS provided for in the purchase order agreement.

1.7. Prohibition of unauthorized conversions and modifications

The user is not entitled to the device to perform its own modifications and adaptations. Modifying parts on the generator by unauthorized persons will void the warranty of the device and the correct functioning cannot be guaranteed.

1.8. Specific accessories required for safety reason

Only use accessories approved by AMETEK CTS for these generators and intended as accessories for these devices. Measuring instruments for the measurement of instrument parameters shall be designed for the maximum voltage and current from the generator. Otherwise safety cannot be guaranteed.

1.9. Procedure in case of hazard

If a hazard could exist due to an unintended condition of the generator, the following procedure is recommended: Disconnect the device- and EUT power supplies from the power supply and ensure that the generator is always earthed via the supply lines or a different ground connection. Wait at least 15 minutes and ground all outputs via a 10 kΩ, 15 W resistor. Call an AMETEK service center.

1.10. Temperatures at the surface

The components that generate heat in the device (resistors, etc.) are arranged inside the device in such a way that no temperatures higher than 50 °C occur on the surface.

1.11. Lifting and carrying instructions

The generators usually weigh more than 20 kg. For lifting and carrying, two persons must always lift the equipment together or use a suitable lifting device. The device must be gripped by the carrying handle and on the underside.

Page 9: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 9 / 50

2. Before Activating the Equipment

2.1. Damage due to Shipment The instrument was tested before shipment and was packed carefully on a transport palette. Each box is marked with a detailed list of the contents.

Before activating the equipment, check for damage which may have occurred during shipment. Check each container as well as the generator itself. In case of physical damage, contact the manufacturer before operating the unit.

Figure 1 – Instrument

Figure 2 – Rack

Figure 3 - System

2.2. Power requirements Prior to turning the equipment on, check that the selected voltage corresponds to the supply voltage.

Page 10: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 10 / 50

The position of the voltage selector must correspond with the mains. If you change the mains voltage, replace the fuses according the recommended value on the number plate.

First - check the 115 V / 230 V voltage selector where applicable!

Do not connect 230 VAC into the 115 VAC units.The result could be a severely damaged unit.

Figure 4 - Power Settings

If power to the equipment of Class I is supplied by an auto-transformer which is connected to a higher supply voltage, the base of the auto-transformer must be connected to the neutral of the power supply.

Equipment of Class I, which is supplied with a 3-phase power cord, must only be connected to a grounded receptacle. The protective ground connection must not be disconnected or otherwise interrupted.

Each interruption or disconnection of the protective ground connection inside or outside the equipment may cause electric shock.

Connecting different units with each other or to a personal computer for remote control may be performed only with cables recommended and/or supplied by the manufacturer.

Page 11: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 11 / 50

2.3. 3-phase EUT power lines The connectors for 3-phase EUT power lines on the devices is realized with CEE connectors. Please take care to connect carefully the correct plugs during installation. Important is a direct connection between neutral and PE from the power supply. Otherwise you get problems with the floating neutral star point and inconvenience or malfunctions during the test. The figures show the correct mapping of the plugs of the CEE connectors.

Figure 5 - Female Receptacle

Figure 6 - Male Plug

2.4. Fuse for the EUT power supply

The pulse generators have no built in fuse for the EUT power supply. It is in the scope of responsibility of the user to protect the EUT external for the rated current.

CAUTION

The design of the external fuse must be matching the following rules:

fuse dimension must be equal or smaller than the rated EUT current of the connected test generator

fuse must be designed for protect the connected EUT device under test in malfunction

Example of external fuse Fuse in the building is designed for 32A. A Fuse box with 16A fuse protection is installed between the building supply and the test generator. Test generator and EUT are now fused for 16A rated current

Figure 7 - External Fusing

Example 16A protection with a 100A installation with a 3-phase fuse box.

The user has protected his Equipment Under Test with this additional fuse box for a rated current of 16A.

Figure 8 - Example of External Fuse

Page 12: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 12 / 50

3. Safety functions

The test area must be organized so that only those involved in the test may enter it. In the case that the safety circuit is used to control the complete area, an additional interlock contact must be used to directly protect the operator from contact with the DUT. Neither the DUT nor any cables or accessories must be touched during the test. During work on the DUT, the test procedure must be stopped and the DUT disconnected from the power mains supply. Coupling devices have no safety functions built in, because these functions are integrated in the generators.

3.1. Safety circuit

The safety circuit will switch off the high voltage and the TEST ON button of the unit.

3.1.1. Safety circuit for compact NX and NSG 3000A series

The safety circuit locks the system and enables the generation of the high voltage impulses in the generators.

Design Each device that has internal relevant high voltage unit, includes a safety circuit. The safety circuit works like an “open collector circuit”, where the external safety loop must be closed for switch on the high voltage.

Safety circuit closed The device will generate high voltage pulses after start

Safety circuit open The device will switch off the high voltage and discharges the high voltage circuit

Figure 9 - Safety Circuit

Figure 10 - Safety Circuit Diagram

3.1.2. Safety circuit on Rack Systems

Rack Systems have a separate safety circuit, driven by the mains supply voltage. Release the Safety Circuit button will disconnect the mains power from the devices and EUT supply. For power on it is necessary to unlock the red colored safety button manually by turning.

Figure 11 - Rack Stop Button

National Standards for safety circuits can differ from the example above.

The user is responsible for the correct design of his safety circuit.

Page 13: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 13 / 50

Important for operation

Connect the delivered Safety Circuit Terminal SCT or the optional Safety

Circuit Adapter SCC AD to the SYSLINK plug.

The generator does not start any test if the safety circuit is not connected and closed.

NOTE: The short circuit is already short-circuited at the delivered Safety circuit terminal SCT and Safety Circuit Adapter SCC AD

.

Figure 12 - Safety and Warning Lamp Connectors

SCT Safety Circuit termination

The safety circuit terminal SCT is part of the delivery and includes the short circuit for the safety circuit. If an external safety circuit is required, please order the optional Safety Circuit Adapter SCC AD.

Figure 13 - SCC AD option for external safety circuit loop

Figure 14 - Safety

Termination

3.2. Warning lamp

3.2.1. Warning lamp for compact NX and NSG 3000A series

The warning lamp offers a voltage free contact that indicates the status of the generator system.

Design Each device with warning lamp function can short thewarning lamp contact. An external powered relays max. 60 V / 2 A is controlling the lamp. The user is responsible for the warning lamp power supply Warning Lamp switch closed (red): - Safety circuit is closed and TEST ON is on

Status: RED Lamp, Power on EUT output, High voltage can be “switched ON” Warning Lamp switch opened (green): - Safety circuit is open and / or TEST ON is off - Generator is switched off

Status: GREEN Lamp, no danger, High voltage is off

Figure 15 - Warning Lamp Connection

Page 14: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 14 / 50

3.3. Earthing of devices Earth Bolt

Generators must be grounded to the reference ground plane. Generally, the generators are equipped with an aluminum earth bolt (8x30mm) at the rear side of the device.

For contact the earth bolt there are the following connections: - Screw M4 x 12 mm tapped hole - Ø 4mm x 20mm holes for 4mm banana plugs The drawing shows the dimensioning of the earth bolt.

Figure 16 - Earth Connection

Earth connection with a tress

The earth connection must be designed as a low ohmic and low inductance connection as illustrated as example in the picture at the right side.

This solution allows the user for fast exchange the device.

Figure 17 - Earth Connection

Tress Earth connection with a brass bar

Alternative method is using a brass bar equipped with banana plugs and additional fixed with a M4 screw as illustrated at the right-side picture.

This is a solution preferred for fix mounted test setup installation.

Figure 18 - Earth with Brass Bar

3.4. FI Fault current protection

The standards recommend decoupling and filter capacitors to PE for decoupling surge pulses. This is the reason for tripping fault current protection relays interrupt the mains to the EUT supply. For eliminate the circumstance use one of the following options: Remove the fault current protection This solution does not limit the current to PE. The surge test as other EMC tests with higher currents to PE are possible. The user must take care to the circuit with no fault current protection. Only trained professional people are allowed to perform such tests. Using of insulating transformers An insulating transformer separates the circuit from the protected path. The synchronization works between line and PE and therefore a connection between neutral and PE is necessary. Without this connection the stray capacitance defines the phase sync.

Figure 19 - Example with Insulating Transformers

Behind the insulating transformer the neutral and PE must be connected for a proper phase synchronization of the generator. In case of a 3-phase system the user must perform similar.

Page 15: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 15 / 50

4. Testing and Precautions

All tests offered by High Voltage or EMC generators are immunity tests on electronic equipment or devices. These tests are potentially dangerous to the operator. It is the responsibility of the user to avoid critical failures and risks to the environment and the operator.

National and International regulations regarding human safety must be followed.

Individuals with certain health conditions (e.g. with a heart pacemakers or similar devices), should not participate in testing.

Long power supply lines to the EUT may radiate energy which may interfere with other instruments unrelated to the generator. It is the responsibility of the user to determine whether to conduct immunity tests in a given area. Generator and coupling/decoupling voltage power supplies must be grounded and connected to reference plane.

4.1. Precautions to be taken • EMC and High Voltage test areas must always be supplied by a decoupled and well-known power supply. • The decoupling can be realized either by

- filtering - or the use of an isolating transformer

• When using isolating transformers, the fault current safety devices normally don’t trip. • EMC and High Voltage test areas must always have a clear grounding strategy. All coupling/decoupling high power

voltages and all LISN’s must be strictly and securely connected to the ground reference plane of the test setup. • Please ensure the correct current back flow to the generator. Missing connections can cause extremely high voltages and

may become a health hazard. Therefore, it is absolutely necessary to use ground reference planes which are connected to a protective ground.

4.2. Failures and Heavy Strain If it is determined that a safe operation of the equipment due to failure or heavy strain is no longer possible, the supply voltage must be disconnected, and the equipment protected from unintentional use.

Unsafe operation is determined as follows: • the equipment shows visible damage • the equipment is not working • the equipment has experienced severe stress during transport • the equipment was stored in an unsuitable environment for an extended period of time.

Please exercise good judgement in determining whether a device is unsafe for operation.

4.3. Coupling Networks • The coupling network has mostly no On / Off switch and no internal fuse for the EUT power supply. This is caused the

different regulation in each country. The device under test must be protected by the user in an adequate safe solution. As an option special adapters and switches can be built in, but the user must specify this special solutions.

• Generators and coupling devices must be grounded and connected to the reference ground.

• For coupling pulses to the lines, the coupling path must be setted.

• If a line has not to be coupled, it is necessary to disconnect or switch off this coupling path.

• Special safety adapter cables are part of the delivery.

Page 16: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 16 / 50

4.4. Interference to the environment

The AMETEK CTS interferences generators are instruments with a functionally emission of electromagnetic interference during the test (e.g., ESD, EFT, conducted RF, etc.). Therefore, a disturbance of the environment cannot be excluded.

The user has an obligation to use a suitable test environment in order to minimize influences to the environment. This may require a suitable screening or to test in a shielded room.

Next should be considered during the test, that nearby are no disturbers such as (switching relays or drives with frequency converters etc.) that can influence through their emission the test.

Page 17: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 17 / 50

5. Testing

The generator can be operated locally from the front panel or remotely from the computer. When it is necessary to observe the EUT during the test, it can be more comfortable to operate locally. The tester is nearer to the EUT and therefore faster to switch off the test generator in EUT fail condition. When the computer fails, the Test On button on the generator front or the Fail 1 input must be used for switching off the test.

5.1. Safety Precautions • The test area must be secured that only authorized test personnel may enter it. • Working alone with high voltages is dangerous. • The high voltages must be switched off when nobody is present. • Neither the EUT nor cables or accessories must be touched during the test. • Make sure all high voltage connections are adequately taped or otherwise insulated to prevent accidental contact by you

or neighboring colleagues. • Keep one hand in your pocket when probing high voltage circuits or discharging capacitors. This reduces the peril to

touch the high voltage with both hands. • While working on the EUT, the test procedure should be stopped and the EUT disconnected from the voltage power

supply. • The EUT must be tested within a safety container or within a protected area. In extreme circumstances the EUT may

ignite or explode as a result of internal damage.

5.2. Danger from EUT

CAUTION

The device being tested may become defective and ignite due to the influence of the applied test signal.

The storraged energy inside the test simulator must be to consider. This energy can destroy or damage the EUT even when the EUT operates under normal conditions.

Therefore, the operator shall take the following precautions:

• As soon as the EUT ceases to operate as intended, the test shall be stopped immediately. • In case of internal damage, the operator may be exposed to high frequency signals of high power (up to 75 Watts and

more) anywhere on the EUT. • Cables and connectors can be overloaded by high voltages or energies. • Due to internal damage of components fire and/or explosion may occur. • Unintended use of the EUT may cause hazardous situations in the vicinity of the test area. • The user is responsible for the correct EUT protection. The device under test is to be secured in such a way that no

dangerous conditions arise.

CAUTION

Never touch the EUT or anything connected to the EUT during a test

It is absolutely necessary to observe and comply with all safety instructions.

Page 18: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 18 / 50

6. Quick start guide compact NX - NSG 3000A

6.1. Scope of delivery Using the following list, check that all the items ordered have been delivered:

Item Name Remark Picture 1a compact NX5 or

compact NX7 compact NX generator including - Single phase coupling network

Including ordered Modules - Burst Module - Surge Module - Ring wave Module (NX7 only) - Power Fail Module - Telecom Surge Module

1b NSG 3040A or

NSG 3060A NSG 3000A generator including - Single phase coupling network

Including ordered Modules - Burst Module - Surge Module - Ring wave Module (NSG 3060A only) - Power Fail Module - Telecom Surge Module

2 Power Mains cable Power Mains cable

- Connectors country coded

3 ESC EUT Supply Cable - Connectors country coded

4 ESA 1 EUT Supply Adapter

- Connectors country coded

5 SCT

#112801 Safety Circuit Terminator (Sys. Link) Short circuit for Interlock (no ext. loop), Alternative with ext. loop: SCC AD (option)

6 SWL AD

#111241 Warning Lamp Adapter

7 Ethernet cable

#107460 Ethernet crossover network cable RJ45, Cat 6, SF/UTP, red

8 Cleaning tissue

9 Safety Manual Safety Manual

200 / 500 / NX Series

10 USB Memory card Files on USB Memory card

11 User Manual User manual (pdf on the memory card)

12 Iec.control Software Software iec.control, (on the memory card)

13 Software license

If ordered on a license sheet - UOC Optolink Converter USB to LWL - Optical fiber cable 5m

6.1.1. Accessories If additional equipment is ordered refer to the user manual of these devices

Page 19: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 19 / 50

Name Remark Picture SLC xxx Sys Link Cable with various cable length UOC #111311

USB Optolink Converter (USB to LWL) Optical Fiber cable, 5m Remark: The USB Optolink is included in the software license

Copper braid Earth band connection NX-generator to coupling NX,

Dimension: 300 x 23 mm, 25mm2, 4 x Screws M4 x 10mm

6.1.2. Options

Name Remark Picture PVF BKIT 1 Pulse Verification Fast Burst Kit 1

PVF 50, Pulse Verification Fast 50 Ohm PVF 1000, Pulse Verification Fast 1000 Ohm PVF AD 1 Pulse Verification Fast Adapter 1 –

Multi Contact (MC) to SHV fix

PVF AD1 PVF 50 or PVF 1000

CCI PVKIT 1 Industrial Capacitive Coupling Clamp Pulse Verification Kit 1

Transducer plate Support PVF AD3 Pulse Verification Fast Adapter 3 – MC to SHV

ESS 1 #111607

ESS 1 Interlock for compact NX5 system

Switches OFF High voltage and EUT power supply

SCC AD #111240

Safety Circuit Adapter (Sys. Link) Short circuit for Interlock for external safety loop Alternative to: SCT Safety Circuit Terminator

coupling NX / CDN 30x3A

External 3-phase coupling network coupling NX: AC voltage: 3 x 400 V / 690 V DC voltage: up to 1000 VDC Current: 16 A up to 200 A CA and or DC

CDN 30x3A: AC voltage: 3 x 480 V DC voltage: up to 500 VDC Current: 16 A up to 100 A AC and or DC

CCI Capacitive Coupling Clamp Industrial For coupling of Burst impulses to signal and data-lines.

BCC 1000 Burst connection cable SHV 1000 mm

SWL #111594

Safety Warning Lamp Safety warning lamp for NX series

SVP CL 100 Calibration load resistor for power fail generators

Low inductive 100 Ohm carbon load resistor

Page 20: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 20 / 50

SVP CL 1700 Inrush current measurement for power fail generators Calibration box 1700uF for inrush current verification

6.2. Installation put in service This chapter includes a checklist with steps that should be taken before the generator is switched on and put into operation.

6.3. Safety instructions for installation and initial installation

National regulations in installation and operation of electrical equipment must be respected.

CAUTION

The compact NX / NSG 3000A test system is not suitable for use in an explosive atmosphere.

CAUTION

Connect the EUT only after the initial system setup has finished.

6.3.1. Qualifications of the staff Knowledge of electrical engineering and electromagnetic compatibility is required to install the unit. The national regulations for installation of electrical equipment as well as the safety at work of electrical equipment must be known to the user.

6.3.2. Installation The test system conforms to protection class 1. Local installation regulations must be respected to ensure the safe flow of leakage currents.

CAUTION

Operation without a ground connection is forbidden!

Two independent ground connections are necessary - one for the test system and one for the EUT. These must be connected back to the local permanent installation or to a fixed, permanent ground conductor.

Operate the equipment only in dry surroundings. Any condensation that occurs must be allowed to evaporate before putting the equipment into operation. Do not exceed the permissible ambient temperature or humidity levels. Use only officially approved connectors and accessory items.

Ensure that a reliable return path for the interference current is provided between the EUT and the generator. The ground reference plane and the ground connections to the instruments, as described in the relevant test standards, serve this purpose well.

The test system may only be opened by a qualified specialist upon specific instruction given by the manufacturer. The equipment works, on principle, with two independent power supplies, one for the generator and one for the EUT. The generator must be disconnected from both sources before any modifications to the test setup are undertaken. Besides the mains connections themselves, certain components also operate at high voltages, and are not provided with any form of extra protection against accidental contact. The system complies with the safety requirements of IEC/EN 61010-1 (Safety requirements for electrical equipment for measurement, control and laboratory use).

It is the user’s responsibility to ensure that the test rig does not emit excessive electromagnetic interference (EMI) that might affect other equipment. The test system itself does not produce any excessive radiation; however, the injection of interference pulses into the EUT can result in the device and/or its associated cables radiating EMI. To avoid radiating unwanted interference the standards organizations recommend that the test setup be located in a Faraday cage.

Since the purpose of the test system is to produce interference signals for interference immunity testing, the requirements in the IEC/EN 61000 series concerning limiting the radiated EMI can only be complied with by operating the test system inside a

Page 21: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 21 / 50

Faraday cage.

6.4. Installation of the compact NX / NSG 3000A

6.4.1. Installation site Place the test system so that there is enough free space around the cooling air inlets on both sides and behind the fan outlet on the rear panel.

6.4.2. Voltage free switching Make sure that the cables can be connected without hindrance and disconnected without problems. The complete voltage disconnection from the device is carried out by disconnecting the cables. Make sure that capacitors inside the device can still be under voltage, even in an emergency.

6.4.3. Grounding and power connection Two independent ground connections are necessary- one for the test system and one for the EUT. These must be connected back to the local permanent installation or to a fixed, permanent ground conductor. To avoid electric shock the power cord protective grounding conductor must be connected to ground.

Figure 20 - Grounding and power connection

System GND The system is connected to GND via the earth wire of the connected mains power cable for the generator and the EUT supply. A separate GND wire from the reference GND plane to the building earth point may be useful to avoid interferences to other areas. EUT GND Ensure that a reliable return path for the interference current is provided between the EUT and the generator. The GND wire from the generator to the reference GND and must be designed as a low inductance connection suitable for high frequencies. The reference ground plane and the ground connections to the instruments, as described in the relevant test standards, serve this purpose well. Protection (optional) A proposal is to separate protection of the EMC system with filters, insulation transformer or fault current protection and other measures may be useful for the EMC installation. The advantage is the separation of the EMC system from all other installation. A shielded room with adequate filters may be the best solution for avoid unwanted conducted and radiated interference to other areas.

Page 22: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 22 / 50

6.4.4. Mains Switch and fuse The mains power voltage indicated on the instrument must correspond with the local supply voltage (mains voltage: 85–265 Vac, universal power unit, mains frequency: 50–60 Hz). Use cables of sufficient current capacity.

Figure 21 - Mains Switch, fuse holder and power input

To replace a fuse: 1) Disconnect the mains cable 2) Pull the fuse holder out of the connector 3) Remove the damaged fuse(s) 4) Insert 1 or 2 fuses (4 A / 115V and 2 A / 230 V slow blow) 5) Replace the fuse holder 6) Plug the mains cable into a power outlet with a solid ground

connection 7) Switch the system on and operate as instructed in this manual

6.4.5. Connecting the generator to the ground reference The earth connection to the GRP should be low inductive and low ohmic. The cross-section of the earth connection should be at least 6 mm2. A braid of stranded wire or a flat copper or brass strip is better.

Figure 22 - Connection to reference ground

Connection to reference ground For burst tests, the generator must be placed on a ground reference plane which is connected to ground.

A low inductive high frequency ground connection between the test system and the ground reference plane (GRP) is absolutely essential for performing burst tests correctly.

Figure 23 - Earth bolt dimension

Earth bolt dimensions:

Screws: Metric M4 , Distance: 10.1 mm x 11 mm Plug: Banana plug, ∅ 4 mm

Page 23: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 23 / 50

6.5. Fuses for EUT with smaller nominal currents

The AMETEK CTS pulse generators have no built-in fuse for the EUT power supply. It is in the scope of responsibility of the user to protect the EUT external for the rated current.

CAUTION

The design of the external fuse must be match the following rules:

Fuse must be equal or smaller than the rated EUT current of the connected test generator

Fuse must be designed for protect the connected EUT device under test in malfunction

Example of external circuit breaker

Circuit breakers in the building are designed for 32 A. A circuit breaker box with 16 A protection is installed between the building supply and the test generator. Test generator and EUT are now fused for 16 A rated current

Figure 24 - External Circuit Breaker

6.6. EUT power ON/OFF with supply Line to Line (400V)

CAUTION

To operate the generator internal CDN with line to line voltages of 400 V,

the generator must be designed for this higher line to line voltage

The internal electrical switch interrupts the power on line L. Using a two-phase supply (L-L), the second line is connected to the Neutral EUT path and can’t be interrupted by the generator. The following setup with an external contactor will interrupt both lines L1 and L2 (connected to N-path). It is controlled via the potential-free contact of the warning lamp. The design and installation of the contactor is in the responsibility of the user.

Page 24: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 24 / 50

Figure 25 - Line-to-Line (400V) Operation

EUT power setup for a test device powered with 400 V line to line

6.7. Front view (model-specific distinctions)

Figure 26 - Operating Elements

1 Active indication 2 Touch screen 3 Knob (Inc. / Dec /Enter) 4 " Test On" 5 CRO I (surge)

6 CRO V (surge) 7 HV pulse Burst output 50 Ω 8 Coupling (Power, Burst, Surge) 9 EUT test supply 10 Ground reference (calibration)

11 CRO Trigger output ↑ 5V 12 Start / Pause 13 Stop 14 Back 15 USB input (Memory stick)

Page 25: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 25 / 50

6.8. Rear view (model-specific distinctions)

Figure 27 - Rear view

1 Phase indication PF1 / PF2 2 EUT supply input 3 Surge output HV - COM 4 Ventilator 5 Mains Switch 6 DC output 0-10V

7 Monitor V, Monitor I 8 Fail, EUT 1, EUT2 9 External Trigger IN 10 Sys Link IN 11 USB A interface 12 USB B interface

13 Ethernet interface 14 Opto Link Interface 15 Sys Link OUT 16 Sync input 17 Mains input device 18 Reference earth connection

6.9. Phase indicator (compact NX-models only)

The phase indication shows the correct connection of the supply to Phase and Neutral input of the compact NX . For the generator hardware, both, L and N paths are potential free and a reverse connection is not relevant for the generator operation. It is in the responsibility of the user to decide to run the test in correct phase-neutral connection. Anyway, with wrong phase connection it is possible and normal to measure a voltage between neutral and earth.

Phase indication

1 compact NX5 generator

2 EUT supply may be a - direct supply from building or via a - tapped or variac transformer

3 Phase indication LED illuminated

4 Phase indication LED inactive

Note: The phase synchronization signal taken from the L

path. Figure 28 - Phase Indication

6.10. Power ON the compact NX - NSG 3000A generator

Approx. 3-4 seconds after power on the generator will drive the ventilator at full speed during few seconds. Then the ventilator returns to a variable speed control concerning the temperature.

Booting

After the welcome screen the software scans the equipment for search all existing internal phenomenon’s and external devices. A list shows and lists all detected modules.

Page 26: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 26 / 50

Figure 29 - Loading

Home screen after booting

The home screen shows all detected symbols of phenomenon with their color.

Figure 30 - The Home Screen

Page 27: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 27 / 50

7. How to use compact NX - NSG 3000A generator

7.1. 1. General safety instructions

7.2. 2. Front connections

7.3. 3. Rear connections

1. Must be operated only by authorized and trained specialists 2. Do not operate without any ground connection 3. Hazardous high voltages are generated on the output terminal

1. EUT (Equipment Under Test) supply 2. Output and coupling options

1. Phase Indication 2. EUT Supply Input (L, N, PE) 3. Mains and Power Switch 4. Monitoring, trigger and other features 5. Interfaces

Page 28: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 28 / 50

7.4. 4. Control elements

7.5. 5. Home screen

7.6. 6. Quick start

1. Phenomenon 2. Advanced Menu 3. Brings you back to the Home Screen 4. Test and Control Buttons 5. Output and coupling options

1. Each phenomenon is color coded. Tap to select.

1. Tapping QUICK START will enable you to start a test 2. Pressing either “BACK” or ← on the screen will take you one menu level back. 3. Pressing “TEST ON” at any time will both enable high voltage pulses as well as provide power to the EUT and enable pulses to be generated.

Page 29: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 29 / 50

7.7. 7. Pulse settings

7.8. Info: Pulse settings

7.9. Info: Pulse settings

1. Tap to select the parameter. Parameters written in white can be changed. Selected parameter is underlined. 2. Pulse Graphic. Press loupe-symbol to enlarge or swipe left and right to naviagte to other views. 3. Press to change adjustment range. Selected range will be shown briefly next to the selected parameter. Push and hold to „zero“ to nearest range. 4. Rotate to increment or decrement a selected parameter in steps corresponding to the selected range.

1. Tapping a selected parameter opens an on-screen setting menu.

1. Long press for detailed information.

Page 30: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 30 / 50

7.10. 8. Starting a test

7.11. 9. The test run

7.12. Info: Button Status indicators

1. Press “TEST ON” to enable High voltage and power to the EUT 2. Press “START/PAUSE” to start a test. 3. Long press “START/PAUSE” for continuous test.

1. Press “START/PAUSE” to pause or press “STOP” to stop a test. 2. Press “START/PAUSE” to continue or start a test. 3. After Stop press “BACK” to return to Menu or tap the ←.

White: HV and EUT Power Disabled Yellow: HV and EUT Power Enabled

Green: Ready or Paused Orange: Testing

Red: Testing Red blinking: Error

Page 31: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 31 / 50

8. Quick start guide coupling NX - CDN 30x3A

8.1. Grounding and power connection Two independent ground connections are necessary- one for the test system and one for the EUT. These must be connected back to the local permanent installation or to a fixed, permanent ground conductor. To avoid electric shock, the power cord protective grounding conductor must be connected to ground.

Figure 31 - Grouding requirements

System GND The system is connected to GND via the earth wire of the connected mains power cable for the coupling, generator and the EUT supply. A separate GND wire from the reference GND plane to the building earth point may be useful to avoid interferences to other areas. EUT GND Ensure that a reliable return path for the interference current is provided between the EUT and the generator. The GND wire from the generator to the reference GND and must be designed as a low inductance connection suitable for high frequencies. The reference ground plane and the ground connections to the instruments, as described in the relevant test standards, serve this purpose well. Protection (optional) A proposal is to separate protection of the EMC system with filters, insulation transformer or fault current protection and other measures may be useful for the EMC installation. The advantage is the separation of the EMC system from all other installation. A shielded room with adequate filters may be the best solution for avoid unwanted conducted and radiated interference to other areas. Connection to reference ground

Figure 32 - Earth bolt dimension

Earth bolt dimensions:

Screws: Metric M4 x 10 mm Distance: 10.1 mm x 11 mm Plug: Banana plug, ∅ 4 mm

Page 32: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 32 / 50

8.2. Mains Switch and fuse The mains power voltage indicated on the instrument must correspond with the local supply voltage (mains voltage: 85–265 Vac, universal power unit, mains frequency: 50–60 Hz).

Figure 33 - Mains Switch, fuse holder and power

input

To replace a fuse: 1) Disconnect the mains cable 2) Pull the fuse holder out of the connector 3) Remove the damaged fuse(s) 4) Insert 2 fuses, (2 x 2 A / 230 V slow blow) 5) Replace the fuse holder 6) Plug the mains cable into a power outlet with a solid ground

connection 7) Switch the system on and operate as instructed in this manual

8.3. Connecting coupling NX with compact NX generator When setting up the test national and international regulations regarding human safety have to be guaranteed.

Figure 34 - Adding coupling

All units, the surge generator, the EFT and the coupling matrix can be installed one above the other. The coupling matrix should be used as central output for the EUT and should be mounted directly onto the ground reference plane. The compact NX5 has to be mounted directly above the coupling NX5 due to the short HV cable. It is recommended to connect the simulator to the ground reference plane of the test set-up.

Figure 6.1: Example of a test rack with compact NX5 and coupling NX5 bs-3-480-16

Connect the following cables between the generator and coupling network

Name Length [m] Connection Remark 1 SLC 500, Sys Link cable 0.5 m, (1.0 m) compact NX5 – coupling NX 26 pole D-Sub High Density 2 BCC 400, HVS Burst cable 0.4 m compact NX5 – coupling NX Coaxial cable 3 HVS Surge cable 0.5 m compact NX5 – coupling NX 2 cable for HV and COM 4 Earth connection 0.3 m compact NX5 – coupling NX 300 mm x 23 mm; 25 mm2 5 Earth connection 0.3 m coupling NX – Ref GND 300 mm x 23 mm; 25 mm2 6 Power mains cable 2 m Mains – Supply coupling 85V to 154 V 50 / 60 Hz 7 EUT mains To define Mains 3-ph – EUT input Ext fuse required

Page 33: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 33 / 50

Figure 35 - Front side connections

Connection front side:

1. BCC 350, Burst connection cable. Coaxial cable for connect the EFT Burst pulse from the generator to the coupling.

2. Burst output to a capacitive coupling clamp CCI with the coaxial connection cable BCC 1000.

3. Output to the EUT. Burst and Surge pulses are coupled to the supply lines.

4. Surge output to external CDN for data lines via a HVS cable.

Figure 6.2: Front side compact NX5 with coupling NX5

Figure 36 - Rear side connections

Connection rear side:

1. Copper braid for earth the generator and coupling

2. HVS Surge cable

3. SCC AD, Safety circuit Adapter (Sys.link)

4. SLC 500 Sys.Link cable (Daisy chain)

5. SWL AD, Safety warning Lamp Adapter (Sys.link)

6. Mains supply (control)

7. EUT supply input

8. Earth connection to reference ground plane with low inductance connection.

Figure 6.3: Rear side compact NX5 with coupling NX5

8.4. Connecting CDN 30x3A-C32 with NSG 30x0A generator When setting up the test national and international regulations regarding human safety have to be guaranteed.

Figure 37 - Example Test

Rack

All units, the surge generator, the EFT and the coupling matrix can be installed one above the other. The coupling network should be used as central output for the EUT and should be mounted directly onto the ground reference plane. The NSG 3060A has to be mounted directly above the CDN 3063A-C32 due to the short HV cable. It is recommended to connect the simulator to the ground reference plane of the test set-up.

Figure 3.1: Example of a test rack with NSG 3060A and CDN 3063A-C32

Connect the following cables between the generator and coupling network

Name Length [m] Connection Remark 1 SLC 500, Sys Link cable 0.5 m, (1.0 m) NSG 3060A – CDN 3063A 26 pole D-Sub High Density 2 BCC 350, HVS Burst cable 0.4 m NSG 3060A – CDN 3063A Coaxial cable 3 HVS Surge cable 0.5 m NSG 3060A – CDN 3063A 2 cable for HV and COM 4 Earth connection 0.3 m NSG 3060A – CDN 3063A 300 mm x 23 mm; 25 mm2 5 Earth connection 0.3 m CDN 3063A – Ref GND 300 mm x 23 mm; 25 mm2

Page 34: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 34 / 50

6 Power mains cable 2 m Mains – Supply coupling 85V to 154 V 50 / 60 Hz 7 EUT mains To define Mains 3-ph – EUT input Ext fuse required

Figure 38 - Front Connections

Connection front side:

1. BCC 350, Burst connection cable. Coaxial cable for connect the EFT Burst pulse from the generator to the coupling.

2. Burst output to a capacitive coupling clamp CCI with the coaxial connection cable BCC 1000.

3. Output to the EUT. Burst and Surge pulses are coupled to the supply lines.

4. Surge output to external CDN for data lines via a HVS cable.

Figure 3.2: Front side NSG 3060A with CDN 3063A-C32

Figure 39 - Rear Connections

Connection rear side:

1. Copper braid for earth the generator and coupling

2. HVS Surge cable

3. SCC AD, Safety circuit Adapter (Sys.link)

4. SLC 500 Sys.Link cable (Daisy chain)

5. SWL AD, Safety warning Lamp Adapter (Sys.link)

6. Mains supply (control)

7. EUT supply input

8. Earth connection to reference ground plane with low inductance connection.

Figure 3.3: Rear side NSG 3060A with CDN 3063A-C32

Page 35: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 35 / 50

8.5. Front view (model-specific distinctions)

Figure 40 - coupling NX front side (model for 16 A / 32 A)

1 Active LED 2 Phenomenon (Burst, Ring Wave, Surge) 3 Coupling indication LED 4 Test Supply AC, DC and Over temperature 5 "TEST ON” button

6 EFT input from compact NX generator 7 EFT output to capacitive coupling clamp 8 EUT Output L1/DC+, L2, L3, N/DC-, PE 9 GND reference for EFT verification 10 Output HV & COM to external Surge CDN

8.6. Rear view (model-specific distinctions)

Figure 41 - Rear side coupling NX

1 Reference earth connection (Screw M4) 2 EUT power L1 & dc+, L2, L3, N & dc-, PE 3 Input HV & COM from NX generator 4 Ventilator

5 Power switch 6 Mains connector and fuse (5 x 20 mm) 7 Sys Link

Page 36: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 36 / 50

8.7. Operation with external CDN or generator internal CDN

The coupling network of the coupling NX is operated and controlled via the compact NX generator. After power on, the compact NX generator scans the system for searching all devices.

Power ON procedure

Result power on

OK

OK

NOGO generator cannot find the coupling NX,

switch OFF / ON compact NX generator

same time second first

same time first second

The test generator realizes that a three phase coupler is connected and shows the possible coupling modes in the display of the generator. The complete system is controlled by the generator from front panel or software. If the external coupling network is connected or switched on later than the generator, the generator must be reset by switching power ON/OFF.

On the front panel of the coupler the user can switch ON/OFF the power supply of the EUT.

Using iec.control software the coupling matrix is completely controlled by software.

Coupling device detection

After power ON the compact NX, the controller searches all external devices.

External CDN detected: Working with external coupling device

External CDN not detected: Working with internal generator CDN

Page 37: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 37 / 50

9. Quick start guide DCD sr- and st-series

9.1. Test Setup for DCD application

Figure 42 - DCD Use

DCD coupling / decoupling network Impulse generator

Earth connections

Earthcable: Low inductance and low ohmic connection ex. Use a copper tresse or a flat copperband. Connection: Connect the earth bolt of the DCD, compaxt NX generator and ground reference together Depending on the application, the sequence of the earth connection DCD - generator - reference earth can be changed as required.

WARNING

The device earth bolt must be connected to the ground reference plane, if there is one. In other case user is responsible that the generator earth bolt is proper connected with the building earth.

Connection HV – COM (impulse cable generator – DCD)

Twisted cable (HV and COM) HV-cable red: HV generator - HV DCD coupling / decoupling network HV-cable black: COM generator - COM DCD coupling / decoupling network

Dataline connection - Input from AE port on rear side - Output to EUT on frontside

Figure 43 - DCD Connection

Page 38: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 38 / 50

Figure 44 - AE Protection

9.2. AE Protection for auxiliary equipment Application

Depends on the application different protection units are awailable to protect the auxiliary equipment.

The protection unit is changeable to get the best protection for the auxiliary equipment

9.3. Device functions and operating The difference between the various models is:

- Number of data lines (DCD sr 4-x series with 4 lines, DCD sr 8-x series with 8 lines) - nominal voltage and current level of the signal- and data lines - various test voltage level of 5 kV and 7 kV

9.4. Coupling network DCD sr

Figure 45 - DCD Connections 1 Signal output to EUT 2 HV COM input (Generator) 3 Signal input (AE port) 4 Protection device 5 PE earth plug to ground 6 Coupling HV to L1 - L4 (S2)

7 Coupling L1 L4 to COM (S1) 8 Connection COM-GND (S1) 9 External coupling 10 Coupling 3 µF Ringwave 11 Coupling 0.5 µF / 40 Ω Surge 12 Coupling GDT (inside the bridge) / 40 Ω Surge

Page 39: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 39 / 50

9.5. Coupling Line to Line

9.5.1. Coupling with 40 Ω resistor and 0.5 µF capacitor

Figure 46 - Line-to-Line

Example for coupling as per IEC 61000-4-5

Coupling: Capacitive with 0.5 µF and 40 Ω resistor Coupling Path: Line L1 (S2) to Line L3 (S1)

9.5.2. Coupling with 40 Ohm and Gas Discharge Tube (GDT)

CAUTION

Use only bridges with built in GDT (gas discharge tube). Otherwise the generator circuit is loaded with the line voltage in series with 40 Ω. This can destroy the generator

Figure 47 - Example for coupling as per IEC 61000-4-5

Coupling: Gas discharge tube (GDT) and 40 Ω resistor NOTE: The GDT is inside the bridge! Coupling Path: Line L2 (S2) to Line L4 (S1)

CAUTION

Max. applied EUT voltage with GDT coupling is given by the GDT breakdown voltage of 90 V. Other line voltages may require an GDT with adapted line voltage.

Page 40: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 40 / 50

9.6. Coupling to GND

9.6.1. Test setup Coupling to PE

Figure 48 - Example for coupling as per IEC 61000-4-5

Coupling: Capacitive with 0.5 µF and 40 Ω resistor Coupling Path: Line L1 (S2) protected earth PE (S1)

9.7. Coupling network DCD st, 4 lines 4 Data line setup:

Each of the 4 lines has an impedance of= 160 Ω All four lines are connected by using a coupling device (CD)

Figure 49 - 4 Data Line Setup

2 Data line setup:

Two lines are switched in parallel. Each of the 2 lines has an impedance of= 80Ω (160 Ω // 160 Ω)

L1 = (L1’ // L3’) L2 = (L2’ // L4’)

Figure 50 - 2 Data Line Setup

For 2 Data line setup: - Line 1 and Line 2 coupeld with a coupling device (CD) to the EUT output. - Lines L1 is parallel with L3 - Lines L2 is parallel with L4

Page 41: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 41 / 50

9.8. Coupling network DCD st, 8 lines 8 Data line setup:

Each of the 8 lines has an impedance of= 320 Ω

Figure 51 - 320 Ω, 8 Lines

4 Data line setup:

Two lines are switched in parallel. Each of the 4 lines has an impedance of= 160Ω (320 Ω // 320 Ω)

L1 = (L1’ // L5’) L2 = (L2’ // L6’) L3 = (L3’ // L7’) L4 = (L4’ // L8’)

Figure 52 - 160 Ω, 4 Lines

2 Data line setup:

Four lines are switched in parallel. Each of the 2 lines has an impedance of= 80Ω (320 Ω // 320 Ω // 320 Ω // 320 Ω)

L1 = (L1’ // L3’ // L5’ // L7’) L2 = (L2’ // L4’ // L6’ // L8’)

Figure 53 - 80 Ω, Lines

9.9. Test setup for 10/700 µs impulses

Due to the nature of the wiring used for unshielded outdoor symmetrical communication lines (twisted pairs), the coupling is always in common mode. The coupling decoupling schematic is shown in Figure A.4. (IEC 61000-4-5 Ed 3.0)

Coupling via arrestors(GDT) or Avalanche breaking diodes (ABD) is the preferred coupling method for unshielded outdoor symmetrical communication lines. The coupling network also has the task of splitting the surge current into multiple pairs in multi-conductor cables. The internal generator matching resistor Rm2 (25 Ω) is replaced by the Rc = 25 Ω in the DCD coupling / decoupling network.

Figure 54 - IEC 61000-4-5 Ed3.0

Page 42: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 42 / 50

9.9.1. Test setup with 4 lines Network DCD ST 4-1

4 Dataline setup: Two symmetrical lines (2 pairs)

Each of the 4 lines has an impedance of= 25 Ω

Bridge (grey): The 25 Ω impedance results by shorting the 135 Ω resistors.

Bridge (CD): Connection for the built-in coupling device (GDT/ABD) for coupling to lines L1 to L4.

Figure 55 - Two Symmetrical Lines

2 Data line setup: One symmetrical lines (1 pair)

Bridge (grey): The 25 Ω impedance results by shorting the 135 Ω resistors of lines L1 and L2.

Bridge (CD): Connection for the built-in coupling device (GDT/ABD) for coupling to lines L1 and L2.

Figure 56 - One Symmetrical Line

Page 43: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 43 / 50

Figure 57 - CCI

10. Quick start guide capacitive coupling clamp CCI

The CCI is used to couple EFT/burst pulses to I/O lines as required in different European and international standards for immunity testing. The coupling of the Electrical Fast Transients EFT/burst to signal lines can usually not be achieved by discrete capacitive coupling without interfering with the signal flow. It is often impossible to contact the required circuit (direct), e. g. coaxial or shielded cables. In this case the coupling is realized by the capacitive coupling clamp. The interference simulator can be connected on both sides of the coupling clamp. Normally the clamp is connected directly to the coaxial output of the generator (50 Ohm output), but it also can be connected and controlled via the external coupling network.

Figure 58 - CCI Connection

To activate the coupling via the coupling clamp, the appropriate coupling must be selected on the generator.

Figure 59 - Generation Settings for CCI

Page 44: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 44 / 50

11. Quick start guide HSC 4-8, CNI 508N2, SPN 508N1

11.1. HSC 4-8 Coupling / decoupling network The HSC 4-8 is a specifically designed CDN for testing high-speed communication lines. Its unique design allows data rates of up to 1,000MBit/s to be transmitted without degradation by the coupling/decoupling network. The decoupling network protects the auxiliary equipment limiting the residual voltage to max. 50V@2kV for common-mode surge test pulses.

Moreover, the HSC 4-8 fully supports PoE and PoE+ standard requirements as per IEEE 802.3.

The HSC 4-8 may also be used for testing shielded cables as per IEC/EN 61000-4-5, Fig. 16.

Figure 60 - HSC 4-8

It simplifies the test set-up and offers proper decoupling of the auxiliary equipment as well as a high safety for the operator as the surge is coupled through the HSC 4-8 rather than being directly applied to the metallic case of the DUT. The HSC 4-8 may also be used to couple Ring Wave impulses (0.5 us/100 kHz) and burst pulses to shielded cables .

11.2. CNI 508 N2 Coupling / decoupling network The CNI 508N2 is the predecessor of the HSC 4-8 and can also be used for testing high-speed communication lines. Its unique design allows data rates of up to 1,000MBit/s to be transmitted without degradation by the coupling/decoupling network. However, unlike the HSC 4-8, it cannot be used for ringwave tests.

Figure 61 - CNI 508N2

11.3. SPN 508 N1 Surge Protection Network The SPN 8 protects the auxiliary equipment with a residual voltage of less than 10V. This Ethernet surge protection device can alternatively use for all other application where the user must protect his devices from any overvoltage.

The SPN 8 has a galvanic isolation of the signal between protected and unprotected side. For decoupling the pulse in direction to the AE port, the surge protection network SPN 508 N1 must mandatory be connected at the opposite side of the EUT at HSC 4-8 or CNI 508N2.

Figure 62 - SPN 508 N1

Page 45: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 45 / 50

11.4. Front and rear view HSC 4-8

1 Ring Wave / SDOW input from Generator 2 Burst input from Burst Generator 3 Switch Ring Wave / Surge 4 Handle open position 5 Output to EUT Port

6 Surge input from Surge Generator to shield 7 Surge input from Surge Generator to lines 8 Handle closed position 9 Output to RJ45 AE Port 10 Earth plugs

11.5. Front and rear view CNI 508N2

1 Burst input from Burst Generator 2 Output to EUT Port 3 Handle open position 4 Surge input from Surge Generator to shield

5 Surge input from Surge Generator to lines 6 Handle closed position 7 Output to RJ45 AE Port 8 Earth plugs

11.6. Front and rear view SPN 508N1

Page 46: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 46 / 50

1 Insulating plate 2 Indication Protected side 3 Output Port AE

4 Earth plugs 5 Indication Unprotected side 6 Output Port EUT via coupling/decoupling network

11.7. General Coupling of the surge pulses is achieved by means of the coupling / decoupling network HSC 4-8. For additional protection of the auxiliary equipment by reducing the residual voltage level the SPN 508N1 is used.

The HSC 4-8 design is based on the IEC 61000-4-5 (Ed 3.0: 2012) figure 11, IEC 61000-4-12 (Ed. 3) figure 10.

The HSC 4-8 allows the following couplings with separate input connectors: - Surge and Ring wave pulses to data-lines - Surge pulses to shield - Burst pulses to shield

For an additional overvoltage protection of the AE port the SPN 508N2 can be used for limiting the residual voltage to approx. 10V.

Figure 63 - Schematic of the HSC 4-8 for Surge to high speed

data-lines

Figure 64 - Schematic of the HSC 4-8 for Ring Wave to high speed data-lines

11.8. Test methods for Surge and Ring Wave to high speed datalines The IEC 61000-4-5 Ed.3 Surge standard shows in figure 11 the coupling network for testing high speed datalines. The solution for coupling/decoupling is based on the existing methods of surge testing. There are two different Methods for testing surge pulses on datalines.

Coupling to shielded lines

Page 47: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 47 / 50

Test to shielded line is adapted from the figure 16 where the surge is applied direct to the shield of the data cable.

Figure 65 - Coupling to Shielded Lines

Coupling direct to high speed lines

Test with coupling direct to the data-lines is based to figure 11 of IEC 61000-4-5 standard. The surge is applied direct to unshielded symmetrical high-speed data-lines. For decoupling and protection, it is necessary to use separate devices of the HSC 4-8 assembly.

Figure 66 - IEC 61000-4-5

11.9. Test setup The SPN 508N1 protects the auxiliary equipment with a residual voltage of less than 10V. Generator output plugs

The Surge and Ringwave output plugs are located at the rear side of the impulse generator. The surge generator impedance is 2 Ω. The additional impedance is inside the HSC 4-8.

Surge, Ringwave: Output HV –COM on rearside of the generator with direct coupling.

Figure 67 - Surge output on rearside

Figure 68 - Burst pulse output

Grounding: The generator and the used devices of the HSC 4-8 assembly must be grounded properly to the reference ground.

11.10. Surge test to shielded datalines This test will couple the surge pulse to the shield of a shielded RJ24 data-cable. The cable to the EUT must be mandatory a shielded cable.

Page 48: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 48 / 50

Figure 69 - Setup for surge test to shielded datalines

The surge current on the cable shield can be very high and is limited by the 2Ω generator impedance. To use the SPN 508N1 protection device will protect the AE port side and all connected devices.

The surge coupling voltage and current to the lines depends on the impedance connected at the HSC 4-8 AE port.

In case of a short circuit as shown in figure 5.4, the coupling to the lines is nearly 1:1 and a current wave 8/20us appears.

Is the at the AE Port a high impedance (open circuit), an impulse waveshape 1.2/50us appears at the EUT Port

Figure 70 - HSC 4-8 with Short circuit at the AE port

Page 49: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 49 / 50

11.11. Surge test to datalines This test will couple the surge pulse direct to the 8 datalines of the RJ45 cable. The cable to the EUT can be a shielded or unshielded data-cable. Figure 5.5 shows the test setup for surge to datalines.

Figure 71 - Surge Test to Datalines

11.12. Surge test to data-lines with SPN 508N1 protection In case of a high protection level of approx. 10V at the AE port, the user can connect additional the SPN 508N1. This protection will separate the AE side the surge by a galvanic insulation from the surge circuit. Figure 5.6 illustrates the setup with using a SPN 508N1 protection device.

Figure 72 - SPN 508N1 Use

11.13. Burst test to data-lines The burst pulse is applied from the 50Ω front side EFT output to the blue HV-BNC plug of the CNI 508N2. The coupling is to the shield. The burst pulse can be measured on the shield as on the data-lines.

Figure 73 - Burst on Data Lines

Page 50: QUICK-START GUIDE Transient Immunity AND SAFETY

AMETEK CTS Quick start guide - Conducted

1.00 50 / 50

12. Maintenance, Adjustments, Replacement of Parts

CAUTION

Electrical maintenance must only be performed by qualified service technicians

The generators do not contain any parts or components requiring special maintenance.

Electrical maintenance must only be performed by experienced and specially trained technicians. Generally, standard maintenance requires only the periodic cleaning of the instrument, verification and calibration of certain parameters.

• When removing the cover or other parts of the equipment, high voltage parts may become exposed. High voltages are potentially lethal.

• For service, repair, adjustment or replacement of parts, the generator must be disconnected from all power supply sources before covers are to be removed.

• The user is not permitted to change or modify any EM TEST / TESEQ generator. Only original EM TEST / TESEQ parts and components shall be used for repair and service. AMETEK CTS is not responsible for accidents or injuries caused through the use of parts or components not sold by AMETEK CTS.

• Maintenance and service must only be performed by qualified service technicians who are trained and familiar with the dangers of servicing the EM TEST / TESEQ generator.

• Only fuses of correct voltage and amperage as specified by the manufacturer are to be used for replacement. The repair of fuses is not permitted.

12.1. Disposal of devices

For the disposal of electronic devices, the country-specific regulations are to be considered. The equipment should be delivered to a specialized waste collection center.

EM TEST and Teseq devices can be returned to AMETEK CTS in Switzerland or to their agency for adequate disposal. Alternatively, the equipment can be handed over to a specialized enterprise for disposal of electronic devices.

Details to used material and components

- Built in capacitors contain no polychlorinated biphenyl (PCB). - Back-up batteries and rechargeable batteries must be disposed separately.

12.2. Calibration and Verification For the calibration and verification of the voltage impulses the using of calibration adapters is recommended. The calibration is at all times without any mains supply on the output to calibrate. Therefore, the following warnings are to adhere strictly.

CAUTION

Before starting the calibration or verification remove the EUT Mains Supply from the generator and from the coupling network