Perspectives in Medicine (2012) 1, 129—131 Bartels E, Bartels S, Poppert H (Editors): New Trends in Neurosonology and Cerebral Hemodynamics — an Update. Perspectives in Medicine (2012) 1, 129—131 journal homepage: www.elsevier.com/locate/permed Post-carotid stent ultrasound provides critical data to avoid rare but serious complications John J. Volpi a,b,, Zsolt Garami b,c , Rasadul Kabir a , Orlando Diaz-Daza a,b , Richard Klucznik a,b a The Methodist Hospital Neurological Institute,6560 Fannin, Suite 802, Houston, TX, 77030, USA b The Methodist Hospital Research Institute, 6565 Fannin, Houston, TX, 77030, USA c Methodist DeBakey Heart and Vascular Center, Houston, 6550 Fannin, Suite 1401, TX,77030, USA KEYWORDS Stroke; Carotid ultrasound; Carotid stenting Summary Carotid stenting is a common procedure for revascularization of carotid artery stenosis. In this study, we evaluated the role of carotid ultrasound post carotid stenting. In a retrospective analysis, we identified 45 patients who received post-stent ultrasound. On routine follow-up we measured a range for peak systolic velocity of 33—150 cm/s and end diastolic veloc- ity 11—52 cm/s. We also identified two cases, where immediate post-stent ultrasound provided critical data that required further intervention, and potentially avoided serious complications. © 2012 Elsevier GmbH. All rights reserved. Introduction Carotid stenting is an accepted form of revascularization in the US and many countries based on the recent results of the CREST trial [1]. The choice of follow-up imaging remains variable for post-stent patients and some patients receiving no post-stent imaging. Ultrasound imaging is a cost effective and simple way to evaluate immediate post-stent patients. Methods We retrospectively reviewed a database for a 2 year period from 2008 to 2010 for patients who had significant carotid stenosis and underwent carotid stenting, and post-stent carotid ultrasound exam. In stent velocities were measured with a General Electric LOGIQ E9 (Milwaukee, WI) with 9 MHz linear probe that was used to evaluate the post stent carotid artery. Corresponding author at: 6560 Fannin, Suite 802, Houston, TX, 77030, USA. Tel.: +1 713 441 3951; fax: +1 713 793 7019. E-mail address: [email protected] (J.J. Volpi). Results Forty-five patients (age between 43 and 75 years) were identified, who received post stent ultrasound. We found a mean peak systolic velocity of 83 cm/s and a mean end diastolic velocity of 24 cm/s in this population, with a range peak systolic velocity 33—150 cm/s and end diastolic veloc- ity 11—52 cm/s. Out of 45, we found 2 cases of immediate complications of stent placement. Discussion Case 1: A 77-year-old woman with no focal neurological deficits underwent elective right carotid stenting at an outside institution. Post stent, she complained of neck pain and was lethargic with fluctuating left side weakness. She was transferred to our facility and was found with low flow in the recently stented vessel (Fig. 1). The stent appeared to be patent and fully deployed, but on follow-up angiogram was found to be in the dissected false lumen of 2211-968X/$ — see front matter © 2012 Elsevier GmbH. All rights reserved. doi:10.1016/j.permed.2012.02.019

Post-carotid stent ultrasound provides critical data to avoid rare but serious complications

  • Upload
    richard

  • View
    214

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Post-carotid stent ultrasound provides critical data to avoid rare but serious complications

Perspectives in Medicine (2012) 1, 129—131

Bartels E, Bartels S, Poppert H (Editors):New Trends in Neurosonology and Cerebral Hemodynamics — an Update.

Perspectives in Medicine (2012) 1, 129—131

j o u r n a l h o m e p a g e : w w w . e l s e v i e r . c o m / l o c a t e / p e r m e d

Post-carotid stent ultrasound provides critical datato avoid rare but serious complications

John J. Volpi a,b,∗, Zsolt Garamib,c, Rasadul Kabira, Orlando Diaz-Dazaa,b,Richard Klucznika,b

a The Methodist Hospital Neurological Institute,6560 Fannin, Suite 802, Houston, TX, 77030, USAb The Methodist Hospital Research Institute, 6565 Fannin, Houston, TX, 77030, USAc Methodist DeBakey Heart and Vascular Center, Houston, 6550 Fannin, Suite 1401, TX,77030, USA

KEYWORDSStroke;

Summary Carotid stenting is a common procedure for revascularization of carotid arterystenosis. In this study, we evaluated the role of carotid ultrasound post carotid stenting. In a

Carotid ultrasound;Carotid stenting

retrospective analysis, we identified 45 patients who received post-stent ultrasound. On routinefollow-up we measured a range for peak systolic velocity of 33—150 cm/s and end diastolic veloc-ity 11—52 cm/s. We also identified two cases, where immediate post-stent ultrasound providedcritical data that required further intervention, and potentially avoided serious complications.

right

R

Fiadpi

o

© 2012 Elsevier GmbH. All

Introduction

Carotid stenting is an accepted form of revascularizationin the US and many countries based on the recent results ofthe CREST trial [1]. The choice of follow-up imaging remainsvariable for post-stent patients and some patients receivingno post-stent imaging. Ultrasound imaging is a cost effectiveand simple way to evaluate immediate post-stent patients.

Methods

We retrospectively reviewed a database for a 2 year periodfrom 2008 to 2010 for patients who had significant carotidstenosis and underwent carotid stenting, and post-stentcarotid ultrasound exam. In stent velocities were measuredwith a General Electric LOGIQ E9 (Milwaukee, WI) with9 MHz linear probe that was used to evaluate the post stentcarotid artery.

∗ Corresponding author at: 6560 Fannin, Suite 802, Houston, TX,77030, USA. Tel.: +1 713 441 3951; fax: +1 713 793 7019.

E-mail address: [email protected] (J.J. Volpi).

D

CdopSlaa

2211-968X/$ — see front matter © 2012 Elsevier GmbH. All rights reservdoi:10.1016/j.permed.2012.02.019

s reserved.

esults

orty-five patients (age between 43 and 75 years) weredentified, who received post stent ultrasound. We found

mean peak systolic velocity of 83 cm/s and a mean endiastolic velocity of 24 cm/s in this population, with a rangeeak systolic velocity 33—150 cm/s and end diastolic veloc-ty 11—52 cm/s.

Out of 45, we found 2 cases of immediate complicationsf stent placement.

iscussion

ase 1: A 77-year-old woman with no focal neurologicaleficits underwent elective right carotid stenting at anutside institution. Post stent, she complained of neckain and was lethargic with fluctuating left side weakness.

he was transferred to our facility and was found withow flow in the recently stented vessel (Fig. 1). The stentppeared to be patent and fully deployed, but on follow-upngiogram was found to be in the dissected false lumen of

ed.

Page 2: Post-carotid stent ultrasound provides critical data to avoid rare but serious complications

130 J.J. Volpi et al.

Figure 1 Post Carotid ultrasound. Flow in the right carotidartery did not appear centered in the vessel lumen.

Figure 2 Dissected lumen. The angiogram confirmed that thestent was deployed into a dissected false lumen.

Figure 3 Plan for repair. This treatment approach was usedfor repair of the stent in the false lumen.

Figure 4 Initial angiogram and post stent ultrasound. Theuntreated right mid internal carotid artery appeared normal onto

tn

csutnm

he routine angiogram, but the post stent ultrasound showed anbvious lumen irregularity.

he carotid (Fig. 2). This was subsequently corrected witho adverse events (Fig. 3).

Case 2: A 40-year-old woman with recent motor vehi-le accident and carotid dissection underwent successfultenting of the affected left carotid artery. On her follow-up

ltrasound, the stented carotid was normal, but the con-ralateral (untreated) carotid artery was found to have aew flow-limiting dissection with clot (Fig. 4). This abnor-ality was not apparent on the initial angiogram during
Page 3: Post-carotid stent ultrasound provides critical data to avoid rare but serious complications

Post-carotid stent ultrasound provides critical data to avoid rare

C

Issost4sts

R

[1] Brott TG, Hobson RW, Howard G, Roubin GS, Clark WM, Brooks W,et al. Stenting versus endarterectomy for treatment of carotid-artery stenosis. N Engl J Med 2010;363:11—23.

Figure 5 Angiogram and stent. The second angiogram showedan obvious dissection and it was corrected with a second stent.

its injection (Fig. 4). A second angiogram was performedand the dissection was easily identified, and this vesselwas also stented subsequently without any adverse events(Fig. 5).

but serious complications 131

onclusion

n all patients, post stent ultrasound provides a baselinetudy for future follow-up. In rare cases, post stent ultra-ound can identify potentially serious complications. Inur study 2 out of 45 patients (4.4%) were found with aignificant abnormality post stenting that could have ledo cerebral ischemia. Interestingly, in the CREST study,.4% of post stent patients suffered stroke or death. Weuggest that post carotid stent ultrasound may yield poten-ially valuable findings to reduce the risk of imminenttroke.

eference