Upload
vuthuan
View
218
Download
2
Embed Size (px)
Citation preview
1
Particulate Matter Counter
Alankar Gupta
B.TECH. – 4th year, Engineering Physics
IIT Guwahati
Arul Goutham
B.TECH. – 4th
year, Electricals and Electronics Engineering
NIT Tiruchirapalli
Project Mentor:
Prof. Navakant Bhat
Department of Electronics and Communication Engineering
Indian Institute of Science, Bangalore July - 2011
2
ACKNOWLEDGEMENT
We wish to express our deep gratitude to our project mentor, Prof. Navakant Bhat for
providing us an opportunity to work under his expert guidance. We would also like to thank
Mr. Amit Kumar Gupta for guiding and directing us throughout the project work at each
and every possible step.
A debt of gratitude to Mr.Dwarkanath and Mrs. Subhashini for their assistance and support
during the course of our project.
The work mentioned in this report was carried out at Characterization Lab - II,
Microelectronics and Photonics Building, Department of Electronics and Communication
(ECE), IISc- Bangalore.
- Alankar and Arul
3
Contents
List of figures 4
Particulate Matter 5
Scattering Theories 6
Proposition on how to count particulate matter 11
Experimental Setup 12
Opacity experiment: 13
90 deg scattering: 14
Details of Photodiode and Laser diode used 18
Numerical Method: 19
Observations and Inferences 23
Proposed calibration method: 26
Conclusions 27
References 28
4
List of figures
1. Light scattering by an induced dipole moment due to an incident EM wave
2. Schematic showing scattering angle w.r.t incident beam
3. Normalized scattering pattern of particles of various sizes in Log scale
4. Setup showing Laser Diode (bottom), sampling tube (center) and Photodiode (left).
5. Setup while calibrating the size of laser beam falling on the sampling area.
6. Rough Schematic of opacity experiment
7. Rough schematic of 90 deg scattering experiment
8. Voltage (v) of PD Vs Samples taken in the absence of blower at 90deg to sample area.
9. Setup with the new Exhaust mechanism – consists of a funnel arrangement and
blower.
10. Voltage of PD Vs samples taken in presence of blower run at lower speed.
11. Voltage of PD Vs samples taken in presence of blower run at higher speed.
12. APC LD ADL-65075TA2 circuit with POT attached to terminal 3.
13. PD OPT101 circuit used to detect scattering.
14. Chi-square method to determine particle size
5
Particulate Matter
Particulates – also known as particulate matter (PM), fine particles, and soot – are tiny
subdivisions of solid matter suspended in a gas or liquid. In contrast, aerosol refers to
particles and/or liquid droplets and the gas together. Sources of particulate matter can be
manmade or natural. Air pollution and water pollution can take the form of solid particulate
matter, or be dissolved [1].Salt is an example of a dissolved contaminant in water, while sand
is generally a solid particulate.
Classification and Composition
PM is classified according to the aerodynamic diameter, which is a physical property of a
particle in a viscous fluid such as air. In general, particles have irregular shapes with actual
geometric diameters that are difficult to measure. Aerodynamic diameter is an expression
of a particle's aerodynamic behaviour as if it were a perfect sphere with unit-density and
diameter equal to the aerodynamic diameter. Such a model has the same terminal settling
velocity.
Fraction Size range
PM10 (thoracic fraction) <=10 μm
PM2.5 (respirable fraction) <=2.5 μm
PM1 <=1 μm
Ultrafine (UFP or UP) <=0.1 μm
PM10-PM2.5 (coarse fraction) 2.5 μm – 10 μm
Table. 1 Particulate Matter classification
Coarse particles (PM10) have an aerodynamic diameter between 2.5µ m and 10µ m. They are
formed by mechanical disruption (e.g. crushing, grinding, abrasion of surfaces); evaporation
of sprays, and suspension of dust. PM10 is composed of aluminosilicate and other oxides of
crustal elements, and major sources including fugitive dust from roads, industry, agriculture,
construction and demolition, and fly ash from fossil fuel combustion. The lifetime of PM10 is
from minutes to hours, and its travel distance varies from <1km to 10 km.
6
Fine particles have an aerodynamic diameter less than 2.5µ m (PM2.5). They differ from
PM10 in origin and chemistry. These particles are formed from gas and condensation of high-
temperature vapours during combustion, and they are composed of various combinations of
sulfate compounds, nitrate compounds, carbon compounds, ammonium, hydrogen ion,
organic compounds, metals (Pb, Cd, V, Ni, Cu, Zn, Mn, and Fe), and particle bound water.
The major sources of PM2.5 are fossil fuel combustion, vegetation burning, and the smelting
and processing of metals. Their lifetime is from days to weeks and travel distance ranges
from 100s to >1000s km. In addition, fine particles are associated with decreased visibility
(haze) impairment in many cities of the U.S.
Physical properties and physiological effects
The capacity of particulate matter to produce adverse health effects in humans depends on its
deposition in the respiratory tract. Particle size, shape, and density affect deposition rates.
The most important characteristics influencing the deposition of particles in the respiratory
system are size and aerodynamic properties. Particles between 2.5 and 10µ m in aerodynamic
diameter correspond to the inhalable particles capable to be deposited in the upper respiratory
tract. Particles with aerodynamic diameter smaller than 2.5µ m (PM2.5) correspond to the
respirable particle fraction capable of penetrating the alveolar region of the lung. Inhaled
particles come in contact with surface of the respiratory system. These particles pass the
proximal airway (throat and larynx) of the respiratory tract, and deposit in the
tracheobronchial conductive airway of the lungs (bronchial and bronchiolar airway) or in the
gas exchange region (respiratory bronchioles, alveolar ducts, and alveoli of the lung
parenchyma).
Scattering theories
The scattering of light may be thought of as the redirection of light that takes place when an
electromagnetic (EM) wave (i.e. an incident light ray) encounters an obstacle or non-
homogeneity, in our case the scattering particle. As the EM wave interacts with the discrete
particle, the electron orbits within the particle’s constituent molecules are perturbed
periodically with the same frequency (νo) as the electric field of the incident wave. The
oscillation or perturbation of the electron cloud results in a periodic separation of charge
within the molecule, which is called an induced dipole moment. The oscillating induced
dipole moment is manifest as a source of EM radiation, thereby resulting in scattered light.
The majority of light scattered by the particle is emitted at the identical frequency (νo) of the
7
incident light, a process referred to as elastic scattering. In summary, the above comments
describe the process of light scattering as a complex interaction between the incident EM
wave and the molecular/atomic structure of the scattering object; hence light scattering is not
simply a matter of incident photons or EM waves “bouncing” off the surface of an
encountered object.
Fig. 1 Light scattering by an induced dipole moment due to an incident EM wave
Formal light scattering theory may be categorized in terms of two theoretical frameworks.
One is the theory of Rayleigh scattering (after Lord Rayleigh) that is, strictly speaking as
originally formulated, applicable to small, dielectric (non-absorbing), spherical particles. The
second is the theory of Mie scattering (after Gustav Mie) that encompasses the general
spherical scattering solution (absorbing or non-absorbing) without a particular bound
onparticle size. Accordingly, Mie scattering theory has no size limitations and converges to
the limit of geometric optics for large particles. Mie theory, therefore, may be used for
describing most spherical particle scattering systems, including Rayleigh scattering.
However, Rayleigh scattering theory is generally preferred if applicable, due to the
complexity of the Mie scatteringformulation [3]. The criteria for Rayleigh scattering is that
α<<1 and |m|<<1, where α is the dimensionless size parameter given by the expression
where a is the spherical particle radius, and λ is the relative scattering wavelength defined as
8
where λo is the incident wavelength with respect to vacuum, and mo represents the refractive
index of the surrounding medium. Finally, m is the refractive index of the scattering particle,
and is commonly represented by the complex notation defined as
In this notation, n indicates the refraction of light (i.e. n equals the speed of light in vacuum
divided by the speed of light in the material), while the complex term is related to absorption.
The commonly used absorption coefficient of the material (cm-1
) is related to the complex
part of the refractive index via the relation
It is noted that the value of k is never exactly zero for any material, but materials with a
value approaching zero are termed dielectrics. The magnitude of the refractive index, |m| , as
needed for the Rayleigh criteria, is given by the expression
The Rayleigh criteria as related above, namely α<<1 and lml<<1, correspond physically to
the assumptions that the particle is sufficiently small such that the particle encounters a
uniform electric field at any moment, accordingly the time for penetration of the electric field
is much less than the period of oscillation of the EM wave. In our case mo = 1 (air)
For our Particle size calculation the size parameter ranges from 48- 12 (for PM10 and PM2.5 ).
Hence Rayleigh criteria is not applicable and we are bound to use the complete MIE theory.
MIE Theory is a pretty complex mathematical theory describing the complete profile of
scattering but has some essential condition which needs to be fulfilled –
Incident light of only a single wavelength is considered.
No dynamic scattering effects are considered.
The scattering particle is isotropic.
There is no multiple scattering.
9
All particles are spheres.
All particles have the same optical properties.
Light energy may be lost to absorption by the particles.
For developing the algorithm for our particle size calculation we will be only touching the
essential parts of the Mie theory which is helpful in serving our purpose. We will not go into
details of calculation of scattering cross sections and many more unnecessary things.
1. Calculation of the complex Mie coefficients -
The Ricatti-Bessel functions Ψ and ξ are defined in terms of the half-integer-order Bessel
function of the first kind (Jn+1/2 (z)), where
The parameter ξn is given by the expression,
where Hn+1/2(z) is the half-integer-order Hankel function of the second kind, where the
parameter Xn is defined in terms of the half-integer-order Bessel function of the second kind,
Yn+1/2(z), Namely
2. With the help of Mie coefficients calculated above we can further calculate Mietheory
scattering amplitudes,
10
Where Ɵ is the angle between the incident ray and the scattered ray called the scattering
angle.
Fig. 2 Schematic showing scattering angle w.r.t incident beam
The values of an and bn are same as defined earlier and the angular dependent functions πn and
τn are expressed in terms of the Legendre polynomials by
Now For a randomly polarized light source, the total scattered light intensity is given by the
term S11:
So now in the end we got S11 as a function of Ɵ (scattering angle). In this derivation of S11 we
are not bothered about the exact magnitude rather we are more concerned about the relative
distribution across the various Ɵ values. So we can just normalize the scattering profile with
some known value and then compare the scattering pattern for various particle size.
11
Proposition on how to count particulate matter
In order to determine the concentration of PM in air sample, we first need to determine the
size of particle. From the scattering theories, it is clear that scattering pattern of light largely
depends on shape, size and optical properties of the particle scattering light. We could very
well assume PM in air (like smoke) to be a sphere of fixed refractive index [2]. With this
assumption, the parameter affecting light scattering is the size of the particle. As explained
earlier, we can obtain Mie pattern for various particle sizes theoretically by employing the
rigorous mathematical equations developed above.
Diameter 1µm Diameter 2.5µm
Diameter 10µm Diameter 20µm
Fig. 3 Normalized scattering pattern of particles of various sizes in Log scale
We can also obtain scattering pattern experimentally by measuring intensity at various angles
around the sampling area. This intensity pattern needs to be normalised for comparison with
the theoretical pattern. We then compare both the patterns for resemblance and hence
establish the particle size from it. This would require some kind of arrangement for
12
measuring intensity at multiple angles but later in this report we have proposed a way to
eliminate this need.
Once we have determined the particle size, we can proceed to find the concentration of the
same. For this we propose to use a constant volume pump which delivers quite stable flow
rate of air. The density of PM in air is usually less and we can obtain single particle scattering
all the time. We get pulses of light and we can categorise them by proper matching. With the
fixed flow rate, we can also get the volume of air flowing or which is being sampled. With
these two data in hand, we can easily obtain the type of particle and its concentration in air.
The variations in intensity are very close to each other for angles close to each other. Even a
slight error in our setup or misalignment in detector optics can make our result totally wrong.
This drawback can be overcome by the use of mirror. The idea is to place our light sensor at
a fixed angle and then use an elliptical mirror which will collect scattered light over an angle
range and focus it to our sensor hence giving us a sum of amplitudes within that angle range.
We can calibrate this setup for various reference particle sizes and hence obtain the count as
well as its size.
Experimental Setup
A rudimentary experimental setup was designed to test our proposition on determining
particle size. To test our proposition we first needed to establish two points – 1. We should be
able to detect light scattering using the equipments we had in hand. 2. We should be able to
detect various densities of smoke using them. On establishing these two points, we can
proceed to determine particle size by our proposition using a better design.
13
Fig. 4 Setup showing Laser Diode (bottom), sampling tube (center) and Photodiode (left).
Fig. 5 Setup while calibrating the size of laser beam falling on the sampling area.
Opacity experiment:
The Photodiode (PD) was placed in line with the collimating lens and Laser diode (LD).
Without the absence of sample, the PD reads high voltage. On introduction of sample
(Incense smoke), the particle scatters light in all directions thus reducing the intensity of light
falling on photodiode.
14
Fig. 6 Rough Schematic of opacity experiment
We used photodiode OPT101 and APC laser diode ADL-65075TA2. We met with success in
this experiment. The major problem in this experimental setup is that the PD quickly reaches
saturation and changes in voltage with scattering are difficult to observe. We needed to
constantly tune the power output of LD to obtain observable changes in PD voltage.
90 deg scattering:
Fig. 7 Rough schematic of 90 deg scattering experiment
We next placed the PD 90 deg to the sampling area. This avoids several problems with the
initial experiment- 1. PD doesn’t reach saturation. 2. LD can be operated at high power
(intense beam). 3. Multiple interactions of smoke particles with the light beam is avoided.
Without the absence of sample, the PD read a low voltage. This could be attributed to the
stray light falling on PD. It can also be observed from fig.2 on the left and right wall of the
cardboard box some reflections of laser beam. This can only be avoided by using a
15
slit/cylinder kind of arrangement in front of the collimating lens. On introducing sample, the
PM scatters light in all directions and we could observe a rise in voltage of PD.
Fig. 8 Voltage (v) of PD Vs Samples taken in the absence of blower at 90deg to sample area.
From the plot we could infer that,
1. There are too many oscillations in the voltage of PD (sampled every 0.5 sec) due to
the unsteady flow of smoke through the sample area.
2. We can clearly differentiate between the absence and presence of smoke with proper
sampling mechanism.
3. We might also need to sample the voltage of PD at a faster rate. It would be better to
use an oscilloscope than a multimeter.
4. We needed to reduce oscillations in voltage before measuring intensities at various
positions for particle size measurement.
We next fitted a exhaust mechanism to our present setup to have a steady smoke flow through
the sample area.
0 10 20 30 40 50 60 70 80 90 1000.4
0.6
0.8
1
1.2
1.4
1.6
1.8
2
samples
Voltage v
16
Fig. 9 Setup with the new Exhaust mechanism – consists of a funnel arrangement and blower.
The blower used had only two levels of speed. Repeating the experiment at the two speed ,
we were able to get improved results, lesser oscillations in voltage.
Fig. 10 Voltage of PD Vs samples taken in presence of blower run at lower speed.
0 10 20 30 40 50 60 70 80 90 100
0.16
0.18
0.2
0.22
0.24
0.26
0.28
samples
Voltage (
V)
17
Fig. 11 Voltage of PD Vs samples taken in presence of blower run at higher speed.
From these plots,
1. We could see the reduction in oscillations of PD voltage. From fig.6 the oscillation is
about 0.04-0.06 V whereas in fig.7 oscillations were around 0.4 V.
2. We could clearly differentiate between the presence and absence of smoke at all
instants except for few when the voltage drops to base value.
We used photodiode OPT101 and APC laser diode ADL-65075TA2 to obtain these plots.
0 10 20 30 40 50 60 70 80 90 100
0.35
0.4
0.45
0.5
0.55
0.6
samples
Voltage (
v)
18
Details of Photodiode and Laser diode used
Laser diode ADL-65075TA2 comes with automatic power control. By varying the resistance
of TrimPot attached to terminal 3 of LD we can get various output power from LD.
Fig. 12 APC LD ADL-65075TA2 circuit with POT attached to terminal 3.
We used a 10 K ohm POT. Usually operated at 4-5 K ohm. The maximum output power with
this LD is 10 mW. For more intense (power) beam we used Sanyo red laser diode DL-6147-
040 with Edmund optics laser driver demo module #56-805 and Edmund optics driver chip
#56-804. We could achieve a power output of 40 mW with this setup.
Photodiode we used in all our experiments were OPT101 Ic. We occasionally used
Hamamatsu Photodiode which has more sensitivity (in the nW range) as compared to
OPT101 (in the micro watt range). This was due to fact that the oscillations we initially
obtained with the setup was too high to use a more sensitive PD (it would just give more
voltage fluctuation).
19
Fig. 13 PD OPT101 circuit used to detect scattering.
We tried with different Rext and Cext values to improve the responsivity of PD once the
oscillations have been reduced.
We also carried out the experiment with different beam widths at sample area.
Numerical Method:
As explained in our proposition, we need to match the theoretical and experimental graphs in
order to determine the particle size. The necessity to measure at different angles before going
for single angle measurement has already been explained in the calibration topic (page: ).
There are various numerical techniques to measure the degree of closeness of a curve with a
set of standard curves. The method we have used is Chi-square method [4]. By this method,
20
we generate the theoretical values t various angles. We also feed the experimental values
measured at the same angles. The program compares the two values using the formula,
(16)
The minimum value of x2 gives the best fit theoretical curve for the experimental data and
hence the particle size. The transfer function for converting the voltage output of PD to
intensity of light is a linear and can be obtained from the plots provided in the corresponding
PD datasheets.
In order to calculate theoretical Mie plot, we need Mie coefficients an and bn values. We do
so by using equations (1) - (7). We used in-built Bessel functions available in matlab. The
initial conditions n=0 has been taken care [5]. We calculated it using recursive functions.
function coeff=zerocool_anbn(x,m) % to calculate coefficients an,bn
inputs: x- size parameter & m- complex refractive index
nmax=round(x+2+(4*(x)^(1/3))); % calculates max. n req for an &
bn n=1:1:nmax;frac=(n+0.5);c=m*x; sqx=sqrt(pi*0.5/x);sqc=sqrt(pi*0.5/c);
bessjx=besselj(frac,x).*sqx; %calculates spherical bessel
functions of first kind & second kind bessjc=besselj(frac,c).*sqc; bessyx=bessely(frac,x).*sqx;
hx=bessjx+i*(bessyx); %defines hankel functions
bessjxl=[(sin(x))/x,bessjx(1:nmax-1)]; % bessel functions for
order=0 to order=nmax-1 bessjcl=[(sin(c))/c,bessjc(1:nmax-1)]; bessyxl=[(-cos(x))/x,bessyx(1:nmax-1)];
derx=x.*bessjxl-n.*bessjx; %derivative of bessel
functions expanded using recursive formula derc=c.*bessjcl-n.*bessjc;
hxd=bessjxl+i*(bessyxl); derh=x.*hxd-n.*hx;
m2=m*m;
an=(m2.*bessjc.*derx-bessjx.*derc)./(m2.*bessjc.*derh-hx.*derc); %
an bn=(bessjc.*derx-bessjx.*derc)./(bessjc.*derh-hx.*derc);
%bn
coeff=[an;bn];
21
We next calculated angular dependent functions πn and τn are using Legendre polynomials.
function angular=zerocool_pitau(u,nmax) %calculates angular
dependent functions pi n and tau n inputs: u= cos(theta) and nmax
%p and t are for pi n and tau n respectively. p(1)=1; p(2)=3*u;t(1)=u;t(2)=6*u*u-3; %defines the initial
conditions for pi n and tau n for n=3:1:nmax pn1=((2*n-1)/(n-1)).*p(n-1); pn2=(n/(n-1)).*p(n-2); p(n)=pn1.*u-pn2; % pi n calculated
using recursive formula
tn1=n.*p(n); tn2=(n+1).*p(n-1); t(n)=tn1.*u-tn2; %tau n calculated
using recursive formula end angular=[p;t];
We use the above two functions to obtain scattering amplitudes S1 and S2.
function s=zerocool_s1s2(m,x,u) % calculates scattering
amplitudes s1 & s2 inputs: complex refractive index % size parameter and
u=cos(theta) theta- scattering angle temp=zerocool_anbn(x,m); % calls an & bn a=temp(1,:); b=temp(2,:);
nmax=round(x+2+(4*(x)^(1/3))); temp1=zerocool_pitau(u,nmax); % calls pi n &
tau n p=temp1(1,:); t=temp1(2,:);
s1=0; s2=0; for n=1:1:nmax s1=s1+((2*n+1)/(n*(n+1)))*(a(n).*p(n)+b(n).*t(n)); %
calculates scattering amplitude - perpendicular s2=s2+((2*n+1)/(n*(n+1)))*(a(n).*t(n)+b(n).*p(n)); %
calculates scattering amplitude - parallel
end s=[s1;s2];
We then obtained the intensity at various angles.
function raiden=zerocool_main(m,x,angledata)
%calculates intensity function and plots intensity versus scattering angle %inputs: complex
refractive index x- scattering parameter step- step size
22
n=0;
for i=1:1:length(angledata) n=n+1; temp=zerocool_s1s2(m,x,cos(angledata(i)*pi/180)); s1=temp(1,:); s2=temp(2,:); s12=s1*s1'; s22=s2*s2'; I1(n)=s12; % calculates magnitude of
s1 square I2(n)=s22; % calculates
magnitude of s2 square end I1=I1./max(I1); %theta=0:stepsize:pi; %sum=0; %for n=390:1:612 % sum=sum+(I1(n)+I2(n))./2; %end %plot(theta*(180/pi),I1); % plots I1 versus
theta %sum %polar(theta,I1); %plot(theta*(180/pi),I2); raiden=[I1];
We then employ our Chi-square method to obtain the best fit curve.
function raiden=zerocool_main(m,x,angledata)
%calculates intensity function and plots intensity versus scattering angle %inputs: complex
refractive index x- scattering parameter step- step size
n=0;
for i=1:1:length(angledata) n=n+1; temp=zerocool_s1s2(m,x,cos(angledata(i)*pi/180)); s1=temp(1,:); s2=temp(2,:); s12=s1*s1'; s22=s2*s2'; I1(n)=s12; % calculates magnitude of
s1 square I2(n)=s22; % calculates
magnitude of s2 square end I1=I1./max(I1); %theta=0:stepsize:pi; %sum=0; %for n=390:1:612 % sum=sum+(I1(n)+I2(n))./2; %end %plot(theta*(180/pi),I1); % plots I1 versus
theta
23
%sum %polar(theta,I1); %plot(theta*(180/pi),I2); raiden=[I1];
We tested this program by feeding data generated (by means of another program) for various
particle sizes (approx) and checked if the program is able to return a close value.
Fig. 14 Chi-square method to determine particle size
The program was quite accurate and its accuracy greatly dependent on the step size entered
for the size parameter in the program. The smaller the step size, the greater is the accuracy
but more run-time required.
Observations and Inferences
Mie scattering:
Mie scattering assumes single particle scattering i.e. incident light is scattered
by a single sphere and detected. No multiple scatterings are allowed.
Experimentally this can be achieved by having thin volume of air in the path of
24
beam. In lab, as we are using incense sticks (which produces dense smoke) we
have to use a variable speed pump to draw it slowly. Practically in the field,
particle concentration is going to be low and it can be fairly assumed to be
single particle scattering. This precaution ensures that the scattering curve we
get is just a multiple of the actual Mie curve.
For size parameter between 0.1<x<100 we use Mie scattering for experiments.
Photodiode:
Keeping a mirror to collect light around 70 to 90 degrees (or in the higher
ranges) could help differentiate particles of 10µm and 2.5 µm. This is due to the
fact that 2.5 µm scatters light in the higher degrees too. The comparison can be
made only upon normalisation if the highest intensities (intensity of incident
beam) are different (usually at 0 deg). The mention of angles here are just to
give an idea. Actual angle needs to be determined. GRIMM also claims that by
proper placing of this mirror any undulations in Mie scattering can be
eliminated to some degree.
It is better to have the window of the photodiode as small as possible. This
ensures high resolution. Hamamatsu photodiode has a smaller window than the
IC version. Further improvement in resolution can also be achieved by closing
the aperture of the diode.
The photodiode can be placed further (not too far) from the point where beam of
light incidents on the particle. This can ensure that laser beam which tends to
diverge (from the datasheet) doesn’t fall on the detector and we observe only the
scattered light. A second collimating lens could also be used for focusing the
beam.
Filters can be used along with photo detectors to remove stray light. It can also
prevent the change in colour (or wavelength) due to motion of the smoke
particles (dynamic light scattering).
The Photo diode should be fast enough to detect pulse of light as it happens. It
is advisable to use an oscilloscope in the measurement of pulses of voltage by
diode as it has a high sampling frequency. In the actual product, we might need
a fast ADC to sample the voltage and provide for further processing.
25
A sensitive photodiode should be used. This is necessary as practically the no
of particles in air would be very less and hence the intensity of scattered light.
This again entails use of amplifiers for significant voltage output.
An array of photodiodes may be used for detecting the scattering pattern.
Charge coupled device may be used in place of normal photodiode.
We faced problem of saturation of photodiode on exposure to unscattered
beam. This might occur even if the photodiode is in region of maxima of the
Mie pattern. Two adjustments can be done regarding this: 1. Adjust the
sensitivity of the diode by adding resistors and capacitors (refer datasheet), 2.
The laser diode light can be attenuated.
o It could also be caused by reflection of laser beam from the walls of the
container. Hence a slit kind of arrangement can be placed in front of the
collimating lens of laser diode to prevent any reflection.
Laser diode:
We placed problem of saturation of photodiode on exposure to unscattered
beam. This might occur even if the photodiode is in region of maxima of the
Mie pattern. Two adjustments can be done regarding this: 1. Adjust the
sensitivity of the diode by adding resistors and capacitors (refer datasheet), 2.
The laser diode light can be attenuated.
Constant power of the laser diode can be ensured by using the driver demo
module. Multiplexing of laser diode with two power levels might be required
depending on the size of the particle to be measured.
Polarisers can be used for perpendicular and parallel polarisations.
Sampling:
A fairly wide beam (wide here means some small multiple of actual particle
size) of light can be made to incident on the sample. This ensures parallel rays
of light are incident on each particle.
Steady flow of smoke would be preferable. This in turn could help in
calculating no of particles in certain sample of air using no of pulses.
Stray light needs to be prevented. Cover the entire setup. Even optical fibre
could be used to introduce light on the sample keeping the laser diode out.
26
We need not consider the bright spot formed at 0 deg for particles of all size.
Our region of interest should be around the region where light intensity
decreases rapidly. (90 deg). This is the region which shows considerable
changes (oscillations) depending on particle size.
Humidity should be low. Preheating of the sampling air can improve accuracy.
Numerical Methods:
Advanced methods such as Monte Carlo and genetic algorithms are used for
solving the scattering pattern. The method we have used is a simple one and
hence introduces some amount of error in the theoretical calculation of
intensity.
There might be light reflected from the interior of the particle. This might
undergo a change in angle while being received at the detector due to change in
refractive index. This might require correction to be made for theoretical Mie
plot. This is not considered in our program.
Proposed calibration method:
o With the initial setup measurements of intensity can be taken at various
angles and fitted with the Mie curve to determine particle size. This requires
setup of the pump system for steady flow, distance of detector from sample,
lens in front of detector, adjustment of collimating lens for proper size of
beam, adjustment of intensity of laser diode, filters (if necessary) and
attenuation (if necessary). This will ensure that our setup is capable of
producing Mie pattern.
o Later the detector can be fixed at particular angle (90 deg) and calibrated
using test particles and thresholds can be set for each particle. These
thresholds can then be programmed in microprocessor of the actual device.
27
o Why this can’t be done at the first place?
It needs to be ensured that the setup (here I mean the various
parameters mentioned above) are right for Mie scattering detection.
It should be made sure that the detector detects only the scattered
light from the sample and not something else.
Conclusions
In future we can follow the suggestions we have mentioned above to obtain an accurate Mie
plot and hence determine the particle size. By replacing the exhaust mechanism with a
constant volume pump, we can eliminate oscillations in PD voltage and also determine the
concentration of PM in air.
28
References:
1. Light Scattering Theory – David W.Hahn – Department of Mechanical and
Aerospace,
University of Florida. ([email protected])
2. George W. Mulholland, Raymond L. McKenzie, Egon Marx, and
Robert A. Fletcher - Refractive Index and Evaporation Rate of Individual Smoke
Droplet. Langmuir Vol. I, No. 3, 1985 .
3. Lorenz-Mie Scattering
4. Particle size determination: An undergraduate lab in Mie scattering
Weiner, M. Rust, and T. D. Donnell Harvey Mudd College, Department of Physics,
Claremont, California 91711
5. Bohren_C.F.,_Huffman D.R. Absorption and Scattering of Light by
Small_Particles(1983)