100
   Q   s   e   r   v    i   c   e              T    h    i   s    D   o   c   u   m   e   n    t    i   s   a   c   o   m   p    l   e    t   e   s   c   a   n    f   r   o   m     t    h   e    O   r    i   g    i   n   a    l     T   e    k    t   r   o   n    i   x    M   a   n   u   a    l               Q   s   e   r   v    i   c   e

Qservice ----- This Document is a Complete

  • Upload
    stefano

  • View
    217

  • Download
    0

Embed Size (px)

Citation preview

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 1/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T

  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 2/100

TEK070-6717-00

PRODUCT GROUP 46

First Printing October 1987

COMMITTED TO EXCELLENCE

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 3/100

Copyright O 1987 Tektronix, Inc. All rights reserved. Contents

of this publication may not be reproduced in any form without

the written permission of Tektronix, Inc.

Products of Tektronix, Inc. and its subsidiaries are covered by

U.S. and foreign patents issued and pending.

Information in this publication supersedes that in all previously

published material. Specification and price change privileges re-

served. TEKTRONIX, TEK, PLOT 10, TEKTEST, SCOPE-MOBILE,

and are registered trademarks. For further information,

contact Tektronix, Inc. Corporate Offices, P.O. Box 500, Beaver-

ton, OR 97077. Phone: (503) 627-71 11; TWX: (910)467-8708;

TLX: 151754. Subsidiaries and distributors worldwide.

Printed in Hong Kong.

INSTR UM ENT SERl JMBERS

Each instrument has a serial number on a panel insert, tag, or

stamped on the chassis. The first number or letter designates

the country of manufacture. The last five digits of the serial

number are assigned sequentially and are unique to each instru-

ment. Those manufactured in the United States have six uniquedigits.

BOO0000 Tektronix, Inc., Beaverton, Oregon, U .S.A.

HK00001 Hong Kong

100000 Tektronix Guernsey, Ltd., Channel Islands

200000 Tektronix United Kingdom, Ltd., London

300000 Sony/Tektronix, Japan

700000 Tektronix Holland, NV, Heerenveen,

The Netherlands

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 4/100

NOTICE to the userloperator:

The German Postal Service requires that Systems assembled by the

operatorluser of this instrument must also comply with Postal

Regulation, Vfg. 104611984, Par. 2, Sect. 1.

HlNWElS fiir den BenutzerIBetreiber:

Die vom Betreiber zusammengestellte Anlage, innerhalb derer dies

Gerat eingesetzt wird, mu0 ebenfalls den Voraussetzungen nach Par. 2,

Ziff. 1 der Vfg. 104611984 genugen.

NOTICE to the userloperator:

The German Postal Service requires that this equipment, when used ina

test setup, may only be operated if the requirements of Postal

Regulation, Vfg. 104611 984, Par. 2, Sect. 1.7.1 are complied with.

HlNWElS fur den BenutzerIBetreiber:

Dies Gerat darf in MeAaufbauten nur betrieben werden, wenn die

Voraussetzungen des Par. 2, Ziff. 1.7.1 der Vfg. 104611984 eingehalten

werden.

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 5/100

Certificate of the ManufacturerIImporter

We hereby certify tha t the2205 OSCILLOSCOPE

AND ALL INSTALLED OPTIONS

complies with the RF Interference Suppression requirements ofAmtsbl.-Vfg 104611984.

The German Postal Service was notified that the equipment is beingmarketed.

The German Postal Service has the right to re-test the series and toverify that it complies.

TEKTRONIX

Bescheinigung des Herstellersllmporteurs

Hiermit wird bescheinigt, daA derldie ldas2205 OSCILLOSCOPE

AND ALL INSTALLED OPTIONS

i n ~be re ins t immun~it den Bestimmungen der Amtsblatt-Verfugung104611984 funkentstort ist.

Der Deutschen Bundespost wurde das lnverkehrbringen dieses Geratesangezeigt und die Berechtigung zur uberprufung der Serie auf Einhaltender Bestimmungen eingeraumt.

TEKTRONIX

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 6/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 7/100

CONTENTS

Page

. . . . . . . . . . . . . . . . . . . . . . . . . . .perator's Safety Summary iv

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .he 2205 Oscilloscope vii

SECTION 1 PREPARATION FOR USE

Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ine Voltage Selection 1-1

Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -4

Power Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .nstrument Cooling 1-4

Initial Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

SECTION 2-OPERATION

. . . . . . . . . . . . . . . . . . . . . . . . . . . .ront Panel Organization 2-1

. . . . . . . . . . . . . . . . .ontrols. Connectors. and Indicators 2-1. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .earning the Controls 2-6

. . . . . . . . . . . . . . . . . . .isplay Controls .... . .... . . . . 2-8

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ertical Controls 2-8

Input Coupling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .hannel Selection 2-9

Add and Invert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .hop or Alt? 2-9

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .orizontal Controls 2-10

. . . . . . . . . . . . . . . . . . . . . . . . . .weep Speed Selection 2-10. . . . . . . . . . . . . . . . . . . . .agnifying Waveform Details 2-10

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 8/100

SECTION 2-OPERATION (CONTINUED)

Trigger Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1 0

Which Mode to Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Slope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Connecting Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Waveform Fidelity and Probe Grounds . . . . . . . . . . . . . .Probe Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . .Probe Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .Coaxial Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

SECTION 3-APPLICATIONS

Amplitude Measurements . . . . . . . . . . . . . . . . . . . . . . . ... 3-1

Peak-to-peak Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1

Instantaneous Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3

Algebraic Addition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5

Common-mode Rejection . . . . . . . . . . . . . . . . . . . . . . . . . 3-6

Amplitude Comparison (Ratio) . . . . . . . . . . . . . . . . . . ... 3-7

Time Measurements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-8Time Duration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8

Period and Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1 0

Rise Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1 0

Time Difference Between Pulses on

Time-Related Signals . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 2

Phase Difference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 4

Television Displays . . . . . . . . . . . . . . . . . . . . . . . . . ..... 3-1 7

TV Field Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1 7

TV Line Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1 8

SECTION 4-CHECKS AND ADJUSTMENTS

Trace Rotation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -1

Probe Compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 9/100

SEC TION 5-PERFORMANC E CHARACTE RISTICS

SEC TION 6-OPTIONS AND ACCESSO RIES

Standard Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1

Optional Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4

APPENDIX A - Magnif ied Sweep Speeds

APPENDIX B - Repackaging for Shipment

APPENDIX C - Glossary

2205 Operator's iii

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 10/100

Operator's Safety Summary

OPERATOR'S SAFETY SUMMARY

The general safety information in this summary is for both oper-ating and servicing personnel. Specific warnings and cautions -found throughout the manual where they apply - do not appear

in this summary.

Terms

In This Manual

CAUTION statements identify conditions or practices that

could result in damage to the equipment or the property.

WARNING statements identify conditions or practices that

could result in personal injury or loss of life.

As Marked on Equipment

CAUTION indicates a personal-injury hazard that is not im-

mediately accessible as one reads the markings, or a

hazard to property, including the equipment itself.

DANGER indicates a personal-injury hazard that is

immediately accessible as one reads the marking.

Symbols

In This Manual

This symbol indicates applicable cautionary or otherinformation. For maximum input voltage see

Table 5-1.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 11/100

Ooerator 's Safetv Summarv

As Marked on Equipment

DANGER-High voltage.

@ Protective ground (earth) terminal.

ATTENTION-Refer to manual.

Power Source

This product is intended to operate from a power source that

does not apply more than 250 V rms between the supply con-ductors or between either supply conductor and ground. A pro-tective ground connection, by way of the grounding conductor in

the power cord, is essential for safe operation.

Grounding the Product

This product is grounded through the grounding conductor of thepower cord. To avoid electrical shock, plug the power cord into

a properly wired receptacle before making any connections to

the product input or output terminals. A protective ground con-nection, by way of the grounding conductor in the power cord, is

essential for safe operation.

Danger Arising From Loss of Ground

Upon loss of the protective-ground connection, all accessible

conductive parts, including knobs and controls that may appearto be insulating, can render an electric shock.

Use the Proper Power Cord

Use only the power cord and connector specified for your

product.

The power cord must be in good condition.

2205 Operator 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 12/100

Operator's Safety Summary

Read Section 1 for power-cord and connector information.

Use the Proper Fuse

To avoid fire hazard, use only a fuse of the correct type, volt-age rating and current rating as specified on the back of your

product and in Table 6-1.

Do Not Operate in an Explosive A tmosp here

To avoid explosion, do not operate this product in an explosive

atmosphere.

Do Not Remove Covers or Panels

To avoid personal injury, do not remove the product covers orpanels. Do not operate the product without the covers andpanels properly installed.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 13/100

Operator's Safety Summary

THE 2205 OSCILLOSCOPE

The TEKTRONIX 2205 Oscilloscope is a rugged, lightweight,

dual-channel, 20 MHz nstrument that features a bright, sharply

defined trace on an 80 by 100 mm cathode-ray tube (cr t) .

Its low-noise vertical system supplies calibrated deflection

factors from 5 mV to 5 V per division at full bandwidth.

Stable triggering is achieved over the full bandwidth of the verti-

cal system. The flexibility and high sensitivity of the trigger sys-

tem provides a range of conveniences such as hands-free trig-

gering with the peak-to-peak automatic mode, normal triggermode, independent selection of TV line and TV field triggering atany sweep speed, and single-sweep triggering. The trigger sig-nal is dc coupled. An external triggering signal or an externalZ-axis modulation signal can be applied via a front-panel

connector and the source-selector switches.

The horizontal system provides calibrated sweep speeds from0.5 s to 100 ns per division. For greater measurement accu-racy, a XI0 magnifier circuit extends the maximum sweep speed

to 10 ns per division.

Accessories

The instrument is shipped with the following accessories: opera-

tor's manual, two signal adapters, a power cord, and a power-

cord clamp. Part numbers for these standard accessories, as

well as for other optional accessories, are located in Section 6,Options and Accessories. The voltage-sensing signal adapterswere designed specifically to complement the performance ofyour 2205.

2205 Operator's vii

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 14/100

For More Information

Should you need additional information about your 2205 Oscillo-

scope or about other Tektronix products; contact the nearestTektronix Sales Office or Distributor, consult the Tektronix prod-

uct catalog, or, in the U S , , all the Tektronix National MarketingCenter, toll free at 1-800-426-2200.

The 2205 Oscilloscope.

viii 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 15/100

SECTION 1 IREPARATION

FOR USE

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 16/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 17/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l

  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 18/100

Preparation for Use

Table 1-1

Power Plugs and Line Voltage Selection

LineVoltageSelector

1 5 V

PlugConfiguration

Powerlption

Std

ReferenceStandardsb

ANSl C73.11NEMA 5-15-PIEC 83UL 198.6

Plug Type

UnitedStates120 v

Europe220 v CEE (7), l, IV, VI I

IEC83IEC 127

UnitedKingdom240 V

BS 1363IEC 83IEC 127

Australia240 V

AS C112IEC 127

NorthAmerica240 V

ANSl C73.20NEMA 6-15-PIEC 83UL 198.6

Switzerland220 v

SEVIEC 127

a A 5 A, Type C fuse is installed inside the p lug of the Option A2 power co rd

b Reference Standards Abbreviations:ANSl --A merica n National Stand ards InstituteAS -Stan dards Asso ciation of AustraliaBS -British Stan dard s InstitutionCE E -International Comm ission on Rules for the Approval of Electric- quipmentIEC -International Electrote chnica l Comm issionNEM A-National Electrical Manu facturer's Asso ciationSEV -Schw eizerische r Elektrotechnischer VereinUL -Underwriters Laboratories Inc.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 19/100

Preparation for Use

PROTECTION REPLACE

O N L Y W IT H S P E C IF IE D

CAUTIONT Y P E A N D R A T E D F U S E

TO AVOID ELECTRIC

S H O C K , T H E P O W E R

CORD PROTECTION

G R O U N D IN G C O N D U C T O R

MU S T B E C O N N E C T E D

T O G R O U N D .

P O W E R

M A X W A T TS

M A X V A

D O N O T R E M O V E

COVERS. REFERS E R V IC IN G T O

[IUALIFIEO PERSONNEL.

LINE VOLTAGE

SELECTORHTEKTRONIX I N C . , BEAVERTON, OREGON, U . S . A

M A D E I N H O N G K O N G

Figure f -1. Voltage selector switch, fuse,and power-cord receptacle.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 20/100

Preparation for Use

LINE FUSE

The fuse holder is located on the rear panel and contains the line

(mains) fuse. Use the following procedure to verify that theproper fuse is installed or to install a replacement fuse.

1. Unplug the power cord from the power source (if applica-

ble).

2 . Use a screwdriver to press in and slightly rotate the fuse-

holder cap counterclockwise to release it.

3. Pull the cap (with the fuse) out of the fuse holder.

4.Verify that the fuse is the same type listed on the back of theinstrument. The two type of fuses listed are not directly in-

terchangeable; they require different types of fuse caps.

5. Reinstall the fuse (or replacement fuse) in the fuse-holder

cap.

6. Replace the fuse holder and cap.

POWER CORD

This instrument has a detachable, three-wire power cord with athree-contact plug for connection to both the power source andprotective ground. The ground contact on the plug connects

through the power-cord to the external metal parts of the instru-

ment. For electrical shock protection, insert this plug only into a

power source outlet that has a properly grounded protective-

ground contact.

Instruments are shipped with the required power cord as or-

dered by the customer. Power cord plug information is pre-

sented in Table 1-1, and part numbers are listed in Table 6-1.

INSTRUMENT COOLING

Maintain adequate airflow to prevent instrument damage from

internally generated heat. Before turning on the power, check

that the spaces around the air-intake holes on the sides of the

cabinet are free of any obstruction to airflow.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 21/100

Preparation for Use

TlAL START-UP

to now, you should have made the following preparation:

Read the safety information.

Verified that the LINE VOLTAGE SELECTOR is set for thesource voltage to be used.

3. Verified the fuse.

4 . Attached the power cord.

5. Ensured adequate ventilation around the instrument.

6. Plugged the power cord into the appropriate power sourceoutlet.

Now turn on your oscilloscope by pressing in the POWER button.Observe that the Power-On indicator, located below the button,illuminates.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 22/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 23/100

SECTION 2

OPERATION

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 24/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 25/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 26/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 27/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l

  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 28/100

Table 2-1Summary of Controls, Connectors, and Indicators

Tit le Function Recommended Use

INTENSITY Adjusts tracebrightness.

Compensate for ambientlighting, trace speed,trigger frequency.

BEAM FIND Compresses displayto within CRT limits

Locate off-screenphenomena.

FOCUS Adjusts for finesttrace thickness.

Optimize displaydefinition.

TRACEROTATION

Adjusts traceparallel to center-line.

Compensate for earth'smagnetic field.

POWER Turns power onand off.

Control power to theinstrument.

PowerIndicator

Illuminates whenpower is turnedon.

Determine powercondition.

VERTICALPOSITION

Moves trace upor down on screen.

Position trace verticallyand compensate for dccomponent of signal.

View either channelindependently or bothchannels simultaneously.

Selects signalinputs for display.

NORM-INVERT Inverts the Channel2 signal display. Provide for differential

I H 1-CH 2) or summecCH 1 CH 2) signals

when ADD is selected.

ADD-ALT-CHOP

ADD shows alge-braic sum of CH 1and CH 2 signals.ALT displays eachchannel alternately.CHOP switches be-tween CH 1 andCH 2 signals duringthe sweep at500 kHz rate.

Display summed or indi-vidual signals.

Selects verticalsensitivity.

Adjust vertical signalto suitable size.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 29/100

Table 2-1 (cont'd)

No.

13

14

15

16

17

18

19

Title

AC-GND-DC

HORIZONTALPOSITION

MAG(X I -X1 0)

Funct ion

Provides continu-ously variable de-flection factorsbetween calibrated

ositions of thebOLTSIDIV switch.Reduces gain by a tleast 2 . 5 : l .

In AC, blocks dccomponent ofsignal. In GND,gives referencepoint and allows

Irecharging ofnput-couplingcapacitor. In DC,couples all com-ponents of signal.

Provides for inputsignal connections.CH 1 gives hori-zontal deflectionwhen SECIDIV isin X-Y; CH 2 pro-vides vertical de-flection.

POSITION movestraces horizontally.

Selects degree ofhorizontal magni-fication.

Selects time-basespeed.

Provides continu-ously variable un-calibrated sweep

speeds to at least2 .5 times the cali-brated setting.

Recommended Use

Match signals forcommon-modereadings. Adjust heightof pulse for rise-timecalculations.

Select method ofcoupling input signals tothe vertical deflectionsystem.

Apply signals to thevertical deflectionsystem.

Control trace positioningin horizontal direction.

Examine smallphenomena in detail.Extends sweep speed to10 nsldiv.

Set horizontal speedmost suited to require-ments.

Extend the slowestspeed to at least1 .25 sldiv.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 30/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 31/100

Operation

Table 2-1 (cont'd)

No.

20

Title

PROBEADJUST

(GroundReceptacle)

SLOPE

LEVEL

TRIG'DREADY

MODE

Function

Provides approxi-mately 0.5-V ,1-kHz squarewave.

Provides safetyground and directconnection tosignal source.

Selects the slopeof the signal thattriggers the sweep

Select trigger-signal amplitudepoint.

Indicator lightswhen sweep istri gered in P-P~ 8 ~ 0 .ORM, orTV FIELD. inSingle Sweep

mode, indicatestriaaer Is armed.

P-P AUTO/TVLINE triggers fromwaveforms andtelevision lineshaving repetitionrates of at least20 Hz. NORMtriggers from ade-quate signal, withno trace in ab-sence of tri ersignal. TV+!iLDtr l ers from TV

fieW ignals; trig-get- polarity mustbe observed. SGLSWP triggerssweep only oncewhen- armed byRESET button:used for displayingor photographingnonrepetltive orunstable signals.

Recommended Use

Allows user to adjustcompensation of 1OXE obe. This source may

e used to check thebasic functioning ofvertical and horizontalcircuits but is not in-tended to check theiraccuracy.

Chassis groundconnection.

Provide ability to triggerfrom positive-going ornegative-going signals.

Select actual point oftrigger.

Indicate trigger state.

Select trigger mode.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 32/100

Operation

Table 2-1 (cont'd)

Title

RESET

SOURCE

EXT INPUT

Function

Arms tri ger cir-wit for 8GL SWP.

CH 1 , CH 2, LINE,and EXT triggersignals are selec-ted directly. InVERT MODE, trig-ger source is de-termined by theVERTICAL MODEselectors asfollows :

CH 1: trigger

comes from Chan-nel 1 signal.CH 2: triggercomes from Chan-nel 2 signal.BOTH-ADD andBOTH-CHOP:trigger is algebraicsum of Channel 1and Channel 2 sia-rials.BOTH-ALT : tria-ger comes f r o 6Channel 1 andChannel 2 onalternate sweeps.

Connectlon forapplying externalsignal for use asa trigger.

Recommended Use

Select source ofsignal that is coupledto the trigger circuit.

Trigger from a sourceother than verticalsignal. Also used forsingle-shot application

Connection forapplying externalsignal for intensitymodulation. Tointensity modulatethe dis lay, set theright TRIGGERSOURCE controlto EXT=Z.

Provide intensitymodulation fromindependent source.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 33/100

LEARNING THE CONTROLS

If you have not read Section 1 yet, you should do so now. Then,after turning the power on, let the oscilloscope warm up for a

few minutes before starting this procedure.

1. Set instrument controls as follows:

Display

INTENSITY

FOCUS

Vertical (both channels)I POSITIONMODE

VOLTSIDIVVOLTSIDIV Varible

Input Coupling

Horizontal

POSITIONMODE

SECIDIVSECIDIV Variable

Trigger

SLOPELEVEL

MODE

SOURCE

Midrange

Set for clearly defined trace

MidrangeCH 1. I v (1X)CAL detent (fully clockwise)

AC

MidrangeX1

0.2 msCAL detent (fully clockwise)

f

Midrange

P-P AUTO

CH 1

2. Connect a signal adapter to the Channel 1 input BNC con-nector (labeled CH 1 or X). Attach the ground lead to the

collar of the EXT INPUT connector and apply the signal

adapter tip to the PROBE ADJUST terminal. If necessary,

adjust the TRIGGER LEVEL control to get a stable display.

3. Change the Channel 1 input coupling to GND and use the

Channel 1 POSITION control to align the baseline trace to the

2-6 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 34/100

center horizontal graticule line. This sets the zero reference

for the display.

4. Return the Channel 1 input coupling to AC. Notice that the

square wave is centered vertically on the screen. Now setthe input coupling to DC and observe what happens to the

waveform. The zero reference is maintained at the center

horizontal graticule line.

NOTE

More information about using the controls is con-

tained at the end of this procedure. Refer to itwhen needed while learning the front-panel con-

t rols.

5. Use the following controls and notice the effect each has on

the displayed waveform as the settings are changed.

Each POSITION control

CH 1 VOLTSIDIV

CH 1 VOLTSIDIV Variable (CAL)

SEC/DIV Variable (CAL)

HORIZONTAL MODE

HORIZONTAL MAG

TRIGGER SLOPE

6. At this point, connect the second signal adapter to the CH2

or Y input connector. Set the VERTICAL MODE selector to

CH 2 and TRIGGER SOURCE to CH 2, then follow steps 2

through 5 again, using the Channel 2 controls.

7 . Now set the VERTICAL MODE selectors to BOTH-NORM-ALT

and return both VOLTSIDIV controls to .2 V (1X). Rotate all

variable controls clockwise to their CAL detents. Set theTRIGGER SOURCE switch to CH 1. Then use the VERTICAL

POSITION controls to position the two traces to convenient

locations on the screen.

8. While watching the Channel 2 waveforms, set the middle

VERTICAL MODE control to CH 2 INVERT and notice the ef-

fect. Then set the right MODE to ADD. What happens to the

control waveforms? Finally, return the middle MODE control

to NORM. What waveform is displayed now?

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 35/100

Operation

Congratulations! You now know how to use the 2205 front-panel

controls to display signals and move them about on the screen.

The remainder of this section gives you more information about

the controls and offers suggestions for their use. Section 3 ex-plains how to make specific types of measurements and how to

use the other controls not covered in the preceding exercise.

DISPLAY CONTROLS

Set the INTENSITY control for comfortable viewing, but not

brighter than you need. Use high-intensity settings to observe

low-repetition-rate signals, narrow pulses in long time intervals,

or occasional variations in fast signals.

VERTICAL CONTROLS

When making voltage measurements, rotate the VOLTSfDIV CALcontrol fully clockwise (in detent). Best accuracy is achieved by

setting the VOLTSIDIV control for the largest display possible.

When a cable or a signal adapter is used to connect a signal, thenumber in the 1X front-panel area indicates sensitivity. When

using a 10X probe, the number in the 10X PROBE front-panel

area indicates sensitivity. For example, the setting of 1 Vldivwith a cable or signal adapter would be 10 Vldiv with a 10X

probe. And, although the knob skirt is not marked, 5 Vldiv at 1X

would be 50 Vldiv if a 10X probe were used.

Input Coupling

For most applications, use DC input coupling. This mode is

compatible with the standard accessory signal adapters and itdisplays logic levels and DC levels of static signals.

Use GND input coupling to show where the zero-volt level will be

located when you shift to DC or AC coupling.

Use AC coupling for the special cases where you need to see

small signals on large DC voltage levels.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 36/100

Operation

Channel Selection

With the three VERTICAL MODE selectors, you can display com-

binations of the two vertical channels. When Channel 1 is se-

lected, the other two MODE switches are not active. When CH 2is selected, the middle MODE switch (NORMICH 2 INVERT) be-

comes active. And when BOTH channels are selected for dis-

play, all three MODE selectors are active.

ADD and INVERT

Select ADD mode to display the algebraic sum of the CHI and

CH 2 signals. When you use ADD, the CH 1 and CH 2 VOLTS/DIV settings should be equal.

Selecting CH 2 INVERT causes the polarity of the CH 2 waveform

to be inverted. This allows you to see the difference between

CH 1 and CH 2 signals on the ADD trace.

CHOP or ALT?

When BOTH channels are selected, the display is time-shared.

The CHOP mode displays each channel for a short time and mul-

tiplexes during the sweep to give the appearance of displaying

both channels at once. This mode (CHOP) works better thanALT for sweep speeds slower than 1 ms per division and for

low-repetition-rate signals that make the display flicker (up to 2psldivision) .The ALT mode displays each channel for the duration of a com-

plete sweep. It gives a cleaner display of multiple channels than

CHOP does and is usually preferred at moderate to high sweep

speeds.

Certain trigger conditions, including composite trigger selection,

can cause a display that implies a phase relationship or evensynchronization of independent waveforms. If in doubt about the

relative timing of CH 1 and CH 2 signals, experienced operators

set the TRIGGER SOURCE to either CH 1 or CH 2.

2205 Operator 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 37/100

Operation

HORIZONTAL CONTROLS

Sweep-Speed Selection

The unmagnified sweep (MAG set to XI) is the horizontal func-

tion needed for most applications. Best measurement accuracyis achieved by setting the SECIDIV control for the fastest sweep

that will display the interval of interest. The variable control

(CAL) should be in its detent (fully clockwise).

Magnifying Waveform Details

The XI0 MAG mode expands the unmagnified trace. A magni-fied trace is useful for observing specific portions of a waveformwhen you are making accurat'e timing measurements or looking

at waveform details.

When selected, X I 0 MAG mode lets you examine details, suchas the entire leading edge of a pulse's second repetition, of a

trace up to 10 screen lengths from the trigger point.

Appendix A lists the sweep speeds for 10X magnification atevery SECIDIV control setting.

The one-division segment of the unmagnified trace centered on

the middle vertical graticule line is expanded and displayed asthe magnified trace. With the center vertical graticule line as the

reference, the investigation of waveform details around any

point on the unmagnified trace, as well as the measurement of

time with greater accuracy, become easy tasks.

TRIGGER CONTROL

For most signals, the trigger-control settings that will yield

hands-off triggering are:

MODE

SOURCE

P-P AUTO

VERT MODE

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 38/100

Operation

Which Mod e to Use

P-P AUTOITV LINE-With this mode set, the range of the LEVEL

control is confined to the values between the triggering-signal

peaks. For example, selecting P-P AUTO and rotating the

LEVEL control to the center of its range establishes a trigger

point about midway between the peaks of the triggering signal.

In this mode, the absence of a triggering signal causes the

sweep to free-run. And with signals below 20 Hz, the P-P AUTO

circuit may not find the correct level.

Whenever TRIGGER SOURCE is set to VERT MODE, the trigger-

ing signal is supplied by the channel being displayed - or as

shown in Table 2-1.

The P-P AUTO mode is effective for monitoring logic signals and

television lines having at least 20 Hz repetition rate. SelectingP-P AUTO at the instrument front panel also sets the TV LINE

triggering mode.

NORM - This mode produces a sweep only when the triggering

signal meets the criteria set by the LEVEL and SLOPE controls.

With NORM mode selected, range of the LEVEL control is suffi-

cient to set any voltage threshold that can be displayed by the

instrument. In the absence of a triggering signal, no sweep

occurs.

Use the NORM mode for viewing infrequent events and erratic

signals.

SGL SWP When this mode is selected, the sweep is triggered

only once. Press the RESET button once to arm the trigger cir-

cuit and illuminate the READY indicator. When a trigger event

occurs, the sweep runs once and the READY light extinguishes.

Use the SGL SWP mode to display or photograph nonrepetitive

or unstable signals.

TV FIELD - This mode triggers the sweep at the beginning of a

television field. To change the TV field being displayed, you

must interrupt the trigger signal by setting the input coupling

switch momentarily to GND then back either DC or AC until the

desired field is displayed.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 39/100

Operation

To display Field 1 and Field 2 at the same time, connect the

same television signal to both the CH 1 and CH 2 inputs; set

VERTICAL MODE to BOTH and ALT; and set the SEC/DIV control

to 0.5 ms or faster sweep speed.

If you magnify the vertical display beyond the graticule, the trig-

ger may be degraded. To avoid trigger overload, use either CH1 or CH 2 for display and use the EXT INPUT channel with an

appropriate video signal as the trigger source. A composite

sync signal can be used for the trigger source as well as com-

posite video.

Source

Choose a single channel as the trigger source to correctly dis-play the timing relationships between two channels. Choose the

channel with the lowest frequency signal to avoid ambiguous dis-

plays.

With TRIGGER SOURCE set to VERT MODE and VERTICAL MODE

set to ADD or CHOP, the triggering signal is the algebraic sum of

the Channel 1 and Channel 2 input signals.

Use a composite trigger source only to view asynchronous

signals simultaneously. To generate a composite trigger: select

VERT MODE TRIGGER SOURCE and BOTH-ALT VERTICAL MODE.

Slope

Use the SLOPE control to select either the rising ( f ) or thefalling(1)dge of the signal to trigger the sweep.

Level

The LEVEL control gives you complete freedom to choose the

most appropriate threshold voltage on a signal to initiate sweeps

whenever any trigger mode except P-P AUTO is selected.

CONNECTING SIGNALS

The signal adapter supplied with the instrument is usually the

most convenient way to connect a signal to the 2205. These

signal adapters are shielded to prevent pickup of electromag-netic interference. When connected to the 2205 input, a signal

2-1 2 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 40/100

Operation

adapter presents 1 M a and about 100 pF impedance to the cir-

cuit under test. If this capacitance is disruptive to the circuit be-

ing tested, use the optional 1OX probe.

Waveform Fidelity and Probe Grounds

When using a probe, its ground lead must be used for accurate

measurements and observations. Use the shortest ground con-nection possible for best waveform fidelity.

In some cases, a separate ground from the unit under test to the

ground receptacle on the oscilloscope front panel can reduceinterference from low-frequency hum and noise. For roughchecks of larger signals, such as 5 V logic, a ground lead sepa-rate from the probe - or even the safety ground connection,

which is shared with the unit under test - may work for a signalground. Fast signal transitions will be highly distorted, and extra-neous noise will be induced without the probe ground connec-tion, and/or with extra ground connections from the 2205 to the

circuit being tested.

Probe Compensation (Optional 1OX Probe)

Misadjustment of probe compensation is a common source ofmeasurement error. Due to variations in oscilloscope input

characteristics, probe compensation should be checked when-ever a 10X probe is moved from one oscilloscope to another orfrom one channel to another on the same oscilloscope. Always

compensate the probe to the channel on which it will be used.See the procedure in Section 4 , Checks and Adjustments.

Probe Handling (Optional 1OX Probe)

Both the probe and the probe accessories should be handledcarefully to prevent damage. Striking a hard surface can dam-

age both the probe body and the probe tip. Exercise care toprevent the cable from being crushed, kinked, or excessively

strained.

2205 Operator 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 41/100

Coaxial Cables

To maintain good waveform fidelity and accuracy, use only high-

quality, low-loss coaxial cables. When you use 50 R or 75 flcoaxial cable, attach a matching external terminator. Some high

frequency response will be lost without external termination.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 42/100

SECTION 3

APPLICATIONS

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 43/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l

  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 44/100

POSITION TO

CENTERLINE

VERTICALDEFLECTION

MEASUREAMPLITUDEFROM A TO 6

6717-04

Figure 3-1. Measuring peak-to-peak voltage of a waveform.

NOTE

If the amplitude measurement is critical or i f the

trace is thick (because of hum or noise on thesignal), a more accurate result can be obtainedby measuring from the top of a peak to the top of

a valley. This will eliminate trace width from the

measurement.

8. Calculate the peak-to-peak voltage, using the following for-mula:

vertical VOLTSIDIV

Vp-P = deflection x setting

(divisions) (1OX Probe) *

* I f a I X probe is being used for the measurement, use theI X VOLTSIDIV setting.

3-2 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 45/100

Applications

EXAMPLE. In Figure 3-1, the measured peak-to-peak verticaldeflection is 4 .4 divisions using a 10X attenuator probe with the

VOLTS/DIV switch set to 5V (1OX PROBE).

Substituting the given values:

Vp+ = 4.4 div x 5 V/div = 22 V.

Instantaneous Voltage

To measure the instantaneous voltage level at a given point on a

waveform, referred to ground, use the following procedure:

1. Apply the signal to either the CHI or the CH2 input connec-

tor and set the VERTICAL MODE selector to the channelused.

2. Verify that the VOLTS/DIV variable control is in the CAL

detent and set input coupling to GND.

3. Vertically position the trace to the center horizontal graticule

line. This establishes the ground reference location.

NOTE

If the measurements are to be made relative to a

voltage level other than ground, set the input cou-pling switch to DC and apply the reference voltage

to the input connector. Then position the trace tothe horizontal reference line.

4 . Set the input coupling control to DC. Points on thewaveform above the ground reference location are positive;points below are negative.

NOTE

If using Channel 2, ensure that the center VERTI-CAL MODE selector (NORM, CH 2 INVERT), is set

to NORM.

5 . If necessary, repeat Step 3 using a different horizontalground reference line that allows the waveform in Step 4 to

be displayed on screen.

2205 Opera to r 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 46/100

Applications

6. Adjust the TRIGGER LEVEL control to obtain a stable display.

7. Set the SECIDIV control to a position that displays at least

one cycle of the signal.

8. Measure the divisions of vertical deflection between the

ground reference line and the point on the waveform at

which the level is to be determined (see Figure 3-2) .

Figure 3-2. Instantaneous voltage measurement.

9. Calculate the instantaneous voltage, using the following for-mula:

Instan- vertical polarity VOLTS/ DIV

taneous = deflection x (+ or -) x setting

Voltage (divisions) (1OX Probe) *

*If the signal adapter is being used for the measurement,use the I X VOLTSIDIV setting.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 47/100

Applications

EXAMPLE. In Figure 3-2, the zero-volt reference is as shown atthe center vertical graticule line, a 10X attenuation probe is be-

ing used, and VOLTSIDIV is set to 2 V (10X PROBE). Two un-known voltage levels, A and B, are marked on the waveform. A

is 1.5 divisions above the zero-volt reference level, and B isthree divisions below the zero-volt reference level. What are the

respective voltages of levels A and B?

Substituting the given values:

lnstantaneous Voltage A = 1.5 div x (+I x 2 Vldiv = 3.0 V.

lnstantaneous Voltage B = 3.0 div x (-1 ) x 2 Vldiv = -6.0 V.

Algebraic Addition

With the three VERTICAL MODE selectors set to BOTH-NORM-

ADD, the waveform displayed is the algebraic sum of the signalsapplied to the Channel 1 and Channel 2 inputs (CHI + CH2). If

the middle MODE selector is then set to CH2 INVERT, thewaveform displayed is the difference between the signals ap-plied to the Channel 1 and Channel 2 inputs (CHI - CH2). When

both VOLTSIDIV controls are set to the same deflection factor,

the deflection factor of the ADD trace is equal to the deflectionfactor indicated by either VOLTSIDIV control.

The following general precautions should be observed when us-ing ADD VERTICAL MODE.

1. Do not exceed the input voltage rating of the oscilloscope.

2. Do not apply signals whose peaks exceed the equivalent of

about +8 imes the VOLTSIDIV settings, because large volt-ages may distort the display. For example, with a VOLTS1

DIV setting of 0.5 V, the voltage applied to that channel

should not exceed approximately 4 V.

3. Position the Channel 1 and Channel 2 waveforms near ten-

ter screen, when viewed separately. This ensures the great-

est dynamic range for ADD mode operation.

4 . To attain similar responses from both channels, set theChannel 1 and Channel 2 input coupling selectors to the

same position.

2205 Operator 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 48/100

Applications

Common-Mode Rejection

The following procedure shows how to eliminate unwanted AC

line frequency components. Similar methods could be used toeliminate other unwanted frequency components or to provide a

DC off set.

1. Apply the signal containing the unwanted line-frequency

components to the CHI input connector.

2. Apply a line-frequency signal to the CH2 input connector. Tomaximize cancellation, the signal applied to Channel 2 mustbe in phase with the unwanted line frequency component ofthe Channel 1 input.

3. Set the three VERTICAL MODE selectors to BOTH-NORM-ALT; set both VOLTSIDIV controls equally to produce dis-plays of approximately four or five divisions amplitude.

4. Adjust the CH2 VOLTSIDIV and Variable (CAL) controls sothat the Channel 2 display is approximately the same ampli-tude as the undesired portion of the Channel 1 display (seeFigure 3-3 top).

5. Now set the middle and right VERTICAL MODE selectors toCH2 INVERT and ADD. Slightly readjust the CH2 VOLTSIDIV

variable control for maximum cancellation of the undesired

signal component (Figure 3-3 bo ttop) .Amplitude Comparison (Ratio)

In some applications, it may be necessary to establish a set of

deflection factors between step settings of the VOLTSIDIV con-trols. This is useful for comparing unknown signals to a refer-

ence signal of known amplitude.

To accomplish this, a reference signal of known amplitude is first

set to an exact number of vertical divisions by adjusting the

VOLTSIDIV and VOLTSIDIV Variable (CAL) controls. Unknownsignals can then be quickly and accurately compared to the ref-erence signal without disturbing the setting of the VOLTSIDIV

Variable controls.

1. Apply the reference signal to either the CHI or the CH2 inputconnector and set the left VERTICAL MODE selector to dis-

play the channel used.

3-6 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 49/100

CH 1 SIGNALWITH UNWANTEDLlNE FREQUENCY

COMPONENT

CH 2-SIGNALFROM LINE

FREQUENCYSOURCE

INVERTED

SlGNAL WITHLlNE FREQUENCY

COMPONENTCANCELED

CH 1AND CH 2 SIGNALS

RESULTANT SIGNAL

67 7-01

Figure 3-3. Common-mode rejection.

2. Set the amplitude of the reference signal to 5 vertical divi-sions by adjusting the VOLTS/DIV and VOLTSIDIV Variable(CAL) controls.

3. Disconnect the reference signal and apply the unknown sig-nal to the same channel input. Adjust the vertical position ofthe waveform so that its bottom edge just touches the 0%line on the crt.

4. Horizontally position the waveform so that its topmost fea-tures cross the center vertical graticule line (see Fig-

ure 3-4 ) .

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 50/100

Applications

READ PERCENTRATIO (30%)

Figure 3-4. Determining voltage ratio.

5. Read the percent ratio directly from the graduations on thevertical centerline, referring to the 0% and 100% percentagemarks on the left edge of the graticule (1 minor division

equals 4% for a 5-division display).

TIME MEASUREMENTS

Time Duration

To measure time between two points on a waveform, use the

following procedure:

1. Apply the signal to either the CHI or the CH2 input connec-

tor and set the VERTICAL MODE selector to display the chan-nel used.

2. Adjust the TRIGGER LEVEL control to obtain a stable display.

3. Set the SEC/DIV control to display between one and twocomplete repetitions of the waveform. Check that the SEC/

DIV variable control is in the CAL detent.

3-8 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 51/100

Applications

4 . Position the display to place the time-measurement points

on the center horizontal graticule line (see Figure 3-5).

- . HORIZONTAL

DISTANCE

1 67 17-08]

Figure 3-5. Measuring time duration.

5. Measure the horizontal distance between the time measure-

ment points.

6. Calculate time duration using the following formula:

horizontal SEC/DIVdistance control

Time -- (divisions) setting

Duration magnification factor

EXAMPLE. In Figure 3-5, the distance between the time meas-urement points is 8.3 divisions, and the SECIDIV setting is 2 msper division, HORIZONTAL MAG is set to X I .

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 52/100

Substituting the given values:

Time Duration = 8.3 div x 2 msldiv = 16.6 ms

Period and Frequency

In the preceding example, you measured the time duration of

one complete waveform cycle. This particular measurement is

called the waveform period (T) . The frequency (f) of a recur-rent signal can be determined from its period as follows:

1. Measure the time duration of one waveform cycle (period)

using the preceding time-duration measurement procedure.

2. Calculate the reciprocal of the period to determine the

waveform frequency.

I'

XAMPLE. The signal in 'Figure 3-5 has a period (T) of 16.6 ms.

Calculating frequency (f) :

1 1f = - -

T - 16.6 I O - ~ S= 60 Hz

Rise Time

Rise time measurements use the same methods as time dura-tion, except that the measurements are made between the 10%and 90% points on the low-to-high transition of the selectedwaveform. Fall time is measured between the 90% and 10%points of the high-to-low transition of the waveform.

1. Apply a signal to either the CHI or the CH2 input connector

and set the VERTICAL MODE selector to display the channel

used.

2. Set the appropriate VOLTSIDIV and VOLTSIDIV Variable(CAL) control for an exact 5-division display.

3. Vertically position the traces so that the zero reference ofthe waveform touches the 0% graticule line and the top of

the waveform touches the 100% graticule line.

4 . Horizontally position the display so the 10% point on the

waveform intersects the second vertical graticule line (Fig-

ure 3-6, Point A).

3-10 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 53/100

7-IGNAL

AMPLITUDE

4ORIZONTAL DISTANCE)C

Figure 3-6. Measuring rise time.

NOTE

You can achieve better accuracy by using the

SECIDIV control, or horizontal magnification, to

expand the waveform transition so that it occupiesfrom four to six horizontal divisions between the10% and 90% amplitude points.

5. Measure the horizontal distance between the 10% and 90%points (Figure 3-6, Points A and B) and calculate time dura-tion using the following formula:

horizontal SECIDIVdistance control

Rise -- (divisions) setting

Time magnification factor

EXAMPLE. In Figure 3-6, the horizontal distance between the

10% and 90% amplitude points is 5 divisions, and the SECIDIVswitch is set to 1 ps per division. HORIZONTAL MAG is set to

X I .

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 54/100

Substituting given values in the formula:

Rise Time =

5 div x 1 psldiv

1 = 5 ps

Time Difference Between Pulses on Time-Related Signals

The calibrated sweep speed and dual-trace features of the 2205

allow measurement of the time difference between two separate

events. To measure time difference, use the following proce-

dure:

1. Set the TRIGGER SOURCE to CHI.

2. Set both input coupling controls to the same position.

3. Using probes or cables with equal time delays, apply aknown reference signal to the CHI input connector and ap-

ply the comparison signal to the CH2 input.

4 . Set both VOLTSIDIV controls for a display of about five divi-

sions.

5. Set VERTICAL MODE to BOTH; then select either ALT orCHOP, depending on the frequency of the input signals.

6. If the two signals are opposite in polarity, set the middle

VERTICAL MODE selector to CH2 INVERT to invert the Chan-

nel 2 display.

7. Adjust the TRIGGER LEVEL control for a stable display.

8. Set the SECIDIV control to a sweep speed that providesthree or more divisions of horizontal separation betweenmeasurement points on the two displays. Center each of

the displays vertically (see Figure 3-7).

2205Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 55/100

Applications

CHANNEL 1 (REFERENCE) CHANNEL 2

IHORIZONTAL

I- DIFFERENCE y-1 I

Figure 3-7. Time difference between pulses ontime-related signals.

9. Determine the horizontal difference between the two signalmeasurement points and calculate the time difference usingthe following formula:

horizontal SECJDIVdifference control

Time - (divisions) settingDifference

magnification factor

EXAMPLE. In Figure 3-7, the SECJDIV control is set to 50 psper division and HORIZONTAL MAG is set to X10. The horizontal

difference between waveform measurement points is 4.5 divi-sions.

Time - 50 mldiv x 4.5 divDifference 10

= 22.5 ps

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 56/100

Applications

Phase Difference

In a similar manner to the preceding measurement, you can

make a phase comparison between two signals of the same fre-quency using the dual-trace feature of the 2205. This method

of phase-difference measurement can be used for signals withfrequencies up to the limit of the vertical deflection system. To

make a phase comparison, use the following procedure:

1. Set the TRIGGER SOURCE selector to CHI.

2. Set both input coupling controls to the same position, de-

pending on the type of input coupling desired.

3. Using probes or cables with equal time delays, apply a

known reference signal to the CHI input connector and ap-

ply the unknown signal to the CH2 input.

4 . Set VERTICAL MODE to BOTH; then select either ALT or

CHOP, depending on the frequency of the input signals. The

reference signal should precede the comparison signal intime.

5. If the two signals are opposite in polarity, set the middleVERTICAL MODE switch to CH2 INVERT to invert the Channel

2 display.

6. Set both VOLTSIDIV and VOLTSIDIV Variable (CAL) controls

to display equal-amplitude waveforms.

7. Adjust the TRIGGER LEVEL control for a stable display and

center the display vertically.

8. Set the SECIDIV control to a sweep speed that displays

about one full cycle of the waveforms.

9. Position the displays and adjust the SECIDIV Variable (CAL)control so that one cycle of the reference signal occupies

exactly eight horizontal graticule divisions at the 50% rise-

time points (see Figure 3-8). Each horizontal division of the

graticule now represents 45" of the cycle (360" divided by 8divisions), and the horizontal graticule calibration can be

stated as 45" per division.

3-1 4 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 57/100

Applications

10. Measure the horizontal difference between corresponding

points on the two waveforms at the 50% rise-time points and

calculate the phase difference using the following formula:

horizontal graticulePhase - difference calibrationDifference - (divisions) (deg/div)

EXAMPLE. In Figure 3-8, the horizontal difference is 0.6 divi-

sion with a graticule calibration of 45" per division.

Substituting the given values into the phase-difference formula:

Phase Difference = 0.6 div x 45Otdiv = 27O.

CHANNEL 1 CHANNEL2(REFERENCE) f (LAGGING)

MEASURETIME FROM A TOB

HORIZONTALDIFFERENCE

Figure 3-8. Phase difference.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 58/100

Applications

More accurate phase measurements can be made by using thehorizontal magnifier function to increase the sweep speed with-

out changing the SECfDIV variable control setting. To do this,

set the HORIZONTAL MAG to X10.

With the sweep speed increased 10 times (MAG set to X I 0) , themagnified horizontal graticule calibration will be 4.5" per division

(45OIdivision divided by 10). Figure 3-9 shows the same signalsillustrated in Figure 3-8, but horizontally magnified by a factor of

10.

EXAMPLE. In Figure 3-9, the 10X magnified display results in a

horizontal difference of six divisions between the two signals.

Substituting the given values into the phase difference formula:

Phase Difference = 6 div x 4.5"Idiv = 27".

I CHANNEL1 (REFERENCE) CHANNEL2 (COMPARISON)

MEASURETIME FROM

ATOB

I- HORIZONTALI DIFFERENCE

+I

6717-1;

Figure 3-9. High-resolution phase difference.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 59/100

Applications

'TELEVlSlON DISPLAYS

T V Field Signals

The 2205 television feature can be used to display TV field sig-

nals. To do so, proceed as follows:

1. Set the TRIGGER MODE selector to TV FIELD and set theSECIDIV control to 2 ms.

2. Apply the television signal to either the CHI or the CH2 input

connector and set the VERTICAL MODE selector to display

the channel used.

3. For positive-going TV signal sync pulses, set the TRIGGERSLOPE to f nd rotate the LEVEL control fully counter-

wise. (Sometimes it may be necessary to move the LEVEL

control slightly clockwise to obtain stable triggering.)

4 . Set the appropriate VOLTSIDIV control to display one divi-sion or more of composite video signal.

5. To change the TV field being displayed, momentarily inter-

rupt the trigger signal by setting the input coupling to GND,

then back to DC or AC until the desired field is displayed.

INOTE

To examine a TV Field signal in more detail, set

HORIZONTAL MAG to X10.

6. To display either Field 1 or Field 2 individually, connect the

television signal to both the CHI and CH2 input connectors

and select BOTH and ALT VERTICAL MODE. Set the SECI

DIV control to 0.5 ms or faster sweep speed (displays lessthan one full field). This will synchronize the Channel 1 dis-

play to one field and the Channel 2 display to the other field.

2205 Opera to r 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 60/100

Applications

TV Line Signals

The following procedure is used to display a TV Line signal.

1. Verify that TRIGGER MODE is set to P-P AUTO/TV LINE.

2. Apply the Television signal to either the CHI or the CH2 input

connector and set the VERTICAL MODE selector to displaythe channel used.

3. Set the SECIDIV control to 10 ps.

4 . For positive-going TV signal sync pulses, set the TRIGGER

SLOPE to f nd rotate the LEVEL control fully clockwise.

For negative-going sync pulses, set the TRIGGER SLOPE to'7, and rotate the LEVEL control fully counterclockwise.

(Sometimes it may be necessary to move the LEVEL control

I

lightly clockwise to obtain stable triggering.)

5. Set the appropriate VOLTS/DIV control to display one divi-

sion or more of composite sync signal.

NOTE

To examine a TV Line signal in more detail, set

HORIZONTAL MAG to X10.

6. To display a selected TV Line pulse, first trigger the sweep

on a TV Field sync pulse. Use the HORIZONTAL POSITION

control to select the desired line pulse, then set HORIZON-TAL MAG to X10.

Z-MODULATION

The Z-modulation system can be used to display time markers,

because it depends entirely upon the accuracy of the signal

source. It can also be used in any situation where external con-

trol of the brightness of some or all of the trace is required.

The Z, or intensity, modulation feature is operated in the follow-ing manner:

3-18 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 61/100

Applications

1. Set the left and right TRIGGER SOURCE controls to EXT and

EXT=Z, respectively.

2. Apply a signal to either the CHI or the CH2 connector and

set the VERTICAL MODE selector to display the channel

used.

3. Apply the Z-modulation signal to the connector labeled EXTINPUT or Z .

When the Z-modulation and the vertical-input signals are syn-chronized, Z-modulation is seen as gaps in the trace at themodulation frequency. The size of the gap depends upon the

mark-to-space ratio of the Z-modulation signal. The positive-

going portion of the Z-modulation signal decreases display

brightness.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 62/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 63/100

SECTION 4

CHECKS AND

ADJUSTMENTS

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 64/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 65/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l

  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 66/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 67/100

Checks and Adjustments

9. Select CH2 VERTICAL MODE. Disconnect the Channel 1

probe tip and attach the Channel 2 probe tip to the PROBEADJUST output terminal.

10. Obtain a 5-division signal display and vertically center thetrace.

11. Repeat step 7 for the Channel 2 probe.

LFCOMP ROTATINGSLEEVE

\@-?

Figure 4-2. Locating the low-frequency

compensation adjustment.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 68/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 69/100

SECTION 5

PERFORMANCE

CHARACTERISTICS

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 70/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 71/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 72/100

TABLE 5-1 (cont'd)

VERTICAL DEFLECTION SYSTEM (cont'd)

Characteristics Perform ance Req uireme nts

CHOP Mode Switching 1 500 kHz 5 30%.

Rate

Bandwidth (-3 dB)

+5"C to t35"C

O°C to +5"C and

+35"C to +40°C

AC Coupled Lower CutoffFrequency

Input Characteristics I

20 MHz or more.

15 MHz or more.

10 Hz or less at -3 dB.

Resistance I 1 M a ? 2 % .I

Capacitance 1 25 pF ?2pF.

Maximum Safe InputVoltage (DC or ACCou~led)

Trace Shift

400 V (dc + peak ac) or 800 Vac p-p to 10 kHz or less.

Common-Mode RejectionRatio (CMRR)

With VOLTSIDIVSwitch Rotation

At least 10 to 1 at 20 MHz.

0.75 division or less (Variablecontrol in CAL detent).

With VOLTSIDIV Variable 1 1.0 division or less.Control Rotation IWith Channel 2 I 1.5 divisions or less.Inverted

2205 Operator's

Channel Isolation Greater than 100 to 1at 10 MHz.

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 73/100

Performance Characteristics

TABLE 5-1 (cont'd)

TRIGGER SY ST EM

Characteristics

Trigger Sensitivity

Performance Requirements

Lowest UsableFrequency in P-P

AUTO Mode

P-P AUTO/TV LINEand NORM Modes

Internal Signal

External Signal

Resistance I 1 M i l 2 10%.

5 MHz

0.3div

40 mV

TV FIELD Mode

External Input

Capacitance 1 25 pF 22.5 F.

30 MHz

1.0div

150 mV

1.0division of composite sync.

Maximum Voltage I 400 V (dc + peak ac) or 800 Vac p-p at 10 kHz or less.

Trigger Level Range I

2205 Operator's

NORM Mode

EXT Source

+_I5 divisions referred to theappropriate vertical input.

At least 2 1.6V, 3.2 V p-p.

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 74/100

Performance Characteristics

TABLE 5-1 (cont'd)

HORIZONTAL DEFLECTION SYSTEM

Characteristics

Sweep Rate

Calibrated Range

Performance Requirements

Accuracy

Variable Control Range

Sweep Linearity

Position Control Range

Registration of Unmagni-fied and Magnified Traces

0.5 s per division to 0.1 ps perdivision in a 1-2-5 sequence.Magnification extends maximumusable sweep speed to 10 nsper division.

Ma nified

+ 3% + 4%

5 4 % 2 5%2ontinuously variable and un-

calibrated between calibratedstep settings. Decreases cali-brated sweep speeds at leastby a factor of 2.5.

Ma nified

2 7%

Start of sweep, to 10th divisionin XI and to 100th division inX10, will position past the centervertical graticule line.

0.2 division or less, aligned tocentral vertical graticule line.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 75/100

Performance Characteristics

TABLE 5-1 (cont'd)

Z-MODULATION

Characteristics Performance Requirements

Sensitivity

Usable Frequency Range

X-Y OPERATION (X I MOD E)

5 V causes noticeable modu-lation. Positive-going inputdecreases intensity.

Dc to 5 MHz.

Maximum Safe InputVoltage

Deflection Factors

400 V (dc + peak ac) or 800 Vac p-p to 10 kHz or less.

Same as Vertical DeflectionSystem with Variable controlsin CAL detents.

Accuracy

X- Axis

Bandwidth (-3 dB) IY- Axis

X- Axis I Dc to at least 1 MHz.I

Same as Vertical DeflectionSystem.

PROBE ADJUSTMENT SIGNAL OUTPUT

Y-Axis

Phase Difference BetweenX- and Y-Axis Amplifiers

Voltage into 1 M i l Load 1 0.5 V 2 5%.

I

Same as Vertical DeflectionSystem.

+ 3 O f rom dc to 50 k ~ z .

Re~etitionRate I 1 kHz It20%.

2205 Operator 's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 76/100

Performance Characteristics

TABLE 5-1 (cont'd)

POWER REQUIREMENTS

Characteristics

Line Voltage Ranges

Performance Requirements

1 15 V Settina

230 V Setting

Line Frequency

Maximum PowerConsumption

Line Fuse

115 V Setting

230 V Setting

95 V ac to 128 V ac.

185 V ac to 250 V ac.

48 Hz to 440 Hz.

40 W (60 VA) .UL198.6, 3AG (114 x 1% inch)

0.75 A Slow

0.5 A, Slow

CATHODE-RAY TUBE

Display Area 1 80 by 100 mm.

Standard Phosphor I GH (P31).I

2205 Operator's

Nominal AcceleratingVoltage

1800 V 510%.

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 77/100

Performance Characteristics

Table 5-2

Environmental Characteristics

Altitude

Operating

Characteristics

Temperature

Operating

Nonoperating

To 4500 m (15,000 ft.). Maxi-

mum operating temperaturedecreases 1" C per 300 m(1,000 ft.) above 1,500 m

(5,000 ft.).

Performance Requirements

0°C to +40 "C (t32"F to+lo4 O F).

-55 O C to +75 O (-67 O F tot167OF).

Vibration (Operating)

Nonoperating

Relative Humidity

Operating

(+30°C to +40 "C)

Nonoperating

(+30°C to +60°C)

Shock (Operating andNonoperating)

- -To 15,250 m (50,000 ft.).

85% to%, -5%

85% +O%, -5%

Radiated and ConductedEmission Requirements

major axes at atotal displace-ment of 0.015 inch p-p (2.4 gat 55 Hz) with frequency variedfrom 10 Hz to 55 Hz to 10 Hzin one-minute sweeps. Hold for10 minutes at 55 Hz in each ofthree major axes. All majorresonances must be above55 Hz.

30 g, half-sine, 11-ms du-ration, three shocks per axiseach direction, for a total of18 shocks.

Meets VDE 0871 Class B.

2205 Operator's 5-7

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 78/100

Performance Characteristics

Table 5-3

Mechanical Characteristics

Characteristics I Performance Requirements--Weight With Power Cord

Height 1 138 mm (5.4 in).

8.25 kg ( 13.7 lbs ) or less.

Domestic Shipping Weight

Width

With Handle

9.1 kg (20.0 Ibs) or less.

Without Handle1

327 mm (12.9 in).Depth I

With HandleExtended

Without Front Cover

With Optional FrontCover

10kHz 50kHz 100kHz 500 kHz 1 MHz 100MHz

FREQUENCY

441 mm ( 17.4 in ).

455 mm ( 17.9 in ).

Figure 5-1. Maximum input voltage versus frequency deratingcurve for CH 1 OR X, CH 2 OR Y, and EXT INPUT connectors.

5-8 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 79/100

SECTION 6

OPTIONS ANDACCESSORIES

2205 Operator's I

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 80/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 81/100

  Q  s  e  r  v  i  c  e  -  -  -  -  -  T  h  i  s  D  o  c  u  m  e  n  t  i  s  a  c  o  m  p  l

  e  t  e  s  c  a  n  f  r  o  m   t

  h  e  O  r  i  g  i  n  a

  l   T  e  k  t  r  o  n  i  x  M  a  n  u  a  l   -  -  -  -  -

  Q  s  e  r  v  i  c  e

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 82/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 83/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 84/100

Options and Accessories

TABLE 6-1

Power Cords and Fuses

Description

Standard (United States)

Power Cord, 2.5 m

Fuse, 0.75 A,250 V,3AG, V4" x 1 /4", Slow

Option A1 (UniversalEurope)

Power Cord. 2.5 m

Fuse, 0.5 A, 250 V ,

3AG, 114"x 1 /d; Slow

Option A2 (UnitedKingdom)

Power Cord, 2.5 m

Fuse, 0.5 A, 250 V,3AG, l14" x 11/4", Slow

Option A3 (Australia)

Power Cord. 2.5 mFuse, 0.5 A, 250 V,3AG, V4" x 1W: Slow

Option A4 (North America)

Power Cord, 2.5 m

Fuse, 0.5 A, 250 V,3AG, 114" x 11/4", Slow

Option A5 (Switzerland)

Power Cord, 2.5 mFuse, 0.5 A , 250 V,3AG, 114" x 11/4", Slow

Part Number

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 85/100

Options and Accessories

OPTIONAL ACCESSORIES

Table 6-2 lists recommended optional accessories for your in-strument.

TABLE 6-2

Optional Accessories

Front Panel Protective Cover

Accessory Pouch

Front Panel Protective Coverand Accessory Pouch

Hand Carrying Case

CRT Light Filter, Clear

Rack Mount Conversion Kit

Viewing Hoods

Collapsible

Polarized

Binocular

Probe, 10X, 2 m, withaccessories

Alternative Power Cords

European

United Kingdom

Australian

North AmericanSwitzerland

2205 Operator's

Part Number

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 86/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 87/100

APPENDIX AMagnified Sweep Speeds

SEClDlVSetting

Magnified Sweep Speed(TimelDivision)

X I 0

50 ms

20 ms

10ms

5 ms

2 ms

1 ms

.5 ms

.2 ms

.1 ms

50 ps

20 ps

10 ps

5 JIS

2 PS

1 JIS

.5 JIS

.2 ps

. I ps

50 ns

20 ns

10 ns

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 88/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 89/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 90/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 91/100

APPENDIX C

GLOSSARY

DISPLAY CONTROLS

BEAM FIND-Finds the beam when it is off the screen.

FOCUS-Adjusts the trace for optimum focus.

INTENSITY-Adjusts the brightness of the trace.

POWER (ON-OFF)-Turns on AC power to the power supply.When the light is on the instrument operates; when the light is

off the instrument is off.

TRACE ROTATION-Adjusts the trace to be parallel with the fixed

graticule. (See OTHERS for graticule) .

VERTICAL CONTROLS

ADD-Combines Channel 1 and Channel 2 algebraically.

ALTERNATE (ALT)-Shows CH 1 and CH 2 alternately; first onesweep of Channel 1 then one sweep of Channel 2.

Alternating Coupling (AC)-Static signal components are

blocked and only the alternating components of the input signal

reach the channel.

BOTH-Displays CH 1 and CH 2 in ALT, CHOP, or ADD mode.

CAL-Changes the variable scale factor up to 2.5 times the

VOLTS/DIV setting.

Channel 1 (CH 1)-Displays only Ch 1 on the screen.

Channel 2 (CH 2)-Displays only Ch 2 on the screen.

Channel 2 Invert (CH 2 INVERT)-Inverts the polarity of CH 2.

(See OTHERS for polarity.)

CHOP-The scope will draw part of CH 1 , then part of CH 2 by

switching back and forth.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 92/100

Channel 1 VoltslDivision (CH 1 VOLTS1DIV)-This switch is the

scale factor for CH 1 .

Channel 2 VoltsIDivision (CH 2 VOLTS1DIV)-This switch is thescale factor for CH 2.

Direct Coupling (DC)-All components of the input signal reach

the input channel.

GROUND (/)I)-Shows where the zero level will be located when

shifted to AC or DC.

Normal (NORM)-CH 1 and Ch 2 both have the same polarity

when set to NORM. (See OTHERS for polarity.)

POSITION-Moves the trace vertically on the screen.

HORIZONTAL CONTROLS

CAL-Reduces the SEClDlV by up to 2.5 :1 .

Ground receptacle-Provides safety earth connection to signal

source.

Magnified times 1 (MAG XI)-Magnifies the SECIDIV by 1.

Magnified times 10 (MAG X10)-Magnifies the SECIDIV by 10.

POSITION-Moves the trace horizontally on the screen.

PROBE ADJUST-Provides a square wave for use as a reference

when adjusting compensation on 1OX probes.

SecondsIDivision (SEC1DIV)-Lets you select the rate at which

the beam sweeps across the screen.

TRIGGER CONTROLS

Channel 1 (CH 1)-Selects CH 1 signals to trigger the time base.

Channel 2 (CH 2)-Selects CH 2 signals to trigger the time base.

External (EXT)-Selects external signals to trigger the time base.

External 10 (EXTI 10) -Selects external signal, divides it by 10,

and connects it to the trigger circuit.

LEVEL-Determines signal level at which the trigger will occur.

C-2 2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 93/100

LINE-Another trigger source (the power line).

Normal (NORM) -Allows selection of any TRIGGER LEVEL within

approximately +I 5 divisions.

Peak to Peak Automatic (P-P AUTO)-Limits TRIGGER LEVEL to

a range determined by peaks of signal. If no signal appears in

about 5 ms, the sweep starts on its own to show a "baseline."

RESET-This button is used only in SGL SWP Mode to reset the

sweep. (See TRIGGER for single sweep.)

Single Sweep (SGL SWP)-A mode in which, when a trigger oc-

curs, only one sweep will run.

SLOPE-Determines whether the trigger point will occur on the

raising or falling edge of the signal.

Trigger Ready (TRIG'D)-Lights when the sweep is triggered.

Television Field (TV FIELD) -Trigger on tv fields.

Television Line (TV LINE)-Trigger on tv lines.

Vertical Mode (VERT MODE)-Trigger source is determined by

the Vertical Mode..

OTHERS

Alternation-A change in polarity, like a positive (+) changing

directions to a negative (-) and back again.

Cathode-Ray Tube (CRT) -The picture tube.

Divisions-The lines that divide CRT display area units into X and

Y components.

Graticule-The square boxes formed by lines on the CRT face.

Hertz-The basic unit of frequency.

Ohm (a)--The basic unit of resistance.

Polarity-The positive (+) or negative (- ) direction of voltage or

current.

2205 Operator's

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 94/100

For further information, contact:

U. S. A. , Asia, Australia, Central & SouthAmerica, JapanTektronix, Inc.

P. 0. Box 1700

Beaverton, Oregon 97075

For additional literature, or the address and phone

number of the Tektronix Sales Office nearest you,contact:

Phone: (800) TEK-WIDE

Telex: (910) 467-8708

TLX: 151754

Cable: TEKWSGT

Europe, Africa, Middle EastTektronix Europe B. V.

European Headquarters

Postbox 827

1180 AV Amstelveen

The Netherlands

Phone: (20) 471 146

Telex: 18312-1 8328

CanadaTektronix Canada inc.

P. 0. Box 6500

Barrie, Ontario L4M 4V3

Phone: (705) 737-2700

Tektronix sales and service offices aroundthe world:Albania, Algeria, Angola, Argentina, Australia, Austria,

Bangladesh, Belgium, Bolivia, Brazil, Bulgaria, Canada,

Peoples Republic of China, Chile, Colombia, Costa Rica,

Czechoslovakia, Denmark, East Afr ica, Ecuador, Egypt,

Federal Republic of Germany, Fiji AWA New Zealand,

Finland, France, Greece, Hong Kong, Hungary, Iceland,

India, Indonesia, Ireland, Israel, Italy, Ivory Coast, Japan,

Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico,

Morocco, The Netherlands, New Zealand, Nigeria,

Norway, Pakistan, Panama, Peru, Philippines, Poland,

Portugal, Qatar, Republic of South Africa, Rumania,

Saudi Arabia, Singapore, Spain, Sri Lanka, Sudan,

Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey,

Tunisia, United Arab Emira tes, United Kingdom, Uruguay,

USSR, Venezuela, Yugoslavia, Zambia, Zimbabwe.

Copyri h t@ 1987, Tektronix , Inc. All rights reserved.

Printej in U. S. A. Tektronix products are covered by U. S.

and foreign patents, issued and pending. Information in

this publication supersedes that in all previously published

material. Specificat ion and price change privileges

reserved. TEKTRONIX, TEK, SCOPE-MOBILE, andTELEQUIPMENT are registered trademarks. For further

information, contact : Tektronix, Inc, Corporate Offices,

P.O. Box 500, Beaverton, OR 97077. Phone (503)

627-7111; TLX: 151754; TW X: (910) 467-8708; Cable:

TEKWSGT; Subsidiaries and distributors worldwide.

COMMITTED TO EXCELLENCE

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 95/100

MANUAL CHANGE INFORMATION

At Tektronix, we continually strive to keep up with latest electronicdevelopments

by adding circui t and component improvements to our instruments as soon as they

are developed and tested.

Sometimes, due to printing and shipping requirements, we can't get these

changes immediately into printed manuals. Hence, your manual may contain new

change information on follow ing pages.

A single change may affect several sections. Since the change info rmation sheets

are carried in the manual until all changes are permanently entered, some

duplication may occur. If no such change pages appear following this page, your

manual is correct as printed.

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 96/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 97/100

Rktronbc+ MANUAL CHANGE INFORMATIOND Q e * n B ) m v

Date:02-01 -90 Change Reference: C1/0290

q-oduct: 2205 Operators Manual Manual Part Number:070-6717-00-- -- - -DESCRIPTION Product Group

EFFECTIVE ALL SERIAL NUMBERS

Change page 6-2 to read:

Option 23

Two P6119 1X- 1OX Selectable-attenuation Probes areprovided in place of the standardP6103 10X Probes.

Option 24

Two P6lO3 ,dc to 50 MHz, 1OX Modular Probes are pro-vided.

Change page6-6:

Table 6-2 (cont'd)

Attenuator Voltage Probes1OX Standard10X Submininature10X Environmental1X-1OX Selectable100X High Voltage1000X High Voltage

Part Number

Page 1 of 1

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 98/100

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 99/100

MANUAL CHANGE INFORMATION

Date: 3-20-90 Change Reference: C1 0390

Product: 2205 OSCILLOSCOPE OPERATORS Manual Part Number: 070-6717-00

DESCRIPTION Product Group 46

EFFECTIVE ALL SERIAL NUMB ERS

I PAGE 5-4

CHANGE TO:

Characteristics

Sweep Linearity

Table 5-1 (cont)

Performance Requirements

MagnifiedI

Linearity measured over any 2of the center8divisions.

Linearity measured over any 2 of thecenter 8divisions,

excluding the first 40 ns.

Page 1 of 1

8/14/2019 Qservice ----- This Document is a Complete

http://slidepdf.com/reader/full/qservice-this-document-is-a-complete 100/100