2/9/2017 1 ©2014 MFMER | slide-1 PET/MR: Technology, Installation and Clinical Practice Robert A. Pooley, PhD, Brad Kemp, Ph.D, Kris Gorny, Ph.D., Anshuman Panda, Ph.D., Gene Paden, M.S., Judy James, Ph.D. FLAAPM Fall Meeting, Daytona Beach Shores September 9, 2016 ©2014 MFMER | slide-2 Disclosure • Nothing to disclose

PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

1

©2014 MFMER | slide-1

PET/MR:Technology, Installation and Clinical Practice

Robert A. Pooley, PhD, Brad Kemp, Ph.D, Kris Gorny, Ph.D., Anshuman Panda, Ph.D., Gene Paden, M.S., Judy James, Ph.D.

FLAAPM Fall Meeting, Daytona Beach Shores

September 9, 2016

©2014 MFMER | slide-2

Disclosure

• Nothing to disclose

Page 2: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

2

©2014 MFMER | slide-3

Catana

PET/MR Systems

GE SIGNA PET/MR

©2014 MFMER | slide-4

Page 3: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

3

©2014 MFMER | slide-5

GE SIGNA PET/MR

• MR• 3T Discovery 750W• 70 cm bore• PET compatible array coils

• PET• Time-of-flight• 60 cm transverse FOV• Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog

SiPM• 10.5% energy resolution• 390 ps average timing resolution• Transverse spatial resolution = 4.5 mm

©2014 MFMER | slide-6

Positron Imaging

Page 4: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

4

©2014 MFMER | slide-7

Positron Imaging

Total Events = Trues + Scatter + Randoms

©2014 MFMER | slide-8

GE SIGNA PET/MR

• MR• 3T Discovery 750W• 70 cm bore• PET compatible array coils

• PET• Time-of-flight• 60 cm transverse FOV• Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog

SiPM• 10.5% energy resolution• 390 ps average timing resolution• Transverse spatial resolution = 4.5 mm

Page 5: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

5

©2014 MFMER | slide-9Kemp

©2014 MFMER | slide-10Kemp

Page 6: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

6

©2014 MFMER | slide-11

Motivation for Combining PET and MR

• MR provides preferred image contrast for some anatomy

• brain, breast, prostate, abdomen• Single modality scan preferred over multiple scans

for patient, with improved co-registration.

• Decrease in dose for at-risk populations• No CT scan

• Improvements in PET image quality• MR based motion correction• MR based partial volume effect correction

©2014 MFMER | slide-12Catana

Page 7: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

7

©2014 MFMER | slide-13

©2014 MFMER | slide-14Catana

Page 8: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

8

©2014 MFMER | slide-15

Catana

©2014 MFMER | slide-16

How MR and PET Can Affect Each Other

• How MR can affect PET• Main magnetic field (PET detector failure)• Gradient fields (eddy currents, heating, vibration)• RF interference (drop in PET count rate)

• How PET can affect MR• Susceptibility artifacts (B0 inhomogeneity, linearity of

gradient fields)• RF interference (artifacts from PET electronics)

Page 9: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

9

©2014 MFMER | slide-17

Challenges

• MRI compatible PET detectors• Typical PET scanners use photomultipliers

which do not operate well in magnetic fields• Development of solid state photo detectors

• Avalanche photodiodes (APDs)• Silicon photomultipliers (SiPMs)• Immune to magnetic fields due to short

path of charge carriers

©2014 MFMER | slide-18Kemp

Page 10: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

10

©2014 MFMER | slide-19Kemp

©2014 MFMER | slide-20Kemp

Page 11: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

11

©2014 MFMER | slide-21Kemp

©2014 MFMER | slide-22Kemp

Page 12: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

12

©2014 MFMER | slide-23

Challenges

• Quantitative imaging• PET attenuation correction requires linear

attenuation coefficients at 511 keV.• CT data from PET/CT can easily be

transformed to linear attn. coefficients• More difficult in PET/MR

• Distinguishing bone from air filled cavities is a particular challenge

• UTE sequences will improve this in future

©2014 MFMER | slide-24

Attenuation Correction Techniques

• MRI – Segmentation

• MRI – Atlas

• MRI – UTE

• PET – Emission

• PET – Transmission

Page 13: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

13

©2014 MFMER | slide-25Kemp

©2014 MFMER | slide-26Kemp

Page 14: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

14

©2014 MFMER | slide-27Kemp

©2014 MFMER | slide-28Kemp

Page 15: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

15

©2014 MFMER | slide-29

MRAC

Kemp

©2014 MFMER | slide-30

MRAC

Kemp

Page 16: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

16

©2014 MFMER | slide-31

PET/MR Installation

• Design considerations• As simple as combining nuclear medicine facility

with MR facility• Shielding

• Magnetic field• RF interference• Acoustic noise• Radiation protection

• MR access – 4 Zones• Ferromagnetic detector• Uptake areas, MR-only patient waiting areas

©2014 MFMER | slide-32

Page 17: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

17

©2014 MFMER | slide-33

©2014 MFMER | slide-34

PET/MR Acceptance Testing

• Primarily separate MR and PET testing

• MR• ACR Phantom• Coil QC• RF noise• Acoustic noise• Magnetic fringe field

• PET• NEMA NU 2-2007. Manual and Software analysis provided with

system.• Image quality, Accuracy of Attenuation, Scatter Corrections• Spatial Resolution• Sensitivity• Scatter Fraction, Count Losses, Randoms Measurement

• Simultaneous PET/MR Tests: co-registration, PET stability with EPI

MR system Passed all tests except one coil that was not installed properly, and excessive acoustic noise in control room.

PET system Passed all tests

Simultaneous PET/MR Passed

Page 18: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

18

©2014 MFMER | slide-35

ACR PET Phantom

©2014 MFMER | slide-36

ACR PET Phantom

Page 19: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

19

©2014 MFMER | slide-37

NEMA - Image Quality

©2014 MFMER | slide-38

NEMA - Accuracy

Page 20: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

20

©2014 MFMER | slide-39

NEMA - Scatter Fraction

©2014 MFMER | slide-40

NEMA - Sensitivity

Page 21: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

21

©2014 MFMER | slide-41

NEMA - Spatial Resolution

©2014 MFMER | slide-42

PET/MR Clinical Practice

• Scan Protocol

• Clinical Indications

Page 22: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

22

©2014 MFMER | slide-43Kemp

©2014 MFMER | slide-44Kemp

Page 23: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

23

©2014 MFMER | slide-45Kemp

©2014 MFMER | slide-46

Catana

Page 24: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

24

©2014 MFMER | slide-47

Clinical Protocols and Indications -Oncologic

• Torso Survey PET/MR plus Focused Brain PETMR:

• Advanced Lung Cancer• CNS Lymphoma• Metastatic Melanoma• Cancer with Brain Metastases

• Torso Survey PET/MR plus Focused Abdomen PET/MR

• Pancreatic Cancer• Hepatic Cancer• Cancer with Hepatic Metastases

©2014 MFMER | slide-48

Clinical Protocols and Indications -Oncologic

• Torso Survey PET/MR plus Focused Pelvic PET/MR

• Endometrial Cancer• Ovarian Cancer• Cervical Cancer• Vulvar Cancer

• Torso Survey PET/MR plus Focused Brachial Plexus PET/MR

• Metastasis to Nervous System• Peripheral nerve sheath tumor of the brachial plexus• Neurolymphomatosis of the Brachial Plexus

Page 25: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

25

©2014 MFMER | slide-49

Clinical Protocols and Indications -Neurologic

• Focused Brain PET/MR• Intractable Epilepsy

©2014 MFMER | slide-50

Clinical Images

Page 26: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

26

©2014 MFMER | slide-51

Indication: Seizures. 5.4 mCi FDG

Broad areas of hypometabolism

©2014 MFMER | slide-52

15-year-old female with diagnosis of primary generalized epilepsy. 4.5 mCi FDG

Extensive area of decreased FDG activity; highly suspicious for a primary seizure focus

Page 27: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

27

©2014 MFMER | slide-53

CNS lymphoma, status post methotrexate, procarbazine, and Rituxan

Significant residual hypermetabolic CNS lymphoma

F-18 FDG, 10.81 mCi

©2014 MFMER | slide-54

Page 28: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

28

©2014 MFMER | slide-55

PET Brain from WB Survey PET Brain from Focused PET

©2014 MFMER | slide-56

Page 29: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

29

©2014 MFMER | slide-57

©2014 MFMER | slide-58

Restaging prostate cancer

Multiple choline avid foci worrisome for local seminal vesicle recurrence and pelvic nodal metastatic disease

C-11 Choline, 10.71 mCi

Page 30: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

30

©2014 MFMER | slide-59

©2014 MFMER | slide-60

Page 31: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

31

©2014 MFMER | slide-61

Radiotracers

• [18F] FDG Clinical Metabolism

• [18F] Fluoride Clinical Bone scan

• [13N] Ammonia Clinical Myocardial perfusion

• [11C] Choline Clinical Prostate cancer

• [18F] FDOPA Research Cancer

• [11C] Acetate Research Myocardial metabolism

• [O-15] Water Research Perfusion

• [15O] Oxygen Research Metabolism

• [18F] FLT Research Proliferation; brain tumors

• [11C] Flumazenil Research GABAergic CNS

©2014 MFMER | slide-62

Radiotracers

• [11C] McN5652 Research SERT CNS

• [11C] HED Research Myocardial Innervation

• [11C] PIB Research Amyloid Beta CNS

• [18F] AV138 Research Amyloid Beta CNS

• [18F] Florbetapir Research Amyloid Beta CNS

• [18F] Flutemetamol Research Amyloid Beta CNS

• [11C] Palmitate Research Metabolism

• [18F] AV1451 Research Tau CNS

• [18F] FTHA Research Fatty acid uptake

Page 32: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

32

©2014 MFMER | slide-63

References

• Catana. “Integrated PET/MRI.” MGH/HST Center for Biomedical Imaging.

• Kemp, Gorny. “PET/MR System Overview.” Mayo Clinic, Rochester, MN.

• Vandenberghe, et. Al. “PET-MRI: a review of challenges and solutions in the development of integrated multimodality imaging.” Phys. Med. Biol. 60 (2015) R115-R154

©2014 MFMER | slide-64

Thank YouMayo Clinic Florida

[email protected]

Page 33: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

33

©2014 MFMER | slide-65

©2014 MFMER | slide-66

Page 34: PETMR Pooley FLAAPM Fall... · • Lutetium based scintillator (4 x 5.3 x 25 mm^3) with analog SiPM • 10.5% energy resolution • 390 ps average timing resolution • Transverse

2/9/2017

34

©2014 MFMER | slide-67