Páginas desdeAPI MPMS 4.8_PROVER RECALIBRATION FREQUENCY.pdf

Embed Size (px)

Citation preview

  • 8/13/2019 Pginas desdeAPI MPMS 4.8_PROVER RECALIBRATION FREQUENCY.pdf

    1/2

    6 CHAPTERPROVING SY STEMS

    objects traveling in the flow stream. They should beinspected and cleaned periodically.Turbine meters are especially susceptible to the effectsfrom deposits because they re velocity devices. Layering orcoating of the meters internais will change the velocity ofthe liquid flowing through the meter and cause the meter toregister incorrectly.Temperature changes can affect the mechanical clearancesof displacement meters, as well as the viscosity of the uidbeing metered. This may result in changes n slippage.4.8.2.8.5 Electronic Equipment andInstrumentation

    All electrical and electronic equipment, such as counters,switches, interconnecting cables, and grounding cables, shallbe periodically inspected for condition and for proper instal-lation and operation. Operating procedures may requirespecial permission or permitsbefore equipment is connected.A counter may miss some of the puises generated by themeter, in which case the counter will read low. Counting toofew puises is usually caused by setting the sensitivity controldeveloped. By adjusting the sensitivity control or by elimi-nating the electrical fault, the trouble can usually becorrected.A counter may include signals from outside sources aspulses. These signals, not generated by the meter, will causethe counter to read high. Signals not generated by the metercan originate from electrical power supplying the counter,electrical welding equipment, radio transmitters, and soforth. These puises may be intermittent and difficult todetect. See API MPMS, Chapter 5 Section5.Signal transmission cables should be kept as far awayfrom power cables s possible and should cross power cablesat right angles. Shielded signal transmission cable isnormally grounded only at the instrument-receiving end toprevent a ground loop current that travels along the shieldand adversely affects the signal transmission).

    on-the counter too low, or by an electrical fault which has

    4.8.2.8.6 Flow Rate VariationsMeter performance is dependent upon flow rate; thus,flow rate during proving shall be maintained at or near thenormal operating flow rate.4.8.2.9 METER REGISTRATION HEAD) CHECK

    Compare the meter register indicated volume) to theproving-counter registration. This can be done by manuallygating starting and stopping) a prover counter connected tothe transmitter, based on a significant volume registered bythe meter counter or register. The pulses displayed on theprover counter are then compared to the volume displayedon the mechanical register. If the meter generates 8400

    puises per barrel, the prover counter should show approxi-mately 84 000pulses for each 10barreis on the register.4.8.2.10 FREQUENCY OF METER PROVING

    The frequency required for proving varies from severaltimes a day to twice a year or even longer depending uponthe value of the liquid, costhenefit to prove, meter provinghistory, meter system stability, and variations of operatingsystems.For large volumes or different liquids, a permanentlyinstalled prover is normally used. The meters should beproved whenever the flow rate, temperature, pressure, APIGravity relative density), or viscosity changes significantly.Normally, time or volume is used to determine when themeter should be proved.When metering a single or similar liquid, the meter factoris normally applied forward to the meters indicated volumeuntil the meter is reproved. Normally, there is a prescribeddeviation limit between consecutive meter factors on thesame or similar liquid. When this deviation limit isexceeded, the previous and the new meter factors arenormally averaged and applied to the indicated volumeduring this period. If the deviation limit is consistentlyexceeded, it may be appropriate to reduce the intervalbetween meter proving. It may also be appropriate to inspectand repair the meter and the proving system.When batching operations permit, a new meter factorshould be determined for each batch. This applies tobatching operations involving different liquids or lengthydown time. When a meter is proved during a batch, the meterfactor should be applied forward until the meter is reprovedduring the batch. If the meter is reproved during the batch, adeviation limit may be installed between consecutive meterfactors, or the meter factors may be averaged. When thisdeviation limit is exceeded, the previous and new meterfactors are normally averaged and applied to the metersindicated volume between these provings. If i t is impracticalto prove each batch, meter factors are normally appliedforward until the next proving, as is the case withnonbatching operations.

    The proving frequency for new systems should start atshort intervals and be extended to longer intervals as confi-dence increases in the system. See API MPMS Chapter 13.2for statistical evaluation of meter proving data.4.8.2.11 PROVER RECALIBRATIONFREQUENCY

    Typically a provers base volume is originally certified atthe manufacturer site by the water draw method in the pres-ence of the purchaser and other interested parties. Provervolumes may change as the result of worn or faulty detectorswitches; the reduction of internal coating thickness; or lossof internal material due to oxidization, abrasion, or the accu-mulation of foreign material such as wax) buildup. Subse-

    yright American Petroleum Instituteoduced by IHS under license with API Licensee=Perez Companc/5955403001

    Not for Resale, 05/09/2005 07:08:49 MDTeproduction or networking permitted without license from IHS

    --`,,,,,`,``,`,```,```````,`,`-`-`,,`,,`,`,,`---

  • 8/13/2019 Pginas desdeAPI MPMS 4.8_PROVER RECALIBRATION FREQUENCY.pdf

    2/2

    Sect ion Operat ion o Proving Systems

    quent calibration is required whenever a change in basevolume could have occurred.Six considerations determine how frequently a provershould be recalibrated. They are usage, time, calibration

    history, calibration costhenefit, contractual requirements,and value of the metered liquids. Usage causes wear, andtime contributes to deteriorationof the prover.For the recommended procedure for calibrating a prover,refer to API MPMS Chapters 4 and 12.2. Recalibration ofprovers should occur when any one of the following condi-tions exist:a. Alterations or repairs which affect the certified volume aremade to the prover.b. A meter control chart indicates a change in provervolume.c. The maximum interval indicated below has elapsed.1. Three years for small volume provers and mobileprovers.

    2. Five years for permanently installed pipe provers.3. Five years for permanentIy installed tank provers.4.Three months for master meter provers.The prover displacer and the inside surface of the prover

    should be inspected periodically. The surface of a sphere orthe contact edge of a piston cup or seal may indicate theinternal condition of the prover. If these surfaces or edgesare scored or worn, this may indicate that the proverrequires further inspection or repair and may require recal-ibration.4.8.2.12 FILLING AND PRESSURING THEPROVER

    This section refers to conventional pipe provers, smallvolume provers, and master meter provers.This section doesnot pertain to tank provers, which are covered in 4.8.5.4.After checking that end closures and any openablefittings are properly fastened and that all vent and drainvalves are closed, proceed with filling the prover in thefollowing sequence:a. Partially open the prover isolation valve to fill the proverb. Observe the system for leaks. Wait until the system iscompletely filled and the connections have been shown to beleak-tight. Verify the seal integrity of all vents, drains,reliefs, and all double block-and-bleed valves.c. Open the vents to allow discharge of air/gas when the fluidis admitted into the prover.d. Fully open the prover inlet and outlet valves.e. Close the valve to divert all flow through the prover.f. Operate the prover and continue to vent the high pointsuntil no air is observed.g. Close the vents when air or vapor is no longer observed.

    slowly.

    4.8.2.13 CERTIFICATIONVerify that the prover has a valid calibration certificateand that the certificate is for the prover being used, by veri-fying the prover serial number with the serial number on the

    certificate. If a conventional pipe prover is being used, checkto ensure that the prover volume between detectors is sufficient to accumulate a minimum of 10,000 pulses. If not,pulse interpolation techniques are required. Since someprovers have more than one calibrated volume, verify thatthe proper calibration certificate is being used.If a tank prover is used, verify that the prover volume isequal to a minimum of one minute of the maximum oper-ating flow rate. See API MPMS, Chapter 4.4.If a master meter is used, all data that is used to developthe master meter factor s), including the prover calibrationreport, certificate, and master meter factor s) reports shouldbe available.If a small volume prover is used, verify that the interpola-tion system has a valid and current calibration certification.Refer to API MPMS, Chapter4 Sections3 and 6.

    4 8 3 Conventional Pipe Provers4.8.3.1 PRINCIPLEOF OPERATION

    The basic principle on which the pipe prover operates isshown in Figure 3. A sphere or piston known as a displ ceris installed inside a specially prepared length of pipe. Whenthe prover is connected in series with a meter, the displacermoves through the pipe and forms a sliding seal against theinner wall of the pipe so that it always travels at the samespeed as the liquid flowing through the pipe.

    In some conventional provers, the displacer is a piston withelastomer or plastic seals. However, in most conventionalprovers, the displacer is an elastomer sphere. To provide goodsealing, the pipe bore must be smooth.At two or more points in Figure 3 there are devices knownas detectors fixed to the pipe wall. These detectors emit anelectric signal when the displacer reaches them. The signalfrom the first detector switch is used to start the electroniccounter, which accumulates pulses from the meter. When thedisplacer reaches the second detector, its signal stops theproving counter. The number of pulses shown on the provingcounter is the total pulses generated by the meter while thedisplacer was travelling between the two detectors. Conven-tional pipe provers both bidirectional and unidirectional)arethose that have a volume between detectors that permits aminimum acculuation of 10 OOo direct unaltered)pulses fromthe meter. Thus a unidirectional prover typically accumulatesa minimum of 10,ooO pulses per proving run, and a bidirec-tional, prover typically accumulates a minimum of 20,000pulses per proving run. Direct unaltered) puises include thosethat are the output of high frequency pulse generators, consid-ered to be a part of the meter. It should also be noted that

    yright American Petroleum Instituteoduced by IHS under license with API Licensee=Perez Companc/5955403001

    Not for Resale, 05/09/2005 07:08:49 MDTeproduction or networking permitted without license from IHS

    --`,,,,,`,``,`,```,```````,`,`-`-`,,`,,`,`,,`---