Page 1: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Part 5: Introduction to Spectroscopy for Chemical Analysis

KR

LSU

Page 2: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI-REVIEW: What we know? What we want to know?

Chemical analysis and spectrophotometry: using radiation objects (light) to measure molecular concentrations.

Basic components of spectroscopic instrumentation:

1)Light sources,

2)Wavelength selection devices,

3)Detectors,

4)Basic methods of measurements, absorption and emission etc.

5).. and we also want to know how to present results of measurements: mean values and uncertainty in Spectrophotometry

Page 3: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

What are radiation objects/photons

Page 4: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Simple picture: radiation objects move from left to right, the electromagnetic field oscillates up-down

Radiation Objects: origin from interaction with material objects

(electrons, protons, atoms, molecules….)

Page 5: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

How do we know that light can be described through wave-like observables: From experiments

Page 6: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Wave-like observables

Page 7: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI:

Radiation objects, also known as photons, can be described using physical properties or observables such as wavelength, frequency, energy, momentum, polarization etc.

Physical observables are obtained from physical measurements of physical objects.

Physical observables can be related to each another, depending on the way experiments measure those objects: wavelength is related to frequency, frequency to energy etc. The so called wave-like observables can be related to the particle-like observables.

Page 8: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

IMPORTANT: Light as well as electrons can be described through what we sometimes call wave-like physical observables

• Waves are described by wavelength, , amplitude, A and frequency, . The frequency, , of a wave is the number of cycles which pass a point in one second.

• The speed of a wave, c, is given by its frequency multiplied by its wavelength:

Speed ( c ) = • C = 2.9979 * 108 m/s (in vacuum)

Page 9: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

IMPORTANT: Relationship between physical observables: frequency and wavelength:

C= *

yellow light : =589nmfrequency = c/ =

=(2.99799 *10 8 m/s / 589nm) (1nm/10-9 m)== 5.09 *1014s-1

Page 10: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Electrons in Atoms: Spectroscopic FYI: Electrons in Atoms: Spectroscopic Investigations and Line Spectra (Hydrogen Atom)Investigations and Line Spectra (Hydrogen Atom)

Line SpectraLine SpectraRadiation composed of only one wavelength is called monochromatic.

Radiation that spans a whole array of different wavelengths is called continuous.

White light can be separated into a continuous spectrum of colors.

Page 11: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

electromagnetic radiation objects: different physical properties ------electromagnetic radiation objects: different physical properties ------IMPORTANT IMPORTANT

Radiation composed of only one wavelength = monochromatic (one color).

Radiation that spans a whole array of different wavelengths = continuous.

White light = continuous spectrum of colors.

Dark= no light.

Page 12: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

RO originate from interactions with material objects (electrons, protons, atoms, molecules ..)

When light interacts with material it can be transmitted, reflected, scattered, absorbedemitted. Depending on the material, the interaction with different radiation will be different.

Page 13: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU
Page 14: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

How do we detect photons/radiation objects

Page 15: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Photons are coming from individual events. Energies of photons emitted from (or interacting with)

electrons in atoms are quantized :

**This is similar to some other (wave-like) systems: musical instruments (piano) etc.

Page 16: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Each interaction event between radiation objects and material

objects can be described with particle-like observables:

Radiation objects (photon) detected in individual events. (DEMOS)

Page 17: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

IMPORTANT: Relationship: energy and frequencyEnergy of yellow photons of 589nm:

E=h * = h * (c/ ) = (6.63 810-34 Js) *(5.09 1014s-1)=3.37*10-19 JFor: 1 2 3 ……..photons:

1*3.37*101*3.37*10-19-19 J 2*3.37*10 J 2*3.37*10-19-19 J 3 3.37*10 J 3 3.37*10-19-19 J J

1 h1 h 2 h 2 h 3 h 3 h ……. energy

Page 18: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

The general paradigm of spectrophotometry:counting molecules with photons!

How???

Atoms, molecules, ions absorb or emit photons!In a single photon regime it is 1: 1, one photon for one molecule of compound, or 1 mole of absorbed

photons per one mole of affected compound molecules! (“Titration with photons”

Page 19: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

How does absorption and emission of light work

Page 20: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Rydberg’s FYI: Rydberg’s analysis of atomic line analysis of atomic line spectra (Na,H):spectra (Na,H):

2211

fi

H

nnhR

hE

*observables using h constantMovie colors of alkali metals

RH= 2.18 10-18 J

Page 21: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU
Page 22: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

IMPORTANT: Main principle of spectrometry for chemical analysis

• If ONE radiation objects is observed from the interaction with ONE material object then we are in the position to count material objects using radiation objects at the atomic, electronic or molecular accuracy.

Page 23: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Absorption and Absorbance

Instrumental configurationand definitions

Page 24: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

IMPORTANT: BASIC COMPONENTS OF SPECTROSCOPIC INSTRUMENTS (and absorption as an example)

-sources of radiation objects (continuum and discrete lamps, lasers, ..)

-selection of radiation objects (filters, monochromators, interferometers, tunable lasers..)

-detectors (different wavelength range photon transducers-phototube, PMT, photodiode, photoconductor, photovoltaic cell, thermocouple, thermistor, pyroelectric, etc. image detectors

-signal processor (electronics, computers..)

Page 25: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Radiant power, P, ~ energy per second per unit area of a light beam 

When light is absorbed by a sample, the radiant power of the light beam decreases.  Transmittance, T, ~ fraction of the original light that passes through a sample. 

T = P/Po and T has the range of 0.0 – 1.0  Absorbance, A, is related to transmittance. 

A = log10(Po/P) = –log10T  

IMPORTANT

Page 26: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Cont.

• Absorbance is more useful quantity because it is directly proportional to the concentration, C, of the light absorbing species in the sample.

•• A k C•  • An absorption spectrum is a graph showing how

A varies with l.•  

Page 27: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Absorbance dependence on the number of molecules/atoms (Beer’s law)

Exponential decrease natural logarithm dependence

FYI

-dP/P = e c dx

- Integral ( dP/P )(from Po to P) = e * c * Int (dx) (from 0 to b) -lnP –(-ln Po) = e c b or ln(P0/P) = e c b

What if small concentration or short legth?????

Page 28: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Beer's Law ( Beer-Lambert Law) 

A = b Cwhere c is the concentration (M), b is the path length (cm), and is the molar absorptivity (M–1 cm–1). is wavelength dependent. 

Beer's Law works for dilute (~ 0.01 M) solutions in which the absorbing species is not participating in a concentration-dependent equilibrium.

Page 29: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Practical issues

Page 30: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Broad absorption peak measurements:

Page 31: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

For Quantitative Analysis

• Absorbance should not be too large or small(~0.4-0.9 units)• Absorbance should be measured at dA/d 0; at the lambda max providing no interference from other species at this lambda.• Monochromator bandwidth should be as large as possible but small compared to the spectral bandwidth. Slit width to small allows less light to reach the analyte, smaller S/N ratio and precision. Too wide a monochromator band width (slit width) distorts the peak shapes of the spectrum. A As shown in the figure above; the instrumental setting with the band width at 0.1 nm and the choice of 609 nm for max would be desirable for

the spectra shown. At the 2.0 nm setting there is no fine structure shown in this spectrum. 

Page 32: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

EXPERIMENTAL PARAMETERS : how to get the most from measurements with absorption spectroscopy

C Choice of the Wavelength and Bandwidth – 

Page 33: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Conditions

• What characteristics we need for absorption spectrophotometers?

• ·  Most spectrophotometers operate best at values for A 0.4 – 0.9

• ·  If absorbance too high, intensity is hard to measure.

• ·  If absorbance too low, hard to distinguish between reference and sample

Page 34: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU
Page 35: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI (cont.)• Slight mismatches between reference and sample cuvettes

leads to systematic error

• ·  Samples must be dust free. Cuvettes must be wiped clean before use.

Page 36: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Example

At a given wavelength, a cuvet filled with a sample solution has a transmittance of 63.1%. A reference cuvet filled with the solvent has a transmittance of 94.7% at that same wavelength. What is the corresponding absorbance of the sample?

 

A = –log10 T so Asample = –log(0.631) = 0.20

Aref = –log(0.947) = 0.02

 

Acorrected = Asample – Aref = 0.20 – 0.02 = 0.18

 

Page 37: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

A  Example

A 0.267 g quantity of a compound with a molecular weight of 337.69 g/mol was dissolved in 100 mL of ethanol. Then 2.000 mL was withdrawn and diluted to 100 mL. The spectrum of this solution exhibited a maximum absorbance of 0.728 at 438 nm in a 2.000-cm cell. What is e? 

 

Procedure for Constructing a Calibration Curve 

Step #1. Prepare known samples of analyte, covering a convenient range of concentration, and measure the response of the analytical procedure to these standards. 

Step #2. Subtract the average response of the three blank samples from each measured responses to obtain the corrected response. The blank measures the response of the procedure when no analyte is present. 

Step #3. Make of graph of corrected response vs. quantity of analyte analyzed.

Page 38: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Example: the following information is obtained during absorbance measurements at 427 nm.

standards corrected blank Absorbance conc(M) Abs readings readings readings

0.1 0.89 0.008 0.898 0.05 0.47 0.01 0.48 0.01 0.079 0.0089 0.0879

0.005 0.04 0.009 0.049 0 0

y = 9.0088x - 0.0015

-0.2

0

0.2

0.4

0.6

0.8

1

0 0.05 0.1 0.15

Ab

s

conc

Beer's calibration plot

Page 39: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI : Some additional information not included in this course

Page 40: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

Relations between observablesare similar for both material and radiation objects:Photoelectron experiments

Compton: electron –photon scattering

Laue, Bragg: X-rays: photon diffraction

Electron diffraction

Black body radiation, etc.

E = h

where h is Planck’s constant (6.626 10-34 J.s).

mvh

The momentum, mv, is a particle-like observable, where as is a wave-like observable.

In this equation de Broglie presented the relation between the observables of waves (wavelength) and particles as they apply to observables mass and speed (velocity ) of objects.

Page 41: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Relationship: wavelength and velocity of electrons

electron velocity v=5.97* 106m/s electron mass me = 9.11* 10 –28 g

= h/(m *v)=(6.63 810-34 Js) /(5.97* 106m/s *9.11* 10 –28 g)( (1kg m2/s2)/1J) * (103g/1kg)=

1.22 10-10 m =0.122nm

Page 42: Part 5: Introduction to Spectroscopy for Chemical Analysis KR LSU

FYI: Relationships between observables and interactions

Energy (E)and time (t), momentum (m*v) and position(x) have also relationships:

We can write them as: dE * dT > hdp *dx > h

*uncertainty principles/relations