Physics of MR Image Acquisition HST-583, Fall 2002 Review: - MRI: Overview - MRI: Spatial Encoding MRI Contrast: Basic sequences - Gradient Echo - Spin Echo - Inversion Recovery HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology Dr. Jorge Jovicich HST.583

Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

  • Upload
    others

  • View
    6

  • Download
    2

Embed Size (px)

Citation preview

Page 1: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Physics of MR Image AcquisitionHST-583, Fall 2002

• Review: - MRI: Overview - MRI: Spatial Encoding

• MRI Contrast: Basic sequences - Gradient Echo - Spin Echo - Inversion Recovery

HST.583: Functional Magnetic Resonance Imaging: Data Acquisition and Analysis Harvard-MIT Division of Health Sciences and Technology Dr. Jorge Jovicich

HST.583

Page 2: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Overview of an MRI procedure

Tissue protons align

IMAGE

(equilibrium state)

RF pulses

Protons absorb

(excited state)

processes

Protons emit RF energy )

detection

Relaxation processes

Repeat

using magnetic field gradients

Subject

Safety screening

RAW DATA MATRIX Fourier transform

00 Bγω = Larmor EquationMagnetic field with magnetic field

RF energy

Relaxation

(return to equilibrium state

NMR signal

Spatial encoding

HST.583 S(t1,t2) S(x1,x2)

Page 3: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Dynamics of the Magnetization

d r

M( t )

dt =

rM( t ) × γ

rBext ( t) −

( Mx i + My j) T2

* − (Mz − M0 )

T1

k

B0

M0

timeEquilibrium

RF

Equilibrium

signal 00 Bγω =

B0 B0 B0 B0

M0

M(t)M(t) M(t)

• Mathematical Description: precession + relaxation (Bloch equations)

NMR

• Geometrical description: damped precession

HST.583

Page 4: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Spatial Encoding in MRI

Key concept: ω( r ) = γ (B0 + G.r)

• Slice Selection

• Frequency Encoding

• Phase Encoding

- Location - Thickness - Rephasing/Refocussing

- Fourier Transform - FOV - Gradient Echo Formation

- Phase / Frequency Equivalency - FOV

HST.583

Page 5: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Slice Selection

• RF excitation (selective bandwidth) + magnetic field gradient

shape & duration bandwidth (bw)• RF pulse properties

frequency (wo)

∆z

z)

z

B Bo bw

w

zo

wo

Gz

ω( r ) = γ B0 + G.r( ) RF

Gz

Phase

• Magnetic field gradient (G • Schematically

Slice thickness (∆z): bw = γ Gz ∆z • Slice parameters Slice center (zo): wo = γ Gz zo

HST.583

Page 6: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

‘Pulse sequence’ so far

RF

“slice select” Gz

MR signalfrom whole S(t)

Selected slice

Sample points

t

t

T2 *

t

HST.583 Courtesy of L. Wald

Page 7: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Frequency Encoding

• After slice selection all spins in the selected plane have the same frequency and phase (a)

• Signal is acquired while a magnetic field gradient Gx is on (b)

Gx

(a) (b) Gradient field y

63 MHz 64 MHz 65 MHz x

x

w(x) = γ Gx x

HST.583

Page 8: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

‘Pulse sequence’ so far

RF

“slice select” Gz

“freq. encode” Gx(read-out)

S(t)

Sample points t

t

t

t Signal dephases rapidly

HST.583 Courtesy of L. Wald

Page 9: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

More signal can be obtained:

RF

“slice select” Gz

“freq. encode” Gx(read-out)

S(t)

Sample pointst

-+

Acquisition window

Dephase

Rephase

We can save some time …HST.583

Page 10: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

‘Pulse sequence’ so far

RF

“slice select” Gz

“freq. encode” Gx(read-out)

S(t)

Sample points

HST.583

t

-

+

Acquisition window

Gradient Echo

Page 11: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Frequency Encoding

Measured signal: Frequency spectrum: time domain 1D x-projection

S(t) Fourier transform

S(ω)

time

bw frequencyrec

δt Receiver bandwidth (bw )rec

Sampling time δt=1/ bwrec Nx: number of samples ω( r ) = γ (B0 + G.r)

Acquisition time (tacq): t = N δt = N / bwacq x x rec

• Relevant parameters Resolution (δx): δx = FOV / Nx x

Field of view (FOVx): bw = γ G FOVrec x x

HST.583

Page 12: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

We could have measured ‘exactly’the same information differently:

RF

“slice select” Gz

“freq. encode” Gx(read-out)

S(t)

Sample points t

-

+

Acquisition window

RF excitation sampling points

Each

So far we’ve used ONE and we’ve acquired Nx

while the read-out gradient was ON

HST.583

Page 13: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Let’s look at one arbitrarysampled point in the MR signal

RF

Gz

Gx

S(t)

HST.583 dephasing during t Same net dephasing

-

+

t -

+ = -

-

RF

S(t)

Gz

Gx =

S(t) depends on net

Page 14: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

We can do the same for all Nx sampled points:

-

RF

Gz

Gx RF

Gz

Gx

+

RF

Gz

Gx

Measure one point after dephasing gradient

Net negative dephasing

Net zero dephasingRepeat experiment Nx

times to sample all points Net positive dephasingHST.583

Page 15: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

We could have measured ‘exactly’the same information differently:

S(t)

Measured signal: time domain

RF S(t)

slice Gzselect time

phase Gx FT

encode Frequency spectrum:

1D x-projection S(ω)

Repeat Nx times frequency

HST.583

Page 16: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Basic Gradient Echo Sequence

RF α α

TR TE

Gz Slice

Gy Phase

Gx Read

HST.583

Page 17: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Conventional spin-warpMRI sequence

θ Repeat Nphase times kphase

RF

GN

GNphase

slice

phase

phase

TE kreadGread

Signal Prephase

timet

0 acq

NNread TR Prephase

read

The signal we measure is in tγspatial frequency space (k-space) k(t ) = G(t ') dt '

2π ∫ 0 HST.583

Page 18: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Pulse Sequences & Image Contrast

Contents: • Definition of Contrast • Contrast parameters • Concept of MRI contrast weighting • Properties of pulse sequences &

sequence parameters

HST.583

Page 19: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Image Contrast Definition

• Goal: maximise the contrast ( USEFUL IMAGES! )

• Contrast: difference in MR signals between different tissues

Tissue 1 Tissue 2

Water protons’ T1, T2, T2*, ρ, T1, T2, T2*, ρ,mobility/environment: Diffusion, Flow Diffusion, Flow

MR signal: S1

noise

SSCNR σ

21 − =

S2

Some examples?Contrast to Noise ratio: HST.583

Page 20: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Image Contrast: What can we manipulate?

Tissue Properties: fixed Experimental Variables

Tissue T1 (ms) T2 (ms) ρ* Fat 260 84 0.90 White Matter 780 90 0.72 Gray Matter 920 100 0.84 CSF 3000 300 1.00

• Pulse sequence

• Pulse sequence parameters

- Repetition time: TR

- Echo time: TE

ρ*: % H2O relative to CSF - Inversion time: TI

- RF flip angle: α

• Contrast agent

What’s the effect of these variables?

HST.583

Page 21: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Image Contrast:Weighting the MR Signal

• General MRI pulse sequence: combination of contrasts

Signal Intensity: S(x,y) = k T1× ρ ×

× T2 × …

• Contrast Weighting: maximize one term, minimize the others

Example: T1-weighting T1

S(x,y) = k ρ× ×

× T2 × …

by choosing adequate sequence & sequence parameters

Some examples?HST.583

Page 22: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Typical MRI sequences

Gradient Echo:- Proton Density weighting - T1 weighting - T2* weighting

Spin Echo:- Proton Density weighting - T1 weighting - T2 weighting - Double echo: PD & T2 weighting

Inversion Recovery:- MP-RAGE: for high T1 contrast - FLAIR: for PD or T2 weighting without CSF signal

HST.583

Page 23: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Typical MRI sequences

Gradient Echo:- Proton Density weighting - T1 weighting - T2* weighting

Spin Echo:- Proton Density weighting - T1 weighting - T2 weighting - Double echo: PD & T2 weighting

Inversion Recovery:- MP-RAGE: for high T1 contrast - FLAIR: for PD or T2 weighting without CSF signal

HST.583

Page 24: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Basic Gradient Echo Sequence

RF

Slice

Phase

Read

α α

TR TE

−−= *

21 0 expexp1sin

T TE

T TRMSGRE αMR Signal:

HST.583

Page 25: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Repetition Time (TR): T1 contrast

TR encodeRF

Voltage(Signal)

Mxy

Mz

WM

GM

encode encode

GM: gray matter (long T1) time WM: white matter (short T1)

HST.583 Courtesy of L. Wald

Page 26: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Gradient Echo Sequence

−−= *

21 0 expexp1sin

T TE

T TRMSGRE α

*

*

Proton Density weighting T1 weighting T2* weighting ( TR >>T1, short TE, sin α ~ 1 ) ( TR ~ T1, short TE, sin α ~ 1 ) ( TR >>T1, TE ~ T2* )

Water density Gray / White matter / CSF Hemorrhage

HST.583

Page 27: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

HST.583

Gradient Echo Sequence

FA = 3o

FA = 5o

FA = 30o

FLASH

Courtesy of A. Dale and B. Fischl

Manipulating contrast with flip angle

Proton Density Weighting T1 Weighting

Page 28: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Typical MRI sequences

Gradient Echo:- Proton Density weighting - T1 weighting - T2* weighting

Spin Echo:- Proton Density weighting - T1 weighting - T2 weighting - Double echo: PD & T2 weighting

Inversion Recovery: - MP-RAGE: for high T1 contrast - FLAIR: for PD or T2 weighting without CSF signal

HST.583

Page 29: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Basic Spin-Echo Sequence

Excitation Pulse Refocusing Pulse

τ τ

180o

90o

RF

Slice

TE Phase

Read

Spin EchoHST.583

Page 30: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Spin-Echo Sequence: The MR Signal

τ τ

TE

TR

S TR TE

SE = M0 1 − exp − exp − T1 T2

HST.583

Page 31: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Spin Echo Sequence

−−=

21 0 expexp1sin

T TE

T TRMSGRE α

Proton Density weighting T1 weighting T2 weighting ( TR >>T1, short TE, sin α ~ 1 ) ( TR ~ T1, short TE, sin α ~ 1 ) ( TR >>T1, TE ~ T2 )

Water density Gray / White matter / CSF Hemorrhage

HST.583

Page 32: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Double Spin Echo Sequence

τ τ τ τ

180o 180o

90o

RF

Slice

TE1 TE2Phase

Read

Echo 1 Echo 2 HST.583

Page 33: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Typical MRI sequences

Gradient Echo:- Proton Density weighting - T1 weighting - T2* weighting

Spin Echo:- Proton Density weighting - T1 weighting - T2 weighting - Double echo: PD & T2 weighting

Inversion Recovery:- MP-RAGE: for high T1 contrast - FLAIR: for PD or T2 weighting without CSF signal

HST.583

Page 34: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Inversion Recovery

Excitation Pulse Refocusing Pulse Inversion Pulse

τ τ TI

180o 180o

90o

RF

Slice

TE Phase

Read

Magnetization Preparation Spin Echo

HST.583

Page 35: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Enhanced T1-weighting: Inversion Recovery

M0

-Mo

Spin Echo

TE

Mz(t)

90o

180o180o TI

CSF GM

180o Inversion: prepare magnetization prior to Spin Echo detection

2/1/1/ )21(Signal eee −−− +−∝ ρ T TE T TR T TI

HST.583

Page 36: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Fluid Attenuation Inversion Recovery (FLAIR)

M0

-Mo

Mz(t)

180o

TI

Spin Echo

TE

90o

180o

CSF

GM

Signal null at: TI = ln(2)* T1

HST.583

Page 37: Physics of MR Image Acquisitiondspace.mit.edu/bitstream/handle/1721.1/36388/HST-583Fall-2002/NR/rdonlyres/Health...Overview of an MRI procedure Tissue protons align IMAGE (equilibrium

Physics of MR Image AcquisitionHST-583, Fall 2002