40
PROTOCOL No. PERFORMANCE QUALIFICATION PROTOCOL FOR DRY HEAT STERILIZER SUPERSEDES NIL EFFECTIVE DATE PAGE No. Page 1 of 40 PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT PROTOCOL FOR QUALIFICATION OF DRY HEAT STERILIZER, FULLY AUTOMATIC EQUIPMENT TAG NO.:……………. AREA: MICROBIOLOGY LAB Document No. Supersedes Effective Date No. Of Pages 41

PROTOCOL FOR QUALIFICATION O D H S F A E T N .:……………. A M L · 2020. 12. 21. · 5.3.2.2 Validation Port The unit is provided with a 1 Validation port for the insertion

  • Upload
    others

  • View
    2

  • Download
    1

Embed Size (px)

Citation preview

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    PROTOCOL

    FOR

    QUALIFICATION

    OF

    DRY HEAT STERILIZER, FULLY AUTOMATIC

    EQUIPMENT TAG NO.:…………….

    AREA: MICROBIOLOGY LAB

    Document No.

    Supersedes

    Effective Date

    No. Of Pages 41

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    PROTOCOL CONTENTS

    S. No. Section Title Page No.

    1.0 Protocol Approval ......................................................................... 3

    2.0 Overview ......................................................................................... 4

    2.1 Objective .............................................................................. 4

    2.2 Purpose and Scope……………………..………………….4

    2.3 Responsibility ...................................................................... 4

    2.4 Qualification Team .............................................................. 4

    3.0 Training Record ............................................................................ 5

    4.0 Qualification Requirements .......................................................... 6

    5.0 System / Equipment Description .................................................. 7

    5.1 System / Equipment details ................................................ 7 5.2 System/Equipment Identification/Description…………...7

    5.2 Generic Design .................................................................... 7

    5.3 Safety features ................................................................... 10

    6.0 Performance Qualification Procedure ....................................... 12

    6.1 Air Velocity Test………………………………………….12

    6.2 HEPA Filter Integrity Testing…………………………...13

    6.3 Air Flow Pattern Test…………………………………….14

    6.4 Heat Distribution Study………………………………….15

    6.5 Heat Penetration Study…………………………………..15

    6.6 Endotoxin Challenge Study……………………………...18

    6.7 Non-viable Air Borne Particle Count.…………………..19

    7.0 Acceptance Criteria ..................................................................... 20

    8.0 Qualification Report .................................................................... 21

    9.0 Approval of Qualification Report .............................................. 21

    10.0 Observed Deviation ..................................................................... 22

    11.0 List of Exhibits/Annexure ........................................................... 23

    11.1 List of Exhibits .................................................................. 23

    11.2 List of Annexure ................................................................ 23

    12.0 Reference Documents ................................................................. 23

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 3 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    1.0 Protocol Approval

    This is a specific protocol for Qualification of Dry Heat Sterilizer (Fully Automatic)

    which is installed, in the New Microbiology Lab.

    Initial Approval

    This protocol has been approved by the following:

    Prepared By:

    Name Designation Department Signature Date

    Checked By:

    Name Designation Department Signature Date

    Final Approval:

    Name Designation Department Signature Date

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 4 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    2.0 Overview

    2.1 Objective

    To establish the methodology for the Qualification of Dry Heat Sterilizer (Fully Automatic).

    2.2 Purpose and Scope

    The purpose of this protocol is to establish Documentary evidence that Dry Heat Sterilizer

    (Fully Automatic) has been qualified to perform Dry Heat Sterilization.

    This Protocol is Applicable for the Qualification of the DRY Heat Sterilizer (Fully Automatic)

    which is installed in the Autoclave and DHS loading room in Microbiology Lab.

    2.3 Responsibility

    • Protocol / Report Preparation: Microbiology Executive

    • Execution of Qualification Activity: Microbiology Executive/ QA Executive

    • Protocol / Report checking : Manager QC / Manager QA

    • Approval of Protocol / Report: Quality Control Manager / QA Manager and above

    2.4 Qualification Team

    • Microbiology Executive / Manager

    • Quality Control Manager/Head

    • Quality Assurance Executive / Manager

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 5 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    3.0 Training Record

    3.1 Purpose

    The purpose of the training is to familiarize the trainees with the overall strategy of

    performance qualification of Dry Heat Sterilizer (Equipment No.:………).

    3.2 Scope

    This Training is applicable to the Dry Heat Sterilizer (Fully Automatic).

    3.3 Topics

    The following topics shall be covered during training:

    • Overall strategy of qualification process.

    • General precautions/guidelines to be followed during qualification.

    • Training records shall be attached with the report as Annexure – 01.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 6 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    4.0 Qualification Requirements

    Following instruments shall be calibrated at the time of the Re-Qualification.

    S.No. Instrument Name Instrument

    Code/Sr. No.

    Calibration

    Certificate No.

    Calibration

    Due on

    1. Inbuilt RTD sensors

    2. Inbuilt RTD sensors

    3. Inbuilt RTD sensors

    4. Inbuilt RTD sensors

    5. Inbuilt RTD sensors

    6. Inbuilt RTD sensors

    7. Inbuilt RTD sensors

    8. Inbuilt RTD sensors

    9. Temperature indicator

    controller of panel

    10. External Data Logger

    11. RTD Sensor of Data Logger

    12. RTD Sensor of Data Logger

    13. RTD Sensor of Data Logger

    14. RTD Sensor of Data Logger

    15. RTD Sensor of Data Logger

    16. RTD Sensor of Data Logger

    17. RTD Sensor of Data Logger

    18. RTD Sensor of Data Logger

    19. RTD Sensor of Data Logger

    20. RTD Sensor of Data Logger

    21. RTD Sensor of Data Logger

    22. RTD Sensor of Data Logger

    23. RTD Sensor of Data Logger

    24. RTD Sensor of Data Logger

    25. RTD Sensor of Data Logger

    26. RTD Sensor of Data Logger

    27. Aerosol Photometer

    28. Non Viable particle counter

    Calibration Certificate shall be attached as Annexure-06.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 7 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    5.0 System / Equipment Description

    5.1 System / Equipment details

    The Dry Heat Sterilizer (Fully Automatic) Equipment ID No.:................ shall be used for

    the Depyrogenation of BET tubes and Sterilization of Other glassware in Microbiology Lab.

    5.2 System /Equipment Identification/Description

    Component Specifications

    Name of equipment Dry heat Sterilizer (Fully Automatic)

    Equipment ID No -----

    Serial Number -----

    Name of the Manufacturer Fabwell Engineers

    Dimensions 600 (W) X 900 (H) X 900 (D) mm

    Working Temperature Up to 2800C

    Chamber Volume 0.486m3

    Doors Hinged

    Details of Purchase Order No. ------

    Equipment Location New Microbiology laboratory

    5.3 Generic Design

    5.3.1 Working Principle

    The Dry Heat Sterilizer (Fully Automatic) can be efficiently used to perform Dry Heat

    Sterilization and Depyrogenation processes: It is base on following two principles.

    5.3.1.1 Heat Conduction:

    Conduction is accomplished either by a molecular interaction from higher energy level to a

    lower energy level or by free electrons. Thus, the ability of solids to conduct heat varies

    directly with the free electron concentration. Pure metals are best conductors and non- metals

    are the poorest conductor of heat.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 8 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    5.3.1.2 Radiant Heating Process:

    Radiant heating is the process in which energy flows from high temperature body to a lower

    temperature. The geometry of both source and the exposure section of the unit will affect

    the uniformity of the radiation density.

    5.3.2 Brief Machine Description

    5.3.2.1 Chamber & Outer Cover

    The outer body of the DHS is made of Stainless steel and inside chamber is also made out of

    stainless steel sheet. The inside of the DHS is tapered to get uniform air circulation.

    Portion between the two walls is having a mild steel framework. Top sheet is of Stainless

    Steel for mounting re-circulation fan motor, Dampers and also for junction box.

    5.3.2.2 Validation Port

    The unit is provided with a 1 Validation port for the insertion of flexible RTD, PT-100

    sensors. The Port end connections is Flange type with gasket for ease of insertion

    5.3.2.3 Insulation

    The chamber is insulated with fibre-glass. The insulation is held in place by a stainless steel

    cover. This ensures less heat transfer to outside & also safety of handling.

    5.3.2.4 Door

    The equipment is having Two Stainless Steel Doors made. S.S. doors will also be insulated

    with fibre glass with and covered with S.S. on the outer side. Doors are rugged and hinged to

    external wall of Sterilizer frame-work for ease of opening.

    5.3.2.5 Door Components

    Other Door components such as Radial Locking Arms, Hinge Plate etc. are also of Stainless

    Steel.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 9 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    5.3.2.6 Door Gaskets

    The door gasket is of Silicone rubber (square cross section) fitted in the Groove provided on

    the door.

    5.3.2.7 Heating System

    A heater of S.S. has been provided inside the chamber. Heaters have been provided in two

    zones. Each Zone is controlled separately.

    5.3.2.8 Air Circulation

    The air inside the sterilizer has been provided Hot Air Circulation arrangement consists of a

    blower fan and motor.

    5.3.2.9 Blower Fan

    Blower fan is made of S.S. 316 & is dynamically balanced. Suitable Silicon gasket has been

    provided for motor and motor mounting plate.

    5.3.2.10 Motor

    It has been supplied with TEFC Motor of 3 HP, 1440 rpm, operable on 3 phase, 415 Volts,

    50 cycles.

    5.3.2.11 Filter Module

    The DHS is provided with pressure module mounted on to the sterilizer which maintains

    positive pressure inside to ensure that no outside air enters the sterilizer. It consists of

    Pre-filter of 5 micron. & HEPA filter of 0.3 micron.

    5.3.2.12 Exhaust Duct

    The sterilizer is provided with exhaust duct with HEPA filter necessary Valves etc. to

    exhaust hot air. The exhaust air can be taken out of the room by means of suitable duct

    (clients supply).

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 10 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    5.3.2.13 Exhaust damper

    An Exhaust damper is provided at the top of the unit. The damper during the heat-up phase

    opens to ensure complete removal of moisture and during the cooling phase.

    5.4 Safety Feature Description

    5.4.1 Alarms and Warnings

    The DHS shall be equipped with several alarms and warnings which alert the operator (or the

    User’s main control system) that system operation is outside of established parameters. Fatal

    alarms switch the system into the shutdown sequence. Non-fatal warnings allow the system to

    continue to run, but it will display “warning” messages and sound the alarm horn.

    Alarms and warnings are separated into “fatal alarms” and “non-fatal warnings” based on the

    Following criteria:

    • Fatal alarms shut the system down when there is a reasonable risk of damage to the

    system or of contamination of, or damage to, the user’s distribution loop or final product or a

    reasonable risk of injury to personnel.

    • Non-fatal warnings do not shut the system down, as there is little or no risk of damage

    to the system or of contamination of, or damage to, the user’s distribution loop or final

    product and little or no risk of injury to personnel.

    5.4.1.1 Fatal Alarms

    An alarm horn shall sound, and alarm message shall be displayed, and operation shall be

    determinate.

    When a fatal alarm occurs. Fatal alarms shall include:

    • Door Open.

    • Chamber high-temperature alarm

    • Chamber low-temperature warning.

    5.4.1.2 Non-Fatal Warnings

    An alarm horn shall sound and a warning message shall be displayed when a non-fatal warning

    occurs, but operation shall continue. Non-fatal warnings shall include:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 11 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    • Compressed air low-pressure warning

    • Other Side Door Open warning

    • Incorrect password entry warning

    6.0 Qualification Procedure

    The following procedure shall be used for the performance qualification of the Dry heat

    Sterilizer Fully Automatic.

    All the Calibration certificates shall be verified as per Exhibit-01 and the calibration

    certificate shall be attached as Annexure-06. The Calibration and Temperature Verification

    Checklist shall be as per Exhibit -E02

    The following procedure shall be used for the Re-qualification of the Dry Heat Sterilization.

    Air Velocity Test

    HEPA Filter Integrity Test (PAO Test)

    Air Flow Pattern Test

    Heat Distribution Study

    Heat Penetration Study

    Endotoxin Challenge Study

    Non Viable Air Borne Particle Count.

    6.1 Air Velocity Test

    6.1.1 Objective:

    To demonstrate that the system is capable of delivering air velocities, as per the requirement, to

    maintain continuous laminarity inside the Dry Heat Sterilizer.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 12 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    6.1.2 Procedure:

    • Operate the Dry Heat Sterilizer as Per SOP, “Operation and Maintenance of Dry Heat

    Sterilizer.

    • Ensure that the blower is switch “ON” at least 20 minutes prior to the start of the

    observations and keep the heaters bank “OFF”.

    • Measure the air velocity at Distance of 6 inches from the filter face of each HEPA filter, at

    5 locations (four corners and centre) with the Anemometer.

    • Calculate the Average Velocity (V, in feet per minute) of each HEPA filter/ riser as:

    Where, i = 1 to 5 and v = Velocity observed at each point.

    Acceptance Criteria

    The Average air velocity shall be within 90 ± 20% FPM.

    Rationale for Acceptance criteria:

    For Circulation of Temperature more air Flow required, hence more Air Required to Maintain

    A Grade area.

    The observations shall be recorded as per Exhibit – E10

    6.2 HEPA Filter Integrity Testing (PAO Test)

    1 2

    5

    3 4 4 3

    2 1

    5

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 13 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    6.2.1 Objective

    HEPA filter leak test shall be checked in order to find out that filters are not damaged and there

    is no leakage from the filter media joints and frames etc.

    PAO test shall be carried out as per test procedure stated below for leakage of HEPA filters. The

    observations shall be recorded as per Exhibit–E08 & Test reports shall be attached as

    Annexure-05.

    6.2.2 Test Procedure:

    • Filter integrity testing shall be checked using cold PAO (Poly alpha olfein) smoke generator

    aerosol photometer

    • Aerosol shall be introduced at the upstream side of HEPA filter, such that it gets the actual

    concentration of 10 μg/L to 30 μg/L

    • Actual upstream concentration shall be set as 100 %.

    • The downstream of the filter shall be scanned with the probe at the speed of 10 FPM.

    6.2.3 Acceptance Criteria:

    Penetration shall not exceed 0.01% at 100 % upstream concentration.

    6.3 Air Flow Pattern Test:

    Objective:

    To verify the unidirectional parallel airflow and non-turbulent up to working zone & movement

    of air flow due the pressure gradient throughout the chamber maintaining the Class 100

    environmental conditions.

    6.3.1 Equipment / Instruments and Chemical Used:

    Water based Smoke Generator

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 14 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    6.3.2 Method applied:

    • The temperature of chamber of the Dry Heat Sterilizer should not be more than 5°C above

    the ambient temperature during the study.

    • Ensure that blower of the Dry Heat Sterilizer is running and heater is OFF.

    • Verify and record the path of smoke as per acceptance criteria.

    • Above procedure can also be performed by using water fogger.

    Acceptance criteria:

    The smoke flow should be unidirectional from the HEPA filter face to return air riser.

    The observations shall be recorded as per Exhibit – E11.

    6.4 Heat Distribution Study

    6.4.1 Objective

    The Dry Heat Sterilizer when operated with Empty Chamber is capable of producing the

    temperature profiles as per the temperature set points in the equipment and the temperature

    distribution is uniform throughout the sterilization period when the chamber is empty.

    6.4.2 Procedure

    6.4.2.1 Heat distribution study of the Dry heat sterilizer shall be done at temperature 250 0C for not less

    than 30 min using 16 flexible RTDs placed inside the chamber. Three runs of Empty cycle shall

    be run at 250 0C for 30 min.

    6.4.2.2 None of the RTD shall touch the body of the chamber of the sterilizer while doing RQ study.

    6.4.2.3 Drawings showing location of the flexible RTDs in the chamber for heat distribution study shall

    be as per Annexure-02.

    6.4.2.4 Detail of each cycle run shall be recorded as per Exhibit-E03 and print outs shall be attached as

    Annexure -07.

    6.4.3 Acceptance criteria:

    There should be uniform distribution of temperature NLT 250°C at each probe in sterilizer

    chamber during sterilization hold period. Upper targeted limit for temperature will be 270°C.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 15 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Based on validation study results for upper limit will be finalized in report.

    6.5 Heat Penetration Study

    6.5.1 Objective

    Purpose of heat penetration studies is to check the penetration of heat into the loaded articles

    when the Dry Heat Sterilizer when operated with loaded chamber is capable of producing the

    temperature profiles as per the temperature set points in the equipment and the temperature

    distribution is uniform throughout the sterilization period.

    6.5.2 Procedure

    6.5.2.1 Heat penetration study shall be done using not less than 16 flexible RTDs with the help of data

    logger at 250 0C for not less than 30 min.

    6.5.2.2 Three runs of sterilization cycle shall be run with loaded chamber to perform the heat

    Penetration study.

    Three runs of sterilization cycle with Minimum Load.

    Three runs of sterilization cycle with Maximum Load.

    6.5.2.3 Each probe shall be inserted inside the articles to be sterilized and one endotoxin challenge vial

    Containing 100000 EU/vial shall be placed along with each sensor.

    6.5.2.4 Detail of the challenge vial recorded as per Exhibit-E05.

    6.5.2.4 Five Biological Indicators strips containing Bacillus atrophaeous ATCC No. 9372 having

    population count not less than106 CFU shall also be placed in each cycle inside the article which

    are kept in the coolest part of the chamber determined during Heat distribution study.

    6.5.2.5 Details of each cycle run has been recorded as per Exhibit-E03.

    6.5.2.6 The observation of Heat Penetration Study have been recorded as per Exhibit E04.

    6.5.2.7 Details of the loads shall be as per Annexure-03.

    6.5.2.8 Drawings showing location of the flexible RTDs in the chamber for heat penetration study

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 16 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    shall be as per Annexure-04.

    6.5.2.9 At the end of the cycle examined the measured temperature and noted the following:

    The parts of the usable chamber space that are the fastest and the slowest to attain the

    sterilization temperature.

    The parts of the usable chamber space that are the Hottest and the coolest during the

    sterilization holding time.

    6.5.2.10 FD at a temperature other than 250°c is the time required to provide the lethality equivalent to

    that provided at 250°c for a started time.

    FD shall be calculated as follows

    Where

    ∆T = Time interval between two successive measurements.

    TA = Actual temperature at that point.

    TB = Base temperature i.e. 121.1oC.

    Z = 45(Change in temperature required to alter the D value by a factor of 10 i.e.1 log

    Reduction).

    6.5.2.11 Calculated the SAL values as per following formula and record the data as per Exhibit-E04.

    FD

    SAL = ------

    D250

    Where D250 is the time required to reduce the microbial population by 90% i.e. 1 log reduction.

    6.5.2.12 D250 shall be calculated as follows:

    TA-TB

    ------------

    Z

    D250 = D160 x 10

    D250 is the D value at temperature 250°c.

    D160 is the D value at temperature 160°c (Certified value).

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 17 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    TA is 160°c

    TB is 250°c

    Z is the z value at 160°c.

    6.5.3 CALCULATION OF FH VALUE

    Definition of FH value (Lethal Rate / accumulated lethality):

    The FH value is a "unit of lethality" and is defined as the number of minutes equivalent of

    depyrogenation at reference temperature (250° C), calculated using a know Z value.

    FH = dt x 10 (T-250) / Z

    Where

    dt: Interval of temperature monitoring in a certain location i.e. 01 minute.

    T: Temperature at a certain time in certain location during depyrogenation hold

    Z: Z Value which indicates relationship of lethality to temperature. (45ºC)

    6.5.4 Acceptance criteria:

    There should be temperature NLT 250°C at each probe in sterilizer chamber during

    sterilization hold period.

    Upper targeted limit for temperature will be 270°C. Based on validation study results for

    upper limit will be finalized in report.

    Minimum of 3 log reduction shall be achieved using Endotoxin challenge vials.

    Minimum of 12 log reduction shall be achieved using Biological Indicators.

    6.6 Endotoxin Challenge Study

    Endotoxin Challenge Study is the part of Heat penetration study and shall be done for the set

    load by placing endotoxin challenge vial next to each temperature sensor.

    6.6.1 Take an Endotoxin challenge vial and prepare the Endotoxin dilution containing

    100,000EU/ml and add 0.1 ml of 100000 EU/ml in each vial.

    6.6.2 Take 18 No. of de-pyrogenated vials for each cycle to be challenged for depyrogenation.

    6.6.3 Air dry the vials, seals them and mark them the respective location and cycle no.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 18 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    6.6.4 Place 16 vials containing 0.1ml of 10,000EU/ml along with each sensor inside the Chamber and

    process at 250°C for not less than 30 minutes.

    6.6.5 Keep two vials as such without processing them as positive control.

    6.6.6 The Endotoxin challenge vial shall be placed along with each temperature sensor as per

    Annexure-04

    6.6.7 After processing let the vials cool to room temperature.

    6.6.8 Reconstitute each processed vial and one unprocessed vial with 1ml of LRW and vortex for 15

    min.

    6.6.9 Take 0.1ml from each processed vial in reaction tubes in duplicate and add 0.1ml of lysate

    (Sensitivity 0.125EU/ml).

    6.6.10 For performing positive control prepare a dilution of 1:10,000 containing 1EU/ml of endotoxin

    content from one unprocessed vial.

    6.6.11 Take 0.1ml of 1EU/ml in reaction tubes in duplicate and add 0.1ml of lysate.

    6.6.12 The details of the dilution preparation is as per Exhibit E06.

    6.6.13 Perform Control curve from the unprocessed challenge vial also.

    6.6.14 Incubate the Reaction tubes at 37±1°C for 60±2 minutes.

    6.6.15 Details of the control curve have been recorded as per Exhibit E07.

    6.6.15 Calculate the minimum log reduction using following formula

    Minimum log reduction= log Endotoxin concentration of the unprocessed control vials- log

    Endotoxin concentration of the processed vials

    Where Endotoxin concentration = Lysate sensitivity x reconstituted volume x dilution

    6.6.16 Observations:

    - A positive reaction indicates Endotoxin content greater than the calculated Endotoxin value.

    - A negative reaction indicates content less than the calculated Endotoxin value.

    - A positive reaction in the 1/10,000 dilution of the unprocessed control vials indicates an initial

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 19 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Endotoxin concentration greater than 1250 EU/ml.

    6.7 Non Viable Air Borne Particle Count

    6.7.1 Objective

    To demonstrate that the Dry Heat Sterilizer meets the requirement for desired cleanliness class

    i.e. Class 100 (Grade A).

    6.7.2 Procedure

    a) This test shall be performed after cleaning of Dry Heat Sterilizer as SOP, “Operation and

    Maintenance of Dry Heat Sterilizer.

    b) Carry out the studies with Empty Chamber.

    c) The temperature of Chamber of the Dry Heat Sterilizer should not be more than 5°C above

    the ambient temperature during the study.

    d) Ensure that blower of the Dry Heat Sterilizer is running and heater is OFF.

    e) Perform the operation of Non-Viable Particle Counter at Centre (Triplicate Sampling) of the

    dry heat sterilizer.

    Acceptance Criteria:

    The number of particles shall not increase more than the limits mentioned below.

    • The observations of nonviable particle count shall be recorded as per Exhibit-E09 and test

    reports shall be attached in Annexure-05.

    6.8 Any deviation observed during performance qualification shall be recorded in the observed

    deviation, corrective action and justification report section.

    6.9 Observed deviation shall be reported to the department head and quality head.

    6.10 If the observed deviation does not have any major impact on the qualification the final conclusion

    shall be provided.

    Class Max. No. of permitted particles Per cubic Meter (At Rest)

    ≥ 0.5 µm ≥ 5.0 µm

    100 3520 29

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 20 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    6.11 If the observed deviation has major impact on the qualification, deviation shall be reported to the

    manufacturer for the corrective action and qualification activity shall be performed again.

    7.0 Acceptance Criteria

    Performance qualification shall be considered acceptable when requirements listed in section 6.0 of

    this protocol has been fulfilled and all the components of Dry Heat Sterilizer, Fully Automatic

    are performing as per our requirement and as per manufacturer recommendations.

    8.0 Qualification Report

    The performance qualification report shall consist of a summary document, in narrative form,

    which shall briefly describe the activity performed along with the observations recorded in

    relevant exhibits.

    This report shall also include the related documents and attachments/annexure which were

    completed at the time of qualification activity.

    9.0 Approval of Qualification Report

    The report shall be evaluated and proper references / conclusions / recommendations shall be

    recorded by quality assurance.

    The performance qualification report shall be evaluated and finally approved by quality

    assurance.

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 21 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    10.0 Observed Deviation

    S.No. Page No. Point

    No.

    Observed Deviation Deviation

    Reported

    By

    Deviation

    Approved By

    Corrective Action Taken Justification of Corrective

    Action

    Corrective action taken

    and justification given by

    Report Approved By

    Department Head Quality Head

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 22 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    11.0 List of Exhibits/Annexure

    11.1 List of Exhibits

    Exhibit No. Exhibit Title No. of Pages

    E01 Calibration Certificate Availability Verification Checklist

    E02 Calibration and Temperature Verification Checklist

    E03 Heat Distribution Study

    E04 Heat Penetration Study

    E05 Details of Endotoxin Challenge Vial

    E06 Dilution Preparation for Endotoxin Challenge Study

    E07 Detail of Control curve of Endotoxin Challenge Vial

    E08 HEPA Filter Integrity Testing

    E09 Non-viable Air Borne Particle Count

    E10 Air velocity Test

    E11 Air Flow Pattern Test

    Total No. of Pages

    11.2 List of Annexure

    Annexure No. Annexure Title No. of Pages

    01 Training Record

    02 Drawing of Locations of Temperature Probes in the

    Chamber for Heat Distribution Study

    03 Details of the Sterilization Loads

    04 Drawing of Locations of Temperature Probes in the

    Chamber for Heat Penetration Study

    05 Test Reports

    06 Calibration Certificates

    07 Print outs of the Data

    Total No. of Pages

    12.0 Reference Documents

    ➢ ISO 14644 of Clean Rooms and Associated Controlled Environments.

    ➢ Validation Master Plan.

    ➢ Schedule - M – “Good Manufacturing Practices and Requirements of Premises, Plant and

    Equipment for Pharmaceutical Products.”

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit – E01

    Calibration Certificate Availability Verification Checklist

    Equipment Name / Description : Dry Heat Sterilizer, Fully Automatic

    Tag No . : QB/DHS/001

    Location : Microbiology Lab (Ground floor)

    S.No. Document

    Description

    Specifications Actual Observations

    1. Calibration Certificates of all the

    instruments attached

    All the calibration

    certificates should

    available

    Remarks: Observations Comply / Not Comply with the specifications.

    Checked By:____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By:____________________ _________ ________

    (Name) (Sign) (Date)

    Exhibit – E02

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Calibration and Temperature Verification Checklist

    Location :

    Set Temperature:

    Pre calibration/Post Calibration study started at:

    Probe Description Temperature Observed (ºC)

    Probe 1

    Probe 2

    Probe 3

    Probe 4

    Probe 5

    Probe 6

    Probe 7

    Probe 8

    Probe 9

    Probe 10

    Probe 11

    Probe 12

    Probe 13

    Probe 14

    Probe 15

    Probe 16

    Acceptance Criteria – Temperature difference among probe to probe shall not be more than

    ±0.5ºC.

    Remarks: _______________________________________________________________

    Checked By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By: ____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit – E03

    Heat Distribution Study

    Date of test

    Cycle No.

    Type of load

    Set sterilization temperature

    Set sterilization period

    Process cycle started at Hrs.

    Dwell period started at Hrs.

    Dwell period ended at Hrs.

    Temperature profile

    Temp. Sterilization temperature of different sensors

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

    Max.

    Min.

    Avg.

    Endotoxin

    reaction of

    vials

    Observed

    log

    reduction

    +: Denote gel clot formation -: Denote no gel clot formation

    Acceptance criteria:

    -There should be uniform distribution of temperature NLT 250°C at each probe in sterilizer chamber

    during sterilization hold period. Upper targeted limit for temperature will be 270°C.

    -Minimum 3 log reduction shall be achieved at each location while doing Endotoxin challenge study.

    Remarks: Dry heat sterilizer complies/does not comply with the above acceptance criteria.

    Checked By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By: ____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit – E04

    Heat Penetration Study

    Date of test

    Cycle No.

    Type of load

    Set sterilization temperature

    Set sterilization period

    Process cycle started at Hrs.

    Dwell period started at Hrs.

    Dwell period ended at Hrs.

    Temperature profile

    Temp. Sterilization temperature of different sensors

    1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16

    FD Value

    SAL

    Endotoxin

    reaction

    of vials

    Endotoxin

    log

    reduction

    +: Denote gel clot formation -: Denote no gel clot formation

    Acceptance criteria:

    -There should be temperature NLT 250°C at each probe in sterilizer chamber during sterilization hold

    period. Upper targeted limit for temperature will be 270°C.

    - Minimum of 3 log reduction shall be achieved using Endotoxin challenge vials.

    - Minimum of 12 log reduction shall be achieved using Biological Indicators.

    Remarks: Dry heat sterilizer complies/ does not comply with the above acceptance criteria.

    Checked By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By: ____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 3 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit – E05

    Details of Endotoxin Challenge Vials

    Make

    Catalogue No.

    Lot No.

    Potency

    Volume to be reconstituted

    Expiry date

    Checked By:____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By:____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit-E06

    Dilution Preparation for Endotoxin Challenge Study

    LAL Reagent Details

    Manufacturer : Lot No. : Expiry Date :

    Sensitivity :

    LAL Water Details

    Lot No. : Expiry Date : Date of

    Opening :

    Dilution Preparation:

    Checked By:____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By:____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit-E07

    Detail of Control curve of Endotoxin Challenge Vial

    Batch No/Lot No of LAL : ______________________ Date of Test : ________________

    Incubation Temperature: 37 ± 1ºC Incubation Period: 60 ± 2 min.

    LAL reagent details

    Manufacturer: Expiry date:

    Sensitivity: Reconstituted on: Use before:

    LAL regent water details

    Lot No: Expiry date: Vial opened on:

    Test Description Tube 1 Tube 2 Inference

    Blank

    Control 2λ (_______EU/ml)

    Control λ (_______EU/ml)

    Control λ/2 (_______EU/ml)

    Control λ/4 (_______EU/ml)

    Geometric mean( GM):

    ∑ e ∑ e= Summation of all endpoints

    GM= antilog ------------- f= number of replicates

    f

    =

    =

    Remarks: Sensitivity of LAL reagent is within/not within ± two fold dilution.

    Acceptance criteria: LAL regent qualifies if GM confirms the label claim sensitivity of the reagent

    within ± two fold dilution.

    +: Denote gel clot formation -: Denote no gel clot formation

    Checked By:____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By:____________________ _________ ________

    (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Exhibit – E08

    HEPA Filter Integrity Testing

    Equipment ID No._____________________ Date of test____________________________

    Name of the Instrument used for test____________________________________________________

    Make_______________________________ Model________________________________

    Date of calibration____________________ Due date of calibration___________________

    Sr. No._____________________________ Concentration of PAO upstream____________

    Filter No. Maximum Downstream

    Concentration of PAO observed Remarks

    Acceptance criteria: NMT 0.01% Poly alpha Olefin shall penetrate across the HEPA filter and no

    leakage shall be observed from the fitment.

    Remarks: ________________________________________________________________________

    _________________________________________________________________________________

    Checked By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Exhibit – E09

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Non Viable Air Borne Particle Count

    Date of test__________________________________ Filter No.________________________________

    No. of locations to be sampled__________________ Witnessed By_____________________________

    Instrument used for test________________________ Make___________________________________

    ___ Model______________________________________ Sr. No___________________________________

    Date of calibration____________________________ Due date of calibration______________________

    Volume of air sampled ________________________ Time for which air sampled__________________

    Location Particle count 0.5μ per m3 Particle count 5.0μ per m3

    1 2 3 Avg. 1 2 3 Avg.

    L1

    L2

    L3

    L4

    L5

    Acceptance criteria:

    Class100: Particles ≥ 0.5 μm =NMT 3520 , Particles ≥ 5.0 μm = NMT 29

    Conclusion: Air sampled complies/does not comply with as per Acceptance criteria

    Checked By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Verified By: ____________________ _________ ________

    (Name) (Sign) (Date)

    Exhibit-E10

    Air velocity

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 3 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Date of test__________________________

    Name of the Instrument used for test______________________________________________________

    Make_______________________________ Model___________________________________

    Date of calibration____________________ Due date of calibration______________________

    Sr. No._____________________________

    Location Air velocity (FPM)

    V1

    V2

    V3

    V4

    V5

    Average in FPM

    Witnessed by

    Acceptance Criteria: The Average air velocity shall be within 90 ± 20% FPM.

    Remarks: Results are within/ not within limit.

    Reviewed By:____________________ _________ ________

    (QC) (Name) (Sign) (Date)

    Verified By:____________________ _________ ________

    (QA) (Name) (Sign) (Date)

    Exhibit- E11

    AIR FLOW PATTERN TEST

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 4 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Instrument

    Name

    Date of Test

    Make Instrument ID.

    Date HEPA filter Location Acceptance Criteria Observation

    Observed

    By

    QA

    Sign/Date

    Right Hand Side HEPA

    Filter

    Fumes should follow the path from

    supply air filter face to the grill of

    return air riser under the chamber.

    Left Hand Side HEPA

    Filter

    Acceptance criteria: The smoke flow should be unidirectional from the HEPA filter face to return air

    riser.

    Remarks: Air flow direction Comply/does not comply as per the acceptance criteria.

    Checked By: ________________________ ___________ ____________

    (QA) (Name) (Sign) (Date)

    Verified By: ________________________ ___________ ____________

    (QA) (Name) (Sign) (Date)

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 01

    Training Record

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 02

    Drawing Of Locations of Temperature Probes in the Chamber

    for Heat Distribution Study

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 1 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 03

    Details of the Sterilization Loads

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 2 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 04

    Drawing of Locations of Temperature Probes in the Chamber for Heat

    Penetration Study

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 3 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 05

    Test Reports

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 4 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 06

    Calibration Certificates

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages:

  • PROTOCOL No. PERFORMANCE QUALIFICATION

    PROTOCOL

    FOR

    DRY HEAT STERILIZER

    SUPERSEDES NIL

    EFFECTIVE DATE

    PAGE No. Page 5 of 40

    PHARMA SCHOLARS MICROBIOLOGY DEPARTMENT

    Annexure – 07

    Print Outs of the Data

    Equipment Name: Dry Heat Sterilizer ( Fully Automatic)

    Equipment Tag No.:

    Location: New Microbiology Lab

    No. of Pages: