Proportional Counters

Embed Size (px)

Citation preview

  • 7/27/2019 Proportional Counters

    1/13

    Ionisation chamber

  • 7/27/2019 Proportional Counters

    2/13

    Air-filled ion chambers are used in portablesurvey meters, for performing QA testing ofdiagnostic and therapeutic x-ray machines,and are the detectors in most x-ray machinephototimers

  • 7/27/2019 Proportional Counters

    3/13

    GM COUNTERS

  • 7/27/2019 Proportional Counters

    4/13

  • 7/27/2019 Proportional Counters

    5/13

    appl

    Often used for inexpensive survey meters

    In general, GM survey meters are inefficientdetectors of x-rays and gamma rays

    GM detectors suffer from extremely long deadtimes seldom used when accurate measurementsare required of count rates greater than a fewhundred counts per second

    Portable GM survey meter may become paralyzedin a very high radiation field should always useionization chamber instruments for measuring such

    fields

  • 7/27/2019 Proportional Counters

    6/13

  • 7/27/2019 Proportional Counters

    7/13

    Proportional counters

    Operates at voltages beyond ionisation chamber

    region ,but below that of geiger region. o/p pulse from the chamber starts to increase

    with the increase in field strength at the central

    electrode. The pulse is amplified by 1000 or more. o/p

    pulse is of few mv.

    Design of the counter and applied voltages are sochosen that a high voltage exists close to the

    anode.

    Preamp is a voltage follower- high i/p imp

  • 7/27/2019 Proportional Counters

    8/13

    Followed by low noise linear amp having gain in the

    range of 500-1000. non overloading type

    Preamp shd be placed near the detector. Pulse height analyser-counts a pulse only if its amp

    falls within specific limits.

    Scaling unit counts down the pulses so that they aredigitally displayed.

    App:

    Commonly used in standards laboratories, healthphysics laboratories, and for physics research

    Seldom used in medical centers

  • 7/27/2019 Proportional Counters

    9/13

    Scintillation counter Scintillator is a crystalline sub which produces minute

    flashes of light in the vis or uv range when it absorbsionising rad.

    Light flashes are of very short duration and detected byPMT, which produces a pulse for each particle.

    Scint + pmt=scintillation counter

    NaI with 0.5% of thallium iodide- r ray detector.

    Anthracene beta

    Scintillators are used in conventional film-screenradiography, many digital radiographic receptors,fluoroscopy, scintillation cameras, most CT scanners,

    and PET scanners

  • 7/27/2019 Proportional Counters

    10/13

  • 7/27/2019 Proportional Counters

    11/13

    PMTs perform two functions:

    Conversion of ultraviolet and visible light photons

    into an electrical signal

    Signal amplification, on the order of millions to

    billions

    Consists of an evacuated glass tube containing a

    photocathode, typically 10 to 12 electrodes called

    dynodes, and an anode

  • 7/27/2019 Proportional Counters

    12/13

  • 7/27/2019 Proportional Counters

    13/13

    Semiconductor detectors

    Silicon & Germanium crystalsalpha and beta

    Func in a similar manner to that of the gas

    ionisation chamber.