21
012-07135B Speed of Light Apparatus SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A Beam-splitter Measuring Microscope s L 1 L 2 M F (Fixed Mirror) M R (Rotating Mirror) Laser

SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

Embed Size (px)

Citation preview

Page 1: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

012-07135B Speed of Light Apparatus

i

SPEED OF LIGHTAPPARATUS

Instruction Manual andExperiment Guide forthe PASCO scientificModel OS-9261A, 62 and 63A

Beam-splitter

MeasuringMicroscope

s L1L2

MF

(Fixed Mirror)

MR

(Rotating Mirror)

Laser

Page 2: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

Speed of Light Apparatus 012-07135B

�ii

The lightning flash with arrowhead,within an equilateral triangle, is intendedto alert the user of the presence ofuninsulated “dangerous voltage” withinthe product’s enclosure that may be ofsufficient magnitude to constitute a riskof electric shock to persons.

CAUTION:TO PREVENT THE RISK OFELECTRIC SHOCK, DO NOTREMOVE BACK COVER. NO USERSERVICEABLE PARTS INSIDE.REFER SERVICING TO QUALIFIEDSERVICE PERSONNEL.

The exclamation point within anequilateral triangle is intended to alertthe user of the presence of importantoperating and maintenance(servicing) instructions in theliterature accompanying theappliance.

CAUTIONRISK OF ELECTRIC SHOCK

DO NOT OPEN

Page 3: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

012-07135B Speed of Light Apparatus

i

Table of Contents

Section Page

Copyright, Warranty, and Equipment Return .................................................. ii

Introduction ..................................................................................................... 1

Measuring the Velocity of Light: History ....................................................... 2

Galileo

Römer

Fizeau

Foucault

The Foucault Method ....................................................................................... 3

A Qualitative Description

A Quantitative Description

Equipment ........................................................................................................ 6

Setup and Alignment ....................................................................................... 8

Alignment Summary ........................................................................... 12

Alignment Hints.................................................................................. 13

Making the Measurement ............................................................................... 14

Notes on Accuracy and Maintenance ............................................................. 16

Page 4: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

Speed of Light Apparatus 012-07135B

�ii

Credits

This manual authored by: Bruce Lee

This manual edited by: Dave Griffith

Copyright Notice

The PASCO scientific 012-07135A Speed of LightApparatus manual is copyrighted and all rightsreserved. However, permission is granted to non-profit educational institutions for reproduction ofany part of the manual providing the reproductionsare used only for their laboratories and are not soldfor profit. Reproduction under any other circum-stances, without the written consent of PASCOscientific, is prohibited.

Limited Warranty

PASCO scientific warrants the product to be freefrom defects in materials and workmanship for aperiod of one year from the date of shipment to thecustomer. PASCO will repair or replace at its optionany part of the product which is deemed to bedefective in material or workmanship. The warrantydoes not cover damage to the product caused byabuse or improper use. Determination of whether aproduct failure is the result of a manufacturingdefect or improper use by the customer shall bemade solely by PASCO scientific. Responsibilityfor the return of equipment for warranty repairbelongs to the customer. Equipment must beproperly packed to prevent damage and shippedpostage or freight prepaid. (Damage caused byimproper packing of the equipment for returnshipment will not be covered by the warranty.)Shipping costs for returning the equipment afterrepair will be paid by PASCO scientific.

Copyright, Warranty, and Equipment Return

Please—Feel free to duplicate this manualsubject to the copyright restrictions below.

Equipment Return

Should the product have to be returned to PASCOscientific for any reason, notify PASCO scientificby letter, phone, or fax BEFORE returning theproduct. Upon notification, the return authorizationand shipping instructions will be promptly issued.

➤ ➤ ➤ ➤ ➤ NOTE: NO EQUIPMENT WILL BEACCEPTED FOR RETURN WITHOUT ANAUTHORIZATION FROM PASCO.

When returning equipment for repair, the units must bepacked properly. Carriers will not accept responsibilityfor damage caused by improper packing. To be certainthe unit will not be damaged in shipment, observe thefollowing rules:

➀ The packing carton must be strong enough forthe item shipped.

➁ Make certain there are at least two inches ofpacking material between any point on theapparatus and the inside walls of the carton.

➂ Make certain that the packing material cannotshift in the box or become compressed, allowingtheinstrument come in contact with the packingcarton.

Address: PASCO scientific10101 Foothills Blvd.

Roseville, CA 95747-7100

Phone: (916) 786-3800

FAX: (916) 786-3292

email: [email protected]

web: www.pasco.com

Page 5: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

1

012-07135B Speed of Light

Introduction

The velocity of light in free space is one of the mostimportant and intriguing constants of nature. Whether thelight comes from a laser on a desk top or from a star thatis hurtling away at fantastic speeds, if you measure thevelocity of the light, you measure the same constantvalue. In more precise terminology, the velocity of lightis independent of the relative velocities of the light sourceand the observer.

Furthermore, as Einstein first presented in his SpecialTheory of Relativity, the speed of light is criticallyimportant in some surprising ways. In particular:

1. The velocity of light establishes an upper limit to thevelocity that may be imparted to any object.

2. Objects moving near the velocity of light follow a setof physical laws drastically different, not only fromNewton’s Laws, but from the basic assumptions ofhuman intuition.

With this in mind, it’s not surprising that a great deal oftime and effort has been invested in measuring the speedof light. Some of the most accurate measurements weremade by Albert Michelson between 1926 and 1929 usingmethods very similar to those you will be using with thePASCO Speed of Light Apparatus. Michelson measuredthe velocity of light in air to be 2.99712 x 108 m/sec.From this result he deduced the velocity in free space tobe 2.99796 x 108 m/sec.

But Michelson was by no means the first to concernhimself with this measurement. His work was built on ahistory of ever-improving methodology.

Page 6: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

2

Speed of Light 012-07135B

Galileo

Through much of history, those few who thought tospeculate on the velocity of light considered it to beinfinite. One of the first to question this assumption wasthe great Italian physicist Galileo, who suggested amethod for actually measuring the speed of light.

The method was simple. Two people, call them A and B,take covered lanterns to the tops of hills that are separatedby a distance of about a mile. First A uncovers herlantern. As soon as B sees A’s light, she uncovers herown lantern. By measuring the time from when Auncovers her lantern until A sees B’s light, then dividingthis time by twice the distance between the hill tops, thespeed of light can be determined.

However, the speed of light being what it is, and humanreaction times being what they are, Galileo was able todetermine only that the speed of light was far greater thancould be measured using his procedure. Although Galileowas unable to provide even an approximate value for thespeed of light, his experiment set the stage for laterattempts. It also introduced an important point: to mea-sure great velocities accurately, the measurements mustbe made over a long distance.

Römer

The first successful measurement of the velocity of lightwas provided by the Danish astronomer Olaf Römer in1675. Römer based his measurement on observations ofthe eclipses of one of the moons of Jupiter. As this moonorbits Jupiter, there is a period of time when Jupiter liesbetween it and the Earth, and blocks it from view. Römernoticed that the duration of these eclipses was shorterwhen the Earth was moving toward Jupiter than when theEarth was moving away. He correctly interpreted thisphenomena as resulting from the finite speed of light.

Geometrically the moon is always behind Jupiter for thesame period of time during each eclipse. Suppose,however, that the Earth is moving away from Jupiter. Anastronomer on Earth catches his last glimpse of the moon,not at the instant the moon moves behind Jupiter, but onlyafter the last bit of unblocked light from the moon reacheshis eyes. There is a similar delay as the moon moves outfrom behind Jupiter but, since the Earth has moved

farther away, the light must now travel a longer distanceto reach the astronomer. The astronomer therefore sees aneclipse that lasts longer than the actual geometricaleclipse. Similarly, when the Earth is moving towardJupiter, the astronomer sees an eclipse that lasts a shorterinterval of time.

From observations of these eclipses over many years,Römer calculated the speed of light to be 2.1 x 108 m/sec. This value is approximately 1/3 too slow due to aninaccurate knowledge at that time of the distancesinvolved. Nevertheless, Römer’s method providedclear evidence that the velocity of light was notinfinite, and gave a reasonable estimate of its truevalue—not bad for 1675.

Fizeau

The French scientist Fizeau, in 1849, developed aningenious method for measuring the speed of lightover terrestrial distances. He used a rapidly revolvingcogwheel in front of a light source to deliver the lightto a distant mirror in discrete pulses. The mirrorreflected these pulses back toward the cogwheel.Depending on the position of the cogwheel when apulse returned, it would either block the pulse of lightor pass it through to an observer.

Fizeau measured the rates of cogwheel rotation thatallowed observation of the returning pulses for carefullymeasured distances between the cogwheel and the mirror.Using this method, Fizeau measured the speed of light tobe 3.15 x 108 m/sec. This is within a few percent of thecurrently accepted value.

Foucault

Foucault improved Fizeau’s method, using a rotatingmirror instead of a rotating cogwheel. (Since this is themethod you will use in this experiment, the details will bediscussed in considerable detail in the next section.) Asmentioned, Michelson used Foucault’s method to producesome remarkably accurate measurements of the velocityof light. The best of these measurements gave a velocityof 2.99774 x 108 m/sec. This may be compared to thepresently accepted value of 2.99792458 x 108 m/sec.

Measuring the Velocity of Light: History

Page 7: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

3

012-07135B Speed of Light

MF

(Fixed Mirror)

MR

(Rotating Mirror)

L2 Beamsplitter L1s

Laser

MeasuringMicroscope

Figure 1: Diagram of the Foucault Method

A Qualitative Description

In this experiment, you will use a method for measuringthe speed of light that is basically the same as thatdeveloped by Foucault in 1862. A diagram of the experi-mental setup is shown in Figure 1, above.

With all the equipment properly aligned and with therotating mirror stationary, the optical path is as follows.The parallel beam of light from the laser is focused to apoint image at point s by lens L

1. Lens L

2 is positioned so

that the image point at s is reflected from the rotatingmirror M

R, and is focused onto the fixed, spherical mirror

MF. M

F reflects the light back along the same path to

again focus the image at point s.

In order that the reflected point image can be viewedthrough the microscope, a beam splitter is placed in theoptical path, so a reflected image of the returning light isalso formed at point s´.

Now, suppose MR is rotated slightly so that the reflected

beam strikes MF at a different point. Because of the

spherical shape of MF, the beam will still be reflected

directly back toward MR. The return image of the source

point will still be formed at points s and s´. The onlysignificant difference in rotating M

R by a slight amount is

that the point of reflection on MF changes.

Now imagine that MR is rotating continuously at a very

high speed. In this case, the return image of the sourcepoint will no longer be formed at points s and s´. This isbecause, with M

R rotating, a light pulse that travels from

MR to M

F and back finds M

R at a different angle when it

returns than when it was first reflected. As will be shownin the following derivation, by measuring the displace-ment of the image point caused by the rotation of M

R, the

velocity of light can be determined.

A Quantitative Description

In order to use the Foucault method to measure the speedof light, it’s necessary to determine a precise relationshipbetween the speed of light and the displacement of theimage point. Of course, other variables of the experimen-tal setup also affect the displacement. These include:

• the rate of rotation of MR

• the distance between MR and M

F

• the magnification of L2, which depends on the

focal length of L2 and also on the distances be

tween L2, L

1, and M

F.

Each of these variables will show up in the final expres-sion that we derive for the speed of light.

The Foucault Method

Page 8: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

4

Speed of Light 012-07135B

To begin the derivation, consider a beam of light leavingthe laser. It follows the path described in the qualitativedescription above. That is, first the beam is focused to apoint at s, then reflected from M

R to M

F, and back to M

R.

The beam then returns through the beamsplitter, and isrefocused to a point at point s´, where it can be viewedthrough the microscope. This beam of light is reflectedfrom a particular point on M

F. As the first step in the

derivation, we must determine how the point of reflectionon M

F relates to the rotational angle of M

R.

Figure 2a shows the path of the beam of light, from thelaser to M

F, when M

R is at an angle θθθθθ. In this case, the

angle of incidence of the light path as it strikes MR is also

θθθθθ and, since the angle of incidence equals the angle ofreflection, the angle between the incident and reflectedrays is just 2θθθθθ. As shown in the diagram, the pulse oflight strikes M

F at a point that we have labeled S.

Figure 2b shows the path of the pulse of light if it leavesthe laser at a slightly later time, when M

R is at an angle

θθθθθ1 = θθθθθ + ∆θ∆θ∆θ∆θ∆θ. The angle of incidence is now equal to

θθθθθ1 = θθθθθ + ∆θ∆θ∆θ∆θ∆θ, so that the angle between the incident and

reflected rays is just 2θθθθθ1 = 2(θθθθθ + ∆θ∆θ∆θ∆θ∆θ). This time we label

the point where the pulse strikes MF as S

1. If we define D

as the distance between MF and M

R, then the distance

between S and S1 can be calculated:

S1 - S = D(2θθθθθ1

- 2θθθθθ) = D[2(θθθθθ + ∆θ∆θ∆θ∆θ∆θ) - 2θθθθθ] = 2D∆θ∆θ∆θ∆θ∆θ (EQ1)

In the next step in the derivation, it is helpful to think of asingle, very quick pulse of light leaving the laser. Sup-pose M

R is rotating, and this pulse of light strikes M

Rwhen it is at angle θθθθθ, as in Figure 2a. The pulse will thenbe reflected to point S on M

F. However, by the time the

pulse returns to MR, M

R will have rotated to a new angle,

say angle θθθθθ1. If M

R had not been rotating, but had

remained stationary, this returning pulse of light would berefocused at point s. Clearly, since M

R is now in a

different position, the light pulse will be refocused at adifferent point. We must now determine where that newpoint will be.

The situation is very much like that shown in Figure 2b,with one important difference: the beam of light that isreturning to M

R is coming from point S on M

F, instead of

from point S1. To make the situation simpler, it is conve-

nient to remove the confusion of the rotating mirror andthe beam splitter by looking at the virtual images of thebeam path, as shown in Figure 3.

The critical geometry of the virtual images is the same asfor the reflected images. Looking at the virtual images,the problem becomes a simple application of thin lensoptics. With M

R at angle θθθθθ1

, point S1 is on the focal axis

of lens L2. Point S is in the focal plane of lens L

2, but it is

a distance ∆∆∆∆∆S = S1 - S away from the focal axis. From

thin lens theory, we know that an object of height ∆∆∆∆∆S inthe focal plane of L

2 will be focused in the plane of point

s with a height of (-i/o)∆∆∆∆∆S. Here i and o are the distancesof the lens from the image and object, respectively, andthe minus sign corresponds to the inversion of the image.As shown in Figure 3, reflection from the beam splitterforms a similar image of the same height.

Figure 2a: When MR is at angle θ, thelaser beam is reflected to point S on MF.

MF

S

θθ 2θ

Laser

θ

MR

Figure 2b: When MR is at angle θ1, thelaser beam is reflected to point S1 on MF.

θ∆θ

θ1 = θ+∆θ

θ+∆θ

θ+∆θ

2(θ+∆θ)

S1

Figure 2 a,b: The Reflection Point on MF

∆S∆s

∆S

D B A

∆s'

L2MR

MF

s1

s

Beamsplitter

Virtualimage of

MF

S1

S

S

S1

Figure 3: Analyzing the Virtual Images

Page 9: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A
Page 10: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

6

Speed of Light 012-07135B

OS-9262 Basic Speed of Light Apparatus

PASCO scientific

MODEL OS-9263HIGH SPEED ROTATING MIRROR

CAUTIONALLOW MOTOR TO STOP

BEFORE CHANGING DIRECTION

MIRROR ROTATION DIRECTION

CCW CW

STOP

STOP MOTORIF LIT MORETHAN 5 SEC.

PUSH FORMAX REV/SEC

60 SEC LIMIT

ADJUST

REV/SEC

0153

OS-9263A High Speed Rotating Mirror Assembly

What You Need to Measure the Speed ofLight

In order to measure the speed of light as described in thismanual, you will need all the items listed below (seeFigure 4). If you have an OS-9261 Complete Speed ofLight Apparatus, everything is included. If you have theOS-9262 Basic Speed of Light Apparatus or the OS-9263A High Speed Rotating Mirror, you will needadditional components, as listed, to make the measure-ment.

The OS-9261 Complete Speed of Light Apparatusincludes:

• OS-9262 Basic Speed of Light Apparatus, whichincludes:

– OS-9263A High Speed Rotating Mirror Assembly

– Fixed Mirror

– Measuring Microscope

The Equipment

• SE-9367 0.5 mW He-Ne Laser

• OS-9103 One-Meter Optics Bench

• OS-9172 Laser Alignment Bench

• OS-9142 Optics Bench Couplers

• OS-9133 Lens (48 mm FL)

• OS-9135 Lens (252 mm FL)

• OS-9109 Calibrated Polarizers (2)

• OS-9107 Component Holders (3)

• OS-8514 Laser Adapter Kit

• Alignment Jigs (2) – Part Number 648-02230

MeasuringMicroscope

Fixed Mirror

OS-9133 Lens (48 mm FL), andOS-9135 Lens (252 mm FL)

OS-9103 One-Meter Optics Bench

SE-9367 0.5 mW He-Ne Laser

OS-8514 Laser Adapter Kit

OS-9142 Optics BenchCouplers

OS-9109Calibrated

Polarizers (2)

OS-9107ComponentHolders (3)

AlignmentJigs (2)

Figure 4: Equipment Included with the OS-9261A Complete Speed of Light Apparatus

Page 11: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

7

012-07135B Speed of Light

About the Equipment

1. High Speed Rotating Mirror Assembly

The High Speed Rotating Mirror comes with its ownpower supply and digital display. The mirror is flat towithin 1/4 wavelength. It’s supported by high speed ballbearings, mounted in a protective housing, and drivenby a DC motor with a drive belt. A plastic lock-screwlets you hold the mirror in place during the alignmentprocedure.

An optical detector and the digital display providemeasurements of mirror rotation to within 0.1% or 1 rev/sec. The display and the controls for mirror rotation areon the front panel of the power supply. Rotation isreversible and the rate is continuously variable from 100to 1,000 rev/sec. In addition, holding down the MAXREV/SEC button will bring the rotation speed quickly toits maximum value at approximately 1,500 rev/sec.

➤➤➤➤➤ CAUTION: Before turning on the motor forthe rotating mirror, carefully read the cautionarynotices in the section of this manual entitled“Making the Measurement”.

2. Measuring Microscope

The 90X microscope is mounted on a micrometer stagefor precise measurements of the displacement of theimage point. Measurements are most easily made byvisually centering the image point on the microscopecross-hairs before and after the displacement. By notingthe change in the micrometer setting, the displacementcan be resolved to within 0.005 mm.

To focus the cross-hairs, slide the eyepiece up or down inthe microscope. To focus the microscope, loosen thelock-screw on the side of the mounting tube and slide themicroscope up or down within the tube.

With the lock-screw loosened, the microscope can also beremoved from the mounting tube. This can be helpfulwhen you are trying to locate the image point. A piece oftissue paper placed over the tube provides a screen thatallows you to view the point without focusing the micro-scope.

In addition to the microscope and micrometer, themicrometer stage also contains the beamsplitter. Thelever on the side of the stage is used to adjust the angle ofthe beamsplitter. When the lever points directly down,the beamsplitter is at a forty-five degree angle.

3. Fixed Mirror

The Fixed Mirror is a spherical mirror with a radius ofcurvature of 13.5 meters. It is mounted to a stand and hasseparate x and y alignment screws.

4. OS-9103 Optics Bench

The 1.0 meter long Optics Bench provides a flat, levelsurface for aligning the optical components. The bench isequipped with a one meter scale, four leveling screws,and a magnetic top surface. The "fence", a raised edge onthe back of the bench, provides a guide for aligningcomponents along the optical axis.

5. SE-9367 Laser with the OS-9172 Alignment Bench

The 0.5 mW, TEM00

mode, random polarization laser hasan output wavelength of 632.8 nm. The Alignment Benchattaches to the Optics Bench for precise, stable position-ing of the laser.

6. Alignment Jigs (2)

These jigs mount magnetically to the Optics Bench.Each has a 2 mm diameter hole that is used to align thelaser beam.

7. Optical Components

The use of the lenses and polarizers is described in theSetup and Alignment section of the manual.

Page 12: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

8

Speed of Light 012-07135B

Setup and Alignment

The following alignment procedure is tailored forthose using the OS-9261A Complete Speed of LightApparatus. For those using only some of the compo-nents in the complete system, the general procedure isthe same, though the details depend on the opticalcomponents used.

➤ IMPORTANT: Proper alignment is critical,not only for getting good results, but for gettingany results at all. Please follow this alignmentprocedure carefully. Allow yourself about threehours to do it properly the first time. Once youhave set up the equipment a few times, you mayfind that the alignment summary at the end ofthis section is a helpful guide.

For reference as you set up the equipment, Figure 5shows the approximate positioning of the componentswith respect to the metric scale on the side of the OpticsBench. The exact placement of each component dependson the position of the Fixed Mirror (M

F) and must be

determined by following the steps of the alignmentprocedure described below.

All component holders, the Measuring Microscope, andthe Rotating Mirror Assembly should be mounted flushagainst the “fence” of the Optics Bench (Figure 6). Thiswill insure that all components are mounted at rightangles to the beam axis.

Fence

Figure 6:Placing Components

Flush Against the Fencefor Proper Alignment

Figure 7: Coupling the Optics Bench and theLaser Alignment Bench

Laser and LaserAlignment Bench

Optics Bench

SideView

TopView

Four screwsincluded withBench Couplers

Bench Couplers

Four levelingscrews fromOptics Benchand LaserAlignmentBench (usetwo, save two)

17 cmLeveling Screw

62.2 cm

L2 (252 mmfocal length)

Optics Bench

L1 (48 mmfocal length) Laser

Laser AlignmentBench

Polarizers

93.0 cm

LevelingScrews

Figure 5: Equipment Alignment

MeasuringMicroscope

82.0 cm*

Rotating MirrorAssembly

(MR)

* Earlier units with themicroscope offset to the rightof center on its base should

be set at 81.0 cm.

Page 13: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

9

012-07135B Speed of Light

To Set up and Align the Equipment:

1. Place the Optics Bench on a flat, level surface.

2. Place the Laser, mounted on the Laser AlignmentBench, end-to-end with the Optics Bench, at theend corresponding to the 1-meter mark of the metricscale.

3. Use the Bench Couplers and the provided screws toconnect the Optics Bench and the Laser AlignmentBench. Details are shown in Figure 7. Do not yettighten the screws holding the Bench Couplers.➤ Note that the leveling screws must be removedfrom the Optics Bench and from the Laser Align-ment Bench to attach the Bench Couplers. Two ofthe removed leveling screws are then inserted intothe threaded holes in the Bench Couplers and areused for leveling.

4. Mount the Rotating Mirror Assembly on the oppositeend of the bench. Be sure the base of the assembly isflush against the fence of the Optics Bench and alignthe front edge of the base with the 17 cm mark on themetric scale of the Optics Bench (see Figure 8).

5. The laser must be aligned so the beam strikes the cen-ter of the Rotating Mirror (M

R). Two alignment jigs

are provided for this purpose. Place one jig at eachend of the Optics Bench as shown in Figure 8, withthe edges flush against the fence of the bench. Whenproperly placed, the holes in the jigs define a straightline that is parallel to the axis of the Optics Bench.

6. Turn on the Laser.

➤➤➤➤➤ CAUTION: Do not look into the laser beam,either directly or as it reflects from either mirror.Also, when arranging the equipment, be sure thebeam path does not traverse an area where someonemight inadvertently look into the beam.

7. Adjust the position of the front of the laser so thebeam passes directly through the hole in the firstjig. (Use the two front leveling screws to adjust theheight. Adjust the position of the laser on the LaserAlignment Bench to adjust the lateral position.)Then adjust the height and position of the rear ofthe laser so the beam passes directly through thehole in the second jig.

Figure 8: Using the Alignment Jigs to Align the Laser

17 cm

MR Alignment Jigs

Leveling Screws: Use to aim the laserbeam through the alignment jigs.

Figure 9: Aligning the Rotating Mirror (MR)

Hole in Alignment JigReflected laser beam

Paper

Page 14: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

10

Speed of Light 012-07135B

13. Place the Measuring Microscope on the Optics Benchso that the left edge of the mounting stage is aligned withthe 82.0 cm mark on the bench (see Figure 5). The leverthat adjusts the tilt of the beam splitter should be on thesame side as the metric scale of the Optics Bench.Position this lever so it points directly down.

➤➤➤➤➤ CAUTION: Do not look through the micro-scope until the polarizers have been placed betweenthe laser and the beam splitter in step 19.

The beamsplitter will slightly alter the position ofthe laser beam. Readjust L

2 on the Component

Holder so the beam is again centered on MR.

14. Place the Fixed Mirror (MF) from 2 to 15 meters

from MR, as shown in Figure 11. The angle between

the axis of the Optics Bench and a line from MR to

MF should be approximately 12 degrees. (If it is

greater than 20-degrees, the reflected beam will beblocked by the Rotating Mirror enclosure.) Also besure that M

F is not on the same side of the optical

bench as the micrometer knob, so you will be able tomake the measurements without blocking the beam.

➤➤➤➤➤ NOTE: Best results are obtained when MF is

10 to 15 meters from MR. See Notes on Accuracy

near the end of the manual.

8. To fix the laser in position with respect to the OpticsBench, tighten the screws on the Bench Couplers.Then recheck the alignment of the laser.

9. Align the Rotating Mirror. MR must be aligned so

that its axis of rotation is vertical and also perpen-dicular to the laser beam. To accomplish this, removethe second alignment jig and then rotate M

R so that

the laser beam reflects back toward the hole in thefirst alignment jig (Figure 9). Be sure to use the re-flective side of the mirror. It helps to tighten the lock-screw on the rotating mirror assembly just enough soM

R holds its position as you adjust its rotation.

If needed, use pieces of paper to shim between theRotating Mirror Assembly and the Optics Bench sothat the laser beam is reflected back through the holein the first jig.

10. Remove the first alignment jig.

11. Mount the 48 mm focal length lens (L1) on the Optics

Bench so that the center line of the ComponentHolder is aligned with the 93.0 cm mark on the met-ric scale of the bench. Without moving the Compo-nent Holder, slide L

1 as needed on the holder to cen-

ter the beam on MR (see Figure 10). Notice that L

1has spread the beam at the position of M

R.

12. Mount the 252 mm focal length lens (L2) on the Op-

tics Bench so the center line of the ComponentHolder aligns with the 62.2 cm mark on the metricscale of the bench. As for L

1 in step 11, adjust the

position of L2 on the Component Holder so that the

beam is again centered on MR.

Figure 10: Positioning and Aligning L1

93.0 cm

L1: Position L1 at 93.0 cm,then adjust its positionon the holder to centerthe beam on MR.

Rotating Mirror (MR)

Page 15: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

11

012-07135B Speed of Light

MeasuringMicroscope

Figure 11: Positioning the Fixed Mirror (MF)

MF

(Fixed Mirror)

L2 Beamsplitter L1s

MR

(Rotating Mirror)

♠ 12° Laser

15. Position MR so the laser beam is reflected toward

MF. Place a piece of paper in the beam path and

“walk” the beam toward MF, adjusting the rotation

of MR as needed.

16. Adjust the position of MF so the beam strikes it ap-

proximately in the center. Again, a piece of paper inthe beam path will make the beam easier to see.

17. With a piece of paper still against the surface of MF,

slide L2 back and forth along the Optics Bench to

focus the beam to the smallest possible point on MF.

18. Adjust the two alignment screws on the back of MF

so the beam is reflected directly back to the center ofM

R. This step is best performed with two people:

one adjusting MF, and one watching the beam posi-

tion at MR.

19. Place the polarizers (attached to either side of asingle Component Holder) between the laser and L

1.

Begin with the polarizers at right angles to eachother, than rotate one until the image in the micro-scope is bright enough to view comfortably.

If you can’t find the point image there are severalthings you can try:

• Vary the tilt of the beamsplitter slightly (no morethan a few degrees) and turn the micrometer knob tovary the transverse position of the microscope untilthe image comes into view.

• Loosen the lock-screw on the microscope. As shownin Figure 13, remove the microscope and place a pieceof tissue paper over the tube to locate the beam. Ad-just the beamsplitter angle and the micrometer knob tocenter the point image in the tube of the microscope.

• Slide the Measuring Microscope a centimeter or soin either direction along the axis of the Optics Bench.Be sure that the Microscope stays flush against thefence of the Optics Bench. If this doesn’t work, re-check the alignment, beginning with step 1.

20. Bring the cross-hairs of the microscope into focus bysliding the microscope eyepiece up and down.

21. Focus the microscope by loosening the lock-screwand sliding the scope up and down. If the appara-tus is properly aligned, you will see the point im-age through the microscope. Focus until the imageis as sharp as pos-sible.

Figure 13: Lookingfor the Beam Image

Lock-screw

Micrometerknob

Tissuepaper

Lever foradjusting

thebeamsplitter

angle

2-15 Meters

Page 16: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

12

Speed of Light 012-07135B

Alignment Summary

(see Figure 14 for approximate component placement)

This summary is for those who are familiar with theequipment and the experiment, and just need a quickreminder of the steps in the alignment procedure. If youhave not successfully aligned the apparatus before, werecommend that you take the time to go through thedetailed alignment procedure in the preceding section.

1. Align the laser so the laser beam strikes the center ofM

R (use the alignment jigs).

2. Adjust the rotational axis of MR so it is perpendicular

to the beam (i.e. as MR rotates, there must be a posi-

tion at which it reflects the laser beam directly backinto the laser aperture).

3. Insert L1 to focus the laser beam to a point. Adjust L

1so the beam is still centered on M

R.

4. Insert L2 and adjust it so the beam is still centered on

MR.

5. Place the Measuring Microscope in position and,again, be sure that the beam is still centered on M

R.

Rotating MirrorAssembly

(MR)

17 cmLeveling Screw

62.2 cm

L2 (252 mmfocal length)

Optics Bench

L1 (48 mmfocal length) Laser

Laser AlignmentBench

Polarizers

93.0 cm

Figure 14: Equipment Alignment

MeasuringMicroscope

82.0 cm*

* Earlier units with themicroscope offset to the rightof center on its base should

be set at 81.0 cm.

LevelingScrews

1 or 2°L2

➤➤➤➤➤ IMPORTANT: In addition to the point image,you may also see some extraneous beam imagesresulting, for example, from reflection of the laserbeam from L

1. To be sure you are observing the

right image point, place a piece of paper betweenM

R and M

F while you watch the image in the

microscope. If the point does not disappear, it is notthe correct image.

Cleaning Up the Image

22. In addition to the point image, you may also see in-terference fringes through the microscope (as well asthe extraneous beam images mentioned above). Thesefringes cause no difficulty as long as the point imageis clearly visible. However, the fringes and extraneousbeam images can sometimes be removed without los-ing the point image. This is accomplished by turningL2 slightly askew, so it is no longer quite at a rightangle to the beam axis (see Figure 12).

Figure 12: Turning L2 Slightly Askew to Clean Up the Image

Page 17: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

13

012-07135B Speed of Light

If the spot you need to measure is significantly off-center,you can move it by adjusting the angle of the beamsplitter.

���������� ��������������

Another common problem is a spot that is “stretched”with no easily discernible maxima. Check first to makesure that this is the spot you need by blocking thebeam path between the moving and fixed mirrors. If itis, then twist L

2 slightly until the image coalesces into

a single spot.

����������

�������� ��

Once the mirror begins to rotate, it is safe to look into themicroscope without the polarizers. You will notice thatyour carefully aligned pattern has changed: now the entirefield is covered with a random interference pattern, andthere is a bright band down the center of the field. Ignorethe interference pattern; there’s nothing you can do aboutit anyway. The band is the image of the laser when, onceeach rotation, the mirror reflects it into the microscopebeamsplitter. This is also unavoidable.

Your actual spot will probably be just to one side of thebright band. You can check for it by blocking andunblocking the beam path between the rotating mirror andfixed mirror and watching to see what disappears.

If you aligned everything perfectly, the spot will behidden by the bright band; in this case, make sure that youhave a spot when the rotating mirror is fixed and isreflecting the laser to the fixed mirror. If you do have thecorrect spot under stationary conditions, then misalign thefixed mirror very slightly (0.004° or less) around thehorizontal axis. This will bring the actual spot out fromunder the bright band.

➤➤➤➤➤ CAUTION: Do not look through the micro-scope until the polarizers have been placed betweenthe laser and the beamsplitter.

6. Position MF at the chosen distance from M

R (2 - 15

meters), so the reflected image from MR strikes the

center of MF.

7. Adjust the position of L2 to focus the beam to a point

on MF.

8. Adjust MF so the beam is reflected directly back onto

MR.

9. Insert the polarizers between the laser and the beamsplitter.

10. Focus the microscope on the image point.

11. Remove polarizers.

Alignment Hints

Once you have the microscope focused, it may still bedifficult to obtain a good spot. There may be several otherlights visible in the microscope besides the spot reflectedfrom the fixed mirror.

��������������� �����

������� ������������ ��

The most common of these are stray interferencepatterns. These are caused by multiple reflections fromthe surfaces of the lenses, and may be ignored. Ifnecessary, you may be able to eliminate them byangling the lenses 1 – 2°.

Stray Spots are most often caused by reflections off thewindow of the rotating mirror housing. To determinewhich spot is the one you must measure, block the beampath between the rotating mirror and the fixed mirror. Therelevant spot will disappear.

Page 18: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

14

Speed of Light 012-07135B

Making the Measurement

The speed of light measurement is made by rotating themirror at high speeds and using the microscope andmicrometer to measure the corresponding deflection ofthe image point. By rotating the mirror first in onedirection, then in the opposite direction, the total beamdeflection is doubled, thereby doubling the accuracy ofthe measurement.

➤➤➤➤➤ Important—to Protect the Rotating MirrorAssembly:

– Before turning on the motor, be sure the lock-screw for the rotating mirror is completely loos-ened, so the mirror rotates freely by hand.

– Whenever the speed of the motor is accelerated,the red LED on the front panel of the motor con-trol box will light up. As the speed stabilizes, thislight should go off. If it does not, turn off the mo-tor. Something is interfering with the motor rota-tion. Check to be sure the lock-screw for M

R is

fully loosened.

– Never run the motor with the MAX REV/SECbutton pushed for more than one minute at a time,and always allow about a minute between runs forthe motor to cool off.

1. With the apparatus aligned and the beam image insharp focus (see the previous section), set the directionswitch on the rotating mirror power supply to CW,and turn on the motor. If the image was not in sharpfocus, adjust the microscope. You should also turn L

2slightly askew (about 1 - 2°) to improve the image. Toget the best image you may need to adjust the micro-scope and L

2 several times. Let the motor warm up at

about 600 revolutions/sec for at least 3 minutes.

2. Slowly increase the speed of rotation. Notice how thebeam deflection increases.

3. Use the ADJUST knob to bring the rotational speedup to about 1,000 revolutions/sec. Then push theMAX REV/SEC button and hold it down. When therotation speed stabilizes, rotate the micrometer knobon the microscope to align the center of the beam im-age with the cross hair in the microscope that is per-pendicular to the direction of deflection. Recordthe speed at which the motor is rotating, turn offthe motor, and record the micrometer reading.

MF

(Fixed Mirror)

MR

(Rotating Mirror)

L2 Beamsplitter L1s

Laser

MeasuringMicroscope

Figure 15: Diagram of the Foucault Method

Page 19: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

15

012-07135B Speed of Light

➤➤➤➤➤ NOTE:

When reversing the direction of movement ofthe micrometer carriage, there will always besome movement of the micrometer knob beforethe carriage responds. Though this source oferror is small, it can be eliminated. Just adjust theinitial position of the micrometer stage so thatyou always turn the micrometer knob in thesame direction as you adjust it.

Reverse the direction of the mirror rotation byswitching the direction switch on the power supplyto CCW. Allow the mirror to come to a completestop before reversing the direction. Then repeatyour measurement as in step 3.

➤ NOTES:

-When the mirror is rotated at 1,000 rev/sec ormore, the image point will widen in the directionof displacement. Position the microscopecross-hair in the center of the resulting image.

- The micrometer on the Measuring Microscopeis graduated in increments of 0.01 mm for thebeam deflections.

5. The following equation was derived earlier in themanual:

When adjusted to fit the parameters just measured,it becomes:

Use this equation, along with the diagram in Figure15, to calculate c, the speed of light. (To measureA, measure the distance between L

1 and L

2, then

subtract the focal length of L1, 48 mm.)

c = 4AD2ω(D + B) ∆s′

c = 4AD2(ωcw – ωccw)(D + B)(s′ cw– s′ ccw)

➤ NOTES: This equation is the same as the original equation

in step 5, but with two differences:

The rotational velocity is expressed in rad/s.

The CCW rotational velocity is expressed as anegative number, reflecting the direction of ro-tation.

c = 8πAD2(Rev/seccw + Rev/secccw )(D + B)(s′ cw – s′ ccw )

Page 20: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

16

Speed of Light 012-07135B

Accuracy

Precise alignment of the optical components and carefulmeasurement are, of course, essential for an accuratemeasurement using this equipment. Beyond this, themain factor affecting accuracy is the distance between thefixed and rotating mirrors.

As mentioned in the alignment procedure, the optimumdistance between M

R and M

F is from 10 to 15 meters.

Within this range, accuracy within 5% is readily obtain-able. If space is a problem, the distance between themirrors can be reduced to as little as 1 meter and propor-tional reduction in accuracy will result.

In general, longer distances provide greater accuracy. MR

rotates farther as the light travels between the mirrors, andthe image deflection is correspondingly greater. Greaterdeflections reduce the percentage of measurement error.

However, the optical components are designed foroptimal focusing of the image point at 13.5 meters (this isthe radius of curvature of M

F). Image focusing is not a

significant problem as long as the distance between themirrors is within about 15 meters. At larger distances theintensity and focus of the image point begins to drop, andmeasurement and alignment are hampered.

Typical sample data taken in our lab gives values for cthat are within 1.5 - 2.5% of accepted values.

Notes on Accuracy and Maintenance

Maintenance

Regular maintenance for this equipment is minimal. Themirrors and lenses should be cleaned periodically.

➤ IMPORTANT: All mirrors and lenses maybe cleaned with lens tissue, except the sphericalmirror (M

F). It has a delicate aluminized front

surface and should only be cleaned with alcoholand a soft cloth. Do not use any cleaning com-pound that contains ammonia (such as Windex);the ammonia will attack the aluminum surface.

If problems arise with the rotating mirror assembly, suchas a broken drive belt, notify PASCO scientific. We donot recommend that you attempt to fix this equipmentyourself. (See the warranty and equipment return infor-mation at the front of this manual.)

Page 21: SPEED OF LIGHT APPARATUS - Astronomy Speed of Light Apparatus i SPEED OF LIGHT APPARATUS Instruction Manual and Experiment Guide for the PASCO scientific Model OS-9261A, 62 and 63A

17

012-07135B Speed of Light

Technical Support

Feedback

If you have any comments about the product or manual,please let us know. If you have any suggestions onalternate experiments or find a problem in the manual,please tell us. PASCO appreciates any customerfeedback. Your input helps us evaluate and improve ourproduct.

To Reach PASCO

For technical support, call us at 1-800-772-8700 (toll-freewithin the U.S.) or (916) 786-3800.

fax: (916) 786-3292

e-mail: [email protected]

web: www.pasco.com

Contacting Technical Support

Before you call the PASCO Technical Support staff, itwould be helpful to prepare the following information:

➤ If your problem is with the PASCO apparatus,note:

- Title and model number (usually listed on thelabel);

- Approximate age of apparatus;

- A detailed description of the problem/sequence ofevents (in case you can’t call PASCO right away,you won’t lose valuable data);

- If possible, have the apparatus within reach whencalling to facilitate description of individual parts.

➤ If your problem relates to the instruction manual,note:

- Part number and revision (listed by month and yearon the front cover);

- Have the manual at hand to discuss yourquestions.