Professor Ioannis Pallikaris Professor Ioannis Pallikaris MD, PhDMD, PhD
Different mediums (Nd:YAG Crystals vs ArF Gas)
Less optics for the shaping of the pulse No use of toxic gasses Improved beam quality and shape Reduced thermal effect and collateral damage Hydration dependence on 193nm
Hydration dependence on 193nm
Corrosive, toxic and high cost gases
Big spot size Absorption
coefficient suggest 193nm affected more by BSS or NaCl 0.9%
Mirror 90%
Medium
Pumping
ArF
193 nm@400hzMirror 100%
Medium
Nd:YAG
1064 nm
213 nm@300Hz
Non-linear optics
All treatments performed by CustomVis Pulzar Z1 & Allegretto 400
Last follow-up to 12 months PRK Mean SEQ = - 4.50 ± 1.81D
[-6.50D to -1.50D Sph and up to –2.50D Cyl] 160 eyes in Excimer and 160 in Solid MMC solution was applied to PRK cases in
both systems according to given attempted correction
LASIK and PRK for treating Myopia and Myopic
Astigmatism. A retrospective study.
89% between + 1.00/-1.00 D at 12 months postop for Excimer
96% between +1.00/-1.00 D at 12 months postop for Solid
35%% of eyes gained more than 1 line in Solid 31% of eyes gained more than 1 line in Eximer
157 eyes
0% of eyes changed more than 0.50D between 3m and last postop
Solid
•Smooth ablation surface
•No edema or distortion of the adjacent corneal stroma
•Normal structure of the endothelial layer in both groups
S
E
S
E E
S
The Descemet’s membrane and the endothelium did not demonstrate any morphological difference with intact tissues
S
E
S
E
Solid
ExcimerS
E
No statistical significant difference found between two groups in any postoperative interval (p>0.05).
Study group, 30 eyes underwent myopic PRK with the CustomVis Pulzar Z1 laser system.
Control group, 30 eyes underwent myopic PRK with the a Allegretto Wavelight 400Hz Excimer system
Coma Y Decrease 0.04
Coma X Increase 0.05
Spherical aberration Increase 0.04
HO rms Increase 0.08
0
0.04
0.08
0.12
0.16
0.2
0.24
0.28
0.32
0.36
0.4
0.44
0.48
0.52 preop
1month
3 months
6 months
12 months
PRKN=40 eyes
Attempted SpEq: -6.45 ± 2.5 D
Pupil diameter 6mm
Alcon LADARVision4000 System Wavefront-guided Excimer Laser Ablation Using PRK Mean attSphEq : -4.21D Diopters
6months postop :Pupil diameter 6mm
RMS HO: decrease 0.29Spherical Aberration: decrease 0.06Third order RMS: increase 0.13
Daniel S. Durrie,, et al JRS 2008
WaveLight ALLEGRETTO Laser Wavefront Optimized Treatments in Surface AblationMean attSphEq : -3.64 ±1.92 Diopters
3months postop:Pupil diameter 6mm
RMS HO: increase 0.03Spherical Aberration: increase 0.01Third order RMS: increase 0.03
Aleksandar Stojanovic, et al JRS 2008
Predictability presented similar results between the two systems The excimer seems to give more
overcorrection effect due to the use of MMC compared to the Solid system
Both system are safe for correction with PRK and LASIK
Stability of refraction is the same for both systems with stability level being equal at 12 moths PostOp period
Histology has proved to have similar results to the samples of Excimer laser
There are no statistical significant changes in corneal endothelium monolayer in post PRK patients (up to one year) with solid state (213 nm) and excimer (193 nm) laser systems.
Similar changes to high order aberration were found between Solid and Excimer