Frédéric Molina 19.03.2018 Phase Noise measurement and Pulse stability measurement COMPANY RESTRICTED

Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

  • Upload
    others

  • View
    32

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Frédéric Molina19.03.2018

Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 2: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Agenda

ı Part 1 : Phase Noise measurement

ı Part 2 : Pulse stability measurement

19/03/2018 2

DUT

Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 3: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

What is Noise?Phase and Amplitude Noise

ı Ideal Signal (Noiseless)

V(t) = A . sin(2t)

A = nominal amplitude = nominal frequency

3

t

FFT

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 4: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

What is Noise?Phase and Amplitude Noise

ı Real Signal

V(t) = [A + E(t)] . sin(2t + t)

E(t) = amplitude fluctuationst = phase fluctuations

4

t

)]1([)___

_1_____log(10)(0

HzdBcPPowerFrequencyCarrier

PHztorelatedPowerSidebandSinglefL SSBm

FFT

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 5: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement MethodsPhase Detector

5

DUT

Reference Source

PDLPF

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 6: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

FSWP : Digital Phase DemodulatorInternal structure

19/03/2018 6

ı PM signal has phase wraps at ±π. No filtering or FFT possibleı FM demodulator preferred to digital PLL due to its simplicity

Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 7: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

• Simultaneous measurement of AM Noise and Phase Noise with Total Noise

Phase Noise

AM Noise

FSWP

Total Noise (PN+AM)

AM Noise, Phase Noise, and Total Noise

19/03/2018 7Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 8: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Special / Complex Measurement Setups1. Pulsed Phase Noise

ı Like a Spectrum Analyser phase noise measurement

ı Automatic PRI detection

ı What about Pulse Desensitisation? Improved dynamic due to… Gating on Pulse Cross Correlation

8

DUT

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 9: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

1/PW = 100kHz

A * 20 log(/T)(pulse desensitization)

PRF Lines

• PRF filter is narrow to pass center line and attenuate all other PRF lines• Center line is lower than pulse level due to pulse desensitization: 20*log(duty cycle)

PW ()10us

PRI (T)100us

Level (A)10dBm

Time

1/PRI = PRF = 10kHz

Pulsed Phase Noise – Pulse Spectrum

19/03/2018 9Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 10: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Pulsed Phase Noise

Pulsed

~6dB

CW

• Max offset limited to PRF/2 for pulsed signals due to PRF lines• Pulsed phase noise ~6dB higher at PRF/2 due to coherent combining

0.5/PRI = 5kHz1/PRI = PRF = 10kHz

19/03/2018 10Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 11: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Pulsed Phase Noise Measurement

Max offset automatically limited to PRF/2 (5kHz)

• Measurement truncated to max offset of PRF/2• Measurement time same as for CW

19/03/2018 11Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 12: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Pulsed vs CW Phase Noise Measurement

Pulsed measurements same as CW for offsets << PRF/2

Pulsed measurement higher than CW at PRF/2

<< PRF/2PRF/2

CW

Pulsed

19/03/2018 12Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 13: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

R&S FSWP – FSWP-K4… and Show Everything on one Sight

l Phase Noisel AM Noisel Pulse in Time

Domainl Pulse in

FrequencyDomain

19/03/2018 13Phase Noise measurementand Pulse stability measurement

Page 14: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Additive Phase Noise – Digital Phase DemodulatorMeasurement Setup

DUT

19/03/2018 14Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 15: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Additive Phase Noise – Digital Phase Demodulator• Internal hardware automatically reconfigures when “Additive” is selected• No phase detector or need for quadrature – no phase shifter• Greatly simplifies measurement setup and calibration• Internal Low Noise Synthesizer as DUT stimulus

FSWP Signal Source AnalyzerLow Noise Synthesizer

DUT

LPF

LPF

Digitizers+ FPGA

LNA

LNA

LPF

LPF LNA

LNA

19/03/2018 15Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 16: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Additive Phase Noise – Digital Phase Demodulator• Internal Low Noise Synthesizer as DUT stimulus

19/03/2018 16Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 17: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Agenda

ı Part 1 : Phase Noise measurement

ı Part 2 : Pulse stability measurement

19/03/2018 17Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 18: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Option K6 : Pulse measurement

ı The K6 Pulse Measurement Application is available on Running on directly on signal analyzers (FSW, FSWP, FPS) Running on PC (VSE) connected to analyzers and scopes

19/03/2018 18Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 19: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Process

19/03/2018 19Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 20: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Process

19/03/2018 20Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 21: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessData Acquisitionl Segmented capture

l Capture more pulses, less noisel Support in HW for same timing accuracy as “normal” capture

19/03/2018 21COMPANY RESTRICTED

Page 22: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Process

19/03/2018 22Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 23: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Process

19/03/2018 23Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 24: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessEnvelope Parameters

Ampl

itude

(% o

f Pul

se T

op)

19/03/2018 24COMPANY RESTRICTED

Page 25: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Process

19/03/2018 25Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 26: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessResult Display

ı Trend Plots column of table Visual recognition of „trends“ Multiple trends with different

parameters X-axis can be Timestamp or

Pulse Number

19/03/2018 26Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 27: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessResult Display

ı Trend as “spectrum” Plots FFT of column of table Visual recognition of

periodicities Multiple plots with different

parameters

19/03/2018 27Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 28: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessResult Display

ı Distribution Plots histogram of table column Visual recognition of „states“ Multiple distributions with different

parameters Y-axis can be Pulse Count or

Occurrence

19/03/2018 28Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 29: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement ProcessResult Displayı Linked markers – inter-pulse and intra-pulse

2919/03/2018COMPANY RESTRICTED

Page 30: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

K60 : Transient Analysis (Hop)

ı Automated measurement per hop Power measurements Frequency/phase Settling time Hop-hop changes

ı Statistics Numerical Trend Histogram

19/03/2018 30Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 31: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

K60 : Transient Analysis (Chirp)

ı Automated measurement per chirp Power measurements Frequency/phase Modulation linearity

ı Statistics Numerical Trend Histogram

19/03/2018 31Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 32: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

FSWP-K6P : Pulse stability measurement

ı Variability over time with respect to a reference value

32

scusi @ 123RF.com scusi @ 123RF.com scusi @ 123RF.com

Time

Reference

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 33: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Stability – Pulsed Radar

ı Variability over time with respect to a reference value

33

Time

Reference

60 dB

0.01 dB

RF Power

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 34: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Stability – Pulsed Radar

ı Variability over time with respect to a reference value

34

Time

RF Power

Reference

3 GHz

0.001 rad

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 35: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Stability – Pulsed Radar

ı In radar, moving targets cause “variability over time” in the radar return To discern a target, the transmit/receive hardware needs to be stable

3519/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 36: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Metrics

Standard deviation, normalized to the average power

Relative to average power,-60 dBc: average power deviation over all pulses is 60 dB smaller than average power level

36

Pulse (n)

Pulse (n)

10 · log dBc

19/03/2018COMPANY RESTRICTED

Page 37: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Metrics

Pulse-to-pulse deviation, normalized to the average power

Relative to average power,-60 dBc: average power deviation pulses-to-pulse is 60 dB smaller than average power level

37

Pulse (n)

Pulse (n)

10 · log dBc

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 38: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Metrics

Standard deviation

Relative to 1 radian,-60 dBc means RMS phase deviation over all pulses equals of 1 milli-radian (0.001rad)

38

Pulse (n)

Pulse (n)

10 · log dBc

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 39: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Metrics

Pulse-to-pulse deviation

Relative to 1 radian,-60 dBc means RMS phase deviation pulse-to-pulse equals of 1 milli-radian (0.001rad)

39

Pulse (n)

Pulse (n)

10 · log dBc

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 40: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement

ı We differentiate between two kinds of measurement:

40

DUT Analyzer DUT Analyzer

“Absolute” “Additive”

19/03/2018Phase Noise measurementand Pulse stability measurement

DUT

COMPANY RESTRICTED

Page 41: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Absolute measurement with Spectrum digitizerı In FSWP for “absolute” measurements,we have two digitizers: Spectrum digitizer (digital I/Q)

Phase Noise digitizer (2x analog I/Q)

41

DUT Analyzer

“Absolute”ATTENRF In

RefChan. 1

SynthChan. 1

LO 1

fADC

LO 1

fADC

PCFPGA

7680 MHz

1280 MHz

ATTEN ADC 0° 90° NCO

9 GHz

LO 1

< 8 GHz

19/03/2018Phase Noise measurementand Pulse stability measurement

ATTENRF In

0° 90°

0° 90°

ADC1,I

ADC1,Q

ADC2,I

ADC2,Q

I1

Q1

I2

Q2

RefChan. 1

SynthChan. 1

LO 1

LO 2

fADC1

fADC1

fADC2

fADC2

LO 1

fADC1

PCFPGA

RefChan. 2

SynthChan. 2

LO 2

fADC2 DUT

COMPANY RESTRICTED

Page 42: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Additive measurement with Phase Noise digitizer

ı In FSWP for “additive” measurements, only thePhase Noise digitizer (2nd analog I/Q mixer) is used

42

DUT Analyzer

“Additive”

LO 2Pulse Switch

ATTENRF Out

ATTENRF In

0° 90°

ADC2,I

ADC2,Q

I2

Q2

LO 2 fADC2

fADC2

PCFPGARef

Chan. 2

SynthChan. 2

fADC2

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 43: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Performance[test measurements: DUT bypass]ı Initial performance results for 3 GHz vs digitizer Pulse width: 5 us

WB for “wideband” SA path, “PN (Abs|Add)” for phase noise path (Absolute | Additive)

43

-100

-80

-60

-40

-20

0

3 GHz Amplitude Stability, T: 10us

WB PN (Abs) PN (Add)

-100

-80

-60

-40

-20

0

3 GHz Phase Stability, T: 10us

WB PN (Abs) PN (Add)

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 44: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Performance[test measurements: DUT bypass]ı Performance summary phase stability (median on pulse), 10 MHz I/Q bandwidth Pulse width: 5 us

WB for “wideband” SA path, “PN” for phase noise path (Absolute), , “PN+” for phase noise path (Additive)

47

Interval 10 us 100 us 1 ms 10 msDigitizer WB PN PN+ WB PN PN+ WB PN PN+ WB PN PN+

3 GHz ‐67.4 ‐74.3 ‐81.6 ‐67.2 ‐73.9 ‐81.6 ‐63.4 ‐64.0 ‐81.5 ‐49.4 ‐47.8 ‐81.2

10 GHz ‐65.0 ‐64.5 ‐75.0 ‐64.5 ‐64.1 ‐74.9 ‐54.1 ‐55.1 ‐74.8 ‐37.6 ‐40.3 ‐74.7

18 GHz ‐60.6 ‐63.7 ‐74.2 ‐60.0 ‐62.6 ‐74.1 ‐49.9 ‐50.0 ‐73.8 ‐33.9 ‐34.2 ‐69.5

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 45: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement – Performance[test measurements: DUT bypass]ı Performance summary amplitude stability (median on pulse), 10 MHz I/Q bandwidth Pulse width: 5 us

WB for “wideband” SA path, “PN” for phase noise path (Absolute), , “PN+” for phase noise path (Additive)

48

Interval 10 us 100 us 1 ms 10 msDigitizer WB PN PN+ WB PN PN+ WB PN PN+ WB PN PN+

3 GHz ‐71.1 ‐84.7 ‐85.3 ‐70.8 ‐84.6 ‐85.1 ‐71.1 ‐84.4 ‐84.9 ‐71.0 ‐84.1 ‐84.5

10 GHz ‐68.4 ‐75.6 ‐78.2 ‐68.3 ‐75.5 ‐77.9 ‐68.3 ‐75.4 ‐77.7 ‐68.3 ‐75.2 ‐76.8

18 GHz ‐64.6 ‐72.1 ‐75.1 ‐64.6 ‐72.0 ‐75.0 ‐64.5 ‐71.9 ‐74.9 ‐64.4 ‐71.3 ‐74.0

19/03/2018Phase Noise measurementand Pulse stability measurement

COMPANY RESTRICTED

Page 46: Phase Noise measurement and Pulse stability measurement...Absolute measurement with Spectrum digitizer ı In FSWP for “absolute” measurements,wehave two digitizers: Spectrum digitizer

Measurement Application

ı The FSWP-K6P Pulse Stability Measurements extends the K6 application on FSWP

ı Multi-burst analysis Groups pulses in “bursts”

ı Burst averaging Focus on in-burst stability Control averaging range Compare positions in burst

4919/03/2018Phase Noise measurementand Pulse stability measurement