Quality assurance of a characterised Ge-detector used for various geometries (mathematical efficiency calibration) Asser Nyander Poulsen Specialkonsulent, Ph.D. [email protected] Strålebeskyttelse i Sundhedsstyrelsen (SIS) Knapholm 7 2730 Herlev, Denmark www.sis.dk

Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Quality assurance of a characterised Ge-detector used for various geometries (mathematical efficiency calibration) Asser Nyander Poulsen Specialkonsulent, Ph.D. [email protected] Strålebeskyttelse i Sundhedsstyrelsen (SIS) Knapholm 7 2730 Herlev, Denmark www.sis.dk

Page 2: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Setup: Characterized Ge detector Electrical cooling Shielded lab + 10 cm Pb. No background peaks except 511 keV

Page 3: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm
Page 4: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

QC Objective

Must keep track of: “Any parameter or setting that can be edited by the user that can affect the measurement result.”

Page 5: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Transfer

Manuel selection

Manual entry Non-editable software

Editable spreadsheet

.ECC

File

.CNF

.NLB

.CAL

.GEO

Genie2K Geometry Composer

.RPT

.PDF

.TPL

.CTF

.TXT

Date, time

File name Sample ID Ref date Vol, mass Uncertainty

Live time

Energy & uncertainty list

HV Gain Stabilizer Filter

Library Editor

Nuclides, t½, keV yield, uncertainty

Library

Nuclides t½

keV yield

uncertainty

Certificate Editor

PC clock

Cascade summing

correction

File name Dimensions Materials Densities

Detector SN Template Beaker

.BKR .PAR

Detector characteri- zation

Dimensions Materials

Beaker Editor

QA Log

Report Viewer

File name

Algorithm settings

Calibration

Report template

Validation parameters

Final report

Sample Container

.PDF

Data report

Test results

Default energies and uncertainties

coi-lib.clb

Page 6: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Hardware and algorithm settings QC

CAM-file parameter checksum

Genie2000 .rpt - file Excel

Page 7: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm
Page 8: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Manually entered parameters

Page 9: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

LabSocs. Sensitivity Analysis

Cs-137 point source, 21 mm from detector 661 keV, net count-rate 762 cps (+/- 0,7 %, k=1)

Hardware settings

Page 10: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

LabSocs. Sensitivity Analysis LabSocs geometry parameters

- Distance is a critical parameter

Page 11: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Validation Do I have a problem ?

Page 12: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Detector reference plane lowered (2mm)

Page 13: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Electric cooler issue “Sudden loss of cooling efficiency after 1 year”

Pressure relief (3-5 h) and cooler restart

Page 14: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Extra material

Page 15: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Can You

prove it? It is so…

Page 16: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Human factor

Page 17: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Geometry

PLEX: Holder ID, absorber

GEC-987: Detector and measurement ID

WHI: Sample container ID

Unique name: GEC-987_PLEXWHI

Name is used for .GEO, .CAL,

Page 18: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Loose end #1 - Genie settings

Page 19: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Genie report template (.tpl)

“miss”, * and @ are moved to the end

Section tag

Page 20: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Loose end # 2 - Nuclide libraries

Data transfer Best data source?

Page 21: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Nuclides listed with gaps! Excel challenge

Page 22: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Drift

Energy sens.

Resolution

Efficiency

Page 23: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Eu-152 LACE

Page 24: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

QC parameter Energy tolerance (Eu-152)

Maximal deviation (keV)

Page 25: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Screenshots for documentation

Density calculated from sample mass and fill height

Page 26: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

1. Assumed sample composition and weight entered

2. Density roughly adjusted to reach approx. fill-height

Page 27: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Gammaspec 2016 sample

9,1698 (10) g

Fill-height ≈ 5 mm

Spectrum data file: GEC-987-K1-….cnf

Detector ID – Measurement ID – Acquisition cycle -

Page 28: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Geometry model (LabSocs) Standard container ”WHITE.bkr”

Page 29: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

Test measurements

“Point” Cs-137, Co-60, Am-241 (Ra-226, Pb-210)

“Beaker” Eu-152

“Custom Beaker” K-40 (KCl)

Page 30: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

𝑧𝑒𝑡𝑎 =𝑎𝑏𝑠(𝐴𝑐𝑒𝑟𝑡 − 𝐴𝑜𝑏𝑠)

𝑢(𝐴𝑐𝑒𝑟𝑡)2 + 𝑢(𝐴𝑜𝑏𝑠)

2

QC parameter

Page 31: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

In case of non-compliance:

Increase efficiency uncertainty.

E.g. changed from factory default (4%)

Page 32: Quality assurance of a characterised Ge-detector used for ... · 2. Density roughly adjusted to reach approx. fill-height . Gammaspec 2016 sample 9,1698 (10) g Fill-height ≈ 5 mm

A (w.m.) > MDA ?

Reporting rule

yes

no “ A < [MDA]”

“ A (best est.) +/- u (best est.)”

To avoid false positive results, we have chosen not to report A when MSA < A < MDA. In some cases, w.m. < 0 and best. est. >0. w.m. = weighted mean best. est. = best estimate (ISO11929)