Principles of Magnetic Resonance

Page 2: Principles of Magnetic Resonance

Review

What is Larmor Frequency?

What is its purpose in image acquisition?

Page 3: Principles of Magnetic Resonance

Larmor Frequency

It is the frequency at which the hydrogen precess

When hydrogen is near an externally applied magnetic field, an increase or decrease causes the precessional frequency or resonant frequency also increase or decrease.

Page 4: Principles of Magnetic Resonance

Larmor equation

ω0 = B0*ƔExample:

If we have an MRI capable of having a magnetic strength of 1.5 Tesla. What is the precessional frequency of hydrogen?

Answer:▪ 1.5T x 42.6MHz/T = 63.9MHz

Page 5: Principles of Magnetic Resonance

What happens when it precesses?

After the atom is “pushed” by an electromagnetic energy and the RF pulse is stopped.

Two things happens; They align with the applied magnetic field

Attain the normal rate of precession.

Page 6: Principles of Magnetic Resonance

What happens…The RF energy that was absorbed

by the atom is released, this energy decays rapidly and is referred to as Free Induction Decay.

A single pulse will not provide enough information for imaging techniques, so this pulse is repeated multiple times.

An interval at which this pulse is repeated is called repetition time or TR and is usually in milliseconds.

Page 7: Principles of Magnetic Resonance

Cont…The repetition time determines

how much longitudinal (T1) occurs. TR controls the T1 weighting of the

image.The FID signal needs to be

focused so a large signal appears at certain time. Echo, and the time it takes to

appear after the RF pulse is applied is called the echo time (TE).

Page 8: Principles of Magnetic Resonance

Echo Is produced from the nuclei

remaining in the transverse plane. Dephasing and precession occurs

at different rates.A refocusing pulse is given at

a given time, usually 1800, at half the echo time (TE/2). It flips the entire system into its

mirror image.

Page 9: Principles of Magnetic Resonance

Refocusing pulse effect on echo• Those nuclei precessing at a

faster rate, but are now going in opposite direction and now are behind the slower precessing nuclei.

• At the echo time, TE, all nuclei have reached the same point and the receiver coil detects a large signal, which refers to as echo.

Page 10: Principles of Magnetic Resonance

A

B

C

RF PulseRefocusing

Pulse

Page 11: Principles of Magnetic Resonance

Refocusing pulse effect on echo

Page 12: Principles of Magnetic Resonance

Relaxation TimeT1 and T2 relaxation time

Page 13: Principles of Magnetic Resonance

Relaxation

The term returning to equilibrium is called relaxation and can be thought of as two-step process. T1 relaxation or Longitudinal relaxation

T2 relaxation or Transverse relaxation

Page 14: Principles of Magnetic Resonance

Relaxation

Page 15: Principles of Magnetic Resonance

T1 relaxation It is controlled by a time constant

referred to as T1. It is the time it takes about 63% of

the nuclei to realign with the external magnetic field.

After the magnetic moment is flipped 900 by the application of a pulse of RF energy, the pulse is turned off. This is followed by a gradual return to equilibrium along the z axis.

Page 16: Principles of Magnetic Resonance
Page 17: Principles of Magnetic Resonance
Page 18: Principles of Magnetic Resonance

T2 relaxation It occurs simultaneously with the T1

relaxation. It involves the rate of precession of

different nuclei.When RF pulse is applied all atoms

are precessing together, in phase, at the exact same rate. Once it stops, each atom finds itself in a slightly different magnetic environment.

Page 19: Principles of Magnetic Resonance

Cont…The time constant controlling

how fast this process occurs is called the T2 relaxation.

As with T1 values, the T2 value is the time it takes for 63% of the nuclei to be out of phase with each other.

The T2 happens more rapidly than the T1.

Page 20: Principles of Magnetic Resonance

Cont…T2 value is much shorter than

the T1 value. Typical T1 for biological tissues

range from 200 to 1,000ms, whereas T2 50 to 150ms.

At higher applied magnetic fields, T1 is increased while T2 are caused by local effects and are unaffected by field strengths.

Page 21: Principles of Magnetic Resonance
Page 22: Principles of Magnetic Resonance

T2* or T-2 starThere are Inhomogeneities in the

local magnetic field around each nuclei which make the actual dephasing go faster.

Inhomogeneities; Variation in the applied magnetic

field Changes in magnetic

susceptibilities Chemical shifts Changes caused by spins of nuclei

moving in vascular tissues.

Page 23: Principles of Magnetic Resonance

Spin-spin Relaxation

During T2 relaxation, all of the nuclide phase their spins by transferring energy to neighboring atoms, this process is referred as spin-spin relaxation.

Page 24: Principles of Magnetic Resonance

SummaryDuring relaxation process,

the atoms emit RF energy of exactly the same energy that they had absorbed; these atomic magnets return to the equilibrium state that they were in before. This energy that is emitted by the atoms is responsible for generating the MR image.

Page 25: Principles of Magnetic Resonance

T1 is primarily affected by the magnetic field of the overall environment, while the T2 relaxation is predominantly affected by the magnetic field in the local vicinity.

The RF signal that is given off by the atoms returning to equilibrium is detected by an antenna and analyzed by computer for MR imaging.