Project 1:

ASTM E647:Standard Test Method of Measurement of

Fatigue Crack Growth Rate

MSE 527L: Mechanical Behavior of Materials Lab10-21-2015

Group 2:Rna WahebRameen HassanzadehRyan OhPavan Kumar NanneSiddhesh SawantDhaval Prajapati

AbstractASTM E647 is a standard test method used for the measurement of fatigue crack growth

rate. The test method involves cyclic loading of notched specimens which have been pre-cracked in fatigue. Crack length is measured, either visually or by an equivalent method, as a function of elapsed fatigue cycles and the data are subjected to numerical analysis to establish the rate of crack growth. Results are expressed in terms of the crack-tip stress-intensity factor range (ΔK).1

IntroductionMaterial tests are used to analyze the strength of a material, which is important in the

selection of materials when designing. When a material undergoes repeated applied loads, it

Page 2: Project1

weakens. This weakening, fatigue, is the progressive and localized structural damage that occurs when a material is subjected to cyclic loading.2 When the loads exceed a certain threshold, microscopic cracks begin to form. When these cracks reach a critical size, they will propagate suddenly and crack. The ASTM E647 test standard is used to help to understand a material's behavior by measuring the fatigue crack growth rate.

The fatigue growth rate is measured from threshold to Kmax, the maximum stress intensity factor. A specimen is prepared based on three possible configuration: compact tension C(T), middle-cracked tension M(T), and eccentrically-loaded single edge tension ESE(T). The specimen is then precracked and tested with a force determined by dA/dN. The crack size is measured as a function of elapsed cycles.

ProcedureSpecimen Preparation Compact tension specimen, C(T), is single edge-notch specimen designed for loads in tension. It requires the least amount of material to evaluate crack behavior. Figure 1 shows the required geometry for a standard C(T) specimen and figure 2 shows the notch details for the specimens. The compact tension specimen would require a clevis and pin assembly at the top and bottom of the specimen to allow in-plane rotation as the specimen is loaded.

.Figure 1. Geometry of Standard C(T) Specimen

Page 3: Project1

Figure 2. Notch DetailsGrips and fixturing for the middle tension specimen M(T) depends on the width. Figure 3 shows the required geometry for a standard M(T) specimen.

Figure 3. Geometry of Standard M(T) Specimen

Page 4: Project1

For tension-tension loading of specimens with W ≤75 mm, a clevis and single pin arrangement is suitable for gripping provided that the specimen gage length is at least 3W. The gage should be at least 1.7W. For tension-tension loading of specimens with W ≥ 75 mm a clevis with multiple bolts is recommended and the minimum specimen gage length should be 1.5W.The specimen may also be gripped using a clamping device instead. This type of gripping is necessary for tension-compression loading. The minimum gage length requirement for clamped specimens is 1.2W.The eccentrically-loaded single edge crack tension specimen ESE(T) is similar to C(T) in its apparatus. Figure 4 shows the required geometry for a standard eccentrically-loaded single edge crack tension specimen.

Figure 4. Geometry of Standard ESE(T) SpecimenGood alignment is important because misalignment can cause non-symmetric cracking,

which may lead to invalid data. However, if it were to occur a strain-gaged specimen would be useful in identifying and minimizing misalignment.

Samples do not always have to come from material with complete stress relief. The residual stress can be minimized through careful selection of specimen shape and size. Symmetrical specimens can also can minimize residual stress.

Page 5: Project1

In order for results to be valid, the specimen must be predominantly elastic at all values of applied force. There is an alternative size requirement for high-strain hardening materials. This is done by defining replacing the yield strength with effective yield strength or flow strength.

Using this alternative size requirement means that plastic deflections would occur in the specimen, which could essentially double the growth rate.

The machined notch may be made by electrical-discharge machining, milling, broaching, or sawcutting.

Table 1. Summary of Notch Preparation

EDM Mill/Broach Grind Sawcut

Notch Root Radius

<0.25 mm <0.075 mm <0.25 mm <0.25 mm

Materials High-strength steels(σ≥ 1175 Mpa),

Titanium, Aluminum

Alloys

Low/medium-strength steels(σ ≤ 1175 Mpa),

Aluminum Alloys

Low/medium-strength steels(σ ≤ 1175 Mpa)

Aluminum Only

If residual stresses are suspected of being present, the distance between two hardness indentations at the mouth of the notch should be measured using a mechanical gage. Data shows that even mechanical displacement change by more than 0.05 mm can significantly change the fatigue crack growth rates.

Once the specimen is prepped, the dimensions are measured to ensure it is within the tolerances of its specific size. Next, it is important that it is precracked to ensure that: a) the effect of the machined starter notch is removed from the specimen K-calibration, and 2) the effects on subsequent crack growth rate data caused by changing crack front shape or precrack load history is eliminated.

To test for fatigue crack growth rates above 10−8 m/cycle, it is preferred that eachspecimen be tested at a constant force range (ΔP) and a fixed set of loading variables (stress ratio and frequency). If force range is incrementally varied it should be done so that Pmax is increased rather than decreased to avoid delaying of growth rates caused by overload effects.

Page 6: Project1

To test for fatigue crack growth rates below 10−8 m/cycle, start cycling at a ΔK and Kmax level equal or greater to the terminal precracking values. Subsequently, forces are decreased as the crack grows and test data are recorded until the lowest ΔK or crack growth rate of interest is achieved.

Make fatigue crack size measurements as a function of elapsed cycles by means of avisual or equivalent technique.

Discussion

Calculations:All calculations for desired parameters focus around Paris's Law for stress concentration,

seen below.

Here, da/dN is the crack growth rate, ΔK is the range of the stress concentration factor, and C and m are proportionality constants that depend on the test methodology. The equation resolves differently depending on the test type, and can be resolved through the data taken. It is important to note that typically the regime of interest for the Paris Law is where da/dN vs ΔK is resolvable, as seen in the center section II in the figure below.

Figure: 5 Sample plot of crack growth rate vs stress concentration range.

As noted, the paris law will be solved differently depending on the test done. For example, for a center crack in an infinite sheet in tension, this resolves to the following:

Page 7: Project1

Because of this, it is very important to keep in mind your test configuration while resolving your desired parameters.

In order to solve different useful material properties, multiple methods can be used. For example, determining the crack growth rate, da/dN, the most straightforward method is to plot the crack length, a, vs the number of cycles, N. da/dN can then be resolved point-to-point, taking the slope of the curve.

However, this point-to-point method has some shortcomings. Since the curve of a vs N may be more complex. In these cases, a second order polynomial can be fit, and da/dN can be empirically solved. This calculation method is shown below, where the regression parameters are denoted by b.

While crack growth rate is valuable information for the current system, it is common the consider any material that has started to crack as failed. Because of this, we can define the parameter ΔKth as the threshold stress concentration factor. This factor is the stress concentration range for which the crack growth rate would be sufficiently small. In the case of this paper, that value is defined to be da/dN = 10-10m/cycle. This value is a commonly accepted “slow” crack rate growth.

In order to solve ΔKth, a plot of log(da/dN) vs log(ΔK) should be taken. A linear regime of this data set should be identified, ideally near the low crack growth rate regime. From this, a linear fit, ΔK at da/dN = 10-10m/cycle can be resolved.

Conclusion

Precision and Error:It is important to note that many sources of error can occur in these testing methods,

which in turn impact the solutions for da/dN, ΔK and ΔKth. To reduce this, it is important to follow the guidelines for acceptable criteria and verify the data integrity as described. However, some errors can be accounted for. For example, if the depth of the crack is shown to have curvature, that results in more than a 5% variance in the calculation for ΔK, the curvature can be accounted for by using a three-point through thickness measurement, measuring at the center of the crack front, the edge of the crack front, and between these and taking the average. Furthermore, if this curvature amount is shown to change with N, an interpolation of the averages can be used, correcting vs two crack contours separated by at least 25% of the sample width.

Page 8: Project1

Figure: Illustration showing the crack front. Curvature of this front can be corrected for.

However, it is always true that not all errors can be accounted for. Therefore, it is also important to identify the possible intrinsic variability and to determine the impact this can have on the precision of the testing. Repeated lab testing has been done in order to estimate the impact to precision due to different factors.

For instance, it has been shown that the amount of force applied under typical test methods commonly has up to a 2% error. The translates proportionately to a 2% error in K, but since da/dN can depend heavily on K, this can result in up to a 10% error.

Other sources can be more difficult to identify, so repeatability tests for variance have been done under close-to-ideal settings. Using a highly homogenous sample in order to minimize material errors, a variance average of 27% was found (ranging from 13-50%). Comparing lab-to-lab in a similar method showed an average variance of 32%.

Similarly, variance in the finding of ΔKth has been explored, resulting in a 3% variance under repeated testing in the same lab, 9% in lab-to-lab comparison. Again, do to da/dN's high dependance on ΔK, this can result in more than an order of magnitude error in da/dN.

Furthermore, one should keep in mind that these precision findings were taken while trying to minimize material error in the test samples. For most real samples, it is believed that sample-to-sample inconsistencies would dominate the error. And, because there is no established material standard for da/dN vs ΔK, it is difficult to truly determine the impact of this type of error.

Because of this, this method should only be used when looking for gross estimates in da/dN. When designing around these findings, one should take these errors into strong considerations, and design in tolerances accordingly.

Reference1. ASTM Standard E647-00, 2001 “Standard Test Method for Measurement of Fatigue

Crack Growth Rates,” ASTM International, West Conshohocken, PA, 2001.<www.astm.org>.