115
Process Instrument Verification for Drinking Water 2014 Plant Operations Conference

Process Instrument Verification for Drinking Water Operations...Process Instrument Verification for Drinking Water 2014 Plant Operations Conference . PIV for Drinking Water ... (color

  • Upload
    others

  • View
    6

  • Download
    0

Embed Size (px)

Citation preview

  • Process Instrument Verification for Drinking Water

    2014 Plant Operations Conference

    http://www.vaawwa.org/�

  • PIV for Drinking Water

    • Overview of Operation, Calibration, Verification • Chlorine Analyzer • Turbidimeter • pH Sensor • Hands-On Lab

  • Operation, Calibration, and Performance Verification of

    CL17 Chlorine Analyzer and 1720 Series Turbidimeters

  • Take-Home Messages

    • Do NOT calibrate the CL17 analyzer unless required by regulatory agencies. – DO verify performance of unit regularly.

    • DO calibrate 1720 Series turbidimeters at

    least quarterly. – DO verify performance regularly.

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric) 1720 Turbidimeters • Calibration and Verification Strategies • Hands-On Lab

  • Process Monitoring

    • Key to cost control • Key to regulatory compliance • Key to process control

  • Laboratory vs On-line Monitoring

    • Laboratory monitoring consists of collecting grab or composite samples and then analyzing them in the laboratory. – Time and labor intensive – Variations in technique at time of sampling and/or

    analysis may cause inconsistent results

  • Laboratory vs On-line Monitoring

    • Automated analysis (on-line) is key to solving critical water problems and reducing labor requirements.

    • Fast detection and correction in each unit process can cut treatment costs and keep a plant’s process in control.

  • Laboratory vs On-line Monitoring

    • In addition to monitoring, on-line instrumentation can also be used to control chemical feed, process automation

  • Laboratory vs On-line Monitoring

    • Laboratory analysis can (and should) be used to supplement the on-line instrumentation – Parameters not needing constant measurement or

    control – Problem solving – Checking calibration of on-line instruments

  • Process Monitoring

    • Process instruments automate analytical tests for on-line, continuous monitoring and control.

  • Advantages to Process Monitoring

    • Detect process problems as they happen • Save operator time • May be used for process automation

  • Selecting an Instrument

    • Considerations in selecting a process instrument – Maintenance requirements – Does it meet the needs of the users? – Reliability – Ease of operation – Features – Cost

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric) 1720D • Calibration and Verification Strategies • Hands-On Lab

  • Color Measurement

    • Chlorine Analyzer uses Color Measurement • Measure the intensity of light passing through

    a colored sample • Convert light intensity measurements to

    concentration

  • Color Measurement

    In this case, the sample absorbs very little of the incident light.

    It is a low concentration sample.

  • Color Measurement

    In this case, most of the incident light is absorbed by the sample.

    It is a high concentration sample.

  • Lamp Monochromator or Filter

    Lens Sample Detector

    Basic Spectrophotometer

  • What is Turbidity?

    • A measure of relative water clarity • An indirect measure of suspended

    solids – Measured by light scattering

    • An indicator of water quality

  • Basic 90o Turbidimeter

  • Key Differences

    • CL17 measures dissolved substance – Color formed is proportional to concentration. – Turbidity is potential interference.

    • 1720 Turbidimeters measure suspended matter

    – Scattered light is proportional to concentration. – Color is an interference.

  • Key Differences

    • CL17 measures transmitted light at 180 degrees to light source.

    • 1720 Turbidimeters measure scattered light at 90 degrees to light source.

  • Key Differences

    • CL17’s analysis is a two-step process, that can compensate for aging light source.

    • 1720 Turbidimeter analysis is a single-step process which cannot compensate for aging light sources.

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric) 1720 Turbidimeters • Calibration and Verification Strategies • Hands-On Lab

  • Operation of (colorimetric) CL17 • Colorimetry is the measurement of color. • The intensity of the color relates to the chlorine

    concentration, based on the instrument calibration. • The CL17 measures transmitted light at 180 degrees

    to the light source. • Two complimentary colors involved (green light and

    reacted pink color)

  • Lamp Filter Sample Detector

    Old Model CL17

    New Model CL17

    Sample Detector LED

  • Operation of (colorimetric) CL17

    • Analysis is a two-step, comparative process. – ZERO then READ

    • The final results are displayed as mg/L. (milligrams/Liter)

  • CL17 Chlorine Analyzer

    Colorimeter

    Pump Module

    Keypad Wiring

  • Operation of (colorimetric) CL17 • “ZERO”

    – Sample cell flushed with fresh sample. – Flow stops and instrument “zeroes out” any color

    or turbidity. • “READ”

    – Reagents added, mixed, and allowed to react. – The pink color formed is measured and compared

    to calibration to determine chlorine concentration.

  • CL17 CHLORINE ANALYZER

    SAMPLE INLET SAMPLE VALVE

    UNDER PRESSURE

    LIGHT SOURCE TO WASTE

    DETECTOR GLASS CELL

    MAGNETIC STIRRER PINCHERS

    INDICATOR

    BUFFER SOLUTION

  • CL17 Operation - Zero

    Sample line opens

    Colorimeter cell flushes with sample.

  • CL17 Operation - Zero

    Sample line closes and flow through colorimeter stops.

    CL17 Zeros on the blank with clear solution (sample) in cell.

  • CL17 Operation - Read

    Sample line closed.

    Reagents dispensed through middle tubes.

  • CL17 Operation - Read

    Reagents mix with sample in sample cell.

    If chlorine is present, pink color forms and persists for ~1 minute.

    CL17 reads sample.

  • CL17 Operation - Read

    CL17 uses factory programmed calibration curve to convert absorbance measurement into mg/L chlorine reading.

  • Light Source Detector

    % Transmittance Absorbance

    100

    10

    0.00

    1.00

    “ZERO”

    60 0.22 “READ”

    40 0.40 “READ”

    “READ”

    (Abs = -log T)

  • Absorbance

    Concentration (m

    g/L) 0.88

    1.60

    4.00

    0.22 0.40 1.00

    Slope or gain of

    calibration

  • Operation of (colorimetric) CL17

    • What happens when light source fades? - Is the accuracy of the reading affected?

    • No. The instrument response is a comparative process. The transmitted light is proportional to incident light, therefore %T is unaffected.

  • Light Source Detector

    % Transmittance Absorbance

    100

    60

    40

    10

    0.00

    0.22

    0.40

    1.00

    “ZERO”

    “READ”

    “READ”

    “READ”

    (Abs = -log T)

  • Light Source

    Detector

    % Transmittance Absorbance

    100

    60

    40

    10

    0.00

    0.22

    0.40

    1.00

    “ZERO”

    “READ”

    “READ”

    “READ”

    (Abs = -log T)

    Faded Light Source

    The amount of light differs, but not the percent transmittance!

  • Common Mistakes (CL17)

    • Incorrect Flow Rate (300 mL/min ideal) – Low flow causes low readings due to incomplete

    cell flushing – High flow causes low readings due to constant

    flushing (color never develops)

  • Common Mistakes (CL17)

    • Reagents/Mixing – DPD powder not used (mix entire bottle of

    powder into indicator solution) – No stir bar, 2 stirbars, stir bar upside down (old) – Failure to change tubing regularly – Failure to prime reagents after tubing change

  • Common Mistakes (CL17)

    • Reagents/Mixing – Use stir bar retriever to remove stir bar from

    sample cell

  • Common Mistakes (CL17)

    • Reagents/Mixing – Use caution not to lose stir bar after removal

  • Common Mistakes (CL17)

    • Problem with optics – Dirty sample cell – Faulty interference filter (old design) – Problem with lamp or photocell

  • Common Mistakes (CL17)

    • “Wrong” Reading – Improper (lab) comparison calibration

    • Check your GAIN – should be 1.00

  • Common Mistakes (CL17)

    Press MENU and scroll to SETUP

    Scroll to GAIN and hit ENTER

  • Common Mistakes (CL17)

    Optimal gain!

  • Common Mistakes (CL17)

    • To check the gain on an old CL17 – Hit the STD key – Gain of 1.00 is optimal

  • Absorbance

    Concentration (m

    g/L) 0.40

    0.10

    (1.00) (2.00)

    (0.50)

    It’s not 0.40 mg/L, it should be 0.20mg/!

  • Absorbance

    Concentration (m

    g/L)

    0.20

    0.10

    (1.00) (2.00)

    (0.50)

    It’s not 0.40 mg/L, it should be 0.20mg/!

    Gain adjusted to “calibrate”

    reading

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric) 1720 Turbidimeters • Calibration and Verification Strategies • Hands-On Lab

  • Operation of 1720 Turbidimeters

    • Turbidity relates to the suspended matter in a liquid, but is not a direct measurement of it.

    • Suspended matter causes light to scatter or be redirected upon contact.

    • Instruments measure scattered light at 90 degrees to light source

  • Operation of 1720 Turbidimeters

    • A one-step process – Turbidimeters don’t have a ZERO key! – Zero turbidity does not exist.

    • Results are displayed as NTU. – Nephelometric Turbidity Units

  • 1720E

  • SAMPLE IN

    SAMPLE OUT

    1720D/E Flow Diagram

    Sample In

    Sample Out

  • Basic 90o Nephelometer

  • 90ø DETECTOR

    LAMP

    LENS

    APERATURE

    WATER LEVEL

    Lamp

    Lens

    Aperture

    Water Level

    90o detector

    Turbidimeter Principle of Operation

  • Operation of 1720 Turbidimeters

    DETECTOR DETECTOR DETECTOR

    LIGH

    T SOU

    RC

    E

  • Amount of Light Scattered

    Turbidity (NTU

    )

    0.20

    0.40

    0.80

    Slope or gain of

    calibration

  • Operation of 1720 Turbidimeters

    • What happens as light source ages and fades - Is the accuracy of the reading affected?

    • Yes! The amount of light scattered is directly related to the calculated turbidity, based on the calibration.

    • An aging bulb causes low readings.

  • Operation of 1720 Turbidimeters

    New Lamp

    0.20 NTU 0.40 NTU 0.80 NTU

    DETECTOR

    LIGH

    T SOU

    RC

    E

    DETECTOR DETECTOR

  • Operation of 1720 Turbidimeters

    0.20 NTU 0.40 NTU 0.80 NTU

    LIGH

    T SOU

    RC

    E

    Aged Lamp

    DETECTOR DETECTOR DETECTOR

  • Amount of Light Scattered

    Turbidity (NTU

    ) 0.20

    0.40

    0.80

    Slope or gain of

    calibration

  • Amount of Light Scattered

    Turbidity (NTU

    )

    0.20

    0.40

    0.80

    Slope or gain of

    calibration

    0.12

    0.48

    0.24

  • Amount of Light Scattered

    Turbidity (NTU

    ) 0.20

    0.40

    0.80

    Original Gain

    Adjusted Gain

  • Operation of 1720 Turbidimeters

    • Measurement is a single step process, and is unable to account for light decay.

    • Periodic calibration corrects the problem by adjusting the relationship between scattered light and corresponding turbidity(calibration gain).

  • Common Mistakes 1720Turbidimeters

    • Improper (lab) comparison calibration – Can cause large errors, since it is usually

    performed at low concentrations – Some state guidelines (CA) do not recommend

    comparison calibration

  • 1720D/E Turbidimeter Calibration

    • User-Prepared Standards – Dilution water and 20 NTU

    Formazin • StablCal Standards

    – 20 NTU Standard Solution • Comparison with Lab Reading

    – not recommended (Hach)

  • Proper Technique For Lab Turbidity Measurements

    • Clean sample cells • Use silicone oil • Degas sample • Maintain instrument • Follow proper calibration procedure

  • Common Mistakes 1720Turbidimeters

    • Improper Maintenance – Replace lamp annually – Clean turb body, bubble trap

    • What can scale do to turbidity values? – Keep optics (lamp, lens, photocell) clean

    • Before calibration, not after!

  • 1720D/E Lamp

    • Instrument provides a warm environment for biological growth, as evidenced below!

  • Common Mistakes 1720Turbidimeters

    • Error in calibration procedure – Contamination – Inaccurate measurement of formazin – Bubbles in dilution water – Failure to clean photodetector – Failure to zero electronics

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric) 1720 Turbidimeters • Calibration and Verification Strategies • Hands-On Lab

  • Calibration/Verification Strategies • CL17 Chlorine Analyzer

    – Calibration not recommended (by Hach) – Verification with lab DPD analysis

    • Lab analysis verified with standard solutions

    • 1720 Turbidimeters – Multiple choices for Calibration – Verification

    • Lab comparison or Ice-Pic Verification Module

  • CL17 Calibration • Calibration/Verification kit

    • Cat # 544900 Complete Kit • Cat # 2835900 Reagents Only

    • Comparison to Laboratory Instrument • Not Recommended by Hach

  • CL17 Verification – Periodically verify lab

    measurement with standard solutions or standard additions

    – Collect grab sample from CL17 – Perform lab analysis on grab

    sample – Compare results – Acceptance +/- 10%

  • 1720D/E Turbidimeter Calibration

    • User-Prepared Standards – Dilution water & 20 NTU

    Formazin • StablCal Standards

    – 20 NTU Standard Solution • Comparison with Lab Reading

    – not recommended (Hach)

  • 1720D/E Turbidimeter Verification • Lab Comparison Method

    – Calibrate lab turbidimeter – Verify lab technique/analysis with 0.5

    or 1.0 NTU StablCal – Collect grab sample from turbidimeter

    body – Measure grab with lab instrument – Compare results – Acceptance +/- 10% or 0.05 below 0.5

    NTU

  • 1720D/E Turbidimeter Verification

    • Use of Ice-Pic Verification Module – Silence alarms, hold outputs – Thoroughly dry turb head – Place module inside turb body – Place turb head inside module, let

    reading stabilize – Compare readings with stored value – Acceptance +/- 10%

    http://ecommerce.hach.com/webapp/commerce/command/ExecMacro/prodlgimage.d2w/report?image=Lpi-5225000-2500-1500-ICPI.jpg�http://ecommerce.hach.com/webapp/commerce/command/ExecMacro/prodlgimage.d2w/report?image=Lpi-5225000-2500-1500-ICPI.jpg�

  • Course Outline

    • Process Monitoring • Key Differences (color and turbidity) • Operation of (colorimetric) CL17 • Operation of (turbidimetric)1720 Turbidimters • Calibration and Verification Strategies • Hands-On Demonstration

  • Operation, Calibration, and Performance Verification of CL17 Chlorine Analyzer and

    1720 Series Turbidimeters

  • Chris Griffin [email protected]

    804-513-6731

    2014 Plant Operations Conference

    mailto:[email protected]�http://www.vaawwa.org/�

  • Operation, Calibration and Verification for pH

    http://www.hach.com/fmmimghach?CODE:NGMP-PHC101-PRBESTND9156|1�

  • pH Theory

    • pH is a measurement of the relative acidity of an aqueous solution

    • pH is a measurement of hydrogen ion concentration

  • pH Theory

    • Acid - increases the hydrogen ion (H+) concentration in a solution

    • Base - increases the hydroxide ion (OH-) concentration in a solution

  • pH Scale

    7 14 0 Acid Base

    Neutral

    Vinegar pH 3

    Ammonia pH 11.5

  • pH Scale

    • pH is a negative logarithmic function • Each decrease in pH unit = 10X increase in acidity

    – Solution at pH4 is 10X more acidic than solution at pH5 – Solution at pH 4 is 100X more acidic than pH6 solution

    7 14 0 6 5 4

    10X

    100X

  • Measuring pH

  • How Does a pH Probe Work?

    • Probe measures hydrogen ion concentration – Two electrodes in probe - sensing half-cell, reference

    half-cell

  • Half-Cells

    • Ion sensing pH half cell – Glass bulb that is sensitive to H+.

    • Reference half-cell – Glass tube filled with salt solution to complete circuit.

  • METER

    Ion Sensing Half-Cell

    Reference Half-Cell

    Ag/AgCl Wire

    Internal Filling Solution

    Reference Electrolyte

    Salt Bridge Junction

  • Reference Half-Cell

    • Dispenses reference solution which completes circuit for meter

  • Sensing Half-Cell

    H+ H+

    H+ H+

    H+

    H+

    H+ H+

    H+

    H+ H+

    H+ H+

    H+ H+

    H+

    Hydrogen ion concentration fixed at

    pH 7

    pH 7 Solution

    H+ conc the same both inside and outside glass bulb

    *No potential develops

  • Sensing Half-Cell

    0mV

    pH 7 Solution

    H+ conc the same both inside & outside glass bulb

    *No Potential develops

  • Sensing Half-Cell

    H+ H+

    H+ H+

    H+

    H+

    H+ H+

    1000H+

    Hydrogen ion concentration fixed at

    pH 7

    pH 4 Solution

    H+ conc 1000x greater outside glass bulb

    *Potential develops

    1000H+ 1000H+

    1000H+ 1000H+

    1000H+ 1000H+

  • Sensing Half-Cell

    pH 4 Solution

    H+ conc 1000x greater outside glass bulb

    *Potential develops

    180mV

  • Sensing Half-Cell

    1000 H+ 1000 H+

    1000H+ 1000H+

    Hydrogen ion concentration fixed at

    pH 7

    pH 10 Solution

    H+ conc 1000x greater inside glass bulb

    *Potential develops

    H+

    H+

    H+

    H+

  • Sensing Half-Cell

    -180mV

    pH 10 Solution

    H+ conc 1000x greater inside glass bulb

    *Potential develops

  • Calibration

    • A calibration curve allows the meter to convert a measured millivolt potential into a pH reading.

    pH

    mV

  • Calibration

    • The optimal slope for pH is –59.16 /decade * • Acceptance criteria = +/- 5% or 3 mV

    * at 25 degrees Celsius

  • Calibration

    mV

    pH

    0

    +180

    -180 4 7 10

  • Calibration

    • -180mV difference measured between pH4 and pH7 • pH4 to pH7 (3 pH units) is 1000x concentration

    change • Decade = 10-fold concentration change = 1pH unit • -180/3 = -60 ≈ -59.16 mV/decade

  • Probe Care and Maintenance

    http://www.hach.com/fmmimghach?CODE:NGMP-PHC101-PRBESTND9156|1�

  • Maintenance

    • New probe • Calibration • Measurement/Storage • Troubleshooting • Cleaning

  • New Probe

    • Condition new pH probe in pH 7 buffer for approximately 30 minutes before initial use

  • Calibration

    • Use fresh calibration solutions • Use temperature compensation • Allow enough time for meter to stabilize on both

    Temp and pH reading • Record mV readings to determine drift in sensor

  • Calibrate

    • Calibrate bench pH meters each shift using two or three fresh buffer solutions

    4.0 7.0 10.0

  • Calibration and Verification

    • On-line sensors comparison check daily – Agreement to (+/-) 0.20 pH units with a calibrated

    laboratory meter • On-line sensors monthly calibrate linearity or per

    regulatory schedule

  • Measurement

    • Place probe into sample, stir, and wait for readings to stabilize

    • Rinse and dry between measurements • Storage between measurements

    – Sample or solution of similar ionic strength to sample – pH4 buffer – Electrode storage solution (i.e. 3M KCl)

  • Troubleshooting

    • mV reading in pH 7 buffer – Should read 0 ± 30 mV in pH 7 buffer

    • Response time – May require cleaning if slow (2 min) in buffered solution

    • Slope – Optimal slope is –59.16 ± 3 mV/decade (+/- 5%)

  • Cleaning

    • Slow response may indicate need for cleaning – Longer than 2 minutes – Immerse and agitate in a warm dilute mild detergent

    solution for a few minutes. Rinse with DI and blot dry before use.

    – Alternate soaking in dilute hydrochloric acid and dilute sodium hydroxide. Rinse with DI water and condition in pH 7 buffer before use.

  • pH - Take Home Messages

    • pH is an electrochemical measurement, useful in many applications.

    • As with many electrochemical analyses, pH requires frequent calibration to achieve accurate results.

    • Proper probe maintenance is essential.

  • Chris Griffin [email protected]

    804-513-6731

    2014 Plant Operations Conference

    mailto:[email protected]�http://www.vaawwa.org/�

    2 PIV Verification of CL17 and 1720 TurbProcess Instrument Verification�for Drinking Water���2014 Plant Operations Conference�PIV for Drinking WaterOperation, Calibration, and Performance Verification of�CL17 Chlorine Analyzer and 1720 Series TurbidimetersTake-Home MessagesCourse OutlineProcess MonitoringLaboratory vs On-line MonitoringLaboratory vs On-line MonitoringLaboratory vs On-line MonitoringLaboratory vs On-line MonitoringProcess MonitoringAdvantages to Process MonitoringSelecting an InstrumentSlide Number 14Course OutlineColor MeasurementColor MeasurementColor MeasurementSlide Number 19What is Turbidity?Basic 90o TurbidimeterKey DifferencesKey DifferencesKey DifferencesCourse OutlineOperation of (colorimetric) CL17Slide Number 27Operation of (colorimetric) CL17CL17 Chlorine AnalyzerOperation of (colorimetric) CL17Slide Number 31CL17 Operation - ZeroCL17 Operation - ZeroCL17 Operation - ReadCL17 Operation - ReadCL17 Operation - ReadSlide Number 39Slide Number 40Operation of (colorimetric) CL17Slide Number 42Slide Number 43Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Common Mistakes (CL17)Slide Number 54Slide Number 55Course OutlineOperation of 1720 Turbidimeters Operation of 1720 Turbidimeters 1720ESlide Number 61Basic 90o NephelometerSlide Number 63Operation of 1720 TurbidimetersSlide Number 66Operation of 1720 TurbidimetersOperation of 1720 TurbidimetersOperation of 1720 TurbidimetersSlide Number 70Slide Number 71Slide Number 72Operation of 1720 TurbidimetersCommon Mistakes 1720Turbidimeters1720D/E Turbidimeter CalibrationProper Technique For Lab Turbidity Measurements Common Mistakes 1720Turbidimeters1720D/E LampCommon Mistakes 1720TurbidimetersCourse OutlineCalibration/Verification StrategiesCL17 CalibrationCL17 Verification1720D/E Turbidimeter Calibration 1720D/E Turbidimeter Verification1720D/E Turbidimeter VerificationCourse OutlineOperation, Calibration, and Performance Verification of�CL17 Chlorine Analyzer and �1720 Series Turbidimeters�����Chris Griffin�[email protected]�804-513-6731���2014 Plant Operations Conference�

    4 PIV pHOperation, Calibration and Verification for pH�pH TheorypH TheorypH Scale pH Scale Slide Number 26How Does a pH Probe Work?Half-CellsSlide Number 29Reference Half-CellSensing Half-CellSensing Half-CellSensing Half-CellSensing Half-CellSensing Half-CellSensing Half-CellCalibrationCalibrationCalibrationCalibrationProbe Care and MaintenanceMaintenanceNew ProbeCalibration CalibrateCalibration and VerificationMeasurementTroubleshootingCleaningpH - Take Home Messages�����Chris Griffin�[email protected]�804-513-6731���2014 Plant Operations Conference�