Preliminary results from 3D CMS pixel detectors

Ozhan Koybasi, Enver Alagoz, Ryan Rivera, Lorenzo Uplegger, Marcos Turqueti, Kirk Arndt, Daniela Bortoletto, Simon Kwan, et al.

Page 2: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

2E Configuration 4E Configuration

Four 3D sensors mounted on plaquettes for testbeam:

2E_WB5_2 : 2E configuration, 280µm substrate thickness4E_WB5_8 : 4E configuration, 280µm substrate thickness4E_WB2-16_5 : 4E configuration, 200µm substrate thickness2E_WB2-16_2 : 2E configuration, 200µm substrate thickness

n+ (readout) p+ (bias)

3D CMS Pixel Layouts

Page 3: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Assembly

• Support wafer made wire-bonding challenging ( especially high voltage wiring)• Sensors mounted on carbon fiber plates (0.5 mm thick)• Cooling was done by a chiller• Sensor temperature measurement: an RTD placed on the carbon fiber (cooling tubes side): ΔΤ = 6 °C

Cooling tubes

sensorPlastic protection

bias wire

Page 4: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

I-V Characteristics

• Measured at room temperature

Page 5: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Noise

• Measured at room temperature• Unable to measure noise at Vbias < 40V for 4E sensors

Page 6: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

TELESCOPE ENCLOSURE

BEAM

FERMILAB NETWORK

CONTROL ROOM

DUT

PIXEL DETECTORS

SCINTILLATORS

3.7V POWER SUPPLY

ACELLERATOR CLOCK

CLOCK AND TRIGGER

DISTRIBUTION

Test beam setup at FNAL

CAPTAN SYSTEM

2x3 plaquettes

2x4 plaquettes

• Test beam at FNAL• 120 GeV protons• No B field

Page 7: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Beam Test Results

Sensor : 2E_WB5_2

2E Configuration n+ (readout) p+ (bias)

280µm substrate thickness

Page 8: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

• Gain calibration done

• ADC to electron conversion:Vcal* [DAC] = ADC x gain - offsetCharge (e-) = Vcal x 65.5 – 410

* 1 Vcal [DAC] = 65.5 electrons

• T ≈ 11 oC on carbon fiber (estimated to be 6 oC higher on the sensor)

Beam spot on 3D

Threshold [DAC]= 50 Vbias = 40V

Page 9: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

• Each point is charge distribution mean

• Convoluted Landau and Gaussian fit to be done later

• Bias scan done only at Threshold [DAC] = 50

Threshold [DAC] = 50

Vbias=40V

Charge collection

Page 10: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Beam Test Results

Sensor : 2E_WB2-16_6

2E Configuration n+ (readout) p+ (bias)

200µm substrate thickness

Page 11: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

• Gain calibration was wrong: it has to be done again• Detector data still requires further studies

Beam spot on 3D

ADC distribution

Vbias=40V

Page 12: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Summary

• Four 3D sensors received and wire bonded• Lab characterization done• 3D sensor assembled for the testbeam• Sensors tested at FNAL with 120 GeV protons• Bias and threshold scans performed• 2Es worked and 4Es failed

Next:• Further analysis• Gain calibration and charge collection with a

radioactive source setup at Purdue

Page 13: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

BACKUP SLIDES

Page 14: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

Correlation between planes

Page 15: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V Sensor : 2E_WB5_2

Page 16: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V

Page 17: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V

Page 18: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V

Page 19: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Sensor : 2E_WB2-16_6Threshold [DAC]= 50, Vbias = 60V

Page 20: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V

Page 21: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V

Page 22: Preliminary results from 3D CMS pixel detectors · 2E_WB5_2 : 2E configuration, 280µm substrate thickness 4E_WB5_8 : 4E configuration, 280µm substrate thickness 4E_WB2-16_5 : 4E

DUT

Threshold [DAC]= 50, Vbias = 60V